key: cord-0787655-mqtjgupy authors: Esmizadeh, Elnaz; Chang, Boon Peng; Jubinville, Dylan; Seto, Curtis; Ojogbo, Ewomazino; Tzoganakis, Costas; Mekonnen, Tizazu H. title: Stability of nitrile and vinyl latex gloves under repeated disinfection cycles date: 2021-04-13 journal: nan DOI: 10.1016/j.mtsust.2021.100067 sha: d911089f9562e435a964ec67b59a02602b3a60f7 doc_id: 787655 cord_uid: mqtjgupy As SARS-CoV-2 transmission by fomites is one of the main concerns of coronavirus disease (COVID-19), the World Health Organisation (WHO) advised on the use of protective gloves for handling contaminated surfaces and fomites. The shortage in the supply of personal protective equipment (PPE) due to the surging demand in conjuncture with the disposal of an unprecedented quantity of contaminated PPEs into the landfill, led to an interest for alternative platforms for the management of PPEs. In this study, we evaluated the potential of re-using gloves after repeated disinfection cycles using six readily available and common sterilization methods: UV, ethanol, heat, steam, bleach, and Quats (Quaternary ammonium compounds) for the inactivation of SARS-CoV-2. For this, two commercially available medical grade gloves, i.e. nitrile and vinyl (polyvinyl chloride) gloves were tested. Both types of gloves showed deterioration in mechanical and thermal performance with the use of quats as sterilization treatment while no remarkable change in properties was observed up to 20 cycles of disinfection for the other sterilization methods. The exceptions were that the vinyl and nitrile gloves did not tolerate steam/dry heat and UV treatment over 10 cycles due to likely dehydrochlorination and thermal degradation, respectively. Subsequent rounds of sterilization caused no significant change in the glass transition temperature (T g ) of either medical gloves, however quats caused a slight reduction in T g due to its plasticizing effect. Overall, the physical sterilization treatments including steam, dry heat, and UV allowed the gloves to retain their thermomechanical performance up to ten cycles of sterilization. On March 11, 2020, the World Health Organization designated the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) disease, COVID-19, a global pandemic [1, 2] . The continuous growth in the number of cases worldwide has induced an unprecedented shortage of personal protective equipment (PPE) such as masks, gloves, gowns, and aprons for front-line health-care workers and the general public in domestic situations [3] [4] [5] . The PPE shortage can also impact other industries that rely on the use of PPE for their regular operation, such as manufacturing, research laboratories, petrochemical industries, food production, etc. On the other hand, a rapid accumulation of potentially infectious PPE in domestic solid waste streams, landfills, and water bodies is already detected [6] . To address this challenge, triggered by surging demand and supply chain issues and waste disposal concerns, urgent solutions are desired that can include efforts to conserve PPEs use [7] through the extended use or safe re-use via effective disinfection protocols. Recently, it was revealed that the filtration efficiency preservation of N95 respirators can be affected by sterilization processes [8, 9] . Bioaerosol contact (e.g. respiratory expelled droplets) [10] , as well as indirect physical contact by touching the contaminated surface (e.g. fomites) [11] are the main potential disease transmission pathways that introduce pathogens in the respiratory tract of the host [12] . While populations generally protect themselves from visible droplet transmissions, a fomite-related J o u r n a l P r e -p r o o f transmission has been overlooked both in hospitals and communities [13] . Fomites play a key role among healthcare workers that may find themselves exposed to contact with contaminated surfaces from a large number of undiagnosed but infected patients [14] . Recent studies on surface persistence of SARS-CoV-2 showed that fomites made of stainless-steel, polymers, and cardboard paper can transmit infection for extended periods (i.e. viable virus can be detected up to 72, 48, and 24 hours after application to these surfaces, respectively) [15, 16] . This highlights the elevated importance of gloving as part of the protection process to reduce the risk of viral transmission [17] . To prevent any accidental risk, the protective gloves need to be made of smooth materials with a high tear strength and tight fit to hands to be effective [18] . Traditionally, latex gloves are employed for single or short-term use in the medical, manufacturing, or domestic environments and discarded. The disinfection and reuse of gloves are not typically practiced because of the concern that the repeated exposure of gloves to various chemical or physical disinfection processes could compromise their barrier properties against viruses or other pathogens. Scientific consensus on the best practice ensuring the retention of gloves' quality after sterilization cycles is scarce [19] [20] [21] . In our previous study, we tested methods that may be suitable for the reuse or extended use of medical-grade protective gloves so that no change in the chemical structure of the gloves was assured without compromising the barrier properties [22] . Here, we have taken this work one step further to study the retention of mechanical and thermal properties after repeated sterilization cycles to avoid the risk of infection due to sudden failure. An understanding of the property retention may have profound impacts on risk-avoidant decision-making for implementation of appropriate disinfection protocols where the reuse of the gloves is practiced. This may also be of interest to the PPE producing industries to design stable medical gloves with extended lifespan. Two powder-free medical examination gloves, vinyl-based (Pro-Medix, Canada) and nitrilebased (The Safety Zone, USA), were chosen to investigate property retention after repeated sterilization cycles [23] . While SARS-CoV-2 virus has been found to persist on gloves for a number of days [15, 24] , it can be rapidly inactivated by surface sterilization procedures recommended by the CDC guidelines [25] . Here, two categories of readily deployed clinical sterilization methods, 3 physical and 3 chemical procedures, were employed for inactivation of SARS-CoV-2 on contaminated medical-grade gloves as according to the guidelines [25] . The former includes ultraviolet (UV) sterilizer cabinet (15 min) [26, 27] , dry heat (85 °C, 30 min) [24, 26] , steam (10 min) [24] and the latter includes commercial grade 75% alcohol spray (Purple Frog Products, Canada) [9, 24] , 2% household bleach solution spray (Clorox®, USA) [24] , and quaternary ammonium compounds (0.15% benzalkonium chloride/0.15 Alkyl ethylbenzyl ammonium chloride) spray (Lloyds Laboratories Inc, Canada) [24] . Scheme 1 presented the six disinfection methods used in this study i.e. 1) UV radiation, 2) dry heat treatment, 3) steam treatment, 4) alcohol, 5) chlorine compounds, and 6) quaternary ammonium compounds. The sample codes for sterilized gloves were defined according to the glove type, sterilization cycle, and sterilization method, e.g., the nitrile gloves sterilized for 10 cycles by steam were defined as Nitrile-10-steam. The effect of sterilization repetition was evaluated for 1, 5, 10, and 20 cycles. J o u r n a l P r e -p r o o f Scheme 1. Six disinfection methods that applied to disinfect the medical gloves in this study for re-use purpose. Rubber vulcanizate swelling in toluene was employed to estimate the changes in crosslink density of nitrile gloves. Samples weighing approximately 0.02 g were cut from each glove specimen and immersed in toluene at room temperature until equilibrium swelling was attained for all the samples (72 h). The weight of each sample was recorded every 24 h by blotting the excess surface toluene with wipes and immediately weighing the sample. The volume fraction of the rubber, , in the swollen network was calculated using Equation 1 [28, 29] : Where is crosslink density in mol per unit volume ( ⁄ ), is volume fraction of rubber in equilibrium swollen vulcanized rubber sample, is molar volume of used solvent at room temperature in ⁄ and ( ) is Flory-Huggins polymer-solvent interaction parameter. The plasticizer migration rate was estimated according to ASTM D552. Dried PVC films weighing approximately 0.02 g were immersed in n-hexane at 50 o C for 2 h. The PVC films were dried and reweighed and the migration rate was calculated according to Equation 3 . where, , and are the sample weight before and after immersion in the solvent, respectively. An AGS-X Shimadzu universal testing machine