key: cord-0714226-xgudbraq authors: Yi, Zhijian; de Dieu Habimana, Jean; Mukama, Omar; Li, Zhiyuan; Odiwuor, Nelson; Jing, Hanzhi; Nie, Chengrong; Hu, Mei; Lin, Zuoxian; Wei, Hongping; Zeng, Lingwen title: Rational Programming of Cas12a for Early-Stage Detection of COVID-19 by Lateral Flow Assay and Portable Real-Time Fluorescence Readout Facilities date: 2021-12-26 journal: Biosensors (Basel) DOI: 10.3390/bios12010011 sha: a7954b2a72d3dc6583a2da945880039f716be5f9 doc_id: 714226 cord_uid: xgudbraq Coronavirus disease 2019 (COVID-19) caused by the SARS-CoV-2 virus has led to a global pandemic with a high spread rate and pathogenicity. Thus, with limited testing solutions, it is imperative to develop early-stage diagnostics for rapid and accurate detection of SARS-CoV-2 to contain the rapid transmission of the ongoing COVID-19 pandemic. In this regard, there remains little knowledge about the integration of the CRISPR collateral cleavage mechanism in the lateral flow assay and fluorophotometer. In the current study, we demonstrate a CRISPR/Cas12a-based collateral cleavage method for COVID-19 diagnosis using the Cas12a/crRNA complex for target recognition, reverse transcription loop-mediated isothermal amplification (RT-LAMP) for sensitivity enhancement, and a novel DNA capture probe-based lateral flow strip (LFS) or real-time fluorescence detector as the parallel system readout facility, termed CRICOLAP. Our novel approach uses a customized reporter that hybridizes an optimized complementary capture probe fixed at the test line for naked-eye result readout. The CRICOLAP system achieved ultra-sensitivity of 1 copy/µL in ~32 min by portable real-time fluorescence detection and ~60 min by LFS. Furthermore, CRICOLAP validation using 60 clinical nasopharyngeal samples previously verified with a commercial RT-PCR kit showed 97.5% and 100% sensitivity for S and N genes, respectively, and 100% specificity for both genes of SARS-CoV-2. CRICOLAP advances the CRISPR/Cas12a collateral cleavage result readout in the lateral flow assay and fluorophotometer, and it can be an alternative method for the decentralized field-deployable diagnosis of COVID-19 in remote and limited-resource locations. SARS-CoV-2 detection; however, they should only be performed by authorized clinical laboratory under the Food and Drug Administration's EUA [23, 24] . Due to its convenience with CRISPR/Cas systems and isothermal amplification techniques, a paper-based diagnostic tool known as the lateral flow strip (LFS) serves as a promising tool that can satisfy the criteria for early-stage diagnosis of any unpredicted pandemic outbreak [25, 26] . An LFS is commercially cheap, portable, and user-friendly, and it allows the high-throughput detection of massive samples in parallel. For instance, several researchers, including Broughton et al. [27] with the DETECTR assay, Ali et al. [28] with the iSCAN assay, Nguyen et al. [29] with the ENHANCE assay, and Ooi et al. [30] with the VaNGuard assay, used Milenia HybriDetect dipsticks for SARS-CoV-2 detection. The HybriDetect dipstick works on the principle of switching a test line with a control line for result interpretation. This switching mechanism of the HybriDetect dipstick has disadvantages in that it can lead to intuitive misinterpretation due to the competitive binding of the reporter to the control and test lines. More recently, Gong et al. [31] demonstrated a novel reporter in a method named TESTOR (Targeting DNA by Cas12a-based Eye Sight Testing in a Onepot Reaction) to solve the unspecificity of the HybriDetect dipstick. The concept of TESTOR is similar to that of the HybriDetect dipstick; however, the difference lies in TESTOR's novel reporter, which was designed with a phosphorothioate in the middle of the digoxin and biotin labels. Zhu et al. [25] also established a CRISPR/Cas12a-based detection system using multiple cross displacement amplification (MCDA) and an LFS for COVID-19 detection. Similar to the HybriDetect dipstick, an FITC-biotinylated reporter was employed; however, the conjugate pad, test, and control lines were modified with streptavidin-gold, biotin-bovine serum albumin, and anti-FITC, respectively. Despite the advancements of the CRISPR/Cas12a-based lateral flow assay (LFA), these assays could lead to nonspecific readouts due to the switching mechanism of the test and control lines. Herein, we report a novel alternative assay for SARS-CoV-2 diagnosis to advance the CRISPR/Cas12a-based LFA. We demonstrate a cost-effective CRISPR/Cas12a approach for COVID-19 detection using RT-LAMP to boost the sensitivity and LFS or real-time fluorescence as an independent parallel system for result readout, termed CRICOLAP. By targeting the surface glycoprotein (S)-and nucleocapsid phosphoprotein (N)-encoding genes of SARS-CoV-2, we design a biotinylated reporter and optimized its complementary capture probe in an LFA. In the presence of the target SARS-CoV-2, the reporter is transcleaved and fails to hybridize with the capture probe, indicating a positive readout. Our novel LFS shows two bands for negative samples and one band for positive samples. The results are also confirmed using a real-time portable Dhelix-Q5 isothermal fluorescence PCR machine (hereafter referred as Dhelix-Q5 for simplicity), but using a fluorescently labeled reporter. Thus, the developed approaches could be used for the rapid, accurate, and inexpensive detection of infectious diseases. All oligonucleotide sequences (Supplementary Materials, Tables S1-S4) were synthesized by Sangon Biotech (Shanghai, China). Bst 2.0 DNA polymerase, Avian Myeloblastosis (AMV) reverse transcriptase, high purity LbaCas12a, NEB buffer 2.1, and deoxynucleoside triphosphates (dNTPs) were obtained from New England Biolabs (Ipswich, MA, USA). Streptavidin (SA), tetrachloroauric(III) acid (HAuCl 4 aq), and trisodium citrate were purchased from Sigma-Aldrich (Steinheim, Germany). Fiberglass, nitrocellulose membrane, and absorbent papers were from Millipore (Billerica, MA, USA). The dispenser machine was purchased from Shanghai Kinbio (Shanghai, China). The portable Dhelix-Q5 isothermal fluorescence PCR machine (portable Dhelix-Q5) was purchased from Guangzhou international biological island (Guangzhou, China). The remaining reagents, such as buffers, were prepared in our laboratory. Both target genes were cloned into the pcDNA3.1+ vector. The N gene was cloned into the pcDNA3.1+ vector at the HindIII/EcoRI site, while the S gene was inserted behind the N gene at the EcoRI/XhoI site to obtain a fused S + N sequence (Table S1 ). The crRNAs targeting the S and N genes (Table S2) were designed using Benchling online software, and the ones with a higher score were selected. Premier Biosoft and PrimerExplorer V5 online software were used to design LAMP primers (Table S3 ) and specifically confirmed using the NCBI-BLAST database. RT-LAMP reaction was performed following the NEB protocol (New England, USA). Typically, 25 µL of each reaction contained 10× isothermal amplification buffer; 6 mM of MgSO4; 8 U/µL of Bst 2.0 DNA polymerase; 7 U/µL of AMV reverse transcriptase; primers F3/B3 (0.2 µM), FIB/FIP (1.6 µM), and Loop F/Loop B (0.4 µM); 1.4 mM dNTPs; 2 µL of RNA target; and water to the final volume. The mixture was sealed and incubated at 65 • C for 25 min in a water bath. For fluorescence detection, SYTO-9 was added, and this was followed by incubation in a Dhelix-Q5 (λ ex = 470 nm; λ em = 520 nm). The CRISPR/Cas12a reaction for target and reporter cleavage was performed following a previous method with minor modifications [21] . Within a 20 µL mix, the reaction included 10 µL of 2× cleavage buffer [1% glycerol, 10 mM MgCl 2 , 150 mM KCl, 0.5 mM DTT, 20 nM HEPES (pH 7.5)], 50 nM of Cas12a, 500 nM of fluorophore-quencher ssDNAlabeled reporter (FQ-ssDNA reporter), 40 nM of activator, 50 nM of crRNA, 1 µL of target amplicons, and sterile nuclease-free water. After mixing thoroughly, the mixture was incubated at 37 • C for 30 min to ensure sufficient cleavage. The trans-cleavage for fluorescence detection was monitored using a Dhelix-Q5. For the LFA, the protocol was the same; however, the FQ-ssDNA reporter was replaced by our customized biotinylated ssDNA reporter (bio-ssDNA reporter) (Table S4) , and the cleavage activity on the LFS was monitored by the naked eye. A complete LFS basically consists of five main parts: the sample pad, conjugate pad, nitrocellulose membrane (NC), absorbent pad, and backing pad [19] . The sample pad fabricated from glass fibers was thoroughly wetted with a buffer of pH 8.0 (2% glucose, 1% Triton, 50 mM boric acid, 1% BSA), dried, and then kept in a desiccant container. The conjugate pad, also made of glass fibers, was embedded with the AuNP-SA solution [32] and dried thoroughly before use. Using a dispenser machine, the test and control