key: cord-0755132-jjhomkfl authors: Golovkine, Guillaume R.; Roberts, Allison W.; Cooper, Chase; Riano, Sebastian; DiCiccio, Angela M.; Worthington, Daniel L.; Clarkson, Jeffrey P.; Krames, Michael; Zhang, Jianping; Gao, Ying; Zhou, Ling; Biering, Scott B.; Stanley, Sarah A. title: Practical considerations for Ultraviolet-C radiation mediated decontamination of N95 respirator against SARS-CoV-2 virus date: 2021-10-12 journal: PLoS One DOI: 10.1371/journal.pone.0258336 sha: ea1a010685f848d586146a05767f9bbdf9bff210 doc_id: 755132 cord_uid: jjhomkfl Decontaminating N95 respirators for reuse could mitigate shortages during the COVID-19 pandemic. Although the United States Center for Disease Control has identified Ultraviolet-C irradiation as one of the most promising methods for N95 decontamination, very few studies have evaluated the efficacy of Ultraviolet-C for SARS-CoV-2 inactivation. In addition, most decontamination studies are performed using mask coupons that do not recapitulate the complexity of whole masks. We sought to directly evaluate the efficacy of Ultraviolet-C mediated inactivation of SARS-CoV-2 on N95 respirators. To that end we created a portable UV-C light-emitting diode disinfection chamber and tested decontamination of SARS-CoV-2 at different sites on two models of N95 respirator. We found that decontamination efficacy depends on mask model, material and location of the contamination on the mask. Our results emphasize the need for caution when interpreting efficacy data of UV-C decontamination methods. The limited availability of N95 respirators during the SARS-CoV-2 pandemic has forced many healthcare workers to reuse respirators designed for one-time use. In these circumstances, the development of safe and efficient methods of decontamination of N95 respirators could be a partial solution to shortages [1] . The U. S. Center for Disease Control (CDC) has identified ultraviolet germicidal irradiation (UVGI), vaporous hydrogen peroxide and moist heat as the 3 most promising methods for N95 decontamination during a crisis capacity situation [1, 2] . The efficacy of UVGI for decontamination of bacteria and viruses on N95 respirators has been extensively investigated [3] . Ultraviolet-C (UV-C) exposure has been identified as an efficient method for inactivation of several viruses [4, 5] , including respiratory viruses such as influenza [6] [7] [8] , SARS-CoV or MERS-CoV [9] . However, important variability was reported a1111111111 a1111111111 a1111111111 a1111111111 a1111111111 depending on the N95 mask model studied [6, 9] . Very few studies directly evaluate UV-C mediated inactivation of SARS-CoV-2 on N95 respirators [10] [11] [12] . Fischer et al. demonstrated that UV-C could effectively decontaminate SARS-CoV-2 on N95 respirator [11] . However, this study was performed using small, flat mask coupons that do not recapitulate angular incidence and shadowing effects caused by the 3D structure of the masks and could therefore underestimate the levels of UV-C irradiation required for effective decontamination of an intact respirator [13] . Recently, Ozog et al. tested UV-C decontamination on whole N95 respirators and reported important variations of decontamination efficacy between different N95 models and material [12] , which correlates with previous results with H1N1 influenza [6, 9] . However, in this study the virus recovery from unirradiated masks for several of the models was not sufficiently above the limit of detection to determine whether effective decontamination was achieved as defined the U.S. Food & Drug Administration (FDA) as a minimum of 3 log 10 reduction in viable virus [14] . The COVID-19 pandemic and the associated shortage in N95 supplies have triggered the rapid emergence of new implementation strategies for decontamination methods and the creation of new UVGI devices [2] . During the first months of the COVID-19 pandemic, the University of Nebraska Medical Center, followed by other groups, published protocols for the implementation of UV-C based decontamination of N95 [15] . In April 2020, during the peak on the COVID-19 pandemic, the Henry Ford Health System and other hospital settings used UV-C to decontaminate N95 respirators for health care workers [12, 16] . However, as of April 2021, only one UVGI device had received an Emergency Use Authorization (EUA) from the FDA [14] . We aimed to determine the efficacy of UVGI for decontamination of SARS-CoV-2 on intact N95 respirators. To evaluate whether the 3D structure of the masks impacted inactivation of SARS-CoV-2, we tested decontamination at several sites on the respirators. We found that the efficacy of decontamination is significantly influenced by the structure of the mask and corresponding differences in irradiation. We also sought to directly evaluate whether the efficacy of decontamination varies between different models of N95 made with different materials. The SARS-CoV-2 strain used was USA-WA1/2020. Viral stocks were obtained from the Biodefense and Emerging Infections Research Resources Repository. Stocks were amplified in Vero-E6 cells (passage 1) and again in Calu-3 cells (passage 2). Virus passage 2 was used for experiments and was determined to have a concentration of 8 x 10 7 TCID 50 /ml. Additional details in S1 File. Five locations (center, top, bottom, right cheek, and strap) on the exterior of the masks were inoculated with a total of 50 μl of virus stock. The locations were selected after consideration of the 3D structure of the mask and corresponding differences in irradiation dose received to cover a wide range of anticipated doses. Aluminum coupons adhered to the center of the masks were used as a smooth and non-porous control surface. Aluminum coupons were bent to follow the shape of the mask and placed in a location calculated to receive irradiation doses equivalent to the "center" mask location. To allow surface tension to contain the virus in droplets on non-horizontal surfaces, virus was inoculated in 3 aliquots of 16.67 μl. The 3 aliquots were inoculated simultaneously and We created a UVGI device for N95 decontamination designed to generate high levels of reflection and enable ease of use via straightforward fixturing and application. The decontamination chamber consists of a metal reflecting box containing high power, commercially available UV LEDs with driver circuitry on metal core printed PCBs mounted on the vertical walls of the chamber (S1 Fig) . Individual masks were placed inside the chamber by attaching their upper and lower head bands to attachment points that ensured consistent placement inside the chamber. Eight LEDs were arrayed on each vertical sidewall in a fashion to optimize exposure dose uniformity across the surface of an N95 respirator and were calibrated to deliver a minimum irradiance of 1 mW/cm 2 across all locations of the mask (S2 Fig and S1 Table) . The temperature of the metal core circuit board, to which the UV LEDs were mounted to, remained below 41˚C for all disinfection runs. Immediately before irradiation, 2 pieces of UV tape were added to the upper right and lower left corners of the mask. Masks were placed one at a time into the device by attaching the straps on mounting points located at the top and bottom. Masks were irradiated for 300 sec or 600 sec. A picture of the mask was taken after irradiation to document UV tape change of color. To ensure that no loss of virus viability occurred due to desiccation time, virus from non-exposed control masks were harvested after all masks were exposed and biopsies were taken. Control masks were not placed into the device. Inoculated regions of the mask were cut out using 12 mm biopsy punches. Samples were incubated in 1.4 ml (mask punches and aluminum coupons) or 2 ml (strap pieces) of DMEM (Sigma-Aldrich) supplemented with 10% fetal bovine serum, 100 U/mL penicillin, and 100 μg/ mL streptomycin for a minimum of 30 minutes and subjected to gentle manual agitation at the beginning and end of the incubation. Virus was quantified by TCID 50 assay by incubating Vero E6 cells in 96 well plates with 10-fold serial dilutions in 8-fold of incubation media. Five days after inoculation, cytopathic effect, defined as any virus induced cell death or change in cell morphology, was scored visually under brightfield illumination using a 4X / 0.13 NA objective. TCID 50 was calculated using the Reed-Muench method. The limit of detection of the assay was 3.16 TCID 50 /ml, which was determined by calculating the TCID 50 at which no CPE is observed in any replicate wells. The maximum log reduction that could be achieved for each mask location is shown in S2 Table. We tested decontamination of