key: cord-0779491-0epadkr7 authors: Levine, C.; Grady, C.; Block, T.; Hurley, H.; Russo, R.; Peixoto, B.; Frees, A.; Ruiz, A.; Alland, D. title: Use, reuse or discard: quantitatively defined variance in N95 respirator integrity following vaporized hydrogen peroxide decontamination during the COVID-19 pandemic date: 2020-08-21 journal: medRxiv : the preprint server for health sciences DOI: 10.1101/2020.08.18.20177071 sha: 31d4b691953c97d3dadd94b8f4bd070fb6893b4f doc_id: 779491 cord_uid: 0epadkr7 Background: COVID-19 has stretched the ability of many institutions to supply needed personal protective equipment, especially N95 respirators. N95 decontamination and reuse programs provide one potential solution to this problem. Unfortunately, a comprehensive evaluation of the effects of decontamination on the integrity of various N95 models using a quantitative fit test (QTFT) approach is lacking. Aims: 1) To investigate the effects of up to eight rounds of vaporized H2O2 (VHP) decontamination on the integrity of N95 respirators currently in use in a hospital setting. 2) To examine if N95 respirators worn by one user can adapt to the face shape of a second user with no compromise of integrity following VHP decontamination. Methods: The PortaCount Pro+ Respirator Fit Tester Model 8038 was used to quantitatively define the integrity, measured by fit, of N95 respirators following decontamination with VHP. Findings: There was an observable downward trend in the integrity of Halyard Fluidshield 46727 N95 respirators throughout eight cycles of decontamination with VHP. The integrity of 3M 1870 N95 respirators was significantly reduced after the respirator was worn, decontaminated with VHP, and then quantitatively fit tested on a second user. Furthermore, we uncovered inconsistencies between qualitative fit test and QTFT results that may have strong implications on the fit testing method used by institutions. Conclusions: Our data revealed variability in the integrity of different N95 models after VHP decontamination and exposed potential limitations of N95 decontamination and reuse programs. 49 50 51 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 21, 2020. . https://doi.org/10.1101/2020.08.18.20177071 doi: medRxiv preprint As of August 2020, over 21.7 million people have been infected and more than 770,000 53 have died from COVID-19 worldwide [1] . Healthcare workers are at high risk of contracting [14] , or ultraviolet 64 germicidal radiation [15] for decontamination of N95 respirators. The integrity of decontaminated 65 N95 respirators, measured by fit, must be maintained following processing. Proper fit is defined 66 as an intact, airtight seal against the user's face that can be measured using either a qualitative 67 fit test (QLFT) or a quantitative fit test (QTFT), as defined by Appendix A of the OSHA Standard 68 1920.134 [16] . A QTFT more accurately identifies proper fit than a QLFT [17, 18] . However, an 69 N95 respirator that is tested using a QTFT cannot be subsequently worn for future protection 70 because this procedure requires puncturing a hole into the respirator to assess fit. On the other 71 hand, N95 respirators examined via a QLFT can be kept and used by the wearer after the test is 72 completed. Current decontamination studies only use a manikin head form to quantitatively assess fit 74 factor and fail to evaluate all respirator models used in hospital settings [19, 20] . However, product 75 scarcity has required the use of many different N95 models, most of which have not been 76 evaluated after decontamination. Additionally, it can be challenging to develop and maintain 77 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 21, 2020. . https://doi.org/10.1101/2020.08.18.20177071 doi: medRxiv preprint systems which ensure that respirators are returned to their original user, as opposed to returning 78 the same model but from a different user post-decontamination. Yet, the integrity of N95 79 respirators worn by multiple persons has not been investigated. Here, we first evaluated the number and type of N95 respirators qualitatively fit tested 81 and distributed by University Hospital (UH) in Newark, NJ before and during the COVID-19 82 pandemic. We then used QTFTs to assess the integrity of 4/7 of these N95 models following 83 sequential rounds of decontamination with VHP. The integrity of decontaminated N95 84 respirators was then tested on a second user using a QTFT approach. Finally, we evaluated the 85 reliability of qualitative fit testing on models that we found were hard to fit quantitatively by 86 comparing QLFT vs. QTFT results. Our findings revealed that the majority of N95 respirators 87 evaluated were able to withstand the VHP decontamination process and that these N95 88 respirators could be returned to new users with no significant decrease in integrity. However, we 89 did observe exceptions to these findings that may have strong implications on which respirator 90 models are suitable for decontamination and reuse programs. Furthermore, differences in QLFT 91 and QTFT results for certain respirator models uncovered potential weaknesses in using QLFTs 92 for measuring the protective ability of N95 respirators. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 21, 2020. . https://doi.org/10.1101/2020.08.18.20177071 doi: medRxiv preprint 5 first, 2) QLFT users were blinded to the order by which qualitative testing was administered 104 (sweet followed by bitter). All users were deemed medically able to complete fit testing 105 beforehand. Second user N95 respirators. Respirators that had previously failed a QLFT were termed 119 "lightly used" because even though the nosepiece had been fit and shaped to a user's face, the 120 respirator did not undergo the extended wear that is representative of a long hospital shift. These lightly used N95 respirators were used for second user experiments. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 21, 2020. . https://doi.org/10.1101/2020.08.18.20177071 doi: medRxiv preprint 6 To analyze the effect of decontamination on N95 integrity we used a one tailed Kruskal-Wallis 130 test. A one-tailed Mann-Whitney U test was used to assess the differences between N95 131 respirators worn by one user and N95 respirators worn by a second user following 132 decontamination. All statistical analysis was run using Prism 8 software. Respirator models and fit testing have increased during the COVID-19 pandemic in a 135 hospital setting. To define the scope of our targeted decontamination study, we examined the Variation in quantitative fit testing across different N95 models. Prior to any VHP 147 decontamination, we conducted QTFTs on the seven N95 models used at UH during the 148 COVID-19 pandemic (Figure 2 ). We defined a QTFT fit factor of >100 to be a passing value. Users who passed a QTFT were considered certified to use the N95 model tested. Respirator (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 21, 2020. . https://doi.org/10.1101/2020.08.18.20177071 doi: medRxiv preprint 7 representative in our volunteer population. However, for our studies we defined these last three 156 models as "hard-to-fit" and did not include them in the decontamination experiments. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (Table I) . Unfortunately, we were unable to find enough participants able to 196 qualitatively fit either Gerson 2130 or Cardinal Health N95 respirators and therefore they were 197 not included in this study (Table SI) (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 21, 2020. . https://doi.org/10.1101/2020.08.18.20177071 doi: medRxiv preprint 9 validation means health care workers could be at an increased risk of receiving inadequate 208 PPE. Thus, in this study we quantitatively evaluated the integrity of multiple N95 models 209 currently in use at UH following decontamination with VHP. From January to May 2020, UH increased qualitative fit testing over tenfold and 211 expanded the models of N95 respirators distributed compared to six years prior. Using unworn 212 N95 respirators, we examined the integrity of these models after decontamination with VHP and It is crucial that decontamination and reuse programs are able to rapidly turnaround 218 clean N95 respirators. One potential time saving approach is to return N95 respirators to new 219 users instead of having to sort and return respirators to their initial user. To assess the 220 adaptability of N95 respirators to a new face we quantitatively fit tested lightly worn respirators 221 on a second user following decontamination with VHP. Our results were limited to respirators 222 that were initially only worn for a QLFT and are thus not representative of the wear and tear 223 associated during an extended hospital shift. Despite these limitations, the QTFT values were 224 lower for 3M 1870 N95 respirators when they were lightly worn and fit tested by a second user 225 compared to the other respirator models tested. It is also possible that some respirators may be 226 less tolerant for reuse even by the same person, however we were unable to assess that in this 227 study. Although, others have found a significant association between number of shifts N95 228 respirators were worn by a single user and failed QLFTs [21] . These data bring to light a 229 significant obstacle for decontamination and reuse programs. A major unexpected result in this study was the inability to find many users who passed 231 QTFTs on Gerson 1730, Gerson 2130, and Cardinal Health N95 models, despite these 232 respirators representing 42% of passing QLFTs at UH. We therefore wondered if differences 233 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 21, 2020. . https://doi.org/10.1101/2020.08.18.20177071 doi: medRxiv preprint existed between QTFT and QLFT results specifically on these hard-to-fit models. Of the ten 234 participants we recruited for QLFTs, six were able to pass at least one of the tasting challenges 235 when qualitatively fit tested on Gerson 1730. Interestingly, only 1/6 of these participants was 236 able to pass on the Gerson 1730 using a QTFT. This discrepancy between QLFTs and QTFTs 237 may be attributed to taste insensitivity which has been shown to increase false positive fit Together these observations suggest the administration of QTFTs may be warranted for fit 242 testing these hard-to-fit models and should be used to assess inconsistencies between QLFT 243 on other N95 models. We acknowledge that this study has several limitations. All fit testing methods can have All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 21, 2020. All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 21, 2020. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 21, 2020. . https://doi.org/10.1101/2020.08.18.20177071 doi: medRxiv preprint An interactive web-based dashboard to track COVID-19 in real 277 time Transmission potential and severity of COVID-279 19 in South Korea Clinical Characteristics of 138 Hospitalized Patients With Novel Coronavirus-Infected Pneumonia in Wuhan, China No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity Decontamination of Surgical Face Masks and N95 Respirators by 309 Dry Heat Pasteurization for One Hour at 70°C A method to determine the available UV-C dose for the 312 decontamination of filtering facepiece respirators United States Department of Labor. Occupational Safety and Health Standards Personal 314 Protection Equipment: Respiratory Protection Comparison of 316 five methods for fit-testing N95 filtering-facepiece respirators Capability of respirator wearers to detect aerosolized qualitative fit 319 test agents (sweetener and Bitrex) with known fixed leaks. Applied occupational and 320 environmental hygiene Federal Drug Administration. Final Report for the Bioquell Hydrogen Peroxide 322 Vapor (HPV) Decontamination for Reuse of N95 Respirators Evaluation of five decontamination 325 methods for filtering facepiece respirators Correlation Between N95 Extended Use and Reuse and Fit Failure in an Emergency Department Fitting 331 characteristics of eighteen N95 filtering-facepiece respirators No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted