key: cord-0772080-8l8vvcw9 authors: Ahberg, Christian D.; Manz, Andreas; Neuzil, Pavel title: Single Fluorescence Channel-based Multiplex Detection of Avian Influenza Virus by Quantitative PCR with Intercalating Dye date: 2015-06-19 journal: Sci Rep DOI: 10.1038/srep11479 sha: 8029841b3e6b0267188b9c527b0e02dc472b9fdc doc_id: 772080 cord_uid: 8l8vvcw9 Since its invention in 1985 the polymerase chain reaction (PCR) has become a well-established method for amplification and detection of segments of double-stranded DNA. Incorporation of fluorogenic probe or DNA intercalating dyes (such as SYBR Green) into the PCR mixture allowed real-time reaction monitoring and extraction of quantitative information (qPCR). Probes with different excitation spectra enable multiplex qPCR of several DNA segments using multi-channel optical detection systems. Here we show multiplex qPCR using an economical EvaGreen-based system with single optical channel detection. Previously reported non quantitative multiplex real-time PCR techniques based on intercalating dyes were conducted once the PCR is completed by performing melting curve analysis (MCA). The technique presented in this paper is both qualitative and quantitative as it provides information about the presence of multiple DNA strands as well as the number of starting copies in the tested sample. Besides important internal control, multiplex qPCR also allows detecting concentrations of more than one DNA strand within the same sample. Detection of the avian influenza virus H7N9 by PCR is a well established method. Multiplex qPCR greatly enhances its specificity as it is capable of distinguishing both haemagglutinin (HA) and neuraminidase (NA) genes as well as their ratio. SYBR Green-based format is non-specific. To determine the specificity of the PCR products using SYBR Green, a subsequent melting curve analysis 8 (MCA) has to be implemented. MCA is often preferred over CE as it is performed in the PCR system immediately once the PCR is completed by sweeping samples' temperature while monitoring amplitude of fluorescence and thus no sample manipulation is required. Multiplex quantitative PCR (qPCR) methods based on Taqman probes as well as FRET-based systems have been demonstrated 9 . Currently, it is routinely done with probes, such as the popular FITC, JOEL, ROX and Cy5, using multiple optical channels. Each channel requires its corresponding light source, filter set and a detector. Can multiplex qPCR be conducted in one channel, while demultiplexing the results in real-time? Probe-based systems cannot be used as there is no technique capable of distinguishing the PCR products in a real time. On the other hand intercalating dye based PCR has shown promising results as the products can be demultiplexed using melting curve analysis (MCA) once the PCR is completed 10 . However, this method can determine serotype of the DNA or RNA but does not provide quantitative information. Continuous monitoring of the amplitude of the fluorescent signal is a powerful technique with a number of different applications 11 . One of them was product differentiation during PCR where a number of MCAs were performed individually after several thermal cycles of PCR 12 . In this contribution we propose and demonstrate a method to dynamically extract melting curves within each thermal cycle of a real-time PCR based on a single intercalating dye. This was accomplished by processing captured data without changing the PCR protocol, resulting in a set of 40 MCAs. From this set we were able to demultiplex quantitative data for different segments of DNA. The proposed method allows multiplex internal positive controls using a single intercalating dye and is an alternative for probe based systems used to detect HA to NA gene ratios. It employs only a single fluorescent optical channel. Typical MCA for intercalating dye-based PCR is performed by scanning the temperature of the sample at the rate of 1 °C/s or lower while recording the corresponding amplitude of fluorescence signal. The slow scanning rate is an important factor as every system with temperature control such as PCR exhibits a delay between temperature at the sensor and the sample. PCR systems can have a delay as long as a few seconds or more which is why slow temperature scanning for precise determination of a melting curve is essential. During PCR the sample is cycled between different temperatures. Especially interesting is the transition from extension to denaturation as it covers the range of expected melting temperatures. All that is needed, then, is to detect the fluorescence amplitude from the sample and determine the corresponding temperature during this phase. The first task has been demonstrated earlier by continuous fluorescence monitoring during the PCR cycling 11 . However the precise temperature monitoring of the sample during temperature transitions is not straightforward as there is discrepancy between