key: cord-0751394-dguff28t authors: Pasquini, Celio; Hespanhol, Maria C.; Cruz, Kaíque A.M.L.; Pereira, Alexandre F. title: Monitoring the quality of ethanol-based hand sanitizers by low-cost near-infrared spectroscopy date: 2020-08-17 journal: Microchem J DOI: 10.1016/j.microc.2020.105421 sha: 4738557877c1e7bfe13e0be621e13d552e1cb714 doc_id: 751394 cord_uid: dguff28t The use of near-infrared spectroscopy (NIRS) based on a low-cost portable instrument for monitoring the quality of the three major formulations of ethanol-based hand sanitizers used for prevention against CoVID-19 disease is described. The quality of the sanitizers was evaluated using two approaches. In the first, a qualitative method was developed to identify gross non-conformities, using NIR spectral data compression by principal components analysis and projection of the spectrum of the tested sample in the principal component space delimited by samples of sanitizers prepared in the laboratory. In the second, a quantitative method was designed to determine the active substance (ethanol) employing multivariate regression based on partial least squares. The results demonstrate that the first approach can be used to detect non-conformities in the sanitizer composition, mostly associated with incorrect ethanol content. The second explores the use of NIRS for determination of the ethanol content in the three formulations aiming the quality control of the sanitizer manufacturing process. The ethanol content can be determined with an absolute root mean square error of prediction (RMSEP) equal to 0.68 % (m/m), 0.83 % (m/m) and 1.0 % (v/v) for the three formulations evaluated. The RMSEP was estimated as 1.3 % (m/m) for the commercial products. The measurement protocol takes approximately 1 min and requires only about 120 µL of a sample. Besides, NIRS was employed to compare the rate of volatilization of the ethanol in the different formulations, an important parameter concerning the efficacy of ethanol-based sanitizers. Hand sanitizers are widely recommended to assist in the prevention of human contamination by micro-organisms such as bacteria and virus [1, 2] . Several efficient formulations are commercialized. However, the most common contains ethanol as an active substance typically in concentrations close to 70 % (m/m). The composition varies from samples containing only the alcohol and water to products containing small quantities of an emollient, such as glycerol, a fragrance, a colourant, or a gel-forming substance. The gel product, knowing as "alcohol gel", is the most common and recommended hand sanitizer by the health agencies, due safety and efficacy issues [1, 2] . The coronavirus disease (CoVID-19) outbreak increased the worldwide consumption of ethanol-based hand sanitizers in an unprecedented way, as it becomes compulsory in all public and commercial places, and at homes [2] . Regulatory national health agencies, such as the Brazilian ANVISA (Brazilian National Agency of Sanitary Vigilance), took an emergency action to face the increasing demand for sanitizers [3, 4] . This measure includes the possibility of manufacturing and commercialize ethanol-based sanitizers by small companies without previous authorization from the regulatory agency. Naturally, the number of manufacturers increased substantially. Thus, the probability of frauds and nonconformities of commercial sanitizers also increased comparably. The ANVISA describes the manufacturing procedures for three formulations attested for their sanitary efficacy [4] . The simple one (F1) is just a mixture of ethanol/water containing 70 % (m/m) of ethanol. Another formulation (F2) contains ethanol at 80 % (v/v) and small amounts of hydrogen peroxide and glycerol acting as emollient. The most commercialized formulation (F3) is also based on ethanol/water solution containing 70 % (m/m) of the ethanol, using a gel-forming water-soluble polymer such as carbomer, after a pH adjustment between 5.0 to 7.0, using triethanolamine. Furthermore, as a consequence of the flexibilization action and also due to the requirement for quality control of the ethanol-based sanitizers, fast, green and low-cost analytical methods become even more necessary. Near-infrared spectroscopy (NIRS) fulfil these characteristics [5] , and was used to determine the amount of ethanol in several commercial products [6] . The ethanol content of beverages and spirits was successfully determined using NIRS [7] [8] [9] . The evaluation of fuel ethanol quality and the quantity of the alcohol additive found in gasoline were performed using NIRS [7, [10] [11] [12] . In common, these works employ costly bench-type NIR spectrophotometers or instrument prototypes [13] , and quantitative results were based on the use of multivariate regression methods. The literature lacks in describing direct analytical methods for quality control of the composition of ethanol-based sanitizers. So far, only one communication very recently published on-line describes the determination of ethanol in hand sanitizers using infrared (IR) and near-infrared spectroscopy (NIRS) [14] . This communication attempted to determine the ethanol content of sanitizers neglecting the effect of interferents and employing costly NIRS instrumentation (above US$ 20,000), which restricts