key: cord-0752986-4g2rg0py authors: Simard, Carl; Richard, Jonathan; Bazin, Renée; Finzi, Andrés; Trépanier, Patrick title: Standardization of a flow cytometry SARS-CoV-2 serologic test date: 2022-01-03 journal: Cytotechnology DOI: 10.1007/s10616-021-00511-1 sha: 1dfee2223b176706fcfc66c617ed917df64d6738 doc_id: 752986 cord_uid: 4g2rg0py The SARS-CoV-2 virus is the causing agent of the coronavirus disease 2019 (COVID-19) pandemic responsible for millions of deaths worldwide. The development of the humoral response to the virus has been the subject of intensive research. A flow cytometry-based assay using native full-length SARS-CoV-2 Spike protein expressed in 293 T cells was recently proposed as a complementary seropositivity assay. The aim of our study was to further develop the flow cytometry assay and to standardize its parameters for reliable inter-laboratory use. We have optimized the protocol, established the Receiving Operating Characteristic (ROC) curve and tested reproducibility using pre-COVID and convalescent, SARS-CoV-2 individual plasma samples. The flow-based assay was simplified and standardized by cultivating the 293 T cells in suspension and expressing results in Mean Equivalent Soluble Fluorochrome (MESF) using an internal antibody positive control. The ROC curve was determined with an area under the curve (AUC) of 0.996 and the assay specificity and sensitivity were established at 100% and 97.7% respectively. Reproducibility was good as determined on multiple cytometers, on different days, and with data acquisition as far as 72 h post-staining. The standardized assay could be used as a high throughput confirmatory assay in flow cytometry laboratories involved in serological testing. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1007/s10616-021-00511-1. simplified and standardized by cultivating the 293 T cells in suspension and expressing results in Mean Equivalent Soluble Fluorochrome (MESF) using an internal antibody positive control. The ROC curve was determined with an area under the curve (AUC) of 0.996 and the assay specificity and sensitivity were established at 100% and 97.7% respectively. Reproducibility was good as determined on multiple cytometers, on different days, and with data acquisition as far as 72 h post-staining. The standardized assay could be used as a high throughput confirmatory assay in flow cytometry laboratories involved in serological testing. The coronavirus disease 2019 caused by the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) pandemic has caused over 2.1 million deaths worldwide as of January 2021 (https://www. worldometers.info/coronavirus). It has been shown that the virus Spike (S) glycoprotein binds the human angiotensin-converting enzyme 2 (ACE2) through its receptor-binding domain (RBD) (Walls et al. 2020; Hoffmann et al. 2020) . Intensive work has been done since the beginning of the pandemic to develop reliable, sensitive and specific immunosorbent assays to characterize the humoral response of humans to the virus. The seropositivity determination allows for seroprevalence studies (Bajema et al. 2020) , for the characterization of plasma in the development of convalescent plasma therapy (Chen et al. 2020 ) and could be used for the study of the overall antibody responses over time (Anand et al. 2021a) , which include vaccine response persistence in due time. Recently, Anand et al. suggested an innovative flow cytometry approach using the native full-length SARS-CoV-2 S protein stably expressed by 293 T cells (Anand et al. 2021b ). This approach circumvents the potential antibody detection limitations consequential of using RBD peptides for seropositivity determination. This protocol also addresses the potential supply chain plastic shortage some country may face by allowing testing using variable tube formats. Given that the described flow cytometry assay may be of use in many laboratories worldwide, we herein propose a simplified and standardized protocol destined to groups interested in importing the assay within their flow cytometry operations. This paper presents a standardized protocol in order to import the semiquantitative assay in a lab, a proposed Mean Equivalent Soluble Fluorochrome (MESF)-to-control-antibody ratio approach to interpret the data and the positivity threshold determination and reproducibility using convalescent donor samples. Overall, this adaptation should facilitate the integration of such a technique in any flow laboratory and sets ground for further development. Pre-COVID plasmas were obtained during regular blood donations in the province of Quebec, Canada, before December 20th 2019. It is assumed that the disease was absent from this area before this date. Convalescents plasmas were randomly chosen from adult participants in the CONCOR-1 study (#NCT04348656) having received an official diagnosis of COVID-19 by the Québec Provincial Health Authority confirmed by polymerase chain reaction (PCR). All patients have been symptomatic without hospitalisation and free of symptoms for at least two weeks before donation. All donors gave consent to participate in this research project which was approved by the Héma-Québec Research Ethic Committee. The 293 T cells stably expressing the SARS-CoV-2 Spike protein (293 T.SARS-CoV-2 Spike) were previously described (Anand et al. 2021b ). The cells, as well as non-transformed 293 T control cells, were adapted to suspension culture in 293 SFM II medium (ThermoFisher #11686029) as per manufacturer protocol. A detailed protocol is provided as supplementary material. Briefly, a mix of 293 T-SARS-CoV-2 Spike stable cell line and 293 T control cells were stained with an anti-RBD CR3022 monoclonal Ab (50 ng/ml) (graciously provided by A. Finzi) as control or plasma (1:400 dilution) for 20 min at room temperature. Optimal plasma dilution was determined by titration curve (Supplemental Data). After washing, AlexaFluor-647-conjugated goat anti-human IgG ? IgA ? IgM Abs (Jackson Immunoresearch 109-605-064) diluted 1:400 (final concentration of 3.75 lg/ml) was used as detection antibody for 20 min at room temperature. Optimal antibody concentration was determined by titration curve (Supplemental Data). After washing, samples were fixed for 10 min in formaldehyde 2% in DPBS before being centrifuged and resuspended in DPBS-BSA 0.2% (BD 554657). Data acquisition was done on BD Bioscience Accuri C6 flow cytometers in standard configuration with compensations obtained from calibration beads. Results were transformed in Alexa Fluor 647 MESF using a standard curve from beads stained with a known quantity of fluorochrome (Quantum MESF Alexa Fluor 647, Bang Laboratories, #647) (Supplemental Data). Results were reported as the ratio of the sample MESF on the CR3022 control MESF. Statistical analyses were done using R 4.0.3, using package OptimalCutpoints 1. 