key: cord-0725622-vxinnxw0 authors: Eberle, Ute; Wimmer, Clara; Huber, Ingrid; Neubauer-Juric, Antonie; Valenza, Giuseppe; Ackermann, Nikolaus; Sing, Andreas title: Comparison of nine different commercially available molecular assays for detection of SARS-CoV-2 RNA date: 2021-01-29 journal: Eur J Clin Microbiol Infect Dis DOI: 10.1007/s10096-021-04159-9 sha: 8e1576a98ad03e2c884241c97ae8f5614c8b7d55 doc_id: 725622 cord_uid: vxinnxw0 To face the COVID-19 pandemic, the need for fast and reliable diagnostic assays for the detection of SARS-CoV-2 is immense. We describe our laboratory experiences evaluating nine commercially available real-time RT-PCR assays. We found that assays differed considerably in performance and validation before routine use is mandatory. The localized outbreak in the province Hubei and surrounding areas in China at the end of December 2019 led to a rapid spread throughout the world increasing numbers in cases and deaths. On March 11, 2020, WHO declared COVID-19 a pandemic. Major strategies to face this pandemic are (i) to minimize social contacts, (ii) to test people suspected for COVID-19, (iii) to perform strict containment measures for SARS-CoV-2 (severe acute respiratory syndrome coronavirus 2)-infected persons, and (iv) to conduct thorough contact tracing. Public health authorities and other diagnostic laboratories are in urgent need for diagnostic tools, which are fast and reliable. The rapid pandemic spread resulted in various shortages not only in safety equipment, but also in diagnostic reagents and supply. Due to imminent problems with limited test resources, a foresighted and dynamic lab strategy is needed to avoid possible shortages in lab supplies including test kits. An important pillar of the strategy chosen by the Public Health Microbiology laboratory at the Bavarian Health and Food Safety Authority (LGL), Germany, was the pro-active decision to diversify diagnostic tools. Moreover, our laboratory was confronted with an increasing number of various test kits offered by very different companies but of unclear quality standards. In this paper, we describe our laboratory experiences comparing nine commercially available real-time RT-PCR assays, using a Bio-Rad CFX 96 cycler. Based on an initial evaluation of SARS-CoV-2-RT-PCR assays available at the very beginning of the pandemic in Germany [1] , we established the RealStar SARS-CoV-2 RT-PCR Kit RUO, Altona Diagnostics on the Bio-Rad CFX96 Touch Real-Time PCR Detection System (Bio-Rad, Feldkirchen, Germany) as routine method for SARS-CoV-2 diagnosis. Between the calendar weeks 5/2020 and 19/2020, approximately 70,000 mainly respiratory samples were analyzed at the LGL. Eight thousand two hundred six thereof Ute Eberle, Clara Wimmer, Nikolaus Ackermann and Andreas Sing contributed equally to this work. were diagnosed as SARS-CoV RNA positive (Fig. 1) . The assay amplifies sequences of the E gene of B lineage betacoronaviruses and of the S gene specific for SARS-CoV-2 that are detected with fluorophores/Fam and Cy5, respectively. Comparison of nine commercially available molecular diagnostic kits for SARS-CoV-2 To compare the performance of further eight PCR kits for SARS-CoV-2 diagnostic, field samples as well as SARS-CoV-2 reference virus genomes in defined genome copy number and genomes of putatively contaminating agents were used. Respiratory samples, mainly nasopharyngeal or pharyngeal swabs were obtained from patients and contact persons (see column "sample size" in Table 1 ). To increase the total sample volume for further testing with the RealStar SARS-CoV-2 RT-PCR Kit RUO, Altona Diagnostics pretested positive and negative samples were diluted 2 to 3 fold with NaCl 0.9%. Control material (positive controls) comprising Human 2019-nCoV RNA (reference number 026N-03889) and SARS-CoV Frankfurt 1 RNA (reference number 004N-02005) was ordered from the European Virus Archive (EVAg) [2] . Cross-reactivity was tested for human coronavirus HCoV 229E 1 , human coronavirus HCoV-OC43 1 , human coronavirus HCoV-NL63 1 , MERS-CoV 1 , human rhinovirus type 30 1 and type 49 2 , 2 samples human metapneumovirus Typ A 1,2 , Parainfluenzavirus type 2 2 and 3 1 , Influenzavirus (A(H1N1)pdm09, A(H3N2), B Yamagata-and Victoria-Lineage) 1 Table 1 )) As expected, analysis of SARS-CoV-1 yielded positive results for lineage B-betacoronavirus (B-βCoV) detection systems and negative results for with detection systems for SARS-CoV-2-specific genetic regions. Sample sizes are given in Table 1 column "sample size." The variation in sample sizes between different assays is due to the different amounts of available PCR reaction. RNA was extracted from samples using the QIAamp Bio Robot kit (QIAGEN) on a Hamilton Microlab Star (Hamilton, Bonaduz, Switzerland). Extractions were performed in triplicates to obtain sufficient elution volume to execute all PCR assays with identical sample preparations. Eluates were pooled and stored in 30-50-μl aliquots at − 80°C before Detection limit, linearity, and PCR efficiency were determined using Human 2019-nCoV RNA (reference number 026N-03889) from the European Virus Archive (EVAg) with replicates, as given in column "number of replicates." 