with a load cell of 0.5 kN was used for the tensile testing of the glove samples before and after treatment. Prior to testing, all samples were conditioned at room temperature and 50% relative humidity in accordance with ASTM D618. The tensile testing was carried out in accordance with ASTM D882-18. Samples with dimensions of 1×7 cm (width × length) were prepared and tested at a strain rate of 50 mm/min. The tensile strength, modulus at 100% strain, toughness, and elongation at break values were recorded after which their mean values and standard deviations (SD) from five specimens of each treatment group were reported. The surface morphology of the different glove types and treatments were investigated using an Olympus BX53M polarized light microscopy (Melville, NY, USA). The optical microscope equipped with a 50× objective lens and a polarized light filter was used. All samples were prepared by cutting rectangular pieces (10 × 50 mm) out of the glove specimens. TGA analysis of the samples was carried out on a TA instrument Q500 analyzer both under nitrogen and air atmosphere with a 40 ml/min flow in a dynamic non-isothermal mode at a heating rate of 20 °C/min from ambient temperature to 700 °C. The loss factor (tan δ) of the gloves were analyzed using a Dynamic Mechanical Thermal Analyzer (DMA) model Q800 from TA Instruments, USA to investigate the change in the glass transition temperature of the gloves as a result of the cyclic sterilization processes. The glove test specimens from each treatment groups were prepared in a rectangular shape with dimensions ~25 mm × 8.5 mm × 0.06 mm (length × width × thickness). The test was conducted in a tensile mode, at a fixed frequency of 1 Hz, strain amplitude of 10 µm, and an applied force of 0.01 N. The samples were equilibrated at -75 °C (using liquid nitrogen) and heated at a rate of 3 °C/min to a final temperature of 50 °C to acquire the Tan δ curves. In order to determine the normal distribution of the replicates (per cycle per glove and per treatment), a Ryan-Joiner (RJ) normality test was performed on the crosslinking density and mechanical property test results. After which, a one-way analysis of variance (ANOVA) test with Tukey error of 5% for comparison was conducted to evaluate significant difference between the means for the different treatments from cycle 0 to cycle 20 for both the nitrile and vinyl gloves. The difference in means was considered to be statistically significant when p <0.05. To evaluate whether the disinfection techniques imparted any significant odour to the gloves, protocols from odor test standards were adapted and conducted herein where a human testing panel of five participants were used to rank the disinfected gloves samples. Disinfected gloves (20 cycles) were placed in a 35 mm diameter cylindrical glass jars, covered with aluminum foil to minimize visual bias, and sealed with lids for a period of 24 hours. Untreated nitrile and vinyl gloves were also placed in the jars as a control samples. Five panelists were recruited, whereby each panelist was required to quickly unseal the cups and smell the contents provide ranking and seal back. A ranking system relative to the control samples odor tests were conducted in accordance with ASTM E2609. An adjusted odor level range was employed here between -2 to 2 was employed here in order to capture feedback that found the odor that perform better than the reference. An odor level of 2 was designated a foul or unpleasant odor, and the reference gloves were given a score of 0. Furthermore, roasted coffee beans and the reference glove samples were provided to allow panelists to reset their sense of smell. J o u r n a l P r e -p r o o f The crosslinking density (CD) of rubber gloves is an important parameter for the performance of gloves, which is commonly determined by swelling the rubber in solvents. Swelling is initiated by the diffusion of small solvent molecules into the rubber chain segments. In the beginning, the surface of the rubber has a high solvent concentration while the concentration of the bulk is zero. The diffusion proceeds until equilibrium swelling is achieved, which can be correlated to the crosslink density by the Flory-Rehner equation. The lower the swelling, the higher the CD [28] . The change in the CD of the nitrile gloves with sterilization treatments is presented in Fig. 1 . In comparison to the untreated glove (nitrile-control), the CD of the treated gloves either remained unchanged or decreased. Dry heat (Fig. 1f) sterilization caused a statistically significant reduction (p < 0.05) in the CD for all the sterilization cycles ( Fig. 1 ) as compared to the control. The decrease in CD could be a result of reversion, a phenomenon that occurs when degradation of the polymer chain and/or desulphurization takes place in rubber, which leads to a loss in the CD accompanied by a decline in physical properties [30] . The CD of nitrile gloves sterilized by dry heat decreased up to 5 treatment cycles and increased thereafter. When rubber is subjected to thermal aging, polysulphide linkages are susceptible to disassociation, and hence a decrease in the CD. On the other hand, free sulfur present in the rubber gloves could create new crosslinks, resulting in an increase in CD [31] . This explains the initial decrease and subsequent increase in the CD up to 5 treatment cycles and after 10 dry heat treatment cycles, respectively (Fig. 1f) . J o u r n a l P r e -p r o o f Vinyls are mostly plasticized with phthalate esters to improve chain mobility and lower the glass transition temperature (T g ) of the plastic. However, the ease of migration of the plasticizer into the environment with aging and treatment renders the vinyl gloves less flexible and reduces their physical properties [32] . In this study, the leaching stability of the plasticizers from vinyl as a result of the sterilization methods and cycles was studied by soaking the glove specimens in n-hexane, which is a good solvent for phthalates [33] . can also be correlated to the mechanical property tests in which a decrease in the elongation at break attributed to the leaching out of the plasticizers was noted (Fig. 3) . Similarly, Rabek et al [34] , Understanding fundamental material properties such as tensile strength, modulus, elongation at break, and tensile toughness is important for the manufacturing, testing, and validation of gloves used in various industries. According to ASTM D412 and ASTM D5250, the mechanical properties of gloves must not decrease or increase more than 25% caused by various photochemical and/or thermochemical degradation mechanisms to ensure their suitability for the intended use. All the mechanical data sets were examined via the Ryan-Joiner (RJ) statistical analysis test and concluded to have a normal distribution due to the values being close to one, ranging from 0.90 to 0.99 (refer Fig. S1 and Table S1 ). Before examining the effect of various sterilization treatments on nitrile and vinyl gloves, the non-treated samples were tested in order to compare them with each other and to generate a baseline. As shown in Fig. 3 (Cycle 0), the vinyl gloves are in general stiffer with higher tensile strength and modulus by 25 and 68%, respectively. While, the nitrile gloves were more ductile allowing for these gloves to have high elasticity and toughness as compared to vinyl gloves by about 45 and 159%, respectively. Similar property values are reported in the literature for vinyl gloves. However, the nitrile gloves are on the lower end of the reported values for this type of product in the literature as it ranges from 300 to 600% strains [35] [36] [37] . The six different treatments resulted in a range of effects on the tensile strength, elongation at break, and the modulus at 100% strain as shown in Fig. 3 and cycle. However, the vinyl gloves displayed an overall increase in their modulus, which was likely due to self-crosslinking [38] . The property deterioration in the nitrile gloves was likely due to sulfur cure reversion or devulcanization as noted in the crosslinking study. The dry heat, steam, ethanol, and bleach treatment yielded similar results to one another for both the nitrile and vinyl gloves. Upon examining the tensile strength, modulus, toughness, and elongation at break there is no significant (p > 0.05) change from the baseline value. This means after 20 consecutive cycles of sterilization, the gloves did not experience catastrophic failures. However, one exception is that the vinyl gloves under dry heat and steam treatments had a significant increase in their modulus after the 20 cycles which could be a result of dehydrochlorination. Fonseca et al. [39] has reported that dry heat and steam treatments