key: cord-0765561-dou24ofs authors: De Acha, Nerea; Socorro-Leránoz, Abián B.; Elosúa, César; Matías, Ignacio R. title: Trends in the Design of Intensity-Based Optical Fiber Biosensors (2010–2020) date: 2021-06-15 journal: Biosensors (Basel) DOI: 10.3390/bios11060197 sha: c9daf811b6a7346d1400321d3f8c59cbf67927ed doc_id: 765561 cord_uid: dou24ofs There exists an increasing interest in monitoring low concentrations of biochemical species, as they allow the early-stage detection of illnesses or the monitoring of the environment quality. Thus, both companies and research groups are focused on the development of accurate, fast and highly sensitive biosensors. Optical fiber sensors have been widely employed for these purposes because they provide several advantages for their use in point-of-care and real-time applications. In particular, this review is focused on optical fiber biosensors based on luminescence and absorption. Apart from the key parameters that determine the performance of a sensor (limit of detection, sensibility, cross-sensibility, etc.), other features are analyzed, such as the optical fiber dimensions, the sensing set ups and the fiber functionalization. The aim of this review is to have a comprehensive insight of the different aspects that must be taken into account when working with this kind of sensors. In recent years, great efforts have been made to enhance our quality of life. Although most of them are devoted to the treatment or curing of severe diseases [1] , their prevention and early detection are not only gaining interest among scientists [2] and companies [3] , but also governments [4] . These goals involve, for instance, the real-time monitoring of water and air quality [5] or the rapid detection of biomarkers related to illnesses [6] . In order to meet these requirements, accurate, highly sensitive and fast biosensing devices have been developed using different technologies. Among them, one of the most studied during the last decades is that based on optical fiber. It offers several advantages such as biocompatibility [7] , the possibility of in-situ monitoring [8] and the possibility of multiparameter sensing [9] , among others. These features facilitate its utilization in pointof-care and/or real-time applications. Optical fiber-based biosensors allow monitoring two main parameters: wavelength shifts and intensity variations [10, 11] . Biosensing devices characterized by the first parameter were deeply analyzed in [12] . That contribution showed the three main ways to design wavelength-based optical fiber biosensors developed during last decade, as well as the benefits of working with wavelength-shift measurements. On this occasion, authors have considered it of great interest to bring together the latest advances in the field of intensity-based optical fiber biosensors within this contribution. In this manner, readers can have a broad view of the degree of development and possible applications of optical fibers as biosensing platforms. Well-established existing technologies, such as ELISAs, fluorescence microscopy, DNA chips or flow cytometry, seem to solve the current needs in clinical analytics. Despite wavelength shifts, colorimetric techniques obtaining the results after a few minutes or strip-based detections where distinguishing the symptoms at first sight (normally by using colors) and then disposing of the active part are more easy-to-handle and more familiar to what is currently used in clinical analytics. In this sense, since intensity-based optical fiber techniques detect color changes and their use is really close to the day-by-day instrumentation in these areas, they constitute an interesting alternative to wavelength-based techniques that should not be underestimated at all. There are still some drawbacks to solve. For instance, the instabilities of the light source while providing the optical power can affect the assays, since they can mask an eventual enhancement of the obtained measurements. However, this is something that can be avoided by normalizing the measurements [13] . Other parameters that can also affect the measurements are temperature, relative humidity, light coupling from the sensor to the photoreceptor or the influence of ambient light. The implementation of intensity-based optical fiber biosensors is quite simple. As it is well-known, every biosensor owns a substrate to guide the detection phenomenon, a bioreceptor to detect the target analyte uniquely and a biofunctionalization interface to attach the bioreceptor to the substrate. In the case of the technology presented in this contribution, the substrate is the fiber itself as a transporting waveguide. This simplifies the optical set-up a lot, since there is no need for dealing with extra material that involves, for instance, fusing fibers or depositing nanofilms onto the fibers to generate resonant phenomena. As long as the evanescent field of the optical fiber is able to reach the bioreceptor-analyte interaction, the main focus is to biofunctionalize the fiber properly and then detect the target analyte. Moreover, the employed instrumentation is simple and cost-effective. Since the measurements are taken at a specific wavelength, just a simple light source (i.e., an LED) and a photodetector centered at that wavelength are needed, no matter if the light is going to be measured using either transmission or reflection set-ups. In view of the previous considerations, this bibliographic review will focus on those optical fiber biosensors that measure intensity variations. As it will be demonstrated, the utilization of these techniques for the detection of biomolecules is widespread [14, 15] , taking advantage of the improvements that optical fiber technology has experienced in the last decade. To this purpose, a first section describing the main operation principles used to detect phenomena based on intensity variations (luminescence and absorption, essentially) will be addressed. Then, the optical instrumentation and configurations that are commonly employed will be revised, going through their most relevant applications. Next, the sensors performance will be analyzed in detail in terms of their limits of detection (LOD), working ranges, sensitivities, cross-sensitivities and regeneration. This information will allow the reader to have a complete overview of the most important features of luminescence and absorption-based optical fiber biosensors, something that will be summarized in the conclusions section. Biosensors are devices that combine specific bioreceptors (e.g., antibodies, enzymes, or DNA strands) with a transducer (in the case of this review, the optical fiber) so that the interaction between the target analyte and the bioreceptor induces a change in the measured magnitude [16] . In the case of intensity-based optical fiber sensors, that magnitude is the intensity of the light transmitted through the fiber, which can be measure using two different detection methods, mainly luminescence and absorption, which are shown in Figure 1 . The first one comprises those sensors based on luminescence, either when the light collected increases or decreases at a certain wavelength as a function of the analyte concentration. The second phenomenon is absorption. This will include those sensors that take advantage of the absorbing properties of the thin films deposited onto the fiber to detect what happens at a certain wavelength, or spectral range. Among the several advantages of employing thin films, the possibility of tailoring their features (selectivity, sensitivity, permeability) in the nanoscale [17] , their easy fabrication and characterization [18] or Biosensors 2021, 11, 197 3 of 33 the possibility of functionalizing them [19] are the most remarkable ones. They will be discussed in the corresponding section. sors usually need a higher light intensity than phase modulated sensors [20] . In th multimode fibers allow transporting high intensities along their thicker core. Mo is possible to increase the light coupling by manufacturing different geometrie optical fiber. The most representative ones are described throughout this section The second reason is the optical working range, since most of the reported bi work on the UV-VIS range. Apart from that, the