key: cord-0767467-rd12xwvl authors: Black, Margaret A.; Shen, Guomiao; Feng, Xiaojun; Garcia Beltran, Wilfredo F.; Feng, Yang; Vasudevaraja, Varshini; Allison, Douglas; Lin, Lawrence H.; Gindin, Tatyana; Astudillo, Michael; Yang, Diane; Murali, Mandakolathur; Iafrate, A. John; Jour, George; Cotzia, Paolo; Snuderl, Matija title: Analytical performance of lateral flow immunoassay for SARS-CoV-2 exposure screening on venous and capillary blood samples date: 2020-11-07 journal: J Immunol Methods DOI: 10.1016/j.jim.2020.112909 sha: a9ff9288a117ef3b8491e8464b0196b60c3e6da9 doc_id: 767467 cord_uid: rd12xwvl OBJECTIVES: We validate the use of a lateral flow immunoassay (LFI) intended for rapid screening and qualitative detection of anti-SARS-CoV-2 IgM and IgG in serum, plasma, and whole blood, and compare results with ELISA. We also seek to establish the value of LFI testing on blood obtained from a capillary blood sample. METHODS: Samples collected by venous blood draw and finger stick were obtained from patients with SARS-CoV-2 detected by RT-qPCR and control patients. Samples were tested with Biolidics 2019-nCoV IgG/IgM Detection Kit lateral flow immunoassay, and antibody calls were compared with ELISA. RESULTS: Biolidics LFI showed clinical sensitivity of 92% with venous blood at 7 days after PCR diagnosis of SARS-CoV-2. Test specificity was 92% for IgM and 100% for IgG. There was no significant difference in detecting IgM and IgG with Biolidics LFI and ELISA at D0 and D7 (p = 1.00), except for detection of IgM at D7 (p = 0.04). Capillary blood of SARS-CoV-2 patients showed 93% sensitivity for antibody detection. CONCLUSIONS: Clinical performance of Biolidics 2019-nCoV IgG/IgM Detection Kit is comparable to ELISA and was consistent across sample types. This provides an opportunity for decentralized rapid testing and may allow point-of-care and longitudinal self-testing for the presence of anti-SARS-CoV-2 antibodies. The novel severe acute respiratory coronavirus, SARS-CoV-2, emerged in Wuhan, China in December of 2019 [1] . The disease caused by SARS-CoV-2, COVID-19, rapidly spread across the globe and was declared a pandemic by the World Health Organization (WHO) on March 11, 2020 [2] . COVID-19 has caused significant morbidity and mortality worldwide, with over 4 million confirmed cases and 284,536 deaths attributed to the disease as of May 11, 2020 [3] . In the United States alone, there are over 1 million confirmed cases and more than 80,000 deaths attributed to COVID-19 [4] . Reverse-transcriptase polymerase chain reaction (RT-PCR) techniques performed with nasopharyngeal samples are the mainstay for diagnosing acute infection with SARS-CoV-2, and multiple molecular testing modalities are now available [5] . However, serologic testing to determine recent infections and potential immunity remain limited. Serologic testing for IgM and IgG antibodies is a useful adjunct for clinical decision making [6] and has important implications for public health and policy decisions [7] . Serology is cost-efficient, fast, simple to perform, and does not require additional materials such as nasopharyngeal swabs and viral transport media required for many PCR-based molecular testing platforms, which have been scarce during the pandemic. With no available vaccine and limited treatment options for COVID-19, the development and validation of rapid serologic testing is urgently required. Serologic assessment provides valuable information on past exposure, although the protective effect of anti-SARS-CoV2 antibodies remains unknown. As demands for laboratory testing have increased exponentially, commercial vendors are developing in vitro diagnostics for detection of SARS-CoV2, and many are applying for and obtaining emergency use authorization (EUA) from the Food and Drug Administration (FDA) [7] . In addition to implementation of tests in the field. However, highly variable sensitivity and specificity of these assays for COVID-19 immunity quickly led to the recognition that regulatory oversight or robust internal validation is required. Tests approved by the FDA under EUA must be verified before they are widely used for clinical diagnosis and decision-making. In the absence of FDA approval, internal rigorous validation of intended assays with establishing proper thresholds is necessary. Herein, we describe clinical validation of a new lateral flow immunoassay (LFI) test intended for rapid screening and qualitative detection of anti-SARS-CoV-2 IgM and IgG in serum, plasma, and whole blood. We also sought to establish the value of LFI testing on capillary blood obtained from a finger stick sample, as there is currently only one FDA EUA approved assay for capillary blood (DPP