key: cord-0789195-ryf8e2bw authors: Jones, Pete R.; Campbell, Peter; Callaghan, Tamsin; Jones, Lee; Asfaw, Daniel S.; Edgar, David F.; Crabb, David P. title: Glaucoma home-monitoring using a tablet-based visual field test (Eyecatcher): An assessment of accuracy and adherence over six months date: 2020-09-01 journal: Am J Ophthalmol DOI: 10.1016/j.ajo.2020.08.039 sha: f9a50b55a4a380b10bcb8b33a2ee1074b0abd5ec doc_id: 789195 cord_uid: ryf8e2bw PURPOSE: To assess accuracy and adherence of visual field (VF) home-monitoring in a pilot sample of glaucoma patients. DESIGN: Prospective longitudinal feasibility and reliability study. METHODS: Twenty adults (median 71 years) with an established diagnosis of glaucoma were issued a tablet-perimeter (Eyecatcher), and were asked to perform one VF home-assessment per eye, per month, for 6 months (12 tests total). Before and after home-monitoring, two VF assessments were performed in-clinic using Standard Automated Perimetry (SAP; 4 tests total, per eye). RESULTS: All 20 participants could perform monthly home-monitoring, though one participant stopped after 4 months (Adherence: 98%). There was good concordance between VFs measured at home and in the clinic (r = 0.94, P < 0.001). In 21 of 236 tests (9%) Mean Deviation deviated by more than ±3 dB from the median. Many of these anomalous tests could be identified by applying machine learning techniques to recordings from the tablets’ front-facing camera (Area Under the ROC Curve = 0.78). Adding home-monitoring data to 2 SAP tests made 6 months apart reduced measurement error (between-test measurement variability) in 97% of eyes, with mean absolute error more than halving in 90% of eyes. Median test duration was 4.5 mins (Quartiles: 3.9 – 5.2 mins). Substantial variations in ambient illumination had no observable effect on VF measurements (r = 0.07, P = 0.320). CONCLUSIONS: Home-monitoring of VFs is viable for some patients, and may provide clinically useful data. • Twenty glaucoma patients performed visual field assessments at home for 6 months • The data were in good agreement with conventional measurements made in clinic • Home monitoring may be viable and useful for some patients • Practical concerns and future directions for home monitoring are also discussed People with glaucoma, or at risk of developing glaucoma, require lifelong monitoring, including periodic (e.g., annual 1 ) visual field (VF) examinations 2 . The volume of outpatient appointments required (> 1 million/year in the UK alone 3 ) is placing glaucoma services under increasing strain: as evidenced by a growing appointment backlog 4 and instances of avoidable sight loss due to treatment delays 5, 6 . Globally, the challenge of glaucoma management is only likely to intensify over the coming decades 7 , with aging societies 8, 9 and calls for increased monitoring 1 and earlier detection 10 . Furthermore, hospital assessments cannot be performed with the frequency required for best patient care. Multiple studies have shown that intensive VF monitoring could help to identify and prioritize individuals most at risk of debilitating sight loss [11] [12] [13] [14] [15] (i.e., younger patients with fast-progressing VF loss 16 ). Frequent (e.g., monthly) monitoring is likely to be of particular benefit for those patients for whom rapid progression is most likely (e.g., optic disk hemorrhage [17] [18] [19] or most costly (e.g., monocular vision 20 ) . In short, the status quo of hospital-only VF monitoring is costly and insufficient. The solution may lie in home-monitoring 14, 21, 22 . By collecting additional VF data between appointments, hospital visits could be shortened, and in low-risk patients appointments could be reduced in frequency or conducted remotely: decreasing demand on outpatient clinics. Home-monitoring would further allow for more VF testing, and more frequent VF testing: both important for rapid, robust clinical decision-making 12, 23 . For these reasons, interest in home monitoring is growing both for glaucoma 14, 21, 22 , as well as for the management of other chronic ophthalmic conditions [24] [25] [26] [27] and in healthcare generally 28 . This interest is likely to intensify following COVID-19, as both hospitals and patients look to further minimize inperson appointments 29, 30 . Technological advances mean VF home-monitoring is now a realistic proposition. Several portable perimeters have been developed that use ordinary tablet-computers (e.g., Melbourne Rapid Fields 31-33 ; Eyecatcher 34 ) or head-mounted displays (e.g., imo 35, 36 , Mobile Virtual Perimetry 37 ). Such devices are small and inexpensive enough for