their extensive use by small companies and by inspection agencies. Besides, the use of NIRS was not fully explored for quality assessment of sanitizers, aiming to provide, for example, a fast qualitative screening of non-conformities, and information about the rate of volatilization of ethanol from different formulations. The present work aims to improve the use of NIRS as a fast, green, and nonexpensive technique to monitor comprehensively the quality of ethanol-based sanitizers. The study was conducted employing a low-cost portable NIR spectrophotometer (~US$ 1,000), and a straightforward sample measurement protocol. The spectral data were analyzed using multivariate methods based on principal component analysis (PCA) and partial least square regression (PLSR) for qualitative, and quantitative purpose, respectively. Samples of sanitizers (25 g for F1 and F3; 25 mL for F2) were prepared according to with the three formulations preconized by ANVISA [4] described in Table 1 and worldwide adopted [1, 2, 15] . The content of ethanol in the stock solution was 92.15 % (m/m), as determined using an automatic densimeter (Anton Parr, DMA 5000M, Austria). Deionized water was used throughout. The ethanol content was varied from 40 -85 % (m/m) for F1 and F3, and from 40 -90 % (v/v) for F2. The quantities of additives were kept constant at the recommended amount by the regulatory agency. The number of samples prepared was 18, 17, and 18 for F1, F2, and F3 formulation, respectively. Additional samples of formulation F2 and F3 were produced changing the content of additives and used for interference evaluation. For F2 the quantity of glycerol (Polytecnica Química, Brazil, 98 % (m/v)), and hydrogen peroxide (Dinamica, Brazil, 3 % (m/v)) were changed. For F3 the quantity of glycerol, carbomer (3V Sigma, USA, 98-100 % (m/m)), and triethanolamine (Nitrogenius, Brazil, 85 % (m/v)) were varied. The ethanol content was kept as recommended (Table 1) . Forty-one commercial samples of sanitizers were purchased in the local commerce of Viçosa, Minas Gerais state, Brazil. Two of these samples were formulated as ethanol/water solutions (F1). All others were formulated as a gel (F3). No commercial sample was found employing formulation F2. Thirty samples are formulated as a gel (F3) employing carbomer (a cross-linked polymer of acrylic acid) as a gel-forming agent, two samples employ Astragalus gummifer (a natural polysaccharide), and three uses hydroxypropyl methyl cellulose for the purpose. Other six gel samples make no mention of the gel-forming substance employed. Some labels products declare they contain a fragrance and/or moisturizing agents such as aloe vera and preservers in non-specified quantity, among several other additives. All gel-based and water/ethanol commercial samples specify a 70 % (m/m) content of ethanol in their labels. The portable NIR -Generic Model (Young-Green, Hsinchu, Taiwan) based on the low-cost NanoNIR (Texas Instruments Inc., Dallas, Texas, USA) was employed. The instrument was previously evaluated for its signal-to-noise ratio for its monitoring spectral region (900 -1700 nm) [16] . The best signal-to-noise ratio spectral region (960 -1664 nm) was employed [16] . The instrument was held in place using a ring stand and a clamp, as shown in the Figure S1 . The optical measurement window was pointed down and kept at 6 mm above the ring stand platform. The instrument is connected to the controller microcomputer running a property software (Texas Instrument, DLP NIRscan Nano GUI v 2.2.0). Spectra were obtained using the Hadamard multiplexed measurement protocol [13] aiming to improve the signal-to-noise ratio (SNR). The spectrophotometer operated with an automatic gain of the detector amplifier. The spectra were acquired as the average of 20 scans with 228 points obtained in the range 900 -1700 nm. The total scan time per sample is 5.7 s. Only one average spectrum was acquired per measured sample. Kinetic data were obtained employing the back-to-back scan facility available in the controlling software. The transflectance sample cell is depicted in the Figure S2 . The reference signal for absorbance calculation was obtained every 20 min time interval, employing the empty closed-cell. Spectra files are named and stored in ASCII format. About 100 mg of F1 and F3 samples (containing ethanol as recommended by ANVISA) and two commercial samples were delivered to the open measurement cell. A reference for absorbance calculation was obtained using the empty open cell. The experiments were conducted at the temperature of (20 ± 1) o C. Spectra sets were obtained using the functionality of the back-to-back scan of the spectrophotometer software during a 20 min in time intervals of 1 min, while the sample volatilizes. The number of scans for average was ten, allowing to reduce the acquisition time to 3.7 s. Principal components analysis and partial least square regression [17] were performed using the Unscrambler 11.0 chemometric package (CAMO, Norway). The default pre-treatment of the spectral data set was performed sequentially by Savitizky-Golay smoothing (2 o. polynomial, 9 points window) of the full spectrum, area normalization, and standard normal variate (SNV) of the spectral region between 960 -1664 nm [17] . Data has been mean-centred