key: cord-0713713-auakr5qu authors: Akashi, Yusaku; Horie, Michiko; Kiyotaki, Junichi; Takeuchi, Yuto; Togashi, Kenichi; Adachi, Yuki; Ueda, Atsuo; Notake, Shigeyuki; Nakamura, Koji; Terada, Norihiko; Kurihara, Yoko; Kiyasu, Yoshihiko; Suzuki, Hiromichi title: Clinical Performance of the cobas Liat SARS-CoV-2 & Influenza A/B Assay in Nasal Samples date: 2022-04-07 journal: Mol Diagn Ther DOI: 10.1007/s40291-022-00580-8 sha: 4c37f15d44009729e0dde71a95a52577c63b3460 doc_id: 713713 cord_uid: auakr5qu BACKGROUND AND OBJECTIVE: Point-of-care type molecular diagnostic tests have been used for detecting SARS-CoV-2, although their clinical utility with nasal samples has yet to be established. This study evaluated the clinical performance of the cobas Liat SARS-CoV-2 & Influenza A/B (Liat) assay in nasal samples. METHODS: Nasal and nasopharyngeal samples were collected and were tested using the Liat, the cobas 6800 system and the cobas SARS-CoV-2 & Influenza A/B (cobas), and a method developed by National Institute of Infectious Diseases, Japan (NIID). RESULTS: A total of 814 nasal samples were collected. The Liat assay was positive for SARS-CoV-2 in 113 (13.9%). The total, positive, and negative concordance rate between the Liat and cobas/NIID assays were 99.3%/98.4%, 99.1%/100%, and 99.3%/98.2%, respectively. Five samples were positive only using the Liat assay. Their Ct values ranged from 31.9 to 37.2. The Ct values of the Liat assay were significantly lower (p < 0.001) but were correlated (p < 0.001) with those of other molecular assays. In the participants who tested positive for SARS-CoV-2 on the Liat assay using nasopharyngeal samples, 88.2% of their nasal samples also tested positive using the Liat assay. CONCLUSION: The Liat assay showed high concordance with other molecular assays in nasal samples. Some discordance occurred in samples with Ct values > 30 on the Liat assay. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1007/s40291-022-00580-8. The severe acute coronavirus 2 (SARS-CoV-2) pandemic has had a detrimental effect on society globally [1] . The introduction of effective vaccines [2] and treatment [3] were expected to effectively control the pandemic; however, the numbers of new infections and deaths have increased in some countries [1] . Currently, population-based screening and early detection and isolation of infected individuals are still key to effective infection control [4] . The standard method for detecting SARS-CoV-2 is molecular testing because of its high diagnostic performance [5] . However, molecular diagnostics are less widely available and less convenient than antigen testing, and the turnaround time is longer [6] . Several molecular point-ofcare tests (POCTs) have been developed [7] and applied in clinical settings to overcome the disadvantages of traditional molecular tests. The cobas Liat system (Roche Molecular Systems, Inc., Pleasanton, CA, USA) is a small analyzer that automatically performs real-time polymerase chain reaction (PCR) testing using simple procedures and provides results with a short turnaround time [8] . The system has been used for point-ofcare testing, and Liat assays are available for influenza virus [9] , respiratory syncytial virus [10] , Clostridioides difficile [11] , group A Streptococcus [12] , and SARS-CoV-2 [8] . The Liat system and cobas Liat SARS-CoV-2 & Influenza A/B (Liat) has been shown to have high sensitivity for detecting SARS-CoV-2 in nasopharyngeal samples [8] , but the diagnostic performance with nasal samples has not been reported. Nasal sample collection is easier and less invasive than the nasopharyngeal sample collection [13] ; therefore, the clinical utility of Liat would be greater if it could accurately detect SARS-CoV-2 in nasal samples. This study aimed to clarify the diagnostic performance of the Liat SARS-CoV-2 assay with nasal samples. The results of Liat assays with nasal samples were prospectively compared with those of other assays with both nasal and nasopharyngeal samples. This study was conducted in a PCR center located in Tsukuba Medical Center Hospital (TMCH) in Japan between 7 and 29 July 2021. The participants were outpatients suspected of having SARS-CoV-2 infection due to their symptoms or close contact history, all of whom underwent in-house PCR testing at the center [14] . They included individuals referred from 51 healthcare facilities and a local healthcare center, and healthcare workers working at TMCH. Clinical information of the participants was routinely recorded along with the in-house PCR. Verbal informed consent was obtained from all participants, and the ethics board of the