key: cord-0788241-c1ra9drg authors: Kallioinen, Noa; Hill, Andrew; Christofidis, Melany J.; Horswill, Mark S.; Watson, Marcus O. title: Quantitative systematic review: Sources of inaccuracy in manually measured adult respiratory rate data date: 2020-10-10 journal: J Adv Nurs DOI: 10.1111/jan.14584 sha: 3d3a0cad2054cab1930bce16f7ac1da5466b6a63 doc_id: 788241 cord_uid: c1ra9drg AIMS: To identify the potential sources of inaccuracy in manually measured adult respiratory rate (RR) data and quantify their effects. DESIGN: Quantitative systematic review with meta‐analyses where appropriate. DATA SOURCES: Medline, CINAHL, and Cochrane Library (from database inception to 31 July 2019). REVIEW METHODS: Studies presenting data on individual sources of inaccuracy in the manual measurement of adult RR were analysed, assessed for quality, and grouped according to the source of inaccuracy investigated. Quantitative data were extracted and synthesized and meta‐analyses performed where appropriate. RESULTS: Included studies (N = 49) identified five sources of inaccuracy. The awareness effect creates an artefactual reduction in actual RR, and observation methods involving shorter counts cause systematic underscoring. Individual RR measurements can differ substantially in either direction between observations due to inter‐ or intra‐observer variability. Value bias, where particular RRs are over‐represented (suggesting estimation), is a widespread problem. Recording omission is also widespread, with higher average rates in inpatient versus triage/admission contexts. CONCLUSION: This review demonstrates that manually measured RR data are subject to several potential sources of inaccuracy. IMPACT: RR is an important indicator of clinical deterioration and commonly included in track‐and‐trigger systems. However, the usefulness of RR data depends on the accuracy of the observations and documentation, which are subject to five potential sources of inaccuracy identified in this review. A single measurement may be affected by several factors. Hence, clinicians should interpret recorded RR data cautiously unless systems are in place to ensure its accuracy. For nurses, this includes counting rather than estimating RRs, employing 60‐s counts whenever possible, ensuring patients are unaware that their RR is being measured, and documenting the resulting value. For any given site, interventions to improve measurement should take into account the local organizational and cultural context, available resources, and the specific measurement issues that need to be addressed. Vital sign observations are a fundamental part of patient monitoring, and respiratory rate (RR), blood pressure, heart rate, temperature, and oxygen saturation are the most common metrics underlying monitoring protocols such as physiological track-and-trigger systems (Brekke et al., 2019) . These protocols can support nurses and doctors in the early detection of patient deterioration and help facilitate rapid responses to adverse clinical situations (Mohammed et al., 2009; Prytherch et al., 2006; Subbe et al., 2007) . There are several types of track-andtrigger system, which vary in complexity, but all essentially apply algorithms to routinely measured vital sign values to prompt appropriate actions, such as clinical escalation or increased monitoring frequency (Christofidis et al., 2015 (Christofidis et al., , 2016 ). An effective track-and-trigger system can also empower nurses to escalate by providing objective evidence of clinical deterioration to support and corroborate their more subjective or intuitive clinical assessments (Andrews & Waterman, 2005) . The accuracy of vital sign data, therefore, can directly impact the usefulness of these systems, as well as potentially impacting clinical judgement and decision-making more broadly. Nevertheless, there have been few systematic reviews or meta-analyses on the topic of vital sign measurement accuracy. Exceptions are Kallioinen et al.'s (2017) investigation of the sources of inaccuracy in blood pressure measurement, and Tysinger's (2015) more general examination of the accuracy of vital sign data. Tysinger's (2015) review presented an overview of inaccuracies arising at the measurement and documentation stages. However, it did not focus specifically on RR and, unlike Kallioinen et al.' s methodology, the review protocol was not designed to comprehensively collate the research evidence for all unique sources of inaccuracy. Consequently, the results were confined to six key papers, which contrasts sharply with the 328 publications included in Kallioinen et al.'s review, despite its narrower focus on a single vital sign. Results from systematic reviews on the topic of vital sign accuracy can help inform the interpretation and use of vital sign data, the design of track-and-trigger systems, and the formulation of clinical guidelines and educational interventions, potentially leading to improvements in the quality of patient care. However, to date, no published systematic review has presented a comprehensive evaluation of the sources of inaccuracy in RR data, although it is known to be both an important clinical indicator and subject to significant measurement issues (Lovett et al., 2005) . Respiratory rate has been identified as a strong indicator of patient condition. In addition to associations with mortality (Barthel et al., 2013; Bleyer et al., 2011; Ma et al., 2011; Sinnecker et al., 2014; Strauß et al., 2014; Subbe et al., 2003) , abnormal RRs are associated with cardiac arrest (Fieselmann et al., 1993; Schein et al., 1990; Subbe et al., 2003) , sequential organ failure (Kenzaka et al., 2012) , and escalation of care (Cardoso et al., 2014; Considine et al., 2009) . There is also some evidence that, when patients' vital signs are taken in hospital wards, RR is the strongest predictor of subsequent clinical deterioration (Churpek et al., 2016) . Since RR is a useful clinical indicator, it is usually incorporated into physiological track-and-trigger systems Smith et al., 2008; The ANZICS-CORE MET dose Investigators, 2012) . However, studies investigating its measurement accuracy have brought into question the trustworthiness of routinely recorded RR values (e.g. Lovett et al., 2005; Philip et al., 2015) . In the present paper, we therefore set out to fill the research gap identified in the Introduction by systematically reviewing the potential sources of inaccuracy in RR data evidenced in the literature. Despite advances in automated vital sign measurement, RR remains the vital sign most commonly measured without the use of an automated device (Ansell et al., 2014; Churpek et al., 2018) . For this reason, the present review focuses exclusively on the accuracy of manually measured RR data. Broadly speaking, the manual measurement process can be considered to be made up of two discrete stages: observation and documentation. Observation involves counting a patient's breaths, through visual inspection or auscultation, to determine their RR in breaths per minute. Best practice is considered to be a 60-s count using a watch or timing device (World Health Organisation, 1992 . Documentation refers to the recording of the patient's RR in the appropriate region of their observation chart or electronic record. Factors that may impact the trustworthiness of RR data recorded in a clinical chart or electronic system can be associated with either of these stages and both were considered in this review. The aims of the present study were to identify potential sources of inaccuracy in manually measured adult RR data and to quantify their effects. A systematic literature review was conducted using Medline and CINAHL databases (via EBSCOHost) and the Cochrane Library. Due to the broadness of the research question and the diversity of study methodologies found, meta-analyses were performed only for selected sources of inaccuracy on a case-by-case basis. Details of the studies included in the review are synthesized in tabulated form and discussed in the Results section. documentation, nursing, quality of care, systematic reviews and meta-analyses, vital signs The databases were searched from inception to 31 July 2019. A broad search strategy aiming to find all English-language publications related to inaccuracies in the measurement of adult RR data was employed. The inclusion criteria were adapted from those of Kallioinen et al. (2017) , and studies were eligible for inclusion if they contained all of the following: 1. Results from an empirical study relevant to the manual measurement of adult patients' RR in clinical settings; 2. Identification of at least one specific potential source of inaccuracy in the observation or documentation of RR; and 3. In relation to each source of inaccuracy identified, quantification of its prevalence or of its independent effect on documented RR values. The database searches and their results are summarized in Table 1 . The final CINAHL, Medline, and Cochrane Library database searches yielded a total of 7,514 results (i.e. S11 in Table 1 ). Figure 1 presents the subsequent study selection procedure. After 695 duplicate records were excluded, 6,819 unique records remained. Of these, a further 6,746 were excluded based on review of their titles or abstracts. Full texts were retrieved for the remaining 73 papers, as well as 11 further publications derived from the reference lists. After all 84 of these full texts were reviewed, 35 were excluded due to failure to meet the inclusion criteria for reasons such as an inappropriate sample (e.g. child patients), lack of quantitative data, or confounding of multiple sources of inaccuracy, such that independent effects could not be assessed. The remaining 49 studies were included. The standard quality assessment criteria for evaluating primary research papers from a variety of fields (Kmet et al., 2004) TI ("vital sign*" OR "vital parameter*" OR "respiratory rate*" OR "breathing rate*" OR "respiration rate*" OR "patient assess*" OR "observation chart*" OR "early warning") AND ("respiratory rate*" OR "breathing rate*" OR "respiration rate*") AND ("measure*" OR "error*" OR "document*" OR "record*" OR "aware*" OR "bias" OR "observ*" OR "assess*" OR "neglect*" OR "missing*" OR "inaccur*" OR "accura*") ALL: English Language 1,126 CINAHL: exclude MEDLINE records S6 ("respiratory rate" OR "respiration rate" OR "breathing rate") AND ("measurement" OR "error" OR "documentation" OR "bias") One reviewer (NK) conducted the initial analysis of the 84 potentially relevant full-text articles. For each study that met the inclusion criteria, details of the study design, observers, patients, and results were collected and tabulated, as was further information specific to the relevant source of inaccuracy. In cases where a single study investigated both manual and device-assisted measurement of RR, only data relevant to manual measurement were extracted. The quality appraisals were also conducted at this stage and the results were tabulated. The entire process of data abstraction (including quality appraisal) was then independently audited by another reviewer (MJC) and disagreements were resolved through discussion until complete agreement was reached. Findings related to each independent source of inaccuracy were tabulated separately and trends were investigated. Discussion of individual studies and trends is presented in a narrative form, while quantitative data, overall results from the quality appraisals, and further details of each study are presented in Tables 2-6. Full data from the quality appraisals can also be found in the Tables S1-S5. Due to the scarcity of comparable studies for some potential sources of inaccuracy, meta-analytic techniques were performed on a case-bycase basis for specific sources only. All such statistical analyses were F I G U R E 1 Study selection flow chart conducted in R (R Core Team, 2018) using the metafor package for meta-analysis techniques (Viechtbauer, 2010) . The 49 Thirteen empirical studies reported the quantitative effects of one or more sources of inaccuracy potentially introduced into manually measured RR data at the observation stage. Substantial effects of observation method and the awareness effect were apparent, along with a smaller effect of inter-observer variability. Nine studies investigated the impact of one or more 'usual care' methods for taking RR observations (e.g. 15 s or 30 s counts) by comparing the accuracy of data obtained using those methods versus a criterion standard (Bianchi et al., 2013; Hill et al., 2018; Hooker et al., 1989; Lovett et al., 2005; Nielsen et al., 2015; Philip et al., 2015; Rimbi et al., 2019; Takayama et al., 2019; Worster et al., 2003) (Table 2 & Table S1 ). In the studies by Philip et al. (2015) and Nielsen et al. (2015) , ob- (Hill et al., 2018; Hooker et al., 1989; Takayama et al., 2019) reported that 15 s counts significantly underestimated RR by around 2 breaths/min on average. Rimbi et al. (2019) reported a smaller mean difference for nurses whose manual observations were supported by a mobile application that they tapped each time they observed an inspiration, and which totalled the breaths automatically. Although this mean difference was statistically non-significant, the researchers nevertheless found that, for patients with severely abnormal RRs in particular, the 15 s counts led to a substantial incidence of underscoring within their track-and-trigger system, resulting in potential failures