key: cord-0756826-4nbhr22q authors: Payandehpeyman, Javad; Parvini, Neda; Moradi, Kambiz; Hashemian, Nima title: Detection of SARS-CoV-2 Using Antibody–Antigen Interactions with Graphene-Based Nanomechanical Resonator Sensors date: 2021-06-01 journal: ACS Appl Nano Mater DOI: 10.1021/acsanm.1c00983 sha: 9bd695479fba9f8aa62d7379be65360d66921cf7 doc_id: 756826 cord_uid: 4nbhr22q [Image: see text] Coronavirus disease 2019 (COVID-19) is a newly emerging human infectious disease caused by severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2). Early diagnosis is essential to reducing the transmission rate and mortality of COVID-19. PCR-based tests are the gold standard for the confirmation of COVID-19, but immunological tests for SARS-CoV-2 detection are widely available and play an increasingly important role in the diagnosis of COVID-19. Nanomechanical sensors are biosensors that work based on a change in the mechanical response of the system when a foreign object is added. In this paper, a graphene-based nanoresonator sensor for SARS-CoV-2 detection was introduced and analyzed by using the finite element method (FEM). The sensor was simulated by coating a single-layer graphene sheet (SLGS) with a specific antibody against SARS-CoV-2 Spike S1 antigen. In the following, the SARS-CoV-2 viruses were randomly distributed on the SLGSs, and essential design parameters of the nanoresonator, including frequency shift and relative frequency shift, were evaluated. The effect of the SLGS size, aspect ratio and boundary conditions, antibody concentration, and the number of viruses variation on the frequency shift and relative frequency shift were investigated. The results revealed that, by proper selection of the nanoresonator design variables, a good sensitivity index is achievable for identifying the SARS-CoV-2 virus even when the number of the viruses are less than 10 per test. Eventually, according to the simulation results, by using SLGS geometry determination, an analytical relationship is presented to predict the limit of detection (LOD) of the sensor with the required sensitivity index. The results can be applied in designing and fabricating specific graphene-based nanoresonator sensors for SARS-CoV-2. In late 2019, a new pneumonia case was reported with clinical manifestations similar to SARS-CoV viral infection in Wuhan, China. 1 This pneumonia is now termed Coronavirus Disease 2019 . COVID-19 is caused by a novel coronavirus from the β-coronavirus family. 2, 3 This coronavirus was named severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), which was previously known as 2019 novel coronavirus (2019-nCoV). 4 Although SARS-CoV-2 has a lower mortality rate than SARS-CoV, its transmission is much higher, and in a short time, it spread all over the world. 5 So far, more than several hundred million people worldwide have been infected with the disease and, unfortunately, about 2% of them have died. Sadly, this deplorable situation may continue for an indefinite period of time. 6 The rapid spread of COVID-19 has become a major global challenge that has posed an international public health crisis. To reduce the transmission rate and mortality of COVID-19, rapid and accurate diagnosis of infected patients can be used to prevent its spread. 7 SARS-CoV-2 has the positive single-strand RNA (+ssRNA) virus genome and is similar to SARS-CoV and MERS-COV, about 79% and 50% in terms of the genomic sequence, respectively. 8 Structurally, SARS-CoV-2 is a large-sized spherical virus, approximately 60−140 nm in diameter. 4 The SARS-CoV-2 genome encodes four major structural proteins, including Spike, Nucleocapsid, Envelope, and Membrane proteins ( Figure 1 ). Spike, Envelope, and Membrane glycoproteins exist on the surface of the virus. 9 Nucleocapsid protein packages the viral genomic RNA into virus particles, and Membrane protein is the most numerous structural protein of the virus envelope that binds to the Nucleocapsid protein. 10 Spike glycoprotein is the largest structural protein of SARS-CoV-2 and plays an essential role in the virus's pathogenesis. This protein binds to the Angiotensin Converted Enzyme 2 (ACE2) receptor on type II alveolar epithelial cells and infects the host cells. 11 So far, SARS-CoV-2 has been reported to have many genomic mutations, and most of the mutations have occurred in the gene encoding Spike glycoprotein. The mutations that occurred increase the virus' penetration into host cells and the incidence rate of the disease. 