key: cord-0790333-3qa9qtwj authors: Balkourani, G.; Brouzgou, A.; Archonti, M.; Papandrianos, N.; Song, S.; Tsiakaras, P. title: Emerging materials for the electrochemical detection of COVID-19 date: 2021-04-23 journal: J Electroanal Chem (Lausanne) DOI: 10.1016/j.jelechem.2021.115289 sha: d1ad55637f99165dc2aaca65b593d6aeda85f058 doc_id: 790333 cord_uid: 3qa9qtwj The SARS-CoV-2 virus is still causing a dramatic loss of human lives worldwide, constituting an unprecedented challenge for the society, public health and economy, to overcome. The up-to-date diagnostic tests, PCR, antibody ELISA and Rapid Antigen, require special equipment, hours of analysis and special staff. For this reason, many research groups have focused recently on the design and development of electrochemical biosensors for the SARS-CoV-2 detection, indicating that they can play a significant role in controlling COVID disease. In this review we thoroughly discuss the transducer electrode nanomaterials investigated in order to improve the sensitivity, specificity and response time of the as-developed SARS-CoV-2 electrochemical biosensors. Particularly, the Au-based and carbon or graphene-based electrodes are the two main material groups that were explored. Additionally, the adopted electrochemical detection techniques are also discussed, highlighting their pros and cos. The nanomaterial-based electrochemical biosensors could enable a fast, accurate and without special cost, virus detection. However, it is necessary further research to be done in terms of various nanomaterials and novel synthesis strategies in order the SARS-CoV-2 biosensors to be commercialized. 1 electrode materials investigated for SARS-CoV-2 detection. In addition, this review also dealswith 2 the electrochemical techniques being used for the SARS-CoV-2 electrochemical biosensors 3 evaluation, which as we noticed constitute a controversial subject in the scientific community. 4 2. Operational principles of biosensors 5 Generally, a typical sensor is a device used for measuring a physical quantity (input), which is 6 converted into a signal (output, electrical or optical) that is then interpreted by the specialists. Accordingly, 7 a biosensor consists of a bioreceptor, which is a biomolecule, such as enzyme, antibody, or other and its role 8 is to capture the target analyte (e.g. a virus) and a transducer, the role of which is to transform the signal 9 produced from the biomolecule in electricity (current or potential) or other kind of signal depending on the 10 type of transducer (Fig. 1Α ). The recorded signal usually is amplified by a signal processing circuit and then 11 is interpreted with the results to be displayed in a computer or a smartphone 26 . 12 13 Figure 1 . (Α) Operational principle of a biosensor;(B) Label-free and label-based detection process of 14 biosensors;(C) Label-free aptamer detection process;(D) Antibody-aptamer sandwich type (a&b), aptamers 15 pair sandwich type (c&d). Reproduced with permission. 16 Biological analytes, due to their intrinsic physical properties, are hard to be detected directly, so 7 1 drawback of the labelling procedure is that it is time-consuming, as well as labour-and cost-intensive. 2 Furthermore, sometimes the labelling of biomolecules may block active binding sites, leading to the 3 modification of binding properties attributed to the unknown degree of affinity between the target analyte 4 and the bioreceptor 27 and consequently reduces biosensors sensitivity, selectivity and reproducibility. 5 On the contrary, label-free biosensors detect the analytes, utilizing intrinsic physical 6 properties (electrical impedance, dielectric permittivity, or refractive index, etc.) of the analytes in 7 their natural form (Fig.1B) . For example, when electrical change is studied, the conduction or 8 resistance or capacity is measured; in the case of mechanical changes mass and frequency changes 9 are measured; in the case of optical elements the adsorption and emission of light are measured. 10 Label-free biosensing methods are rapid and of low cost, using small reaction volumes. Moreover, 11 they can be integrated into lab-on-chip platforms, detecting in real time the concentration of the target 12 analyte and reducing the adsorption of non-specific molecules 27 . Especially the new one aptamers 13 (oligonucleotides)-based strategy in label free detection 28 , offers the opportunity to a biosensor of 14 analysing various target analytes, even those with small molecules, which if being analysed with the 15 common label-free sensing method would require more time and special attention. The aptamers that are 16 in vitro synthesized, usually form non-covalent bonds with the target analyte, increasing biosensor's 17 specificity and sensitivity towards a precise target. Generally, aptamers (usually a single-stranded 18 DNA/RNA oligonucleotide sequence) are immobilised onto transducer solid surface (electrode 19 transducer) and usually bind to the virus protein 29 (Fig.1C ). When the aptamer binds to the target-analyte 20 acts as a barrier that increases the sensing surface resistance, preventing the electrons transfer to the 21 electrode. The amount of the target sample is proportional to the charge resistance 30 . 