key: cord-0724464-vfv7in7q authors: Lamb, Christopher C.; Haddad, Fadi; Owens, Christopher; Lopez-Yunez, Alfredo; Carroll, Marion; Moncrieffe, Jordan title: Updated Clinical Evaluation of the CLUNGENE(®) Rapid COVID-19 Antibody Test date: 2021-08-30 journal: Healthcare (Basel) DOI: 10.3390/healthcare9091124 sha: 0a252fd68ec76dac0c4125326370a0c06ad2cc5b doc_id: 724464 cord_uid: vfv7in7q Background: COVID-19 antibody testing has been shown to be predictive of prior COVID-19 infection and an effective testing tool. The CLUNGENE(®) SARS-COV-2 VIRUS (COVID-19) IgG/IgM Rapid Test Cassette was evaluated for its utility to aide healthcare professionals. Method: Two studies were performed by using the CLUNGENE(®) Rapid Test. (1) An expanded Point-of-Care (POC) study at two clinical sites was conducted to evaluate 99 clinical subjects: 62 positive subjects and 37 negative subjects were compared to RT-PCR, PPA, and NPA (95% CI). Sensitivity was calculated from blood-collection time following symptom onset. (2) A cross-reactivity study was performed to determine the potential for false-positive results from other common infections. Results: The specificity of subjects with confirmed negative COVID-19 by RT-PCR was 100% (95% CI, 88.4–100.0%). The sensitivity of subjects with confirmed positive COVID-19 by RT-PCR was 96.77% (95% CI, 88.98–99.11%). In the cross-reactivity study, there were no false-positive results due to past infections or vaccinations unrelated to the SARS-CoV-2 virus. Conclusion: There is a need for a rapid, user-friendly, and inexpensive on-site monitoring system for diagnosis. The CLUNGENE(®) Rapid Test is a useful diagnostic test that provides results within 15 min, without high-complexity laboratory instrumentation. The COVID-19 pandemic SARS-CoV2 COVID-19 has infected over 140 million people worldwide and has caused approximately 3.89 million deaths as of 23 June 2021 [1] ; however, some studies suggest that the actual number of global COVID-19 deaths may be about 6.9 million, which is more than double the recorded amount [2] [3] [4] . In response to the pandemic, the US Food and Drug Administration (FDA) authorized the use of COVID-19 serological tests through Emergency Use Authorizations (EUAs) to make COVID-19 in vitro diagnostic tests widely available to help identify individuals with an adaptive immune response indicating recent or prior infection [5] . Serology tests, or immunoassays, play a significant role in the fight against COVID-19 [6, 7] . A prior history of SARS-CoV-2 infection is associated with a lower risk of infection, with an estimated protective effect of up to seven (7) months following primary infection; this supports the conclusion that convalescent plasma with specific antibodies to SARS-CoV-2 has powerful antiviral activity, which can reduce the viral load and mortality in patients with active COVID-19 infection [8] [9] [10] . There is an urgent need for a rapid, user-friendly, and inexpensive on-site monitoring system for diagnosis [11] . The CLUNGENE ® SARS-COV-2 VIRUS (COVID-19) IgG/IgM Rapid (15 min) Test Cassette has been commercially available in the US under an FDAapproved Emergency Use Authorization (EUA201121) [12] and Europe (CE Mark reference 02PBJ267 dated 9 March 2020). The CLUNGENE ® test has been previously studied, including the use of the test in the offices of general practitioners, evaluating the presence of antibodies in convalescent plasma donors, and how the test performs at a point of care facility [13] [14] [15] . The aim of this research was to better understand the sensitivity and specificity of the CLUNGENE ® assay and the potential for false-positive results related to infections or vaccinations not linked to the SARS-CoV-2 virus. In an initial study, a single Point-of-Care (POC) facility was used to estimate the sensitivity and specificity of the CLUNGENE ® test [16] . The study was expanded to a second independent site. The two sites were Sharp Healthcare, a not-for-profit multi-center regional healthcare group located in San Diego, CA; and Alivio Medical Center, an urgent care/primary care center in Indianapolis, IN. Samples used for RT-PCR were nares swabs. Positive subjects were symptomatic for SAR-CoV-2 Virus infected and confirmed with RT-PCR positive tested nares swabs. Negative subjects were asymptomatic, from high-risk areas, and confirmed with negative RT-PCR tested nares swabs. Finger-prick whole-blood samples were used for SARS-COV-2 Virus IgG/IgM detection. Subjects were excluded if they were unable to provide informed consent due to mental or cognitive disabilities. The CLUNGENE ® Point-of-Care test was run according to the manufacturer's instructions (see Figure 1 ). The test result was read after 15 min. Days from symptom onset were captured from an electronic medical record which documented self-reported data from patients reporting on the number of days they had been sick at the time of study enrollment. Categorical variables were compared using the chi-squared or Fisher exact test, and continuous variables were compared using the Student t-test or Mann-Whitney U