key: cord-0786593-p83vifca authors: Heger, Lukas Andreas; Elsen, Nils; Rieder, Marina; Gauchel, Nadine; Sommerwerck, Urte; Bode, Christoph; Duerschmied, Daniel; Oette, Mark; Ahrens, Ingo title: Clinical analysis on diagnostic accuracy of Bosch Vivalytic SARS-CoV-2 point-of-care test and evaluation of cycle threshold at admission for COVID-19 risk assessment date: 2022-05-23 journal: BMC Infect Dis DOI: 10.1186/s12879-022-07447-7 sha: fd75445dcabf86b2a126dd116b5eccb7f4d3bf0c doc_id: 786593 cord_uid: p83vifca BACKGROUND: Point-of-care (POC) polymerase chain reaction (PCR) tests have the ability to improve testing efficiency in the Coronavirus disease 2019 (COVID-19) pandemic. However, real-world data on POC tests is scarce. OBJECTIVE: To evaluate the efficiency of a novel severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) POC test in a clinical setting and examine the prognostic value of cycle threshold (CT) on admission on the length of hospital stay (LOS) in COVID-19 patients. METHODS: Patients hospitalised between January and May 2021 were included in this prospective cohort study. Patients’ nasopharyngeal swabs were tested for SARS-CoV-2 with Allplex™2019-nCoV (Seegene Inc.) real-time (RT) PCR assay as gold standard as well as a novel POC test (Bosch Vivalytic SARS-CoV-2 [Bosch]) and the SARS-CoV-2 Rapid Antigen Test (Roche) accordingly. Clinical sensitivity and specificity as well as inter- and intra-assay variability were analyzed. RESULTS: 120 patients met the inclusion criteria with 46 (38%) having a definite COVID-19 diagnosis by RT-PCR. Bosch Vivalytic SARS-CoV-2 POC had a sensitivity of 88% and specificity of 96%. The inter- and intra- assay variability was below 15%. The CT value at baseline was lower in patients with LOS ≥ 10 days when compared to patients with LOS < 10 days (27.82 (± 4.648) vs. 36.2 (25.9–39.18); p = 0.0191). There was a negative correlation of CT at admission and LOS (r[44](s) = − 0.31; p = 0.038) but only age was associated with the probability of an increased LOS in a multiple logistic regression analysis (OR 1.105 [95% CI, 1.03–1.19]; p = 0.006). CONCLUSION: Our data indicate that POC testing with Bosch Vivalytic SARS-CoV-2 is a valid strategy to identify COVID-19 patients and decrease turnaround time to definite COVID-19 diagnosis. Also, our data suggest that age at admission possibly with CT value as a combined parameter could be a promising tool for risk assessment of increased length of hospital stay and severity of disease in COVID-19 patients. Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) with the resulting Coronavirus disease 2019 pandemic has presented hospitals with considerable challenges [1] . These are partly aggravated by long COVID-19 test turnaround times, associated with centralised laboratory polymerase chain reaction (PCR) testing, resulting in poor patient flow and nosocomial transmission [2, 3] . Standard real-time (RT)-PCR for COVID-19 is the gold standard but requires specialized materials, equipment, personnel and transportation to a centralized laboratory [4] . Point-of-care (POC) PCR tests have the ability to improve testing efficiency by reducing steps between sample collection and results, ultimately contributing to better COVID-19 containment [3] . Numerous assays are suggested for use in clinical practice. However, with some of them just recently released, data on their clinical performance is limited [5] . We evaluate in a prospective clinical setting the Bosch Vivalytic (Bosch Healthcare Solutions GmbH; Stuttgarter, Germany) SARS-CoV-2 test for Bosch Vivalytic One and the COVID-19 ELISA Elecsys Anti-SARS-CoV-2 assay (Roche Holding AG; Basel, Swiss) with a centralised laboratory PCR test (Seegene [Seegene, Inc; Seoul, South Korea] AllplexTM2019-nCoV) carried out by an extern laboratory (Laboratory Dr. Wisplinghoff, Horbeller Str. 18-20 Koeln, Germany) as gold standard. The Vivalytic analyser developed by Bosch is a portable device for molecular diagnostics, able to perform PCR procedures fully automatically. The increment has to be collected in a guanidine thiocyanate-based medium (eNAT ® [COPAN Diagnostics Inc.; Murrieta, USA)]), which stabilizes the viral RNA and completely inactivates the microbial viability. After that the composite is loaded into a specific Vivalytic SARS-CoV-2 cartridge. The Bosch Vivalytic SARS-CoV-2 is singleplex test, which targets the E gene sequence for COVID-19 detection. It takes 39 min to be performed on a specimen volume of 300 µl. Current data point towards a plausible positive correlation between the amount of detected virus measured by the proxy PCR cycle threshold (CT) value and the severity of disease [6] . Previous evidence from influenza suggests that the higher the initial viral load, the worse the clinical development [7] . However, firm