1-4 (López-Ratón et al. 2014) .The determination of the optimal seropositivity threshold was based on the Youden's index. The distribution of results expressed as MESF to CR3022 control antibody ratio from the plasma of N = 56 pre-COVID donors and of N = 44 convalescent donors is shown in Fig. 1A . The data obtained from the assay show a clear separation of anti-S negative and positive groups. The means of MESF ratio were 0.30 ± 0.14 and 13.95 ± 14.39 for pre-COVID and convalescent samples respectively. In order to assess the performance of the test as a classifier for COVID-19 antibodies presence and determine a positivity threshold, we established a Receiving Operating Characteristic (ROC) curve (Fig. 1B) . The area under the curve (AUC) of 0.996 indicates excellent discrimination between negative and positive samples. The optimal threshold point of 1.1 determined from Younden's index has a sensitivity of 97.7% (IC95: 88.0%-99.9%) and a specificity of 100% (IC95: 93.6%-100%). Therefore, a MESF ratio greater or equal to 1.1 relative to CR3022 50 ng/ml is considered to be indicative of seropositivity (Fig. 1A , dashed grey line). In order to validate the standardized protocol, the reproducibility was evaluated by analyzing samples on three different Accuri C6 flow cytometers, on two different days ( Fig. 2A) . The mean CV of 0.070 supports highly reproducible results. In order to accommodate higher volume of testing and more flexibility, the post-staining stability was evaluated by analyzing samples stored at 4°C in the dark after 24 h and 72 h (Fig. 2B) . The mean CV of 0.023 suggests that samples can be analyzed up to 72 h after staining without any loss of signal, and therefore, allows for an The SARS-CoV-2 pandemic has accelerated the development of many serodiagnostic platforms and inspired the establishment of alternative methods of analysis to circumvent some technical and/or supply chain challenges, such as the flow cytometry-based assay to detect SARS-CoV-2 Spike specific antibodies described by Anand et al. (2021b) . In this study, we report a protocol optimization regarding the cell culture in suspension and the quantitative measurements using a ratio of MESF and of an internal control, and propose standardization steps to facilitate the implementation of the method in any flow cytometry laboratory. In addition, we present the positivity threshold determination by comparing non-COVID-19 to confirmed-COVID-19 samples and describe how sensitive (97.7%) and specific (100%) the flow cytometry assay is. Finally, with the suggested standardization steps, we support the potential use of this flow assay as a quick and potentially high throughput confirmation test in a non-clinical setting. Given our understanding of the challenges associated with flow cytometry and our experience in interlaboratory standardization of flow assays, we ought to provide the community with a standardized baseline which could be adapted for further characterization of the SARS-CoV-2 humoral response. For example, laboratories who only have residual or low sample volume could adapt this assay to measure the IgG1/2/ 3/4, IgA and IgM distinct response all within one tube using multiple fluorescent antibodies. Also, while using 293 T cells in suspension facilitated cell culture and time management, our adaptation only allows for about 20 samples to be done from start to finish, by one technician, during a regular day. This capacity suited our testing requirement as we mostly used the technique to resolve unclear, near threshold ELISA results. Laboratories who require a higher throughput could setup an automated sampler protocol on their flow cytometer, and/or use our described 72 h poststaining stability to batch flow cytometry data acquisition. Overall, our study demonstrates that the adapted flow cytometry assay is robust enough to be used as a seropositivity confirmation assay. Altogether, our findings set the standardization basis for a SARS-CoV-2 flow cytometry assay using Spikeexpressing transfected 293 T cells and confirm the possibility of using the assay as an indicator of the serological response to the virus. Acknowledgements The authors are grateful to the convalescent plasma donors who participated in this study and the Héma-Québec team involved in convalescent donor Funding This work was supported by le Ministère de l'É conomie et de l'Innovation du Québec, Programme de soutien aux organismes de recherche et d'innovation and by the Fondation du CHUM to A.F. This work was also supported by Canada's COVID-19 Immunity Task Force (CITF), in collaboration with the Canadian Institutes of Health Research (CIHR) and a CIHR foundation grant #352417 to A.F. A.F. is the recipient of Canada Research Chair on Retroviral Entry no. RCHS0235 950-232424. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript. Data availability The datasets generated during and analysed during the current study are available from the corresponding author on reasonable request. High-throughput detection of antibodies targeting the SARS-CoV-2 Spike in longitudinal convalescent plasma samples Longitudinal analysis of humoral immunity against SARS-CoV-2 spike in convalescent individuals up to 8 months post-symptom onset Estimated SARS-CoV-2 seroprevalence in the US as of Convalescent plasma as a potential therapy for COVID-19 SARS-CoV-2 cell entry depends on ACE2 and TMPRSS2 and is blocked by a clinically proven protease inhibitor OptimalCutpoints: An R package for selecting optimal cutpoints in diagnostic tests Structure, function, and antigenicity of the SARS-CoV-2 spike glycoprotein Publisher's Note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations