key: cord-0726381-w2dk0ec7 authors: Princeton Open Ventilation Monitor Collaboration,; Bourrianne, P.; Chidzik, S.; Cohen, D. J.; Elmer, P.; Hallowell, T.; Kilbaugh, T. J.; Lange, D.; Leifer, A. M.; Marlow, D. R.; Meyers, P. D.; Normand, E.; Nunes, J.; Oh, M.; Page, L.; Pereira, T.; Pivarski, J.; Schreiner, H.; Stone, H. A.; Tank, D. W.; Thiberge, S.; Tully, C. title: Inexpensive multi-patient respiratory monitoring system for helmet ventilation during COVID-19 pandemic date: 2020-06-30 journal: nan DOI: 10.1101/2020.06.29.20141283 sha: cdc7c2f17f0c4fb049e0873aac4f84f6cd9794a3 doc_id: 726381 cord_uid: w2dk0ec7 Helmet non-invasive ventilation (NIV) is a form of continuous positive applied pressure that has emerged as a useful therapy for COVID-19 patients who require respiratory support but may not require invasive ventilation. Helmet NIV has seen an increase in use during the COVID-19 pandemic because it is low-cost, readily available, and provides viral filters between the patient and clinician. Helmet NIV may also provide better patient outcomes by delaying or eliminating the need for invasive ventilation. Its widespread adoption has been limited, however, by the lack of a respiratory monitoring system that is needed to address known safety vulnerabilities and to provide clinicians with a respiratory profile of the patient. To address this safety need, we have developed an inexpensive respiratory monitoring system that is based on readily available commercial components and is suitable for rapid local manufacture. The system is designed for use in conjunction with the COVID-19 Helmet developed by Sea-Long Medical Systems, but is modular and can be used with other ventilation systems. The monitoring system comprises one or more flow and pressure sensors and a central remote station that can be used to remotely monitor up to 20 patients simultaneously. The system reports flow, pressure, and clinically relevant metrics including respiratory rate, tidal volume equivalent, peak inspiratory pressure (PIP), positive end-expiratory pressure (PEEP) and the ratio of inspiratory time to expiratory time (I:E). The device will sound alarms based on clinician-set thresholds. In bench tests using a commercial ventilator and artificial lung system, our device performs comparably to a commercial single-patient respiratory monitor. Results are presented from human-subject tests on a healthy volunteer undergoing helmet non-invasive ventilation. Detailed design and manufacturing documents are provided. invasive ventilation [5] , making non-invasive ventilation (where applicable) uniquely 42 valuable during the pandemic. 43 Helmet non-invasive ventilation (NIV) is a form of CPAP that uses a clear, polyvinyl 44 chloride (PVC), bubble-like helmet, attached to a soft collar that seals around a 45 patient's neck [6] . The helmet delivers an air-oxygen mixture to the patient at slightly 46 higher than atmospheric pressure, to keep the patient's airways open and the patient 47 oxygenated. Helmet NIV is appealing for pandemic use because it can be run directly 48 from a constant flow of air-oxygen, which is readily available in hospitals [7] . Helmets 49 cost less than 300 USD, are straightforward to manufacture, and have been available 50 during the pandemic, despite supply chain disruptions for other ventilators [1] . 51 Importantly, Helmet NIV is an enclosed system that uses viral filters to protect 52 clinicians and other patients from droplets or aerosolized SARS-CoV-2 viral particles 53 shed by the patient [8] . Helmet NIV is permitted by the FDA for COVID-19 54 treatment [9] . 55 Despite these advantages, lingering safety concerns have prevented wider adoption of 56 the helmet in the United States. Steady airflow is required to clear the patient's expired 57 CO 2 from accumulating in the helmet [10] . An unexpected drop of airflow caused by a 58 disruption in gas supply or blockage in the circuit could lead to rebreathing of CO 2 and 59 asphyxiation [11] . Therefore, one of the greatest concerns when using helmets is the lack 60 of an effective monitoring and alarm system. Antisuffocation valves can help mitigate 61 this risk, but ultimately a dedicated monitoring and alarm system is needed to employ 62 helmet NIV safely [12] . Partly for this reason, helmets in the US have been primarily 63 restricted to the ICU, where clinician-to-patient ratios are high enough to allow 64 clinicians to observe patients around-the-clock. 65 We have addressed these safety risks and clinical needs by developing an inexpensive 66 respiratory profile monitoring system, which we call the Princeton Open Ventilator 67 Monitor (POVM), that is based on readily available commercial components and is 68 suitable for rapid fabrication during a pandemic. The system is designed for use in 69 conjunction with non-invasive helmet ventilator systems, such as the COVID-19 Helmet 70 developed by Sea-Long Medical Systems, LLC [1] , but is modular and can be used in 71 other systems. The system comprises one or more flow and pressure sensors per helmet, 72 and crucially includes a central station that can be used to remotely monitor up to 20 73 patients simultaneously. The system reports to clinicians the flow and pressure in the 74 helmet, the patient's equivalent tidal volume (≈TV), respiratory rate (RR), ratio of 75 inspiratory to expiratory time (I:E ratio), peak inspiratory pressure (PIP), and peak end 76 expiratory pressure (PEEP). The system allows helmets to be deployed more broadly on 77 the wards, since a single clinician can now monitor many patients simultaneously. The 78 respiratory profile provided by the system allows clinicians to track disease progression 79 and informs treatment decisions, including decisions about when to intubate. 