conclusion on the relationship between initial CT value and individual prognosis is still missing and data for the most part limited to nasopharyngeal viral load samples processed through external laboratories [8] . Whether CT values derived from POC testing could yield similar results is yet to be established. Different scoring systems have been suggested for COVID-19 risk-assessment, with several of them being derived from established assays [9, 10] . Although being influenced by comorbidities and patient characteristics, Length Of hospital Stay (LOS) has been shown to be a rational choice to measure severity of disease and level of care needed in hospitalized patient [11, 12] . Therefore, we strive to evaluate cycle threshold (CT) in POC reverse-PCR tests in confirmed COVID-19 patients as possible marker for LOS and severity of disease. Such early indicators providing insight into potential disease progression are critical in order to properly select patients requiring special care, especially when exponentially increasing numbers of infections occur in a short time interval and hospitals become overcrowded. Not only the heterogeneous clinical presentations of COVID-19 poses a challenge for attending physicians but also the prevalence of a variety of late-effects can be difficult to manage [13] . Divided into two different entities, "post-acute COVID-19" where symptoms extend beyond 3 weeks and "chronic COVID-19" where symptoms extend beyond 12 weeks, long Keywords: COVID-19, Point of care, PCR test, Length of stay, Patient flow COVID-19 is estimated to affect up to 80% of COVID-19 patients [14, 15] . However, real-world data for hospitalized patients and long COVID-19 is limited. We report data from an investigator-initiated, single-center, prospective registry study to evaluate Vivalytic ™ SARS-CoV-2 Multiplex POC (Bosch Healthcare Solutions GmbH; Stuttgarter, Germany) COVID-19 test in clinical practice and to evaluate CT value at admission for risk stratification in hospitalized patients conducted at the Severinenklösterchen Hospital-University of Cologne. The protocol of this study conforms to the ethical guidelines of the 1975 Declaration of Helsinki and was approved by the institutional ethical committee of the Ärztekammer Nordrhein (20211001). All-comers admitted to the hospital over 18 years were eligible for inclusion. Hospital rules dictated that every patient admitted had to be tested. Tested patients were picked randomly for participation. The decision to perform a PCR-test for SARS-CoV-2 was made independently of study inclusion by the treating physician and patients were asked to participate before the test results were available. Written informed consent was obtained from all patients prior to study inclusion. Patients with a positive PCR-test for SARS-CoV-2 were allocated to the "positive" group, patients with a negative PCR-test for SARS-CoV-2 to the control group (Fig. 1) . Every patient was then tested two additional times using Vivalytic SARS-CoV-2 Test, and SARS-CoV-2 Rapid Antigen Test respectively. All nasopharyngeal swabs were taken by nurses or trained personnel. Patients admitted to the hospital between January and May 2021 were randomly asked to participate in the study. If patients agreed to participate, overall characteristics such as sex, age and medical history were recorded. To characterize the severity of the course of disease in COVID-19, patients were divided according to the Lost to follow-up (n=19) Fig. 1 Schematic representation of the allocation of the 120 participants to the positive or negative group. Nineteen participants of the SARS-CoV-positive cohort were lost to follow-up and 5 participants died. The flow diagram is based on the template of the CONSORT flow diagram [16] length of stay (LOS): A prolonged LOS was defined as equal to or greater than the median length of hospital stay. All clinical data gathered during this period was obtained from the electronic patient file. During followup no interventions were applied for the purpose of this study and all therapeutic and diagnostic procedures were applied as part of standard care at the discretion of the treating physicians. Finally, participants were contacted by phone and asked about the course of disease using a standardized questionnaire after 3 months. Herein, the patients were asked to rank their current physical fitness on a scale from 0 to 100% as they felt it affected by their COVID-19 disease. SARS-CoV-2 testing using Allplex ™ 2019-nCOV Assay (Seegene, Seoul, Korea: polymerase chain reaction assay (PCR) targeting envelope protein-(E-), RNA-dependent RNA polymerase-(RdRP-), and N-genes) was performed by an external laboratory with Applied Biosystems ™ 7500 Real-Time PCR System according to the manufacturer's guidelines. For analysis with Bosch Vivalytic