12−15 When SARS-CoV-2 enters the host body, the infected adaptive immune system of the host responds to the virus and produces antibodies against it to indirectly eliminate SARS-CoV-2. Wang et al. 16 demonstrated high titers of Spike and Nucleocapsid protein-specific antibodies in patients with severe COVID-19 (5 and 7 times, respectively). The produced antibodies bind specifically to virus proteins. 17 This binding describes a specific response between the antigen and antibody. Currently, anti-SARS-CoV-2 Spike, receptor-binding domain, or Nucleocapsid antibodies are used in the available diagnostic tests of COVID-19 and many studies on this disease. The SARS-CoV-2 detection methods are mainly classified into two categories: nucleic acid-based tests and immunological tests. The gold standard test for COVID-19 is the real-time quantitative RT-PCR (qRT-PCR), which detects the virus using the gene targets Nucleocapsid, Envelope, Spike, RdRP, and ORF1ab. 18 Currently, several alternative antigenand antibody-based tests are also in development. 19−21 Although nucleic acid-based tests are high-sensitivity and -accuracy tests, they also are time-consuming and costly techniques that require an RNA preparation step and biosafety conditions. Besides, false-negative results may occur because of the genomic diversity in viral RNA or low viral RNA in the specimens of patients. 3 Other technologies, such as graphenebased biosensors, are rapidly emanating to address these deficiencies. 22 The development of graphene-based biosensors able to indirectly detect SARS-CoV-2 specific antigens without the preparation step can help in the early diagnosis of COVID-19. There are two critical points in selecting a suitable antigen: the antigen must be a highly immunogenic, and there should be no cross-reactivity with other viruses. Seo et al. reported a field-effect transistor (FET)-based biosensor for detection of the Spike antigen of SARS-CoV-2 in clinical specimens. 23 The development of biosensors capable of detecting SARS-CoV2 or biomarkers associated with SARS-CoV2 infection will be an achievement of rapid diagnosis and pandemic management. 24 According to previous studies, 25−27 analytical methods such as biosensors can solve the problem of rapid virus detection. Mechanical nanosensors can be used to quickly detect small objects such as gas molecules, viruses, etc. Mechanical nanosensors are based on measurement of the changes in properties such as frequency, wave velocity, pressure, and strain in the presence of foreign objects. External objects adhere to the surfaces of nanosensors and cause significant changes in their natural frequency and wave velocity. By measurement and calibration of these changes, the presence of a foreign object can be quickly identified. In the past decade, a lot of research has been done on the use of graphene and carbon nanotubes (CNTs) as mechanical nanosensors. 28, 29 Graphene is a widely used and well-known two-dimensional material in a variety of biosensors. The extremely large surface area in biosensors makes it easier for analytes, such as viruses, antibodies, and nucleic acids, to bind to their surface. 30 In addition to the antimicrobial and antiviral application of graphene, 31 single-layer graphene sheet (SLGS) has very high resonant frequencies because of its very high stiffness and low weight. 32−35 This advantage improves the detection limit and operational simplicity of SARS-CoV-2 virus detection. This property makes it possible to calculate the frequency shift of tiny masses of about 1 zg (1 zg = 10 −21 g) placed on the SLGS, and its resonant frequency changes can be well identified. 28, 32 This high-speed mechanism, which can also be used to detect viruses, works with the following steps: First, the SLGS is functionalized with 1-pyrenebutyric acid N-hydroxysuccinimide ester (PBASE), and a suitable antibody and its resonant frequency are evaluated. Then, this functionalized sheet is exposed to the specimen of a patient containing SARS-CoV-2. If the Spike S1 antigens bind to the immobilized antibodies− SLGS, they alter the SLGS's resonant frequency and identify the desired antigen. 28 Simulation of the above mechanism, due to the uniqueness of the specific SARS-CoV-2 antigen to the anti-SARS-CoV-2 antibody, is a significant contribution to the design of a biosensor. There are three general approaches to predict the mechanical properties of hexagonal carbon structures such as graphene and CNTs, including molecular dynamics (MD), molecular mechanics (MM), and continuum mechanics (CM). 