key: cord-0759377-66yabbcx authors: van der Vossen, Jos M.B.M.; Fawzy, Ahmed; Ouwens, Anita; van Doornmalen, Joost P.C.M.; de Samber, Marc; Driessens, Rene; Heerikhuisen, Margreet; Montijn, Roy C. title: Respirator undersupply issues can safely be mitigated by applying UVC disinfection as demonstrated in challenges with Geobacillus stearothermophilus spores and SARS-CoV-2 virus date: 2022-02-10 journal: J Hosp Infect DOI: 10.1016/j.jhin.2022.01.021 sha: 3adc24d16f2eae0cbd9c90f8fb5d5b86866eae3e doc_id: 759377 cord_uid: 66yabbcx BACKGROUND: The global Covid-19 pandemic accompanied by spikes in the number of patients in hospitals, required substantial amounts of respiratory protective devices (respirators) thereby causing shortages. Disinfection of used respirators by applying UVC light may enable safe reuse, reducing shortages. Here, UVC light is explored for disinfection of respirators. AIM: To determine whether UVC disinfection is applicable to enable repeated safe reuse of respirators. METHODS: The UVC chamber (BioShift™, Signify, The Netherlands) equipped with LP mercury discharge lamps emitting at 254 nm, was used to determine the sporicidal and virucidal effects. Respirators challenged with spores and viruses were exposed to various UVC energy levels. Deactivation of the biological agents was studied and UVC effects on particle filtration properties and respirator fit. FINDINGS: A 5-log reduction of G. thermophilus spore viability by a UVC dose of 1.1 J/cm(2) was observed. By simulating spores present in the middle of the respirators, a 5-log reduction was achieved at a UVC dose of 10 J/cm(2). SARS-CoV-2 viruses were inactivated by 4 logs upon exposure to 19.5 mJ/cm(2) UVC. In case UVC must be transmitted through all layers of the respirators to reach the spores and virus, a more than 5-log reduction was achieved using a UVC dose of 10 J/cm(2). Exposure to a six times higher UVC dose did not significantly affect the integrity of the fit nor aerosol filtering capacity of the respirator. CONCLUSION: UVC was shown to be a mild and effective way of respirator disinfection allowing for reuse of the UVC treated respirators. Background: The global Covid-19 pandemic accompanied by spikes in the number of patients in 24 hospitals, required substantial amounts of respiratory protective devices (respirators) thereby causing 25 shortages. Disinfection of used respirators by applying UVC light may enable safe reuse, reducing 26 shortages. Here, UVC light is explored for disinfection of respirators. Aim: To determine whether UVC disinfection is applicable to enable repeated safe reuse of 29 respirators. Methods: The UVC chamber (BioShift™, Signify, The Netherlands) equipped with LP mercury 32 discharge lamps emitting at 254 nm, was used to determine the sporicidal and virucidal effects. Respirators challenged with spores and viruses were exposed to various UVC energy levels. Deactivation of the biological agents was studied and UVC effects on particle filtration properties and 35 respirator fit. Findings: A 5-log reduction of G. thermophilus spore viability by a UVC dose of 1.1 J/cm 2 was observed. By 39 simulating spores present in the middle of the respirators, a 5-log reduction was achieved at a UVC 40 dose of 10 J/cm 2 . SARS-CoV-2 viruses were inactivated by 4 logs upon exposure to 19.5 mJ/cm 2 41 UVC. In case UVC must be transmitted through all layers of the respirators to reach the spores and 42 virus, a more than 5-log reduction was achieved using a UVC dose of 10 J/cm 2 . Exposure to a six 43 times higher UVC dose did not significantly affect the integrity of the fit nor aerosol filtering capacity of 44 the respirator. The current Covid-19 pandemic crisis caused a large demand for personal protective equipment 54 especially respiratory protective devices (respirators) [1] . Apart from shortages of respirators, 55 considerable amounts of energy and resources are needed for production [2] and disposable use, 56 resulting in massive quantities of plastic waste [3] . In view of the shortages, some countries released 57 guidelines for using respirators by healthcare workers and public [4, 5] . To mitigate shortage, 58 decontamination procedures were explored to allow for safe reuse [6 -9] . A recent study [8] 59 demonstrated that steam sterilization at 121 °C was not a generic option for respirator disinfection and 60 appeared detrimental to many respirator brands. In-house testing was therefore advised to verify 61 respirator quality for which simple tests have been established [8, 10] . Disinfection by UVC irradiation 62 might be less harmful to respirator integrity. UVC is effective in inactivating microorganisms, microbial spores and viruses including SARS-CoV-2 [9, 11 -15]. Typically, UVC light is generated by low-pressure (LP) mercury discharge lamps, that emit a 254 nm peak wavelength, close to the optimal UVGI wavelength of 260-265 nm. Wavelengths 66 in the UVC range of 200 -280 nm are suitable to deactivate microorganisms [16] . The mechanism of 67 UVC deactivation of viruses and bacteria is primarily a photoreaction, which causes cross-linking of 68 nucleic acids [17] [18] [19] , impeding transcription and replication of DNA. UVC light also induces 69 photochemical reactions in proteins resulting in loss of host cell recognition of viruses and damage to 70 membranes and envelopes [16] . In this study, the UVC light has been applied on test samples using a closed cabinet, BioShift TM UVC 73 disinfection chamber (Signify, The Netherlands). This is a system intended for object disinfection. The 74 experimental work on this UVC disinfection chamber is aiming at providing an insight into the efficacy 75 of UVC for respirator disinfection. In addition, it should provide results on the effect of UVC on the 76 filtering capabilities of respirators to assure safe reuse of respirators for health care workers. For this study, bacterial spores of Geobacillus stearothermophilus were used as a target because 78 they represent a more robust biological agent compared to vegetative growing microorganisms and 79 viruses [10, 21] . In addition, virucidal effect of UVC against SARS-CoV-2 were explored in this study 80 because of being the main reason to disinfect respirators for reuse in the current pandemic. The BioShift UVC disinfection chamber generates UVC by classical LP mercury discharge lamps of 85 Philips, emitting a narrow spectral line at 254 nm. The cabinet is closed by doors with locks, that will 86 prevent opening during disinfection cycles, to guarantee safe use. For measuring the UVC light 87 irradiance, a ILT2400-UVGI-NB sensor, calibrated for narrow band sources such as LP mercury 88 lamps, was used. The irradiance was measured for the various positions at which spore or virus- To simulate post breathing spore contamination, a 50 µL spore suspension of 4.5 x 10 7 cfu/mL was 135 applied as a spot on inside or outside of the two types of respirators, dried during 15 min at 45°C and 136 subsequently sucked into the respirator layers with an airspeed of 28 L/min simulating inhaling or 137 conversely exhaling conditions. The spore loaded respirator materials were exposed during 1 hour to 138 a dose of UVC irradiation in the BioShift of 10 J/cm 2 . Negative controls that were not exposed to UVC 139 were included. Each condition was tested in triplicate. modifications. DMEM medium (Gibco # 41966-029) was supplemented with 5% v/v fetal calf serum 144 and contained 100 IU/mL Penicillium and 100 µg/mL Streptomycin. The load of viral infectious 145 particles was determined by 50% endpoint titre in serial dilution assay in combination with 50% tissue 146 culture infective dose (TCID50) calculation by the Spearman-Karber method [22] . For testing, 50 µL The SARS-CoV-2 loaded disks were exposed to 1.3 mW/cm 2 UVC in duplicate to 0, 15 and 60 152 seconds in the transport box placed in the BioShift chamber. In addition, the virus loaded disks 153 covered individually with material of respirator 1 and respirator 2 were exposed under identical 154 conditions in the transport box for 125 minutes thereby being exposed to 10 J/cm 2 UVC. To verify the 155 limit of UVC energy needed to realize almost complete eradication the virus on the disks covered by 156 respirator material 1, UVC exposure was done for 5.5 minutes in the transport box in the BioShift 157 chamber. The difference between the recovered infectious counts (TCID50) of non-exposed virus and 158 UVC exposed viruses was used to calculate the reduction in infectivity of the virus. Comparative testing of filter efficiency and fit of respirator material pre-and post UVC treatment The quality of the respirators, respirator 1 and respirator 2, were evaluated pre-and post UVC 162 treatment for face-fitting and total inward leakage of particles by the particle penetration test using the 163 method described by van Wezel et al. [10] 164 165 UVC measurements -calibration / UVC transmission through respirators 167 Table 1 shows details of UVC measurements including the comparison for the two dual layer stacks (Table 1) . To simulate the