key: cord-0773262-ccjxdhpl authors: Gdoura, Mariem; Abouda, Imen; Mrad, Mehdi; Ben Dhifallah, Imen; Belaiba, Zeineb; Fares, Wasfi; Chouikha, Anissa; Khedhiri, Maroua; Layouni, Kaouther; Touzi, Henda; Sadraoui, Amel; Hammemi, Walid; Meddeb, Zina; Hogga, Nahed; Ben Fadhel, Sihem; Haddad-Boubaker, Sondes; Triki, Henda title: SARS-CoV2 RT-PCR assays: In vitro comparison of 4 WHO approved protocols on clinical specimens and its implications for real laboratory practice through variant emergence date: 2022-03-28 journal: Virol J DOI: 10.1186/s12985-022-01784-4 sha: 8835ed8c72eea146da8e5b65ba5da3c68c0b74d0 doc_id: 773262 cord_uid: ccjxdhpl INTRODUCTION: RT-PCR testing on nasopharyngeal swabs is a key component in the COVID-19 fighting, provided to use sensitive and specific SARS-CoV2 genome targets. In this study, we aimed to evaluate and to compare 4 widely used WHO approved RT-PCR protocols on real clinical specimens, to decrypt the reasons of the diverging results and to propose recommendations for the choice of the genome targets. METHODS: 260 nasopharyngeal samples were randomly selected among the samples tested between Week-16, 2020 and week-16 2021, in the Institut Pasteur de Tunis, Tunisia, one of the referent laboratories of COVID-19 in Tunisia. All samples were tested by Charité, Berlin protocol (singleplex envelop (E) and singleplex RNA-dependent RNA polymerase (RdRp)), Hong Kong Universiy, China protocol (singleplex nucleoprotein (N) and singleplex Open reading frame Orf1b), commercial test DAAN Gene® (using the CDC China protocol), (triplex N, Orf1ab with internal control) and Institut Pasteur Paris protocol (IPP) (triplex IP2(nsp9) and IP4 (nsp12) with internal control). For IPP, a selection from samples positive by IP2 but negative with IP4 was re-tested by exactly the same protocol but this time in singleplex. New results were described and analyzed. RESULTS: In vitro analysis showed discordant results in 29.2% of cases (76 out of 260). The most discordant protocol is DAAN Gene® due to the false positive late signals with N target. Discordant results between the two protocol’s targets are more frequent when viral load are low (high Ct values). Our results demonstrated that the multiplexing has worsened the sensitivity of the IP4 target. CONCLUSION: We provide concise recommendations for the choice of the genome targets, the interpretation of the results and the alarm signals which makes suspect a gene mutation. The Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV2) is responsible for the coronavirus disease . Rapidly after its emergence at the end of 2019 in Wuhan, China, COVID-19 has being spread globally and becoming a massive pandemic [1, 2] . COVID-19 pandemic has provided a great recognition to the key role played by the laboratories in the diagnostic testing; for screening, monitoring and contact-tracing [3] . Indeed, molecular testing is the angular stone to prevent and control virus circulation: it is reliable, rapid and accurate. Through this pandemic, millions of RT-PCR tests were carried out daily, which demonstrate how suitable it is for large scale testing. As of 29 September 2021, 244.69 per 1000 inhabitants were tested by RT-PCR in Tunisia (total of 2.92 millions) which is a high rate as compared to other countries [4] . Soon after the emergence of the SARS-CoV2, the first full sequence was published on January 10th, 2020 and the first RT-PCR detection assay was published on January 23rd, 2020 [5] . Today, whole panoply of protocols and kits is available within reach of all laboratories; many other protocols have been published by research groups, number of them has been approved by the WHO in March 2020 and a huge number of commercial tests have been created by companies [6, 7] . These tests are diverse by targeting a range of SARS-CoV2 specific genome regions. Obviously, genome targets impact the analytical performances; the sensibility of detection by RNA copy per milliliter and the specificity to detect unambiguously the SARS-CoV2 [3] . The biggest challenge for laboratories was to have a sufficiently sensitive RT-PCR assay to be able to detect the virus in pre-symptomatic individuals, who harbor often low viral loads [8] . Oppositely, high sensitivity may lead to false positive results especially for patients having recovered from SARS-CoV2 infection, due to genomic debris but no viable virus [9] . Many authors focused on the evaluation and the improvement of available RT-PCR protocols and tests, but especially for commercial tests. In this study, we aimed to evaluate and to compare 4 widely used WHO approved RT-PCR protocols on real clinical specimen, to decrypt the origins behind the diverging results and to propose recommendations for the choice of genome targets. We focused on the impact of