key: cord-0788488-i4kg1gbv authors: Pettit, M.; Boswell, S.; Qian, J.; Novak, R.; Springer, M. title: Accessioning and automation compatible anterior nares swab design date: 2020-09-25 journal: nan DOI: 10.1101/2020.09.24.20200386 sha: 7c9a4a1761cb78d9ec610e2a16f3c04dd58b66f3 doc_id: 788488 cord_uid: i4kg1gbv The COVID-19 pandemic has resulted in an unparalleled need for viral testing capacity across the world and is a critical requirement for successful re-opening of economies. The logistical barriers to near-universal testing are considerable. We have designed an injection molded polypropylene anterior nares swab, the RHINOstick, with a screw cap integrated into the swab handle that is compatible with fully automated sample accessioning and processing. The ability to collect and release both human and viral material is comparable to that of several commonly used swabs on the market. SARS-CoV-2 is stable on dry RHINOstick swabs for at least 3 days, even at 42{degrees}C, and elution can be achieved with small volumes. The swab and barcoded tube set can be produced, sterilized, and packaged at < 2 USD per unit and can easily be adopted by large research institutes to increase throughput and dramatically reduce the cost of a standard SARS-CoV-2 detection pipeline. dry RHINOstick swabs for at least 3 days, even at 42°C, and elution can be achieved with small volumes. 23 The swab and barcoded tube set can be produced, sterilized, and packaged at < 2 USD per unit and can 24 . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) preprint The copyright holder for this this version posted September 25, 2020. . https://doi.org/10.1101/2020.09.24.20200386 doi: medRxiv preprint easily be adopted by large research institutes to increase throughput and dramatically reduce the cost of a 25 standard SARS-CoV-2 detection pipeline. 26 As of September 8 th 2020, at least 27 million cases of COVID-19 and 890,000 deaths have been 29 reported world-wide (1) . To determine if a patient has COVID-19, in most cases, a nasopharyngeal (NP) 30 swab is collected by a trained professional. The swab is then deposited in 1-3 mL of transport media 31 followed by RNA purification and RT-qPCR. NP swabs are around 15 cm in length with a collection 32 head coated with short synthetic filaments, flock, or spun fibers (2); collection is often an uncomfortable 33 process. The high demand for testing during this pandemic has outstripped the supply of NP swabs (and 34 many other critical reagents for testing) resulting in a testing bottleneck (3). These supply limitations 35 together with a drive towards patient self-collection has spurred the development of alternatives to the 36 standard NP swab. A promising alternative is anterior nares (AN) swabs, commonly referred to as nasal 37 swabs. AN swabs offer a testing sensitivity similar to NP swabs (4, 5) but are easier to use and more 38 comfortable for the patient. 39 The choice of swab and collection device can have a major impact on the testing speed in clinical 40 labs. Upon receiving samples, a typical procedure in a testing facility is to first accession the delivered 41 patient samples by scanning the barcoded label to upload relevant patient data into the system, then swabs 42 are manually removed from each collection tube and disposed of. The sample transport media is then 43 processed to purify RNA, which is used as input for RT-qPCR. The initial steps in this procedure are hard 44 to automate, slow, and expose staff to infection. Standard 1D barcoding systems and the manual removal 45 of swabs is time consuming and thus costly. There are machines that can perform the entire procedure 46 from accessioning to results, one tube at a time, e.g. a cobas® 8800, but this process is slow, 1056 tubes 47 per 8-hour shift (6), and the machines are expensive. 48 In an effort to meet the dramatic increase in demand for nasal swabs, several groups have 49 designed and 3D printed new swabs (2, 7). The performance of these swabs is comparable to that of 50 . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) preprint The copyright holder for this this version posted September 25, 2020. . https://doi.org/10.1101/2020.09.24.20200386 doi: medRxiv preprint standard swabs; however, they aim to reproduce the existing status quo, rather than to address some of the 51 limitations caused by the standard swab design. An ideal swab would be one that is comfortable for 52 patients to self-administer without sacrificing performance, while also allowing for automated specimen 53 accessioning and processing. Additionally, the swab would be made from non-absorbent material, 54 allowing samples to be diluted into smaller volumes of transport media than those used in the current 55 procedure, rendering the sample more concentrated and allowing for more sensitive detection of viral 56 RNA. Here we present the RHINOstick, a swab that 1) performs as well as existing AN swabs; 2) is 57 compatible with direct input to RT-qPCR for extraction free SARS-CoV-2 detection; and 3) is compatible 58 with a collection system (swab and tube) that enables automated sample accessioning and processing. 59 Swab design. The swabs were designed in SolidWorks (Dassault Systèmes) and manufactured 62 using single-shot rapid injection molding (Protolabs) from medical grade FHR P5M4R polypropylene 63 (Flint Hills), a material compatible with autoclaving (121°C, 20 min), ethylene oxide, gamma radiation, 64 and e-beam sterilization. The stacked rings of the swab head enable collection of nasal matrix without the 65 need for an absorbent coating, using a design previously developed at the Wyss Institute. The cap cavity 66 is compatible with automated decapping robot systems using a square profile adapter head, while the 2 67 mm pitch external threading mates with the interior threading of sample collection tubes from several 68 major manufacturers (e.g. Matrix, Micronics, and LVL). As the new swab we developed in this study is 69 useful for the collection of nasal samples for diagnostic tests we call it the RHINOstick swab. Absorption of liquid by swab. The swabs used in this study were weighed on an analytical 71 balance before and after a 15 second incubation in 1 mL of nuclease free water. Six replicates were 72 measured and results are reported in Table 1 . Anterior nares self-swabbing to compare swab performance. We compared several swab 74 types for performance in anterior nares (AN) specimen collection: the RHINOstick prototype, Proctor & 75 Gamble (P&G) blue prototype, a Wyss Institute flocked prototype, Puritan hydraflock, Puritan foam, 76 . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) preprint The copyright holder for this this version posted September 25, 2020. . https://doi.org/10.1101/2020.09.24.20200386 doi: medRxiv preprint Puritan polyester, US Cotton, and Microbrush®. Per CDC guidelines, volunteers were instructed to insert 77 the swab 0.5 inch into a nostril, rotate three times along the membrane of the nose firmly and leave in 78 place for 10 to 15 seconds, remove, and then repeat this procedure on the other nostril with the same swab 79 to collect nasal matrix (8). The volunteer was then instructed to place the used swab in a dry 1. Table S1 . 93 Recovery of human mRNA from AN swabs. SARS-CoV-2 negative volunteers performed AN 94 swabbing as directed with each type of swab tested (Fig. 2, Table 1 ) to collect nasal matrix. There were 95 three biological replicates for each AN swab measurement, taken on at least two different days. For every 96 condition in which a swab was tested, an unused swab, without nasal matrix, was processed in parallel as 97 a negative control. To recover the sample from the swabs, all swabs were suspended in 200 µL of 1x 98 PBS, vortexed for 10 sec, spun down in a microcentrifuge, and input directly to the RT-qPCR for 99 GAPDH mRNA detection (Fig. 2C) . 100 AccuPlex SARS-CoV-2 verification panel v2 (Seracare), a packaged synthetic virus, containing the N 102 . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) preprint The copyright holder for this this version posted September 25, 2020. . https://doi.org/10.1101/2020.09.24.20200386 doi: medRxiv preprint near the limit of detection (Fig. 2D, S2C ). 10 µL of 100 copies/µL packaged synthetic virus was directly 104 applied to the collection head of each swab. Swabs were left in a fume hood for about 20 min until the 105 swabs appeared dry to the eye indicating absorption of the packaged synthetic virus into the collection 106 material. At least three biological replicates were used for every swab tested and replicate data was 107 collected on at least two different days. Swabs were then inserted into a 1.5 mL microcentrifuge tube 108 that were de-identified by BiocaBiolistics with their IRB reviewed and approved SOP for de-linking 114 specimens. These NP swabs arrived in 1-3 mL of viral, multitrans, or universal transport media (VTM, 115 MTM, or UTM). 