22 The main advantages of the aptamers are as follows: they can be chemically modified for 23 increasing their stability and affinity towards the analyte target and can be synthesized in large amounts, 24 when their nucleic sequence becomes known 31 . However, even though aptamers are considered as the 25 next generation bioreceptors, the up-to-date commercialized ones do not use a single aptamer, since it 8 1 alternative, either the pair antibody-aptamer sandwich type [ (Fig.1D 3 From the four SARS-CoV-2 structural properties(spike, envelope, matrix and nucleocapsid) the 4 spike protein (S1 and S2) is usually chosen as the domain receptor, since it constitutes a trans-membrane 5 protein of the virus-cell with high immunogenicity, having the ability to induce immune responses and 6 facilitating SARS detection 33 . Moreover, the spike protein has affinity with the angiotensin-converting 7 enzyme 2 (hACE2), which is used as a receptor infecting human cells 34, 35 . Currently the aptamers, 8 having affinity with the SARS-Co-V-2 spike protein, were identified through the known Systematic 9 Evolution of Ligands by Exponential enrichment system or "SELEX" system 36 . According to this 10 process, a variety of polynucleotides (aptamer pool) are exposed to the target molecule and the aptamers 11 with higher affinity are in vitro selected and then eluted, while the rest is disregarded ( Fig. 2A ). This process is repeated many times until a small number of aptamers with high affinity and 2 specificity will remain to the desired target. 3 Today, one of the SARS-CoV-2 diagnostic tools is based on an enzyme-linked immunosorbent 4 assay (ELISA) biosensor. The operational key principle is based onto the target-analyte interaction 5 with the enzyme-antibody complex, followed by an enzymatic reaction, which is responsible for the 6 signal production. In Figs 2Bi and 2Bii are depicted the direct and indirect detection processes, 7 respectively. At the direct detection, the target-analyte is adsorbed directly on the microplates' surface 8 (Figs 2Bi) , with the risk that undesired proteins (albumin) will also be adsorbed. On the other hand, at 9 the indirect detection process, the indirect adsorption (sandwich structure: antibody-analyte-antibody, 10 seen in Fig.2Bii ), exhibits higher specificity as the adsorption is more selective. In both cases the 11 analyte amount is specified via the colour change from the enzyme reaction 37 . An electrochemical transducer is characterized of the lower cost, simpler construction, 14 higher specificity, lower sensitivity and portability. Their operational principle is based onrecording 15 of the electrical signal, produced when the bioreceptor selectively reacts with the target-analyte 16 (labelled or not labelled),which is proportional to the analyte's concentration 38 . According to the identificationprocess of the target-analyte, the electrochemical biosensors 18 are classified into: i) biocatalytic and ii) affinity sensors. A biocatalytic electrochemical biosensor 19 usually include enzymes, the role of which is to identify the target-analyte producing electroactive 20 species. On the other hand, the operation of an affinity electrochemical biosensor is based on the 21 selective interaction between a bioreceptor (antibody, enzyme, nucleic acid or other receptor) and the 22 consequent production of an electrical signal (Fig.1A) . 23 For the virus diagnosis, the affinity electrochemical biosensors are chosen. Usually the 24 binding process includes antibody-antigen interaction (Fig.3A ) or DNA hybridization (Fig.3B ). 25 Immunosensors operation is based on the interaction between the antibody, immobilized onto the 26 sensing surface of the electrode, and the antigen of a specific analyte. When the binding process 1 takes place, an electrical current is produced accordingly to the antigen's concentration of target 2 analyte, which then it is amplified and displayed 39 . The challenges of the immunosensors, however, 3 led also to the development of bio recognition procedure, using DNA or RNA fragments 40 . At the hybridization procedure the single-stranded DNA segments, placed onto the surface of 8 the sensing electrode (DNA probes), bind selectively with the complementary part of the target-9 analyte's DNA (pair-base procedure), producing an electrical current/signal. Afterwards, the 10 confirmation of a successful hybridization process occurs: i) using electroactive complexes (redox 11 indicators), such as ferrocenyl naphthalene diimide that present different affinity towards a single 12 stranded DNA, than to a double strand (successful hybridization) 41 or ii) using enzymatic labels, such 13 as horseradish peroxidase, and from the occurring redox reactions the produced current is measured. 14 Lately, nanoparticles are investigated as electrochemical labels, because compared to 15 enzymes own higher stability and conductivity, offering faster electron transfer from the immobilized 16 target to the electrode. At the DNA hybridization procedure, in absence of labels, the guanine base is 17 oxidized on the electrode's surface, while the produced current can be amplified by the aid of redox 18 mediators [e.g. Ru(bpy) 3 +2 ] 42 . At this method, the success of the hybridisation procedure is checked 19 by measuring the changes of transducer's capacitance or conductivity 43 . The shape as well as the size of the binding site defines the degree of affinity, sensitivity and 21 selectivity and it is critical for each kind of biosensor. The main challenge of immunosensors is the 22 antibody (Ab) production for a specific virus antigen (Ag) to a host organism, which is a time 1 consuming and costly process. The immobilization of the antibody on the solid surface of the 2 electrode is another challenge to overcome. In case the antibodies are immobilised with wrong 3 orientation they may lose their binding ability being unable to provide a highly sensitive biosensor. 