the SARS-CoV2 Alpha variant of cocern (VOC) detection by the Institut Pasteur Paris protocol and how to manage it. The Laboratory of Virology of the Institut Pasteur de Tunis, Tunisia, is one of the national referent laboratories for the molecular and serological diagnosis of SARS-CoV2 since March 2020 and one of the national referent laboratories for the SARS-CoV2 variant emerging surveillance in Tunisia. From the bio-bank that contains remnant and retrospective clinical samples previously collected from individuals suspected to have, or diagnosed with SARS-CoV2 and stored at − 80 °C, we randomly selected 260 nasopharyngeal samples. These samples were collected between 13/04/2020 and 19/04/2021 (Week-16 2020 and week-16 2021), covering a year of virus circulation in Tunisia, through which, Alpha VOC was introduced. All samples underwent, first, RNA extraction using the QIAamp Viral RNA Mini Kit (QIAGEN ® , Germany) then were tested, during 2 months period (May-June 2021)by the 4 WHO approved protocols compared in the present study [7] : Charité, Berlin protocol (singleplex envelop (E) and singleplex RNA-dependent RNA polymerase (RdRp), Hong Kong Universiy, China protocol (singleplex nucleoprotein (N) and singleplex open reading frame (Orf1b)), CDC China protocol (commercial test DAAN Gene ® , (triplex N, Orf1ab with internal control) and Institut Pasteur Paris protocol (triplex IP2 (nsp9) IP4 (nsp12) with internal control). (for the rest of the manuscript they will be called BERLIN, HKU, DAAN Gene ® and IPP, respectively). Between different RT-PCR tests, the extracted RNAs were stored at + 4 °C for periods less than 24 h and at − 80 °C for longer periods. The interpretation of the results was done according to the respective protocol's instructions. Interpretation took into consideration both the Cycle threshold (Ct) value and the shape of the obtained curves. Overall, when both targets per protocol gave positive amplification (Ct value ≤ 39), the result was positive, when both gave negative amplification (Ct value > 39), the result was negative, however, when just one target per 2 gave a positive amplification, the result was retained as "partial positive". Each protocol results were described and analyzed. We conducted a comparison between the 4 WHO approved protocols taking into consideration at least one positive target per protocol. Concordant and discordant results were rigorously classified and deeply analyzed. Then we compared the 8 targets independently of the protocols. A selection from "partial positive" samples was retested by exactly the same protocol but this time in singleplex. New results were described and analyzed. The MedCalc ® V18.2.1 Software was used for comparative analysis of the final results with the different RT-PCR protocols. The agreement between the different protocols' results was determined by the Cohen's Kappa coefficient. The concordance between two observations increases when Kappa coefficient is closer to 1. The concordance between the targets' results was evaluated as well as the Ct values pairwise comparisons among the 4 protocols. The correlation between the Ct values of different targets was determined by the Spearman rank correlation coefficient. The calculation of the mean, median, standard deviation, maximum and minimum Ct values for each target was also performed. For the comparison of the two targets in each protocol, the t-test was used for the pairwise comparison of their Ct values. A p-value equal to or less than 0.05 was considered as statistically significant. For the 260 samples, results for each protocol were represented in Table 1 . The protocol that gave the most positive results was DAAN Gene ® (55% positive results), the one that gave the most negative results was HKU (50.4% negative results) and the one that gave the most partial positive results was IPP (16.9%). Ct values for both targets of DAAN Gene ® and BERLIN were different (p < 0.0001) the mean of difference was 1.5 ± 2 and 2.3 ± 2 respectively. The Ct value of BERLIN E was often lower than the BERLIN RdRP one and DAAN Gene ® N was often lower than DAAN Gene ® Orf 1ab. However, HKU and IPP gave similar Ct values for both targets (p > 0.001). For "partial positive" results, the targets that gave usually positive signals were E for BERLIN, N for HKU, N for DAAN and IP2 for IPP. In these partial positive results, the Ct values of the positive target were high for all the cases ( Ct value mean higher than 33.5) except for the IP2 (Ct value mean equal to 25.2). Concordance and different groups of discordant results were represented in Table 2 The 76 remaining samples (29.2%) gave different discordant results being totally negative by at least one protocol and clearly positive by at least one protocol. The Ct values of