test, as appropriate. All tests were two tailed, and p < 0.05 was considered statistically significant. SPSS Statistics, IBM SPSS software, version 27.0 (SPSS, Inc., Chicago, IL, USA) was used for all calculations. The CLUNGENE ® test can be stored at ambient temperature between 4 and 30 degrees Celsius. The product is packaged in a box with 25 test kits, with each kit wrapped in an individual foil patch for proper storage and transportation. The test kit itself should be kept away from direct sunlight. An initial cross-reactivity study was performed in which various common infectious agents were tested for potential false-positive results ( Table 1 ). All tests were negative. In addition, as recommended by the US Food and Drug Administration [17] , a follow-up study tested the cross-reactivity of the CLUNGENE ® device to antibodies to common coronaviruses that are not SARS-CoV-2: anti-229E, anti-NL63, anti-OC43, and anti-HKU1; and those for which there is a high rate of vaccinations and/or infection in the US, i.e., anti-Haemophilus influenzae IgG and IgM. The testing was performed at Medcomp Sciences, an independent clinical medical laboratory [18] . The CLUNGENE ® Point-of-Care test was run according to same manufacturer's instructions as was performed with Study #1. Samples (50) used for the cross-reactivity study were obtained from Trina Bioreactives, Ag; Trina Bioreactives, Ag, (Grabenstrasse 8, 8606 Nänikon, Switzerland) is an ISO certified company providing specialized Sars-COV-2 in vitro diagnostic biomaterials [19] . The samples were collected under a protocol approved by the ethics committee of the National Medical Association, Baden-Wurttemburg, Germany (file #F-2012-027, "Plasma Samples for Studies"). Test results were read after 15 min. Five (5) serum samples that were positive for IgM and five for IgG were analyzed in three (3) separate CLUNGENE ® Point-of-Care test lots with each of the below antibodies: Cross-reactivity was determined by using Beckman Coulter UniCel DxI Access Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids ( Table 2 ). All QC materials were supplied and used as indicated by the manufacturer. A minimum of 2 levels of QC were processed per run in accordance with Westgard rules for 6-sigma quality requirements. QC results must fall within 3 standard deviations of historical data, as recorded in Levy-Jennings plots by the system. All QC results were documented and verified by Clinical Laboratory Scientist before processing samples (Table 3) . Qualitative results are reported as Positive/Reactive or Negative/Non-Reactive, depending on whether the analyte in question is above or below the established signal cutoff value (S/CO). Each analyte tested may have a different cutoff value, but each cutoff value is determined during calibration of the instrument, using the value of the calibrator. An analysis was run on 99 patients who completed the study (Table 4) . Thirty-seven (37) patients who had negative COVID-19 RT-PCR were tested and found antibody negative by using the CLUNGENE ® test (95% CI, 90.60~100.00%). Three (3) out of four (4), or 75%, of RT-PCR positive COVID-19 subjects tested prior to day 7 of symptom onset were antibody positive (95% CI, 30.06~95.44%). Twenty-three (23), or 100%, of RT-PCR positive COVID-19 subjects tested positive between day 8 and 14 from symptom onset (95% CI, 85.69~100.00%). A total of 33 out of 35, or 94.28%, of RT-PCR positive COVID-19 subjects tested positive after day 14 from symptom onset were antibody positive (95% CI, 30.06~95.44%). In all 62 patients with confirmed COVID-19 with RT-PCR, the combined sensitivity of IgM and IgG was 96.77% (95% CI, 88.98-99.11%), meaning that there was 96.77 positive agreement between a positive RT-PCR test and a positive antibody test. The specificity was 100% (95% CI, 88.4-100.0%), meaning there was 100% agreement between a negative RT-PCR test and 100% negative antibody result. These results are displayed in Table 5 . The positive predictive value can be calculated, but the result is dependent on the prevalence of disease in the community. If a test for a disease has 96.77% sensitivity and 100% specificity, and the disease prevalence is 10%, the positive predictive value (PPV) is 100%, and the negative predictive value (NPV) is 99.64%. At the time of this publication, the positivity rate for the nares SARS CoV2 RT-PCR tests was between 8.2% and 10% in San Diego, CA (San Diego County), and Indianapolis, IN (Marion Country) [20] . The test results of negative quality-control samples were all negative, and the test results of the positive quality-control samples were all positive. The consistency rate of cross-reactivity of negative samples was 100%. In the first study, the performance characteristics of CLUNGENE ® were evaluated and showed a specificity of 100% and a sensitivity of 96.77%. In the second study, there were no false-positive results due to past infections or vaccinations unrelated to the SARS-CoV-2 virus. These results are in line with the new European Commission's Medical Device Coordination Group requirements for rapid