SARS-CoV-2 assay (Bosch Healthcare Solutions GmbH, Waiblingen, Germany) using the Vivalytic SARS-CoV-2 analyzer, samples were collected in a guanidine thiocyanate-based medium (eNAT ® [COPAN Diagnostics Inc.; Murrieta, USA]), stabilizing the viral RNA and completely inactivating the microbial viability. Vivalytic SARS-CoV-2 analyzer is a portable device and works fully automatically. Roche SARS-CoV-2 Rapid Antigen Test is an immunochromatographic assay for qualitative detection of SARS-CoV-2 infection in nasopharyngeal swabs. The presence of viral antigens in sufficient concentration enables binding to specific mouse monoclonal anti-SARS-CoV-2 antibodies, which is then reflected by the appearance of a visual indicator after the clinical specimen is collected and then deposited in a pre-filled extraction buffer container with the solution being put on the test sample according to manufacturer's guidelines. Inter-assay variability was performed conducting repetitive tests on four samples throughout 3 days in triplicates. Intra-assay variability was performed conducting four consecutive assays in duplicates in one run. For intra assay precision we used inactivated whole pathogen sample (AMPLIRUN ® TOTAL SARS-CoV-2 RNA Control [Vircell SA, Granada, Spain]) and for inter assay precision we used both, inactivated whole pathogen sample as well as patient samples. 15 000 AMPLIRUN ® SARS-CoV-2 RNA CONTROL copies were reconstituted in guanidine thiocyanate-based medium eNAT ® (COPAN Diagnostics Inc.; Murrieta, USA) to generate the control sample. Then a dilution series of consecutive samples was performed (1:2; 1:10; 1:100; 1:1000). Both, the results from Bosch Vivalytic (Bosch Healthcare Solutions GmbH; Stuttgarter, Germany) SARS-CoV-2 test for Bosch Vivalytic One as well as the results from the COVID-19 ELISA Elecsys Anti-SARS-CoV-2 assay (Roche Holding AG; Basel, Swiss) were compared to the results from the current gold standard at Severinenklösterchen Hospital -University of Cologne, a centralised laboratory PCR test (Seegene [Seegene, Inc; Seoul, South Korea] AllplexTM2019-nCoV) carried out by an extern laboratory (Laboratory Dr. Wisplinghoff, Horbeller Str. 18-20 Koeln, Germany) to calculate inter-and intraassay precision. Continuous patient data were compared using a Mann-Whitney U-test or unpaired t-test. Categorical differences between patient groups were compared using Fishers exact test. Continuous variables are presented as mean ± standard deviation (SD) if found to follow a Gaussian distribution according to the D' Agosstino-Pearson omnibus normality test or as median ± lower and upper quartiles if found to follow a non-Gaussian distribution. Categorical patient characteristics are presented as percentages. Spearman nonparametric correlation was used to evaluate the degree to which LOS and CT value at admission move in relation to each other. Independent predictors of LOS above or below median length of stay ( Descriptive analyses were performed using Graph Pad Prism Version 9.0 (Prism 9 for Mac OS X; GraphPad Software. Inc. La Jolla. CA). We included 120 patients hospitalized between January and May 2021 in this prospective cohort study. Of those, 46 (38%) patients tested positive for COVID-19 infection via the external laboratory. The COVID-19 positive group was younger when compared to the negative control group (55. 5 Inter-and Intra-assay coefficients of variability for Vivalytic ™ SARS-CoV-2 Multiplex analyzer were below 15% (Tables 2, 3, 4). In relation to the results of the external laboratory RT-PCR test, the POC test had a test-sensitivity for SARS-CoV-2 infection of 88% with a specificity of 96%. The positive predictive value was 90% and the negative predictive value 95%. The antigen test had a sensitivity for COVID-19 infection of 65% with a specificity of 100%. The positive predictive value for the antigen test was 100% and the negative predictive value 88% when compared to the All-plexTM2019-nCoV assay. Computing a nonparametric spearman correlation showed a high correlation between the CT Values from the Vivalytic SARS-CoV-2 assay and the AllplexTM2019-nCoV assay (r[38] = 0.6158, p ≤ 0.0001). To further assess the agreement between the two methods, Bland-Altman plot was used. After eliminating CT values above the maximum number of cycles (CT < 35; N = 16), Bland-Atman shows good agreement between both test methods (limits of agreement 0.2125 [− 1,777,113-1,819,613] ). Patients with positive SARS-CoV-2 tests had an overall median length of hospital stay of 9.5 (4.75-15.25) days. To quantify severity of disease we divided COVID-19 patients into two groups. A prolonged LOS was defined as equal to or greater than the median length of hospital years; p = 0.001) and had higher levels of haemoglobin (13.6 (± 1.7) g/dl vs. 11.9 (± 2.3) g/dl; p = 