COVID-19 IgM/IgG System, Chembio Diagnostic Systems, Inc.) [8] . Plasma, serum, whole blood, and capillary blood (finger stick) samples from patients diagnosed with COVID-19 by PCR were tested with the 2019-nCoV IgG/IgM Detection Kit (Colloidal Gold) (Biolidics Ltd.), and the results were correlated with those obtained by enzyme-linked immunosorbent assay (ELISA), the gold standard for serologic detection of antibodies. This retrospective study assessed the sensitivity and specificity of a commercially available lateral flow immunoassay for detection of IgM and IgG antibodies specific to SARS-CoV-2. Samples from 62 patients used in this study were discarded clinical samples collected for routine laboratory tests and research samples collected at NYU Langone Medical Center under Institutional Review Board approval (IRB#: S16-00122). Whole blood, plasma, or serum samples collected by venous blood draw and capillary blood collected via finger stick were used. Patients were defined as SARS-CoV-2 positive if RT-qPCR (cobas, Roche FDA J o u r n a l P r e -p r o o f EUA) performed on nasopharyngeal samples was positive for SARS-CoV-2 sequence. Plasma and venous whole blood samples were collected from hospitalized SARS-CoV-2 positive patients on the day of the initial positive nasopharyngeal PCR test (D0; n=24) or ≥7 days (D7; n=26) after the positive PCR test (D7). For 11 patients, serum samples at two time points separated by 7 days were available (D0 and D7). Plasma samples from hospitalized patients with SARS-CoV-2 not detected in nasopharyngeal sample by PCR, but positive for other respiratory viruses (human metapneumovirus, rhinovirus/enterovirus, coronavirus NL63, Influenza A H1N1), were also collected at D0 (n=4). In addition, plasma samples of presumably SARS-CoV-2 negative patients hospitalized in January-April of 2019 (pre-SARS-CoV-2 era) that were previously collected and stored at -80°C were used (n=20). Samples collected for validation of capillary whole blood samples were obtained by finger stick from 14 patients who had recovered from COVID-19 at 18-46 days after positive SARS-CoV-2 positive RT-qPCR testing. Nasopharyngeal samples were tested for clinical care using the cobas® SARS-CoV-2 molecular diagnostic test (Roche) following the manufacturer's instructions per FDA approved EUA protocol. Venous plasma, serum, whole blood, or capillary (finger stick) blood was used for the 2019-nCoV IgG/IgM Detection Kit (Colloidal Gold) (Biolidics Ltd., Singapore) following the manufacturer's instructions. This lateral flow immunoassay test is based on solid-phase immunochromatography using recombinant colloidal-gold labeled SARS-CoV-2 antigen viral antigen. Briefly, one drop (20 µL) of room temperature (18-28 °C) plasma, serum, whole blood, or capillary finger stick blood was added to the sample well of a test cassette using the provided dropper, and then 3 drops of diluent were added into the same sample well. Results were read after 10 minutes by assessing visual color changes in the testing 29%, 21%, and 29%, respectively (Table 1A) . Sensitivity at D7 increased to 54% for IgM, 88% for IgG, and 92% for IgM or IgG (Table 1A) . Of 24 negative plasma samples, 22 tested negative for IgM and IgG antibodies (92%) by LFI. Of 4 negative plasma samples obtained in March 2020, no IgM or IgG was detected. Interestingly, of 20 plasma samples obtained from before the SARS-CoV-2 outbreak, 2 samples tested positive for IgM, however no samples tested positive for IgG. The overall specificity of the LFI was 92% for IgM, 100% for IgG, and 92% for IgM and IgG combined (Table 1B) . Eleven patients with SARS-CoV-2 were tested sequentially by Biolidics LFI at D0 and D7. One patient was positive for IgM and IgG antibodies at D0 and also at D7. Of the 10 patients who were negative for antibodies at D0, 4 showed IgM antibodies at D7 and 8 showed IgG antibodies at D7 time point. Overall, 9 of 11 sequentially tested patients showed IgM and/or IgG antibodies at D7 by Biolidics LFI (Figure 2A , Supplemental Table 1 ). Results of serologic LFI were correlated with results obtained by ELISA. ELISA was performed with plasma or serum samples from 15 patients with SARS-CoV-2 detected on nasopharyngeal swab and 24 negative control samples (Supplemental Table 3 (Table 2) . LFI and ELISA were compared at both D0 and D7 for detecting IgM, IgG, and IgM or IgG ( Figure 2 ). There was no significant difference between LFI and ELISA in any comparison except borderline significance for detecting IgM at D7 (p = 0.04; Table 3 ). Correlation of MGH ELISA and Biolidics LFI was analyzed to establish a 95% confidence threshold indicating that at least 95% of positive results by Biolidics LFI correspond