lines on the NC membrane were sprayed with a customized DNA capture probe and biotinylated rabbit polyclonal anti-IgG antibody solutions, respectively. The distance between the test and control lines was estimated to be 7 mm, and then the NC membrane (25 mm in width) was dried and kept at room temperature for 12 h and 4 • C until further use. Lastly, all the LFS parts were collectively attached, each overlapping by 2 mm. Finally, the LFS with a 4 mm width was readily obtained using an LFS cutter. Nasopharyngeal swabs stored in viral transport medium (VTM) at the Wuhan Institute of Virology (WIV), Chinese Academy of Sciences, were inactivated in a biosafety level 2 laboratory by heating at 95 • C for 5 min. The total RNA was extracted from the inactivated samples using a magnetic bead-based extraction kit, FineMag Quick Viral DNA/RNA (Genfine Biotech, Changzhou, China), on an automated device, Purifier Modesty (Genfine). China CDC authorized WIV to handle COVID-19 patients' samples during the outbreak in Wuhan, China. The RT-qPCR was performed using a commercial kit from Sansure Biotech (Changsha, China) officially approved for COVID-19 testing. The kit targets ORF1ab and the N gene of SARS-CoV-2 together with the human internal control gene (IC). RT-qPCR was conducted in a 50 µL reaction tube consisting of 26 µL of reaction solution, 4 µL of enzyme mixture, and 20 µL of template RNA. The reaction was performed on a CFX96 DeepWell real-time system (BIORAD, Hercules, CA, USA). The thermocycling protocol was set as follows: reverse transcription was proceeded at 50 • C for 30 min, and cDNA pre-denaturation occurred at 95 • C for 1 min, followed by 45 cycles at 95 • C for 15 s. The annealing and fluorescence acquisition steps were conducted at 60 • C for 30 s. A sample was considered positive with a cycle threshold (Ct) of less than 40, and the test was valid only if the cycle threshold (Ct) was less than 35 (Ct < 35). In the present study, CRICOLAP was established by integrating the CRISPR/Cas12a coupled with RT-LAMP to boost the sensitivity and two independent parallel systems (LFS and real-time fluorescence) for result readout. This study targeted the S and N conserved genes of SARS-CoV-2 (Scheme 1A). As shown in Scheme 1B, the extracted RNAs were RT-LAMP pre-amplified (Scheme 1B, a) and applied to Cas12a reaction for collateral cleavage (Scheme 1B, b1 and b2) and then to the LFS or real-time fluorescence reader for results readout. For the LFS, we customized a novel LFA based on the CRISPR/Cas12a trans-cleavage mechanism. Researchers have shown that a DNA probe of more than 15 bases generates a strong signal on the LFS via DNA hybridization [33] . Furthermore, the CRISPR/Cas12-based FQ-ssDNA reporter (FQ-TTTATT/FQ-3TA2T) was reported to release fluorescence upon trans-cleavage by Cas12a [21] . Therefore, we sought to customize the fluorescent reporter for use in our LFS. From various attempts (Table S4) , we realized that a biotinylated-ssDNA reporter of 11 bases (Bio-TTTTTTTTATT/bio-8TA2T) could strongly hybridize with a 32-base ssDNA probe sequence (Table S5 ) on the LFS test line; however, it was undetectable once cleaved by Cas12a ( Figure S1 ). Therefore, when the CRISPR/Cas12a mixture is subjected to the sample pad, it migrates to the conjugate pad containing the AuNP-SA complex and flows toward the test line and control line across the NC membrane. The aggregation of the bio-ssDNA reporter at the test line reveals a negative test (Scheme 1B, c1), while the disappearance of a visual red-like test line indicates a positive test resulting from efficient reporter trans-cleavage (Scheme 1B, c2). In parallel, for fluorescence analysis, the FQ-ssDNA reporter remained intact due to the absence of the target (Scheme 1B, d1), while in the presence of the target, the FQ-ssDNA reporter fluoresced gradually as a result of trans-cleavage (Scheme 1B, d2). The RT-LAMP and CRISPR/Cas12a experimental conditions were optimized to improve the trans-cleavage efficiency. Firstly, six and seven different primer sets (Table S3) for the S and N genes were used in RT-LAMP. Of the six sets of the S gene, the S1 and S4 primer sets demonstrated excellent amplification in electrophoresis ( Figure S2a ) and real-time fluorescence assays ( Figure 1A1 ). However, S1 showed a fast response and was selected for further experiments. For the N gene, the N1 set was selected ( Figure S2b and Figure 