to identify at-risk patients. In two of these studies (Hill et al., 2018; Rimbi et al., 2019) , 30 s counts were also compared against the criterion-measured counts, yielding the same general pattern of results but with RR underestimated to a lesser degree. In the fifth study, Worster et al. (2003) found no significant difference between an unspecified 'standard practice' assessment and a less rigorous criterion measure based on 30 s auscultation. However, mean differences only indicate potential systematic biases, not the extent to which an individual usual care observation may vary from a criterion standard. For this, the limits of agreement can be informative. Wider limits of agreement indicate greater variation in the association between usual care and criterion standard observations. Among the nine studies, the widest reported range of 95% limits of agreement for usual care observations was −30.9-+20.0 (Philip et al., 2015) . Since the relevant criterion value was known to be 72 breaths/min, this indicates that 95% of all usual care observations were somewhere in the broad range between 41.1 and 92 breaths/min. On the other hand, the narrowest range was −0.49-+1.83 for a criterion value of 30 breaths/min (Nielsen et al., 2015) , which bucked the general trend for the limits of agreement to be wider for higher known values in a given study (Nielsen et al., 2015; Philip et al., 2015) . Philip et al. (2015) also evaluated the accuracy of "spot check" estimates (opportunities for 12 s counts without a timer) and found that the limits of agree- Two studies investigated inter-observer variability (Dinh et al., 2013; Lim et al., 2002) and/or intra-observer variability (Lim et al., 2002) in RR measurements ( Three studies reported the effect of awareness of measurement on individuals' RRs (Han et al., 1997; Hill et al., 2018; Western & Patrick, 1988 ; Table 4 and Table S3 ). Hill et al. (2018) investigated the RRs measured manually from video recordings of healthy volunteers. The other two studies employed automated devices to measure RR. Nevertheless, these studies were still regarded as meeting the cri- because individuals' actual RRs were expected to be affected by this source of inaccuracy and hence it would have an impact on RR data regardless of whether the measurements were taken automatically or manually. In the study by Western and Patrick (1988) , participants' awareness of their own respiration was heightened by asking them to count their breaths in threes, while in the studies by Hill et al. (2018) and Han et al. (1997) , participants were explicitly told that their RRs were being observed or recorded. RRs recorded during these periods were compared with those recorded without intervention. Hill et al. (2018) reported a significant decrease in mean RR (−2.13 breaths/min) when patients were aware of the measurement. Han et al. (1997) and Western and Patrick (1988) both reported significant increases in inspiratory time and Han et al. (1997) reported a significant increase in expiratory time, when awareness of breathing was heightened. In combination, the mean changes in inspiratory and expiratory time in these two studies corresponded to mean decreases in RR of −2.4 and −1.1 breaths/min respectively. Hence, all of the available evidence, regardless of whether RR was measured manually or automatically, suggests that awareness of observation causes a decrease in measured RR as an artefact of the observation process, such that it becomes less reflective of the patient's clinical condition. However, no meta-analysis was conducted on these studies due to their small numbers and diverse methodologies. Thirty-seven publications that identified sources of inaccuracy related to the documentation of RR were included. Two specific sources were apparent: value bias and recording omission. It has been suggested that particular values are often over-represented in manually measured RR data because they represent estimates rather than actual measurements ( (Table 5 and Table S4 ). Only three of these studies provided direct evidence of value bias by comparing the recorded prevalence of over-represented values with criterion-measured comparison data (Bianchi et al., 2013; Mukkamala et al., 2008; Semler et al., 2013) . In studies that lacked criterion-measured comparison data, weaker indirect evidence of value bias was presented in the form of clusters of particular values that were often seen as inherently unlikely (e.g. Cooper et al., 2013) or else compared with expected distributions and found to deviate (i.e. Badawy et al., 2017; Pedersen et al., 2018) . We conducted a meta-analysis on frequency data from all eight studies (employing an unweighted random effects model using the double arcsine transformation; Freeman & Tukey, 1950; Miller, 1978 Taken together, these studies (especially those that employed criterion-measured comparisons) provide clear support for the suggestion that there is a tendency to bias RR data by recording values that are thought to be common or 'normal' (e.g. Badawy et al., 2017; Semler et al., 2013) . However, no study reported detailed characteristics of the observers, such as experience. This impedes attempts to pinpoint the reasons for such a tendency and potential targets for interventions. Thirty-two studies were identified that reported frequencies for the omission of RRs from patient charts or electronic records. Two of the studies were large-scale audits ( However, the 13 studies that reported smaller-scale general audits of hospital inpatients' vital sign documentation yielded omission rates ranging from 9.4-100% (excluding rates obtained after interventions; Table 6 and Table S5 ). Among these, five studies employed 12, 24, or 48 hr audit periods (Cretikos et al., 2007; Hall et al., 2003; McBride et al., 2005; Odell et al., 2007; Van Leuvan & Mitchell, 2008) , four used longer audit periods of at least 1 week (Cahill et al., 2011; Edwards & Murdin, 2001; Helliwell et al., 2002; McGain et al., 2008) , and three examined the entirety of patients' stays in the ward (Rosen et al., 2015; Smith & Oakey, 2006) or the emergency department (Parkes, 2011) . In addition, a single point prevalence study examined Table 7 & Table S6 ). These included adverse events in general (Chen et al., 2009; Cretikos et al., 2007; MERIT study investigators, 2005) , cardiac and/or respiratory arrest (Hodgetts et al., 2002; Kenward et al., 2001; Nurmi et al., 2005) , admission of inpatients to the intensive care unit (Goldhill et al., 1999; Jonsson et al., 2011) , or clinical concern about potential deterioration requiring overnight medical review (Gordon & Beckett, 2011) . With the exception of one study that only reported omissions made in the 15 min immediately preceding an event (MERIT study investigators, 2005) , most investigated the shift or 24-hr period preceding an event (sometimes excluding the final 15 min). A meta-analysis was also conducted on the proportion of omissions reported in these studies. The meta-analysis yielded a predicted proportion of 47.8% [95% CI: 26.8%, 69.2%; Prediction Interval: 0%, 99.5%, τ 2 = 0.122] (Figure 3b) . Notably, a further study that focussed on the emergency department treatment of asthma attacks found a substantial omission rate of 72.9% at the severity evaluation stage (Linares et al., 2006) . RRs from observations taken during triage or hospital admission (Bergrath et al., 2011; Considine et al., 2006; Cooper et al., 2013; Crandon et al., 2008; Gerdtz et al., 2013; O'Reilly et al., 2012) , or immediately after admission (Armstrong et al., 2008; Keene et al., 2017) . Omission rates ranged from 0.8%-81.5% (excluding any postintervention rates; Table 8 and Table S7 ). Data from these studies were included in a third meta-analysis of proportions, which yielded a predicted proportion of omissions of 21.6% [95% CI: 6.2%, 42.7%; Prediction Interval: 0.0%, 84.7%; τ 2 = 0.108] (see Figure 3c ). For studies that provide separate data on multiple systems/charts, subgroups or timepoints, the bracketed numbers in this column can be used to cross-reference Figure 3a . b This study was not included in the meta-analysis (see Results section). For studies that provide separate data on multiple systems/charts, subgroups or timepoints, the bracketed numbers in this column can be used to cross-reference Figure 3b . b This study was not included in the meta-analysis (see Results section). TA B L E 8 Studies reporting the percentage of respiratory rate measurements omitted from records during triage or admission to hospital, or immediately after admission criterion-measured comparison data consistently found that over half of all recorded RR values were potentially the product of bias and all studies showed that the two most common values appeared to be suspiciously prevalent (Table 5 and Figure 2 ). Finally, widespread evidence of recording omission was also revealed (Table 6) . Meta-analyses showed that omission rates tended to be substantially higher among hospital-based general audits of inpatient observations (58.1%) and patient records completed prior to inpatient adverse events or clinical deterioration (47.8%), compared with data collected during triage, admission to hospital, or immediately after admission (21.6%; Figure 3 ). In comparison with value bias, however, there was also substantially more variation between study sites. In interpreting the findings of this review, it is worth noting that some of the identified sources of inaccuracy could potentially have cumulative effects on individual RR observations. For example, if a 15-s count is conducted on a patient who is aware that their respiration is being observed, then the combination of an artefactual reduction in their actual RR due to the awareness effect and underscoring due to the observation method could potentially create a substantially misleading impression of the patient's clinical condition (Hill et al., 2018) . This issue could also be exacerbated by interor intra-observer variability. Furthermore, the prevalence of value bias may call into question the trustworthiness of all recorded RR data at affected sites, while recording omission has a further impact on the utility of RR as a clinical indicator. Consequently, hospitals that rely on the manual measurement of RR may need to address all of these issues to maximize the clinical utility of RR data, including its use in the context of track-and-trigger scoring systems designed to facilitate the early recognition of deteriorating patients. In terms of nursing practice, the results of this review emphasize the value of conducting 60-s counts whenever it is practicable to do so and of ensuring that patients are unaware that their RR is being measured. They also highlight that recording omission is a widespread and important problem, and the need to count RR rather than simply estimating it. Hence, it is vital that nursing education addresses all of these points, as well as instilling an appropriate counting technique to minimize unwanted variability in RR data. Nevertheless, despite being necessary, training alone may not be sufficient to eliminate the measurement issues identified in this review if their underlying causes are not considered and addressed. For studies that provide separate data on multiple systems/charts, subgroups or timepoints, the bracketed numbers in this column can be used to cross-reference Figure 3c. improvements unless other elements of the surrounding system are optimized first (Russ et al., 2013) . Hence, knowledge of the underlying mechanisms for the sources of inaccuracy identified in this review may allow for better targeted and more effective interventions. A particular challenge in determining the underlying causes of these issues is the marked lack of reported information about the individuals whose observations were studied (e.g. nurses). Authors typically reported sufficient information on patient participants (when applicable), yet this was rarely the case for observers, even when observation method or inter-/intra-observer variability was the main focus of the study. While there may be pragmatic obstacles to its inclusion, particularly for large-scale chart-review studies, detailed observer information would be a useful addition to future primary research publications on these topics. There is some suggestion in the literature that many instances of value bias may reflect estimates-rather than measurements-of RR (Badawy et al., 2017; Keene et al., 2017; Semler et al., 2013) . Similarly, some instances of recording omission may also represent failure to measure RR in the first place. Recent studies have pointed to a range of potential underlying causes for the general neglect of RR measurement in hospital care (Ansell et al., 2014; Elliott, 2016; Hogan, 2006) . These include: perceived or actual lack of time; lack of knowledge and training; lack of automated measurement; and the perceived unimportance of RR (despite actually being a strong predictor of clinical deterioration; Churpek et al., 2016) . Similarly, time constraints may cause clinicians who do conduct formal RR counts to favour shorter count durations (e.g. 15 s). This may cause RR to be underestimated, as demonstrated in the present review. In addition, neglect of vital sign monitoring more generally has been found to be influenced by shift length, with longer shifts leading to less frequent monitoring (Dall'Ora et al., 2019) and therefore more recording