to antibody concentration of at least the confidence threshold on ELISA (Figure 3 , black dotted line). Our data suggest that Biolidics LFI positivity can be used as a substitute for high level antibody levels by ELISA. Of Table 3B ). Results of serologic LFI were compared using both D7 plasma and D7 venous whole blood samples from 22 patients with SARS-CoV-2 detected on nasopharyngeal samples ( Figure 4A , 4B). With D7 plasma samples, 21 of 22 patients (95%) tested positive for antibodies (IgM and/or IgG). The same 21 patients tested positive for antibodies using D7 venous whole blood, indicating no significant difference between sample types for detecting IgM and IgG (p = 0.33, p = 1.00; Table 4A , 4B). Plasma and whole blood results for IgM and IgG separately were equivalent with one exception. In one patient (P35), IgM and IgG were both positive with D7 plasma, while IgM was negative and IgG was positive with D7 venous whole blood. In 14 patients who recovered from SARS-CoV-2, results of LFI with capillary (finger stick) samples were assessed at least two weeks after PCR confirmation (range, 18-46 days; median, 30.5 days) and symptom J o u r n a l P r e -p r o o f onset (range, 19-61 days; median, 32 days). Six patients tested positive for IgM, 12 tested positive for IgG, and 13 tested positive for IgM or IgG, for corresponding sensitivities of 43%, 86%, and 93%, respectively (Table 4C , 4D; Figure 4B ). One patient was excluded from the sensitivity analysis due to concurrent diagnosis of an autoimmune disorder (systemic lupus erythematosus). Interestingly, this patient did not exhibit IgM or IgG antibodies on LFI. The COVID-19 pandemic has rapidly altered the landscape of clinical testing validation and regulatory Patient results were compared to those of negative control samples from hospitalized patients with negative SARS-CoV-2 PCR results as well as control samples archived before the emergence of SARS-CoV-2. We also sought to establish the sensitivity of the Biolidics test kit using capillary blood obtained via a finger stick, as LFI tests have potential utility in community testing, point of care testing, or even self-testing. Overall, we show that the Biolidics LFI kit has clinical sensitivity of 92% at 7 days after PCR diagnosis of SARS-CoV-2. Test specificity was 92% for IgM and 100% for IgG. Results of the LFI are similar to those obtained with gold standard ELISA testing, providing evidence that robust LFI tests can have utility comparable to that of ELISA with additional benefits of easy testing and low cost. There was no significant difference in detecting IgM and IgG with Biolidics LFI and ELISA at D0 and D7 (p=1.00), except for marginally significant detection of IgM at D7 (p=0.04). Nevertheless, IgM detection is more prone to false positive results by both ELISA and LFI. Results obtained from plasma and whole blood samples were compared and show reproducibility between sample types, which removes the need for plasma or serum preparation. Furthermore, we show J o u r n a l P r e -p r o o f that capillary whole blood obtained by finger stick shows comparable sensitivity for detecting anti-SARS-CoV-2 IgM and IgG antibodies as venous blood samples. This provides an opportunity for communitybased testing, rapid point of care testing, and potentially self-testing for the presence antibodies, similar to glucose self-monitoring. While ELISA remains a gold standard for establishing the level of antibodies, we show that strong correlation between LFI and ELISA can provide an estimate of antibody levels between closely correlated clinical assays. This close correlation decreases the need for additional testing by ELISA in settings where the financial cost of ELISA is a limiting factor prohibitive to broad serologic screening. Serology provides important complementary information to PCR testing that is useful to evaluate the immunity status of a patient. Production of specific IgM antibodies is one of the body's first lines of defense against viruses; IgG antibodies are produced 1-2 weeks later and provide long-term immunity. The presence of IgM antibodies can be used to indicate recent exposure, while IgG antibodies indicate previous exposure and often portend immunity [6] . In our study, detection of IgM was less specific than IgG detection by both Biolidics LFI and ELISA, and sensitivity for detection of IgM and IgG antibodies increased substantially with time. The most sensitive LFI results were obtained when both IgM and IgG were considered more than 7 days after PCR testing. Both LFI and ELISA at D7 showed positive IgM and/or IgG results for 8 of 9 sequentially tested patients. One patient did not show IgM or IgG antibodies on LFI or