COVID-19 antibody tests and consistent with previously published results [21] [22] [23] [24] [25] . Antibody testing is a useful aid to confirm past infection [26] . Recent findings confirm that antibody testing is predictive of prior COVID-19 infection, and rapid screening methods-even from finger pricks-are effective testing tools [27] . However, we see the potential for a much broader use and recommend a combined approach that uses both RT-PCR and serological testing. The advantage of the CLUNGENE ® antibody test is its simplicity, since there is no need for specialized laboratory personnel to perform and interpret results. The low rate of false positivity makes this test ideal to rule in disease and eliminate the need for further RT-PCR testing if seroconversion occurs, since the CLUNGENE ® antibody test can diagnose most infected COVID-19 patients. If the test is negative, a recommendation should be made to have a follow-up RT-PCR test. Serology testing also has the potential to monitor the presence of antibodies. Studies confirm that a prior history of SARS-CoV-2 infection is associated with a lower risk of infection, with an estimated seven (7) month protective effect [9, 10] . The association of SARS-CoV-2 Seropositive Antibody Test with Risk of Future Infection has now been established [28] . It is clear that titers of IgM and IgG antibodies against the receptor-binding domain (RBD) of the spike protein of SARS-CoV-2 decrease significantly between one (1) and seven (7) months, and concurrently, neutralizing activity decreases [29] . Monitoring the presence of antibodies from past infection could assist healthcare professionals in assessing the likely presence of neutralizing activity and immunity when managing patient care. Given widespread availability, COVID-19 serology testing, similar to other infectious diseases, can now become routine [30] . Consideration should be given to adding COVID-19 antibody testing to the WHO List of Essential In Vitro Diagnostics (EDL), which now consists of 122 test categories, including most serious infectious diseases [31] . A rapid 15-min COVID-19 assay offers the ability to do this testing quickly and efficiently-tests are now available for less than $5 and routinely reimbursed in the US by public and private insurers [5] . In addition, the ability of the CLUNGENE ® antibody test to detect antibodies to the coronavirus's spike protein's receptor binding domain means it has the potential to assess the efficacy of most vaccines, as well as convalescent plasma therapy [32] . Countries in Europe are now using antibody testing to determine if a second COVID-19 vaccine dose is required if a patient has a prior infection based on a positive antibody test [33] . Recently, airports and Blood Banks have been providing COVID-19 antibody testing services to determine whether a person has developed immunity to COVID-19 through vaccination or through contracting the virus previously [34, 35] . Some countries, including China, require an antibody test. Limited evaluation of the CLUNGENE ® antibody test has confirmed positive antibody test results following patients who have been vaccinated [36] . Pfizer's recent data suggest that its vaccine is efficacious for only 6 months and that a third shot within 12 months is likely needed [37] . Furthermore, there is a potential issue regarding vaccine efficacy for recipients who do not receive a full dose; the US CDC estimates that 3% of vaccinated people who received a first dose did not receive a recommended second dose [38] . COVID-19 vaccination also fails to stimulate an immune response in many blood-cancer patients or those otherwise immunocompromised [39, 40] . Additional studies are needed to confirm the efficacy of serology testing to monitor vaccine effectiveness. Limitations of the study include a small sample size from two geographic areas. The study also did not include special groups, such as pregnant women or children. The subjectivity of symptom reporting by patients can be a confounding factor in determining the duration of illness. Some patients may have been symptomatic for a different time period than they recalled. Lastly, the CLUNGENE ® antibody test has not been compared with another test in the study. In a pandemic crisis with significant economic and health implications, this study confirms the utility of serological testing for COVID-19 disease diagnosis providing rapid test results with a relatively high degree of sensitivity and specificity. Furthermore, given recent data regarding the relationship between positive serology and immunity, routine testing can be a useful tool to monitor antibody status for optimal patient care. Tests such as the CLUNGENE ® SARS-COV-2 VIRUS (COVID-19) IgG/IgM Rapid Test Cassette can assist healthcare professionals to help identify individuals with an adaptive immune response indicating recent or prior infection, as intended by the US FDA under an EUA. Funding: This research was partially funded by Clongene and Speranza. Institutional Review Board Statement: The study was conducted according to the guidelines of the