4 Moreover, in the case of DNA biosensors the orientation and the amount of the DNA probe are very 5 important, confirming the significant role the electrode material plays. For this reason, the DNA 6 molecules should be immobilized in a specific orientation to maintain stability and activity, and the 7 other characteristics as well 44 . Moreover, in order to avoid the binding of undesired species, the 8 surface of the electrode should be covered with blocking agents. When the aptamer production is adopted, a series of DNA or RNA oligonucleotides 10 (aptamers) are produced, which detect and bind selectively to a three-dimensional surface, such as 11 the one of proteins (Fig.1C ). The sequences of the aptamers are produced via a polymerase chain 12 reaction (PCR). 13 Therefore, for achieving the optimal characteristics for biosensors operation, the choice of 14 electrode material along with its treatment procedure is of great importance. Notably, the 15 immobilization procedure has been improved by far due to the development of nanomaterials 45 At the amperometric detection method, a constant potential (vs a reference electrode) is 5 applied at the working electrode due to which an electrochemical oxidation or reaction happens that 6 produces current. The produced current that is measured (vs a counter electrode) is proportional to 7 the analyte target's concentration into the sample 48 . Especially, when the measurement is conducted 8 with the rotating disk electrode technique 49 , in which the mass transport of the analyte onto the 9 working electrode surface is enhanced. This method is mainly adopted for both the biocatalytic and affinity biosensors as they own 11 the advantage of very low detection limit and very high selectivity towards analyte target. Moreover, 12 the applying of an exact potential that causes the electrochemical oxidation or reduction reaction, is a 13 time-consuming detection procedure 50 . 14 On the other hand, according to the voltametric detection procedure, the working electrode is 15 imposed to work under a potential range (vs a reference electrode) into which an electrochemical 16 oxidation or reduction reaction occurs, and the produced current is measured (vs a counter electrode). The 17 current response is proportionally to the analyte concentration applying mainly the differential pulse 18 voltammetry, cyclic voltammetry, square-wave voltammetry and linear sweep voltammetry 50 . Nowadays, 19 aptamer based biosensors 51 which are considered the next generation biosensors, tend to adopt the 20 voltametric detection techniques, exploiting the robustness and the very low detection limits that they 21 present even with a small amount of analyte volume. According to the electrochemical impedance technique, a small AC excitation signal (2-10) 24 mV under a specific frequency (Hz) range, is emitted to the electrode under investigation, and by 1 conductivity are estimated. 52 . At high frequency values the electron transfer can be estimated, while 2 the mass transfer is estimated at low frequency values. 3 The EIS is primarily used for the evaluation of the affinity label-free electrochemical 4 immunosensors for the directly control of the antibody-antigen binding process that happens onto the 5 electrode's surface. 6 Indicatively, before the measurement the working electrode's surface is covered (via the aid 7 of electrodeposition method) with a polymeric film with high conductivity onto which the 8 bioreceptors are incorporated too. The detection process takes place by applying a known potential 9 value to the working electrode which causes current production and consequently electron transfer. 10 The electron transfer resistance at the electrode and the bulk of analyte interface, change as the target 11 analyte binds onto the bioreceptor. This change of the resistance that is measured with EIS is 12 proportional to the antigen concentration. In this way the formation of the antibody-antigen bonding 13 layer can be monitored appropriately and so there is no need labels to be used 53 . 14 The low cost, the quicker response of assays, the higher signal to noise ratio, are some of the 15 main advantages of the electrochemical impedance detection. Nevertheless, the main disadvantage of 16 this type of detection is the low likelihood of regenerating the working electrode 54 . 18 According to the potentiometric method the sensor is consisted of two reference electrodes 19 that are fixed onto a layer whose potential change is measured when it reacts selectively with the 20 charged ion of the analyte target. The charged ion that is produced by the aid of an enzyme, which 21 catalyzes a specific reaction causes the potential difference that the membrane presents before and 22 after the charge ion is produced. 23 The field effect transistors-based biosensors 55 belong to this category; consisted of a 24 biological recognition compartment and a field-effect transistor (semi-conducting material). 