key: cord-0761556-yuqorgzd authors: Baradoke, Ausra; Santos, Adriano; Bueno, Paulo R.; Davis, Jason J. title: Introducing polymer conductance in diagnostically relevant transduction date: 2020-10-22 journal: Biosens Bioelectron DOI: 10.1016/j.bios.2020.112705 sha: b57d4a1dde48c102b827f7a674c554b1b88c793e doc_id: 761556 cord_uid: yuqorgzd In this work we demonstrate that an impedance derived capacitance method is able to cleanly resolve the resonant conductance characteristics of an electrode-confined polymer film. In decorating the film with receptors, this conductance is thereafter modulated by the capturing of specific targets, demonstrated herein with C-reactive protein. This entirely reagentless and single step marker quantification is relevant to the drive of moving assays to a scaleable format requiring minimal user intervention. The considerable growth, aging and increased mobility of the human population is driving a progressive demand for more responsive, lower cost, health services additionally able to support earlier intervention (Cėpla et al. 2020; Guo 2012; Piliarik et al. 2009 ). Infection outbreaks, such as those caused by influenza virus (H1N1) (Dotis and Roilides 2009 ) and SARS-CoV-2 (Bedford et al. 2020) , are very specific recent examples of the need to screen and monitor public health; any such capability must necessarily be underpinned by a reliable and scaleable assaying platform that requires minimal user intervention. Within this space, technologies which are "label-free" are particularly attractive and a range of such methods are available based on optically resolved dielectric change (surface plasmon resonance (Cėpla et al. 2020; Guo 2012) , piezoelectric based mass detection (quartz crystal microbalance) (Speight and Cooper 2012) and electrochemistry (Baradoke et al. 2019b; da Silva et al. 2017; Grieshaber et al. 2008; Kimmel et al. 2011; Ronkainen et al. 2010) . Of the electrochemical methods, those based on impedance spectroscopy (EIS), can be exceedingly sensitive and derived capacitive methods (ECS) equally so without any requirement for a redox probe in solution Garrote et al. 2019) . Where traditional EIS sensors primarily monitor changes in the charge transfer resistance ( ) upon a target binding event (Baradoke et al. 2019a; Lisdat and Schäfer 2008; Santos et al. 2014) these require the addition of a large excess of diffusively mobile redox probe to solution and are limited in terms of scope of supporting film (making the use of multifunctional or highly nonfouling interfaces, more than a few nanometres in thickness, difficult). ECS, on the other hand, monitors changes in the charging state of the interface. We have shown that this can be evaluated by using the electrochemical capacitance ( ) signal of the interface Bueno et al. 2012; Fernandes et al. 2013) . is mesoscopic in essence (Bueno 2018a (Bueno , b, 2019 Bueno and Davis 2020) and related to the redox density-of-states (DOS) of film-confined redox addressable functionalities. This faradaically derived capacitive signal can be used in derived clinically relevant biomarker assays with ferrocene based J o u r n a l P r e -p r o o f monolayers, mixed films, peptides and Prussian blue decorated interfaces (Cecchetto et al. 2020; Fernandes et al. 2014; Santos et al. 2014 ). Herein we demonstrate that resonant conductance of the film associated to the redox capacitive states can also be used to transduce local binding events in an entirely "reagentless" manner. In order to charge/discharge electrode-confined redox states, such as those presented by a faradaically-active polymer film, they must be electronically addressable from the electrode such that the imposition of a specific promotes a resonant exchange of electrons. If a sinusoidal wave is superimposed on this bias the interfacial capacitance is readily mapped, as is the conductance across the "molecular bridge" spanning the space between electrode and redox site (Bueno and Davis 2020) . The chemical potential of this bridge is related to the potential gradient through = − μ/ , where is the elementary charge. For an ideal quantum mechanical single electrical channel, the current is = − μ( / )( / μ), where is the number of electrons, and the charge velocity component along the channel length . Importantly, in this mesoscopic regime of scale, the "bridge" electronic properties show a dependency on the bridge DOS through ( / μ) = 2 /ℎ (ℎ is the Planck constant). This DOS is, in turn, spatially and energetically coupled to that of the (faradaically charging) redox sites in a manner that governs the charge and discharge characteristics of the film, where the associated electric current can be expressed as = (2 /ℎ) . Consequently, one can write = , where = 2 /ℎ and = 2 /ℎ is the quantum of conductance. For a (realistic) transmittance through multiple channels, the arising mesoscopic conductance is given by = ∑ (μ) = , where ∑ (μ) represents the overall transmission probability across the "molecular bridge" and equates to if the transmittance of the individual channels is ideal. A key part of this formalism, is that any change in (such as that induced by a local target capture event) will alter the mesoscopic conductance of the bridge. Traditional EIS, encompassing an analysis of the charge transfer between electrode and free species in solution (in parallel to double layer capacitance), does report on resonant conductance/resistance because