key: cord-0713026-4lu1n0s7 authors: Gorzalski, Andrew J.; Tian, Honglin; Laverdure, Chris; Morzunov, Sergey; Verma, Subhash C.; VanHooser, Stephanie; Pandori, Mark W. title: High-Throughput Transcription-mediated amplification on the Hologic Panther is a highly sensitive method of detection for SARS-CoV-2 date: 2020-06-10 journal: J Clin Virol DOI: 10.1016/j.jcv.2020.104501 sha: f9d899add666be69dec68296efdef3dadf6d425c doc_id: 713026 cord_uid: 4lu1n0s7 BACKGROUND: As the demand for laboratory testing for SARS-CoV-2 increases, additional varieties of testing methodologies are being considered. While real time polymerase chain reaction (RT-PCR) has performed as the main method for virus detection, other methods are becoming available, including transcription mediated amplification (TMA). The Hologic Aptima SARS-CoV-2 Assay utilizes TMA as a target amplification mechanism, and it has only recently received Emergency Use Authorization (EUA) by the Food and Drug Administration (FDA). OBJECTIVES: We sought to compare the sensitivity and specificity of the Aptima SARS-CoV-2 Assay to RTPCR as a means of SARS-CoV-2 detection in a diagnostic setting. STUDY DESIGN: We performed a limit-of-detection study (LoD) to assess the analytical sensitivity of TMA and RT-PCR. This preceded a comparison of the methods using previously evaluated clinical specimens (nasopharyngeal swabs) using 116 human specimens tested by both methodologies. Specimens included sixty-one (61) specimens found reactive by real-time PCR, fifty-one (51) found non-reactive, and four (4) deemed inconclusive. RESULTS: The Aptima SARS-CoV-2 Assay showed a markedly higher analytical sensitivity than RT-PCR by LoD study. Evaluation of clinical specimens resulted in fewer inconclusive results by the SARS-CoV-2 assay, leading to potentially higher clinical sensitivity. CONCLUSIONS: The higher analytical sensitivity may explain the assay’s ability to ascertain for the presence of SARS-CoV-2 genome in human specimens deemed inconclusive by real-time PCR. TMA provides an effective, highly sensitive means of detection of SARS-CoV-2 in nasopharyngeal specimens. The foundation of COVID-19 diagnostic testing has included direct detection of viral genomes through nucleic acid amplification methods. For tests that have been utilized under an Emergency Use Authorization by the Food and Drug Administration (FDA), the sole mechanism of amplification used in such tests to date of this manuscript has been real-time PCR (RT-PCR). Assessments of the sensitivities of these tests have been generated and provided to the scientific community, including comparisons of CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel, the Hologic Panther Fusion and the Diasorin Simplexa, among others (1) (2) (3) (4) . Tests that utilize mechanisms of amplification other than the polymerase chain reaction are now becoming available. Among them is the Hologic Panther (non-Fusion) Aptima platform which utilizes transcription mediated amplification (TMA) to amplify the number of target genomes. The Hologic Panther is not uncommonly located in public health and clinical laboratories. This is primarily due to its ease-of-use and high-throughput capability. At the time of preparation of this manuscript, the Hologic Panther is located in 54% of public health laboratories, making it a strong candidate for contributing to the high-throughput testing needs of states and counties seeking to reopen their businesses and institutes (5) . Additionally, while throughput of testing is a key feature that is sought in detection assays for SARS-CoV-2, sensitivity is likely also a very significant trait. Reports of false negative results from real time PCR testing indicate that even the high analytical sensitivity of RT-PCR may be challenged by the pathology of COVID-19 (6, 7) . For this reason, any new tool in the diagnostic arsenal should be assessed for its ability to detect sensitively this virus. Herein, we present data from a comparison study between RT-PCR and TMA as employed by the Hologic Aptima SARS-CoV-2 Assay. This study includes nasopharyngeal swabs taken from individuals tested for SARS-CoV-2 either in a medical or public health screening setting. Additionally, an assessment of analytical sensitivity and limit-of-detection was performed using a quantitated solution of viral genome diluted in transport matrices. specificity relative to one another, with 100% concordance through 20 specimens. Real-time specimens may be "reactive", "non-reactive" and "inconclusive". Inconclusive results by the utilized RT-PCR assays are caused when only one of two or three genomic targets shows amplification for a particular specimen. The cause of this is due either to the presence of very low numbers of viral genomic targets for detection (2) , or false, low-level reactivity for a singular primer/probe set. Patient specimens were stored at -80°C for 1-5 weeks prior to assessment by transcription-mediated amplification. Transcription-mediated amplification (TMA) was performed on the Hologic Panther using RUO versions of the SARS-CoV-2 detection assay and included the following steps: 500 µl of VTM was added to 710 µl of Hologic lysis tube solution. Such specimens were loaded onto the Hologic Panther and tested by programmed protocol, which includes the analysis of 360 µl of lysed specimen. Specimens resulted by the Panther are recorded as "positive", "negative" or "Invalid" ("invalid" did not occur during the course of this study). With consideration of inconclusive specimens: true positivity is ascertained by reactivity by