28,29,36−39 Although the MD method is exceptionally accurate and considers all of the interactions between carbon atoms, it is suitable for a SLGS of an approximate size of less than 10 nm. Although the MM method overcomes this problem to some extent, by increasing the size of the SLGS by more than 1000 nm, which is the minimum size required for detecting a virus with an average diameter of 100 nm, we are again faced with an increase in the computational time and resource. In the CM method, which is a fast method, the SLGS is considered to be an entirely uniform sheet with equivalent thickness and mechanical properties. 32, 40 In this study, we simulate a graphene-based biosensor functionalized with the specific anti-SARS-CoV-2 spike S1 ACS Applied Nano Materials www.acsanm.org Article simple analytical relationship is provided to predict the LOD of nanomechanical sensors that can be utilized for other viruses. 41, 42 Another important point of the present paper is to consider the statistical analysis to design sensors that help to make a more accurate simulation of reality. Also, consideration of the geometric and physical properties of the SARS-CoV-2 virus particles in the biosensor design has made the present work special for this virus. It is hoped that these study results will give a useful reference for design of the nanoresonator sensor for the rapid detection of SARS-CoV-2. ■ METHODOLOGY Selection of the Antigen. Among the four structural proteins in SARS-CoV-2, Nucleocapsid and Spike are antigen targets for virus detection. Spike glycoprotein is less expressed and less conserved than Nucleocapsid. 16 This glycoprotein has two polypeptides, the S1 and S2 subunits ( Figure 1 ). The S1 subunit is a specific antigen and has high immunogenicity for SARS-CoV-2, 7,17 and it can be ideal for the design of highly specific diagnostic tests. The SARS-CoV-2 Spike S1 antigen is specifically binding to the anti-SARS-CoV-2 Spike S1 antibody. Therefore, the anti-SARS-CoV-2 Spike S1 antibody is selected as the Spike S1 receptor on the SLGS surface. The virus detection procedure was performed in four necessary steps, according to Figure 2 . PBASE, a bifunctional linker molecule, and the anti-SARS-CoV-2 spike S1 antibodies were uniformly distributed on the SLGS (Figure 2a) . PBASE results in the immobilization of antibodies on the SLGS surface. The aromatic pyrenyl group of PBASE binds to the basal plane of the SLGS through noncovalent Π−Π interaction. 3, 22 The antibodies were then attached to the PBASE linker and immobilized on the SLGS surface. In the first step, after PBASE functionalization and immobilization of the antibody on the SLGS, the frequency of the functionalized graphene was calculated using the FEM (Figure 2b ). In the second step, several desired virus samples were randomly distributed on the sheet, using the Poisson point process (Figure 2c ,d). This model was then entered into the FEM solver, and the vibrational properties of the SLGS were recalculated (Figure 2e ). The frequency shift was obtained from the difference between the two FEM analyses, which is the criterion for virus detection. The entire sheet was covered with antibodies to trap viruses on the SLGS. Because of the increase in the probability and strength of antibody binding to the SARS-CoV-2 antigen, it could be assumed that the antibodies were entirely uniformly distributed on the SLGS and increase its concentration. The size of the anti-SARS-CoV-2 spike S1 antibody is 15 nm × 2.5 nm × 10 nm, and its molecular weight is 147400 g/mol. 43 Therefore, in this study, we assumed that, at the maximum concentration, each anti-SARS-CoV-2 spike S1 antibody with a mass of 244.77 zg occupies a circular space with a diameter of 15 nm on the SLGS. The decrease in the antibody concentration was also modeled by increasing this circular space. Calculating the average mass of each coronavirus for use in vibration analysis is an important and fundamental point that must be done carefully. According to the images reported in the literature, 4, 44 the shape of the virus can be considered to be spherical. Because the size of the virus diameter is reported to be between 60 and 140 nm, the average diameter is considered to be 100 nm in the simulations. According to the 1 g/mL density reported for the virus and with an average diameter of 100 nm, 44 the mass of each virus was calculated to be 524 ag (1 ag = 10 −18 g). Generation of a Random Distribution of the Virus. The number and distribution of viruses on the SLGS surface are completely random and depend on many variables. To investigate this problem, we need the appropriate probabilistic distribution functions. In statistics and probability, the Poisson distribution is a discrete probability distribution that describes the probability that an event will occur in a certain number at a fixed time or place, provided that these events occur at a certain average rate and are independent of the time of the last incident. 