before multivariate analysis. Before starting the multivariate analysis, the best pre-treatment of the spectral data was established. Although the spectrum acquisition of a sample is very rapid and simple, it is difficult to maintain constant the sample quantity delivered into the measurement cell. As a consequence, its optical path changes as a function of the sample quantity and pressure imposed by the operator on the glass window. This fact generates a source of undesirable spectral variability. Figure 1 shows three sets of five spectra each of F1 (ethanol, 70 % (m/m)), F2 (ethanol 80 % (v/v)) and F3 (ethanol, 70 % (m/m)) sanitizers before ( Figure 1A) and after ( Figure 1B ) transformation by smoothing, area normalization and standard normal variate [17] . The transformation is effective to correct the spectra set for unwanted sources of variability, keeping the original spectrum profile. Therefore, all data set was pre-treated this way before conducting the chemometric analysis. In Figure 1B is possible to observe that the F1 and F3 formulation samples containing the same recommended amount of ethanol (70 % (m/m)) show very similar spectra (indistinguishable in Figure 1B) , despite F3 being formulated as a carbomer gel. F2 sample shows slightly different spectra, probably due to the addition of glycerol, whose molecule presents three O-H groups, absorbing strongly and overlapping the ethanol spectrum in the ethanol characteristic absorption spectral regions. The larger ellipse was arbitrarily inserted to delimit the ethanol content tolerance between 62 -75 % (m/m). This range was found to be capable of inactivating the coronavirus, causing CoVID-19 disease [15] . Commercial samples whose scores are located outside the large ellipse can be screened as non-conform or fraudulent products. For instance, commercial samples sold as alcohol gels labelled as "a", "b", "c-e", "f" and "g" in Figure 3B , present a relatively high distance from the scores of the samples with correct composition. These samples can be screened as non-conforming samples. Gross frauds, exemplified by sample labelled as "a" in Figure 3B , found very close to an F3 with only 52 % (m/m) ethanol, can be easily identified. Five samples were detected to have an ethanol content somewhat higher than the recommended. This samples may have a higher content of ethanol or may have employed glycerol as an additive. Three of those samples declare in their labels to contain glycerol in a non-specified quantity. In contrast, among the seven samples identified as non-conform, four declares to contain glycerol. As this substance exert a positive interference on the content of ethanol detected by NIRS, it consists of additional evidence of fraud. Each of the three sets of samples prepared in the laboratory was split into two subsets. Besides the samples prepared according with the composition recommended by ANVISA, the calibration and validation sets included x additional F3 samples containing variable contents of the additives. One set was employed for PLSR model development aiming at the determination of the active substance (ethanol) in the sanitizers of the three formulations; the other was used for model validation. Table 2 The determination of ethanol content in the three formulations may suffer interference of other additives and/or the change in the content of substances such as glycerol, hydrogen peroxide and carbomer already present in formulations F2 and F3 as recommended by ANVISA. The variety of composition found in the commercial alcohol gel samples is astonishing. Thirty-seven additives were reported in the labels of the commercial products used in this study. The most common are glycerol, denatonium benzoate (denaturant), aloe barbadenis, aminomethyl propanol, propylene glycol, and fragrances. Though the manufacturers mention the composition, they do not mention the quantities of the additives. Nevertheless, they refer to substances that must be present in small quantities in the sanitizer formulation [2] . Therefore, they should not affect significantly the determination of the ethanol content using the PLSR models developed in this work. The ethanol content of the prepared samples was predicted using the previously validated model for each formulation. The results are shown in Table 3 . As can be observed, glycerol can be identified as the most interfering concomitant. Overall, the interference caused by variable quantities of additives in formulations F2 and As previously mentioned, positive interference caused by the glycerol can be anticipated, as its molecule present three O-H groups and its spectrum overlaps that of ethanol almost along the entire NIR spectral region employed in this work. This fact can explain the higher ethanol content (77 % (m/m)) obtained for only one sample whose label declares to contain glycerol. However, the confidence interval (called in the Unscrambler as "deviation" [17] ) of this result shows that the sample must be classified as a prediction outlier. This fact reveals that the multivariate regression model can detect commercial samples whose composition are significantly different of the recommended by the regulatory agency. On the other hand, as the quantity this substance added in the gel formulations must not exceed 