the latter exists only when molecules or films are physically or chemically attached to the electrode. Even in the absence of a traditional solution phase redox probe, a standard impedimetric Nyquist analysis of an electrode coupled conductive film fails to resolve any meaningful film resistance (Figure 1c ; where only solution resistance, , is resolvable). For this we revert to a capacitance analysis. We note first that is given by the diameter of a capacitive Nyquist semicircle ( Figure 1d ) and that dominates over , for redox addressable films (where the double layer charging is heavily suppressed). The resistive component of the charging element is accessed through a Bode analysis noting that ! "# = 1/( + ) , where this inverse results in a RC time-constant corresponds to the peak frequency & ' = 1/(2(!) (Figure 1e ). Film conductance is, of course, = 1/ . This , recorded when film redox states are resonant with electrode states, corresponds to the higher limit of the conductance (Figure 2b ), but can thereafter be modulated by any change in bridge chemical potential; these changes are induced by localised (recognition) effects at the electronically coupled redox sites. can be obtained at high frequencies () = 2(&), as shown in (c) can be then estimated by the diameter of the semicircle. (e) Bode plot for the imaginary part of the capacitance as the function of frequency. The peak frequency & can be used to estimate . In prior work we have demonstrated that the capacitive analysis of redox addressable and target receptive films can be applied in signal transduction (Fernandes et al. 2013; Oliveira et al. 2019; Piccoli et al. 2018; Santos et al. 2015) , and that there is an intrinsic association between resolved molecular film capacitance and film conductance (Bueno 2018a, b; Bueno and Davis 2020) . Herein we demonstrate that, with suitably prepared conductive polymer films, the recruitment of targets is associated with a measurable change in film chemical potential and thus conductance. Such "conductometric" analyses have classically been restricted to field-effect transistor (FET) configurations. Although these possess an inherent amplification, they also require very substantial control over microfabrication and can be problematic to reproducibly scale. The conductance analyses herein are "single electrode contact" in basis and very readily scaled. We exemplify this here with a redox charging conductive polyaniline film and a quantification of C-reactive protein biomarker (CRP), a cardiac, acute phase and inflammatory marker (Black et al. 2004 ). Details about the used screen-printed electrodes and reagents are described in Supplementary Information (SI-1) and the device is shown in Figure SI All electrochemical experiments were performed with a portable PalmSens potentiostat in a three-electrode configuration, consisting of a screen-printed graphene nanoplate working electrode (SPE), a screen-printed silver reference electrode, and a screen-printed gold counter electrode. All experiments were performed in 0.1 M phosphate buffer (PB), pH 7.4, unless otherwise stated. EIS was performed between 9 kHz to 0.1 Hz (20 frequencies logarithmic arranged), with a sigmoidal AC perturbation of 10 mV (peak-to-peak). These experiments were performed at DC "redox-in" or formal potential of PANI film (−0.2 V). Impedance-derived capacitance spectroscopic was obtained via * ()) = , ** /)|,| and ** ()) = , * /)|,| , where ) is the angular frequency J o u r n a l P r e -p r o o f () = 2(&). The of the interface was obtained graphically as the diameter of the semicircle of the capacitive Nyquist plots. The conductance was calculated from the resistance determined from the peak frequency & ' of the imaginary capacitance Bode plot (! "# = 1/[/ + 0 ] = 2(& ' ) as shown in Figure 1e as example). Note that = 1/ , and that additionally can be obtained from the impedimetric Nyquist plot at high frequencies region (as shown in Figure 1c ) (Lisdat and Schäfer 2008) . The relative response (%) for the signal transduction was obtained as follows (%) = ( [ 3456 ] − 7 3 8 )/ 7 3 8 , where [ 3456 ] is the signal transduction for a certain target concentration, and 7 3 8 is the signal of the blank. The limit of detection (LOD) of the assay was determined according to 9: = 3 ?@ , where < is the standard deviation of the y-intercept of the standard plot ( (%) versus target concentration) (Shrivastava and Gupta 2011 Electrodes were washed with copious amounts of water, followed by electrochemical polishing in 0.1 M KOH, between −1.0 V to 1.3 V for 20 cycles. Electropolymerisation of pythic acid doped polyaniline (PANI-Pa) was performed in an aqueous solution of 1 mL 98% aniline and 2 mL of 50% phytic acid (w/w in H 2 O) with the addition of 17 mL of MilliQ water by applying a current density of 10 μA/cm 2 for 10 min. The film was rinsed with 0.1 M PB and then equilibrated in the same buffer for 10 min before further electrochemical characterization. Covalent immobilization of the receptor was performed by exposure of the PANI-PA films to 2.5% glutaraldehyde (in PB) for 30 min, followed by rinsing with copious amounts of PB and exposure to 5 μg mL -1 of anti-CRP or 5 μg mL -1 anti-D-dimer for 30 min. After rinsing with 0.1 M PB (pH 7.4), remaining active sites were deactivated with 10 mM ethanolamine in 0.1 M PB (pH 7.4) for 30 min, followed by rinsing with PB. Sensor surfaces were first equilibrated in PB for 