45, 46 To simulate the random distribution of the viruses, the Poisson point process was selected, which is employed by some researchers to predict an utterly random manner of discrete phenomena. 47, 48 According to this approach, the probability of the number (k) of the attached virus happening on the SLGS can be estimated as 47 Here A, N(A), and λ are the SLGS area and number and density of viruses per area, respectively. The effect of the virus density on the probability distribution of the attached viruses is shown in Figure 2c . Obviously, the number of attached viruses increases by enhancing the density of the viruses. Two typical SLGSs with different virus densities are illustrated schematically in Figure 2c . For example, for a graphene with an area of 100 μm 2 , we consider the virus density λ = 0.5 μm −2 . Here, the average number of viruses attached to the SLGS will be N = 0.5 × 100 = 50. Although the average number of viruses attached to the surface is 50, according to Figure 2c , the viruses' number in the range varies between 20 and 80. This means that any number of them have a chance of attaching independently. For example, in this case and according to Figure 2c , the probability of 40 viruses attaching is about 2%. If we select 1000 samples to investigate the statistical vibration behavior of the sensor, in 20 samples, 40 viruses must be attached to the SLGS surface. According to the simulation run time and literature, 100 random samples have been generated, which is an acceptable number to assess the statistical behavior of the nanomechanical sensor function. 47 In this study, samples were generated with a 10−1000 range of viruses per test and a virus density range from 0.1 to 10 μm −2 . In the previous study, the average viral loads in different specimens have been reported. 49 Sputum specimens include 17429 ± 6920, nasal swabs 651 ± 501, and throat swabs 2552 ± 1965 copies per test. 49 Therefore, it was estimated that the minimum number of viruses in the clinical sample is 150 copies per test, and the maximum is 17429 copies per test. Accordingly, the virus number range of 10−1000 viruses was considered to evaluate the nanosensor frequency shift. A code was written in Python to randomly distribute the location of the viruses on the SLGS. According to the number obtained from the Poisson point process, the viruses are distributed on the sheet so that they do not collide with each other. Samples of random particle scattering with different densities of the viruses are shown on the SLGS in Figure 2c . Graphene Vibration Analysis. SLGS can be modeled as a sheet with the equivalent thicknesses and mechanical properties shown in Table 1 . 50 The sheet is meshed with 10000 (100 × 100) elements using the shell element, and the constraints are applied according to different boundary conditions. Here, ω i and q ̅ i are the resonant frequency and associated ith mode shape of the vibration, respectively. The geometry and boundary conditions of SLGS can be very effective parameters in the vibrational behavior of the biosensor. 51 Therefore, the size (L) and aspect ratio (M = ratio of the width to the length) of the SLGS have been investigated in the 5−100 μm and 0.2− 5 ranges, respectively. The effect of the four conventional This process was repeated for all 100 generated samples to find the stochastic vibration characteristics of the biosensor. In this section, the effect of the number of viruses, antibody concentration, SLGS geometry, and SLGS boundary conditions on the frequency shift and the relative shift frequency of the biosensor's first three frequencies were investigated. The values of the fixed parameters in the analysis are given in the caption of the corresponding figures. In total, more than 3000 simulations were performed on a personal computer with a 4core 2.4 GHz CPU and 8 GB RAM. Depending on the SLGS geometry and boundary conditions, the run time of each simulation varied between 20 and 80 s. The first three frequencies, frequency shift, and relative frequency shift of the biosensor versus the number of viruses are plotted as semilogarithmic and log−log graphs, respectively (see the Supporting Information, Table 1S , and Figure 3a ,b). The intersection of the curve with the Y axis is the resonant frequency of the SLGS in the presence of the anti-SARS-CoV-2 spike S1 antibody. As expected, as the number of viruses increases, the resonant frequency decreases as the total system mass increases. 