1 % (m/m) [2] , the interference can be considered negligible. Therefore, is possible that this sample may also have a higher ethanol content. Considering that the efficiency of the gel sanitizer to inactivate the SARS -Cov The time interval the active substance, ethanol in the present case, remains in contact with the hands during sanitization is a relevant factor regarding the sanitizer efficiency [2, 15] . Ethanol is volatile at room temperature, and it is lost during the sanitization procedure. Several factors, such as the presence of additives, composition and the viscosity of the gel, can influence the volatilization rate of ethanol [2] . Therefore, the F1 and F3 formulations and two commercial products (S18 and S19) employing different gel-forming (Astragalus gummifer, and hydroxypropyl methylcellulose), containing ethanol close to 70 % (m/m), as recommended by ANVISA, were evaluated for their volatilization rate using NIRS. Figure 5A and 5B show the pre-treated spectrum sets obtained for the volatilization experiment carried out using the F1 sample and the commercial S18, which employs Astragalus gummifer as gel-forming, respectively. The content of ethanol associated with each spectrum obtained during the volatilization experiments was determined using the joint F1+F3 PLSR model. The use of this model should cause a bias in the ethanol concentration results because the model was constructed using the closed-cell. Nevertheless, the results permit to evaluate the volatilization rate in a relative way. As can be seen in Figure 5C , the F3 formulation presents a lower volatilization rate of ethanol than F1. This characteristic may be attributed to the gel matrix. Sample S18, based on Astragalus gummifer gel-forming, shows the lowest rate of volatilization. Although based on a preliminary study, the results obtained demonstrate that NIRS can be used to assess other parameters associated with the efficacy of the ethanol-based sanitizer. Not only the content of active substance can be determined by NIRS. The effect of other quality parameters associated with concomitant factors such as formulation as a gel or not, and gel characteristics can also be assessed. Considering these preliminary results, perhaps, under more controlled conditions, the volatilization rate assessed by NIRS may be used as a quality test for ethanol-based sanitizers in the future. The use of near-infrared spectroscopy (NIRS) for quality control of the most frequent The cost of the spectrophotometer (~ US$ 1,000) and the sample cell employed in this work is affordable to small manufacturers which are helping to fight the CoVID-19 outbreak. The instrument has been previously evaluated by its long-term performance with excellent results considering the cost/benefit ratio [16] . The chemometrics software package "R" [19] available free of charge can be used to perform the necessary multivariate analysis of spectral data in case a commercial software is not available. The analytical procedure is straightforward, and manufacturers can produce their models (with the desired composition, including additives not studied in this work) to improve the performance, and effectively control the quality of the raw substances employed for Haskell Ltda for supplying most of the reagents employed in this study. Hand sanitisers amid CoViD-19: A critical review of alcohol-based products on the market and formulation approaches to respond to increasing demand Ministério da Saúde -MS Agência Nacional de Vigilância Sanitária -ANVISA Resolução de diretoria colegiada -RDC Nº 350 de 19 de Orientações Gerais para Produção de Formulações Antissépticas Alcoólicas Near infrared spectroscopy: A mature analytical technique with new perspectives: A review Near infrared spectroscopic determination of alcoholssolving non-linearity with linear and non-linear methods Determination of ethanol in fuel ethanol and beverages by Fourier transform (FT)-near infrared and FT-Raman spectrometries Comparison of non-invasive NIR and Raman spectrometries for determination of alcohol content of spirits Alcohol determination in distilled alcoholic beverages by liquid phase fourier transform mid-infrared and near-infrared spectrophotometries Simultaneous determination of methanol and ethanol in gasoline using NIR spectroscopy: Effect of gasoline composition Classification and determination of alcohol in gasoline using NIR spectroscopy Optimised determinations of water I ethanol by encoded photometric near-infrared spectroscopy: A special case of sequential standard addition calibration Dual-beam near-infrared hadamard spectrophotometer Determination of ethanol in gel hand sanitizers using mid and near infrared spectroscopy Thoughts on what chemistry can contribute to fighting SARS-CoV-2 -A short note on hand sanitizers, drug candidates and outreach Feasibility of a portable, low-cost nearinfrared spectrophotometer for the quality screening of omega-3 dietary supplements Multivariate data analysis Practical guide to interpretative near-infrared spectroscopy CP and MCH proposed the use of NIR low-cost portable instrument for monitoring the quality ethanol samples ☒ The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work The authors acknowledge the Instituto Nacional de Ciências e Tecnologias Analíticas