University of Tsukuba approved the study (approval number: R03-41). One nasopharyngeal and two nasal samples were collected from each participant. The nasopharyngeal samples were collected as described previously [15] . Nasal samples were collected from the anterior nostrils according to recommended procedures [16] . Briefly, the swab was inserted to a depth of approximately 2 cm and rotated four times against the nasal mucosa. The procedure was repeated in the other nostril using the same swab. One nasal swab sample was used for molecular testing using Liat, and the other nasal swab sample was used for the evaluation of antigen testing in another study. All nasal sample collections for molecular testing followed those for antigen testing. Both the nasal and the nasopharyngeal swab samples were diluted in Universal Transport Medium (UTM TM ; Copan Diagnostics Inc., Murrieta, CA, USA) and stored at -80 °C. The nasopharyngeal samples were frozen after being used for an in-house PCR test. After being thawed, all nasal samples underwent PCR testing according to three methods: (i) cobas Liat system and cobas Liat SARS-CoV-2 & Influenza A/B (Liat), (ii) cobas 6800 system and cobas SARS-CoV-2 & Influenza A/B (cobas; Roche Molecular Systems, Branchburg, NJ, USA) [17] , and (iii) a national standard method developed by the National Institute of Infectious Diseases (NIID), Japan [18] . If a sample showed positivity on one of the methods and negativity on the other two, they were tested using the Gene Xpert system and the Xpert Xpress SARS-CoV-2 (Xpert Xpress; Cepheid, Sunnyvale, CA, USA) [19] . The nasopharyngeal samples were tested using Liat and the NIID test for the evaluation, and were not tested using the cobas system due to an insufficient amount of residual UTM sample. If the Liat and NIID results were discordant, the samples were tested using Xpert Xpress, in a similar manner to the nasal samples. Liat, cobas, and Xpert Xpress perform sample preparation (purification and the extraction of RNA), real-time PCR, and the detection of the viruses using a fully automated process. For the Liat assay, a total of 200 μL of UTM TM sample was loaded into the test cartridge, which was then inserted into the system. The Liat test targets the ORF1a/b and N gene and shows positive results if one or both genes are detected. We re-tested samples if invalid results occurred. For the cobas assay, we used 1 mL of UTM TM sample (400 μL for the analysis and 600 μL for the dead space). The targeted regions were ORF1a/b and E gene, and the cobas test provided separate results for each target. The results of cobas were considered positive for SARS-CoV-2 when samples tested positive for one of the two targets. The Xpert Xpress used 300 μL of UTM TM sample and targeted the E and N2 genes. Similar to cobas, Xpert Xpress showed separate results for each target. All three assays were conducted according to the manufacturer's instructions. For the NIID test, purification and RNA extraction were performed using MagNA Pure 96 total NA Isolation Kit and the MagNA Pure 96 Instrument (Roche Molecular Systems, Branchburg, NJ, USA) from 140-µL aliquots of UTM TM sample. The NIID test targets the N2 region, and the equipment used for the RT-PCR included the PCR LightCycler ® 480 Instrument II (Roche Diagnostics International Ltd, Rotkreuz, Switzerland), the QuantiTect ® Probe RT-PCR Kit (QIAGEN, Hilden, Germany), and a SARS-CoV-2 positive control (Nihon Gene Research Laboratories, Sendai, Japan). The RT-PCR was performed in duplicate. The Ct values were automatically calculated after the detection of SARS-CoV-2 for all test methods used in this study. Apart from detecting SARS-CoV-2, the Liat assay can also detect influenza virus; however, the results were not re-evaluated and validated by confirmatory molecular assays in this study. The results of the Liat assay were compared with those of the other molecular assays, and the concordance rate was calculated with 95% confident intervals (CIs), using the Clopper and Pearson method. The sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of the Liat assay with nasal samples were also calculated. SARS-CoV-2 were considered positive when one of molecular assays showed positive on corresponding nasopharyngeal samples. The Wilcoxon signed-rank test was used to compare the median Ct values. The correlations of the Ct values between two molecular assays were assessed using Pearson's productmoment correlation coefficient. The statistical analyses were performed using R version 3.14.1 (R Foundation for Statistical Computing, Vienna, Austria), and the figures were created using Python version 3.8.12 (Python Software Foundation, Wilmington, DE, USA). All codes and dataset used for the analyses are available in the Online Supplementary Material (OSM). p values < 0.05 were considered statistically significant. The study included a total of 843 participants, of whom 29 provided only nasopharyngeal samples. Therefore, we obtained 814 nasal and 843 nasopharyngeal samples. The Liat assay could not be performed on one nasopharyngeal sample due to an insufficient amount of residual sample (case ID: T0725). Supplementary Table 1 (OSM) presents data on the symptoms of those who were tested with the Liat assay of nasal samples. Four nasal samples (0.49%) and 23 nasopharyngeal samples (2.7%) exhibited invalid results on the first test of the Liat assay. In the second (repeat) test, all four nasal samples tested negative, one nasopharyngeal sample tested positive, and 13 nasopharyngeal samples tested negative. Due to insufficient residual sample volume, the second test could not be performed on one of the nasopharyngeal samples. The nasopharyngeal sample was excluded from the comparative analysis of the Liat and the NIID assay (case ID: T0613). The remaining eight nasopharyngeal samples tested negative following the third test after being diluted threefold with UTM TM . Finally, we compared the results of 814 nasal and 841 nasopharyngeal samples obtained in the Liat assay to those of the other molecular assays. The Liat assay was positive for SARS-CoV-2 in 113 (13.9%) of the nasal samples and 151 (18.0%) of the nasopharyngeal samples. A nasal sample was positive for influenza virus as detected by the Liat assay. No co-infection cases were identified in this study. The Liat assay with nasal samples showed sensitivity, specificity, PPV, and NPV of 88.6%, 99.4%, 96.5%, and 98.0%, respectively, when compared to the results of molecular assays with their responding nasopharyngeal samples. The results of the Liat assay are compared with the results of the cobas and NIID assays performed using nasal samples in Tables 1 and 2 , respectively. The total, positive, and negative concordance between the Liat and cobas assays were 99.3%, 99.1%, and 99.3%, respectively. The comparison was also performed stratified by the presence of relevant symptoms. In the symptomatic participants, the total, positive, and negative concordance was 99.4%, 100%, and 99.3%, respectively. In the asymptomatic participants, the total, positive, and negative concordance was 99.2%, 97.9%, and 99.3%, respectively. Five nasal samples were positive only on the Liat assay, with Ct values ranging from 31.9 to 37.2. One of the five samples tested positive on additional analyses using Xpert Xpress with the Ct values of 39.6 for the E gene and 37.3 for the N2 gene. When compared to the results of the NIID test, 13 sets of samples showed discordant results, all of which were Liat positive/NIID negative ( Table 2) . Eight of the 13 samples were positive on cobas assay. There were six sets of discordant results between the cobas and Liat assays, of which one sample was Liat-negative/cobas-positive and five samples were Liat positive/cobas negative. The correlation of the Ct values of the Liat assay and the other assays is shown in Fig. 1a- Of the 841 nasopharyngeal samples included in the final analysis, the results of the Liat and NIID assays were positive for SARS-CoV-2 in 151 and 144 samples, respectively, with a concordance of 99.2%. There were seven discordant samples, all of which were Liat positive/NIID negative. The Ct values of the two assays were significantly correlated (p < 0.001, Fig. 1d ). Both nasopharyngeal and nasal samples were obtained from 813 participants. The test results of the Liat assay performed on the two sample types were compared. Both sample types tested positive in 108 samples, 14 pairs of samples tested positive on the nasopharyngeal sample and negative on the nasal sample, and four pairs of samples tested positive on the nasal sample and negative on the nasopharyngeal sample. The correlation of the Ct values between the two sample types is shown in Fig. 2a, b. Table 3 summarizes the results of the discordant cases; the left panel shows nasal samples with positive results on only one of the three molecular assays (Liat, NIID, and cobas) and the right panel shows nasopharyngeal samples with discordant results of the Liat and NIID assay. Additional analyses using Xpert Xpress were performed for these cases, although one sample (case ID: T0198) could not be tested due to the lack of residual sample volume. In nasal samples, five samples were positive only using the Liat assay (Ct value range 31.9-37.2), and one sample was positive only using the cobas assay (Ct value 37.8). Of the five Liat-positive discordant samples, one sample tested positive using the Xpert Xpress assay, and another was positive on the corresponding nasopharyngeal sample from the same participant on both the Liat and the NIID assays. In the nasopharyngeal samples, all seven discordant samples were Liat positive and NIID negative (Ct value range 30.6-36.8). The Xpert Xpress assay results were positive in six of the seven discordant samples. This prospective evaluation showed the high concordance in the nasal samples of the Liat assay results with those of other molecular assays. The Ct values of the Liat assay were significantly lower than those of other molecular assays, but were significantly correlated, although some discordant results were observed. Most of the cases of discordance occurred in samples with Ct values > 30 on the Liat assay. In nasopharyngeal samples, the total, positive, and negative agreement of the results of the Liat and the cobas 6800/8800 assays were 98.6%, 100%, and 97.4%, [8] . Similarly, this study showed high concordance between the results of the Liat assay and other molecular assays in nasal samples. Nevertheless, we observed some invalid results on the Liat assays, which may be caused by the presence of inhibitors in nasal mucosa or discharge. The Liat assays were successfully completed in all samples that initially provided invalid results after they were diluted. The dilution of samples reduces the influence of the inhibitors on molecular assays [20] . Some discordant results were observed in both nasal and nasopharyngeal samples, the majority of which were Liat positive and NIID/cobas negative. The discordant samples generally had Ct values > 30 on the Liat assay, indicating low viral concentrations. The Liat, cobas, and NIID assays have all been shown to have high analytical performance in previous studies [8, 17, 21] . However, the analytical performance on clinical specimens may be different due to the quality of the RNA extraction, the presence of inhibitors, genomic mutations, and stochasticity observed in samples with very low viral concentrations [22, 23] . A previous study comparing Liat and cobas reported that all discordant samples were Liat positive/cobas negative [8] , which is consistent with the results of this study. The Ct values of the Liat assay were strongly correlated with those of the cobas and NIID assays, but were significantly lower. Determining the Ct values is crucial to identify which patients with SARS-CoV-2 infection are most likely to transmit the virus, with higher Ct values indicating lower infectivity [24, 25] . However, Ct values can vary depending on the reagents and equipment used, even if the same samples are tested [25] ; thus, the Ct values provided by each set of equipment should be carefully interpreted. The nasal samples were more likely than nasopharyngeal samples to provide negative results. In this study, 88.2% of nasal samples tested positive on the Liat assay in participants whose nasopharyngeal samples tested positive using the Liat assay. A meta-analysis found that the sensitivity of RT-PCR of nasal samples was 82% compared to RT-PCR of nasopharyngeal samples [26] , which is consistent with our study. The Ct values between both nasal and nasopharyngeal samples were strongly correlated but varied (r = 0.68 and 0.70 for Liat and NIID assay, respectively), despite using the same collection media and assay procedures for both samples. The viral load is generally lower in the nostrils than in the nasopharynx [26] , and previous studies found a similar variance of Ct values between those samples [27, 28] . The procedures for sample collection, the difference in viral dynamics between participants, and the conditions of samples may also have caused the fluctuation of their Ct values [25] . Our study has some limitations. First, we were unable to perform cobas testing on nasopharyngeal samples. The cobas assay has a high sensitivity and has been widely used worldwide [29, 30] . The level of discordance may vary depending on the equipment used for comparison. Second, we did not evaluate performance of the assays in samples from individuals infected with SARS-CoV-2 variants with gene mutations. The emergence of new variants could affect the diagnostic performance of the test. Third, the nasal samples used for molecular examination were collected after acquiring those used for antigen testing. The viral load in the nasal samples may have reduced due to the order of the procedure, which may have led to the variations in results obtained in the molecular examination. Finally, we did not use fresh samples for the evaluation of molecular examinations. Although to a miniscule degree, the storage process involving freezing and thawing also reportedly affects the viral load in samples [31] . In conclusion, the results of the Liat assay showed a high concordance with those of the other molecular assays in both nasal and nasopharyngeal samples. The findings of our study suggest that the Liat assay is suitable for use in a variety of clinical situations, primarily where accurate detection of SARS-CoV-2 is necessary. The online version contains supplementary material available at https:// doi. org/ 10. 1007/ s40291-022-00580-8. Coronavirus disease (COVID-19) Weekly Epidemiological Update and Weekly Operational Update Efficacy and safety of COVID-19 vaccines: a systematic review and meta-analysis of randomized clinical trials. Vaccines (Basel) COVID-19) Treatment Guidelines COVID-19 Treatment Guidelines Testing at scale during the COVID-19 pandemic World Health Organization. Recommendations for national SARS-CoV-2 testing strategies and diagnostic capacities Infectious Diseases Society of America In Vitro Diagnostics EUAs -Molecular Diagnostic Tests for SARS-CoV-2 Clinical performance of the point-of-Care cobas Liat for detection of SARS-CoV-2 in 20 minutes: a multicenter study Analytical and clinical evaluation of a point-of-care molecular diagnostic system and its influenza A/B assay for rapid molecular detection of the influenza virus In-house validation of the cobas Liat influenza A/B and RSV assay for use with gargles, sputa and endotracheal secretions Accurate detection of streptococcus pyogenes at the point of care using the cobas Liat Strep A nucleic acid test Diagnostic performance and characteristics of anterior nasal collection for the SARS-CoV-2 antigen test: a prospective study A prospective evaluation of the analytical performance of GENECUBE ® HQ SARS-CoV-2 and GENECUBE ® FLU A/B How to Obtain a Nasopharyngeal Swab Specimen Interim Guidelines for Collecting and Handling of Clinical Specimens for COVID-19 Comparison of two high-throughput reverse transcription-PCR systems for the detection of severe acute respiratory syndrome Coronavirus 2 Development of genetic diagnostic methods for detection for novel Coronavirus 2019(nCoV-2019) in Japan Multicenter evaluation of the Cepheid Xpert Xpress SARS-CoV-2 Test PCR inhibitors-occurrence, properties and removal Comparative performance and cycle threshold values of 10 nucleic acid amplification tests for SARS-CoV-2 on clinical samples Clinical evaluation of the BioFire® Respiratory Panel 2.1 and detection of SARS-CoV-2 Molecular diagnostic technologies for COVID-19: limitations and challenges Viral cultures for COVID-19 infectious potential assessment-a systematic review Viral dynamics and Real-Time RT-PCR Ct values correlation with disease severity in COVID-19 Performance of saliva, oropharyngeal swabs, and nasal swabs for Clinical Performance of the cobas Liat SARS-CoV-2 & Influenza A/B in Nasal Samples SARS-CoV-2 molecular detection: a systematic review and metaanalysis Swabs collected by patients or health care workers for SARS-CoV-2 testing Comparing nasopharyngeal and mid-turbinate nasal swab testing for the identification of SARS-CoV-2 Clinical evaluation of the cobas SARS-CoV-2 test and a diagnostic platform switch during 48 hours in the midst of the COVID-19 pandemic Roche's cobas SARS-CoV-2 Test to detect novel coronavirus receives FDA Emergency Use Authorization and is available in markets accepting the CE mark Prolonged unfrozen storage and repeated freeze-thawing of SARS-CoV-2 Patient samples have minor effects on SARS-CoV-2 detectability by RT-PCR Acknowledgements We thank Yoko Ueda, Mio Matsumoto, Masaomi Matsubayashi, Yumiko Tanaka, Mika Yaguchi, Naoko Minagawa, and the staff at the Tsukuba Medical Center Hospital for their intensive support in this study. The author Yuki Adachi and the technical support staff at Roche Diagnostics laboratory performed the reference RT-PCR testing. The authenticity of the data obtained can be guaranteed by the fact that all equipment used for PCR examinations were fully automated and that authors Yusaku Akashi and H. Suzuki individually validated appropriate data handling under the supervision of the University of Tsukuba. All dataset and codes used in this study are available in the Online Supplementary Material. Yusaku Akashi 1,2,3 · Michiko Horie 4 · Junichi Kiyotaki 5 · Yuto Takeuchi 1,6 · Kenichi Togashi 4 · Yuki Adachi 7 · Atsuo Ueda 8 · Shigeyuki Notake 8 · Koji Nakamura 8 · Norihiko Terada 6 · Yoko Kurihara 6 · Yoshihiko Kiyasu