SARS-CoV-2 on two masks models, 3M 1860 and 3M 8210. Although both masks are approved for healthcare worker use during the COVID-19 pandemic, their outer layers have different shapes and are comprised of different materials. We wondered whether these differences would impact the efficacy of UV-C based decontamination. We analyzed decontamination of 5 different inoculated mask locations (center, top, bottom, right cheek and strap, Fig 1 panel A) as well as a control aluminum coupon adhered to Each site was inoculated with 50 μl of 8e 7 TCID 50 /ml virus, applied as three aliquots of 16.7 μl. Inoculated masks were allowed to dry for 3.5 hours at room temperature in a biosafety cabinet before masks were exposed to UV-C irradiation. UV tape was adhered to masks to confirm irradiation. center of the mask. Masks were exposed to UV-C for 0, 300 or 600 seconds. The minimal doses received at each location were greater than 300 and 600 mJ/cm 2 for the 300 and 600 second exposures, respectively. Importantly, the UV-C doses used in our study were considerably lower than the cumulative doses reported to degrade N95 material [17] . The CDC has issued specific recommendations on the reuse of N95 respirators as a crisis management strategy. Accordingly, the FDA requires "Tier 3" devices (bioburden reduction) to yield at least 3 log 10 inactivation of various pathogens on N95 respirators [14] . Both UV-C doses achieved close to a 5 log 10 reduction in virus on the aluminum control coupons (Fig 1, panel B and Table 1 ), validating the efficacy of our UV-C device to eliminate SARS-CoV-2 on non-porous material. However, the 300 second exposure was insufficient for decontamination when averaging locations across the masks (Fig 1, panel B and Table 1 ). The 600 second exposure effectively decontaminated the 3M 1860 masks but failed to decontaminate 3M 8210 masks (Fig 1, panel B and Table 1 ). Notably, there was little difference between the 300 and 600 second doses on the 8210 masks regardless of location, suggesting that increased exposure time does not achieve higher levels of decontamination of this mask surface. While the reduction averaged across the entire mask was greater than 3 log 10 for the 3M 1860 mask at 600 seconds, there were important variations between different locations on the mask (Fig 1, panel C and Table 1 ). Irradiation doses were calculated using a representative 1860 model N95 mask with integrated irradiance sensors (S2 Fig). We determined that the smaller reduction in viral titer at the bottom location correlates with a lower irradiation dose received at this location. Conversely, the Right location received the highest irradiation dose and demonstrated effective decontamination of the 3M 1860 masks at the 300 sec and 600 sec exposures. However, neither dose at the Right location was sufficient for decontamination of the 3M 8210 masks (Fig 1, panel C) , despite receiving the largest dose. Furthermore, the straps were difficult to decontaminate, with large variability (Fig 1, panel C) , a result observed for other viruses [8] . We hypothesize that this is due to the strap material and potential shadowing effects caused by twists in the strap during exposure to UV-C. These results suggest that some N95 respirator models are not compatible with UV-C based decontamination. Importantly, dose validation experiments were performed with both mask models and showed similar irradiances at each location. Therefore, differences in decontamination efficacy do not result from variations in the 3D shape of the masks but are likely due to differences in mask material. The 1860 model is designed to be fluid resistant and has a smooth polypropylene outer layer while the 8210 model is not considered fluid resistant and has a polyester outer layer. Our results suggest that the 1860 facepiece is more appropriate for UV-C decontamination than the 1860 strap (braided polyisoprene), the 8210 facepiece, or the 8210 Fig) , it is possible that desiccation could increase susceptibility of the virus to UV-C. Interestingly, we noted that the outer layer of the 8210 mask presents a rougher surface than the 1860 mask, which perhaps shields the virus from sufficient UV-C exposure and prevents efficient UV-C decontamination of this mask model. Although our device was designed to expose the entirety of the masks to UV-C, our