key: cord-0744139-77avzyuq authors: Sagami, Ryota; Nishikiori, Hidefumi; Sato, Takao; Murakami, Kazunari title: Endoscopic shield: barrier enclosure during the endoscopy to prevent aerosol droplets during the COVID-19 pandemic date: 2020-05-11 journal: VideoGIE DOI: 10.1016/j.vgie.2020.05.002 sha: 7e61aaa5ab83be1bcc95a383a5626b9b81938772 doc_id: 744139 cord_uid: 77avzyuq nan Endoscopic shield: barrier enclosure during the endoscopy to prevent aerosol droplets during the COVID-19 pandemic Coronavirus disease 19 (COVID-19) refers to human infection with severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2). The World Health Organization (WHO) declared COVID-19 a pandemic on March 11, 2020. By April 21, 2020, the number of confirmed COVID-19 cases had increased to more than 2,397,000 globally, with an overall mortality rate of 6.8%. 1 COVID-19 is mainly spread through direct contact or aerosol droplets. 2, 3 Whether endoscopy represents an aerosol-generating procedure remains unclear, but insufflation during endoscopic procedures could cause splash and aerosol droplets due to processes such as reflex vomiting, sneezing, and coughing. Contamination by such aerosol droplets may increase the risk of SARS-CoV-2 transmission. In areas with many infected patients, about 10% of healthcare personnel (HCP) have been shown to contract COVID-19, and protection of HCP is recommended. 4 However, HCP involved in endoscopy face substantial risks due to the short physical distance between patients and personnel. Many endoscopic centers have thus reduced normal endoscopic activities. 5 Triage and assessment of risk from patients with suspected or confirmed COVID-19 before endoscopy, regular monitoring of the supply and use of personal protective equipment (PPE), and performance of endoscopies in a negative pressure room, when available, with strict isolation precautions for suspected or confirmed cases of COVID-19 are recommended. 2 In addition, urgent 2 endoscopies, such as for upper GI bleeding or severe cholangitis, should be performed by strategically assigned HCP to minimize risks of concomitant exposure. 2 Protection against COVID-19 infection during endoscopic procedures due to splash and aerosol droplets should be considered from not only the side of the HCP, but also the side of the patient. Newly developed barrier enclosures for use during endotracheal intubation may reportedly provide additional protection as an adjunct to standard PPE, from the perspective of preventing the spread of aerosol droplets. 6 With reference to that report, the Endoscopic Shield was developed. This plastic cube barrier was designed to cover the head of the patient during upper GI endoscopy. This device is a plastic cube with sides of 45 centimeters. Four square 1.5-mm-thick plastic plates were glued together, with 2 holes in the plate on the facial side. With the patient in a left lateral position, the foot side and left side of the cube are open, and the head of the patient is covered, especially on the facial side (Fig. 1A) . In our situation, a 170-cm tall endoscopist in standard PPE performed upper gastrointestinal endoscopy on a mannequin with the mouthpiece placed as usual. The facial side of the cube contains 2 small holes, each 2 cm in diameter, and the endoscope is inserted through one of the holes, selected according to the size of the patient's face and the height of their mouth ( Fig. 1B and C) . The other hole is closed with medical tape (Fig. 1C) . The size of the hole was determined based on the diameter of the TJF-260V side-view endoscope (Olympus Medical Systems, Tokyo, Japan) with a diameter of 13.5 mm and working space because that endoscope has the largest diameter of any endoscope in Japan and is used for urgent endoscopic retrograde cholangiopancreatography. A single episode of reflex vomiting or cough was simulated by bursting a small nitrile rubber balloon containing 10 mL of fluorescent dye. The balloon was 3 attached to a tube and placed in the hypopharynx of the mannequin, then oxygen was pumped through the tube inside the mannequin until the balloon burst. Simulated endoscopy was then performed without and with placement of the endoscopic shield. The scene for each simulation was illuminated with ultraviolet light to visualize the area of scattered dye droplets. In the situation without the endoscopic shield, dye was clearly identified on the right glove, arm, upper chest, abdomen, and socks of the endoscopist (Fig. 2A) . Barely any dye was identified on the face mask or eye shield. Contamination of the floor occurred within approximately 1.5 m from the head of the bed (Fig. 2B) . With the endoscopic shield, most dye was identified on the inner front surface of the cube, on the inner upside surface to a lesser extent, and not on other surfaces (Fig. 3) . A little dye was found only on the right glove, arm, and chest of the endoscopist, with none on the abdomen, socks (Fig. 2C ), face mask, or eye shield. Contamination of the floor was also minimal (Fig. 2D) . The endoscopic shield thus appeared to protect against the widespread dispersal of aerosol droplets. Given this situation, the endoscopic shield could prevent the majority of splash and aerosol droplets from a patient, markedly reducing the exposure of HCP including endoscopists and the surrounding environment such as floors. Next, use of the endoscopic shield for a live patient was evaluated. One of the authors underwent upper gastrointestinal endoscopy with the endoscopic shield without conscious sedation, after informed consent was obtained. The patient was placed in a left lateral position, and the endoscopic shield was placed to cover the head. There was no discomfort, and the vital signs did not change. When the endoscopic procedure is performed with conscious sedation, the cube can be placed after the start of sedation. The endoscope was inserted through one of the holes on the facial side of the cube. A disposable rubber check valve was attached to one hole to suppress infection 4 with contact of the endoscope. During the procedure, head repositioning was performed without difficulty because of the sufficient working space in the cube, and a nasal cannula was smoothly inserted through the foot side of the cube during the procedure. The standard gastrointestinal screening was completed in about 10 minutes without problems, and there was no interference between the endoscope and the endoscopic shield. After the procedure, many droplets due to reflex vomiting and coughing were confirmed on the inner front surface of the cube. The attached check bulb was abandoned, and the cube was carefully washed and wiped with sodium hypochlorite. Because the endoscopic shield is an external device, the device could be reusable with such disinfection. The device may be useful for all patients who undergo upper gastrointestinal endoscopy, including urgent endoscopic procedures. However, the device may have little to no clinical applicability for patients undergoing general endotracheal intubation, because aerosolization would not be a concern in such cases. The new device was easy to use, and no particular training was required. Of course, our simulation only partially reproduced the true situation of endoscopy and could not identify whether the shield prevented widespread dispersal of smaller aerosol particles that could still prove infectious. In addition, access to the patient may be restricted by the device depending on the situation. However, the shield has the possibility of providing an additional level of protection against COVID-19 infection spreading during necessary endoscopic procedures. The prototype of the endoscopic shield was easy and inexpensive to produce, at a price of $40 USD; therefore, commercialization of this device is now planned. However, the shape and durability should be considered in detail for it to be commercialized. The endoscopic shield could prevent widespread dispersion of aerosol droplets, and HCP, including endoscopists and surrounding environments such as floors, were exposed only slightly 1 Video demonstration of the way to use the endoscopic shield and the effectiveness of the device from the perspective of preventing the spread of aerosol droplets. World Health Organization Practice of endoscopy during COVID-19 pandemic: position statements of the Asian Pacific Society for Digestive Endoscopy Aerosol and Surface Stability of SARS-CoV-2 as Compared with SARS-CoV-1 COVID-19: protecting health-care workers Low risk of covid-19 transmission in GI endoscopy Barrier Enclosure during Endotracheal Intubation with endoscopic shield use. Therefore, the endoscopic shield may reduce the spread of COVID-19 infection during endoscopy.