80 We have produced and tested 50 such devices in a matter of weeks at a marginal June 29, 2020 4/29 . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted June 30, 2020. The flow meter consists of a differential pressure sensor that measures the pressure drop 97 across a flow channel. This pressure drop is then used to calculate flow through the 98 channel. The differential pressure sensor was chosen to be capable of measuring 99 relatively small pressure drops (±500 Pa) with good dynamic range. Stable operation in 100 a humid environment was also a key consideration, as were cost and availability. The 101 device selected was the Sensirion SDP31 differential pressure sensor. This part works by 102 diverting a small portion of the flow and measuring the change in temperature as the 103 gas passes over a small internal heating element. The temperature change in the gas as 104 it flows through the sensor is related to the mass flow rate of the diverted flow stream, 105 which in turn is related to the pressure difference across the sense ports. Readings from 106 the SDP31 are routed to the controller over an I2C bus at a rate of 120 readings per 107 second. The SDP31 flow sensor, which comes in a surface mount package, is mounted to a 109 small printed circuit board (PCB, see Supplementary Design File 1) that is fixed to the 110 flow block as shown in Fig. 3 . To prevent condensation from forming in the narrow 111 flow-sensor channel, a heater has been added. Heat is generated in a TIP32 transistor 112 housed in a TO-220 package. The TIP32 is mounted on the opposite side of the printed 113 circuit board from the SDP31. Operating the transistor at a power of less than 1 W 114 suffices to raise the temperature in the SDP31 sensor to 45 • C, which is above the dew 115 point of the air in the system. The operating temperature is regulated using a 116 pulse-width modulation signal from the readout controller. . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) The copyright holder for this preprint this version posted June 30, 2020. . Flow-sensor assembly. A flow and pressure sensor are exposed to the flow element. The electronic sensor elements are mounted on circuit boards that connect to an external RaspberryPi controller via a standard RJ-45 jack. June 29, 2020 6/29 . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) The copyright holder for this preprint this version posted June 30, 2020. The flow element, shown in Fig. 4 , is a 19-mm-diameter tube, partially obstructed by a 130 10 mm thick wall perforated by nineteen 3.175-mm-diameter channels arranged in a 131 hexagonal pattern. One of the channels is partially obstructed and is used to guide a 132 small fraction of the air flow through the flow sensor. The obstruction is only partial, as 133 a portion of the air flow can directly cross the central wall through a small aperture. 134 We tuned the diameter of this aperture in order to match the dynamic range of the 135 sensor to the typical flow rates to be measured. We found that a 1.59-mm-diameter 136 aperture is well adapted to a 500 Pa dynamic range differential pressure sensor. example, see Fig. 10 ). Drill wander was minimized by using stiff bits and starting the 147 hole by "pecking." Deburring is also important, as there is a tendency for residual 148 Delrin "hair" to form at the entry and exit of the hole, see 5. This was reduced by 149 touching up the holes with a slightly larger bit. With proper jigging and after some 150 practice, one block could be machined in under 20 minutes. The yield is about 80%. Cleaning the blocks to "medical grade" required some care. The primary goal was to 152 remove all residual Delrin hair. This was most successfully done by a commercial 153 company [13], using a frozen CO 2 deburring technique, see Fig. 5 . The hair could also 154 be melted away by application of a heat gun. In addition, we had success by mounting a 155 3.175-mm-diameter "low scratch" tube brush dipped in 80 grit water-based slurry in a 156 hand drill and then polishing each hole. These last two methods require some care with 157 the spinning brush being preferable. After the mechanical processes, blocks were 158 cleaned with soap and water and then sonicated in isopropanol for 10 minutes. encoder, and a red-colored "Silence" pushbutton. On the sensor readout side, the digital bus interfaces were restricted to comply with 172 3.3V I2C and SPI and the control elements to 3.3V PWM and GPIO. The data 173 transmission distance of the I2C bus imposes a proximity requirement on the interface 174 box to the flow element of approximately 2 m. The time-base of the readout is derived 175 from the RPi system clock using timer interrupts. The human interface is based on 176 asynchronous signal interrupts. The SDP31 differential pressure sensor is placed into continuous, average until read, 178 mode. The RPi reads the SDP31 over I2C at a rate of 50 Hz, where each reading is the 179 June 29, 2020 8/29 . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) The copyright holder for this preprint this version posted June 30, 2020. . June 29, 2020 9/29 . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) The copyright holder for this preprint this version posted June 30, 2020. . arithmetic average of 40 individual readings taken at 0.5 ms intervals. The MP3V5004G 180 pressure sensor has an analog output that is digitized by an MCP3202 12-bit ADC from 181 Microchip. The MCP3202 is read out at a rate of 200 Hz over SPI. A 4-sample 182 arithmetic average is formed and recorded at a rate of 50 Hz on the same time-base as 183 the SDP31. The temperature of the SDP31 sensor is read out at 1 Hz. The preferred operating temperature of the flow block sensors is 45 • C to avoid 185 condensation from hot, humid vapor. A heating circuit located on the flow block is 186 adjusted using a convergent "take-back-half" servo algorithm operating on a PWM The addition of a single, customized two-layer PCB was found to be sufficient to The initial system software for the RPi is installed on a 32GB microSD memory card 205 with a separate data partition. The disk image is then replicated using multiple USB3 206 hubs running in parallel to provide identical systems for all interface boxes. The custom 207 enclosure allows the microSD card to be removed and replaced, but to minimize 208 handling of the microSD card, the preferred method of software update is through a 209 local repository on the remote monitoring station. The alarm is based on a 10 s running average of the flow rate and pressure, as well 211 as the average respiratory rate, calculated using a decaying weight on the last several 212 breaths. Maximum and minimum values are entered directly via the pushbutton rotary 213 knob. Rotation of the knob cycles through alarm setting menus, and a "push and turn" 214 operation is used to change the alarm threshold settings. The sounding of the alarm generates a PWM volume-controlled buzzer and 216 introduces a red back light to the LCD display. A red silence pushbutton on the 217 interface box may be used to temporarily silence the audible alarm, turning the back 218 light to orange. Silence can be invoked preemptively or post-alarm. The silence 219 duration defaults to 120 s, but can also be manually adjusted upon invoking the silence 220 button. The alarm status and all monitoring data is continuously streamed to the 221 remote monitoring station through a wired ethernet interface. The unique digital 222 identification of the flow element sensors, RPi ethernet MAC address, and a unique 223 interface box naming system allows the remote monitoring station to unambiguously 224 assign alarms to patients, as described in Section 2.1.14. June 29, 2020 10/29 . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted June 30, 2020. The remote monitoring station is the primary means of providing clinicians with 227 detailed patient information. The remote monitoring station comprises a networked 228 computer and a 27 inch monitor. Both an Intel NUC miniPC and a Raspberry Pi were 229 tested, although any modern PC running Python on any operating system will suffice. 230 The remote monitoring station provides a simultaneous overview of all patient 231 interface box data and alarm states. For each patient it shows flow, pressure and 232 volume waveforms and a set of derived quantities, including respiratory rate and 233 equivalent tidal volume (see Fig. 1 ). In a "drilldown" view it provides full-screen plots 234 of the waveforms for a single patient as well as a pressure-volume plot and derived 235 quantities, while still displaying alarm status information on all patients, see Fig. 15 . The remote monitoring station receives data from interface boxes over a wired or 237 wireless network. In a hospital setting we expect an air-gapped wired network to be 238 used. Our system was tested on an ethernet network with a 24 port hub. The 239 monitoring station acts as an always-on appliance and handles the automatic discovery, 240 pairing and logging of patient interface boxes, described in Section 2.1.14. Because the 241 interface boxes lack battery-backed real-time clocks, the remote monitoring station 242 maintains a master time record for recorded data streams. . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted June 30, 2020. software integrates together the data collection and handling in the interface boxes and 250 remote monitoring stations, provides an analysis suite for monitoring and alarm 251 determination, drives the rotary-dial interface of the interface box and implements the 252 graphical user interface of the remote monitoring station. It has been designed to work 253 with minimal computational resources, to be easily portable, to have few external 254 dependencies, and to use standard, well supported, tools when external dependencies 255 are used. The software is written entirely in Python and uses only open-source components. On top of a base Python 3.7 installation, additional requirements are currently NumPy, 258 SciPy, PyQt5, pyqtgraph, pyzmq, confuse/pyyaml, zeroconf and the Python Raspberry 259 Pi tools. Qt was selected for both its ability to display real-time graphics with minimal 260 computation resources, and for its portability across computing platforms including for 261 embedded devices. The sections below discuss the software design and implementation 262 of the interface boxes, the remote monitoring station and the communication between All communication from the device loop to the patient loop is one-way; and all 296 communication from the patient loop to the remote monitoring station is also one-way. 297 This provides a simpler architecture that was easy to make robust, and ensures that one 298 box can be monitored by multiple stations if needed. One side effect of this is that the 299 realtime clocks on the patient stations are not synchronized, so synchronization is made 300 possible by the timestamps on the remote monitoring station. Configuration is handled through several YAML files. There is a 302 processor/config defaults.yaml file that stores the defaults for most settings. There is a top-level file povm.yml that stores copies of settings that might be commonly 304 configured in a deployment. And then, when running, the patient software stores a 305 povm-live.yml file in the data directory that keeps track of runtime changes to the 306 alarm limits, so that the alarm settings get restored when the system is restarted. This 307 file is not intended to be edited directly, and is only searched for the rotary settings. If 308 this file were to be corrupted, this is not a failure -the settings are just not restored. A 309 dedicated thread stores the rotary settings every 10 seconds if they have changed. . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted June 30, 2020. . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted June 30, 2020. . https://doi.org/10.1101/2020.06.29.20141283 doi: medRxiv preprint measurements are derived from these breath-based measurements as exponentially The analysis functionality is implemented using the rolling buffer functionality The back-end software is shared between the interface box and the remote monitoring 369 station. A Generator is the core class, which handles the analysis thread and most of 370 the data structure. The subclass that does the on-device collection and rebroadcast 371 thread as described in Section 2.1.10 is the Collector. There is also a RemoteGenerator 372 subclass of Generator, and this instead only reads from an interface box broadcast in a 373 thread. This is used by the main Graphical User Interface (GUI) to monitor each 374 connected interface box. The remote monitoring station GUI is implemented in PyQt [14] . The GUI main 376 window displays a grid of patient stations, each of which shows waveforms for flow, 377 pressure, and volume and derived quantities. The grid can be rearranged via drag and 378 drop. Clinicians can adjust the waveform axes. Clicking the status button shows 379 advanced options, such as removing a disconnected sensor. Alarm status is displayed 380 using both colors and symbols, as described in Table 3 . Clicking any of the grid tiles on the main screen will bring up a "drilldown" screen of 382 information for that interface box (see Fig. 15 ). The drilldown screen shows a larger set 383 of waveforms and additional derived quantities. It also includes a Volume vs. Pressure 384 plot and additional information about alarm limits and averaging times, the duration . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted June 30, 2020. Table 3 . Alarm status GUI scheme for remote monitoring station. When patient interface boxes are connected to the network, they are automatically 394 discovered by the remote monitoring station using the python-zeroconf framework. Clinicians then need only to associate the interface box with a given patient by entering 396 the patients name, room number or other identifier. To reduce the chance of error when associating interface boxes with patients, each The names of the first 20 devices we produced are given in Table 4 . . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted June 30, 2020. ranging from 10 L/min to 100 L/min were imposed while the differential pressure ∆P 427 at the flow sensor was measured using our device (see Fig. 8 ). The details of the sweep 428 were chosen to probe both the steady-state relation between Q and ∆P and temporal 429 dynamics during changes of applied flow. The 0 to 100 L/min flow range accessible to 430 our MFC is clinically relevant to the helmet NIV application and corresponds to 431 differential pressure ∆P readings from 0 to 150 Pa (Fig. 8b) . The relation between ∆P 432 (Fig. 8b , red) and Q (Fig. 8a, blue) , averaged over many flow elements, gives the 433 calibration in Fig. 9b . Flow elements with obvious manufacturing defects had Q-∆P 434 relations that deviated (see Fig. 10 ), and were removed from the global calibration. Our device calculates the reported flow Q from a differential pressure measure ∆P 443 across a portion of the flow element. A single global lookup table is used for all devices 444 to convert from ∆P to Q (see Fig. 9 ). The relationship between Q and ∆P matches . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted June 30, 2020. given the length L and diameter D of a pipe, and a Darcy friction factor f D that have a growing boundary layer. In such laminar flow cases we expect ∆P ∝ Q 3/2 . We 460 note that our calibration lookup table follows the ∆P ∝ Q 7/4 for high flows and the 461 ∆P ∝ Q 3/2 for lower flows in agreement with expectations ( Fig. 9 ). To compare our device against commercial systems, the April 22nd version of our device 468 was tested on a commercial test lung (IngMar ASL5000) driven by a commercial 469 ventilator (GE Avance CS2 Anaesthesia System) in series with a commercial respiratory 470 monitoring system (Philips Respironics NM3). Our device and the commercial subject was in a standing position. The subject's spontaneous breathing was recorded 501 on our respiratory monitor device over two 20 min trials as the subject was instructed 502 to breath normally. A water manometer in our air delivery system appeared to 503 contribute to dynamic changes to the volume in the helmet circuit coincident with the 504 subject's breathing, so the manometer was inactivated during portions of the recording. 505 We placed a prototype of our device in a circuit with a commercial single-patient 509 respiratory monitoring system (Philips NM3) on a test lung (ASL-5000) driven by a 510 commercial ventilator. We compared recordings made by our device to those made by 511 the test lung and the commercial monitoring system. Recordings from our device show 512 strong agreement with the test lung for flow and volume. See Figs. 12, 13 and 14. Our device's flow measurement showed closer agreement to the test lung than the 514 commercial monitoring system that was measuring flow simultaneously. There was a 515 scaling factor of about 5% discrepancy between our device and the test lung's measured 516 flow, while the commercial monitor reported flows that differed from our device by a 517 scaling factor of about 15% and an offset of 2 L/min. The commercial respiratory 518 monitor also would occasionally show a flat line for flow or pressure during a breath, 519 while our system captured dynamics of all breaths (see artifact at approximately 158 s 520 in Fig. 12 and as a horizontal spread around zero in Fig. 14) . In measures of pressure, our system had close agreement to the commercial 522 respiratory monitor. We note that our system seemed to avoid quantization artifacts 523 found in the commercial monitoring system at low pressures (see steps in pressure 524 around 80 s in NM3 recording in Fig. 12 ) . We also compared derived quantities-the equivalent tidal volume, PIP, PEEP, and 526 breathing rate-with those reported by the ventilator at the same time. Our system 527 largely agrees with the ventilator's own settings (see Table 5 ). June 29, 2020 21/29 . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted June 30, 2020. . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted June 30, 2020. . https://doi.org/10.1101/2020.06.29.20141283 doi: medRxiv preprint 3.2 Human volunteer study 529 A healthy consenting adult volunteer was placed in a helmet and received medical air at 530 80 L/min while undergoing monitoring with our device. The device captured reasonable 531 waveforms for flow, pressure and equivalent tidal volume, as shown in Fig. 15 Fig 15. Respiratory profile of healthy human subject in helmet. Top: Pressure, flow and integrated flow for gas on the expiratory path of a helmet worn by a healthy human subject is shown. Positive flow is defined to be gas leaving the helmet through the expiratory path. Bottom: Derived quantities, including tidal volume, respiratory rate, etc are visible on the remote monitoring station (shown here in "drilldown" screen). June 29, 2020 23/29 . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted June 30, 2020. . https://doi.org/10.1101/2020.06.29.20141283 doi: medRxiv preprint 3.3 External sources of variability 536 We tested the influence of relative humidity on our sensors, Fig. 8(d) . We challenged 537 our device by measuring 20 hours of continuous flow of saturated humid air 538 (RH = 100%). Both sensors performed even in the presence of significant condensation 539 in the flow element, which suggests that our heater circuit functions effectively. In tests 540 with variable humidity, we observed that the flow Q increased slightly by less than 3% 541 while the humidity rose from 5 to 100%. Finally, the presence of curvature in tubing has been reported to affect measures of 543 downstream flow [16, 17] . This is relevant because our device may dangle off the helmet 544 at an angle, as can be seen in Fig. 1 . We introduced a 90 • bend into the tubing a 545 variable distance L upstream of our device and observed the change in our device's 546 reported flow Q under a constant imposed flow of 100 L/min, Fig. 8(e) and Fig. 16 . We 547 observed a systematic increase in the flow Q reported by our device for bends less than 548 30 cm upstream of our device. The largest deviations from the imposed flow occurred 549 when the bend was closest to the device (5% for L = 5 cm). 551 We completed a production run of 50 devices. Producing devices at scale posed 552 additional challenges regarding calibration and quality control. In particular 553 manufacturing defects in the flow element altered the relationship between the 554 measured pressure difference ∆P and the flow Q across the device, as shown in Fig. 10 . 555 Initial determination of ∆P vs. Q for each flow element with a fixed sensor array 556 allowed us to eliminate outliers from the determination of the average calibration. After 557 each flow element is integrated into its final flow-sensor assembly, it will undergo the 558 acceptance testing described in Sec. 2.2.6. An example test result is shown in Fig. 17 . We designed and built an inexpensive respiratory monitoring device from conception to 561 production in a six week period coincident with the initial peak of the COVID-19 562 pandemic in our region. Our device shows similar, and in some aspects better, 563 agreement with a test lung than a commercial monitoring system that is currently 564 unavailable. Measurements of a healthy volunteer in a helmet taken by our device are 565 reasonable and match prior reports. Our device has the potential to improve patient 566 June 29, 2020 24/29 . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted June 30, 2020. . outcomes and to mitigate well-documented safety concerns regarding CO 2 re-breathing 567 that have prevented the helmet's widespread adoption. Our device also introduces the 568 ability to monitor many patients simultaneously. Importantly, our device uses 569 off-the-shelf components that have shown themselves to be robust to supply chain 570 disruptions. The device is well suited for local manufacture during a pandemic. The 571 device is designed for helmet systems but we have also successfully tested it in 572 traditional invasive ventilator settings. We currently are seeking FDA emergency use 573 authorization. For future versions of the device we are exploring adding CO 2 -and O 2 -sensors, and 575 possibly adding hardware, such as an EEPROM, with each sensor board to enable 576 per-sensor calibration curves. . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) The copyright holder for this preprint this version posted June 30, 2020. . https://doi.org/10.1101/2020.06.29.20141283 doi: medRxiv preprint of Penn Health Tech. We thank Sebastian Seung and his volunteers for feedback. We 592 thank Theodor Brasoveanu and his students at the Princeton Day School for supplies. 593 We thank Meghan Testerman of Princeton University Libraries for assistance. Table S1 . . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) The copyright holder for this preprint this version posted June 30, 2020. . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) The copyright holder for this preprint this version posted June 30, 2020. . https://doi.org/10.1101/2020.06.29.20141283 doi: medRxiv preprint Medical Systems website Critical Care Utilization for the COVID-19 Outbreak in Lombardy, Italy: Early Experience and Forecast During an Emergency Response Intensive Care Management of Coronavirus Disease 2019 (COVID-19): Challenges and Recommendations. The Lancet Respiratory Medicine Effect of Noninvasive Ventilation Delivered by Helmet vs Face Mask on the Rate of Endotracheal Intubation in Patients With Acute Respiratory Distress Syndrome: A Randomized Clinical Trial Presenting Characteristics, Comorbidities, and Outcomes Among 5700 Patients Hospitalized With COVID-19 in the New York City Area New Treatment of Acute Hypoxemic Respiratory Failure: Noninvasive Pressure Support Ventilation Delivered by Helmet-a Pilot Controlled Trial Video: Non-Invasive CPAP by Helmet Setup COVID-19, Maurizio Franco Cereda, MD Minimise Nosocomial Spread of 2019-nCoV When Treating Acute Respiratory Failure. The Lancet Enforcement Policy for Ventilators and Accessories and Other Respiratory Devices During the Coronavirus Disease 2019 (COVID-19) Public Health Emergency Continuous Positive Airway Pressure Delivered with a "Helmet": Effects on Carbon Dioxide Rebreathing Danger of Helmet Continuous Positive Airway Pressure during Failure of Fresh Gas Source Supply Performance of Different Continuous Positive Airway Pressure Helmets Equipped with Safety Valves during Failure of Fresh Gas Supply Friction factors for pipe flow The London, Edinburgh, and Dublin Philosophical Magazine and Turbulent pipe flow downstream of a 90 • bend MP3 V3-Loading Top.pdf PDF Human Layout for Assembly MP3 V3-Loading Bot.pdf PDF Human Layout for Assembly LCD Interface Loading-Top.pdf PDF Human Layout for Assembly LCD Interface Loading-Bot.pdf PDF Human Layout for Assembly