positive targets for this group ranged from 13 to 39 (median = 36, mean = 33 ± 8.9) and were significantly higher than those of the concordant group (p < 0.0001). These 76 samples divided into 3 groups (Table 2) : Group D1 contained 21 samples that gave positive results with 3 out of 4 protocols, almost equally distributed between the 4 protocols: HKU, IPP, DAAN Gene ® and BERLIN failed to detect 7, 6, 4 and 4 samples, respectively. Group D2 contained 44 samples that gave negative results with 3 out of 4 protocols and a positive result with only one protocol (DAAN gene in most of cases, 43 out of 44). Group D3 contained 11 samples that gave a positive result by two protocols and a negative result by two protocols. To further focus on concordance between protocols, we calculated agreement between the results as represented in Table 3 . The strength of agreement was very good and similar for the couples BERLIN/HKU, IPP/ BERLIN and IPP/HKU, however, agreement was moderate between DAAN Gene ® and the three other protocols. In order to evaluate the 8 targets regardless of their respective protocols, the Ct values for each target for the discordant results were represented in Fig. 1 . The lowest Ct value was obtained by DAAN Gene ® N target (Ct = 13) followed by the IP2 target (Ct = 15). The lowest means were those of IP2 (Ct = 27) followed by HKU N and HKU Orf 1b (Ct = 28). Generally, when the samples were amplified by the 8 targets, the samples with low viral load had high Ct values with all of them and those with high viral load gave low Ct values with all of them. The mean variation was of 7Ct ± 2.7 for each sample. A big panel of discordant patterns was found between the 8 targets. The most interesting patterns were the divergent DAAN Gene ® N and the IP4 from all others. The DAAN Gene ® N target gave positive signal for 28 cases (Ct value median of 38 [36-39]) whereas the 7 other targets were negative. DAAN Gene ® N target had a poor agreement with all other targets ( In the present study, we provided a concise description of discordant results between 4 WHO approved RT-PCR protocols to detect the SARS-CoV2 genome: BERLIN, HKU, DAAN gene and IPP, through testing 260 real clinical specimens. The DAAN gene ® commercial test used the same CDC China protocol approved by the WHO. In vitro analysis showed discordant results in 29.2% of cases (76 out of 260). The most discordant protocol is DAAN Gene ® due to false positive late signals with N target. Discordant results between the two protocol's targets are more frequent when viral load are low (high Ct values). Our results demonstrated that the multiplexing has worsened the sensitivity of the IP4 target. We provide concise recommendations for the choice of the targets, the interpretation of the results and the alarm signals which makes suspect a gene mutation. Globally, molecular testing by PCR has revolutionized the diagnosis of infectious diseases. In the context of the COVID-19 pandemic, high performing tests has allowed it to identify infected people and to decide of their discharge. Different targets were proposed and validated, classically; the preferred targets of pathogens include the conserved specific genes as the nonstructural genes like RdRp or genes that are expressed abundantly such as the structural S and N genes. In our study, the protocols were approved since March 2020 by the WHO. Many authors studied these protocols to focus on sensibility and specificity by in vitro or in silico studies [10] [11] [12] . Our study explored in vitro clinical specimen. Importantly, our study used a large number of clinical specimens from patients with confirmed COVID-19, oppositely, the majority of published studies explore cell culture supernatant or RNA transcripts, or sometimes a very small number of real samples collected in a precise time [13] . In our study we found that 29.2% of tested samples gave discordant results (76 out of 260), occurring more often with high Ct values and probably due to false positive amplifications. Sule et al. suggested that the Ct values over than 28 are probably related to non-specifically precipitated sequences due to an inactivation of the Taq polymerase and proposed that Ct values > 33.33 or 35, or ≥ 39.2 or 40 could be considered as negative [14] . In our study, Ct value less than 40 is considered positive, according to the respective protocol authors. Among all discordant cases, false negative results represent the most disturbing cases, for our study we obtained 21 samples positive by 3 protocols but one (Group D1 Table 2 ). This is problematic because not only it may underestimate the COVID-19 incidence but, perhaps more acutely, will lead to infectious individuals remaining as a source of infection in the community and undermine the effectiveness