10-fold serial dilutions were performed carrying out PCR assays. All PCR assays were performed according to the instructions of the respective manufacturers. During the evaluation phase, the threshold was manually set to 150 RFU (Bio-Rad CFX96 Touch Real-Time PCR Detection System) irrespective of the assay validated and the fluorophore. Amplification plots with exponential curve increases were considered positive. A summary of the results and features of the nine commercially available PCR assays tested is shown in detail in Table 1 . The limit of detection was determined using Human 2019-nCoV RNA (reference number 026N-03889) from the European Virus Archive (EVAg) containing defined genome copy numbers. The PCR assays of Altona Diagnostics (AT), Mikrogen Diagnostik (MG), Shanghai Fosun Long March Medical Science Co (SF), and BGI reached a detection limit of about 10 copies/reaction (see column "limit of detection" in Table 1 ). Notably, due to the limited amount of PCR reagents available for testing, the number of achievable replicates was limited to two-four (Table 1) , preventing detailed statistical Table 1 ). The specificity of each assay was tested in order to discriminate SARS-CoV-2 signals from other respiratory pathogens such as other coronaviruses. All assays gave no false-positive test results (see columns "specificity" and "number of FP" in Table 1 ). Due to the limited amount of reagents available for testing sensitivity and specificity performances of BGI, Shanghai Fosun Long March Medical Science Co., Ltd. (SF), and Primerdesign Ltd (PD) test kits, results should be interpreted with caution (see column "sample size" in Table 1) . Additionally, a multi-correlation analysis was conducted focusing on the specific results for SARS-CoV-2. For assays showing more than one specific result, the smallest Ct value was included in the analysis. Most tests showed Pearson standard correlation coefficients above 95%. The distribution of Ct values is shown in the histograms in Fig. 2 Fig. 2 and Table 1 ). A main task for public health laboratories is to deliver reliable results for emerging pathogens of global threat to prevent spread. Since on January 7, 2020, a novel coronavirus was identified, and shortly after, the first sequence of the new strain was published. Shortly afterwards [3, 4] , our laboratory faced the difficult task to decide on appropriate diagnostic assays. Due to an enormous demand for diagnostic reagents, we were confronted with a shortage of PCR kits to detect SARS-Cov-2. Furthermore, we had to deal with offers for PCR kits from different origins and of unclear quality standards. To decide which of the kits available on the market at the time would be best for our diagnostics, we tested a number of kits that had given promising results in other studies [5] and targeted two or more SARS-CoV-2 genes in one PCR setup. This was the case in all but one commercial kit (WELLS BIO, INC., WB). For two of the kits (BGI and Primerdesign (PD)), it was not possible to retrieve information on how many genes were targeted and which genetic regions were amplified. In our view, the detection of more than one gene seemed crucial as different target genes may differ in specificity and sensitivity. In our hands, the kits of Fast Track Diagnostics (FTD), Mikrogen Diagnostik (MG), Altona Diagnostics (AT), and Shanghai Fosun Long March Medical Science Co., Ltd. (SF) gave the best results. Since for all assays, identical extracted material was used, we assume that the observed differences between kits are due to differences in targeted genome regions and/or primer design. Further complications were due to the fact that the kits of BGI, WELLS BIO, INC., WB, and SolGent Co., Ltd. contained the internal control as part of the Mastermix and could not be used as an extraction control resulting in lack of a control system for the extraction procedure. In outbreak situations, it is crucial to efficiently optimize all workflows even with limited resources of human workforce, reagents, and devices. Comparable results were published for Altona Diagnostics, Mikrogen, and BGI in other studies [5] . The quality of the tested PCR assays showed very different results. Therefore, it is very important to control commercially available kits for their performance characteristics prior to use in routine laboratory diagnosis. Even in situations with enormous pressure due to reagent shortage, quality must not be neglected. Fast and reliable results are crucial dealing with the COVID-19 pandemic. The Fast Track Diagnostics (FTD), Mikrogen Diagnostik (MG), Altona Diagnostics (AT), and Shanghai Fosun Long March Medical Science Co., Ltd. (SF) outperformed all other tested mentioned above. In addition to correct results, it is recommended to use assays with at least two different target regions either in one channel or two to have a more robust assay for a fast evolving pathogen. Rapid establishment of laboratory diagnostics for the novel coronavirus SARS-CoV-2 in Bavaria The European Virus Archive goes global: a growing resource for research Identification of a novel coronavirus causing severe pneumonia in human: a descriptive study A new coronavirus associated with human respiratory disease in China Comparison of seven commercial RT-PCR diagnostic kits for COVID-19 Publisher's note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations Authors' contributions UE: planned experiments, data analysis, conceptualized the laboratory work, and wrote the manuscript draft. CW: conducted experiments and data analysis. NA: conceptualized the laboratory work and participation in manuscript preparation. AS: conceptualized the laboratory work and participation in manuscript preparation. IH: data analysis. ANJ: data analysis and participation in manuscript preparation. GV: data analysis.Funding Open Access funding enabled and organized by Projekt DEAL. Conflict of interest The authors declare that they have no conflict of interest.Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.