51 Conversely, increasing the number of viruses enhanced the frequency shift, which is our main criterion for detecting the presence of viruses. This behavior has been reported in other researches that have added only one varied mass to the SLGS. 28 ,32,50 They added a concentrated particle to the center of the plate and examined the frequency shift as its mass increased. The reported values for the frequency shift are on the order of gigahertz, with the main reason being the minimal size of the SLGS in the range of 10−20 nm. The frequency shift histograms for the first three frequencies are shown in parts c−e of Figure 3 , respectively. As shown in the figures, an acceptable frequency shift is obtained. For example, the average Δf 1 of about 0.8−25 kHz with varying numbers of 10−1000 viruses per test shows the excellent performance of the current biosensor. 35, 52 Also, having values greater than 5% for f relative can be a readable region, an excellent criterion for nanosensor design. 28 the massive spread of viruses on the SLGS, which is an acceptable value. This value has decreased to 0.3 kHz in the first frequency. This biosensor's low sensitivity to this noise, which may be caused by the high dispersion of viruses, also can be due to the high elastic modulus of the SLGS. The effect of the anti-SARS-CoV-2 spike S1 antibody concentration on the frequency shift and relative frequency shift of the first three frequencies of the biosensor is shown in Figure 4a . As shown, f relative is in the readable range (>5%) for all antibody concentration values. In the histograms of Figure 4b −d, it is found that increasing the antibody concentration from 8.5 × 10 −29 to 8.5 × 10 −27 mol nm −2 reduces Δf 1 from about 16.4 to 4.9 kHz. In choosing the antibody concentration, we face two paradoxical problems as follows: (1) If the antibody concentration is low, it will increase the sensitivity of the system and improve the efficiency and sensitivity of the sensor. This problem is quite evident in Figure 4 , and the reason is the weight loss of the system due to the low number of antibodies attached to it and thus increasing frequency changes. However, this increases the scatter results. (2) According to case 1, although a low antibody concentration is an advantage, in practice, reducing the antibody concentration reduces the chance of viruses bonding to the SLGS surface and thus reduces the chance of virus detection in the samples. According to the mentioned items 1 and 2, in this figure we sought to achieve sufficient sensitivity along with high concentration. According to Figure 4 , the value of the antibody concentration was chosen as 6.38 × 10 −27 mol nm −2 . The frequencies of the SLGSs with different antibody concentrations also are summarized in Table 2S . The effect of the SLGS length on the frequency shift and relative frequency shift is illustrated in Figure 5 . A linear relationship between Δf and the SLGS length is found in the log−log graph (left side of Figure 5 ). With increasing length due to decreasing system stiffness, the resonant frequency decreased. 50,51 This effect ultimately reduces the sensitivity of the biosensor, resulting in a reduced frequency shift. In quantities of less than 20 μm, f relative is more than 5%, which is a good range for biosensor design. The histograms in Figure 5b −d show the effect of the SLGS length on the frequency shift in the first three frequencies. Figure 5a demonstrates that increasing the length of the SLGS from 5 to 100 μm reduces the frequency shift from 65.3 to 0.1 kHz and increasing the length to more than 20 μm shows a slow decrease in f relative . Although reducing the length of the SLGS increased f relative , the limitations of making a nanosensor with smaller dimensions must also be considered. The results also are presented in Table 3S . We generally know that decreasing the plate aspect ratio due to an increase in the system stiffness increases the resonant frequency. 51 The effect of the aspect ratio of the SLGS on the frequency shift and relative frequency shift of the biosensor shows an exponential trend (Figure 6a ). According to Figure 6a , in the M < 3 range, f relative is more than 5%, which is a good range for the design. Although the effect of this parameter is not crucial in M > 1, in M < 1, it is very significant and obviously increases the frequency shift and thus improves the performance of the biosensor. Reducing the aspect ratio from 1 to 0.2 increased the frequency shift by more than 50 times, which demonstrated the importance of this parameter in the biosensor design (Figure 6b−d) . However, production limitations of the device must also be considered. Another interesting point is that the frequency ranges of