of vinyl may lead to a dehydrochlorination by a quasi-ionic mechanism at low temperatures (< 200 °C). The dehydrochlorination typically occurs at thermally labile sites such as allylic chloride structures present on the polymer backbone. As the dehydrochlorination process increases the polymer structure begins to form a cross-linked structure [39] . Since the plasticizer migration study as a result of quats treatment displayed increase in the nhexane soluble plasticizers in vinyl gloves (Fig. 2) , the observed reduction in elongation at break could also be a direct result of leaching out of plasticizers. Fig. 4 and Fig. 5 depicts the visual observation and optical microscope images of the gloves surfaces after 20 times of disinfections cycles for nitrile and vinyl gloves, respectively. From the visual inspection, no significant changes were found for the disinfected nitrile gloves (Fig. 4a) . The gloves colour remains similar as the untreated control samples. For the vinyl gloves, no significant changes after disinfection up to 20 cycles were observed except for quats disinfection method. The vinyl glove becoming clear and transparency after treatment as compared to the control vinyl glove sample (Fig. 5a) . This might be due to the leaching out of the plasticizer as evidenced from the crosslink density and mechanical properties results discussed above. The surface morphology of the disinfected gloves was further inspected with optical microscope and the optical images are showed in Fig. 4b-h and Fig. 5b -h for the different disinfection methods of nitrile and vinyl gloves, respectively. Surface micro-cracks and voids were observed for the nitrile gloves after disinfections. Bleach, UV, heat and quats disinfected nitrile gloves showed noticeable micro-cracks as compared to other samples (Fig 4c,d,f, and g) . 5b -h showed the optical images of the vinyl gloves' surface after exposed to the various disinfection methods. As compared to untreated sample, it can be seen that the vinyl gloves experienced some extent of swelling after 20 cycles of disinfections. UV, ethanol, heat and quats disinfected vinyl gloves showed higher extent of swelling as compared to other samples (Fig 5d,e,f, and g) . Overall, both nitrile and vinyl glove' surfaces indicated minor to mild surface changes after the physical and chemical disinfection treatments up to 20 repeated cycles of disinfection. TGA/DTG analysis was performed to evaluate the effect of various sterilization treatments on the thermal stability of the gloves. The effect of twenty cycles of various sterilization methods on TGA and DTG thermograph of nitrile gloves under an inert atmosphere is shown in Fig. 6 (a and a') . The results confirmed that untreated nitrile gloves underwent one major degradation stage (300-600 o C). Any possible weight loss below that can be attributed to the evaporation of surfactants, stearic acid, and wax that are typically used in glove formulations [41] . The main J o u r n a l P r e -p r o o f weight loss step was initiated at around 300 o C, indicating that the gloves were thermostable up to 300 o C and contain negligible amount of moisture. This degradation step could be related to the decomposition of the other heavy carbonaceous organic compounds [42] . Fig. 6 (b and b') displays TGA/DGT data of sterilized nitrile gloves under air. Oxygen as a degrading media creates a different mechanism and affects the overall thermal degradation by producing a twostep degradation. As noted from Fig. 6 , sterilization of nitrile gloves by quats, ethanol, and bleach led to a shift in the degradation thermograms towards lower temperatures under both oxidative and nonoxidative environments. This shows that the involvement of chemical sterilizations has caused a minor reduction in the thermal stability of nitrile gloves. The emergence of a small degradation peak with quats-and ethanol-treated nitrile gloves at temperatures below the main degradation suggest the generation of small molecular weight materials. Contrarily, the employment of physical sterilization treatments such as steam, dry heat, and UV did not change the degradation behavior of nitrile gloves as compared to the non-treated baseline gloves. Overall, the impact of the sterilization treatments on the thermal degradation behavior is in agreement with the mechanical property changes ( Fig. 3 and Table