use of molecular labels in th luminescence-based biosensors, which emit in the visible range, needs a low-lo guide to where to couple and transport the luminescent emission produced. Luminescence is a well-established detection mechanism in different biom search areas. In fact, it is one of the most important sensing procedures in clin biological applications. It can be found when performing PCRs [21] , ELISAs [22] cence microscopy [23] or gene expression in DNA chips [24] , among others [25] . ing principle is based on the Jablonski's diagram schematic shown in Figure 2 . T certain substances, called luminophores, that absorb the energy provided by pho wavelengths located at the blue, violet and UV part of the electromagnetic spectr energy is absorbed by the existing electrons in the different orbitals, which are p to higher energy orbitals in a process called "excitation" (1). Then, since the electr to return to their lowest energy state, they will progressively go back to their or bitals (2), thus emitting part of the energy they have absorbed in the form of lowe photons (3) . This means at wavelengths located within the green to red and e spectrum [26] . These transitions from the excited states to the ground states are called radia sitions. However, there also exist certain electron transitions where light emiss not take place. They are the non-radiative transitions (dashed arrows in Figure occur through several mechanisms, such as vibrational relaxation, intersystem cr internal conversion [27] . Both radiative and non-radiative transitions influence t tum efficiency, this means the ratio between the absorbed photons and those em luminescence. This is another important parameter that determines the beha fluorophore and, therefore, its choice when performing luminescence-based ments. A common characteristic to every biosensor based on these techniques is the fact of working with multimode fibers. The first reason to justify this is that intensity-based sensors usually need a higher light intensity than phase modulated sensors [20] . In this sense, multimode fibers allow transporting high intensities along their thicker core. Moreover, it is possible to increase the light coupling by manufacturing different geometries on the optical fiber. The most representative ones are described throughout this section. The second reason is the optical working range, since most of the reported biosensors work on the UV-VIS range. Apart from that, the use of molecular labels in the case of luminescence-based biosensors, which emit in the visible range, needs a low-loss waveguide to where to couple and transport the luminescent emission produced. Luminescence is a well-established detection mechanism in different biomedical research areas. In fact, it is one of the most important sensing procedures in clinical and biological applications. It can be found when performing PCRs [21] , ELISAs [22] , fluorescence microscopy [23] or gene expression in DNA chips [24] , among others [25] . Its working principle is based on the Jablonski's diagram schematic shown in Figure 2 . There are certain substances, called luminophores, that absorb the energy provided by photons with wavelengths located at the blue, violet and UV part of the electromagnetic spectrum. This energy is absorbed by the existing electrons in the different orbitals, which are promoted to higher energy orbitals in a process called "excitation" (1). Then, since the electrons tend to return to their lowest energy state, they will progressively go back to their original orbitals (2), thus emitting part of the energy they have absorbed in the form of lower energy photons (3) . This means at wavelengths located within the green to red and even NIR spectrum [26] . These transitions from the excited states to the ground states are called radiative transitions. However, there also exist certain electron transitions where light emission does not take place. They are the non-radiative transitions (dashed arrows in Figure 2 ), which occur through several mechanisms, such as vibrational relaxation, intersystem crossing or internal conversion [27] . Both radiative and non-radiative transitions influence the quantum efficiency, this means the ratio between the absorbed photons and those emitted as luminescence. This is another important parameter that determines the behavior of a fluorophore and, therefore, its choice when performing luminescence-based measurements. For radiative transitions, the duration of the emission time is known as lifetime, and it allows distinguishing between fluorescence (lifetimes from 10 −10 to 10 −5 s) and phosphorescence (lifetimes from 10 −4 to 10 4 s). The wavelength range between the emission and the absorption peaks is known as Stokes shift. The larger it is, the simpler is the experimental set-up required, as there is no need for utilizing optical filters. However, when working with continuous light sources, if both peaks are close, optical filters are necessary in order to isolate the excitation from the emission light at the photoreceptor. For radiative transitions, the duration of the emission time is kno it allows distinguishing between fluorescence (lifetimes from 10 −10 to 1 rescence (lifetimes from 10 −4 to 10 4 s). The wavelength range between th absorption peaks is known as Stokes shift. The larger it is, the simpler set-up required, as there is no need for utilizing optical filters. Howe with continuous light sources, if both peaks are close, optical filters are to isolate the excitation from the emission light at the photoreceptor. Furthermore, the luminescent intensities and lifetimes of some lum on the presence of certain analytes or on the conditions of the surroun they are of great interest for the development of optical biosensors. T attaching the luminophores to the bioreceptor [29, 30] or to the target a cases, it is the binding between the bioreceptor and the analyte that the surrounding medium of the luminophore [28] . This normally lea enhancement processes (i.e., shorter or longer lifetimes). Another poss analyte with a luminophore, so when it joins the bioreceptor, the lum can be captured. Therefore, luminescence quenching as well as lum ment biosensors can be developed. To the first group belong all the se sity decreases in the presence of the target analyte. In the second case nomenon takes place: the luminescent emission increases due to the pr analyte. Figure 3 shows the working principle of a luminescence quenchi ber biosensor. The fiber is first functionalized with a luminophore-labe that is illuminated at the absorption wavelength of the luminophor analyte concentration (Ci) increases, the luminophores are quenched intensity (Ii) decreases (d). The dynamic response of the sensor is dep whereas the obtained calibration curve is shown in Figure 3f . In the ca Furthermore, the luminescent intensities and lifetimes of some luminophores depend on the presence of certain analytes or on the conditions of the surrounding media [28] , so they are of great interest for the development of optical biosensors. This can be done by attaching the luminophores to the bioreceptor [29, 30] or to the target analyte [31] . In some cases, it is the binding between the bioreceptor and the analyte that induces changes in the surrounding medium of the luminophore [28] . This normally leads to quenching or enhancement processes (i.e., shorter or longer lifetimes). Another possibility is to label the analyte with a luminophore, so when it joins the bioreceptor, the luminescent emission can be captured. Therefore, luminescence quenching as well as luminescence enhancement biosensors can be developed. To the first group belong all the sensors whose intensity decreases in the presence of the target analyte. In the second case, the opposite phenomenon takes place: the luminescent emission increases due to the presence of the target analyte. Figure 3 shows the working principle of a luminescence quenching-based optical fiber biosensor. The fiber is first functionalized with a luminophore-labelled bioreceptor (a) that is illuminated at the absorption wavelength of the luminophore (b). As the target analyte concentration (C i ) increases, the luminophores are quenched (c), so the emitted intensity (I i ) decreases (d). The dynamic response of the sensor is depicted in Figure 3e , whereas the obtained calibration curve is shown in Figure 3f . In the case of luminescence enhancement sensors, the opposite process occurs: as the analyte concentration increases, so does the luminescent emission. Only the first process is shown for simplicity. As it has been explained, the luminescence emission by the luminophores occurs due the absorption of energy in the form of light, which requires their exposure to a light source. If this exposure is long-term, it can lead to a photochemical modification of the The purpose of this schematic is to show the generic behavior of this kind of sensors, so the graphs do not correspond to experimental or simulated data, nor to any existing material. As it has been explained, the luminescence emission by the luminophores occurs due the absorption of energy in the form of light, which requires their exposure to a light source. If this exposure is long-term, it can lead to a photochemical modification of the luminophore, inhibiting the transition of the electrons from the ground state to the excited ones, and, in consequence, decreasing the emitted number of photons. This phenomenon is known as photobleaching and, although it can be mathematically modeled and compensated [13] , its effect must be reduced as much as possible because it damages the structure of the luminophore [32, 33] . Among the different elements that can be used as bioreceptors, aptamers (chemically synthesized DNA or RNA strands or oligonucleotides) [34] , DNA strands, proteins and antibodies are the most common. Regarding the applications of these sensors, DNA [35] , glucose [36] and metal ions detection [37] are the most widespread. Despite them, other bioanalytes such as proteins [38] or bacteria [30] can also be detected. The performance of luminescence quenching-based sensors are usually modeled by the Stern-Volmer equations, as they allow the analysis of the distribution of the luminophore inside the sensing film. When it is homogeneously distributed, the quencher affects the whole luminophore population equally, so the calibration curve follows a linear tendency, given by the Stern-Volmer Equation (1) [26] : where [Q] represents the concentration of the target analyte (quencher), K SV is the quenching constant, I 0 the luminescent intensity for 0% of quencher concentration and I the luminescent intensity for a given analyte concentration. In the cases in which the luminophore population is heterogeneously distributed, the calibration curve I 0 /I follows the Demas model [39] according to Equation (2): where f 1 and f 2 ( f 2 = 1 − f 1 ) correspond to the populations of the fluorophore, and K SV,1 and K SV,2 to their quenching constants, respectively. These equations are of great utility when analyzing luminescence-based systems, as they allow the investigation of bioreceptor and quencher distribution, association, diffusion and reaction at the molecular level [40] . Absorption-based sensing has been commonly employed in a wide variety of applications, ranging from gas detection [41] to water-quality monitoring [42] , but also biodetection [43] . The transduction principle is determined by the energy levels of the different materials: as electrons can only exist in discrete energy levels, to be moved from the ground state to higher energy levels, they must absorb enough energy, which is acquired from the absorbed light, as it is depicted in Figure 4 . Thus, the absorption spectrum of a certain material represents the wavelengths at which light provides to its electrons enough energy to reach higher energy levels. There are several ways of detecting the presence of analytes using this technique. However, it is important to classify them according to how the absorption takes place. A first group of absorption-based sensors is that in including a selective group of substances, which due to their internal structure, conformation, polarization or even their own chemistry or physics, they can absorb light at a specific wavelength. This is the case of substances such as acetone, ozone, sulfur dioxide or transition metals, which present an absorption band within the UV range [44, 45] . Their detection can be carried out with no sensitive material coatings onto the substrate (i.e., the optical fiber) [46] . materials: as electrons can only exist in discrete energy levels, to be m ground state to higher energy levels, they must absorb enough energy, w from the absorbed light, as it is depicted in Figure 4 . Thus, the absorptio certain material represents the wavelengths at which light provides to its e energy to reach higher energy levels. A second group is comprised of those sensors that need the functionalization of a substrate with a bioreceptor or a sensing film onto it [47] . They are often based on evanescent wave absorption (EWA), which consists of the interaction of part of the light transmitted through the substrate with the surrounding medium [48] . Specifically, the evanescent wave is modified by changes on the refractive index of the surrounding medium, which increases or decreases the light coupled to the cladding modes of the fiber substrate. A critical factor here is the penetration depth of the evanescent wave (d p ). If d p reaches the bioreceptor-analyte area, it will be possible to detect the magnitude of the biological interaction. In this case, the presence of substances that absorb light in the range of the working wavelengths will cause a reduction in the transmitted intensity. The last option is also based on EWA, but it has been considered apart due to the recent developments and high number of contributions recently published. In some specific cases, optical fiber structures designed to detect wavelength shifts are used as absorption-based biosensors. That is the case of Localized Plasmon Surface Resonances (LSPR) using metallic nanoparticles (NPs), usually made of gold (Au) or silver (Ag) [49] . These NPs are of great interest as they can be biofunctionalized with a specific bioreceptor to detect the target analyte [50] . In these cases, once the LSPR is located at the desired wavelength, the univocal bioreceptor-analyte interaction induces more or less absorption [12] as a function of the increasing analyte concentration, allowing the characterization of this kind of biosensors. Figure 5 shows the working principle of an absorption-based biosensor developed using a metallic nanoparticle functionalized with a bioreceptor (a). This sensing probe presents an original absorption band centered at a determined wavelength (b). As the concentration of the target analyte increases (c), it binds the bioreceptor and this makes the absorbance increase (d), which can be taken into account to monitor the dynamic response of the biosensor as well as to obtain its calibration curve (e and f). This working principle can be applied to other absorption-based sensors that do not require a previous Biosensors 2021, 11, 197 8 of 33 functionalization of the optical fiber. In those cases, the initial absorbance will increase as a function of the analyte concentration. The purpose of this schematic is to show the generic behavior of this kind of sensors, so the graphs do not correspond either to experimental or simulated data, or to any existing material. The response of an optical fiber biosensor is influenced by the optical instrumentation The response of an optical fiber biosensor is influenced by the optical instrumentation and also by the number of correctly attached bioreceptors to the optical substrate and their performance. Regarding the optical instrumentation, the combination of the light source, the photodetector and the optical fiber geometries should discriminate minimum intensity variations, which increases the system resolution. In relation to the bioreceptors, they play a key role as they are responsible for the detection of the target analytes. An optimized combination of both of bioreceptors and instrumentation will determine the response of the optical biosensors, as it defines their parameters, i.e., the sensitivity, LOD, dynamic range, regeneration and cross-sensitivity. All of them are going to be analyzed in this section. The calibration curve of an optical