30 min. Prior to a ten-minute exposure to defined target concentrations (from 0.25 µg mL -1 to 16 µg mL -1 ) or control, both spiked into 1% fetal bovine serum (FBS). Sensors were then rinsed in PB and , !, and recorded in pure PB in order to obtain conductance of the film for each target concentration. Results and discussion Results and discussion Results and discussion In generating antibody supporting conductive films, aqueous phase electrodeposition generates a stable electroactive film presenting an oxidation potential peak ( AB ) at about -0.15 V, and reduction potential peak ( 46 ) at about -0.25 V in PB solution in Supplementary Information) . is resolved to be, as expected, maximal at the electrochemical formal potential (∼ 32 µF from the semicircle diameter in Figure SI-1b) . In the absence of PANI-Pa a resolved (geometric) double layer capacitance is ∼ 0.3 µF. A subsequent standard glutaraldehyde coupling of anti-CRP results in a decrease in ( Figure SI -2), related to the attenuation of the redox signal as expected (Fernandes et al. 2013; Fernandes et al. 2014; Lehr et al. 2014) . As shown in Figure 2a and Figure SI -3, thereafter decreases as target concentrations increases (visualized at ~ 0.182 Hz, the frequency where the real component of the complex capacitance produces the semicircle from which is obtained). By analyzing the circuit elements from the Nyquist and Bode capacitive plots and frequency peak (& ' ), film conductance can J o u r n a l P r e -p r o o f be resolved to decrease on specific target recruitment (Figure 3c and Table SI-1) . This conductance decrease/resistance increase is consistent with the observed decrease in coupled electrochemical capacitance ( Figure SI-3) which in turn directly reflects the current through the "bridge", noting that = − μ( / )( / μ), where / μ is DOS. A plot of interfacial relative response ( (%) = ( [ 3456 ] − 7 3 8 )/ 7 3 8 , generates a standard curve (Figure 2d) , with a resolved sensitivity of 30.9% per decade, higher than that typically reported for capacitance sensors (Baradoke et al. 2020; Piccoli et al. 2018 ) and a limit of detection (LOD) of 1.8 µg mL -1 (10 nmol L -1 ) that is clinically relevant (Sonawane and Nimse 2017) . Interfacial conductance responses show good levels of selectivity ( Figure SI-4b) and can be accurately ( F 0.99) described by the Langmuir-type isotherm plot ( Figure SI-4a) , with a dissociation constant entirely consistent with previous reports with a J = 17 LM (Baradoke et al. 2020 ). We have herein demonstrated that the conductance of an electrode-confined redox polymer can be used as a transducer signal for biosensing. This conductance is directly correlated to the electrochemical or redox capacitance of the film that arises from its embedded faradaically-active sites. Following a relationship theoretically outlined in previous work (Bueno 2018a, b; Davis 2020) these principles represent a highly accessible, readily scaleable, single electrode contact, field effect sensing ability. Reagentless redox capacitive assaying of CRP at a polyaniline interface Properties of Anti-CA125 antibody layers on screen-printed carbon electrodes modified by gold and platinum nanostructures Screen-printed GPH electrode modified with Ru nanoplates and PoPD polymer film for NADH sensing: Design and characterization COVID-19: towards controlling of a pandemic C-reactive protein Common Principles of Molecular Electronics and Nanoscale Electrochemistry The Nanoscale Electrochemistry of Molecular Contacts Nanoscale Origins of Super-capacitance phenomena Charge transport and energy storage at the molecular scale; from nanoelectronics to electrochemical sensing Elucidating Capacitance and Resistance Terms in Confined Electroactive Molecular Layers Capacitance Spectroscopy: A Versatile Approach To Resolving the Redox Density of States and Kinetics in Redox-Active Self-Assembled Monolayers Serological point-of-care and label-free capacitive diagnosis of dengue virus infection Photografting and Patterning of Poly(ethylene glycol) Methacrylate Hydrogel on Glass for Biochip Applications Electrochemical Biosensors in Point-of-Care Devices: Recent Advances and Future Trends H1N1 influenza A infection Label free redox capacitive biosensing Comparing label free electrochemical impedimetric and capacitive biosensing architectures Perspectives on and Precautions for the Uses of Electric Spectroscopic Methods in Label-free Biosensing Applications Electrochemical biosensors -Sensor principles and architectures Surface plasmon resonance based biosensor technique: A review Electrochemical Sensors and Biosensors Label-free Capacitive Diagnostics: Exploiting Local Redox Probe State Occupancy The use of electrochemical impedance spectroscopy for biosensing Pseudocapacitance phenomena and applications in biosensing devices Redox Capacitive Assaying of C-Reactive Protein at a Peptide Supported Aptamer Interface Surface Plasmon Resonance Biosensing Electrochemical biosensors Fundamentals and Applications of Impedimetric and Redox Capacitive Biosensors Redox-tagged peptide for capacitive diagnostic assays Methods for the determination of limit of detection and limit of quantitation of the analytical methods C-Reactive protein: a major inflammatory biomarker A Survey of the 2010 Quartz Crystal Microbalance Literature J.J.D. and A.B. would like to thank Innovate UK for funding. P.R.B. and A.S. would like to thank FAPESP and CAPES.