key: cord-0714707-vic6rbnb authors: Ko, Dae-Hyun; Kim, Hyun Soo; Hyun, Jungwon; Kim, Han-Sung; Kim, Jae-Seok; Park, Kyoung Un; Song, Wonkeun title: Comparison of the Luminex xTAG Respiratory Viral Panel Fast v2 Assay With Anyplex II RV16 Detection Kit and AdvanSure RV Real-Time RT-PCR Assay for the Detection of Respiratory Viruses date: 2017-06-20 journal: Ann Lab Med DOI: 10.3343/alm.2017.37.5.408 sha: cfb3735e8a8570efe9bc510fb05fe73441b00314 doc_id: 714707 cord_uid: vic6rbnb BACKGROUND: The accurate and rapid identification of the causative viruses is important for the timely diagnosis and management of respiratory infections. Multiplex molecular diagnostic techniques have been widely adopted to detect respiratory viruses. We compared the results of a newly upgraded, multiplex, molecular bead-based respiratory viral panel (RVP) assay with the results of Anyplex II RV16 detection kit and AdvanSure RV real-time RT-PCR assay. METHODS: We tested 254 respiratory specimens and cultured viral strains using the Luminex xTAG RVP Fast v2 assay (Luminex Molecular Diagnostics, Canada) and Anyplex II RV16 detection kit and compared the results. Specimens showing discordant results between the two assays were tested with a AdvanSure RV real-time RT-PCR assay. RESULTS: Of the 254 respiratory specimens, there was total agreement in the results between the xTAG RVP Fast v2 assay and the other real-time PCR assay in 94.1–100% of the specimens. The agreement levels were relatively low (94.1–97.6%) for specimens of adenovirus, coronavirus NL63, and parainfluenza type 3. In comparison to the other assay, the xTAG RVP Fast v2 assay detected a higher number of parainfluenza type 3 (4 cases) and metapneumovirus (9 cases). CONCLUSIONS: The xTAG RVP Fast v2 assay showed comparable capabilities compared with the other assays; it will be useful for identifying respiratory viral infections in patients with respiratory symptoms. Clinicians should be aware of the characteristics of the assays they use, since different assays show different detectability for each virus. Respiratory viral (RV) infection is one of the most common infectious diseases worldwide [1] [2] [3] . The rapid and accurate diagnosis of the underlying pathogen is crucial for establishing good clinical practices aimed at reducing morbidity and mortality [4] [5] [6] . Diagnosing RV infections involves detection of the causative viruses in specimens obtained from patients, such as sputum and/or nasopharyngeal swabs. Traditionally, virus cultures and/ or direct fluorescent-antibody assays have been used to detect the viruses [7] [8] [9] . Recently, multiplex PCR-based methods have been developed and introduced for clinical use, given their ad-vantages of decreased hands-on time and improved sensitivity [8, 9] . Luminex technology has been widely adopted in clinical laboratories for various tests such as multiplex PCR, human leukocyte antigen tests, and protein assays [10, 11] . The Luminex xTAG respiratory viral panel (RVP) and RVP FAST v1 assays (Luminex Molecular Diagnostics, Toronto, Canada) proved to be clinically useful in previous studies [12] [13] [14] [15] , and a revised version (xTAG RVP Fast v2) (Luminex Molecular Diagnostics) was recently released in 2013. The Luminex xTAG RVP Fast v2 assay is a multiplex reverse transcriptase (RT)-PCR assay that is incorporated with the Luminex tagging system. Following the first release of the Luminex xTAG RVP assay, Luminex Molecular Diagnostics issued a revised version. The new assay have several advantages over the other assays: 1) detect more types of viruses (19 strains); 2) distinguish among influenza A subtypes (H1, H3, 2009 H1N1), coronavirus HKU1, and parainfluenza type 4; 3) handle the largest number of specimens in one run (94 specimens); 4) use a small specimen volume (10 μL of nucleic acids); and 5) produce semi-quantitative results as the fluorescence intensity. However, it is unclear whether the new version of the Luminex system has similar or superior virus detection abilities compared with other available multiplex RT-PCR assays. In the present study, we evaluated the clinical performance of the Luminex xTAG RVP Fast v2 assay for detecting respiratory viruses compared with that of Anyplex II RV16 detection kit (Seegene, Seoul, Korea) and AdvanSure RV real-time RT-PCR assay (LG Life Sciences, Seoul, Korea). This study was performed retrospectively by using stored specimens. The specimens were collected in Dongtan Sacred Heart Hospital, Korea in April 2014, and from December 2014 to April 2015. During these