intensity-based biosensor is a function that presents the evolution of the emitted or absorbed light by this biosensor as the analyte is progressively detected. As shown in Figure 6 , after obtaining this grapH it is possible to deduct three main magnitudes that can describe the performance of the biosensor: the sensitivity, the LOD and the dynamic range. These three together with other interesting parameters will be analyzed below, in order to establish the different classifications that will be made in the next section. Biosensors 2021, 11, x FOR PEER REVIEW 9 of 34 deduct three main magnitudes that can describe the performance of the biosensor: the sensitivity, the LOD and the dynamic range. These three together with other interesting parameters will be analyzed below, in order to establish the different classifications that will be made in the next section. • Sensitivity: it is the slope of the calibration curve. That is, the ratio between the intensity (or absorbance) variation and the analyte concentration variation, typically measured at every point of the calibration curve. In the case of luminescence-based biosensors, the corresponding expression is given by (3), while for absorption-based biosensors, the intensity increment should be substituted by an absorption increment. while intensity and absorbance are always expressed in arbitrary units (AU), the analyte concentrations can be expressed in different units (mol/L, colony forming units (CFU), etc.) depending on the way the analyte concentration is measured. Moreover, the responses of the sensors tend not to be linear but sigmoidal. Normally, after tak- • Sensitivity: it is the slope of the calibration curve. That is, the ratio between the intensity (or absorbance) variation and the analyte concentration variation, typically measured at every point of the calibration curve. In the case of luminescence-based biosensors, the corresponding expression is given by (3), while for absorption-based biosensors, the intensity increment should be substituted by an absorption increment. while intensity and absorbance are always expressed in arbitrary units (AU), the analyte concentrations can be expressed in different units (mol/L, colony forming units (CFU), etc.) depending on the way the analyte concentration is measured. Moreover, the responses of the sensors tend not to be linear but sigmoidal. Normally, after taking the data it is typical to perform a data conditioning to logarithmic scales before calculating their sensitivities [51] . • Limit of detection (LOD): it is the lowest amount of analyte that can be detected but not necessarily quantitated by the sensor [52] . It is typically calculated as the mean of the values of the reference base line plus three times their standard deviation (4): where y LOD is the LOD of the sensor, y blank corresponds to the average of the reference samples and σ blank is the standard deviation. • Dynamic range: is the range of analyte concentrations between the limit of quantification (LOQ) and the upper limit. The first one is the minimum analyte concentration that can be detected and measured, while the second corresponds to the maximum concentration that the sensor can detect without being saturated. • Cross-sensitivity: it is defined as its sensitivity towards other substances different from the target analyte [53] . As it is an indicator of the selectivity of a sensor, it is a key parameter when characterizing it. • Resolution: it is the minimum change of the measured magnitude that can be detected [54] . • Detection media: apart from detecting biomolecules in standard conditions (i.e., ultrapure water or buffered solutions), biosensors should be capable of working in serum samples, real samples or, at least, solutions that mimic them. • Regeneration: an effective manner of reducing the cost per test is the possibility of reusing the biosensors. Thus, regenerating the sensors surface has become of great interest and several mechanisms have been developed to reach this goal. Luminescent sensors monitor the concentration of the target analyte by measuring the intensity variations of the employed luminophore. Due to the reduced dimensions of the optical fiber, light coupling from the luminophore to the photoreceptor is critical [55] . Thus, several optical schemes are employed, although the reflection architecture is the most widespread [13] . In particular, tapering the tip of the optical fiber increases the area of the sensing surface, allowing to maximize the luminescence coupling and to minimize the reflection of the excitation light [56] . On the contrary, using a transmission set-up hinders the light coupling from the fiber to the sensing film as well as from the sensing film to the photodetector. In spite of that, some authors still use that architecture [57] . The previous three sensing schemes are depicted in Figure 7 . With the same goal of increasing the interaction area between the probe and the analyte, multimode fibers (made of silica or a polymeric material) are used as substrates when fabricating these sensors. The typical fiber diameters employed range from 400 µm to 1.96 mm [35, 58] . Lower fiber core diameters are less common, but are still used [59] . DNA detection is based on the high affinity shown by complementary DNA strands [60] . These sensors consist of the adsorption of the bioreceptor (i.e., a single-stranded DNA) which is complementary to the target DNA strand, onto the surface of the optical fiber [29, 35] . For instance, Long et al. developed a biosensing platform capable of detecting 3.2 aM of the target DNA [31] : the single-stranded DNA that served as bioreceptor was immobilized onto a silanized 600 µm-core tapered optical fiber using streptavidin and the heterobifunctional cross-linker N-(4-maleimidobutyryloxy) succinimide (GMBS). The sensor was exposed to different quantum dot (QD)-labeled DNA strands: complementary DNA, one-base mismatched complementary DNA, and non-complementary DNA. Due to the hybridization between the DNA strand and the QD-labeled complementary DNA, the fluorescence intensity increased as the concentration of the complementary DNA did (Figure 8 ), while no fluorescent emission was detected in the presence of non-complementary DNA. Furthermore, the surface of the sensor was regenerated using sodium dodecyl sulfate (SDS) at pH 1.9, which allowed its reuse at least 30 consecutive times, with a decrease in the maximum fluorescence intensity lower than 8%. Luminescent sensors monitor the concentration of the target analyte by measur the intensity variations of the employed luminophore. Due to the reduced dimensions the optical fiber, light coupling from the luminophore to the photoreceptor is critical [5 Thus, several optical schemes are employed, although the reflection architecture is most widespread [13] . In particular, tapering the tip of the optical fiber increases the a of the sensing surface, allowing to maximize the luminescence coupling and to minim the reflection of the excitation light [56] . On the contrary, using a transmission set-up h ders the light coupling from the fiber to the sensing film as well as from the sensing fi to the photodetector. In spite of that, some authors still use that architecture [57] . The p vious three sensing schemes are depicted in Figure 7 . Luminescent indicators can be attached to the bioreceptor instead of to the target DNA strand. It is so in the case of Giannetti et al. [29] , where the molecular beacon 5 -(ATTO647N)GAGAAAGGGCTGCCA(Thiol)-3 was immobilized onto a 600 µm-core tapered optical fiber. The sensor was exposed to a DNA sequence complementary to that acting as bioreceptor as well as to random DNA strands: the luminescent emission increased as the presence of the target DNA did. However, no variation of the luminescent intensity was observed in the presence of random DNA sequences. Apart from the detection of complementary DNA strands [61] , the monitoring of other analytes, for instance, mercury (Hg 2+ ) or lead (Pb 2+ ) ions is also a widespread application of fluorophore-labeled DNA sequences [62, 63] . In the first case, the detection of Hg 2+ ions is due to the conformational change induced by these ions into thymine(T)rich oligonucleotide (ON) sequences: in the presence