patients to take home, and several appear capable of approximating conventional SAP when operated under supervision 32, 38, 39 . What remains unclear is whether VF home-monitoring works in practice. Are glaucoma patients willing and able to comply with a home-testing regimen (adherence)? And do 'personal perimeters' continue to produce high-quality VF data when operated at home and unsupervised (accuracy)? To investigate these questions, twenty people with established glaucoma were given a tablet perimeter (Eyecatcher) to take home for 6 months. They were asked to perform one VF assessment a month in each eye. Accuracy was assessed by comparing measurements made at home to conventional SAP assessments made at the study's start and end. Adherence was quantified as the percentage of tests completed. Eyecatcher is not yet available for general use, however the sourcecode is freely available online, as detailed in the Methods. To reflect the likely clinical reality of home-monitoring, we used only inexpensive and commonly available hardware (~$350 per person). Ten participants were given no practice with the test before taking it home. The other 10 performed the test once in each eye under supervision. During hometesting, the tablet-computer's forward-facing camera recorded the participant. This allowed us to confirm the correct eye was tested, to record variations in ambient illumination, and to investigate whether 'affective computing' techniques (e.g., head-pose tracking and facial-expression analysis to recognize human emotions) could identify suspect tests 40 . Participants were 20 adults (10 female) aged 62-78 years (Median: 71), with established diagnoses of: primary open angle glaucoma (N=18, incl. 6 normal tension), angle closure glaucoma (N=1), or secondary glaucoma (N=1). Participants lived across south England and Wales (see Supplemental Figure S1 ), and were under ongoing care from different consultant ophthalmologists. Participants were the first 20 respondents to an advertisement placed in the International Glaucoma Association newsletter (IGA News: https://www.glaucoma-association.com/about-the-iga/what-we-do/magazine), and were assessed by a glaucoma-accredited optometrist (author PC) who recorded: ocular and medical histories, logMAR acuity, and standard automated perimetry using a Humphrey Field Analyzer 3 (HFA; Carl Zeiss Meditec, CA, US; Swedish Interactive Threshold Algorithm (SITA) Fast; 24-2 grid). All patients exhibited best corrected logMAR acuity < 0.5 in the better eye, and none had undergone ocular surgery or laser treatment within six months prior to participation. Severity of visual field loss in the worse eye, as measured by HFA Mean Deviation (MD), varied from -2.5 dB (early loss 41 ) to -29.9 dB (advanced loss), although the majority of eyes exhibited moderate loss (Median: -8.9 dB). All HFA assessments (4 per eye) are shown in the Results, and all exhibited a False Positive rate below 15% (Median: 0%). Written informed consent was obtained prior to testing. Participants were not paid, but were offered travel expenses. The study was approved by the Ethics Committee for the School of Health Sciences, City, University of London (#ETH1819-0532), and carried out in accordance with the tenets of the Declaration of Helsinki. As shown in Figure 1A , participants were asked to perform one VF home-assessment per eye, per month, for 6 months (12 tests total per participant). Beforehand, participants attended City, University of London, where they were issued with the necessary equipment, including: a tablet computer ( Fig 1B) , an eye-patch, screen wipes, and a brief set of written instructions. All participants performed two HFA assessments in each eye (24-2 SITA Fast). Ten participants (50%) were also randomly selected to practice the Eyecatcher test once in each eye under supervision. During the 6-month home-testing period, participants had access to support via telephone and email, and received an email reminder once a month when the test was due. After the home-monitoring period was complete, participants returned to City, University of London, and again performed two HFA assessments in each eye. They also completed a semi-structured interview, designed to assess the acceptability of home monitoring, and to identify any potential barriers to use. A qualitative assessment of these interviews will be reported elsewhere. One participant (ID16) was unable to return due to the COVID-19 quarantine. They instead mailed their computer, and performed their interview via telephone. No follow-up HFA assessment could be performed in this individual, but given their