sample and heater temperature as explained above. Is there another way to monitor actual sample temperature during ramping from annealing step to denaturation step? We believe there is a method to measure the sample temperature with high precision. This temperature is given by a thermal control system typically using proportional integrated derivative (PID) method of feedback, the thermal mass of the sample H and the thermal conductance G between the heating block and the sample. The time constant τ of the temperature delay is given by τ = H/G. Values of H, G and PID constants do not change during the PCR process so the temperature profile of the sample is a repetitive function of time during the PCR cycling. This implies that it is sufficient to determine the temperature profile of the sample only during a single cycle. We have used a modified version (see Fig. 1 ) of the virtual reaction chamber (VRC) system for this experiment 13 . The heat transfer and the sample temperature profile was simulated by finite element analysis (FEA) and the results were experimentally verified 14 . The model gave us a correlation between the temperature of the heater and the sample during transition from annealing step to denaturation step required for the MCA extraction. The real sample experiment exhibited faster response than the model probably due to convection in the aqueous sample decreasing the temperature gradient minimally in the transition period. This also provided a homogenous temperature distribution, crucial to the method, since it suppresses data smearing. Here we introduce a method of multiplex qPCR using a single fluorescence channel in the presence of Eva-Green intercalator. It is based on recording both temperature T and fluorescence F as a function of time t and cycle number n. From both functions [F = f(t,n) and T = f(t,n)] we have then eliminated time and generated fluorescence as function of temperature [F = f(T,n)] with PCR cycle number as a parameter. We have thus performed a melting curve analysis (MCA) during each PCR cycle. Subsequently we plotted the first negative derivative of fluorescence with respect to temperature as a function of temperature − dF/dT = f(T). For each amplicon a distinct melting temperature was obtained, with its amplitude corresponding to its concentration at each cycle. PCR amplification curves were then created by plotting the amplitude of the respective amplicon as function of cycle number. This way we can quantitatively detect one or more different amplicons using only a single intercalating dye. We only need to perform continuous fluorescence detection at an acquisition rate at or above 1 kHz, thus conventional reagents and protocols can be still used. The only prerequisite for the methods are a high enough data acquisition rate of fluorescence amplitude, a small sample volume and a sufficiently large difference in melting temperature of the genes for differentiation. Scientific RepoRts | 5:11479 | DOi: 10.1038/srep11479 Preparation of PCR mixture and thermal cycling. PCR mixture consisted of 6μ L Roche LightCycler TaqMan Master Mix (Roche Diagnostics, Germany), 1 μ L 20X EvaGreen Dye in water (TATAA, Sweden) as well as forward and reverse primers 15 (Table 1 )(MWG Eurofines, Germany) at a final concentration of 1.8 μ M. To this solution synthetic complementary DNA (cDNA) templates (ATG Biosynthesis, Germany) for haemagglutinin (HA) and neuraminidase (NA) for the avian influenza virus (H7N9) were added at different concentrations (see Table 1 ). Lastly the solution was adjusted to a volume of 20 μ L using water obtained from a Milli-Q ProgradT3 column (Millipore, Germany). A droplet with a volume of 150 nL of the PCR solution was placed on a hydrophobically coated microscope glass cover slip and covered with 1.5 μ L of mineral oil 9405 (Sigma Aldrich, Germany) thus forming a VRC. This glass was then placed on a micromachined silicon chip integrated with both a heater and a sensor, similar to a design shown earlier 13 . Here the silicon chip has a size of only 15 × 15 mm to better fit a handheld PCR device we are developing (see Fig. 1 ). The small thermal mass of the VRC together with the silicon heater resulted in heating rates > 20 °C/s and similar cooling rates achieved only by passive cooling. The PCR protocol consisted of a hot start for 10 min at 95 °C followed by 40 cycles each consisting of two steps, denaturation for 5 s at 95 °C and annealing/extension for 30 s at 60 °C. Fluorescence amplitude was continuously monitored using an Axiotron II microscope (Zeiss, Germany). We have used a blue LED model M470L3-C4 LED with principal wavelength of 470 nm (Thorlabs, Germany) for excitation. The LED was powered with square wave pulses at a frequency of 2710 Hz and a duty cycle of 5%. Light from the LED as well as emitted light from the specimen was filtered with a filter set model 49002 -ET -EGFP (FITC/Cy2) (Chroma Optical Corp, USA). Emitted