Declaration of Helsinki, and approved by the Institutional Review Board of Sharp Center for Research (protocol ID: IgG/IgM COVID19; IRB#: 2005801; 21 May 2020). This was a formal IRB approved clinical study conducted within Sharp Healthcare, a not-for-profit multicenter regional healthcare group located in San Diego, California. Subjects were included if hospitalized or recently discharged following a SARS-CoV-2 RT-PCR nares test. A study protocol and informed consent were initiated and approved by the Sharp Institutional Review Board. Subjects were included if they were >18 years of age and understood the study and its requirements. Patients who had impairment of cognition or decision-making capacity were excluded. Subjects were screened by research coordinators to determine if they had a nares SARS-CoV-2 RT-PCR test result, and then consent was requested to enroll them in the study. Informed consent was obtained from all subjects involved in the study. Written informed consent was obtained from the patient(s) to publish this paper. The data presented in this study are available on request from the corresponding author. The data are not publicly available. Data. Total Confirmed COVID-19 Deaths and Cases Has Caused 6.9 Million Deaths Globally, More Than Double What Official Reports Show Estimation of Total Mortality Due to COVID-19 Estimation of the fraction of COVID-19 infected people in US states and countries worldwide COVID-19 diagnostic testing: Lessons learned for innovative product development during a public health emergency Interim Guidelines for COVID-19 Antibody Testing Evaluating the accuracy of different respiratory specimens in the laboratory diagnosis and monitoring the viral shedding of 2019-nCoV infections Low serum neutralizing anti-SARS-CoV-2 S antibody levels in mildly affected COVID-19 convalescent patients revealed by two different detection methods Convalescent Plasma Use in the United States was inversely correlated with COVID-19 Mortality: Did Convalescent Plasma Hesitancy cost lives? medRxiv 2021 Risk of SARS-CoV-2 Transmission from Newly Infected individuals with Documented Previous Infection or Vaccination SERS based lateral flow immunoassay for point-of-care detection of SARS-CoV-2 in clinical samples FDA COVID19 antibody detection using lateral flow assay tests in a cohort of convalescent plasma donors Observational Study of SARS-CoV-2 Antibody Immune Response in a Cohort of Patients at a North Suburban Chicago, Illinois Physician's Practice Expanded Clinical Evaluation of the CLUNGENE®Rapid Covid-19 Antibody Test Point-Of-Care Clinical Evaluation of the Clungene ® SARS-CoV-2 Virus IgG/IgM 15-minute rapid test cassette with the Cobas ® Roche RT-PCR platform in patients with or without Covid-19 FDA MedComp Sciences. 2021. Available online COVID-19 in San Diego County MDCG 2021-2: Guidance on State of the Art of COVID-19 Rapid Antibody Tests Clinical and laboratory evaluation of SARS-CoV-2 lateral flow assays for use in a national COVID-19 seroprevalence survey Assessment of SARS-CoV-2 serological tests for the diagnosis of COVID-19 through the evaluation of three immunoassays: Two automated immunoassays (Euroimmun and Abbott) and one rapid lateral flow immunoassay Four point-of-care lateral flow immunoassays for diagnosis of COVID-19 and for assessing dynamics of antibody responses to SARS-CoV-2 Multiplex reverse transcription loop-mediated isothermal amplification combined with nanoparticle-based lateral flow biosensor for the diagnosis of COVID-19 Antibody tests for identification of current and past infection with SARS-CoV-2. Cochrane Database Syst COVID-19 Antibody Tests, Even Rapid Finger Pricks, are Effective, New Study Finds Association of SARS-CoV-2 seropositive antibody test with risk of future infection Evolution of antibody immunity to SARS-CoV-2 SARS-CoV-2 antibodies rapid tests: A valuable epidemiological tool in challenging settings Verder uitstellen van de tweede dosis dan wel inzet van één dosis New Covid Antibody Testing Service Opens at Shannon Airport The Blood Connection. NEW COVID-19 Antibody Testing. 2021. Available online Rapid Tests at Schiphol: Take a COVID-19 Test before Departure Pfizer and Biontech Confirm High Efficacy and No Serious Safety Concerns through up to Six Months Following Second Dose in Updated Topline Analysis of Landmark Covid-19 Vaccine Study COVID-19-Vaccine Second-Dose Completion and Interval Between First and Second Doses Among Vaccinated Persons-United States Suboptimal response to COVID-19 mRNA vaccines in hematologic malignancies patients Case Fatality Rate of Cancer Patients with COVID-19 in a New York Hospital System The author acknowledges Ryan Dagenais for editorial support with the manuscript; his efforts were funded by BioSolutions Services LLC. Christopher C. Lamb, PhD, has worked with the manufacturers of SARS-CoV-2 tests for Emergency Use Authorization submissions to the US FDA. The funders had no role in the design of the study; in the collection, analyses, or interpretation of data; in the writing of the manuscript; or in the decision to publish the results.