key: cord-0785175-nhg1td1r authors: Fiedler, Melanie; Holtkamp, Caroline; Dittmer, Ulf; Anastasiou, Olympia E. title: Performance of the LIAISON(®) SARS-CoV-2 Antigen Assay vs. SARS-CoV-2-RT-PCR date: 2021-05-26 journal: Pathogens DOI: 10.3390/pathogens10060658 sha: b6e992527f0bee4bccf3fb4f526535c4f0fd5668 doc_id: 785175 cord_uid: nhg1td1r We aimed to evaluate the LIAISON(®) SARS-CoV-2 antigen assay (DiaSorin), comparing its performance to real-time polymerase chain reaction (RT-PCR) for the detection of SARS-CoV-2 RNA. 182 (110 PCR-positive and 72 PCR-negative) nasopharyngeal swab samples were taken for the detection of SARS-CoV-2. RT-PCR and antigen assay were performed using the same material. The sensitivity and specificity of the antigen assay were calculated for different cut-offs, with RT-PCR serving as the reference method. Stored clinical samples that were positive for other respiratory viruses were tested to evaluate cross-reactivity. One third (33/110, 30%) were falsely classified as negative, while no false positives were found using the 200 TCID(50)/mL cut-off for the SARS-CoV-2 antigen as proposed by the manufacturer. This corresponded to a sensitivity of 70% (60–78%) and a specificity of 100% (94–100%). Lowering the cut-off for positivity of the antigen assay to 22.79 or 57.68 TCID(50)/mL increased the sensitivity of the method, reaching a sensitivity of 92% (85–96%) vs. 79% (70–86%) and a specificity of 81% (69–89%) vs. 99% (91–100%), respectively. The antigen assay reliably detected samples with high SARS-CoV-2 viral loads (≥10(6) copies SARS-CoV-2/mL), while it cannot differentiate between negative and low positive samples. Cross-reactivity toward other respiratory viruses was not detected. Since 2019, worldwide healthcare systems are being challenged by the COVID-19 pandemic, which is caused by the severe acute respiratory syndrome coronavirus (SARS-CoV-2). Because of the high infectiousness of the virus, it is important to test for it on a large scale, including with potential asymptomatic carriers [1] . Optimizing test strategies during the SARS-CoV-2 pandemic is an important and challenging goal. Hence, we aimed to evaluate the LIAISON ® SARS-CoV-2 antigen assay (Ag assay) (DiaSorin), comparing its performance to real-time polymerase chain reaction (RT-PCR) for the detection of SARS-CoV-2 RNA in nasopharyngeal swabs. The LIAISON ® SARS-CoV-2 Ag assay is a chemiluminescence sandwich-immunoassay (CLIA)-based technology for the determination of nucleocapsid antigen from SARS-CoV-2 samples in upper respiratory specimens. We used two different RT-PCRs: The RealStar ® SARS-CoV-2 RT-PCR kit (Altona), which targets the SARS-CoV-2 genes S (spike) and E (envelope), and the Alinity m SARS-CoV-2 Assay (Abbott), which targets the SARS-CoV-2 genes RdRp (RNA dependent RNA polymerase) and N (nucleocapsid). RT-PCR is actually considered the gold standard for the detection of SARS-CoV-2 [2] . Nevertheless, there are some disadvantages with using RT-PCR. It is time-consuming, expensive, prone to contamination, and trained laboratory staff as well as specialized equipment are needed [1] . In addition, during the pandemic, reagents have often been scarce or even unavailable [3] , and the demands on specialized laboratories (providing RT-PCR diagnostics) consequently increased. Hence, we evaluated whether using an antigen assay could be an adequate back up system for the RT-PCR. A benefit of using the antigen assay is the availability of the LIAISON ® in many laboratories, which could potentially expand our diagnostic capabilities. Previous studies have already compared different antigen rapid diagnostic tests with RT-PCR: Corman et al. [3] showed that antigen rapid diagnostic tests were able to detect SARS-CoV-2 viral load of between 10 6 and 10 7 copies per swab. The objective of our study was the comprehensive evaluation of the LIAISON ® SARS-CoV-2 antigen assay for the detection of SARS-CoV-2 in respiratory tract samples, focusing on a head-to-head comparison with the gold-standard method, RT-PCR. According to the manufacturer's instruction for the LIAISON ® SARS-CoV-2 antigen assay, the deactivation of the sample should be performed within 12 h after sampling. Collecting, distributing, and processing the samples within this timeframe can be a logistical challenge for many healthcare providers. To test if the antigen measurement remains stable over a longer timeframe, we used samples from 10 patients, with 400 µL of each sample being inactivated from days 1 to 4. The original material was kept at 4 • C for the fourday period. All samples had a viral load of >10 6 copies/mL, ranging from 1.5 × 10 6 to 1.9 × 10 8 copies/mL. As shown in Figure 1a , the antigen levels remained remarkably stable over the first three days, which is in line with our daily routine in the laboratory setting. When comparing the first and fourth day, we observed a significant increase in the normalized Tissue Culture Infectious Dose (TCID 50 /mL) values of 27%, with an interquartile range of 0.5-37% (p = 0.02). All other comparisons between different timepoints did not reach statistical significance. rce or even unavailable [3] , and the demands on specialized laboratories (providing PCR diagnostics) consequently increased. Hence, we evaluated whether using an ann assay could be an adequate back up system for the RT-PCR. A benefit of using the igen assay is the availability of the LIAISON ® in many laboratories, which could potially expand our diagnostic capabilities. Previous studies have already compared difnt antigen rapid diagnostic tests with RT-PCR: Corman et al. [3] showed that antigen id diagnostic tests were able to detect SARS-CoV-2 viral load of between 10 6 and 10 7 ies per swab. The objective of our study was the comprehensive evaluation of the LIAISON ® SARS--2 antigen assay for the detection of SARS-CoV-2 in respiratory tract samples, focuson a head-to-head comparison with the gold-standard method, RT-PCR. According to the manufacturer's instruction for the LIAISON ® SARS-CoV-2 antigen y, the deactivation of the sample should be performed within 12 h after sampling. lecting, distributing, and processing the samples within this timeframe can be a logisl challenge for many healthcare providers. To test if the antigen measurement remains le over a longer timeframe, we used samples from 10 patients, with 400 µL of each ple being inactivated from days 1 to 4. The original material was kept at 4 °C for the r-day period. All samples had a viral load of >10 6 copies/mL, ranging from 1.5 × 10 6 to × 10 8 copies/mL. As shown in Figure 1a , the antigen levels remained remarkably stable r the first three days, which is in line with our daily routine in the laboratory setting. en comparing the first and fourth day, we observed a significant increase in the norlized Tissue Culture Infectious Dose (TCID50/mL) values of 27%, with an interquartile ge of 0.5-37% (p = 0.02). All other comparisons between different time-points did not ch statistical significance. To test the linearity of the assay, we performed a series of dilutions (1:2, 1:4, 1:8, 1:16, , 1:64, 1:128, 1:256 and 1:512) for two samples. Each dilution and the undiluted sample e tested in duplicate. The assay demonstrated linearity, especially for values over the TCID50/mL cut-off (Figure 1b) . Values are presented as median (interquartile range). The comparison between the gr was performed using the Friedmann test (p = 0.02). A significant difference can be observed w comparing the first to the fourth day (p = 0.02). All other comparisons did not reach statistical nificance. (b) Serial dilution and linearity of the assay. The line represents the actual dilution v ues on the x-and y-axis, and the triangles and dots represent the measured values for each di tion. The triangles and dots present the average value from two technical duplicates. The assa demonstrates linearity, especially for values over the 200 Tissue Culture Infectious Dose (TCID50/mL) cut-off. We tested 182 nasopharyngeal swab samples by PCR and the LIAISON ® SARS-2 antigen assay to detect SARS-CoV-2. The PCR was considered the gold standard fo classification of the antigen results as true/false negative and positive. Our cohort sisted of 110 SARS-CoV-2 PCR-positive and 72 negative samples. A third (33/110, 30 the samples were falsely classified as negative with the antigen assay. No false posi We tested 182 nasopharyngeal swab samples by PCR and the LIAISON ® SARS-CoV-2 antigen assay to detect SARS-CoV-2. The PCR was considered the gold standard for the classification of the antigen results as true/false negative and positive. Our cohort consisted of 110 SARS-CoV-2 PCR-positive and 72 negative samples. A third (33/110, 30%) of the samples were falsely classified as negative with the antigen assay. No false positives were found using the cut-off of 200 TCID 50 /mL for the SARS-CoV-2 antigen as shown in Table 1 , corresponding to a sensitivity of 70% (60-78%), a specificity of 