worst-case condition in which the spores are most far away from UVC irradiance The virus loaded disks covered individually with all layers of respirator 1 and respirator 2 were The effect of a UVC dose of 60 J/cm2 equivalent to 6 times the UVC dose to deactivate bacterial 223 spores in the respirators did not to destruct the respirators to a level where particle penetration is 224 impaired (Figure 2 ). Moreover, a negative effect on the fit could not be detected. Although the inclusion of bacterial spores of G. stearothermophilus were shown to be a suitable 255 candidate as worst-case biological agent, SARS-CoV-2 was included in this study as well. Virucidal 256 activity of UVC radiation must also be proven for respirator disinfection for later safe use. Dried 257 SARS-CoV-2 viruses on stainless steel disks showed a 4-log reduction in TCID50 upon 15 seconds of 258 exposure to UCV, which represents about 19.5 mJ/cm 2 UVC. As seen for spore-inactivation, there is a 259 relation between UVC transmission through the respirator material and the exposure time thus the 21.5 mJ/cm 2 , respectively. Comparing the results of this study with recently published information on SARS-CoV-2 eradication by 264 UVC, showed that the 4-log reduction we observed upon exposure to 19.5mJ/cm2, is in a comparable 265 range as the previously seen 2.7 log reduction by 3.4 mJ/cm 2 [26], the reduction of 3 x 10 6 dried 266 SARS-CoV-2 units by 16mJ/cm2 UVC [27] , the 3 log SARS-CoV-2 reduction by 3.7 mJ/cm 2 [13] and 267 the 5 x 10 6 SARS-CoV-2 reduction by 1048mJ/cm 2 [14] . Considering the influenza virus H1N1, a UVC Special attention should be paid to the construction of the respirators. A few respirators are on the 275 market that have their strap passed through a folded part of the mask which forms a crevice that is 276 probably less accessible for UVC . These type of respirators may not be a logical choice for UVC 277 disinfection. Since UVC is also known to be destructive to organic chemical polymers, the effect of 6 278 times higher UVC energy (60 J/cm 2 ) exposure of the respirators for disinfection was shown not to be 279 destructive to a level where particle penetration and fit of respirators were impaired to non-acceptable 280 levels. UVC is therefore concluded to be a preferred alternative method for respirator disinfection as 281 means to prevent shortages in health care institutions as it allows for a safe respirator for reuse from 282 a microbiological -and integrity perspective. J o u r n a l P r e -p r o o f Table 2 . Direction of suction Side of UVC exposure Average log reduction upon 10 J/cm2 Respirator 1 inside to outside inside 5.5 Respirator 1 outside to inside outside 5.6 Respirator 2 inside to outside inside 5.7 Respirator 2 outside to inside outside 4.6 J o u r n a l P r e -p r o o f J o u r n a l P r e -p r o o f Respirator 1 reference Respirator 1 UVC exposed Respirator 2 reference Respirator 2 UVC exposed Average particle penetration (%) Percentage particle penetration pre and post UVC exposure (60 J/cm 2 ) 0,3 μm 0,5 μm 1,0 μm 3,0 μm 5,0 μm FFP1 (22%) FFP2 (8%) FFP3 (2%) He Z. Ming W-K. Facemask shortage and the coronavirus disease In-hospital verification of non-CE-marked respiratory protective devices to ensure safety of healthcare 325 staff during the COVID-19 outbreak Inactivation of the coronavirus that induces 328 severe acute respiratory syndrome, SARS-CoV Inactivation study of Bacillus subtilis, Geobacillus stearothermophilus Alicyclobacillus acidoterestris and Aspergillus niger spores under Ultra-High Pressure 333 Innovative Food Science & Emerging 334 UV-C irradiation is highly effective in inactivating SARS-336 CoV-2 replication Susceptibility of SARS-CoV-2 to UV irradiation A pandemic influenza preparedness study: Use of energetic methods 343 to decontaminate filtering facepiece respirators contaminated with H1N1 aerosols and droplets Ultraviolet Germicidal Irradiation Handbook Ultraviolet (193, 216 and 254 nm) 348 photoinactivation of Escherichia coli strains with different repair deficiencies Determination of 50% endpoint titer using a simple formula Inactivation of bacterial spores by UVC light Spores of Bacillus subtilis: their resistance to and killing by radiation, heat and 365 chemicals Portable UV light as an 368 alternative for decontamination Rapid and complete inactivation of SARS-CoV-2 by 371 ultraviolet-C irradiation 2020 Rapid and efficient inactivation of surface dried SARS UVC mJ/cm 2 ∆ log TCID50 Respirator 1 419 ≥5,6 Respirator 2 21,5 5,15 382 Appendix Fig. A.1