0.007) when compared to patients with a LOS > 10 days. There was a negative correlation of CT at admission and LOS (r[44] s = − 0.31; p = 0.038). There was no statistically significant difference between both groups regarding other baseline characteristics collected. In the group with LOS < 10 days, 15 (75%) patients were affected by the virus variant B1.1.7 and there were 16 (61.5%) patients affected in the group with LOS ≥ 10 days (p = 0.4834). (Table 5) . Of 46 patients who tested positive, 22 completed the follow-up and 5 COVID-19 patients died (4 patients due to COVID-19 on day 19, day 5, day 14 and day 5 after inclusion and 1 patient due to severe heart failure). All patients died while being hospitalized. At 3 months after hospital discharge, 15 (68%) COVID-19 patients had persistent depreciation of quality of life incidental with COVID-19 disease (80 [70-90] % of the normal physical capacity). In this single centre prospective cohort study we evaluate real-world diagnostic accuracy of point-of-care testing using Bosch Vivalytic SARS-CoV-2 test for Bosch Vivalytic One and a commercially available antigen test by comparing sensitivity and specificity to an established RT-PCR test performed by an external laboratory. Our prospective and comparative evaluation shows that Bosch Vivalytic SARS-CoV-2 POC test is a valid strategy for COVID-19 detection with the potential to decrease door-to-diagnosis-time allowing earlier detection of COVID-19 positive patients [3, 17] . With its advertised turnaround time of approximately 39 min it far exceeds those of established external laboratories using RT-PCT by far [5] . The agreement of the POC test and RT-PCR is good with robust inter-and intra-assay reliability and good detection of positive samples. However, recently published data indicates that in the very early phase of the infection with high CT values, in which the RT-PCR can already detect the infection, the antigen level is not high enough for the POC test, necessitating repeated testing [18, 19] . While RT-PCR should still be the gold standard, POC testing could help with patient triage and improve patient flow dynamics [20] . The considerably higher sensitivity of the POC test when compared to antigen detection alone especially qualifies it to reduce the number of missed infections with COVID-19, a serious risk when using the antigen test as reported in other publications [21] . The prognostic value of CT in COVID-19 patients has been suggested by several studies [22] . Prospective clinical data as well as data depending on CT values acquired using POC tests however is scarce. Our data shows that on admission POC CT values are lower in patients with prolonged length of hospital stay for COVID-19 patients and that CT value at admission correlates with LOS. In a multivariate analysis however, only age was associated with LOS and variability of CT values of up to 8 cycles were observed for the same specimen material tested in a different laboratory in a recent publication [23] . Therefore, risk assessment using CT values should only be considered if tests are acquired from only one laboratory using the same assay and can only add to initial risk assessment. Nevertheless, through their easier implementation, cost effectiveness and accuracy POC COVID-19 test could help monitoring epidemiological features of the COVID-19 pandemic through testing of large cohorts [24] . Long-term health effects of COVID-19 called Long-COVID or COVID long-haulers are challenging for clinicians through their heterogeneity [13] . It is vital to understand the percentage of patients suffering from long-term effects. Our data shows, that up to 68% of discharged patients experience a COVID-19 related decrease in quality of life at three months after hospital discharge. This is in line with several other studies and the recommendation for patients discharged after [25, 26] . Our data depicts the third wave of the COVID-19 pandemic in Germany early 2021 with SARS-CoV-2 Variant B.1.1.7 being the predominantly detected variant [27] . This mutant strain prompted border closures, lockdowns and new restrictions ultimately cumulating in further deaths and economic loss. However, with a large proportion of the elderly being already inoculated following the guidelines of the Standing Committee on Vaccination (STIKO) we report COVID-19 patients to be younger then regular hospital all-comers forming our control group [28] . This all-the-more emphasises the need for rapid inoculation to protect patients. Several routine laboratory markers have been suggested in the COVID-19 diagnostic algorithm [29] . In a recent meta-analysis LDH has been suggested as a cost-effective biomarker in patients with COVID-19 [30] . An important variable in our study is that two distinct patient groups were analysed one