100% (94-100%), a positive predictive value (PPV) of 1 (0.94-1), and a negative predictive value (NPV) of 0.69 (0.59-0.77). Table 1 . Stratification of the patients according to their SARS-CoV-2 RNA and antigen status (cut-off of 200 TCID 50 /mL) in nasopharyngeal swabs. We investigated whether using a different cut-off for sample positivity could improve sensitivity. The 100 TCID 50 /mL was selected for evaluation because the range from 100-200 TCID 50 /mL is defined as borderline in newer versions of the information sheet for the LIAISON ® SARS-CoV-2 antigen assay. In addition, a receiver operating curve (ROC) was calculated to pinpoint optimal cut-offs ( Figure 2 ). The area under the curve (AUC) was 0.932 (0.895-0.969) (p < 0.001). Sensitivity, specificity, PPV, and NPV were calculated for these four different cut-offs ( Table 2) . Raw values for Table 2 can be found in the Supplementary Section. The first two cut-offs were selected because they provided a sensitivity of over 90%. The use of 22.79 or 23.56 TCID 50 /mL increased the sensitivity, increasing the correct classification of positive samples from 77/110 (70%) to 101 (92%) and 100/110 (91%), respectively. It also led to the misclassification of 14/72 (19%) and 12/72 (17%) negative samples as false positive, respectively. The third cut-off (57.68 TCID 50 /mL) was calculated according to the maximum value of the Youden Index for providing the best combination of sensitivity (79%) and specificity (99%) [4] . The fourth (100 TCID 50 /mL) provided a specificity of 100%, while the sensitivity was only 75%, leading to a misclassification as false negative for 27 of the 110 positive samples. Next, we looked at a potential correlation between the antigen levels in TCID 50 /mL and the viral load measured through SARS-CoV-2 PCR. A linear relationship was observed between the logarithms of antigen concentration and viral load as shown in Figure 3a Sensitivity, specificity, PPV, and NPV were calculated for these four different cut-offs ( Table 2) . Raw values for Table 2 can be found in the Supplementary Section. The first two cut-offs were selected because they provided a sensitivity of over 90%. The use of 22.79 or 23.56 TCID₅₀/mL increased the sensitivity, increasing the correct classification of positive samples from 77/110 (70%) to 101 (92%) and 100/110 (91%), respectively. It also led to the misclassification of 14/72 (19%) and 12/72 (17%) negative samples as false positive, respectively. The third cut-off (57.68 TCID₅₀/mL) was calculated according to the maximum value of the Youden Index for providing the best combination of sensitivity (79%) and specificity (99%) [4] . The fourth (100 TCID₅₀/mL) provided a specificity of 100%, while the sensitivity was only 75%, leading to a misclassification as false negative for 27 of the 110 positive samples. Table 2 . Sensitivity, specificity, PPV, and NPV for four different SARS-CoV-2 antigen level cut-offs for SARS-CoV-2 detection in nasopharyngeal swabs. Plotting the antigen levels in PCR-negative and all PCR-positive samples, as well as the more differentiated view of the positive samples with more than 10 4 (low positive), 10 5 (moderately high positive), 10 6 (high positive), and 10 7 (very high positive) copies of SARS-CoV-2 per mL revealed that the antigen assay can reliably differentiate between negative samples and samples with high viral loads. However, we observed a significant overlap between negative and low positive samples (Figure 3b ). One of the advantages of RT-PCR is the detection of viral RNA in different materials. We aimed to test whether the LIAISON ® SARS-CoV-2 antigen assay could also be used to test different respiratory tract materials. We evaluated 28 pharyngeal wash samples, 8 bronchoalveolar lavage (BAL) samples, and 1 sputum sample (data not shown). The measurement could not be completed due to pipetting errors in two BAL samples and the sputum sample. We included 24 RT-PCR-positive pharyngeal wash samples, with a viral load ranging from 78 to 1.7 × 10 5 copies/mL and 4 negative ones. Using the 200 TCID 50 /mL cut-off for negativity, all 4 PCR-negative samples were correctly classified as negative, while 23 samples were misclassified as negative and 1 sample with a viral load of 1.7 × 10 5 copies/mL was correctly classified as positive. Using the 57.68 TCID 50 /mL cutoff for negativity all samples, positive and negative, were classified as positive. From the remaining 6 BAL samples, 1 was positive with a viral load of 6 × 