ocular history their VF is expected to have been stable. [ Figure 1 About Here -Legend at end of manuscript] J o u r n a l P r e -p r o o f Visual fields were assessed using a custom screen-perimeter (Fig 1B) , implemented on an inexpensive HP Pavilion x360 15.6" tablet-laptop (HP Inc., Paolo Alto, CA, United States). The test was a variant of the 'Eyecatcher' visual field test: described previously 34, 39 and freely available online at https://github.com/petejonze/Eyecatcher. It was modified in the present work to more closely mimic conventional static threshold perimetry; most notably by: employing a ZEST thresholding algorithm 42 , a central fixation-cross, and a button press response. The software was implemented in MATLAB using Psychtoolbox v3 43 , and used bit-stealing to ensure >10-bit luminance precision 44 . The display measured 34.5 x 19.5 cm (34.8 x 20.1 degrees visual angle, at the nominal viewing distance of 55 cm), and extensive photometric calibration was performed on each device to ensure luminance uniformity across the display (see Supplemental Material for technical details regarding screen calibration). During the test, participants were asked to fixate a central cross, and to press a button when they saw a flash of light (Goldmann III dots with Gaussian-ramped edges). As in conventional perimetry, targets were presented against a 10 cd/m 2 white background. Unlike conventional perimetry, participants received visual feedback (a 'popping' dot) at the true stimulus location after each button press. This feedback was intended to keep participants motivated and alert during testing and was generally wellreceived by participants, though four reported being sometimes surprised when feedback appeared at an unexpected location. Testing was performed monocularly (fellow eye patched). The right eye was always tested first, and participants could take breaks between tests. Participants were asked to position themselves 55 cm from the screen (a distance marked on the response-button cable), and to perform the test in a dark, quiet room. In practice, we had no control over fixation stability, viewing distance, or ambient lighting. In anticipation that these may be important confounding factors, participants were recorded during testing using the tablet's front facing camera (see Results). Note that the 55 cm viewing distance is farther than the conventional perimetric distance of 33 cm. This was partly for consistency with previous versions of Eyecatcher 34, 39 (versions that incorporated near-infrared eye-tracking, which required a ~55cm viewing distance). It was also partly to reduce (by a factor of ~1.5), the extent to which any head-movements affected retinal stimulus size/location (i.e., given the lack of chin rest). Note, however, that 33 cm has been used successfully by other tablet perimeters 32 , and would have allowed for the whole horizontal-extent of the 24-2 grid to have been tested. As shown in Fig 1C, the output of each Eyecatcher assessment was 4 x 6 grid of differential light sensitivity (DLS) estimates, corresponding to the central 24 locations of a standard 24-2 perimetric grid (±15° horizontal; ± 9° vertical). For analysis and reporting purposes, these values were transformed to be on the same decibel scale as the HFA = 10 3183.1 / ⁄ . Due to the limited maximum-reliable luminance of the screen (175 cd/m 2 ), the measurable range of values was 12.6 dB to 48 dB (HFA dB scale). Sensitivities below 12.6 dB could not be measured, and were recorded as 12.6 dB. Note, it has been suggested that with conventional SAP, measurements below ~15 dB are unreliable and of limited utility [45] [46] [47] . The MRF iPad app has shown promising results under laboratory settings 32 , and was considered for the present study. We chose to use our opensource Eyecatcher software primarily for practical reasons (i.e., we were familiar with it, and could modify it to allow camera recordings and individual screen calibrations). Where appropriate, and as indicated in the text, pointwise DLS values from the HFA were adjusted for parity with Eyecatcher by setting estimated sensitives below 12.6 dB to 12.6 dB. MD values were then recomputed as the weighted-mean difference from age-corrected normative values 48 J o u r n a l P r e -p r o o f III. RESULTS Figure 2 shows Mean Deviations (MD) for all eyes/tests. Adherence (percentage of tests completed) was 98.3%. Nineteen of 20 individuals completed the full regimen of 6 home-monitoring sessions. Participant 20 discontinued home-testing after 4 sessions/months following consultation with the study investigators. This was due to the