the shift of the first three frequencies are almost in the same range and were independent of the aspect ratio. Table 4S also summarizes the values of the SLGS frequencies with different aspect ratios. In Figure 7 , the effect of the boundary conditions on the vibrational modes of the nanosensor is observed. In all cases, a bending mode that has different horizontal and vertical nodes is observed. The effect of the SLGS boundary conditions on the frequency shift and relative frequency shift of the biosensor ACS Applied Nano Materials www.acsanm.org Article are plotted in Figure 8 . From a structural point of view, increasing the SLGS constraints enhances the system stiffness and increases the resonant frequency. 51 This ultimately increases the frequency shift, thus improving the sensitivity of the biosensor. The CCCC and CFFF boundary conditions show the highest and lowest frequency shifts in the first three frequencies, respectively, with the four and one sides of the SLGS being clamped. The frequency shift improves from 1 to 10 kHz with a change of the boundary conditions from CFFF to CCCC. The CFCF boundry condition is usually used in the construction of these types of biosensors. 35, 53 CFCF Δf 1 is close to the CCCF value and is located between the CFFF and CCCC values. It can be deduced from Figure 8 that the f relative values of the first three frequencies are almost in the same range and are independent of the boundary conditions. The results also are listed in Table 5S . Given the effects of the SLGS length (L), SLGS aspect ratio (M), and number of viruses (N) in the above results, f relative can be estimated by the following equation: where α, β, θ, and γ are dimensionless constants, which were determined using a curve-fitting technique as follows: α = 18.2118, β = 0.6227, θ = 0.7730, and γ = 1.4064. If we consider the sensitivity index ( f relative ) to be a value of 5%, we can easily calculate LOD in terms of the SLGS length for different aspect ratios. In semilogarithmic Figure 9 , LOD is plotted for 23 reported 242 copies per test as the LOD of a FET biosensor. We demonstrated a high sensitivity for SARS-CoV-2 detection, about 10 viruses per test LOD. Therefore, our simulated graphene-based nanosensor platform provides rapid and highly sensitive detection of SARS-CoV-2. To manage COVID-19 and prevent its wide spread, we will need efficient diagnostics methods to quickly and accurately diagnose infected patients. Antigen detection assays promise a rapid turnaround time, point-of-care implementation. Point ofcare tests have the advantage of rapid, accurate, portable, and nonspecific device requirements, which provide great help for disease detection. One of the mechanical biosensors that works based on a change of the mechanical response of the system, when a foreign object is added is graphene-based nanomechanical sensors. Recent research shows that it is possible to detect very small objects using these types of sensors. Designing and calibrating these types of nanosensors to rapidly identify SARS-CoV-2 is one of the most essential and fundamental steps in its construction. In summary, we introduced a graphene-based nanoresonator sensor to diagnose COVID-19 patients and analyze using FEM. The anti-SARS-CoV-2 Spike S1 antibody was immobilized to a SLGS surface, which was used as a sensing area. Also, the effective parameters in detecting SARS-CoV-2, including the number of viruses, antibody concentration, and SLGS properties (size, aspect ratio, and boundary conditions), were investigated. It was found that the sensitivity index of the sensor was highly dependent on the geometry of the SLGS. Therefore, by careful selection of the geometry of graphene, a very low LOD could be obtained. The results showed that the sensor could detect SARS-CoV-2 ranging from 10 to 1000 viruses per test with a passable frequency shift. Finally, according to the simulation results, an analytical relationship was presented in which, by determination of the geometry of the SLGS, the LOD of the sensor could be predicted with the required sensitivity index. In conclusion, we demonstrated high sensitivity for SARS-CoV-2 detection, 10 viruses per test LOD. Therefore, our simulated graphene-based nanosensor platform provides rapid and highly sensitive detection of SARS-CoV-2. So far, no method has been introduced with higher accuracy than the gold standard technique for SARS-CoV-2 detection (qRT-PCR). Our simulated method is the first report of a method that can detect the SARS-CoV-2 virus with very high accuracy (LOD = 10 virus/test) using changes of the mechanical resonance of the nanoparticles. ■ ASSOCIATED