1B1 ). For the temperature assessment, RT-LAMP was tested from 55 to 65 • C, and a maximum signal was obtained at 65 • C ( Figure S2c The RT-LAMP and CRISPR/Cas12a experimental conditions were optimized to improve the trans-cleavage efficiency. Firstly, six and seven different primer sets (Table S3) for the S and N genes were used in RT-LAMP. Of the six sets of the S gene, the S1 and S4 primer sets demonstrated excellent amplification in electrophoresis ( Figure S2a ) and realtime fluorescence assays ( Figure 1A1 ). However, S1 showed a fast response and was selected for further experiments. For the N gene, the N1 set was selected ( Figures S2b and 1B1 ). For the temperature assessment, RT-LAMP was tested from 55 to 65 °C, and a maximum signal was obtained at 65 °C ( Figure S2c Three crRNAs for each target gene were designed, and the ones with a higher score and target matching efficiency were selected ( Figure S3a,b) . For the CRICOLAP validation, we performed separate reactions by omitting individual components in the CRISPR reaction (crRNA, activator, Cas12a enzyme, reporter, and the target). Our assay was validated by confirming the ability of crRNA to guide Cas12a to trans-cleave the reporter in the presence of all reaction components. There was no reporter trans-cleavage in the absence of one reaction component except for the S gene activator ( Figure 2A1 ), presuming that the S gene activated Cas12a per se [21, 34] . Moreover, the cleavage could not occur in the absence of the activator for the N gene ( Figure 2B1 ). The optimum reporter concentrations were 1000 nM and 500 nM for the S and N genes, respectively ( Figure 2A2 ,B2). Three crRNAs for each target gene were designed, and the ones with a higher score and target matching efficiency were selected ( Figure S3a,b) . For the CRICOLAP validation, we performed separate reactions by omitting individual components in the CRISPR reaction (crRNA, activator, Cas12a enzyme, reporter, and the target). Our assay was validated by confirming the ability of crRNA to guide Cas12a to trans-cleave the reporter in the presence of all reaction components. There was no reporter trans-cleavage in the absence of one reaction component except for the S gene activator ( Figure 2A1 ), presuming that the S gene activated Cas12a per se [21, 34] . Moreover, the cleavage could not occur in the absence of the activator for the N gene ( Figure 2B1 ). The optimum reporter concentrations were 1000 nM and 500 nM for the S and N genes, respectively ( Figure 2A2,B2) . Subsequently, the reporter concentrations varying from 0 nM to 100 nM were assessed for LFA. An intense test line was observed at 25 nM and 10 nM for S and N genes, respectively, in the absence of the target and no test line in the presence of the target ( Figure 2A3,B3) . Finally, we assessed the required time for a complete trans-cleavage that can result in a non-detectable test line upon hybridization of the reporter to the capture probe. Forty and twenty minutes were enough to ensure the complete cleavage of a bio-ssDNA reporter ( Figure 2A4,B4) . However, a detectable signal at the LFS resulting from hybridization could be observed within 5 min. Subsequently, the reporter concentrations varying from 0 nM to 100 nM were assessed for LFA. An intense test line was observed at 25 nM and 10 nM for S and N genes, respectively, in the absence of the target and no test line in the presence of the target (Figure 2A3,B3) . Finally, we assessed the required time for a complete trans-cleavage that can result in a non-detectable test line upon hybridization of the reporter to the capture probe. Forty and twenty minutes were enough to ensure the complete cleavage of a bio-ssDNA reporter ( Figure 2A4,B4) . However, a detectable signal at the LFS resulting from hybridization could be observed within 5 min. We assessed the efficiency of four and six LAMP primers. The four primers alone without loop primers achieved a limited sensitivity, while the sensitivity drastically increased with loop primers (Figure 3A1 ,B1; Figure S4a ,b). This activity could be attributed to the LAMP's rolling cycle amplification [35] . Next, the overall CRICOLAP sensitivity was further evaluated using real-time fluorescence analysis and LFA. First, various concentrations of RT-LAMP amplicons ranging from 0 to 10 6 copies/µL were investigated. Our method achieved one copy sensitivity for both the S and N genes ( Figure 3A2,B2) . Second, by replacing FQ-ssDNA with a bio-ssDNA reporter, similar sensitivity was achieved for LFA except for the S gene, which could not attain one copy resolution. CRICOLAP's LFS identified as low as five copies/µL ( Figure 3A3 ) and one copy/µL ( Figure 3B3 ) for the S and N genes, respectively. Intriguingly, there was no significant difference in sensitivity between the synthetic RNA and the RNA extracted from patients' samples, which is in accordance with Ding et al. [36] . We assessed the efficiency of four and six LAMP primers. The four primers alone without loop primers achieved a limited sensitivity, while the sensitivity drastically increased with loop primers (Figure 3A1,B1; Figure S4a ,b). This activity could be attributed to the LAMP's rolling cycle amplification [35] . Next, the overall CRICOLAP sensitivity was further evaluated using real-time fluorescence analysis and LFA. First, various concentrations of RT-LAMP amplicons ranging from 0 to 10 6 copies/µL were investigated. Our method achieved one copy sensitivity for both the S and N genes ( Figure 3A2,B2) . Second, by replacing FQ-ssDNA with a bio-ssDNA reporter, similar sensitivity was achieved for LFA except for the S gene, which could not attain one copy resolution. CRICOLAP's LFS identified as low as five copies/µL ( Figure 3A3 ) and one copy/µL (Figure 3B3 ) for the S and N genes, respectively. Intriguingly, there was no significant difference in sensitivity between the synthetic RNA and the RNA extracted from patients' samples, which is in accordance with Ding et al. [36] . B3) . For RT-LAMP and CRISPR cleavage assays, each reaction was performed using 2 µL and 1 µL, respectively, using the following serial dilutions: 0, 1, 10, 10 2 , 10 3 , 10 4 , 10 5 , and 10 6 copies/µL. Figure 3 . CRICOLAP sensitivity evaluation. RT-LAMP sensitivity with or without loop primers for S (A1) and N genes (B1). CRICOLAP sensitivity monitoring via real-time fluorescence analysis (A2,B2) and LFA (A3,B3). For RT-LAMP and CRISPR cleavage assays, each reaction was performed using 2 µL and 1 µL, respectively, using the following serial dilutions: 0, 1, 10, 10 2 , 10 3 , 10 4 , 10 5 , and 10 6 copies/µL. CRICOLAP sensitivity aligns with the recent CRISPR/Cas12-isothermal amplificationbased techniques, but with different concepts [37, 38] . For instance, two different groups [27, 28] reported an LOD of 10 copies/µL using CRISPR/Cas12a, LAMP, and commercial Milenia HybriDetect dipsticks for colorimetric detection and a laboratory-centered fluorophotometer. Similarly, by substituting LAMP with RPA, Talwar et al. [2] and Gong et al. [31] also achieved LODs of 10 and 5 copies/reaction, respectively. In the HybriDetect dipstick, a test line is immobilized with an anti-goat antibody that captures AuNPs with goat anti-FITC antibody, whereas the control line is coated with a streptavidin that captures the FITC-biotinylated reporter. Upon trans-cleavage, a biotin-labelled reporter fails to capture AuNPs-goat anti-FITC antibody, thereby forming a test line. This switching format has disadvantages in that it can lead to intuitive misinterpretation due to the competitive binding of the reporter to the control and test lines, a limitation that CRICOLAP overcomes. To assess the specificity of CRICOLAP, various species (Streptococcus pneumoniae, Pseudomonas aeruginosa, Mycobacterium tuberculosis, Staphylococcus aureus, Bacillus spp., and Escherichia coli) that usually infect the human respiratory tract and impair the breathing system in a similar manner to SARS-CoV-2 were examined. As shown in Figure 4A1 ,A2,B1,B2, the cleavage occurred only in the SARS-CoV-2 target containing samples, proving no crossreactivity with other species. We also used MEGA X software to align the crRNAs (Scr-RNA1 and NcrRNA2) with SARS-CoV-2 variants and other human coronavirus genomes ( Figure 4A3,B3) . Interestingly, we found that our crRNAs completely matched with all the SARS-CoV-2 variants (SARS-CoV-2 B1.1.7 alpha, SARS-CoV-2 B.1.351 beta, SARS-CoV-2 P.1 gamma, and SARS-CoV-2 B.1.617 delta), indicating a strong specificity of our crRNAs. Moreover, we noticed that they were not related to other viral sequences. Unfortunately, NcrRNA2 appeared to perfectly align with SARS NC_004718.3 ( Figure 4B3) ; however, NcrRNA3 could not (result not shown), which may suggest the avoidance of false positives. Furthermore, because crRNAs have been