test exacerbating chronic symptoms of vertigo (also experienced following SAP). [ Figure 2 About Here -Legend at end of manuscript] MD scores were strongly associated between VFs measured at home (mean of 6 Eyecatcher tests) and those measured in the lab (mean of 4 HFA tests), with a correlation of r 38 = 0.94 [ Fig 3A; Pearson Correlation; P < 0.001] and a 95% Coefficient of Repeatability of ±3.4 dB (Fig 3B) . For reference, mean agreement between random pairs of HFA assessments was 2.2 dB (CI 95 : 1.8-2.6 dB; 20,000 random samples). As shown in Figure 4 , there was also good concordance between individual VF locations [Pearson Correlation; r 878 = 0.86, P ≪ 0.001]. Some individual tests produced implausible data (e.g., Fig 2: ID8 test 3, ID12 test 5). In total, there were 21 tests (9%) where MD deviated by more than ±3 dB from the average (median of all 6 tests). Of these, 13 (62%) occurred in the right eye (tested first), and 7 (33%) deviated by more than ±6 dB. As described in Supplemental Figure S6 , these statistical outliers could be identified with reasonable sensitivity/specificity [Area Under the ROC Curve: 0.78] by applying machine learning techniques to recordings from the tablets' front-facing camera. To quantify the extent to which regular home monitoring reduced VF measurement error (between test variability), Figure 5 shows the estimated rate-of-change (least-squares slopes) at each VF location. We assume that for the 6-month study period the true change in sensitivity was approximately zero, and so any non-zero slope estimates represent random error. This assumption is reasonable given the relatively short timeframe, that all participants were believed to be perimetrically stable, and the fact that when all four HFA tests were considered, almost as many points exhibited positive slopes (increasing sensitivity, Fig 5A, When only a single (randomly selected) pair of HFA pre-and post-test results was considered (i.e., the current clinical reality following two hospital appointments), mean absolute error was 1.96 dB (CI 95 : 1.7 -2.3; Fig 5B, grey shaded region) . As progressively more home-monitoring tests were also considered ( Fig 5B, filled circles) 16) , mean absolute error [MAE] was smaller when home-monitoring data were included, with MAE reducing by more than 50% in 90% of eyes (Median reduction: 85%, CI 95 : 82 -87%). For reference, a reduction of 20% in variability is generally considered clinically significant, and allows progression to be detected one visit earlier 49 . If we consider the homemonitoring data alone (i.e., without any HFA data included; Fig 5B, unfilled squares) , measurement error was still smaller after 6 home monitoring tests (0.78 dB; CI 95 : 0.6 -1.1) versus two HFA tests alone (1.96 dB), with a median reduction in MAE of 68% (CI 95 : 57 -76%). [ Figure 5 About Here -Legend at end of manuscript ] Either with or without HFA data included, there was no significant difference in MAE between the eyes of participants who received initial practice with Eyecatcher, and those who did not [Independent samples t-test; P with = 0.864, P without = 0.812]. In some individuals (e.g., ID3, ID13), MDs measured at home were systematically higher, in both eyes, than those measured in clinic. This difference was not significant across the group as a whole [Repeated measures t-test of MD: t 39 = -1.08, P = 0.286], and may indicate individual differences in fixation stability or viewing distance. They are not likely due to ambient illumination levels, which tended to be highly variable (both within-and between-individuals), but with little apparent effect on the data (see Supplemental Figures S10-11) . Median test duration for Eyecatcher was 4.5 mins (Quartiles: 3.9 -5.2 mins) and did not vary systematically across the 6 sessions [F (5,227) = 0.808, P = 0.547; see Supplemental Figure S12 ]. For comparison, median test duration for the HFA (SITA Fast) was 3.9 mins (Quartiles: 3.3 -4.6 mins), and was faster than Eyecatcher in 30 of 40 eyes (despite the HFA testing over twice as many VF locations). Home-monitoring has the potential to deliver earlier and more reliable detection of disease progression, as well as service benefits via a reduction in in-person appointments. Here we demonstrate, in a preliminary sample of 20 volunteers, that glaucoma patients are willing and able to comply with a monthly VF home-testing regimen, and that the VF data produced were of good quality. 