periods, respiratory specimens, sputum, or nasopharyngeal swabs, and any nucleic acids from the specimens that showed positive results according to the Anyplex II RV16 detection kit were collected and stored at -70°C until further analysis. Stored specimens underwent further evaluations, as explained in the following sections. When the stored specimens were unavailable or the quantity was too small, then the stored nucleic acids were used for the experiments. During the collection periods, 248 stored respiratory specimens and 6 stored nucleic acids were used. The six nucleic acids were used because the frequency of coronavirus 229E and parainfluenza 2 was low and the number of positive specimens for these viruses was too small. Nucleic acids were extracted by using a QIAamp Viral RNA mini kit (Qiagen, Hilden, Germany) and a QIAcube platform (Qiagen). This study was conducted under the approval of the Institutional Review Board of Dongtan Sacred Heart Hospital (2015-069), Hwaseong, Korea. Luminex xTAG RVP Fast v2 assays were performed on the collected specimens or stored nucleic acids after thawing. If the Luminex xTAG RVP results were inconsistent with the results of the Anyplex II RV16 assay, then we repeated the Anyplex II RV16 assay for the same specimens. If the repeated Anyplex II RV16 results were different from the results of the first assay, then the Luminex xTAG RVP Fast v2 tests were repeated and/or a third PCR assay, i.e., the AdvanSure RV real-time RT-PCR assay was performed for comparison. The Anyplex II RV16 and AdvanSure RV real-time RT-PCR assays were performed by using the CFX96 real-time PCR system (Bio-Rad Laboratories, Hercules, CA, USA) and SLAN-96 real-time PCR detection system (LG Life Sciences), respectively. For specimens with complete results from all three assays, we considered that the results were more likely to be true when two of the three assays showed consistent results. All of the tests were performed according to the manufacturers' recommendations. The characteristics and detectable viruses of each kit are summarized in Table 1 . The virus culture supernatants from the adenovirus strains (types 1, 2, 3, 4, 5, 8, 11, 12, 18, 23, 31, 35, 40, and 41 ) and the enterovirus strain (Coxsackie B5) were tested by using the three kits to evaluate their ability to detect each strain. Adenovirus types 1, 2, 3, 4, 5, 8, 18, 23, and 35 were obtained from the Korea Bank for Pathologic Viruses (http://knrrb.knrrc.or.kr), whereas adenovirus types 11, 12, 31, 40, and 41, and the enterovirus (Coxsackie B5) were obtained from the American Type Culture Collection (http://www.atcc.org). Inter-rater agreement statics (Cohen's kappa) were used to compare the results of the Anyplex II RV16 and Luminex xTAG RVP fect agreement, respectively [16] . Statistical analyses were performed by using the VassarStats website (http://vassarstats.net/) and Excel 2010 (Microsoft, Redmond, WA, USA). The final results of the assays for the collected specimens are summarized in Table 2 . Among the repeated specimens, 79 specimens showed discrepancies between the two assays. In these discrepant cases, we compared the final Anyplex II RV16 assay results to those of the other two assays. When there were discrepancies among the three results, we used the results of the second (repeated) assay, because the specimens' status was considered more comparable and reliable in the second assay. The total agreement between the xTAG RVP Fast v2 and Anyplex II RV16 assays ranged from 94.1% to 100% ( Table 2 ). The Cohen's kappa value ranged from 0.66 to 1.00, and was particularly low for the adenovirus, coronavirus NL63, and parainfluenza type 3 specimens (kappa 0.66-0.77, indicating moderate agreement). All of the other viruses showed strong or almost perfect agreement between assays. When compared with the results of the Anyplex II RV16 assay, the Luminex xTAG RVP Fast Discrepant results were interpreted by considering the results of the first two assays along with those of the AdvanSure RV assay, as shown in Table 3 . All seven specimens showing positive results for coronavirus with the AdvanSure RV assay were also positive with Anyplex II RV16 but were negative with Luminex xTAG RVP Fast v2. By contrast, specimens that were positive for metapneumovirus (N = 8) or parainfluenza type 3 (N = 4) with the Ad-vanSure RV assay also showed positive results with Luminex xTAG RVP Fast v2 but were negative with Anyplex II RV16 (Table 3 ). Table 4 shows the results of each assay for the viral supernatants of various strains. The three