key: cord-0771390-cssss6zw authors: Khodare, Arvind; Padhi, Abhishek; Gupta, Ekta; Agarwal, Reshu; Dubey, Shantanu; Sarin, Shiv Kumar title: Optimal size of Sample Pooling for RNA Pool Testing: An Avant-Garde for Scaling up Severe Acute Respiratory Syndrome Coronavirus-2 Testing date: 2020-07-31 journal: Indian Journal of Medical Microbiology DOI: 10.4103/ijmm.ijmm_20_260 sha: 1b0c39cb46b5154154de01cea9295d8b4fad23c5 doc_id: 771390 cord_uid: cssss6zw Abstract Background and Objectives: Timely diagnosis is essential for the containment of the disease and breaks in the chain of transmission of severe acute respiratory syndrome coronavirus-2 (SARS-CoV-2). The present situation demands the countries to scale up their testing and design innovative strategies to conserve diagnostic kits and reagents. The pooling of samples saves time, workforce and most importantly diagnostic kits and reagents. In the present study, we tried to define the pool size that could be applied with acceptable confidence for testing. Materials and Methods: We used repeatedly tested positive clinical sample elutes having different levels of SARS CoV 2 RNA and negative sample elutes to prepare seven series of 11 pools each, having pool sizes ranging from 2 to 48 samples to estimate the optimal pool size. Each pool had one positive sample elute in different compositions. All the pools were tested by SARS CoV 2 reverse transcriptase quantitative polymerase chain reaction. Results: Out of the 77 pools, only 53 (68.8%) were found positive. The sensitivity of pools of 2–48 samples was decreased from 100% (95% confidence interval [CL]; 98.4–100) to 41.41% (95% CL; 34.9–48.1). The maximum size of the pool with acceptable sensitivity (>95%) was found to be of six samples. For the pool size of six samples, the sensitivity was 97.8% and the efficiency of pooling was 0.38. Conclusions: The pooling of samples is a practical way for scaling up testing and ultimately containing the further spread of the CoV disease 2019 pandemic. contained and the surveillance machinery can be activated for contact tracing and ultimately break in the chain of transmission of the virus. [5] Densely and heavily populated countries such as India and other developing countries have abysmally low test rates per capita in the world, [6] and the number of cases is expected to shoot up post lockdown. Hence, the present situation demands countries to scale up their testing and design innovative strategies to conserve diagnostic kits and reagents. The pooling of diagnostic tests, which has previously been applied in other infectious diseases, [7, 8] is one such method where the samples are mixed and tested at a single pool and subsequent individual tests are made only if the pool tests positive. Pooling saves time, workforce, and most importantly diagnostic kits and reagents. In the present study, we tried to define the pool size that could be applicable to a particular population with acceptable confidence and avoiding drop of positive cases which is a major risk of pool testing. Combined nasopharyngeal and oropharyngeal swabs were collected and transported in viral transport media (VTM) maintaining the proper cold chain and sent to the virology laboratory of the Institute of Liver and biliary sciences, New Delhi, India. A volume of 200µl of the sample was further processed for viral nucleic acid extraction by Qiasymphony Diagnostic sample preparation (DSP) Virus/Pathogen mini kit (Qiagen GmbH, Germany) as per the manufacturer's protocol in elutes of 60 µl each. [9] Each sample was subjected to the addition of 10 µl of spiked extraction control at the time of extraction itself, to check the validity of the extraction procedure. The 5 µl of the extracted RNA elute/sample was subjected to RT-qPCR for the qualitative detection of SARS-CoV-2 RNA utilising with AgPath-IDTM One-Step RT-PCR Reagents (Thermo Fisher Scientific) using an Applied biosystem (ABI) 7500 Real-Time PCR system (Thermo Fisher Scientific) and LightMix ® SarbecoV E-gene (TIB MOLBIOL). Reactions were heated to 55°C for 5 min for reverse transcription, denatured in 95°C for 5 min, and then, 45 cycles of amplification were carried out for 95°C for 5 s and 60°C for 15 s using fluorescein amidite parameter for E gene. This assay targets the detection of the E gene for SARS as well as nCoV-2. The primer details are given in Table 1 . All samples that were screened positive for the E gene were confirmed by the performance of RT-qPCR for the detection of specific RdRp gene of SARS-CoV-2 using LightMix® Modular SARS-CoV-2 RdRP (TIB MOLBIOL) using similar PCR conditions as described above. [10] Scheme of pooling of RNA elutes Arbitrarily 7 twice tested SARS CoV 2 E and RdRp gene positive RNA elutes with varying cycle threshold values and 48 twice tested negative elutes for SARS CoV 2 E and RdRp gene tested by utilising AgPath-IDTM One-Step RT-PCR Reagents were selected. A total of 48 negative sample elutes were used to make eight series of 11 pools each (total of 77 pools) in an equal volume (3 µl each) of 2, 4, 6, 8, 10, 12, 16, 20, 24, 32 and 48 sample elutes. Each of the 11 pools was mixed with 1 positive elute to make its seven dilutions series of 1: The frequency of Ct values distribution of individually tested positive samples reported from our laboratory during this pandemic was estimated and was categorised into ≤25, >25-29, >29-31, >31-33, >33-35, >35-38 and >38-40 ranges. The frequency of positive pools falling in these ranges was used to calculate the sensitivity of the pooling of sample