of infection control measures. Clinicians should not hesitate to re-sample the high suspected patients when laboratories return negative result. The N gene target was the most problematic. From the one hand, this is due to the false negative results caused by mismatches with the primers and probes [15] . Thanks to many in vitro and in silico analysis, researchers around the world are tracing the ongoing evolution of the N gene and demonstrated that it was particularly prone to mutations, more than all other targets. These findings affected the HKU, CDC China and Japan NIID N targets [11, [15] [16] [17] . Wang et al. concluded in 14 september 2020 that the N gene is the most non conservative gene giving non uniform performances between different primers and probes, thus, the N gene may not be an optimal [11] . This has become more preoccupant after the emergence of the Alpha VOC that caused a N gene dropout and N gene Ct value shift comparing to the wild strain [18] . For the other hand, N gene target was widely criticized to give persistent late signals for convalescent patients, which make confusion between false positivity or low amount of virus, related to convalescent patients [19, 20] . More importantly, this caused a dilemma regarding the discharge of isolation policies making the WHO canceling the need to 2 negative RT-PCR results [21] . Moreover, soon after the publication of the United States of America CDC protocol, Lee et al. declared that the N gene target should be canceled and disused because of false positive reactivity of N3 [22] This was found in our study for the DAAN gene ® N target with late Ct values, which underlines that late positive signal for N gene alone is more likely to be false positive and needs to be interpreted with caution. In our series, the HKU N target didn't show so many false positive results; this would be related to the primers and probes design as they do not target the same region in the N gene. Since its publication in January 2020, the BERLIN protocol was widely used worldwide [5] . It was reported that the E gene was the most sensitive and the most conservative target [11, 23] apart from some mutations affecting its sensitivity leading to false negative results by a commercial test using this protocol [10] . BERLIN protocol was criticized for the low sensitivity of the RdRp target which was proposed to confirm all BERLIN E gene positive results [24, 25] It is obvious that multiple RT-PCR reactions require more reagents, controls, thermo cyclers and labor, not adapted to a pandemic context. Multiplexing offers the possibility of two or more target detection by just one reaction, which have become attractive thanks to reducing significantly reagent consumption and time. Many authors suggested and demonstrated many successful . A large in silico study based on the theoretical production of RT-PCR signals by the SCREENED software, has evaluated the impact of the Alpha VOC on the RT-PCR protocols used in our study; the authors detected only one mutation (C12778T) in the Alpha VOC within the IPP IP2 target amplicon that has no impact on its sensitivity, as it is not located in the primers and probes annealing sites [33] . Our study argued the need of using at least two independent virus key regions to avoid the false positive and negative results. The N gene is better to be avoided or should be interpreted with much caution. Otherwise, N late signals should be confirmed by testing another non structural region. The E target is the best one in terms of specificity; it presents the best agreement with non structural targets like IP2 and HKU Orf 1b with significant lower Ct values which may substitute the BERLIN RdRp to confirm positive E samples. Biologists should be aware about some alarm signals that should indicate close monitoring and investigating technical or molecular causes i.e. when using IPP or HKU protocol, any divergence between targets of more than 2 Ct values, when one target gives high positive signal while the other is totally negative and when a gene dropout is obtained for gene multiplex assays. Here, re-testing in singleplex should be performed as well as indicating sequencing. As the introduction of variants has worried the researchers about the reliability of RT-PCR protocols, which were established at the beginning of 2020 before the emergence of variants mutations, we recommend that all laboratories perform regular in silico analysis in order to assess the test performance. This could be limited for laboratories using commercial tests that do not specify exactly the primers and probes used. In conclusion, the data presented in this study show the importance of regularly assessing the used RT-PCR protocols especially in the context of new emerging variants. Our