98% of tests were completed successfully (adherence), and the data from 6 home-monitoring tests were in good agreement with 4 SAP tests conducted in clinic (accuracy). This is consistent with previous observations that experienced patients can perform VF testing with minimal oversight 50 , as well as with recent findings from the Age-Related Eye Disease Study 2 (AREDS2)-HOME study group, showing that home-monitoring of hyperacuity is able to improve the detection of neovascular agerelated macular degeneration (AMD) 51 . The use of home-monitoring data was shown to reduce measurement error (between-test measurement variability). When home-monitoring data were added to 2 SAP assessments made 6 months apart (the current clinical reality), measurement error decreased by over 50% in 90% of eyes. Given that a 20% reduction in measurement variability is generally considered clinically significant (i.e., allows progression to be detected one hospital-visit earlier 49 ), this suggests that, even with present technology, home monitoring could be beneficial for routine clinical practice (e.g., support more rapid interventions). Furthermore, while we assume that ancillary home-monitoring, designed to supplement and augment existing SAP, would be the generally preferred model, it was encouraging that robust VF estimates were obtained even when home-monitoring data were considered in isolation. This suggests that home-monitoring may be viable in situations where hospital assessments are impractical, such as in domiciliary care, or in the wake of pandemics such as COVID-19 52 . Home monitoring could also assist with clinical trials. For example, the recent UKGTS trial 53 required 516 individuals to attend 16 visual field assessments over 24 months: a substantial undertaking, of the sort that can make new treatments prohibitively costly to assess 54, 55 . By allowing more frequent measurements of geographically diverse individuals, home-monitoring could lead to cheaper, more representative trials, and could potentially reduce trial durations (i.e., evidence treatment effects sooner). There were, however, individual instances where the home-monitoring test performed poorly. In 21 tests (9%), MD deviated by more than ±3 dB from the median (7 deviating by more than ±6 dB). As has been shown elsewhere by simulation 14 , the effects of these anomalous tests were largely compensated for by the increased volume of 'good' data. However, poor quality data should ideally be averted at-source, and it was encouraging that many of these 21 anomalous tests could be identified by applying machine-learning techniques to recordings of participants made using the tablets' frontfacing camera (see Supplemental Figure S6 ). It is also notable that when interviewed at the end of the study, some participants already suspected some tests of being anomalous (e.g., due to a long test duration, or a feeling that they had not performed well). Consideration may therefore need to be given in future as to whether participants should have the ability to repeat tests or provide confidence ratings. Regarding adherence, one participant (ID20) was advised by the study team to discontinue homemonitoring after 4 months, after reporting that the test was compounding chronic symptoms of dizziness (though interestingly their data appeared relatively accurate and consistent up to this point; see Fig 2) . This adverse effect was not unique to Eyecatcher, and the participant reported similar reactions following conventional SAP. However, this highlights that it may be helpful to tailor the use and frequency of home-monitoring to the needs and abilities of individual patients, in contrast to the current "one size fits all" approach to VF monitoring 12, 56, 57 . A full qualitative analysis of participants' views on the benefits and challenges of home-monitoring is in preparation, and will be reported elsewhere. The present study is only an initial feasibility assessment, examining a small number of self-selecting volunteers. It remains to be seen how well home-monitoring scales up to routine clinical practice or clinical trials. It will be particularly important to establish that home-monitoring is sustainable over longer periods, and is capable of detecting rapid progression [11] [12] [13] [14] [15] . Cost-effectiveness of glaucoma home-monitoring has also yet to be demonstrated, and it would be helpful to perform an economic evaluation of utility, similar to that reported recently for AMD homemonitoring 58 . For this, it would be instructive to consider not just home-monitoring of VFs alone, but also in conjunction with home-tonometry, which also appears increasingly practicable 59 . Long-term, there are even signs that optical coherence tomography 60 (OCT) and smartphone-based fundusimaging [61] [62] [63] are