CONTENT * sı Supporting Information The Supporting Information is available free of charge at https://pubs.acs.org/doi/10.1021/acsanm.1c00983. Tables of frequencies of the SLGSs (PDF) Clinical features of patients infected with 2019 novel coronavirus in Wuhan COVID-19 infection: Origin, transmission, and characteristics of human coronaviruses Laboratory diagnosis of coronavirus disease-2019 (COVID-19) A novel coronavirus from patients with pneumonia in China Presumed Asymptomatic Carrier Transmission of COVID-19 Dependence of the nanoresonator LOD on the SLGS size and aspect ratio in SARS-CoV-2 detection COVID-19). World Health Organization Situation Report Early Detection and Assessment of Covid-19. Front Correction to: Genomic characterization of the 2019 novel human-pathogenic coronavirus isolated from a patient with atypical pneumonia after visiting Wuhan (Emerging Microbes & Infections Neutralizing Antibodies against SARS-CoV-2 and Other Human Coronaviruses Veesler, D. Structure, Function, and Antigenicity of the SARS-CoV-2 Spike Glycoprotein The Natural History, Pathobiology, and Clinical Manifestations of SARS-CoV-2 Infections COVID-19 Coronavirus spike protein analysis for synthetic vaccines, a peptidomimetic antagonist, and therapeutic drugs, and analysis of a proposed achilles' heel conserved region to minimize probability of escape mutations and drug resistance González-Candelas, F. SARS-CoV-2 and COVID-19: A genetic, epidemiological, and evolutionary perspective SARS-CoV-2 neutralizing antibody responses are more robust in patients with severe disease Early detection of antibodies against various structural proteins of the SARSassociated coronavirus in SARS patients Molecular Diagnosis of a Novel Coronavirus (2019-nCoV) Causing an Outbreak of Pneumonia Testing for SARS-CoV-2 (COVID-19): a systematic review and clinical guide to molecular and serological in-vitro diagnostic assays Evaluation of current diagnostic methods for COVID-19 Assay Techniques and Test Development for COVID-19 Diagnosis Interaction of graphene electrolyte gate field-effect transistor for detection of cortisol biomarker Rapid Detection of COVID-19 Causative Virus (SARS-CoV-2) in Human Nasopharyngeal Swab Specimens Using Field-Effect Transistor-Based Biosensor Severe acute respiratory syndrome coronavirus-2 (SARS-CoV-2): a global pandemic and treatment strategies Nanobiosensors for virus detection in the environment. Nanomaterials for Air Remediation 2020 Development of a portable and disposable NS1 based electrochemical immunosensor for early diagnosis of dengue virus A review on applications of carbon nanotubes and graphenes as nano-resonator sensors Effect of temperature on vibrations and buckling behavior of carbon nanotube-based mass sensors using a new temperature-dependent structural model Two-Dimensional Material-Based Biosensors for Virus Detection Can graphene take part in the fight against COVID-19? Nano Today 2020, 33, 100883. (32) Natsuki, T. Theoretical analysis of vibration frequency of graphene sheets used as nanomechanical mass sensor On the molecular mechanics of single layer graphene sheets Mechanical properties of graphene and graphene-based nanocomposites Performance of monolayer graphene nanomechanical resonators with electrical readout Continuum models of multi-walled carbon nanotubes Young's modulus of graphene: A molecular dynamics study Molecular dynamics study on the thermal buckling of carbon nanotubes in the presence of pre-load Vibrational properties of two and three junctioned carbon nanotubes Linear and nonlinear free and forced vibrations of graphene reinforced piezoelectric composite plate under external voltage excitation Design and finite element simulation of metal-core piezoelectric fiber/epoxy matrix composites for virus detection A Review of Piezoelectric and Magnetostrictive Biosensor Materials for Detection of COVID-19 and Other Viruses Matching cellular dimensions with molecular sizes SARS-CoV-2 (COVID-19) by the numbers Review: Frank A. Haight, Handbook of the Poisson Distribution Effect of randomly occurring Stone-Wales defects on mechanical properties of carbon nanotubes using atomistic simulation Mechanical Behaviors of Carbon Nanotubes with Randomly Located Vacancy Defects Defect clustering viewed through generalized Poisson distribution (50) Andelic, N.; Car, Z.; Čanadija, M. NEMS Resonators for Detection of Chemical Warfare Agents Based on Graphene Sheet Formulas for Dynamics, Acoustics and Vibration Graphene mechanical oscillators with tunable frequency Detecting Ultrasound Vibrations with Graphene Resonators ACS Applied Nano Materials www.acsanm.org Article