study focused on the decontamination of the outer layer of the mask only, which limits the conclusions to single-user applications. Although a system shown to effectively decontaminate both the interior and exterior of masks could streamline decontamination and allow for multipleusers, the use of respirators by single-users poses less risk of unintended transmission of SARS-CoV-2 or other pathogens [14] and could represent a viable crisis management strategy to help alleviate N95 shortages. It is important to note that UV-C decontamination methods should also ensure that N95 masks retain their fit and filtration capacities after UV-C exposure [18] , which was not tested in this study. While UV-C is an attractive method for decontamination of PPE when applied at appropriate doses that do not compromise material integrity and device functionality, our findings suggest that efficacy for individual mask models should be evaluated for a given UV-C device. Our results as well as the recent study by Ozog et al. [12] indicate that while the facepieces of some mask models can be successfully decontaminated using UV-C, others seem incompatible with this method of SARS-CoV-2 decontamination. Important factors to consider are the 3D structure of the mask and corresponding differences in irradiation dose received in some mask locations which can significantly influence the efficacy of decontamination. The straps may be particularly difficult to decontaminate and may require the use of a secondary method of decontamination in addition to UVGI [19] . However, UV-C LED technology is improving rapidly, and future devices will offer higher irradiation levels, improving penetration of UVGI and/or shortening exposure times. The identification of existing N95 models that are most suited for UV-C based decontamination or the creation of new mask models for this purpose would be important milestones that could help mitigate future N95 shortages. Implementing Filtering Facepiece Respirator (FFR) Reuse, Including Reuse after Decontamination, When There Are Known Shortages of N95 Respirators Ultraviolet-C and other methods of decontamination of filtering facepiece N-95 respirators during the COVID-19 pandemic Current Understanding of Ultraviolet-C Decontamination of N95 Filtering Facepiece Respirators Development of a Test System To Evaluate Procedures for Decontamination of Respirators Containing Viral Droplets! A method to determine the available UV-C dose for the decontamination of filtering facepiece respirators Ultraviolet germicidal irradiation of influenza-contaminated N95 filtering facepiece respirators Effectiveness of Three Decontamination Treatments against Influenza Virus Applied to Filtering Facepiece Respirators A pandemic influenza preparedness study: Use of energetic methods to decontaminate filtering facepiece respirators contaminated with H1N1 aerosols and droplets Research to Mitigate a Shortage of Respiratory Protection Devices During Public Health Emergencies Decontamination of N95 masks for reuse employing 7 widely available sterilization methods Early Release -Effectiveness of N95 Respirator Decontamination and Reuse against SARS-CoV-2 Virus The Effect of Ultraviolet C Radiation Against Different N95 Respirators Inoculated with SARS-CoV-2 The importance of the minimum dosage necessary for UVC decontamination of N95 respirators during the COVID-19 pandemic. Photodermatol Photoimmunol Photomed Recommendations for Sponsors Requesting EUAs for Decontamination and Bioburden Reduction Systems for Surgical Masks and Respirators During the Coronavirus Disease 2019 (COVID19) Public Health Emergency N95 Filtering Facepiece Respirator Ultraviolet Germicidal Irradiation (UVGI) Process for Decontamination and Reuse Rationale and process for N95 respirator sanitation and re-use in the COVID-19 pandemic Effects of Ultraviolet Germicidal Irradiation (UVGI) on N95 Respirator Filtration Performance and Structural Integrity The Importance of Fit-Testing in Decontamination of N95 Respirators: A Cautionary Note Sequential UV-C and Thermal Decontamination for Reuse of N95 Face Masks during the COVID19 Pandemic We thank Verily employees Greg Arcenio, Beth Bosworth, Warren Cai, Mike Chen, Junjia Ding, Tim English, Chopin Hua, David Heinz, Kyle Nichols, Supriyo Sinha for their valuable contributions and feedback.