becoming straightforward enough to be administered by lay persons, and these might also be explored in future home-monitoring trials. Further, it may be that targeted home-monitoring ---focused on high-risk/benefit glaucoma patients ---is cost-effective, even if the indiscriminate home-monitoring of all patients is not 14 . Thus, it may be best to concentrate home-monitoring resources on those patients whose age 11 or condition [17] [18] [19] makes them most likely to experience debilitating vision loss within their lifetime. It may also be worth considering the potential secondary benefits of home-monitoring, such as improved patient satisfaction and retention 56, 64 , or better treatment adherence. Thus, it is well established many glaucoma patients find hospital visits stressful and inconvenient 56, 65 , and home-monitoring might be welcomed as a way of saving time, travel, and money. Treatment adherence is known to increase markedly prior to a hospital appointment 66 ("white-coat adherence"), or when patients receive automated reminders 67, 68 , and it is conceivable that the anticipation of regular home monitoring could provide a similar impetus. Following COVID-19, home-monitoring of VFs may also be desirable from a public health perspective, as a way of reducing the time each patient spends in clinic, and as a way of reducing the risk (real or perceived) of infection from conventional SAP apparatus. The test itself (Eyecatcher) was intended only as a proof of concept, and was crude in many respects. In fact, we consider it highly encouraging ---and somewhat remarkable ---that the results were as promising as they were, given the low level of technical sophistication. Alternative measures are being developed elsewhere [31] [32] [33] [35] [36] [37] (in particular, the MRF), and there are several ways in which the present test could also be improved in future. The test algorithm (a rudimentary implementation of ZEST 42 ) was relatively inefficient, and could be made faster and more robust: most straightforwardly by using prior information from previous tests, and by using a more efficient stimulus-selection rule 69 . Increased efficiency might be necessary if, for example, attempting to test all 54 locations in a standard 24-2 grid. The source code for the present test is freely available online for anyone wishing to view or modify it. Interestingly, while Anderson and colleagues 14 anticipated home tests would be brief, the relatively long durations in the present study (Median: 4.5 mins per eye) were not cited as a concern by participants (although two individuals observed that test durations were longer and more variable than conventional SAP). It may be that when it comes to home-monitoring, less focus should be placed on test duration than in conventional perimetry (i.e., given the time saved by not having to travel to and wait in clinic). Instead, focus should be directed more towards usability (e.g., the ability to pause, resume, or restart tests). A further key limitation of the present test is that only para-central vision was assessed (±15° horizontal; ± 9° vertical; the most central 24 points of the 24-2 grid). While this seemed sufficient to assess the feasibility of home-monitoring in principle, such a limited field of view would in practice hinder clinician's ability to determine progression in the size or shape of field loss, and key areas of loss may be missed altogether (i.e., much of the superior and inferior arcuate nerve fiber bundle areas were not tested). A wider field of view could be achieved by using a larger screen 38, 70 (at the cost of reduced portability), by requiring the user to fixate different areas of the screen throughout the course of the test [31] [32] [33] 39 (at the cost of increased complexity), or by reducing viewing distance [31] [32] [33] (at the cost of greater measurement error due to head movements; see Methods). Alternatively, the future use of head-mounted displays (or 'smart-glasses') would allow for widefield testing, and would also obviate many practical concerns regarding uncontrolled viewing distance, improper patching, screen glare, or variations in ambient lighting. These potential confounds did not appear to be limiting factors in the present study, but could be problematic in less compliant individuals, or those disposed to cheat or malinger. Other ways in which the present hardware could be improved are by using eyetracking to monitor fixation; using near-infrared facial imaging systems, such as the iPad's TrueDepth camera, to track viewing distance with millimeter accuracy; and/or by integrating iris-scanning to ensure that the correct eye/person is always tested. Long-term, test data will need to be integrated securely into medical records systems, and consideration given how to maintain accurate screen calibrations over extended periods of use 70 . ('Eyecatcher'). During each Eyecatcher assessment, live recordings of the participant were made via the screen's front facing camera (purple arrow). Participants were asked to fixate the central red cross throughout, and to press the button when a white (Goldmann III) dot was seen (C) Output: Example measures of visual field loss from a single participant, with same-patient data from the HFA for comparison. Greyscales were generated using MATLAB code available at: https://github.com/petejonze/VfPlot. Summary of visual field loss (Mean Deviation; MD) for all eyes/tests. Each panel shows the complete data from a single participant. Numbers in the top-left of each panel give participant ID, with asterisks denoting the 10 individuals who received initial practice with Eyecatcher. The right eye (red circles) was always tested first, followed by the left eye (blue squares). Lightfilled markers show the results for monthly Eyecatcher home-monitoring assessments. Dark-filled markers show the results of two HFA pre-tests and two HFA post-tests (all tests performed consecutively, same day). For parity, HFA values were computed using only the same 22 (paracentral) test locations as Eyecatcher, and any estimated sensitivities below 12.6 dB were set to 12.6 dB (to reflect the smaller dynamic range of the Eyecatcher test). Small unfilled markers show the unadjusted MD values as reported by the HFA (i.e., using all 52 test points, and using the full dynamic range). These unfilled markers are most visible (i.e., deviated from the adjusted values) only when field loss was severe. Note that participant 20 chose not complete the final two home monitoring tests, while participant 16 was unable to perform the final HFA assessments due to COVID-19 (see Body Text for details). and last columns show mean-averaged data from two 'pre' and two 'post' reference tests, performed in clinic using a Humphrey Field Analyzer 3 (24-2, SITA Fast). The solid grey regions in the Eyecatcher plots denote those regions of the 24-2 grid not tested due to limited screen size. Only half of participants were randomly selected to complete a supervised practice test. , and show how measurement error decreased as Eyecatcher data were added to a random pair of HFA pre /HFA post assessments (i.e., 'Ancillary Home Monitoring Scenario'). Unfilled markers show measurement error if Eyecatcher data were considered in isolation, without any HFA data (i.e., 'Exclusive Home Monitoring Scenario'). Error bars denote 95% confidence intervals, derived using Bootstrapping (Bias-corrected accelerated method, N=20,000). The shaded region highlights the 95% confidence interval (CI 95 : 1.7 -2.3 dB) given only a single random pair of HFA assessments (i.e., the current clinical reality after two appointments). Are practical recommendations practiced? A national multi-centre cross-sectional study on frequency of visual field testing in glaucoma Tackling the NHS glaucoma clinic backlog issue Surveillance of sight loss due to delay in ophthalmic treatment or review: frequency, cause and outcome Healthcare Safety Investigatory Branch. Lack of timely monitoring of patients with glaucoma: Healthcare Safety Investigation Challenges in the management of glaucoma in developing countries Global prevalence of glaucoma and projections of glaucoma burden through 2040: a systematic review and meta-analysis How to defuse a demographic time bomb: the way forward? Prospective study of glaucoma referrals across Europe: are we using resources wisely? Examining visual field loss in patients in glaucoma clinics during their predicted remaining lifetime Frequency of testing to detect visual field progression derived using a longitudinal cohort of glaucoma patients Frequency of testing for detecting visual field progression Can Home Monitoring Allow Earlier Detection of Rapid Visual Field Progression in Glaucoma? More frequent, more costly? Health economic modelling aspects of monitoring glaucoma patients in England Rates of glaucomatous visual field change in a large clinical population Risk factors for rapid glaucoma disease progression Factors affecting rates of visual field progression in glaucoma patients with optic disc hemorrhage Evaluating the Impact of Uveitis on Visual Field Progression Using Large-Scale Real-World Data The Only Eye Study (OnES): a qualitative study of surgeon experiences of only eye surgery and recommendations