S2 and S3) and barrier performance [22] of nitrile gloves. J o u r n a l P r e -p r o o f The effect of sterilization treatments on the thermal stability of vinyl gloves is presented in Fig. 7 . It was noted that all untreated and treated vinyl gloves showed two thermal degradation stages. The first stage, in the temperature range of 200 to 350 o C, was attributed to dehydrochlorination of vinyl. Cyclization of conjugated polyene sequences occurred at the second stage (400-500 o C) [43] . It can be observed from the insert that the decomposition of ethanol, bleach and quats- In order to study the effect of multiple sterilization cycles on the thermal stability of the nitrile and vinyl gloves, the evolution of TGA/DTG curves is shown in Fig. 8 and Fig. 9 for ten and J o u r n a l P r e -p r o o f 21 twenty sterilization cycles. An interesting phenomenon noticed from these graphs was that the degradation initiation temperatures for the samples treated by chemical sterilization decreased gradually with increasing sterilization cycles. It can be seen that the degradation of the nitrile gloves was more pronounced as compared with vinyl ones indicating higher stability of the vinyl-base gloves against chemical sterilization. J o u r n a l P r e -p r o o f J o u r n a l P r e -p r o o f The changes in the dynamic mechanical properties of the nitrile and vinyl gloves after different sterilization methods and cycles were examined using DMA and the loss factor, Tan δ, curves are presented in Fig. 10 and Fig. 11 , respectively. The Tan δ curves of the nitrile glove only display one peak in the range of -75 to 50 °C due to the maximum chain relaxation i.e. glass transition temperature (T g ). A minor shift in the T g towards higher temperatures (~1-2 °C) was observed for the UV, steam, and ethanol sterilization methods with increasing sterilization cycles ( Fig. 10b-d) . This suggests that the treatments caused minor changes to the crosslinking density of the nitrile rubber after sterilization. The UV, steam, and ethanol sterilization methods cause slight degradation overtime. No significant changes in T g for the heat and bleach sterilizations were observed (Fig. 10a, and 10e) . The T g of the nitrile glove shifted to a lower temperature after treatment with quats solution (Fig. 10f) , which indicates the plasticizing effect of quats compounds to the nitrile glove. The shifting of T g towards lower temperature is mainly attributed to the plasticizing effects [44, 45] . The chain scission after different sterilization methods and cycles causes an increase in the Tan δ magnitude, except for the bleach method. The exposure of moisture from different sterilization methods causes the hydrolysis process which attracts the hydroxyl group in the chemical compounds by the hydrophobic polymer matrix that weakens the polymer chain [46] . For the vinyl gloves, a single maximum chain relaxation T g was detected in the tested temperature range. However, the Tan δ curve of the vinyl glove is much broader with lower amplitude than the nitrile gloves. This shows that nitrile gloves exhibited greater molecular mobility as compared to vinyl gloves [47] . This is in agreement with the higher elongation result for nitrile as compared to the vinyl gloves (Fig. 3) . There was no noticeable shift in the T g of the J o u r n a l P r e -p r o o f 3.7. Odor intensity test Fig. 12a and b presented the odor test ranking setup and the average rankings of sample odors caused by the various disinfection methods based on a group of five panelists, respectively. The gloves used in these tests were conditioned after disinfection to ensure the gloves were dry, resulting in a diminished odor as opposed to when still wet in the case of the chemical disinfection methods. It must be noted that each set of data for the nitrile and vinyl gloves were tested against their own reference sample, as the magnitude of odor intensity differed between nitrile and vinyl gloves. Overall, the intensity of odor from nitrile gloves were higher as Several sterilization methods including physical and chemical procedures were employed for the sterilization and reuse of medical-grade gloves to simultaneously mitigate the shortage and reduce the waste generated from single-use gloves in the current time of COVID-19 pandemic. Our study showed that medical-grade gloves retain their