omissions. Clearly, some of these causes are rooted in systemic and cultural factors and consequently cannot be remedied through training alone. A range of options are available to hospitals seeking to improve the accuracy of their RR data. In this section, we discuss these in light of our results. However, it is important to appreciate that there is no 'one size fits all' solution. Rather, the best solution-or combination of solutions-for a particular site will depend on the organizational and cultural context, the available resources, and the extent to which particular measurement issues prevail locally. For example, it is clear from the review that, although recording omission is a serious and widespread problem and it is generally more prevalent in inpatient wards, some individual ward sites have very low omission rates (and vice versa for emergency departments). It is also important to consider any unintended consequences for patient care of any proposed intervention or change. For sites with substantial and pervasive issues around the manual measurement or documentation of RR, one potential option that addresses many of the possible underlying causes outlined above is to transition to a fully automated measurement method. In theory, the right technology could eliminate all sources of error discussed in this review (provided it is also interoperable with an electronic record system). Even the awareness effect could potentially be avoided if it is not readily apparent when measurements are occurring, or if continuous monitoring allows the patient to habituate to measurement. Unfortunately, current mainstream methods for automated RR measurement (i.e. spirometry, capnometry, and impedance pneumography) are not practical for the mass routine monitoring of inpatients due to a range of factors, such as patient discomfort (e.g. capnometry), interference to natural breathing (e.g. spirometry), and resource intensiveness (all methods; Liu et al., 2019) . However, RR measurement is ripe for digital disruption and over a dozen alternative technological methods have already been proposed (for a recent review, see Liu et al., 2019) . Several of these can potentially be incorporated into clothing or other body-worn devices and some do not even require physical contact with the patient (Liu et al., 2019) . Although many of these technologies are in the early stages of development and only a couple can be implemented at relatively low cost (Liu et al., 2019) , progress to date suggests that unobtrusive automated monitoring of all ward patients' RRs will be a viable option for some sites in the relatively near future. As well as directly addressing the RR measurement issues outlined in this review, such technologies may also have additional indirect benefits if they alleviate some of the time pressure experienced by nursing staff in busy hospital environments, freeing up more time for high-quality nurse-patient interactions. These interactions, in turn, may lead to the identification of additional care needs (Cardona-Morrell et al., 2016) , improved recovery times (Castillo & Sánchez-Sosa, 2002) , and greater patient satisfaction (Evans, 2016) . However, even when such technologies become widely available, they will not necessarily be a panacea for sites where monitoring is currently suboptimal. First, although these devices may be affordable for some hospitals, they are likely to be prohibitively expensive for many, particularly in the developing world. Second, there may be a risk that nurse-patient contact actually decreases if none of the patients' vital signs need to be measured manually and time pressures continue to increase. Finally, there is a risk of deskilling if clinicians are no longer well-practiced at manual measurement. This could potentially place deteriorating patients at elevated risk if and when technologies fail. For example, some clinical information systems have proven vulnerable to prolonged network failures (Berinato, 2003; Flinders, 2015) or ransomware attacks (Collier, 2017) . Indeed, the catastrophic failure of most or all computers, mobile phones, and other sensitive electronic equipment-as well as sustained power outages-are very real possibilities if a solar superstorm on the scale of the 1859 Carrington Event were to strike the modern world (Jonas, 2015) . As unlikely as this may sound to some readers, the current (at time of writing) COVID-19 pandemic illustrates the need for health systems to be resilient to low-probability, high-impact events such as this. Hence, technological solutions do not eliminate the need for clinicians to acquire and maintain skills in the manual measurement of vital signs, including RR. Various devices have also been developed to aid, rather than replace, the manual measurement of RR. Some are relatively lowtechnology devices, such as the World Health Organization's ARI Timer and ARI Timer MK2 (Gan et al., 2015; WHO &WHO, 1992 . These devices are simple timers that indicate when 60 s have elapsed. The target users are primarily community healthcare workers in developing countries, and there has not been widespread adoption in hospitals. More recently, however, more sophisticated mobile phone applications have been developed. Most of these use the 'one-press-per-breath' (OPB) concept, where an observer presses a button on the phone each time the patient takes a breath. The application, rather than the observer, keeps track of the breath count. Notably, some of these applications, such as RRate , do not require an entire 60-s count. Instead, the RR is derived from the average amount of time between breaths . There is evidence that this method can potentially allow for adequately accurate RR measurement in a fraction of the 60 s recommended for manual measurement (Gan et al., 2015) . Further, in a study by Black et al. (2015) , four mobile phone applications were compared: three that employed OPB (of various lengths or number of breaths) and a simple 60-s timer. A 20-breath OPB and a 60-s OPB produced the most accurate measurements and both out-performed the 60-s timer. Because mobile applications can be used on existing Android or iOS personal devices, there is a smaller barrier to adoption than with special devices like the ARI Timer, at least in the developed world. Importantly, the benefit of accurate measurement in shorter counting periods may also overcome one of Elliot's (2016) reported reasons for the neglect of RR, namely lack of time. Hence, the use of such applications may reduce the prevalence of recording omissions and value bias, as well as reducing inter-and intra-observer variability. Nevertheless, the awareness effect remains a potential issue. Compared with taking a simple count unobtrusively