of Hg 2+ ions, T-Hg 2+ -T mismatches are formed [64] , so the T-rich sequences acquire a hairpin structure [65] . Furthermore, those T-Hg 2+ -T base pairs quench the luminescent emission of the indicator labeled to the ON sequences [66] . For instance, the sensor described in [51] exhibited an LOD of 4.73 × 10 −13 M Hg 2+ ions in a phosphate buffered solution (pH 7.4). Due to the high affinity of Hg 2+ ions to T bases, the sensor responded in less than 25 s for the highest Hg 2+ concentrations in a reversible manner (Figure 9a) , and it showed a low cross-correlation to other metallic ions (Figure 9b ). The sensor also allowed the measurement of 5 × 10 −12 M Hg 2+ ions in ultrapure and tap water. Aptamers present the advantage of being highly specific bioreceptors [67] , so they are of great interest for the development of biosensors. Among many other applications, they can be employed for the detection of endocrine disrupting compounds, such as 17βestradiol [68] . Taking advantage of this, N. Yildirim and co-workers [69] reported an optical fiber biosensor capable of detecting 17β-estradiol concentrations from 5 × 10 −9 to 75 × 10 −9 M, with a LOD of 2.1 × 10 −9 M. Its regeneration was carried out with a 0.5% SDS solution (pH 1.9) for 90 s. Furthermore, the biosensor was not only tested in Tris-HCl buffer, but also in wastewater treatment effluent samples. Aptamers present the advantage of being highly specific bioreceptors [67] , so th are of great interest for the development of biosensors. Among many other applicatio they can be employed for the detection of endocrine disrupting compounds, such as 1 estradiol [68] . Taking advantage of this, N. Yildirim and co-workers [69] reported an o tical fiber biosensor capable of detecting 17β-estradiol concentrations from 5 × 10 −9 to 7 10 −9 M, with a LOD of 2.1 × 10 −9 M. Its regeneration was carried out with a 0.5 % S solution (pH 1.9) for 90 s. Furthermore, the biosensor was not only tested in Tris-H buffer, but also in wastewater treatment effluent samples. Antibodies can be also labeled with fluorophores. Wang et al. detected up to 5.9 ± pM concentrations of interleukin-6 (IL-6) in serum samples from lupus patients [70] . T sensor was developed by immobilizing Alexa Fluor 488-labeled anti-IL-6 antibodies o the silanized surface of a 600 μm-core tapered optical fiber, using a reflection architectu It showed a linear behavior in the pM range, as it can be observed in Figure 10 , an specificity of 100% in serum samples. Antibodies can be also labeled with fluorophores. Wang et al. detected up to 5.9 ± 0.6 pM concentrations of interleukin-6 (IL-6) in serum samples from lupus patients [70] . The sensor was developed by immobilizing Alexa Fluor 488-labeled anti-IL-6 antibodies onto the silanized surface of a 600 µm-core tapered optical fiber, using a reflection architecture. It showed a linear behavior in the pM range, as it can be observed in Figure 10 , and a specificity of 100% in serum samples. Highly sensitive and specific biosensors based on fluorophore-labeled antibodies can be achieved taking advantage of the combination of different optical techniques. It is the case of localized surface plasmon coupled fluorescence (LSPCF)-based sensors, which consist of the generation of an SPR with AuNPs, whose electromagnetic field excites the emission of the indicator labelled to the antibodies. An LSPCF optical fiber biosensor was developed by Chang et al. [71] utilizing protein A-adsorbed AuNPs and Atto633-labeled a-H1 antibodies, which were mixed and deposited along the core of a 1 mm-diameter plastic optical fiber. The sensor was based on the detection of hemagglutinin (HA) proteins, which contain the antigenic regions of the Swine-origin influenza A (H1N1) virus (S-OIV). The reported LOD was 13.9 pg/mL of HA, which is 103-fold lower than that obtained using the conventional capture ELISA, and its response was linear from 5 to 50 ng/mL. The sensor also improved the detection sensitivity of S-OIV up to 50-fold in PBS and 25-fold in mimic solution. estradiol [68] . Taking advantage of this, N. Yildirim and co-workers [69] reported an op tical fiber biosensor capable of detecting 17β-estradiol concentrations from 5 × 10 −9 to 75 10 −9 M, with a LOD of 2.1 × 10 −9 M. Its regeneration was carried out with a 0.5 % SDS solution (pH 1.9) for 90 s. Furthermore, the biosensor was not only tested in Tris-HC buffer, but also in wastewater treatment effluent samples. Antibodies can be also labeled with fluorophores. Wang et al. detected up to 5.9 ± 0.6 pM concentrations of interleukin-6 (IL-6) in serum samples from lupus patients [70] . Th sensor was developed by immobilizing Alexa Fluor 488-labeled anti-IL-6 antibodies onto the silanized surface of a 600 μm-core tapered optical fiber, using a reflection architecture It showed a linear behavior in the pM range, as it can be observed in Figure 10 , and specificity of 100% in serum samples. Highly sensitive and specific biosensors based on fluorophore-labeled antibodies can be achieved taking advantage of the combination of different optical techniques. It is th case of localized surface plasmon coupled fluorescence (LSPCF)-based sensors, which Proteins are also bioreceptors that can be used for the detection of other biomolecules [72] . It is the case for glucose binding protein (GBP), which has been widely employed for the detection of that analyte [58] . The binding of glucose to GBP induces conformational changes in that protein, which can be monitored with polarity-sensitive fluorescent probes. The one chosen by C. Tiangco et al. for the development of a glucose biosensor was 6-bromoacetyl-2dimethylaminonaphthalene (BADAN): the authors immobilized BADA-labelled GBP onto Ni-NTA agarose beads for their deposition onto the tip of an optical fiber [36] , fabricating a biosensor that was tested in vitro in PBS and in Yucatan minipig skin, which was used as surrogate for human skin. The sensor detected 2 × 10 −6 M glucose in Yucatan minipig skin. Apart from that, it presented a reversible and repetitive (relative standard deviation of 4.65% for 6 µM and 7.04% for 10 µM glucose) behavior in PBS and a response time of 15 s. The potential of luminescence-based optical fiber biosensors for real applications is shown in [73] . Here, a portable microarray biosensing platform based on four optical fiber sensors working in parallel was developed. Using optical switches, it was possible to illuminate all the fibers and to collect their fluorescent emission using just one photodiode and a single photodetector. Each of the fibers were functionalized using Cy5.5-labelled Microcystin-LR (MC-LR), 2,4-Dichlorophenoxyacetic acid (2,4-D), Atrazine (ATZ) and Bisphenol A (BPA) antibodies, respectively, in order to detect the four pollutants. In buffered solutions the sensors presented LODs of 0.04 µg/L, 0.09 µg/L, 0.02 µg/L, and 0.03 µg/L, respectively. Furthermore, the platform was also tested in real samples, where the presence of other molecules produced a negligible influence on the performance of the sensors. Apart from that, it was possible to regenerate their surfaces using 0.5% (w/w) SDS solutions. The devices reported in this section, as well as many other applications developed during the last years in the field of luminescence-based optical fiber biosensors are summarized in Tables 1-4 according to the type of bioreceptors and analytes: aptamers or DNA strands for the detection of other aptamers or DNA strands (Table 1) , aptamers or DNA strands for the detection of other analytes (Table 2) , antibodies as bioreceptors (Table 3) and other bioreceptors (Table 4 ). Ref. As explained in Section 2.2, absorption-based optical fiber sensors use the interaction between the light transmitted through the optical fiber with the surrounding media to monitor the presence of the target analyte. A crucial parameter of these sensors is the penetration depth [91] . In this sense, in order to improve their sensitivity, several parameters of the optical fiber are typically tailored, for instance, the fiber core diameter, the fiber bending, or the fiber tapering [92] , as it is depicted in Figure 11 . These sensors are usually fabricated onto multimode plastic cladding silica fiber, using a transmission architecture, which simplifies the optical set-up. 19 of 34 As explained in Section 2.2, absorption-based optical fiber sensors use the interaction between the light transmitted through the optical fiber with the surrounding media to monitor the presence of the target analyte. A crucial parameter of these sensors is the penetration depth [91] . In this sense, in order to improve their sensitivity, several parameters of the optical fiber are typically tailored, for instance, the fiber core diameter, the fiber bending, or the fiber tapering [92] , as it is depicted in Figure 11 . These sensors are usually fabricated onto multimode plastic cladding silica fiber, using a transmission architecture, which simplifies the optical set-up. Figure 11 . Different sensing schemes developed for absorption-based optical fiber biosensing: (a) the U-bent region of the optical fiber is coated with the sensing film, (b) the sensing film is deposited onto a tapered region of the fiber and (c) the sensing region is located along the fiber core. In these cases, when using optical fiber as a waveguide, evanescent wave absorbancebased techniques are employed [93, 94] . In order to enhance the interaction between the light and the sensing elements, different parameters of the optical instrumentation can be adjusted, such as the fiber diameter, the bending radius, the numerical aperture or the In these cases, when using optical fiber as a waveguide, evanescent wave absorbancebased techniques are employed [93, 94] . In order to enhance the interaction between the light and the sensing elements, different parameters of the optical instrumentation can be adjusted, such as the fiber diameter, the bending radius, the numerical aperture or the working wavelength [95] . By means of a U-bent PMMA fiber probe in order to increase the penetration depth, a label-free optical fiber biosensor platform was developed and tested against Goat antihuman IgG (GaHIgG) [96] and E. coli cells [97, 98] . First, after analyzing the relationship between the fiber diameter and the absorbance (the second increased with the decrease in the first), a 5 cm-length section of a 200 µm-core optical fiber was uncladded and U-bent. Then, it was functionalized (with either Human IgG antibodies (HIgG) or monoclonal antibodies against E. coli) and illuminated with an LED centered at 280 nm, in order to measure absorbance at that wavelength. When immobilizing HIgG onto the optical fiber for the detection of GaHIgG, the absorbance increased as the analyte concentration did. In particular, the sensor was able to detect GaHIgG from 0.1 to 50 µg/mL, with an LOD of 0.1 µg/mL GaHIgG. The utilization of metallic NPs allows the development of LSPR-based sensors, which offer the advantage of monitoring both absorbance intensity and wavelength shift [99] . Furthermore, these sensors work in the visible range, which also cheapens the cost of the optical instrumentation. For the detection of GaHIgG, Punjabi et al. [100] labelled both the antibody and the antigen with AuNPs of different sizes: an LSPR centered at 535 nm was generated when immobilizing the AuNPs-tagged HIgG onto the optical fiber, whose intensity increased and wavelength was red-shifted in the presence of AuNPs-labelled GaHIgG. The dynamic range was from 0.5 to 10 µg/mL. In the case of [101] , an optical fiber sensor for the detection on interleukin-1β (IL-1β) in synovial fluids is reported: it is based on Au NPs modified by a self-assembled monolayer of 6-mercapto-1-hexanol (MCH) and 11-mercaptoundecanoic acid (MUA), functionalized with anti-IL-1β antibodies. The sensor was capable of detecting IL-1β concentrations in the range from 4.98 × 10 −11 M to 9.95 × 10 −9 M, with a LOD of 1.2 × 10 −12 M and a sensitivity of 5.5 × 10 10 AU/M. Another example reporting the utilization of AuNPs for the development of an LSPRbased optical fiber biosensor is that presented in [102] by Xu et al. After comparing different fiber shapes in order to enhance the sensitivity towards refractive index variations, the sensor was developed onto a 600 µm-core Ω-bent fiber by functionalizing with 3-APTMS and AuNPs. Then, DNA strands for the detection of S. typhimurium were immobilized on the AuNPs. As is shown in Figure 12 , the absorbance of the sensor increased as the concentration of the target bacteria did. The sensor performance was analyzed in a Tris-HCl buffer (pH 7.4), presenting an LOD of 128 CFU/mL, a dynamic range which varied from 5 × 10 2 to 1 × 10 8 CFU/mL and a sensitivity of 0.013 AU/log(CFU/mL), and it was also capable of detecting S. typhimurium in chicken samples. Furthermore, it was tested against other bacteria, such as S. aureus, E. coli, S. enteritis and Shigella, without any cross-sensitivity. AuNPs functionalization with glucose oxidase (GOx) can be used to fabricate LSPRbased glucose biosensors, as the enzyme allows the real time detection of the target analyte [103] . The device presented in [104] , based on a U-bent fiber, exhibited a decrease in the absorbance at 540 nm with the increase in the glucose concentration, as a consequence of the change in the polarization properties of the AuNPs. Furthermore, the authors analyzed the effect of the bending radius of the sensor onto its sensitivity for a given glucose concentration of 100 mg/mL, concluding that the optimal bending radius was 0.982 mm. K. Li et al. [105] , who developed a reversible optical fiber biosensor for the detection of a cancer biomarker (alpha-fetoprotein) in serum, also studied the relationship between the dimensions of the fiber and the Au NPs: their calculations indicated that, when using an unbent fiber, the sensitivity of the sensor could be enhanced by decreasing the diameter of the fiber or by increasing the size of the Au NPs. K. Li et al. [105] , who developed a reversible optical fiber biosenso of a cancer biomarker (alpha-fetoprotein) in serum, also studied the rela the dimensions of the fiber and the Au NPs: their calculations indicated an unbent fiber, the sensitivity of the sensor could be enhanced by decrea Apart from metallic NPs [104] , polymeric thin films [106] can also be functionalized with GOx. For instance, Pahurkar et al. [107] immobilized GOx through cross-linking via 1% glutaraldehyde onto a polyaniline (PANI) layer that had been previously in-situ deposited onto the fiber core for the development of an EWA-based glucose biosensor. The interaction between GOx and glucose enhanced the π-π* transitions in PANI, producing absorption peak at 272 nm: as the glucose concentration of the samples increased, so did the absorbance at that wavelength, as it is depicted in Figure 13 . The sensor responded to glucose concentrations from 10 nM to 100 nM, presenting an LOD of 10 nM in 0.1 M PBS. Besides, it exhibited a stable time response, performing the same 36 days after its fabrication. Biosensors 2021, 11, x FOR PEER REVIEW Figure 13 . Absorbance spectra of the sensor for glucose concentrations in the rang 100 mM. Reprinted with permission from [107] . Apart from the applications described in this section, many others, A, taurine or urea detection, have been carried out during recent years. T rized in the following tables. Just for the sake of clarity, these tables hav according to the bioreceptors employed: IgG or HIgG antibodies (Table 5 ies (Table 6) , enzymes (Table 7 ) and other bioreceptors (Table 8 ). Apart from the applications described in this section, many others, such as vitamin A, taurine or urea detection, have been carried out during recent years. They are summarized in the following tables. Just for the sake of clarity, these tables have been classified according to the bioreceptors employed: IgG or HIgG antibodies (Table 5) , other antibodies (Table 6) , enzymes (Table 7 ) and other bioreceptors (Table 8) . This comprehensive review has analyzed the main luminescence and absorptionbased optical fiber biosensors that the scientific community has published in the literature during the last decade. The main objective has been to discover what kind of optical fibers and configuration set-ups are normally utilized, the main parameters to consider when fabricating and characterizing them and what applications they are used for. After all, the following lines summarize the main conclusions that arose and the trends in the state of the art when designing this kind of biosensors. From an optical point of view, most of the biosensors based on intensity measurements are designed in the UV-VIS range, using thick diameter multimode fibers and with simple configuration set-ups based typically on an LED, a bifurcated fiber and a photoreceptor. The fact of working with this kind of equipment is, surely, a good choice for real applications. An investment to implement these set-ups should not be an issue. In the case of luminescence-based biosensors, they are built on multimode fibers, which are often tapered in order to achieve a better light coupling from the fluorophores to the fiber waveguide. Here, the reflection set-up is the most used since it facilitates the light coupling from the luminescent molecules to the fiber core. Most of the reported contributions show that luminescence can be detected as long as the Stokes shift between excitation and emission peaks is higher than 20 nm. Moreover, the detections can be direct or indirect, depending on whether the luminophore is the bioreceptor directly or whether it indicates a higher or lower number of bioreceptor-analyte interactions. There is a wide range of potential applications for luminescent biosensors that can cover from the detection of DNA strands to that of water pollutants with very high sensitivity and low LOD. Regarding absorption-based biosensors, several sensing schemes have been described in order to enhance the light coupling from the sensing film towards the fiber core. Unlike luminescence techniques, a great part of the reported set-ups are in transmission, even using bent fibers to couple as much light as possible to the evanescent wave. The typical absorption wavelengths are located between 500 and 650 nm, due to the use of gold nanoparticles mainly. Although some analytes can be detected by their specific absorption at certain wavelengths, most of them are monitored thanks to the utilization of antibodies deposited onto metallic nanoparticles, directly transducing the biointeractions into light absorption. Taking into account both the wavelength ranges and the possibility of working with metallic nanoparticles, fabricating LSPR-based biosensors seems to be the most common strategy when developing absorption measurements. To sum up, the evidence shown within this review reveals the great potential of intensity-based optical fiber biosensors. Different types of biocompatible fibers; multiple sensing schemes and optical configurations to be utilized and adapt to different experimental requirements; a wide variety of bioreceptors and the possibility of labelling them with fluorophores or metallic nanoparticles. These facts, and the possibility of dealing with colorimetric techniques already used in clinical analytics and even in point-of-care applications, render this technology of high consideration and an interesting alternative for the development of simple and cost-effective future biomedical devices. Funding: This research was funded by the Spanish AGENCIA ESTATAL DE INVESTIGACIÓN (AEI), through the project with reference PID2019-106231RB-I00-ESPADA and also by the Regional Government of Navarra, through the project with reference 0011-1365-2020-000061-BREATHALYSER (FEDER). Data Availability Statement: Data is contained within the article. The authors declare no conflict of interest. Evolution of cancer pharmacological treatments at the turn of the third millennium Prevention: Public Healthcare, Nutrition, Physical Activity, Vaccination. In Learning Geriatric Medicine Start-ups focus on early-stage cancer diagnosis. Financial Times A roadmap for the early detection and diagnosis of cancer The imperial county community air monitoring network: A model for community-based environmental monitoring for public health action Integrated analysis of exosomal protein biomarkers on alternating current electrokinetic chips enables rapid detection of pancreatic cancer in patient blood Tailoring sol-gel-derived silica materials for optical biosensing Optical fiber-based sensor for in situ monitoring of cadmium sulfide thin-film growth Fiber-optic multi-sensor array for detection of low concentration volatile organic compounds Volatile organic compounds optical fiber sensor based on lossy mode resonances Optical fiber humidity sensors based on Localized Surface Plasmon Resonance (LSPR) and Lossy-mode resonance (LMR) in overlays loaded with silver nanoparticles Trends in the design of wavelength-based optical fibre biosensors Layer-by-Layer assembly of a water-insoluble platinum complex for optical fiber oxygen sensors Fluorescent biosensors-Probing protein kinase function in cancer and drug discovery Laccase-based biosensors for detection of phenolic compounds Nanophotonic biosensors for point-of-care COVID-19 diagnostics and coronavirus surveillance Micro and nanostructured materials for the development of optical fibre sensors Straightforward nano patterning on optical fiber for sensors development Fast and all-optical hydrogen sensor based on gold-coated optical fiber functionalized with metal-organic framework layer Fiber Optic Sensors: Short Review and Applications Detection and analysis of nucleic acid in various biological samples of COVID-19 patients Single chip SPR and fluorescent ELISA assay of prostate specific antigen Biological Applications of Nanoparticles in Optical Microscopy Advantageous antibody microarray fabrication through DNA-directed immobilization: A step toward use of extracellular vesicles in diagnostics Metal-enhanced fluorescence: An emerging tool in biotechnology Principles of Fluorescence Spectroscopy Encyclopedia of Spectroscopy and Spectrometry Site-directed fluorescence labeling of a membrane protein with BADAN: Probing protein topology and local environment Optical fiber nanotips coated with molecular beacons for DNA detection Repetitive immunosensor with a fiber-optic device and antibody-coated magnetic beads for semi-continuous monitoring of Escherichia coli O157:H7 Ultrasensitive quantum dots-based DNA detection and hybridization kinetics analysis with evanescent wave biosensing platform Minimizing the photobleaching of self-assembled multilayers for sensor applications Study and Optimization of the Photobleaching in Self-Assembled Optical Fiber pH Sensors Based on HPTS Using DABCO Antifading Agent Principles of Immunotherapy Developing mixed films of immobilized oligonucleotides and quantum dots for the multiplexed detection of nucleic acid hybridization using a combination of fluorescence resonance energy transfer and direct excitation of fluorescence Fiber optic biosensor for transdermal glucose based on the glucose binding protein Triple functional DNA-protein conjugates: Signal probes for Pb 2+ using evanescent wave-induced emission Erratum: Development of combination tapered fiber-optic biosensor dip probe for quantitative estimation of interleukin-6 in serum samples Photophysics and photochemistry of oxygen sensors based on luminescent transition-metal complexes The centenary of the Stern-Volmer equation of fluorescence quenching: From the single line plot to the SV quenching map Gas detection using portable deep-uv absorption spectrophotometry: A review Research on ultraviolet-visible absorption spectrum preprocessing for water quality contamination detection Novel synthesis and structural analysis of zinc oxide nanoparticles for the non enzymatic glucose biosensor Ultraviolet Absorption Spectra of Purified Virus Preparation. In Characterization of Plant Viruses UV-visible and infrared absorption spectra of transition metals-doped lead phosphate glasses and the effect of gamma irradiation Quantitative detection of H 2 S and CS 2 mixed gases based on UV absorption spectrometry Optical fiber evanescent wave absorption spectrometry of nanocrystalline tin oxide thin films for selective hydrogen sensing in high temperature gas samples Review of the Use of the Optical Fibers for Safety Applications in Tunnels and Car Parks: Pollution Monitoring, Fire and Explosive Gas Detection Optical fiber sensor based on localized surface plasmon resonance using silver nanoparticles photodeposited on the optical fiber end A wavelength-modulated localized surface plasmon resonance (LSPR) optical