light was captured by a PMT photosensor module H10722-20 (Hamamatsu Photonics K.K., Germany). The PMT signal was processed by a lock-in amplifier 7230 DSP (Ametek, USA) and its output recorded by a digital oscilloscope model DPO 3014 (Tektronix, Germany) with data rate of 2500 measurements per second. The temperature of the PCR chip was also recorded by the oscilloscope at the same data rate. For comparison, samples were also analyzed using the Roche LightCycler Carousel-Based system (Roche Diagnostics, Germany) followed by a standard melting curve after thermal cycling. Data analysis. Captured fluorescence data were analyzed using a custom written Matlab-script. First the data were filtered by a fast Fourier transform filter (FFTF). Subsequently, the fluorescence signal was divided into the individual cycles based on the captured temperature profile. The fluorescence amplitude during transition from extension/annealing step to denaturation step was then extracted for each cycle and fluorescence as function of temperature was plotted with cycle number as a parameter. MCA was formed by numerical differentiation of the fluorescence with respect to temperature and its negative value was plotted as a function of temperature. The melting curve captured at the first cycle was subtracted from all melting curves to suppress a background fluorescence effect we have observed. This background fluorescence could be caused by autofluorescence from the adjacent printed circuit board as well as fluorescence of the temperature dependence of the fluorophore as well as unspecific binding of the dye onto single stranded DNA 16 . PCR amplification curves were created by plotting amplitudes at the MCA of each amplicon as a function of the cycle number. We have run a 40 cycle PCR using the master mix manufacturer's specifications. A typical fluorescence profile was obtained (see Fig. 2A ). In Fig. 2B we show the single cycle number 23 to demonstrate the extracted data for the MCA. This procedure was applied to a sample containing only a single amplicon (HA) with melting temperature of 76 °C (see Fig. 3A,B) . We have then added to the sample a second amplicon (NA) with a melting temperature of 68 °C. A second peak corresponding to this amplicon can be observed in the respective MCA (Fig. 3C,D) . Increasing the amount of NA added compared to HA lead to an even faster increase of the magnitude of peak at 68 °C (Fig. 3E,F) . Extracting the fluorescence amplitude from the derived curves for each cycle at 68 °C and 76 °C respectively, results in two PCR amplification curves. Figure 4A shows the two amplification curves extracted for the experiments with a higher concentration of NA than HA (experiments A1 to A5 in Table 2 ). The sample spiked with the cDNA of the HA gene yielded only a single amplification curve with melting temperature of 76 °C, which was expected. For all other experiments we were able to extract two amplification curves, one at a temperature of 68 °C and the second one at temperature of 76 °C. Threshold cycles (C T ) were defined in the usual way. It was found that the difference in C T (∆C T ) of the two curves corresponded to the expected difference according to the ratio of concentrations of HA and NA. An overview of all conducted experiments is presented in Table 1 . The results indicate that ∆C T corresponds to the expected values. We can then conclude that the method can be used for multiplexed quantitative PCR. As an example, (see Table 1 ) sample A5 has the concentration of NA 19 times higher than the concentration of HA. With 100% PCR efficiency one would expect a ∆C T of 4.25 (log 2 19) . We have found the ∆C T value to be 4.0 ± 0.4 which is efficiency of 84%, reasonably close to the ideal value. We have observed MCA peak broadening in our experiment due to applied conditions. It is necessary to run several replicates of the same experiment to suppress random errors and improve the result precision. The MCAs are recorded at a scanning rate of over 20 °C/s while the commercial systems for high resolution melting curve analysis (HRMCA) 17 are usually operating with low temperature scan rates between 0.1 °C/s and 0.5 °C/s. Differentiation of amplicons with melting point differences ΔT m as small as 1.2 °C have been demonstrated in HRMCA. However the method presented in this contribution requires greater ΔT m due to peak broadening caused by the high heating rate. We assume that a difference in amplicons melting temperatures' of at least 5 °C would be sufficient to differentiate them from each other. This could be achieved by primer design since the melting temperature is a function of amplicon length and ratio of CG/AT content. Therefore an extension temperature of 60 °C and a denaturation temperature of 95 °C would allow for 6 different amplicons to be detected in the same droplet, assuming a ΔT m of 5 °C. Furthermore, the temperature measured by the sensor in HRMCA corresponds