in one's head, tapping each time the patient breathes may actually draw more attention to the fact that RR is being assessed. Hence, future research should investigate whether the use of OPB applications increases the awareness effect. As with fully automated systems, the potential for technological failures also means that manual measurement skills cannot be replaced entirely. The introduction of paper-based or electronic track-and-trigger systems may also have an impact on the measurement of RR. For example, as well as providing baseline data on recording omission rates, three studies included in this review also investigated whether introducing a track-and-trigger system would reduce them (Cahill et al., 2011; McBride et al., 2005; Odell et al., 2007; Table 6 ). In all three studies, the frequency of missing RR values was shown to decrease, falling by up to 78% (Odell et al., 2007) . Furthermore, in a simulation study by Ludikhuize et al. (2011) , hospital nurses were more likely to seek out the RRs of mock patients if they had prior experience using a track-and-trigger system in their ward. This may indicate that the importance of RR as a clinical indicator is better understood when observers have been exposed to track-and-trigger systems. However, in each of these studies, it is unclear to what extent the results were attributable to the track-and-trigger system itself, or to the training that accompanied its introduction. Although a track-and-trigger system (especially an electronic system) may provide a forcing function for the recording of vital signs (Hogan et al., 2019) , including RR-thus potentially reducing recording omissions-there is no guarantee that any manually measured RR values documented therein are accurate. Unfortunately, none of the studies of recording omission discussed above assessed the accuracy of the values that were recorded against criterion-measured data. However, the general point is illustrated by Cooper et al.'s (2013) study, which yielded one of the lowest omission rates (2.1%) but also one of the highest combined percentages for the two most common commonly recorded values (82.76%). If nursing staff are time-poor, an unintended consequence of using a track-and-trigger system may be that there is a perverse incentive to estimate RR, resulting in an increased incidence of value bias. This may potentially undermine the track-and-trigger system itself by causing RR to be scored incorrectly (Badawy et al., 2017) . A further five of the studies that provided baseline recording omission data also investigated the impact of alternative interventions involving changes to clinical work systems (Chen et al., 2009; Considine et al., 2006; Gerdtz et al., 2013; Kenward et al., 2001; Rosen et al., 2015; Table 6 ). However, none investigated whether reduced omission rates were accompanied by a concomitant decrease in accuracy or an increase in the prevalence of value bias. Two of these studies examined the impact of introducing a Medical Emergency Team (MET), yielding mixed results for recording omissions. The study by Chen et al. (2009) produced weak evidence of slightly lower omission rates at sites where a MET had been introduced. In contrast, Kenward et al. (2001) found a substantial reduction in recording omissions after the introduction of a MET; however, they attributed it primarily to the accompanying education around the early detection of clinical deterioration. One study investigated the effect of redesigning an electronic triage interface to better align with the clinical decision-making process and to make data entry less effortful (Gerdtz et al., 2013) . Documentation of RR more than doubled following the interface change and further increased after triage education sessions were conducted that emphasized the importance of vital signs for recognizing ill health and deterioration at triage. Another study investigated the effect of an intervention comprising the introduction of written nursing practice standards for the initial assessment of emergency department patients with supporting education, but found an unexplained increase in RR recording omissions (Considine et al., 2006) . Finally, one study employed a Failure Mode and Effects Analysis process, where the researchers worked with frontline staff to: (a) identify current and potential risks to the completion of vital sign monitoring and documentation in their workplace; (b) articulate the potential effects of these risks; and (c) identify potential solutions appropriate to that specific site (Rosen et al., 2015) . The proposed solutions were then presented to the facility's leadership, and RR recording omissions decreased significantly in the following 2 months. Presently, sources such as nursing textbooks provide conflicting advice about the duration of manual RR counts (Hill et al., 2018) . However, it is clear from the observation method results that 60-s counts should be preferred. Nevertheless, enshrining this advice in clinical training will not ensure that it is adhered to if issues such as time pressure and interruptions make it impractical to do so, in which case a technologybased solution may be preferable. Similarly, in the context of manual measurement, inter-and intra-observer variability can be addressed via training in the appropriate protocols, but it may not always be possible to enact these protocols consistently given competing pressures. Of all five sources of inaccuracy, the awareness effect is probably the most amenable to a training-based solution. Indeed, nursing texts are relatively consistent in advising that patients should not be made aware that their RR is being measured (Hill et al., 2018) . It is also clear from some of the studies discussed above that education may provide a useful adjunct to interventions that involve improvements to clinical work systems, such as those aimed at reducing recording omissions. In particular, there was some evidence that there is value in training staff not just in protocols for the accurate measurement of vital signs, but also their central importance to the detection of clinical deterioration (e.g. Gerdtz et al., 2013; Kenward et al., 2001) . Similarly, another of the included studies showed that education provided by critical care outreach teams reduced the incidence of recording omissions on wards already using a track-and-trigger scoring system (Hall et al., 2003; Table 6 ). Finally, a further potentially fruitful use of training is for familiarizing staff with new technologies (Russ et al., 2013) , such as automated measurement devices or measurement aids. This systematic review is the first to focus on identifying and quantifying potential sources of inaccuracy in manually measured RR data. The