10 4 copies/mL and 5 negative. When using the 200 TCID 50 /mL cut-off for negativity, all BAL samples were negative, while when using the 57.68 TCID 50 /mL cut-off, all samples were classified correctly. Next, we looked at a potential correlation between the antigen levels in TCID₅₀/mL and the viral load measured through SARS-CoV-2 PCR. A linear relationship was observed between the logarithms of antigen concentration and viral load as shown in Figure 3a . After excluding values outside the quantified range of antigen concentration (22-100,000), we calculated a linear regression model (r 2 = 0.836, p < 0.001) with the following equation: log(viral load) = 2.51 + [1.1 × log(antigen)]. Plotting the antigen levels in PCR-negative and all PCR-positive samples, as well as the more differentiated view of the positive samples with more than 10 4 (low positive), 10 5 (moderately high positive), 10 6 (high positive), and 10 7 (very high positive) copies of SARS-CoV-2 per mL revealed that the antigen assay can reliably differentiate between negative samples and samples with high viral loads. However, we observed a significant overlap between negative and low positive samples (Figure 3b ). Samples positive for endemic human coronaviruses (HCoVs) (HCoV NL63, HCoV OC43, HCoV 229E, HCoV HKU), influenza A and B, respiratory syncytial virus (RSV), and samples from patients with detectable Epstein-Barr Virus (EBV) in blood were tested using the LIAISON SARS-CoV-2 antigen assay (Ag assay) (data not shown). We observed no cross-reactivity, taking into account the manufacturer s cut-off of 200 TCID 50 /mL. The RSV, influenza, HCoV NL63, HCoV OC43 had a value of less than 22 TCID 50 /mL, while the TCID 50 /mL values for HCoV 229E and HKU amounted to 22.2 and 49.2 TCID 50 /mL. The TCID 50 /mL values of the samples of EBV positive patients ranged from 30.9 to 35.7 TCID 50 /mL. Lastly, we tested 10 nasopharyngeal swab samples (viral load >10 6 copies/mL) from patients carrying the B.1.1.7 variant. All had the detectable SARS-CoV-2 antigen when taking the 200 TCID 50 /mL cut-off into account (data not shown). In this study, we compared the performance of the LIAISON ® SARS-CoV-2 antigen assay vs. the gold standard for SARS-CoV-2 detection, RT-PCR [2] . Our cohort consisted of individuals tested by the Public Health Department of Essen. In case of outbreaks, this also included asymptomatic individuals. Unfortunately, we cannot provide precise information since none is available to us. The absence of such information, however, is not unusual for samples sent to a laboratory for SARS-CoV-2 testing, but it is still a limitation of the study. A third of the samples with detectable SARS-CoV-2 RNA by RT-PCR were falsely classified as negative, with the antigen test taking the manufacturer's recommended cut-off of 200 TCID 50 /mL. The antigen test is, indeed, inferior to the RT-PCR in terms of sensitivity, which is not surprising. RT-PCR is considered the gold standard for SARS-CoV-2 detection. However, even the gold standard is not without disadvantages: it is expensive, time consuming, requires trained laboratory staff as well as specialized equipment [1] , and may not always be available (for example, due to lack of reagents) [3] . However, no false positives were found, amounting to 100% specificity and 70% sensitivity. Antigen tests, especially antigen rapid diagnostic tests, are known to be less sensitive than RT-PCR [5] and may be comparable or inferior in terms of sensitivity to other rapid detection methods, such as loop-mediated isothermal amplification assay [6, 7] . The WHO recommends the use of antigen tests with sensitivity and specificity greater than 80% and 97%, respectively [8] . Using a ROC analysis, we identified different cut-offs, optimized for maximal sensitivity and/or a combination of sensitivity and specificity, and compared three of them. As expected, lowering the cut-off for the antigen assay increased sensitivity. In fact, the two lowest cut-offs increased sensitivity to over 90%. This is significantly higher than the sensitivity of various antigen rapid diagnostic tests, which range from 73% to 82%, as reported in a meta-analysis by Brümmer, et al. [5] , and may be as low as 62.5% in the case of asymptomatic subjects. However, the higher sensitivity, attained by lowering the cut-off, comes at the cost of decreased specificity. Using the cut-off between 22 and 23 TCID 50 /mL was found to lead to false positive results for almost a fifth of the negative