kits had comparable performances for most of the tested strains. However, the Anyplex II RV16 assay was unable to detect adenovirus types 12 and 31, while adenovirus type 18 was not detected by the Luminex xTAG RVP Fast v2 assay. The AdvanSure RV assay missed detecting three types of adenoviruses (31, 40, and 41), and erroneously typed the enterovirus as a rhinovirus (Table 4) . In the present study, we evaluated the ability of the Luminex xTAG RVP Fast v2 assay to detect common respiratory pathogens. Its performance was compared with that of the Anyplex II RV16 assay, and any discrepancies in the results were confirmed by using a third multiplex PCR assay, AdvanSure RV. In general, the Anyplex II RV16 assay detected more pathogens than did the Luminex xTAG RVP Fast v2 assay, which was not surprising considering that only specimens that showed at least one posi-tive result by the Anyplex II RV16 assay were collected for this comparison. Nonetheless, the Luminex xTAG RVP v2 assay detected more metapneumovirus and parainfluenza type 3 specimens. The xTAG RVP v2 and Anyplex II RV16 assays showed relatively poor concordance for adenovirus, coronavirus NL63, and parainfluenza type 3 specimens. Taking into account the results of the third assay, we found that the Anyplex II RV16 assay detected more adenovirus, bocavirus, and respiratory syncytial virus type A specimens. Moreover, metapneumovirus, parainfluenza type 3, enterovirus, and rhinovirus were more likely to be detected by the Luminex xTAG RVP Fast v2 assay. The virus culture supernatants were also tested to evaluate the abilities of the kits to detect various strains. In these experiments, none of the three kits was able to detect all of the adenovirus strains. As there are more than 60 serotypes of adenoviruses [17, 18] , it may be challenging to design assays that can detect all of the subtypes. In addition, the AdvanSure RV assay classified the enterovirus as a rhinovirus because this assay was not originally designed to detect enteroviruses. Among the 254 specimens tested, 92 showed multiple infections with either the Anyplex II RV16 assay or the Luminex xTAG RVP Fast v2 assay. For cases with multiple infections, discrepancies between the two assays were revealed in 40 cases (43.5%). The discrepancy mainly resulted from the inability to detect ac-companying viruses, mostly adenovirus (13 cases), metapneumovirus (7 cases), and coronavirus NL 63 (6 cases). Among the specimens with three complete sets of assay results, there were slightly more discrepancies for cases of multiple infections than for single-infection cases (26 vs 17). However, we were unable to examine this difference statistically owing to the small sample size. The discrepant results might be due to several factors, including differences in the target genes and primers and differences in the analytical sensitivities of the assays. Previous studies have shown that different targets for amplification can affect the analytical sensitivities and specificities of assays [12, 19] . In contrast to the Anyplex II RV16 assay, the Luminex xTAG RVP FAST v2 assay could differentiate among three subtypes of the influenza A virus; thus, this assay may be more helpful, especially during the influenza season. Moreover, only the Luminex xTAG RVP Fast v2 assay could distinguish between coronavirus OC43 and HKU1 specimens. However, it should be noted that it was challenging to directly compare the detection capabilities of the assays because Luminex xTAG RVP Fast v2 reports its capabilities in units of copies/mL, whereas the other assays report the results in units of tissue culture infective dose 50 (TCID50)/ mL. Therefore, laboratory personnel should be aware of the characteristics and differences of the assays they use and should be cautious when performing the assays and reporting the results. In particular, laboratory staff and physicians should maintain close communication to ensure the best use of the assays. For the specimens with discordant results, many showed weak positive reactions ("+" reactions, based on the cycles at which the melting curve analyses revealed the presence of the targets) in the initial Anyplex II RV16 assay. A previous study using serial dilutions of plasmid standards of several viruses showed that greater discrepancies existed in low-copy specimens [13] . However, the cycle threshold values for the AdvanSure RV assay and the fluorescence intensities for the Luminex xTAG RVP Fast v2 assays were highly variable; thus, the discrepant results may be related, at least in part, to