integrity of the review was ensured by adhering to the PRISMA guidelines and the use of two reviewers. In addition, all included papers were subjected to a standardized quality appraisal, the full results of which have been included in the supplementary materials for transparency. The identification of sources of inaccuracy is an important first step towards improving measurement quality, with the potential to inform interventions and further investigations. Nevertheless, existing research on the topic of RR measurement accuracy is notably sparse. While there were a total of 49 studies identified, 32 of these related to a single source of inaccuracy, namely recording omission. The remaining sources of inaccuracy were only investigated in two to nine studies each. Thus, the sources of inaccuracy presented in this review should be taken primarily as a representation of the available literature and not an exhaustive list of potentially influential factors. In addition, it should be noted that all studies of the awareness effect were conducted with healthy volunteers. Hence, it would be particularly valuable for future research to replicate these findings with clinical populations to clarify the extent to which they will generalize, given that inpatients are likely to have more abnormal RR values and more irregular breathing patterns than healthy participants. RR is an important clinical indicator, but this review has identified five sources of inaccuracy related to its observation or documentation. When not subject to recording omission, manually measured RR data may still be inaccurate due to value bias, the observation method used, inter/intra-observer variability, or the awareness effect. In some cases, a single RR measurement may even be affected by several of these factors. Hence, clinicians should interpret recorded RR data cautiously unless systems are in place to ensure its accuracy. For nurses, this includes ensuring that patients are unaware that their RR is being measured, counting rather than estimating RRs, employing 60-s counts whenever possible, and documenting the resulting value. Future research should investigate the underlying causes of the sources of inaccuracy identified in this review, other potential sources that have not previously been investigated, and methods to improve observation and documentation quality. For any given site, the solutions employed should be tailored to take into account the local organizational and cultural context, the available resources, and the specific measurement issues that need to be addressed. The fifth author has one relevant patent issued (methods and means of physiological monitoring), but none of the articles cited in this review relate directly to it. All other authors declare no conflict of interest. All authors have agreed on the final version and meet at least one of the following criteria (recommended by the ICMJE (http://www.icmje. org/recom menda tions/)): (1) substantial contributions to conception and design, acquisition of data, or analysis and interpretation of data; (2) drafting the article or revising it critically for important intellectual content. https://orcid.org/0000-0002-9111-3240 Packaging: A grounded theory of how to report physiological deterioration effectively Why don't nurses consistently take patient respiratory rates? Recording of vital signs in a district general hospital emergency department Is everyone really breathing 20 times a minute? Assessing epidemiology and variation in recorded respiratory rate in hospitalised adults Respiratory rate predicts outcome after acute myocardial infarction: A prospective cohort study Is paper-based documentation in an emergency medical service adequate for retrospective scientific analysis? An evaluation of a physician-run service All systems down Revitalizing a vital sign: Improving detection of tachypnea at primary triage Can simple mobile phone applications provide reliable counts of respiratory rates in sick infants and children? An initial evaluation of three new applications Longitudinal analysis of one million vital signs in patients in an academic medical center The value of vital sign trends in predicting and monitoring clinical deterioration: A systematic review Introduction of a new observation chart and education programme is associated with higher rates of vital-sign ascertainment in hospital wards Vital signs monitoring and nurse-patient interaction: A qualitative observational study of hospital practice Respiratory rate at intensive care unit discharge after liver transplant is an independent risk factor for intensive care unit readmission within the same hospital stay: A nested case-control study Well-being and medical recovery in the Critical Care Unit: The role of the nurse-patient interaction The impact of introducing medical emergency team system on the documentations of vital signs Less is more: The design of early-warning scoring systems affects the speed and accuracy of scoring Observation chart design features affect the detection of patient deterioration: A systematic experimental evaluation Accuracy comparisons between manual and automated respiratory rate for detecting clinical deterioration in ward patients NHS ransomware attack spreads worldwide Can written nursing practice standards improve documentation of initial assessment of ED patients? Predictors of critical care admission in emergency department patients triaged as low to moderate urgency Respiratory rate records: The repeated rate Inter-hospital transfer of trauma patients in a developing country: A prospective descriptive study The objective medical emergency team activation criteria: A case-control study Nurses' 12-hour shifts and missed or delayed vital signs observations on hospital wards: Retrospective observational study Level of agreement between prehospital and emergency department vital signs in trauma patients Respiratory rate-an under-documented clinical assessment Why is respiratory rate the neglected vital sign? A narrative review. International Archives of Nursing and Health Care Exploring the nuances of nurse-patient interaction through concept analysis: Impact on patient satisfaction Respiratory rate predicts cardiopulmonary arrest for internal medicine inpatients Network failure crashed frontline services at London hospital Transformations related to the angular and the square root The performance of a mobile phone respiratory rate counter compared to the WHO ARI timer Systematic review and evaluation of physiological track and trigger warning systems for identifying at-risk patients on the ward Evaluation of a multifaceted intervention on documentation of vital signs at triage: A before-and-after study Physiological values and procedures in the 24 h before ICU admission from the ward Significant deficiencies in the overnight use of a Standardised Early Warning Scoring system in a teaching hospital Respiratory rates measured by a standardised clinical approach, ward staff and a wireless device Choosing between random effects models in meta-analysis: Units of analysis and the generalizability of obtained results Waiting to exhale: Critical Care Outreach and recording of ventilatory frequency Influence of awareness of the recording of breathing on respiratory pattern in healthy humans Documentation of respiratory rate for acutely sick hospital inpatients -An observational study The effects of awareness and count duration on adult respiratory rate measurements: An experimental study Incidence, location and reasons for avoidable in-hospital cardiac arrest in a district general hospital A qualitative study of staff perspectives on measures to prevent patient deterioration Why don't nurses monitor the respiratory rates of patients? Respiratory rates in emergency department patients Solar flares Nursing documentation prior to emergency admissions to the intensive care unit Sources of inaccuracy in the measurement of adult patients' resting blood pressure in clinical settings: A systematic review Efficiency of respiratory rate measurements: Comment on Black et al The effect of the quality of vital sign recording on clinical decision making in a regional acute care trauma ward Time to put the R back in TPR Importance of vital signs to the early diagnosis and severity of sepsis: Association between vital signs and sequential organ failure assessment score in patients with sepsis Standard quality assessment criteria for evaluating Alberta Heritage Foundation for Medical Research Respiratory rate measurement in adults-how reliable is it? Respiratory Medicine Medical audit on asthma in an emergency department Recent development of respiratory rate measurement technologies The vexatious vital: Neither clinical measurements by nurses nor an electronic monitor provides accurate measurements of respiratory rate in triage Measuring adherence among nurses one year after training in applying the Modified Early Warning Score and Situation-Background-Assessment-Recommendation instruments Predictors of in-hospital mortality of older patients admitted for community-acquired pneumonia Long-term effect of introducing an early warning score on respiratory rate charting on general wards Introduction of the medical emergency team (MET) system: A cluster-randomised controlled trial The inverse of the Freeman-Tukey double arcsine transformation Improving accuracy and efficacy of early-warning scores in acute care R = 20: Bias in the reporting of respiratory rates Inter-observer agreement in measuring respiratory rate Observations and warning signs prior to cardiac arrest. Should a medical emergency team intervene earlier? The effect of a critical care outreach service and an early warning scoring system on respiratory rate recording on the general wards Rate of respiration: The forgotten vital sign A critical assessment of early warning score records in 168,000 patients The accuracy of respiratory rate assessment by doctors in a London teaching hospital: A cross-sectional study Calculating early-warning scores -A classroom comparison of pen and paper and hand-held computer methods R: A language and environment for statistical computing (Version 3.5.2) Differences in prehospital patient Assessments for pediatric versus adult patients Respiratory rates observed over 15 and 30 seconds compared with rates measured over 60 seconds. Practice-based evidence from an observational study of acutely ill adult medical patients during hospital admission Engaging staff to improve quality and safety in an austere medical environment: A case-control study in two Sierra Leonean hospitals The science of human factors: Separating fact from fiction Clinical antecedents to in-hospital cardiopulmonary arrest Flash mob research: A single-day, multicenter, resident-directed study of respiratory rate Assessment of mean respiratory rate from ECG recordings for risk stratification after myocardial infarction Incidence and significance of errors in a patient "track and trigger" system during an epidemic of Legionnaires' disease: Retrospective casenote analysis The prognostic significance of respiratory rate in patients With pneumonia Effect of introducing the Modified Early Warning Score on clinical outcomes, cardio-pulmonary arrests and intensive care utilisation in acute medical admissions Reproducibility of physiological track-and-trigger warning systems for identifying at-risk patients on the ward Hospital overnight and evaluation of systems and timelines study: A point prevalence study of practice in Australia and New Zealand A comparison of methods to count breathing frequency Rapid Response Team composition, resourcing and calling criteria in Australia A framework for the meta-analysis of Bland-Altman studies based on a limits of agreement approach How vital are vital signs? A systematic review of vital sign compliance and accuracy in nursing Missed opportunities? An observational study of vital sign measurements Conducting meta-analyses in R with the metafor package Effects of focusing attention on breathing with and without apparatus on the face Programme for the control of acute respiratory infections: Fifth programme report, 1990-1991. Switzerland. World Health Organisation Reliability of vital signs measured at triage Quantitative systematic review: Sources of inaccuracy in manually measured adult respiratory rate data JAN) is an international, peer-reviewed, scientific journal. JAN contributes to the advancement of evidence-based nursing, midwifery and health care by disseminating high quality research and scholarship of contemporary relevance and with potential to advance knowledge for practice, education, management or policy Online Library website: www.wileyonlinelibrary.com/journal/jan Reasons to publish your work in JAN: • High-impact forum: the world's most cited nursing journal, with an Impact Factor of 1.998 -ranked 12/114 in the • Most read nursing journal in the world: over 3 million articles downloaded online per year and accessible in over 10,000 libraries worldwide Positive publishing experience: rapid double-blind peer review with constructive feedback • Rapid online publication in five weeks: average time from final manuscript arriving in production to online publication Online Open: the option to pay to make your article freely and openly accessible to non-subscribers upon publication on Wiley Online Library, as well as the option to deposit the article in your own or your funding agency's preferred archive