40 µL of each sample was aliquoted and frozen at -80°C to limit freeze-thawing of 116 Contrived samples from a clinical source spiked onto swabs with nasal matrix. Nasal matrix 118 was collected from volunteers as described above using RHINOstick and Puritan foam swabs. 5 µL of 119 clinical sample, with either a higher (~1600 copies/µL), or lower titer (~140 copies/µL), were applied to 120 the collection head of used swabs, and swabs were air dried in the BSL2+ biosafety cabinet for 20 min. 121 Each swab was then placed in a 1.5 mL microcentrifuge tube containing 200 µL of 1x PBS, manually 122 spun for 10 sec in the media, and 2 µL was directly input to the RT-qPCR for both N gene (Fig. 2E) and 123 GAPDH mRNA detection (Fig. S2C) . To assess maximum possible viral recovery from the swab, the Three biological replicates were performed for each titer and type of swab tested. 127 . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. swabs that can be directly paired with automation-compatible tubes for an effective collection system. As 152 part of the design, RHINOstick swabs have a cap that can be directly screwed onto a 96-well format 153 . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) preprint The copyright holder for this this version posted September 25, 2020. . https://doi.org/10.1101/2020.09.24.20200386 doi: medRxiv preprint automation-compatible tube, such as a 1.0 mL Matrix tube (Thermo Fisher Scientific) (Fig. 1A) . The 154 swabs were made by single shot injection molding with medical grade polypropylene ( Fig. S1 and 155 Methods). Injection molding of swabs allows for high volume production at low prices. While the swabs 156 can fit onto many tubes, we believe the optimal design is in collection tubes pre-labeled (by the 157 manufacturer) with a serialized Type 128 1D barcode plus human readable code on the side with a 158 matching 2D data matrix barcode on the bottom. (Fig. 1) . This design allows for the collection tube and 159 swab to be accessioned and used by the patient in an unobserved manner without having to pre-register 160 each barcode manually, reducing costs and labor. In addition, the matching 2D barcode on the bottom 161 allows a whole rack of tubes to be accessioned in seconds by a barcode reader. 162 Swab performance. We compared the RHINOstick to several other swabs on the market or 163 under development ( Fig. 2A, Table 1 ). First, we tested for absorption of water. Water absorption is 164 sometimes used as a proxy for the amount of material that a swab will collect (9, 10), although it does not 165 necessarily correlate with effective collection of cells and viral particles. The RHINOstick, as well as the 166 Proctor and Gamble (P&G) blue swab absorbed very little water compared to the majority of available 167 swabs on the market or prototypes. This lack of absorption is likely because polypropylene is more 168 hydrophobic than the other collection materials, such as cotton and spun polyester. 169 To test swab performance more directly, we measured the performance of 8 different AN swabs 170 using several approaches (Fig. 2) . We tested collection and recovery of 1) human mRNA in nasal matrix 171 from swabs, 2) mRNA from viral particles added to swabs, and 3) mRNA from viral particles added to 172 swabs coated in nasal matrix (Fig. 2B) . Human mRNA was used as a process control to assess successful 173 collection and recovery of cells from swabs. The process control also assesses the efficiency of the 174 reverse transcription (RT) reaction as the primers span two exons to ensure the assay quantifies mRNA 175 not DNA (11). A single volunteer swabbed with 8 different brands of AN swabs in triplicate (Fig. 2B, 176 scheme I) and the eluent was used as direct input for RT-qPCR for GAPDH mRNA to quantify the 177 amount of human mRNA recovered (Fig. 2C ). All 8 swabs performed similarly in this assay and no 178 . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this this version posted September 25, 2020. . https://doi.org/10.1101/2020.09.24.20200386 doi: medRxiv preprint we performed direct RT-qPCR on the swab eluant without RNA purification. 180 synthetic SARS-CoV-2 viral particles (Seracare reference) to an unused swab for each of the 8 AN swabs 182 tested (Fig. 2B, scheme II) . The packaged synthetic virus was dried onto the swab and eluted into PBS by 183 vortexing. In a similar experiment, we found that elution into PBS by gentle swirling of the swabs 184 releases the virus at equivalent or superior levels to vortexing in the same amount of time ( Fig. S2A and 185 B). The level of viral particles released by each swab was quantified by RT-qPCR for the SARS-CoV-2 N 186 gene (Fig. 2D) . The RHINOstick performed as well as the other swabs tested, and released an equivalent 187 number of viral particles to the positive control (Fig. 2D) . The lower detection of viral RNA for other 188 swabs such as the Puritan foam is likely due to the fact that these swabs absorb significant volumes of 189 liquid (Table 1) GAPDH mRNA was detected by RT-qPCR using the PBS/swab solution as direct input to RT-qPCR 199 (Fig. 2E, Fig. S2C) . Additionally, the equivalent performance of the RHINOstick to the positive control 200 demonstrates the robustness of RT-qPCR to nasal matrix. The clinical sample titers were determined 201 using an N gene standard curve (Fig. S2D, Supplementary Methods) . RHINOstick swabs were not 202 statistically distinguishable from the positive control at either titer, but the Puritan foam swabs showed 203 . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this this version posted September 25, 2020. . https://doi.org/10.1101/2020.09.24.20200386 doi: medRxiv preprint lower recovery (P < 0.001 by an independent t-test). Replicate Ct values shows the high reproducibility of 204 the qPCR data ( Fig. S2E and F) . 205 A key issue with swabs is the stability of viral particles on the swabs 206 during transport from the collection site to the test lab. To test the stability of SARS-CoV-2 on swabs 207 over time we added SARS-CoV-2 from clinical samples to swabs containing nasal matrix (Fig. 3A) . The 208 contrived samples were left wet or dry at 25°C as well as dry at 42°C, to simulate storage in a hot car or 209 truck, for up to 72 hours before elution into PBS. The presence of both SARS-CoV-2 N gene RNA and 210 GAPDH mRNA was detected by using the swab eluent as direct input into RT-qPCR ( Fig. 3 and S3) . was left out for 72 hours (Fig. 3D and S3A) . Overall, GAPDH detection was more consistent for both the 215 RHINOstick and Puritan foam swabs ( Fig. 3C and D, S3B-D) across all conditions in the time course, but 216 was slightly more variable for the Puritan foam swabs stored wet at room temperature for 72 hours (Fig 217 3E ). The variability in the N gene as well as GAPDH data collected from Puritan foam swabs during the 218 time course is also observed when comparing the Ct's between two technical replicates in the RT-qPCR 219 data ( Fig. S3E and F) . 220 Our improved AN swab is comfortable to use, allows patients to perform swabs for themselves, 223 and enables rapid accessioning and processing. The RHINOstick performs comparably to currently 224 available swabs, releasing similar amounts of human and viral material into solution after use (Fig. 2) . We 225 found that RHINOstick and Puritan foam swabs detected similar levels of GAPDH mRNA (Fig. 3) , while 226 SARS-CoV-2 was detected more consistently from the RHINOstick swab with lower titer contrived 227 samples ( Fig. 2D and E) or after long periods of storage (Fig. 3) . All RT-qPCR reactions performed in 228 . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) preprint The copyright holder for this this version posted September 25, 2020. . https://doi.org/10.1101/2020.09.24.20200386 doi: medRxiv preprint this study used direct input of swab eluant to the reaction mix without any RNA extraction and we were 229 able to detect as low as 10 molecules per assay (Fig. 2D) . 230 SARS-CoV-2 viral particles on the RHINOstick swab proved to be very stable with no 231 statistically significant loss of Ct under all the conditions tested (Fig. 3) . One of the key design elements 232 of the RHINOstick swab is the ability for a patient to self-collect their AN swab for sample processing. 