25 Between the analyte solution and the semiconductor there is an insulating (chemically and 26 electrically) layer that is usually a polymer onto which the bioreceptors are placed. When the binding 1 procedure (bioreceptor-target analyte) becomes successfully, the surface of the electrolyte-insulator 2 layer electrostatic potential changes is recorded, which accordingly cause an electrostatic gating 3 effect to the semiconductor as well as a current difference between the source and the drain 4 electrodes that is also recorded. 5 2.3 Efficiency characteristics of electrochemical biosensors 6 The efficiency of an electrochemical biosensor is determined by its: i) accuracy, ii) response 7 time, iii) selectivity, iv) repeatability, v) reproducibility, vi) detection limit, vii) sensitivity, viii) 8 recovery time and ix) linear range 38 , characteristics that are explained in the Table 1. 9 Table 1 . Determination factors of the efficiency 38 . The amount of uncertainty in a measurement, compared to a reference. The accuracy is important in order to evaluate the correct quantity (usually molarity) of the target-analyte. The time interval between the moment the analyte gets into contact with the bioreceptor and the moment the electric signal is received. A quick response time is preferable if the sensor is used as a device of point-of-care. The ability of a sensor to preferably analyze the desired analyte in presence of other substances. The high selectivity ensures higher lifetime and accuracy of a sensor. It ensures that an electrochemical sensor will give the same output for over a period. It represents a factor along with the reproducibility that determines a sensor's reliability. The ability of an electrochemical sensor to give the same signals under varied measurement conditions. The lowest detection limit is the lowest concentration of the target-analyte that can be tracked with accuracy. The lowest difference identified by a sensor, after the change of analyte's concentration. A sensor with very high sensitivity can even detect very small changes of analyte concentration. The minimum time between two successive measurements. Before the second measurement starts, the sensor should return into its initial condition. The electric signal received from the sensor should be linearly dependenton the analyte concentration. So, the minimum and maximum concentration that could be analysed by a sensor, are determined. Emerging transducer electrode materials for viruses' electrochemical detection 2 The development of nanotechnology has helped the growth of biotechnology, emerging new 3 possibilities that have greatly improved the efficiency characteristics of biosensors 56 . The adoption 4 of nanomaterials for the transducer electrode electrochemical biosensors structure enhanced their 5 sensitivity, selectivity, response time, making them even more approachable for virus detection. 6 Specifically, as many researchers 57 support, the transducer electrode that combines an aptamer 7 or an antibody with nanomaterial (hybrid material) transmit higher signal, especially when the 8 nanomaterial is used as label, presenting also increased sensitivity and affinity towards the target 9 analyte 57 . Among the investigated nanomaterials, gold nanoparticles 58, 59 are mostly investigated as 10 they are biocompatible, with excellent electrical and optical properties. Moreover, gold nanoparticles 11 improve the immobilization of the bioreceptors onto the electrode transducer, providing a larger 12 surface area, which allow the immobilization of higher number of bioreceptors 57 . Therefore, today the 13 scientists pay more attention to further modify the gold nanoparticles either by exploring different 14 supports, such as multiwall carbon nanotubes, graphene, etc, or adopting other preparation strategies. In the next sub-section, we discuss indicatively, some interesting methods that were followed 25 small reaction volumes, allowing the simultaneous hybridization and sequential detection offering 1 more than twenty assay sites. During the detection, the complementary strand of a specific SARS-2 DNA sequence was labeled with thiol group and immobilized onto the Au surface. 3 The electrochemical detection amplified with alkaline phosphatase-labelled streptavidin 4 substance; while for the expulsion of the undesired molecules from electrode surface, albumin and 1-5 hexanethiol are tested as blocking agent, as depicted in Fig.4A . Moreover, the operational parameters of 6 the biosensor were evaluated with the aid of the following electrochemical methods: i) cyclic 7 voltammetry, ii) differential pulse voltammetry, iii) alternating current and iv) square wave voltammetry. 8 It is remarkable that the lowest detection limit of the biosensor was found to be 6 pM (pM =10 -12 M) 60 . 27 Therefore, the Au-nanoparticles at the same time managed to retain their electrocatalytic activity, 1 immobilize the analyte molecules and to amplify the DNA detection ability. This threefold function 2 was also responsible for the less analysis time and the stability of the biosensor. 3 The evaluation of the as-fabricated biosensor was achieved by using cyclic voltammetry (Fig. 4 4Ci) and by estimating the produced current from the electrochemical oxidation of the deposited metallic 5 Ag, which acted as electroactive species. As seen in Fig. 4Cii , before the detection procedure, the silver 6 ions contained in the solution are reduced to metallic silver (Ag) and deposited next to the enzymatic 7 label AP (alkaline phosphatase), using a reducing agent produced by the 3-indoxyl phosphate substrate. 