two orthogonal molecular tests (in this case, TMA and RT-PCR). As a result, two of the evaluated inconclusive specimens in this study are deemed true positives and two are considered true negatives. Analytical sensitivity / limit-of-detection analysis was performed using quantitated genomic RNA from SARS-CoV-2, Isolate USA-WA1/2020 (BEI Resources, Manassas, VA). The stock solution of 5.5X10e7 genome equivalents/ml was diluted in 10-fold series in viral transport media or Aptima Multitest Collection fluid. Dilutions were tested in replicates of five. To assess differences in analytical sensitivity between real-time PCR and TMA, we performed a limit-ofdetection (LoD) study by creating a dilution series of purified / quantified SARS-CoV-2 genomic material either in VTM or APTIMA collection matrix. Specimens at each dilution were tested in replicates of five each by either detection method. As shown in Table 1 Noting the sensitivity difference demonstrated by the LoD study, we sought to assess the performance of TMA on specimens previously tested by RT-PCR for SARS-CoV-2. One hundred and sixteen (116) specimens which were nasopharyngeal swabs, and for which sufficient volume existed were selected for testing. This included specimens that were tested by real-time PCR that were: reactive (51), nonreactive (61) and inconclusive (4) . Reactive specimens, as evaluated by RT-PCR for the N1 gene had a Ct for real-time PCR. Percent agreement for specimens found non-reactive ("negative" by TMA) by realtime PCR was 100% and was 98% for those found reactive ("positive" by TMA). The Hologic Panther SARS-CoV-2 transcription mediated amplification test showed higher analytical sensitivity when compared to real time PCR for the detection of SARS-CoV-2. It additionally provided fewer inconclusive results when evaluating human specimens -perhaps as a result of enhanced J o u r n a l P r e -p r o o f sensitivity. In the assessment of clinical specimens, there was one specimen found to be reactive by RT-PCR and non-reactive by TMA. This result is surprising considering the observed higher analytical sensitivity of TMA. We hypothesize that the freezing and thawing, and perhaps storage of specimens may have had a detrimental effect on the specimen, prior to TMA testing. It is notable as well that specimens (collected in VTM) were subject to some dilution prior to testing on the Hologic Aptima system. Using manufacturer's recommendations, 500 µl of specimen material was first added to 710 µl of lysis buffer prior to testing. This means that the sensitivity of the Aptima test may have been underestimated in the testing of human specimens. The finding of two specimens deemed inconclusive by RT-PCR but positive by TMA match the findings of higher analytical sensitivity. A differential in sensitivities between the two mechanisms is not surprising. Previous studies have shown higher sensitivities associated with TMA based assays compared to RT-PCR for viral detection (8, 9) . Specimens subjected to TMA include a process where 360 µl of collected specimen (transport medium) enter the detection process and is tested. For real time PCR, while 200 µl enters the testing process (extraction), only 5 or 10 µl are subjected to PCR. Moreover, the mechanisms of amplification are very different. PCR involves a doubling of target nucleic acid in each of 40 cycles while transcriptionmediated amplification includes the generation of potentially thousands of transcribed copies of target which can each be subsequently turned into transcriptional templates. An additional benefit of the Panther (TMA) platform, not quantified in this study, was its ease-of-use relative to RT-PCR. Samples need only be loaded by random access to the Panther device, which extracts and amplifies/detects in a fully automated fashion. RT-PCR can only be effectively utilized in a batch testing system, which includes extensive extraction prior to detection. This ability to perform high-throughput testing, on a sensitive molecular detection platform has potential to make a significant contribution to the COVID-19 pandemic, where such high demands exist for testing. The Hologic Panther SARS-CoV-2 assay appears to have these traits. Gorzalski: establishment of RT-PCR assays and test performance Chris Laverdure: test performance, limit of detection Honglin Tian: TMA, limit of detection Stephanie VanHooser: coordination of project Sergey Morzunov: laboratory support Subhash Verma: review, analysis Mark Pandori: Conceptualization, data curation, formal analysis, project administration, writing Rapid and sensitive detection of SARS-CoV-2 RNA using the Simplexa™ COVID-19 direct assay US CDC real-time reverse transcription PCR panel for detection of severe acute respiratory syndrome coronavirus 2. Emerg Infect Dis An assessment of real-time RT-PCR kits for SARS-CoV-2 detection Comparison of Four Molecular In Vitro Diagnostic Assays for the Detection of SARS-CoV-2 in Nasopharyngeal Specimens COVID-19 Public Health Laboratory Capacity and Capabilities Weekly Data Report | Week False-negative of RT-PCR and prolonged nucleic acid conversion in COVID-19: Rather than recurrence Novel Coronavirus Outbreak Research Team.Epidemiologic Features and Clinical Course of Patients Infected With SARS-CoV-2in Singapore Analytical characteristics and comparative evaluation of Aptima HIV-1 Quant Dx assay with Ampliprep/COBAS TaqMan HIV-1 test v2.0 Comparison of transcription mediated amplification (TMA) and reverse transcription polymerase chain reaction (RT-PCR) for detection of hepatitis C virus RNA in liver tissue