samples. The third and optimal cut-off (57.68 TCID 50 /mL), calculated according to the maximal Youden Index, showed an acceptable combination of sensitivity (79%) and specificity (99%). This cut-off may be considered by the manufacturer. In addition, this cut-off roughly fulfilled the WHO standard for SARS-CoV-2 antigen tests [8] . Recently, a Japanese group evaluated the performance of another quantitative, fully automated antigen assay with nasopharyngeal swabs. The Lumipulse ® SARS-CoV-2 Ag test (Fujirebio Inc., Tokyo, Japan) is also a chemiluminescence enzyme immunoassay, which tests the nucleocapsid antigen [9] . They described a sensitivity of 91.7% and a specificity of 98.5% using an optimized cut-off. The specificity is roughly in line with our results, but the sensitivity is substantially higher. This may reflect the superiority of the Fujirebio assay compared to the LIAISON ® SARS-CoV-2 antigen assay. A head-to-head comparison of the two assays would be necessary to conclusively evaluate the point, since the two studies focused on different cohorts. Virus concentrations of 10 6 and 10 7 copies/mL are almost always detected using the cut-off of 57.68 TCID 50 /mL for the LIAISON ® SARS-CoV-2 antigen assay. Virus isolation is possible in only 20% of samples with virus concentrations within this range [10] . In addition, Corman et al. [3] suggested that a negative antigen test result obtained a week after the first symptoms is associated with loss of infectiousness. The LIAISON ® SARS-CoV-2 antigen assay could probably be useful on this aspect and could be the basis of lifting isolation in long-term hospitalized COVID-19 patients. However, more studies are necessary to correlate TCID 50 and viral loads with absence of infectiousness in vitro. Hence, testing in cell culture might be an interesting additional investigation. According to the manufacturer s instructions for the LIAISON ® SARS-CoV-2 antigen assay, it is necessary to perform the sample deactivation within 12 h after taking the sample. Practical knowledge showed that this is usually impossible in a clinical daily routine. To overcome this problem, we evaluated antigen stability over time and found that the antigen measurement remains stable over three days. For other antigens, stability over time has been shown before, e.g., the stability of the hepatitis B surface antigen over 12 months stored at −20 • C [11] . Apart from nasopharyngeal swab samples, which are most frequently used in the daily routine [12] , SARS-CoV-2 diagnostics includes the evaluation of other materials. Here, the LIAISON ® SARS-CoV-2 antigen assay performed suboptimally. It could not reliably detect SARS-CoV-2 in other materials, such as bronchoalveolar lavage, pharyngeal wash, or sputum, which is a limitation of this test in comparison to RT-PCR. Cross-reactivity toward other respiratory viruses was not observed. This underlines the reliability of the LIAISON ® SARS-CoV-2 antigen assay, since at the beginning of the pandemic, false positive results were generated by RT-PCR due to cross-reactivity [13] . In conclusion, the LIAISON ® SARS-CoV-2 antigen assay can reliably detect samples with high SARS-CoV-2 loads. Hence, the antigen assay might be a useful back-up system if performing RT-PCR is not an option (for example, due to lack of reagents). In addition, it might be a helpful tool for lifting the isolation of hospitalized patients. We included 182 nasopharyngeal swabs for the detection of SARS-CoV-2, tested from 16 November to 21 December 2020 at the Institute for Virology, University Hospital Essen, in our study. Our cohort consisted of individuals tested by the Public Health Department of Essen. The sample collection was performed when patients informed the Public Health Department of their COVID-19-specific symptoms. Generally, this was briefly after the onset of symptoms. In case of outbreaks, this also included asymptomatic individuals. The swab collection kits contained a viral transport medium, which was used for both the PCR and the detection of SARS-CoV-2 antigen, as described below. Detection of SARS-CoV-2 RNA was performed using the RealStar ® SARS-CoV-2 RT-PCR kit (Altona Diagnostics, Hamburg, Germany), which targets the SARS-CoV-2 genes S (spike) and E (envelope) for 155 samples and Alinity m SARS-CoV-2 Assay (Abbott, Wiesbaden, Germany), which targets the SARS-CoV-2 genes RdRp (RNA dependent RNA polymerase) and N (nucleocapsid) for 27 samples. For the Altona assay, the Ct values