8 The amount of the deposited Ag was measured by the aid of anodic stripping voltammetry. 1 labelled DNA, aiming at the SARS-CoV detection. The use of the CNTs enables the methylene blue 2 utilization as an electrochemical label that is covalently attached to a single-stranded DNA, thus 3 facilitating the higher electron transfer, as seen in Fig. 4Di , compared to methylene blue molecule 4 without DNA-attachment (Fig. 4Dii) . 5 More specifically, the chemistry of the CNTs surface allows the adsorption of such molecules, 6 which in their turn amplify the analytical signal. Following this strategy, it has been managed seven 7 times higher sensitivity, fact that led to use differential pulse voltammetry and square wave 8 voltammetry, which are very sensitive electrochemical detection techniques and are usually applied 9 for the measurement of very low current values. However, comparable to Au transducer electrodes, 10 the sensitivity (0.2864 μA μM −1 ) was not very high, and the limit of detection remained at the 11 nanoscale (800 nM), indicating that further research is necessary. 12 Recently, for the simultaneous detection of various CoV viruses were used for the first time The square wave voltammetry technique and ferro/ferricyanide redox-couple (electroactive 25 substance) were employed 63 in order to evaluate both the antigen binding onto the working 26 electrodes (after being modified with serum albumin, BSA that prevents undesired molecules to 27 bound onto the electrode) and the investigation of the novel immunosensors operational parameters. 1 In case that the peak reduction current (derived from the redox reactions) was decreased, means that 2 the antibody was successfully bound to the immobilized antigen (spike protein S1), blocking the 3 access of the ferro/ferricyanide redox couple to the working electrode. Thus, even after many years 4 of research, the Au nanoparticles-based electrochemical biosensors, which are considered better 5 devices as compared to the other types of biosensors, present some limitations, such as the lower 6 limit of detection, which depends on the amount of the antigen (consequently the virus antibody) 7 immobilized on the nanomaterials 57 , and few others. The as-suggested electrode was evaluated measuring the electrical conductivity change with 24 the aid of Ag ion (Ag + ) reduction which played the role of the charge carrier, in the case of mismatch 25 between pairs or virus detection. For the Ag + reduction reaction the cyclic voltammetry technique 1 Ag + , which is reduced 66 , as can be seen in Fig. 4E .The formation of distinct peak current curves 2 clearly defines that the Ag + can sufficiently play the role of the redox element that accepts electrons 3 in free-label DNA electrochemical biosensors. Despite the main advantage of the as-suggested 4 biosensor, the optimum detection time was one hour, indicating that most research should be done, to 5 reduce the response time. 6 The above-mentioned limitations, such as the concentration of antibodies, the need for higher 7 sensitivity, the binding of undesired species and the low reproducibility of an electrochemical 8 biosensor, require a more mannered treatment (or functionalization) of the surface of nano catalyst or 9 the exploitation of other nanomaterials than gold nanoparticles to be further investigated. 19 Table 2 lists the Au-based SARS-CoV-2 electrochemical biosensors, also reporting their 20 response time, the used electrochemical techniques for each biosensor evaluation characteristics and 21 the low limit of detection. As above-mentioned, antibodies-sensors can play a significant role mainly in 22 identifying the prevalence of a disease, which in turn could help the doctors to decide the best treatment 23 and the scientists to understand the kinetics (evolution) of a virus over time and so providing them more 24 information for developing the appropriate vaccines. Chrono-potentiometry SARS-CoV-2 spike S1 protein / SARS-CoV-2 Spike S1 antibody 3 min Artificial target:1 pg/L Au nanoparticles/RGO 74 Electrochemical impedance spectroscopy (EIS) SARS-CoV-2 spike S1 protein and RDB antibodies/ SARS-CoV-2 spike S1 & RBD antigens 11.5sec Artificial targets: 1) 1 pM for spike S1 protein The technique of electrochemical impedance spectroscopy was used for the evaluation of the as-25 fabricated sensor. More specifically, when the analysis sample does not contain any virus molecule, an 1 redox molecules play the role of label, which displays a signal when the DNA hybridization is detected) 2 and iv) an auxiliary probe, hexane-1-thiol (HT) (to ensure the analyte molecule capture). The as-fabricated biosensor 75 was tested using both laboratory and clinical samples, as well. 13 Its detection ability was estimated by employing the technique of differential pulse voltammetry 14 (DPV); once the analysis sample contains the virus (target) a certain current is produced (Fig.6Bii) . 15 As Fig. 6Biii shows the current is linearly proportional to the concentration of the analyte (current vs 16 logC analyte ), ranging from 10 −17 to 10 -12 M (mol/L) exhibiting a lower limit of detection (LOD) close 3 detection. However, in the case of clinical samples analysis, the 200 copies/mL was the lowest 4 detection limit measured, achieving an accuracy of 100% 75 . Moreover, the ability of the biosensor to 5 be connected directly to a smartphone, with no need of nucleic acid amplification and reverse 6 transcription, makes it more accessible as a diagnostic tool. 24 nanoparticles ascertained that the charge transfer kinetics is significantly enhanced, and so the 32 1 N-protein (Fig. 7Aii) . This difference was attributed to the fact that in SARS-CoV-2 cell the 2 concentration of S-protein is three-times lower than that of N-protein 97 . In addition, the LOD is 3 estimated to be 19 ng/mL (ng =10 -9 g/mL) and 8 ng/mL for N and S-protein, respectively (Table 3) . 4 However, the as suggested biosensor required a relatively high analysis time close to 30 min, which 5 should be ameliorated with further research. 6 The modification of carbon (or graphene) structure or the alteration of its surface composition 7 strongly improves the properties of carbon (or graphene), motivating more and more research groups 8 to explore them in many versions 98 . Surface functionalization, which is the most common strategy, 9 can be succeeded either by physical interactions or chemical conjugations 100, 101 . 10 Table 3 . Carbon or graphene-based transducer electrode for the SARS-CoV-2 electrochemical detection. Amperometric technique SARS-CoV-2 Biomolecules (N-protein, S1-IgM, S1IgG, and C-reactive protein) / blood and saliva 10 min 1) 0.1×10 12 pg/L and 0.5×10 9 pg/L for N-protein in artificial and real sample 2) 20×10 12 pg/L and 0.2×10 12 pg/L for S1-IgG in artificial and real sample 3) 20×10 12 pg/L and 0.6×10 12 pg/L for S1-IgM in artificial and real sample 4)10×10 12 pg/L and 0.1×10 12 pg/L for C-reactive protein in artificial and real sample Graphene-sheets coated / field-effect transistor 104 -1) SARS-CoV-2 virus' Genome / SARS-CoV-2 spike protein -1) 2.42 ×10 2 copies/mL (in real samples) 2) 100 pg/L (in artificial targets) 11 The functionality of a nano-material increases the ratio of the electrode surface area to the volume, 21 functionalization, assuring that the graphene structure and consequently its electronic structure will 22 not be disturbed when the required receptor is attaching on its surface. When disruptions appear, the 23 receptor and consequently the detection ability of the biosensor will be partially deactivated 111 . The step-by-step functionalization of graphene was initially evaluated by applying differential 25 pulse voltammetry and electrochemical impedance, under open circuit conditions. It was observed 26 that after each modification step the peak current was decreased, while the resistance increased, 27 proving the successful functionalization. The performance of the biosensor platform 103 was then 1 In the case of carbon or graphene-based electrodes, the surface functionalization constitutes 2 the main strategy that was followed. Especially graphene and its derivatives that are considered the 3 most promising materials, does not contain chemically reactive functional groups that could help 4 immobilizing analyte biomolecules. Thus, we observe its surface or structure alteration by: i) doping 5 graphene with another(bio)element, or ii) creating structure defects, or iii) being used as they are to 6 modify screen printed carbon electrodes. Among the applied detection techniques electrochemical impedance spectroscopy, 8 amperometry and differential pulse voltammetry were the most used ones. Meanwhile, in case of 9 amperometric technique there is a concern about the 'real' current response of the sensor when being 10 in an environment with high virus concentrations, as may diffusion phenomena prevail. 11 Electrochemical impedance spectroscopy, square wave voltammetry and differential pulse 12 voltammetry detection techniques are more sensitive and reliable, especially to the very low 13 concentration values of the target analyte. However, for acquiring the 'real' charge the optimum 14 operational conditions are set each time (Hz or voltage step, or scan rate, etc) according to the virus 15 concentration. Therefore, exploiting over time the nanomaterials potencies by applying emerging 16 approaches, such as functionalization, for their development we express the optimistic view that the 17 above challenge that detection techniques present will be overcome and soon it is possible the first 18 commercial SARS-CoV-2 electrochemical biosensors to be emerged. 19 1 2 Thus, the 2 biosensor developed constitutes a suggestion. However, further research about the response time, lower 3 limit of detection and other efficiency characteristics should be explored for the above sensor electrochemical analysis field, was proposed by Mavrikou et al. 73 in order to electrochemically detect They electrodeposited SARS-CoV-2 Spike S1 antibody into kidney cells SARS-CoV-2 bound onto the immobilized antibody, the potential 9 of the cell-membrane is altered and appropriately measured, as shown in Fig. 5Dii. For a quick reading of 10 the results, the novel biosensor was also connected to a smartphone or a tablet. More precisely, the 11 authors 73 achieved 1 fg/mL (fg/mL=10 -15 g/mL) lower detection limit and a good selectivity for SARS at concentrations higher 14 than 10 fg/mL the sensor response decreases. This means that maybe is necessary more cells to be bound 15 onto the Au surface in order to facilitate more targets or independently target-molecules competing each 16 other for the binding sites. The main advantage of this biosensor 73 was the ability to detect the virus 17 antigen within 3 min, without need any prior sample processing. However, the main limitation of the as-18 suggested electrochemical detection method is the necessity for the cell culture, using special equipment 19 and staff. Moreover, the factors of cost and stability of the cell culture However, when a virus is present in the test sample, by applying a specific potential the targeted 2 antibodies bind with the antigens already bound onto the r-GO nanoflakes, increasing the double layer 3 thickness, thus changing its capacitance, as presented in Fig. 6Aii&Aiii. The binding event is validated by This 8 arises questions like whether such sensors would extract real quantitative results when the target's 9 concentration is higher than the saturation value. In this case, high concentrations might cause 10 competitive adsorption phenomena between the antibodies on the antigens' surfaces and maybe the 11 diffusion phenomena prevail causing false charge measurements 54 . All the above parameters affect 12 the behavior of electrochemical biosensor and require 75 , exploiting the unique properties of r-GO, designed for the detection of SARS CoV-2 RNA a portable electrochemical biosensor. More precisely, they functionalized the r-GO 16 surface with p-sulfocalix[8]arene (SCX8) onto which they deposited Au@Fe 3 O 4 nanoparticles Αs schematically seen in Fig. 6Bi, before the deposition of Au nanoparticles on the modified r Among the materials developed to be applied as 22 transducer electrodes for SARS-CoV-2 electrochemical biosensors (Table 2), the one above-presented 23 exhibited the lowest response time, as well the lowest detectable concentration limit 75 . More precisely, 24 the as-suggested sensor is based on nucleic acid following the super sandwich-sensing type, which is 25 consisted of: i) a capture probe (CP, is captured on the catalytic surface area, usually an antibody), ii) a 26 target sequence (usually oligonucleotides or DNA or RNA, when the complementary sequence binds 27 with the target sequence, then an electric signal is produced), iii) a label probe (organic dyes, enzymes or Very recently Mahari et al. 96 , modified the screen For comparison reasons they also fabricated a home-made biosensor equipped with a Au-4 based fluorine doped tin-oxide (FTO) electrode. The detection procedure of the as-proposed platform is 5 based on the conductivity changes, caused by the binding events between the immobilized on the 6 electrode nCovid-19 monoclonal antibody (nCovid-19Ab) and the nCovid-19 antigen Using the technique of differential pulse voltammetry, they found that both biosensors exhibited high 8 sensitivity in then Covid-19Ag concentration range between 10 -15 M-10 -6 M. However, the low detection 9 limit of eCovSens was calculated at 10 fM (fM =10 -15 M), while the Au-FTO one estimated at 120 fM 96 Since the magnetic beads 15 have been set as biocompatible materials 98 , their combination with carbon black is another biosensor 16 configuration that has become increasingly exploited. The magnetic beads (placed under the working 17 electrode) were operated as the support of the immunological chain and they were able to: i) accept 18 higher antibody loading due to the high surface to volume ratio, ii) be moved to a specific volume of 19 analyte and washed to remove interfering species -2 were used as target analytes, bonded with the 22 immobilized-antibodies on the carbon surface. For achieving the optimum bonding between the 23 SARS-CoV-2 protein and the antibody, the latter was labeled with alkaline phosphate enzyme (1-of the sensor: i) its selectivity towards the S-Receptor Binding Domain (SRBD) protein, ii) 4 its sensitivity Ai) S-protein detection and (Aii) P-protein detection 97 ; (Bi) TiO 2 nanotubes 8 functionalized by Co (Co-TNTs), as electrode material for a SARS-CoV-2 biosensor; (Bii) Linear response 9 for concentration range between 14 nM to 1400 nM and a detection response time of ~30 sec 102 Schematic illustration of -NH 2 groups of the SARS-CoV-2 proteins or antibodies attached to 1-Pyrenebutyric 11 acid (PBA); (Cii) Current response of real samples detection of the N-proteins biosensor; (Dii) For laboratory sample the lower limit of detection of SARS protein, 13 was recorded at 1 fg/mL and 100 fg/mL 104 Titania nanotube arrays are already 2 reported as an effective material for application into electrochemical biosensors 105 . Their good sensing 3 ability is mainly attributed to their special morphology that offers high area to volume ratio, more pathways 4 for charge transfer and good adhesion to the substrate 106 . Moreover, their surface functionalization and the 5 addition of defects, seem to create more oxygen vacancy sites Graphene alteration by engraving before functionalization its surface with laser, is suggested as an 8 effective strategy for producing high-sensitive biosensors 108 103 fabricated a wireless, portable SARS-CoV-2 electrochemical For 13 accepting the analyte molecule, graphene surface was functionalized and modified, using 1-pyrenebutyric 14 acid (PBA), as the functional group (linker). The π-electrons and the carboxylic groups of PBA form 15 hydrophobic or non-covalent interactions (π-stacking) with the aromatic rings of graphene (electrostatic 16 bonds). Then, the -NH 2 groups of the SARS-CoV-2 proteins or antibodies are attached to the PBA, while 17 bovine serum albumin (BSA) protein was also introduced Contrary to the common functionalization, in which the functional groups directly bind the sp 2 20 carbon atom by modifying its edges and by introducing defects; pyrene, can provide a non-covalent 1 evaluated, using the amperometric technique and by increasing the analyte concentration and 2 recording the current response. The detected target-analytes were: i) proteins (N-protein and C-3 reactive protein), existing in the virus genome and ii) specific immunoglobulins IgM), produced from the immune response of the infected patient The specific lower limit of detection, recoded for each one of the analytes More precisely, the as-fabricated biosensor 103 was evaluated using electrochemical impedance 7 spectroscopy (EIS) and differential pulse voltammetry (DPV). Fig. 7Cii, shows the current response of 8 the biosensor being tested with real samples, containing N-proteins: S1-IgG, S1-IgM and CRP Furthermore, the lower detection limits were as follows: i) 0.1 g/L and 0.5 mg/L for N-protein in 10 blood and saliva respectively, ii) 20 g/L and 0.2 g/L for S1-IgG g/L for S1-IgM in blood and saliva respectively, and iv) of 10 g/L and 0.1 g/L for C-12 reactive protein in blood and saliva, respectively. Additionally, the as-proposed sensor 103 showed a 13 quick response and accuracy in a time In general, graphene, as a two-dimensional material, presents the ability to adsorb the analyte 15 molecule without any pretreatment; however, running the risk to be easily contaminated 112 . Thus, in 16 order to overcome this challenge, different strategies for changing graphene's properties to the 17 desired ones, including surface modification as well as functionalization in 19 which the last few years graphene was adopted as the active material 113-115 . Recently, a promising 20 electrochemical FET SARS-CoV-2 biosensor, applying graphene sheets as sensing material, was 21 fabricated by Seo et al. 104 . The operation principle of the as-fabricated biosensor is based on the 22 detection of SARS-CoV-2 spike protein bound on its antibody, which can be immobilized on graphene 23 sheets through a probe linker (1-pyrenebutyric acid N-hydroxysuccinimide ester, denoted as PBASE) 24 (Fig.7Di). The PBASE inclusion in the graphene sheets caused its chemical functionalization, creating 25 π-π bonds between them. The operational parameters of the suggested biosensor were studied 1 for real sample Looking for the most effective COVID-19 detection ways, electrochemical biosensors could be 5 very promising tools for: i) accurate, ii) low cost, iii) of high sensitivity, iv) quick and v) portable, 6 detection tools, acting as supplementary to the current diagnostic tools for controlling the COVID disease 7 expansion. In the current review we report the up-to-date developed SARS-CoV-2 electrochemical 8 biosensors Up to date, two main groups of materials have been thoroughlyexplored as transducer 10 electrodes: i) the Au-based and ii) the carbon or graphene-based ones. Both present faster response 11 time (within a few seconds) along with higher accuracy than the current detection methods, and most 12 of them has also higher sensitivity Comparing the two groups we conclude that the carbon or graphene-based ones can compete The use of nanomaterials into the electrochemical biosensors it is known 18 that improves their characteristics by far. However, the synthesis method of them, the combination of 19 the different nanomaterials, their nanostructure (nanoparticles, nanotubes, sheets, functionalized or 20 not), even their size and the way that they are used into an electrochemical biosensor (labels, support, 21 or auxiliary elements for signal enhancement) significantly affect its characteristics Au was mainly used in the form of nanoparticles onto alternative support (polymer based or other) or 24 supported onto reduced graphene oxide before being deposited onto the basic platform. The inclusion 25 of the r-GO to the Au nanoparticles significantly ameliorates SARS-CoV-2 sensor characteristics as References Applied Catalysis B: 13 Environmental Current 6 Opinion in Electrochemistry Bioinspired Materials for Medical Applications Label-free biosensing: advanced materials, devices and 18 applications Electroanalysis: An International Journal Devoted to Fundamental 23 and Practical Aspects of Electroanalysis TrAC Trends in 13 Analytical Chemistry Bioelectrochemistry: fundamentals, experimental techniques and applications Sensors and Actuators B: 22 Chemical Practical Approaches to Biological Inorganic Chemistry Nanoscience and its Applications Smart materials and structures 93. I. Novodchuk, M. Bajcsy and M. Yavuz Nanoscale materials help to improve electrode-analyte interface for developing label-free biosensors. 12 -Carbon or graphene materials functionalization enhances specificity and sensitivity towards SARS-CoV-2 13 S-protein Au-nanoparticles in combination with r-GO improve the sensitivity and detection time of SARS-CoV-2 Electrochemical impedance spectroscopy (EIS) is the most sensitive electrochemical detection technique In EIS, sometimes impedance changes can be caused by undesired binding events, leading to false results