fiber sensor for sensitive detection of mercury(II) ion by gold nanoparticles-DNA conjugates Development of an aptamer based luminescent optical fiber sensor for the continuous monitoring of HG 2+ in Aqueous media Methods for the determination of limit of detection and limit of quantification of the analytical methods Cross sensitivity between ciprofloxacin and levofloxacin for an immediate hypersensitivity reaction Fiber Optic Sensors. Current Status and Future Possibilities Comparative study of layer-by-layer deposition techniques for poly(sodium phosphate) and poly(allylamine hydrochloride) A fiber optic biosensor: Combination tapered fibers designed for improved signal acquisition Highly sensitive optical detection of specific protein in breast cancer cells using microstructured fiber in extremely low sample volume A Noninvasive Fluorescent Biosensor for Transdermal Glucose Photochemical properties and sensor applications of modified yellow fluorescent protein (YFP) covalently attached to the surfaces of etched optical fibers (EOFs) Aptamer-based electrochemical sensors with aptamer-complementary DNA oligonucleotides as probe A portable optic fiber aptasensor for sensitive, specific and rapid detection of bisphenol-A in water samples Reusable evanescent wave DNA biosensor for rapid, highly sensitive, and selective detection of mercury ions A portable DNAzyme-based optical biosensor for highly sensitive and selective detection of lead (II) in water sample Mercury II -mediated formation of thymine-Hg II -thymine base pairs in DNA duplexes Development of novel oligonucleotide-based sensors which are highly Hg(II) selective and are insensitive to other heavy metal ions Photoinduced electron transfer mediated by π-stacked thymine-Hg 2+ -thymine base pairs Nucleic acid aptamers: Clinical applications and promising new horizons Electrochemical detection of 17β-estradiol using DNA aptamer immobilized gold electrode chip Aptamer-based optical biosensor for rapid and sensitive detection of 17β-estradiol in water samples Development of combination tapered fiber-optic biosensor dip probe for quantitative estimation of interleukin-6 in serum samples Detection of swine-origin influenza A (H1N1) viruses using a localized surface plasmon coupled fluorescence fiber-optic biosensor The Glucose Binding Protein as Glucose Sensor Portable and automated fluorescence microarray biosensing platform for on-site parallel detection and early-warning of multiple pollutants Portable evanescent wave fiber biosensor for highly sensitive detection of Shigella Evanescent wave aptasensor for continuous and online aminoglycoside antibiotics detection based on target binding facilitated fluorescence quenching Aptamer Based E-coli Detection in Waste Waters by Portable Optical Biosensor System Nanoporous layer fiber biosensing platform for real time culture-and separation-free detecting bacterial pathogens and measuring their susceptibility to antibiotics T-T mismatch-driven biosensor using triple functional DNA-protein conjugates for facile detection of Hg 2+ Integration of black phosphorus and hollow-core anti-resonant fiber enables two-order magnitude enhancement of sensitivity for bisphenol A detection Borosilicate glass fiber-optic biosensor for the detection of Escherichia coli Highly sensitive and selective optofluidics-based immunosensor for rapid assessment of Bisphenol A leaching risk Development of dual-color total internal reflection fluorescence biosensor for simultaneous quantitation of two small molecules and their affinity constants with antibodies Integrated multichannel all-fiber optofluidic biosensing platform for sensitive and simultaneous detection of trace analytes Reusable optofluidic point-of-care testing platform with lyophilized specific antibody for fluorescence detection of cholylglycine in serum First clinical evaluation of a new percutaneous optical fiber glucose sensor for continuous glucose monitoring in diabetes The Fiber Optic Biosensor for Noradrenaline Concentration Detection Based on CuPc Catalysis Fluorescent protein-based optical biosensor for copper ion quantitation A Novel Fiber Optic Biosensor for Nitric Oxide Determination Based on Vicinal Diaminobenzozcridine Fluorescent Probe Skin ethanol gas measurement system with a biochemical gas sensor and gas concentrator toward monitoring of blood volatile compounds Development of an enzymatic fiber-optic biosensor for detection of halogenated hydrocarbons Effect of taper geometries and launch angle on evanescent wave penetration depth in optical fibers Evanescent Wave Absorption Based Fiber-Optic Sensor-Cascading of Bend and Tapered Geometry for Enhanced Sensitivity U-Bent Plastic Optical Fiber Based Plasmonic Biosensor for Nucleic Acid Detection Immunogold-Silver Staining (IGSS) Based U-Bent Fiberoptic Sandwich Biosensor Optical Characteristics of Fiber Optic Evanescent Wave Sensors Label-free fiber optic biosensor based on evanescent wave absorbance at 280 nm UV Photodiode Based Portable Fiber Optic Biosensor Development of evanescent wave absorbance-based fibre-optic biosensor Localized surface plasmon resonance sensors based on wavelength-tunable spectral dips Biosensing Scheme for an On-Fiber Immunoassay Fiber-optic particle plasmon resonance sensor for detection of interleukin-1β in synovial fluids ω-shaped fiber-optic probe-based localized surface plasmon resonance biosensor for real-time detection of Salmonella typhimurium Effective immobilization of glucose oxidase on chitosan submicron particles from gladius of Todarodes pacificus for glucose sensing Localized surface plasmon resonance-based fiber optic u-shaped biosensor for the detection of blood glucose Gold nanoparticle amplified optical microfiber evanescent wave absorption biosensor for cancer biomarker detection in serum Gamma irradiated Py/PVA for GOx immobilization on tapered optical fiber for glucose biosensing Evanescent wave absorption based polyaniline cladding modified fiber optic intrinsic biosensor for glucose sensing application Augmenting Labeled Bioassay Sensitivity-Exploiting Nanogold Optical Properties on An optical Fiber Biosensor Evanescent wave absorption based S-shaped fiber-optic biosensor for immunosensing applications Evanescent wave absorbance based U-bent fiber probe for immunobiosensor with gold nanoparticle labels U-bent fiber optic plasmonic biosensor platform for ultrasensitive analyte detection Investigation of a low cost tapered plastic fiber optic biosensor based on manipulation of colloidal gold nanoparticles Hand-held optical sensor using denatured antibody coated electro-active polymer for ultra-trace detection of copper in blood serum and environmental samples A plasmonic fiberoptic absorbance biosensor for mannose-capped lipoarabinomannan based tuberculosis diagnosis Surface plasmon resonance and bending loss-based U-shaped plastic optical fiber biosensors Biconical Tapered Optical Fiber Biosensor for Real-Time Monitoring of Bovine Serum Albumin at Femtogram/mL Levels on Antibodyimmobilized Tapered Fibers Plasma enhanced label-free immunoassay for alpha-fetoprotein based on a U-bend fiber-optic LSPR biosensor Fiber optic nanogold-linked immunosorbent assay for rapid detection of procalcitonin at femtomolar concentration level Development and Testing of a Plastic Optical Fiber Grating Biosensor for Detection of Glucose in the Blood A highly selective LSPR biosensor for the detection of taurine realized on optical fiber substrate and gold nanoparticles Fabrication and evaluation of evanescent wave absorption based polyaniline-cladding modified fiber optic urea biosensor Development and Study the Performance of PBA Cladding Modified Fiber Optic intrinsic Biosensor for Urea Detection PANI-ZnO cladding-modified optical fiber biosensor for urea sensing based on evanescent wave absorption Development of uric acid biosensor using gold nanoparticles and graphene oxide functionalized micro-ball fiber sensor probe Detection of Bacteria Using Bacteriophage with Hollow Gold Nanostructures Immobilized Fiber Optic Sensor Novel SPR based fiber optic sensor for vitamin A using Au@Ag core-shell nanoparticles doped SiO 2 -TiO 2 -ZrO 2 ternary matrix A low-cost and portable dual-channel fiber optic surface plasmon resonance system