key: cord-0773774-25c68o09 authors: Arellano, Daniel H.; Tobar, Eduardo A.; Lazo, Marioli T.; Rojas, Veronica A.; Gajardo, Abraham I.J.; Montecinos, Nicolás; Regueira, Tomás; Cornejo, Rodrigo A. title: Assessment of a splitter for protective dual-patient ventilation in patients with acute respiratory distress syndrome date: 2022-02-21 journal: Br J Anaesth DOI: 10.1016/j.bja.2022.02.007 sha: afd9748768cbd179b0b6e2762654ee708f92c162 doc_id: 773774 cord_uid: 25c68o09 nan inspiratory resistance in one patient during dual-patient ventilation with a splitter has been only assessed on test lungs. [7] [8] [9] Shortages of mechanical ventilators during the COVID-19 pandemic prompted development of mechanical ventilators and splitter prototypes, including under the "A breath for Chile" initiative (sponsored by the Ministry of Sciences). In this study, we assessed the performance of the splitter after the electromedical safety inspection. Our objective was to evaluate dual-patient ventilation, without and with a splitter, when one subject develops sudden changes in respiratory mechanics, extreme air leaks, airway disconnection or airway occlusion in experimental and clinical assessments. First, dual-patient ventilation was used to ventilate two test lungs (IMT Analytics ® , SmartLung 2000), without and with a splitter, using a mechanical ventilator (PB 840, Medtronic ® ; manufacturer, location). Pressure-controlled mode was programmed to deliver a VT of 400 ml to each test lung at the study onset, with a respiratory rate Second, dual-patient ventilation with a splitter (NeyunSplit, DTS ® , city, Chile) was performed to ventilate both a test lung and a patient. Five patients older than 18 yr with COVID-19-associated acute respiratory distress syndrome (ARDS), haemodynamic stability and deep sedation, were included. The active humidification system was replaced by a heat and moisture exchanger filter (HMEF). The mechanical ventilator was set to pressurecontrolled mode to deliver a VT of 6 ml kg -1 to the patient and a similar VT to the test lung. Dual-patient ventilation without a splitter generated significant changes in VT, PEEP, plateau pressure and peak pressure of test lungs (p-value <0.001). However, dual-patient ventilation with a splitter did not generate changes in VT, and only minimal changes were observed in PEEP, plateau pressure and peak pressure ( Figure 1A Patient VT, PEEP, plateau pressure and peak pressure remained similar despite Raw/Crs combinations set in the test lung (p > 0.05, Figure 1B) . PaO2:FiO2, heart rate and mean arterial pressure remained unchanged during these assessments, and PaCO2 increased in 3 patients. Both patients synchronously ventilated through dual-patient ventilation with a splitter maintained stability in their ventilatory, haemodynamic and gas exchange parameters after 3 h of assessment ( Figure 1C ). When dual-patient ventilation was used to ventilate a patient and a test lung, the splitter was able to maintain stable and differentiated VT, PEEP and airway pressures in the patient independent of the changes in Crs and Raw of the test lung, even with air leak and occlusion manoeuvres. On the contrary, the same changes in respiratory mechanics during dual-patient ventilation without a splitter generated significant (clinically relevant) changes in VT, PEEP and airway pressure in lung simulators; VT was higher for the lung with better compliance and lower resistance. As in other reports 7-10 , a splitter was able to provide a defined and different VT to two patients despite differences in respiratory mechanics. Of note, we tested for the first time in humans whether sudden changes in lung compliance and/or airway resistance in one simulated patient (test lung) affect the other patient (ARDS patient), including extreme conditions such as massive leak and airway occlusion. The NeyunSplit allows adjustments to a flow limiting valve to keep the VT stable; notwithstanding the extreme changes in compliance and resistance, only two patients needed adjustments, achieving the desired VT in a few seconds. J o u r n a l P r e -p r o o f 5 The increase in PaCO2 was mainly due to the change from active humidification to HMEF by increasing dead space. 11 The VT and RR were not modified during the study. Under real conditions, it is feasible to make changes to compensate for the increase in PaCO2. When initiating dual-patient ventilation, higher initial set VT in the mechanical ventilator is required to ventilate two patients. A greater VT at high RR can induce entrapment and auto-PEEP by limiting expiratory time; this phenomenon can be compensated by adjusting the splitter flow limiter. Use of splitters requires meticulous assembly but can be potentially useful in pandemic scenarios. Improvised automatic lung ventilation for unanticipated emergencies A single ventilator for multiple simulated patients to meet disaster surge Multiplex Ventilation: A Simulation-Based Study of Ventilating 2 Patients with a Single Ventilator Ventilator sharing during an acute shortage caused by the COVID-19 pandemic One ventilator for two patients: feasibility and considerations of a last resort solution in case of equipment shortage A simulated single ventilator/dual patient ventilation strategy for acute respiratory distress syndrome during the COVID-19 pandemic Ventilator Sharing Using Volumecontrolled Ventilation during the COVID-19 Pandemic A rapidly deployable individualized system for augmenting ventilator capacity Regulating inspiratory pressure to individualise tidal volumes in a simulated two-patient, oneventilator system Ability and safety of a heated humidifier to control hypercapnic acidosis in severe ARDS plateau pressure (PPL), peak pressure (Ppeak), and tidal volume showing changes during dual-patient ventilation (DPV) under different combinations of airway resistance (Raw) and compliance (Crs) (R5/C75, R5/C60, R5/C25, R20/C75, R20/C60, and R20/C25), and during air leak or airway obstruction. In A, DPV with or without a splitter was delivered to two test lungs (lung A [dashed line] and B [solid line]). In B, DPV with a splitter was delivered to a patient (dashed line) and a test lung (solid line). In C, DPV with a splitter was delivered to two patients (dashed and solid lines) for 3 h We thank Diego Ugalde for his assistance in figure design. The authors declare no competing interests related to this work.