for patient safety Home monitoring as a useful extension of modern teleophthalmology Home monitoring for glaucoma Frequency of visual field testing when monitoring patients newly diagnosed with glaucoma: mixed methods and modelling Randomized trial of a home monitoring system for early detection of choroidal neovascularization home monitoring of the Eye (HOME) study The current state of teleophthalmology in the United States Monitoring for neovascular age-related macular degeneration (AMD) reactivation at home: the MONARCH study Home Monitoring of Retinal Sensitivity on a Tablet Device in Intermediate Age-Related Macular Degeneration Towards a smart medical home Overview of Digital Transformation and Telemedicine during COVID19 Implementing and Monitoring At-Home VR Oculo-kinetic Perimetry During COVID-19 Performance of iPad-based threshold perimetry in glaucoma and controls Six-month Longitudinal Comparison of a Portable Tablet Perimeter With the Humphrey Field Analyzer A comparison of perimetric results from a tablet perimeter and Humphrey field analyzer in glaucoma patients Portable Perimetry Using Eye-Tracking on a Tablet Computer --A Feasibility Assessment Visual field testing with head-mounted perimeter `imo' Comparison of head-mounted perimeter (imo®) and Humphrey Field analyzer Low-cost, smartphone-based frequency doubling technology visual field testing using a head-mounted display An open-source static threshold perimetry test using remote eye-tracking (Eyecatcher): Description, validation, and normative data Using an open-source tablet perimeter (Eyecatcher) as a rapid triage measure in a glaucoma clinic waiting area The Human Touch: Using a Webcam to Autonomously Monitor Compliance During Visual Field Assessments Categorizing the stage of glaucoma from pre-diagnosis to end-stage disease Properties of perimetric threshold estimates from full threshold, ZEST, and SITA-like strategies, as determined by computer simulation What's new in Psychtoolbox-3. Perception Bit stealing: how to get 1786 or more gray levels from an 8-bit color monitor Assessment of the reliability of standard automated perimetry in regions of glaucomatous damage Effect of restricting perimetry testing algorithms to reliable sensitivities on test-retest variability Reducing variability of perimetric global indices from eyes with progressive glaucoma by censoring unreliable sensitivity data Normal variability of static perimetric threshold values across the central visual field What reduction in standard automated perimetry variability would improve the detection of visual field progression? Continuous visual field test supervision may not always be necessary Effectiveness of Different Monitoring Modalities in the Detection of Neovascular Age-Related: Macular Degeneration: The HOME Study The COVID-19 pandemic will redefine the future delivery of glaucoma care Latanoprost for open-angle glaucoma (UKGTS): a randomised, multicentre, placebo-controlled trial Glaucoma treatment: by the highest level of evidence Glaucoma neuroprotection trials are practical using visual field outcomes 21st century glaucoma care A view on glaucoma --are we seeing it clearly? Economic evaluation of a home-based age-related macular degeneration monitoring system The Icare HOME (TA022) study: performance of an intraocular pressure measuring device for self-tonometry by glaucoma patients Use of optical coherence tomography by non-expert personnel as a screening approach for glaucoma Evolution of optic nerve photography for glaucoma screening: a review Comparison of smartphone ophthalmoscopy with slitlamp biomicroscopy for grading vertical cup-to-disc ratio Undilated versus dilated monoscopic smartphone-based fundus photography for optic nerve head evaluation Treatment for glaucoma: adherence by the elderly A qualitative investigation into patients' views on visual field testing for glaucoma monitoring Adherence and persistence with glaucoma therapy Automated telecommunication-based reminders and adherence with once-daily glaucoma medication dosing: the automated dosing reminder study Interventions improve poor adherence with once daily glaucoma medications in electronically monitored patients QUEST+: A general multidimensional Bayesian adaptive psychometric methodWatson Plug and play perimetry: Evaluating the use of a self-calibrating digital display for screen-based threshold perimetry This study evaluates the feasibility of visual field home-monitoring. It indicates that glaucoma patients (N = 20) are willing and able to perform regular home-monitoring, and that such tests are capable of delivering clinically useful data.