the weak positive reactions, although other unexplained factors may also be involved. Future studies should investigate these possibilities. Several studies have investigated the clinical utility of the Luminex xTAG RVP assay for detection of respiratory pathogens. While the Luminex assay showed generally higher sensitivities than traditional assays, it could detect more influenza, respiratory syncytial virus, metapneumovirus, and rhinovirus/enterovirus cases, while detect less adenovirus and coronavirus cases [12] [13] [14] [15] . And some have reported that great discrepancies were discovered for adenovirus and influenza B virus cases. Similar results were obtained in the present study, with the Luminex xTAG RVP assay displaying high detection capabilities for detecting rhinovirus/enterovirus and metapneumovirus cases but low detection capabilities for coronavirus cases. In fact, many commercial kits tend to produce discrepant results for adenovirus specimens [20, 21] . The main limitation of this study is that we chose our specimens retrospectively; thus, selection bias may have existed and affected the results. In addition, we were unable to assure the stability of the stored specimens. Indeed, some of the specimens showed discrepant results between the initial and follow-up Anyplex II RV16 assays. As sequencing analysis was not performed for the specimens, we compared the results of the assays for concordance, making it difficult to directly determine the analytical sensitivity and/or specificity of these assays. In conclusion, we evaluated the clinical utility of the Luminex xTAG RVP Fast v2 assay. We found that this assay could detect various respiratory pathogens simultaneously and that the performance was comparable to that of the existing methods in terms of agreement. The Luminex xTAG RVP Fast v2 assay may be a good alternative method in clinical laboratories for the diagnosis of RV infections. Systematic review and meta-analysis of respiratory viral coinfections in children The role of respiratory viruses in the etiology of bacterial pneumonia: an ecological perspective Influenza and other respiratory virus infections in solid organ transwww Lower respiratory viral illnesses: improved diagnosis by molecular methods and clinical impact Identification of respiratory virus in infants with congenital heart disease by comparison of different methods Early diagnosis of lower respiratory tract infections (pointof-care tests) Accuracy of rapid influenza diagnostic tests: a meta-analysis Molecular diagnosis of respiratory virus infections Strengths and weaknesses of FDA-approved/cleared diagnostic devices for the molecular detection of respiratory pathogens An evaluation of the Luminex xTAG Gastrointestinal Pathogen Panel assay for the detection of multiple diarrheal pathogens in fecal samples in Vietnam Analysis of Pregnancy-Induced Anti-HLA Antibodies Using Luminex Platform Comparison of Anyplex II RV16 with the xTAG respiratory viral panel and Seeplex RV15 for detection of respiratory viruses Comparison of three multiplex PCR assays for the detection of respiratory viral infections: evaluation of xTAG respiratory virus panel fast assay, Respi-Finder 19 assay and RespiFinder SMART 22 assay Development of a respiratory virus panel test for detection of twenty human respiratory viruses by use of multiplex PCR and a fluid microbeadbased assay Comparison of the Biofire FilmArray RP, Genmark eSensor RVP, Luminex xTAG RVPv1, and Luminex xTAG RVP fast multiplex assays for detection of respiratory viruses Interrater reliability: the kappa statistic Computational analysis of two species C human adenoviruses provides evidence of a novel virus Evaluation of an immunochromatographic assay for the rapid and simultaneous detection of rotavirus and adenovirus in stool samples Evaluation of multiple commercial molecular and conventional diagnostic assays for the detection of respiratory viruses in children Comparison of the AdvanSure real-time RT-PCR and Seeplex((R)) RV12 ACE assay for the detection of respiratory viruses Comparison of the Anyplex(TM) II RV16 and Seeplex((R)) RV12 ACE assays for the detection of respiratory viruses We thank Ji Sun Noh for excellent technical assistance. Luminex Corporation provided the xTAG RVP Fast v2 assay (Luminex Molecular Diagnostics) kits and additional financial support. The funders had no role in the study design, data collection, and interpretation, or the decision to submit the work for publication. No potential conflicts of interest relevant to this article were reported.