233 To best use this feature, it is preferred to use dry swabs in which the swab is put into the collection tube 234 after self-collection in the absence of any buffer. This swab may then be mailed in or collected at a central 235 location without the need for concern over sample leakage in transport. The stability of SARS-CoV-2 on 236 the RHINOstick swab for up to 72 hours before processing ( Fig. 3B and C) demonstrates the feasibility of 237 the dry swab method which is consistent with other studies (12). An additional advantage of the new 238 swab design is the ability to elute the sample in a low volume of liquid (200 µL), potentially increasing 239 the sensitivity of the direct RT-qPCR method by 5-15 fold compared to standard methods. Most 240 commercial swabs cannot be used with this low elution volume, due to the high volume of liquid 241 absorbed by the swab (Table 1) . 242 We envision the RHINOstick swabs being used in the following workflow: the patient will scan 243 the side of the barcode on the side of the tube using a cellphone app, phone-accessed website, or scanner 244 and an ID card at the collection site to link the patient and sample together. After swabbing with the 245 RHINOstick swab, the patient would screw the swab into the barcoded tube. The sample would then be 246 packaged for Category B compliant transport. In an unsupervised self-collection setting, the tube could be 247 rescanned at the sample deposition site to help track sample custody. The tubes would be deposited in a 248 lockbox at the site, which would periodically be sent to the testing center. All swabs would be stored and 249 transported dry avoiding the risk of liquid leakage. In the testing facility, the samples would be received 250 and loaded into 96-well racks by hand (Fig. 1B) . Each rack of tubes would then be put onto a robot that 251 scans the 2D matrix codes on the bottom of the tubes thereby linking the sample ID to each plate and 252 plate location in seconds. After accessioning, the samples can pass to a de-capping robot which removes 253 the caps and the samples can then be eluted, inactivated, and processed for viral quantitation. 254 . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. nasal matrix on a swab. II; unused swab, without nasal matrix, was either treated with packaged synthetic 316 SARS-CoV-2 virus or left untreated (clean, unused swab). III; SARS-CoV-2 negative volunteer self-317 collected nasal matrix on a swab which was then treated with packaged synthetic SARS-CoV-2 or SARS-318 CoV-2 clinical sample (Methods). All samples were eluted in PBS and used as direct input to RT-qPCR 319 assays. Images created with BioRender.com. (C) RT-qPCR quantitation of human GAPDH mRNA from 320 used swabs containing nasal matrix (pink bars) or matched unused swabs (grey bars). (D) RT-qPCR 321 quantitation of the SARS-CoV-2 N gene from packaged synthetic virus applied to clean, unused swabs. 322 The grey bar is the negative control, PBS input into RT-qPCR. The pink line is a guideline for complete 323 recovery based on the positive control. (E) RT-qPCR quantitation of SARS-CoV-2 N gene from swabs in 324 . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) preprint The copyright holder for this this version posted September 25, 2020. . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) preprint The copyright holder for this this version posted September 25, 2020. . https://doi.org/10.1101/2020.09.24.20200386 doi: medRxiv preprint COVID-19 Map -Johns Hopkins Coronavirus Resource Center Here we demonstrate that the RHINOstick, a newly designed injection molded polypropylene 255 swab with a screw cap integrated into the swab handle, performs equal to several commonly used AN 256 swabs on the market at capturing and releasing SARS-CoV-2 viral particles from AN swabs. This 257 innovative AN swab design has the potential to expedite SARS-CoV-2 diagnostic testing while 258 significantly reducing costs. We anticipate that these swabs will be generally useful for pathogen panel 259 The grey bar is the negative control, PBS input into RT-qPCR. The pink line is a guideline for complete 367 . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) preprintThe copyright holder for this this version posted September 25, 2020. . https://doi.org/10.1101/2020.09.24.20200386 doi: medRxiv preprint