elutes for RT-qPCR. Known positive and negative sample elutes were used for making of pools, but for calculation, it was considered that if following a pooling algorithm after obtaining the results of pool testing, the same previously obtained results will get after individual testing of samples. The result of individually tested samples was considered the gold standard for the calculation of sensitivity, specificity, negative predictive value (NPV) and positive predictive value (PPV). In this study, clinical samples previously established as positive for the SARS CoV-2 virus were chosen to determine whether they can detectable when their elutes mixed with negative samples elutes in different dilutions. Positive sample elutes for SARS CoV-2 with a different range of Ct values of E gene target were chosen, serially diluted with negative sample elutes, and RT-qPCR was performed. Overall, 88 pools were made which included 77 pools having one positive sample elute in different concentrations and 11 pools having no positive sample elutes. Out of the 77 pools of up to 48 times serial dilutions of positive samples, only 53 (68.8%) were found positive [ Table 2 ]. Positive samples with Ct values ranging from 25 to 31 were detected in pools till 1: 48 dilution, while pools containing samples of Ct value 33 were detected till 1:32 dilution. Pools with higher Ct values became negative in lower dilutions. Sample with Ct values of 35, 38 and 39 was detected till 1:8, 1:6 and 1:4 dilutions, respectively. As the Ct value of an individual positive sample in the pool was increasing, the probability of detection of the positive pool was decreasing and as the size of the pool increased, Ct was attained later, as expected due to the dilution effect [ Table 2 and Figure 1 ]. The frequency of Ct values (E gene only) distribution of individually tested samples was derived from 227 SARS CoV 2 E and RdRp gene positive samples detected from the 1 st of March to 30 th April 2020 using the same PCR reagents used for pool testing [ Table 3 ]. Sensitivity of pool testing was defined as the probability that a true positive individual sample will be declared positive. The sensitivity of pooling was calculated by taking into consideration of the distribution of a particular Ct value in the results obtained during routine diagnostic testing of individual samples received. We defined seven ranges of Ct values and for pooling, we took samples with Ct value equal to the upper limit of the range as a more conservative approach [ Table 3 and Similarly, the sensitivity of other pool sizes was calculated and shown in Table 2 . The sensitivity of pool testing decreased as the pool size increased and also as the Ct value of the positive sample in the pool increased. The acceptable sensitivity (>95%) of pool testing was found for pooling of up to six samples [ Table 3 and Figure 3 ]. Specificity of pool testing was defined as the ability of the pooling algorithm to correctly identify those without the infection. For the calculation of specificity of pooling, a total of 11 pools of 2, 4, 6, 8, 10, 12, 16, 20, 24, 32 and 48 SARS CoV 2 E and RdRp gene negative samples The NPV was defined as the probability that an individual specimen identified as negative at the end of a pooling algorithm was truly negative. The overall NPV was found to be 97.8%. The NPV for a pool size of 2 and 4 was 100% and found no false negative results, but for the pool sizes of 6, 8, and 10, NPV gradually decreased from 97.2% to 95.45% due to gradual increase in false negative results. For pool sizes from 12 to 48, NPV was progressively increased due to stable false-negative results but increasing pool size. The PPV was defined as the probability that a specimen identified as positive at the end of a pooling algorithm was truly positive. The PPV of pooling was 100%, and no false positive results were found. 1 25 TD TD TD TD TD TD TD TD TD TD TD 2 29 TD TD TD TD TD TD TD TD TD TD TD 3 31 TD TD TD TD TD TD TD TD TD TD TD 4 33 TD TD TD TD TD TD TD TD TD TD TND 5 35 TD TD TD TD TND TND TND TND TND TND TND 6 38 TD TD TD TND TND TND TND TND TND TND TND 7 39 TD TD TND TND TND TND TND TND TND TND The efficiency of a pooling algorithm was defined as the expected number of NAATs required per individual specimen evaluated. An efficiency of sample pooling was calculated 0.38 using the online calculator considering the two-stage Dorfman mini-pool strategy of pool testing with the conservative predictions of a sensitivity of 95%, a specificity of 99%, 4% prevalence of SARS-CoV-2 infection and pool size of 6 samples (http://www.bios.unc.edu/~mhudgens/optimal. pooling.b.htm). Most positive pools reach the threshold at a later Ct as they are further diluted. A pool size of six samples having a sensitivity of 97.8% and NPV of 97.2% was considered acceptable for the pooling of sample elutes for RT-qPCR. While individual level NAAT remains the gold standard for the diagnosis of COVID-19 infection, a limited supply of diagnostic kits and reagents remains the major bottleneck in expediting testing of COVID-19 in the community especially during surge testing. [11] Furthermore, mass testing of samples should be done to control the COVID-19 pandemic at the earliest. [12] Hence, a testing method where a large number of samples can be tested and consuming minimal testing kits and reagent is the need of the hour. In this study, effect of pool size on the detection of SARS CoV 2 and its accuracy in the population was assessed. We found that a single clinical sample with SARS-CoV-2 RNA can be consistently detected in a pool of a maximum of four samples. The Ct values of individually tested RNA elute and that of the pooled group were in sync. As evident from Figure 1 , pooling in this pattern leads to a gradual increase in the threshold cycle of the pooled group. Furthermore, additional amplification cycles could lower the detection limit allowing better detection of pools having samples of higher Ct values. [13] The pooling of samples is essentially important in monitoring the infection in cohesive groups such as quarantine facilities, health-care workers, community surveillance and diagnosing the asymptomatic cases. The infection load in these groups may be low, but even a single positive case among such groups can activate the surveillance system and quarantine the affected group and prevent the further spread in the community. The pooling is better suited for low-prevalence populations, and it could fasten the testing over a very short time that would be able to identify new hotspot areas for infection before becoming condition worst. The pooling of samples can also be done before RNA extraction that is at the time of sample collection, putting the nasopharyngeal/oropharyngeal swabs into a common VTM and after sample collection by pooling of VTM. By doing so, the major bottleneck of RNA extraction can be removed. The pooling of samples leads to the dilution of nucleic acid present in the pool leading to a decrease in the sensitivity. Hence, the pool size should be optimised for the agent to be detected and assay to be used so that even low-positive samples would not be missed in the pool testing. Gupta et al. showed that the increase in the cellular material, including nucleic acid, due to the pooling of multiple samples did not affect the detection of the SARS-CoV-2 virus by RT-qPCR. [14] The size of the pooling of samples should be such that it should give a positive result if it has at least one positive sample that has the lowest amount of viral nucleic acid so that the sensitivity does not get compromised. We determined the pool size by considering of amount of viral nucleic acid (as Ct value is usually proportional to the amount of nucleic acid) that is more prevalent in the community during this pandemic. A study by Arons et al. has shown that SARS-CoV-2 N1 targeted median Ct values for the four symptom status groups were similar (asymptomatic residents, 25.5; pre symptomatic residents, 23.1; residents with atypical symptoms, 24.2; and residents with typical symptoms, 24.8). [15] The median Ct value for samples tested positive in this pandemic was 32 (34.7-29.0). Nearly 97.8% of samples had Ct value of <38 for the 45 cycle protocol of RT-qPCR, and all the pools having a positive sample of Ct value <38 were tested positive up to the pool size of 6 samples, so we determined the optimal pool size of 6 for pool testing of SARS-CoV-2 RT-qPCR. The sensitivity (97.8%) and Negative predictive value (NPV) (97.22%) value of the pool of six samples were taken to consider it optimal pool size. The samples that had a Ct value of >38 were only 2.2%. The clinical significance of such low-positive samples could not be determined because of the unavailability of the gold standard viral culture methods. The efficiency of individual testing is 1, and efficiency of <1 indicates that the pooling algorithm will require fewer tests on average than individual testing. By calculating on online calculator, we found the efficacy of pool testing of six samples 0.38, i.e., for one sample testing, only 0.38 test reagents are required (>95% sensitivity of the test, a prevalence rate of infection 4% and two-stage Dorfman mini-pool strategy used for calculation). Many pool testing strategies for pooling are known. The two-stage Dorfman mini-pool strategy is one of them which are more convenient for the pools of small sizes. In the first stage of this strategy, the master pool comprising all specimens is tested; if the master pool tests positive, all component individual specimens are tested (the second of the two stages). [16] concluSIonS As the COVID-19 pandemic continues spreading, there is tremendous stress on logistics. Hence, pooling of samples is a practical way for scaling up testing and ultimately containing the further spread of the COVID-19 pandemic. This study suggests that a pool size of six samples would be optimal with acceptable confidence. Nil. There are no conflicts of interest. A novel coronavirus from patients with pneumonia in China Naming the Coronavirus Disease (COVID-19) and the Virus that Causes it The laboratory diagnosis of COVID-19 infection: Current issues and challenges COVID-19) Situation Summary People power: How India is attempting to slow the coronavirus Evaluation of the pooling of swabs for real-time PCR detection of low titre shedding of low pathogenicity avian influenza in turkeys Estimating community prevalence of ocular Chlamydia trachomatis infection using pooled polymerase chain reaction testing Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR Faces Such a Shortage of Coronavirus Test-Kits that People say Getting One is Like "Winning the Lottery Covid-19 mass testing facilities could end the epidemic rapidly Evaluation of COVID-19 RT-qPCR test in multi-sample pools. medRxiv Pooled RNA sample reverse transcriptase real time PCR assay for SARS CoV-2 infection: A reliable, faster and economical method Presymptomatic SARS-CoV-2 infections and transmission in a skilled nursing facility The detection of defective members of large populations