demonstrated to be sensitive to single-base mismatch, the crRNAs targeting the S and N genes of SARS-CoV-2 had many mismatches (more than three mismatches) to other coronaviruses, indicating the excellent specificity of CRICOLAP. To validate the feasibility of CRICOLAP, 60 clinical patients' samples were examined (namely, 40 positive and 20 negative samples) initially verified by a commercial RT-PCR kit ( Figure S5) . As shown by a heat map, Figure 5A1 , the fluorescence intensity drastically increased within all the positive clinical samples due to the FQ-ssDNA reporter transcleavage, implying the presence of SARS-CoV-2 ( Figure S5 ). The distribution of corresponding data is indicated by the box and whisker plots ( Figure 5A2,B2) and is depicted based on the detection threshold generating fluorescence intensity in a timely manner dependent on the CRISPR reaction (7 min, n = 6). The significant threshold fluorescence value (denoted with a horizontal dashed line, Figure 5A2 ,B2) calculated based on the three standard deviations above the mean background fluorescence (BGF) [39] was obtained at 7 min of the validated negative controls (n = 20), the highest recorded BGF being 12.3 and To validate the feasibility of CRICOLAP, 60 clinical patients' samples were examined (namely, 40 positive and 20 negative samples) initially verified by a commercial RT-PCR kit ( Figure S5) . As shown by a heat map, Figure 5A1 , the fluorescence intensity drastically increased within all the positive clinical samples due to the FQ-ssDNA reporter trans-cleavage, implying the presence of SARS-CoV-2 ( Figure S5 ). The distribution of corresponding data is indicated by the box and whisker plots ( Figure 5A2,B2) and is depicted based on the detection threshold generating fluorescence intensity in a timely manner dependent on the CRISPR reaction (7 min, n = 6). The significant threshold fluorescence value (denoted with a horizontal dashed line, Figure 5A2 ,B2) calculated based on the three standard deviations above the mean background fluorescence (BGF) [39] was obtained at 7 min of the validated negative controls (n = 20), the highest recorded BGF being 12.3 and 96 for S and N genes, respectively. Therefore, the sample was confirmed positive if it was above 17 for the S gene and above 103 for the N gene, and negative otherwise. Of the 40 positive samples examined, patient sample numbers 38 and 39 were negative for the ORF1ab gene tested by a commercial RT-PCR kit ( Figure S5a) ; however, they were identified to be positive for N and IC genes as well as S and N genes by the use of CRICOLAP. Biosensors 2021, 11, x FOR PEER REVIEW 12 of 16 above 17 for the S gene and above 103 for the N gene, and negative otherwise. Of the 40 positive samples examined, patient sample numbers 38 and 39 were negative for the ORF1ab gene tested by a commercial RT-PCR kit ( Figure S5a) ; however, they were identified to be positive for N and IC genes as well as S and N genes by the use of CRICOLAP. In parallel, the LFS showed no visual test lines for the ten positive clinical samples ( Figure 5A3,B3) . However, a visible test line was detected by the naked eye in the ten negative samples due to the failure of trans-cleavage, which resulted in the hybridization of the reporter and capture probe at the test line ( Figure 5A4,B4) . CRICOLAP showed 97.5% and 100% sensitivity for S and N genes and 100% specificity consistency with the validated commercial RT-PCR data for both negative and positive patient samples (Table S6 ). In the wake of the raging COVID-19 pandemic, an increasing demand for highly effective diagnostics for SARS-CoV-2 detection has been observed. With the breakthrough utilization of CRISPR/Cas in diagnosis based on the trans-cleavage of the ssDNA reporter In parallel, the LFS showed no visual test lines for the ten positive clinical samples ( Figure 5A3,B3) . However, a visible test line was detected by the naked eye in the ten negative samples due to the failure of trans-cleavage, which resulted in the hybridization of the reporter and capture probe at the test line ( Figure 5A4,B4) . CRICOLAP showed 97.5% and 100% sensitivity for S and N genes and 100% specificity consistency with the validated commercial RT-PCR data for both negative and positive patient samples (Table S6 ). In the wake of the raging COVID-19 pandemic, an increasing demand for highly effective diagnostics for SARS-CoV-2 detection has been observed. With the breakthrough utilization of CRISPR/Cas in diagnosis based on the trans-cleavage of the ssDNA reporter in the presence of a target [19] , we successfully customized a novel CRISPR reporter for use in LFA and fluorophotometry in the assay termed CRICOLAP. The assay accurately identified SARS-CoV-2 from 60 clinical patients' samples initially confirmed positive or negative by a commercial RT-PCR kit approved by China CDC. Our novel LFS shows one band and two bands for SARS-CoV-2-positive and -negative patients' samples, respectively, unlike the conventional LFA. Coupling any detection assay with an on-site visual readout is essential, particularly when equipment is not affordable. CRICOLAP LFS is cheap (USD 0.1, excluding the CRISPR reagents) ( Table 1) and can be deployed for on-site testing. According to Milenia [40] and Gang et al. [31] , the detection mechanism of switching the test and control lines in Milenia HybriDetect may sometimes lead to nonspecific signal readouts at the test line, because both test and control line signals depend on the reporter trans-cleavage. This may make discerning the bands of weak positive patient samples from the bands caused by nonspecific faint bands challenging. To solve this, we successfully customized a reporter that complements an optimized capture probe. As such, the signal at the test line could solely be obtained without switching the test and control lines to prevent intuitive misinterpretation. More importantly, since CRISPR/Cas12a machinery relies on three main elements (Cas12a, crRNA, and the reporter) to achieve nucleic acid detection, we were able to design and transduce the signal directly from the CRISPR/Cas12a machinery without using an auxiliary carrier, such as biotin, streptavidin, or digoxin, as observed in several CRISPR/Cas12a-based LFA [2, 25, [27] [28] [29] [30] [31] ( Table 1 ). The concept of CRICOLAP LFS can also serve as a state of the art to harness the trans-cleavage activity, which we believe can increase sensitivity and specificity, because the signal reporter of the CRISPR/Cas machinery is directly captured by a signal amplifier (capture probe) at the test line. On the other hand, the Dhelix-Q5 Isothermal Fluorescence PCR instrument is portable (~4.3 kg) and ten to fifteen times cheaper than RT-qPCR, and it can monitor real-time fluorescence detection with a high signal-to-background. Moreover, it does not require thermocycling protocols as it uses a constant temperature to generate the results. Both methods of CRICOLAP hold potential due to their abilities to detect few copies of S or N genes of SARS-CoV-2 in clinical nasopharyngeal swabs. CRICOLAP LFS may be hampered by false positives from spurious amplicons and unsuccessful reporter trans-cleavage. These can be prevented by appropriate reagent manipulation and proper incubation time monitoring, respectively. Additionally, the LAMP cross-contamination can be eliminated through a two-step loading of primers, with forward, backward, and loop primers within the first 10 min, followed by inner primers. This system stability in SARS-CoV-2 detection promises one-pot feasibility toward minimizing the risks of spurious contamination and repeated reagent manipulation. Another attractive feature to achieve one-pot detection would be the integration of a thermostable enzyme such as Alicyclobacillus acidophilus Cas12b (AapCas12b) and LAMP since AapCas12b retains optimum nuclease activity over a wide range of temperatures (31-59 • C) [41] . This can also facilitate automation, using a custom, handheld cartridge to streamline the testing workflow, thereby enabling POC testing in remote settings. In the future, CRICOLAP could emerge as a robust alternative assay to provide timely support in the battle against infectious diseases in remote and limited-resource areas. USD < 3 [28] Supplementary Materials: The following are available online at https://www.mdpi.com/article/ 10.3390/bios12010011/s1, Figure S1 : Test line signal optimization. (a) Assessment of different lengths of reporters with thymine and adenine base extension. (b) Assessment of different lengths of capture probes with adenine base extension, Figure S2 : CRICOLAP RT-LAMP optimization (primer selection, temperature optimization, and reaction time optimization), Figure S3 : Selection of crRNAs, Figure S4 : CRICOLAP sensitivity evaluation. RT-LAMP sensitivity with or without loop primers, Figure S5 : Validation of 60 clinical patient nasopharyngeal swabs using a commercial RT-PCR kit, Figure S6 : LFA validation for S gene, Figure S7 : LFA validation for N gene, Table S1 : Selected oligonucleotide sequences (optimal) used in this study targeting S and N genes of COVID-19, Table S2 : crRNAs used in this study, Table S3 : LAMP primers used in this study, Table S4 : List of reporter sequences used in this study, Table S5 : List of capture probes used in this study, Table S6 : Concordance comparison between CRICOLAP and RT-qPCR for 60 patients' samples. Institutional Review Board Statement: The study was conducted according to the guidelines of the Declaration of Helsinki, and approved by the Ethics Committee of Wuhan Institute of Virology, Chinese Academy of Science (WIVH17202001, 24 July, 2020). Syncytia formation during SARS-CoV-2 lung infection: A disastrous unity to eliminate lymphocytes Detection of Infectious Viruses Using CRISPR-Cas12-Based Assay FDA COVID-19 Response At-A-Glance Summary as of Clinical Utility of Biosensing Platforms for Confirmation of SARS-CoV-2 In-fection Asymptomatic carrier state, acute respiratory disease, and pneumonia due to severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2): Facts and myths Risking further COVID-19 waves despite vaccination Diagnosing COVID-19: The Disease and Tools for Detection Molecular Diagnosis of COVID-19: Challenges and Research Needs Point-of-Care PCR Assays for COVID-19 Detection Rapid Detection of COVID-19 Coronavirus Using a Reverse Transcriptional Loop-Mediated Isothermal Amplification (RT-LAMP) Diagnostic Platform CRISPR/Cas12a-Based Ultrasensitive and Rapid De-tection of JAK2 V617F Somatic Mutation in Myeloproliferative Neoplasms Programmable RNA editing with compact CRISPR-Cas13 systems from uncultivated microbes Detect and destroy: CRISPR-based technologies for the response against viruses Development of CRISPR-Cas systems for genome editing and beyond Detection of CRISPR-Cas9-Mediated Mutations Using a Carbon Nanotube-Modified Electrochemical Genosensor Nucleic Acid Detection Using CRISPR/Cas Biosensing Technologies Specific Detection of Influenza A and B Viruses by CRISPR-Cas12a-Based Assay Functionally diverse type V CRISPR-Cas systems Multiplexed and portable nucleic acid de-tection platform with Cas13, Cas12a, and Csm6 Exploring the Trans-Cleavage Activity of CRISPR-Cas12a (cpf1) for the Development of a Universal Electrochemical Biosensor CRISPR-Cas12a target binding un-leashes indiscriminate single-stranded DNase activity Nucleic acid detection with CRISPR-Cas13a/C2c2 Accelerated Emergency Use Authorization (EUA) Summary, SARS-COV-2 FDA (Food and Drug Administration). Sherlock CRISPR SARS-CoV-2 Kit Rapid, Ultrasensitive, and Highly Specific Diagnosis of COVID-19 by CRISPR-Based Detection Clinical validation of a Cas13-based assay for the detection of SARS-CoV-2 RNA Sotomayor-Gonzalez, A. CRISPR-Cas12-based detection of SARS-CoV-2 iSCAN: An RT-LAMP-coupled CRISPR-Cas12 module for rapid, sensitive detection of SARS-CoV-2 Enhancement of trans-cleavage activity of Cas12a with engineered crRNA enables amplified nucleic acid detection An engineered CRISPR-Cas12a variant and DNA-RNA hybrid guides enable robust and rapid COVID-19 testing An enhanced method for nucleic acid detection with CRISPR-Cas12a using phosphorothioate modified primers and optimized gold-nanopaticle strip Gold nanoparticle-streptavidin conjugates for rapid and efficient screening of aptamer function in lateral flow sensors using novel CD4-binding aptamers identified through Crossover-SELEX Lateral flow nucleic acid biosensor for Cu2+ detection in aqueous solution with high sensitivity and selectivity Integration of logic gates to CRISPR/Cas12a system for rapid and sensitive detection of pathogenic bacterial genes Evaluation of CRISPR/Cas12a-based DNA detection for fast pathogen diagnosis and GMO test in rice Ultrasensitive and visual detection of SARS-CoV-2 using all-in-one dual CRISPR-Cas12a assay CRISPR-based detection of SARS-CoV-2: A review from sample to result CRISPR is a useful biological tool for detecting nucleic acid of SARS-CoV-2 in human clinical samples Ultra-sensitive and high-throughput CRISPR-Powered COVID-19 diagnosis Lateral Flow Readout for CRISPR/Cas-Based Detection Strategies Repurposing CRISPR-Cas12b for mammalian genome engineering The authors declare no conflict of interest.