findings emphasize that although WHO approved, RT-PCR protocols should be evaluated by each laboratory. Molecular testing by RT-PCR is a double trapped weapon that any mis-interpretation may lead to false positive or negative results, impacting the COVID-19 surveillance. A pneumonia outbreak associated with a new coronavirus of probable bat origin Chronologie de l'action de l'OMS face à la COVID-19 COVID-19 diagnosis-a review of current methods Coronavirus pandemic (COVID-19). Our World Data Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR The diagnostic accuracy of seven commercial molecular in vitro SARS-CoV-2 detection tests: a rapid meta-analysis False-negative results of initial RT-PCR assays for COVID-19: a systematic review Positive RT-PCR test results in patients recovered from COVID-19 A recurrent mutation at position 26340 of SARS-CoV-2 is associated with failure of the E gene quantitative reverse transcription-PCR utilized in a commercial dual-target diagnostic assay Mutations on COVID-19 diagnostic targets The emerging concern and interest SARS-CoV-2 variants Improved Molecular Diagnosis of COVID-19 by the Novel, Highly Sensitive and Specific COVID-19-RdRp/Hel Real-Time Reverse Transcription-PCR Assay Validated In Vitro and with Clinical Specimens Real-time RT-PCR for COVID-19 diagnosis: challenges and prospects Sequence mismatch in PCR probes may mask the COVID-19 detection in Nepal Identification of a polymorphism in the N gene of SARS-CoV-2 that adversely impacts detection by reverse transcription-PCR A novel point mutation in the N gene of SARS-CoV-2 may affect the detection of the virus by reverse transcription-quantitative PCR SARS-CoV-2 N gene dropout and N gene Ct value shift as indicator for the presence of B.1.1.7 lineage in a commercial multiplex PCR assay Detection of SARS-COV N2 gene: very low amounts of viral RNA or false positive? False negative of RT-PCR and prolonged nucleic acid conversion in COVID-19: rather than recurrence Criteria for releasing COVID-19 patients from isolation: scientific brief Analysis of the initial lot of the CDC 2019-Novel Coronavirus (2019-nCoV) real-time RT-PCR diagnostic panel Comparative performance of SARS-CoV-2 detection assays using seven different primer-probe sets and one assay kit Letter to the editor: SARS-CoV-2 detection by real-time RT-PCR. Euro Surveill Bull Eur Sur Mal Transm Eur Commun Dis Bull Authors' response: SARS-CoV-2 detection by realtime RT-PCR Development and evaluation of a duo SARS-CoV-2 RT-qPCR assay combining two assays approved by the World Health Organization targeting the envelope and the RNA-Dependant RNA Polymerase (RdRp) coding regions Triplex real-time RT-PCR for severe acute respiratory syndrome coronavirus 2 Enhanced throughput of the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) realtime RT-PCR panel by assay multiplexing and specimen pooling Covid-19 Réactifs/Evaluations. Société Française de Microbiologie Sequencing using a two-step strategy reveals high genetic diversity in the S gene of SARS-CoV-2 after a high-transmission period in Tunis Molecular Epidemiology of SARS-CoV-2 in Tunisia (North Africa) through Several Successive Waves of COVID-19 Deepening of in silico evaluation of SARS-CoV-2 detection RT-qPCR assays in the context of new variants Publisher's Note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations The authors gratefully acknowledge The Ministry of Health for logistical support and Mr. Ahmed Chakroun for technical advises. Authors' contributions MG: conceptualization, data curation, formal analysis, methodology, project administration, supervision, validation, writing original draft. IA: data curation, formal analysis, supervision, validation, visualization, writing original draft. MM: investigating, data curation, validation. IBD: data curation, formal analysis, visualization. ZB: data curation, formal analysis, visualization. WF: investigation. AC: investigating. MK: investigation. KL: investigation. HT: investigation. AS: investigation. WH: investigation. ZM: investigation. NH: investigation. SBF: investigation, data curation. SHB: investigation and review and editing of the final draft. HT: conceptualizationdata curation, methodology, writing, reviewing. All authors read and approved the final manuscript. This study was funded by the Tunisian Ministry of Higher Education and Scientific research (PRFD5P1-LutteCovid19, LR20IPT10). Not applicable. The samples were collected as part of diagnosis activity. The use of these samples was approved by the Institutional Review Board of Pasteur Institute of Tunis under ID number: 2020/27/I/LR16IPT. According to the approval and as samples were fully anonymized, the Bioethics committee waived the requirement for informed consent. Not applicable. Not applicable.