mechanical, thermal, and Supporting the health care workforce during the COVID-19 global epidemic Skin Reactions to Nonglove Personal Protective Equipment: An Emerging Issue in the COVID-19 Pandemic Risk of COVID-19 among front-line health-care workers and the general community: a prospective cohort study COVID-19 epidemic: disentangling the re-emerging controversy about medical facemasks from an epidemiological perspective Critical Supply Shortages -The Need for Ventilators and Personal Protective Equipment during the Covid-19 Pandemic COVID-19 Creating another problem? Sustainable solution for PPE disposal through LCA approach Rational use of personal protective equipment for coronavirus disease ( COVID-19) and considerations during severe shortages: interim guidance Can N95 Respirators Be Reused after Disinfection? How Many Times? Assessment of N95 respirator decontamination and re-use for SARS-CoV-2 An overview of the effect of bioaerosol size in coronavirus disease 2019 transmission Potential of combating transmission of COVID-19 using novel self-cleaning superhydrophobic surfaces: part I-protection strategies against fomites Environmental and Aerosolized SARS-CoV-2 Among Hospitalized COVID-19 Patients Interrupting COVID-19 transmission by implementing enhanced traffic control bundling: Implications for global prevention and control efforts COVID-19 Surface Persistence: A Recent Data Summary and Its Importance for Medical and Dental Settings Aerosol and surface stability of SARS-CoV-2 as compared with SARS-CoV-1 Stability of SARS-CoV-2 on critical personal protective equipment AGA Institute rapid recommendations for gastrointestinal procedures during the COVID-19 pandemic Advantages and disadvantages of gloves, Kanerva's Occup Changing gears: Medical gloves in the era of coronavirus disease 2019 pandemic Disinfection of gloved hands for multiple activities with indicated glove use on the same patient Effect of multiple alcohol-based hand rub applications on the tensile properties of thirteen brands of medical exam nitrile and latex gloves Can Medical-Grade Gloves Provide Protection after Repeated Disinfection ? Key elements of preparedness for pandemic coronavirus disease 2019 (COVID-19) in nuclear medicine units Stability of SARS-CoV-2 in different environmental conditions Guideline for disinfection and sterilization in healthcare facilities Decontamination of face masks and filtering facepiece respirators via ultraviolet germicidal irradiation, hydrogen peroxide vaporisation, and use of dry heat inactivates an infectious SARS-CoV-2 surrogate virus SARS-CoV-2 UV Dose-Response Behavior Characterisation of Crosslinks in Vulcanised Rubbers: From Simple to Advanced Techniques Detection of Crosslink Density by Different Methods for Natural Rubber Blended with SBR and NBR The effects of compounding variables on the reversion process in the sulphur vulcanization of natural rubber Influence of rubber composition on change of crosslink density of rubber vulcanizates with EV cure system by thermal aging Natural-based plasticizers and biopolymer films: A review Effects of Solvent Molecular Structure on Diffusion through Plasticized Poly(vinyl chloride) Films The effect of plasticizers on the erosion and mechanical properties of polymeric films Tensile Properties and Integrity of Clean Room and Low-Modulus Disposable Nitrile Gloves: A Comparison of Two Dissimilar Glove Types PVC Based Medical Gloves and Formulations Therefor Accelerator Free Nitrile Gloves Thermal dehydrochlorination of poly(vinyl chloride) in the presence of Jatropha seed oil Effects of steam on the thermal dehydrochlorination of poly(vinyl chloride) resin and flexible poly(vinyl chloride) under atmospheric pressure Benzalkonium chlorides: uses, regulatory status, and microbial resistance Effect of magnetic and thermal properties of iron oxide nanoparticles (IONs) in nitrile butadiene rubber (NBR) latex Conversion of waste biomass and waste nitrile gloves into renewable fuel Internally plasticized PVC materials: Via covalent attachment of aminated tung oil methyl ester The impact of imidazolium ionic liquids on the properties of nitrile rubber composites The effect of epoxidized sunflower oil on the miscibility of plasticized PVC/NBR blends Durability of GFRP composite exposed to outdoors weathering Dynamic mechanical analysis of fiber-reinforced phenolics