corresponding to the E-gene were taken into account for further analysis. Furthermore, we tested two SARS-CoV-2 standards provided by INSTAND (Berlin, Germany) of 10 6 and 10 7 SARS-CoV-2 copies/mL, which corresponded to Ct values of 23.11 (E-gene, Altona) and 22.42 (Alinity) and 20.18 (E-gene, Altona) and 18.67 (Alinity), respectively. We diluted the 10 6 standard and measured the Ct values corresponding to 10 5 and 10 4 copies/mL, amounting to Ct 26.3 and 31.4 (E-gene, Altona) and 25.1 and 28.74 (Alinity), respectively. Linear regression was performed for both series of values, and the viral load in copies/mL was calculated based on the equation deriving from it. SARS-CoV-2 antigen detection was performed using the LIAISON ® SARS-CoV-2 Ag assay (Diasorin, Saluggia, Italy), a chemiluminescence sandwich-immunoassay (CLIA)based technology for the determination of nucleocapsid antigen from SARS-CoV-2 samples in upper respiratory specimens. Samples positive for endemic coronaviruses (HCoV NL63, HCoV OC43, HCoV 229E, HCoV HKU), influenza A and B, RSV, and samples from patients with detectable EBV in blood were tested using the LIAISON SARS-CoV-2 Ag assay to evaluate cross-reactivity. The cut-off for sample positivity given by the manufacturer is 200 TCID 50 /mL. The assay was performed according to the manufacturer s instructions with two notable exceptions. For the inactivation of the samples, we mixed a 400 µL sample with a 400 µL inactivation buffer, instead of using 1 mL plus 1 mL. Many swab sample collection kits do not contain enough viral transport media to allow for the testing of the sample with the antigen assay, RT-PCR, and a potential retesting (for example, in case of inhibition or pipetting errors). The applied volume was within the range of the minimum sample volume given by the manufacturer (100 µL sample + 300 µL dead volume). Furthermore, we used samples taken up to 36 h before inactivation, since this reflects laboratory and clinical routine more accurately. The samples were cooled during transport and storage. The ethics committee of the medical faculty of the University of Duisburg-Essen approved the assessment of test samples for the improvement of diagnostic procedures (20-9512-BO). Statistical analysis was performed using SPSS software (v23, SPSS Inc., Chicago, IL, USA), GraphPad Prism 6.0 (GraphPad, CA, USA) and the platform VassarStats (http://vassarstats.net (accessed on 23 March 2021). Two-tailed p values less than 0.05 were considered to be statistically significant. Informed Consent Statement: Assessment of test samples for the improvement of SARS-CoV-2 diagnostic procedures has been approved by the ethics committee of the medical faculty of the University of Duisburg-Essen. Under the above-mentioned circumstances, patient consent was waived. The data presented in this study can become available in part on request from the corresponding author. The data are not publicly available due to data protection issues. Nanozyme chemiluminescence paper test for rapid and sensitive detection of SARS-CoV-2 antigen COVID-19) in Suspected Human Cases: Interim Guidance Comparison of seven commercial SARS-CoV-2 rapid point-of-care antigen tests: A single-centre laboratory evaluation study Youden Index and optimal cut-point estimated from observations affected by a lower limit of detection The accuracy of novel antigen rapid diagnostics for SARS-CoV-2: A living systematic review and meta-analysis Fast detection of SARS-CoV-2 RNA directly from respiratory samples using a loop-mediated isothermal amplification (LAMP) test. Viruses A colorimetric RT-LAMP assay and LAMP-sequencing for detecting SARS-CoV-2 RNA in clinical samples Antigen-Detection in the Diagnosis of SARS-CoV-2 Infection Using Rapid Immunoassays: Interim Guidance Clinical validation of quantitative SARS-CoV-2 antigen assays to estimate SARS-CoV-2 viral loads in nasopharyngeal swabs Virological assessment of hospitalized patients with COVID-2019 Stability of hepatitis B surface antigen over time: Implications for studies using stored sera SARS-CoV-2 RNA identification in nasopharyngeal swabs: Issues in pre-analytics Laboratory testing methods for novel severe acute respiratory syndrome-coronavirus-2 (SARS-CoV-2). Front We acknowledge support by the Open Access Publication Fund of the University of Duisburg-Essen. We also thank Birgit Goitowski, Jessica Ackermann, Fabian Wenskat and Dominik Wermann for their excellent technical support. The authors declare no conflict of interest.