ELISA at D7; this patient previously underwent a solid organ transplant and was taking immunosuppressant therapy ( Figure 2 ). Taken together, our results indicate that IgM and IgG serologic testing must be evaluated within the proper clinical context, and serial serologic assessment or follow-up testing with a more sensitive method assay may be useful in some circumstances. While RT-PCR remains the gold standard for acute diagnosis of SARS-CoV-2, serologic tests provide supplementary diagnostic information and are more practical for use in large-scale screening [7, 9, 10] . Molecular-based PCR testing requires skilled technicians, relatively invasive sample collection, expensive technology, high complexity laboratories, and supplies that were in short supply due to high global demand. These molecular tests are useful for symptomatic patients who require triage and J o u r n a l P r e -p r o o f treatment, but antibody detection may be more useful for population-wide screening protocols. ELISA, the gold standard for antibody detection and quantification, requires trained technicians and relatively complex laboratory procedures that are difficult to quickly scale. In contrast, rapid LFI antibody tests are easy to use, require minimal training for performance and interpretation, are scalable for use in population-wide screening protocols, and are not limited to a laboratory setting. In our study, use of the 2019-nCoV IgG/IgM Detection Kit (Colloidal Gold) did have important limitations. First, detection of antibodies increased from 29% at D0 to 92% at D7, indicating that sensitivity increased significantly with time. This is imperative to consider when implementing largescale screening protocols. Repeat testing may be necessary to determine patients who have developed antibodies if they were first tested soon after exposure to the virus. This is especially true for asymptomatic carriers who never show symptoms but may still develop antibodies. Further large-scale longitudinal studies are needed to determine the expected timeline of antibody development in SARS-CoV-2. It is also imperative to determine whether long-term antibody production is maintained and whether antibodies provide effective immunity against SARS-CoV-2. Interestingly, we identified two patients in our capillary validation cohort who had similar symptoms of PCR-confirmed SARS-CoV-2 at the same time who lived in the same household. One patient had IgG detected by LFI over one month from diagnosis, while the other household member was negative for both IgM and IgG antibodies ( Figure 4C Importantly, neither LFI nor ELISA identified IgG antibodies in negative control specimens, suggesting that IgG is more accurate in identifying prior SARS-CoV-2 exposure. IgM antibody results should be J o u r n a l P r e -p r o o f interpreted in the proper clinical context, and repeated serology testing or immediate PCR testing to exclude asymptomatic infection may be warranted with detection of IgM alone. In summary, rapid serology allows an effective tracking method for asymptomatic carriers and patients with mild disease who do not require sensitive molecular-based diagnosis to guide acute care. Here, we validated the use of Biolidics 2019-nCoV IgG/IgM Detection Kit (Colloidal Gold) and showed consistent results across multiple sample types compared to ELISA. However, the large variability of LFI tests requires that each test is either validated in-house or that proper FDA review and approval is conducted. From an interpretational standpoint, internal validation is recommended as different LFI tests may display different intensity of positive bands. Recording images and utilization of image software or artificial intelligence based interpretation of the results may also decrease inter-observer variability, which is particularly important in the setting of population-wide self-testing. Large-scale population based antibody screening with validated testing platforms will provide additional understanding of the human to human transmission rate, prevalence, incidence, and mortality of SARS-CoV-2, which remain largely uncertain in the rapidly evolving global landscape. Independently validated serology tests will be integral to obtaining this information. J o u r n a l P r e -p r o o f J o u r n a l P r e -p r o o f A Review of SARS-CoV-2 and the Ongoing Clinical Trials World Health Organization. WHO Characterizes COVID-19 as a Pandemic Cases of Coronavirus Disease (COVID-19) in the US Diagnostic Testing for Severe Acute Respiratory Syndrome-Related Coronavirus-2: A Narrative Review SARS-Cov-2 infection: Response of human immune system and possible implications for the rapid test and treatment SARS-CoV-2 Testing