practically to the sample temperature due to the slow scanning rate. In our approach the sample temperature lags behind the temperature measured by the sensor underneath the sample 18 . This can be seen from the comparative end point HRMCA done with the LightCycler, Fig. 4B shows the fluorescence amplitude plotted as function of temperature with a temperature gradient of 0.1 °C/s. In Fig. 4C the first negative derivative of this curve with respect to temperature is shown, displaying two distinct peaks. Both melting temperatures found on the LightCycler are 8 °C higher than the temperatures we measured in our system. Nevertheless this T m offset has no influence on quality of the results especially when this lag is known and the resulting temperature offset is calibrated. Ideally the test should be performed with a clinical sample containing H7N9 virus starting with sample preparation followed with reverse transcription. VRC-based RT-PCR demonstrating almost all those steps was shown earlier, detecting both RNA from H5N1 avian influenza virus 19 as well as from virus of severe acute respiratory syndrome (SARS) 20 . Processing clinical sample containing virus of avian influenza or SARS would also require laboratory classified as biosafety hazard level 3 which we do not have available. In previous work we developed the system based on spiking blood sample with RNA, here in this work we spiked PCR master mix with two types of cDNA. We have developed a method for quantitative, multi target PCR using only a single intercalating dye and thus only one fluorescence channel. Our approach allows researchers to perform affordable multiplex PCR using simple tools as well as reagents. Another advantage compared to conventional methods is small consumption of reagents due to small sample volume. Fluorescent amplitude sampling at rate above 100 samples per second is, however, required for constructing MCA during transition from annealing/extension to denaturation. This method might be particularly interesting for determination of serotype of virus such as dengue fever or an avian influenza. Other potential applications include measurement of viral loads in a sample, differentiation between H7N9 and H7N5, and quantitative detection of co-infections. We have experimentally verified the method using a homemade PCR system primarily for convenience as we have 100% control of the tool. In principle any commercial real-time PCR device would be capable of performing the same protocol as long as it has available fluorescent amplitude sampling rate of 1kHz or higher and can process samples with small volume of a 1 μ L or below. Enzymatic amplification of beta-globin genomic sequences and restriction site analysis for diagnosis of sickle cell anemia Primer-directed enzymatic amplification of DNA with a thermostable DNA polymerase Kinetic PCR analysis: real-time monitoring of DNA amplification reactions Allelic discrimination by nick-translation PCR with fluorogenic probes The real-time polymerase chain reaction. Molecular aspects of medicine Analysis of ligase chain reaction products amplified in a silicon-glass chip using capillary electrophoresis Multicolor combinatorial probe coding for real-time PCR Theory of DNA melting curves Paired multiplex reverse-transcriptase polymerase chain reaction (PMRT-PCR) analysis as a rapid and accurate diagnostic tool for the detection of MLL fusion genes in hematologic malignancies Multiplex real-time PCR detection of Vibrio cholerae Continuous fluorescence monitoring of rapid cycle DNA amplification Product differentiation by analysis of DNA melting curves during the polymerase chain reaction Disposable real-time microPCR device: lab-on-a-chip at a low cost Nanoliter-sized Overheated Reactor Specific detection by real-time reverse-transcription reaction assays of a novel avian influenza A(H7N9) strain associated with human spillover infections in China Characterization of EvaGreen and the implication of its physicochemical properties for qPCR applications High-resolution melt analysis for the detection of a mutation associated with permethrin resistance in a population of scabies mites Non-contact fluorescent bleaching-independent method for temperature measurement in microfluidic systems based on DNA melting curves Catching bird flu in a droplet Clockwork PCR including sample preparation P. Neužil acknowledges partial financial support by Central European Institute of Technology (CEITEC), grant number CZ.1.05/1.1.00/02.0068. Authors would also like to acknowledge the editorial help of Eric Castro. C.A. ran the lab experiment and prepared all necessary reagents as well as the data processing in Matlab environment, A.M. helped with the experiment preparation and P.N. came with the original idea of realtime multiplexing using intercalating dye. All authors reviewed the manuscript. Competing financial interests: The authors declare no competing financial interests.