composed from regular all-comers and one of COVID-19 patients. In our prospective data, elevated LDH was a distinct feature of COVID-19 patients when compared to the control group. However, LDH was not associated with LOS. Larger clinical trials are needed to confirm the value of LDH at baseline for COVID-19 diagnosis. POC PCR is a qualitative not quantitative measurement and it is not possible to quantify the virus load via CT values. CT values vary in-between different assays. We used only a single negative RT-PCR to confirm the absence of COVID-19 infection, risking missing infection. A lot of patients were lost to follow-up mainly due to the quantity of measures taken during the pandemic and the hence decreasing will of people to participate. We did not assess quality of life at the time point of inclusion. The fact, that patients with LOS > 10 days were older could possibly diminish the interpretability of our data. Multivariate analysis showed only age being associated with LOS. However, we argue, that within the complex pathology that is COVID-19, CT value could be a valid tool to identify patients at high risk. Larger clinical trials are needed to validate this. This data from a single centre provides prospective and comparative data indicating, that POC testing using Bosch Vivalytic SARS-CoV-2 is a valid strategy to identify COVID-19 patients and decrease turnaround time during COVID-19 pandemic. Also, our data suggets that the on admission CT value, maybe combined with on admission age, might be a marker for length of hospital stay and severity of disease in COVID-19 patients. Our data emphasise the increased risk for long-term health effects of COVID-19 patients and the consecutive need for individual follow-up. Impact of COVID-19 pandemic on utilisation of healthcare services: a systematic review convenient online submission • thorough peer review by experienced researchers in your field • rapid publication on acceptance • support for research data, including large and complex data types • gold Open Access which fosters wider collaboration and increased citations maximum visibility for your research: over 100M website views per year submit your research ? Clinical characteristics and diagnostic challenges of COVID-19: an update from the global perspective Clinical impact of molecular point-of-care testing for suspected COVID-19 in hospital (COV-19POC): a prospective, interventional, nonrandomised, controlled study Challenges and controversies to testing for COVID-19 Comparative analysis of point-of-care, high-throughput and laboratorydeveloped SARS-CoV-2 nucleic acid amplification tests (NATs) Diagnostic SARS-CoV-2 cycle threshold value predicts disease severity, survival, and six-month sequelae in COVID-19 symptomatic patients A dose-finding study of a wild-type influenza A(H3N2) virus in a healthy volunteer human challenge model A systematic review of the clinical utility of cycle threshold values in the context of COVID-19 Predictive performance of SOFA and qSOFA for in-hospital mortality in severe novel coronavirus disease Performance of pneumonia severity index and CURB-65 in predicting 30-day mortality in patients with COVID-19 COVID-19 length of hospital stay: a systematic review and data synthesis Composite length of stay, an outcome measure of postoperative and readmission length of stays in pancreatoduodenectomy More than 50 Long-term effects of COVID-19: a systematic review and meta-analysis. medRxiv Persistent symptoms in patients after acute COVID-19 Long COVID: an overview The CONSORT statement Rapid, point-of-care antigen and molecular-based tests for diagnosis of SARS-CoV-2 infection Rethinking COVID-19 test sensitivity-a strategy for containment Performance of a point-of-care test for the rapid detection of SARS-CoV-2 antigen Patient flow dynamics in hospital systems during times of COVID-19: cox proportional hazard regression analysis Implementing SARS-CoV-2 rapid antigen testing in the emergency ward of a Swiss University Hospital: the INCREASE study Prognostic value of "Cycle Threshold" in confirmed COVID-19 patients To interpret the SARS-CoV-2 test, consider the cycle threshold value Analysis of cycle threshold values in SARS-CoV-2-PCR in a long-term study Understanding the long-term health effects of COVID-19. EClinicalMedicine COVID-19 discharge and follow-up recommendations Rapid increase of SARS-CoV-2 Variant B.1.1.7 detected in sewage samples from England between STIKO vaccination recommendations: vaccination of immunodeficient patients and vaccination against COVID-19 Routine laboratory testing to determine if a patient has COVID-19 Prognostic value of elevated lactate dehydrogenase in patients with COVID-19: a systematic review and meta-analysis Publisher's Note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations