key: cord-0712886-j4675p96 authors: Salomé, Arthur; Stoclin, Annabelle; Motamed, Cyrus; Sitbon, Philippe; Bourgain, Jean-Louis title: End-Tidal Carbon Dioxide Pressure Measurement after Prolonged Inspiratory Time Gives a Good Estimation of the Arterial Carbon Dioxide Pressure in Mechanically Ventilated Patients date: 2021-11-27 journal: Diagnostics (Basel) DOI: 10.3390/diagnostics11122219 sha: 819d93ca4f245a2d74cdcbd06ac9f8e066846bea doc_id: 712886 cord_uid: j4675p96 Background: End-tidal carbon dioxide pressure (PetCO(2)) is unreliable for monitoring PaCO(2) in several conditions because of the unpredictable value of the PaCO(2)–PetCO(2) gradient. We hypothesised that increasing both the end-inspiratory pause and the expiratory time would reduce this gradient in patients ventilated for COVID-19 with Acute Respiratory Distress Syndrome and in patients anaesthetised for surgery. Methods: On the occasion of an arterial blood gas sample, an extension in inspiratory pause was carried out either by recruitment manoeuvre or by extending the end-inspiratory pause to 10 s. The end-expired PCO(2) was measured (expiratory time: 4 s) after this manoeuvre (PACO(2)) in comparison with the PetCO(2) measured by the monitor. We analysed 67 Δ(a-et)CO(2), Δ(a-A)CO(2) pairs for 7 patients in the COVID group and for 27 patients in the anaesthesia group. Results are expressed as mean ± standard deviation. Results: Prolongation of the inspiratory pause significantly reduced PaCO(2)–PetCO(2) gradients from 11 ± 5.7 and 5.7 ± 3.4 mm Hg (p < 0.001) to PaCO(2)–PACO(2) gradients of −1.2 ± 3.3 (p = 0.043) and −1.9 ± 3.3 mm Hg (p < 0.003) in the COVID and anaesthesia groups, respectively. In the COVID group, PACO(2) showed the lowest dispersion (−7 to +6 mm Hg) and better correlation with PaCO(2) (R(2) = 0.92). The PACO(2) had a sensitivity of 0.81 and a specificity of 0.93 for identifying hypercapnic patients (PaCO(2) > 50 mm Hg). Conclusions: Measuring end-tidal PCO(2) after prolonged inspiratory time reduced the PaCO(2)–PetCO(2) gradient to the point of obtaining values close to PaCO(2). This measure identified hypercapnic patients in both intensive care and during anaesthesia. Lung involvement during SARS-CoV-2 infection frequently dominates the clinical picture of severe forms with both interstitial pneumonia related to the release of proinflammatory cytokines and vascular obstruction of thromboembolic origin [1] . According to international recommendations for Acute Respiratory Distress Syndrome (ARDS) management, protective ventilation is the rule. This practice often induces hypercapnia and requires regular checks of arterial CO 2 pressure (PaCO 2 ) to adjust the level of ventilation. In healthy anaesthetised patients, the monitoring of end-tidal CO 2 pressure (PetCO 2 ) is mandatory even if it underestimates the PaCO 2 , with a gradient around 4.5-13 mm Hg [2] . Lung diseases are associated with pulmonary heterogeneity, and the PaCO 2 -PetCO 2 gradient rises in unpredictable proportions [3] [4] [5] [6] [7] [8] . Therefore, the PetCO 2 value cannot be taken as a good estimate of alveolar PCO 2 and hence cannot be used as an estimate of PaCO 2 . In previous studies in patients with acute lung injury, longer inspiratory time without a change in respiratory rate enhanced CO 2 exchange [8] . During controlled ventilation in adults, the "expiratory plateau" is usually flat, but several factors, such as age and pulmonary disease, are associated with an increase in the slope of the "expiratory plateau". A prolonged expiration manoeuvre improves the prediction of PaCO 2 from end-tidal PCO 2 [9] . Our hypotheses were that end-tidal PCO 2 measured immediately after an inspiratory time (PACO 2 ) prolonged by an inspiratory pause or recruitment manoeuvre could improve the estimation of PaCO 2 and identify severe hypercapnia during mechanical ventilation in COVID ARDS patients or in patients anaesthetised for major surgery. This study complied with the National Medical Ethics Regulation and was approved by the institution's Clinical Research Commission on May 2020. Patients or relatives received information that their anonymous data would be used for research purposes and did not object. This study was prospective and observational, involving routine care of a unicentric cohort, and conducted at Gustave Roussy, Villejuif, France, in both the ICU and operating theatre. Included were patients with a COVID-19 diagnosis hospitalised for ARDS (COVID group), according to the Berlin definition [10] , and patients anaesthetised for major cancer surgery requiring invasive arterial monitoring (anaesthesia group). Patients under protective mechanical ventilation were deeply anaesthetised, with or without muscle relaxation. Increased inspiratory time was used in routine care in two situations: recruitment manoeuvre or prolongation of end-inspiratory pause to measure static compliance. In both groups, measurements were performed only if a bad tolerance of apnoea or recruitment was not predicted because of haemodynamic conditions, and if it would not hinder the surgeon. In both groups, a 10 s (arbitrarily fixed) period of apnoea was obtained using either the end-inspiratory pause function for static compliance evaluation or an automatic recruitment manoeuvre on the intensive care respirator (Evita) or Perseus anaesthesia respirator. With the Zeus anaesthesia respirator, the manoeuvre consisted of sustained manual inflation of the anaesthesia reservoir bag to a peak inspiratory pressure of 30 cm H 2 O for 10 s. The fraction of inspired oxygen concentration (FiO 2 ) was not changed during the procedures. Expiration time was set to 4 s by adjusting the respiratory rate to 10 cycles per minute, and an I:E ratio around 1:2, to obtain an expiratory plateau on CO 2 recording. PCO 2 was measured at the end of the 4 s expiration following the inspiratory pause or recruitment manoeuvre (PACO 2 ). The same investigator carried out all measurements, using a side-stream CO 2 sensor, to ensure their comparability. An arterial blood gas measurement was collected prior to each measurement of PACO 2 , to calculate the PaCO 2 -PetCO 2 and the PaCO 2 -PACO 2 gradients and the PaO 2 /FiO 2 ratio. All patients already needed arterial catheters: none were placed expressly for the study. As the patients' respiratory and haemodynamic status varied over time, several measurements could be conducted for the same patient. FiO 2 , basal EtCO 2 , tidal volume (Vt), respiratory rate (RR), inspiratory plateau pressure (Pplate), settled positive end expiratory pressure (PEEP), central temperature, and when available, level of muscle relaxation (T4/T1), haemodynamic profile including the heart rate (bpm), mean arterial pressure (mm Hg), and norepinephrine dose (milligrams per hour), were recorded for each measurement. Data are presented as mean values ± standard deviations. PaCO 2 values were temperature-corrected, according to the Siggaard-Andersen equation [11] . The normal distributions of PetCO 2 , PACO 2 , and PaCO 2 data were assessed using the Q-Q plot method. When PACO 2 was available from both the inspiratory pause and recruitment, the first measurement was considered for gradient analyses. Indeed, a residual effect of the first recruitment may bias the second measurement. A paired, two-tailed Student's t-test was used to compare the different values of PCO 2 . Correlation coefficients were obtained using the Pearson method to assess the correlation between pairs of PaCO 2 , PetCO 2 , and PACO 2 . PaCO 2 -PetCO 2 , PaCO 2 -PACO 2 , and PACO 2 -PetCO 2 gradient data were compared using a two-tailed unpaired t-test. Bland-Altman plots were used to test for paired PetCO 2 , PACO 2 , and PaCO 2 agreement and reported with the 95% and 99% confidence intervals, for the separate and combined groups. The statistical performance of PACO 2 to detect normocapnia and hypercapnia was calculated for different thresholds of PACO 2 . It was calculated for the separate and combined groups. All statistical results were considered significant if the p-value was less than 0.05. All calculations were performed using Microsoft Excel data processing software (2019 version). From 8 April to 28 May 2020, 67 PACO 2 were obtained for 34 patients (35 measures in 7 patients in the COVID group; 32 measures in 27 patients in the anaesthesia group). Table 1 shows demographic data and baseline characteristics of the study day. Data were similar, except for PaCO 2 , with more hypercapnia in the COVID group (21 vs. 10), as expected. The COVID group presented worse pulmonary characteristics, with a lower compliance (24 vs. 71 mL cm H 2 O -1 in the anaesthesia group), and a lower PaO 2 /FiO 2 ratio (264 vs. 337 mm Hg). In the COVID group, only 29 basal PetCO 2 values were available, due to an initial protocol breach (6 missing data for 2 patients). No PaCO 2 and no PACO 2 values were missing. Five patients were ventilated using an Evita intensive care respirator (corresponding to twelve measurements), and two patients by use of a Perseus anaesthesia respirator, due to a shortage of Evita respirators in the COVID context. In the anaesthesia group, no PetCO 2 , PACO 2 , or PaCO 2 values were missing. PetCO 2 underestimated PaCO 2 in both groups, with a mean level of ∆(a-et)CO 2 of +11 ± 5.7 mm Hg (95% CI 8.8-13) and +5.7 ± 3.4 mm Hg (4.4-6.9) in the COVID and anaesthesia group, respectively (p < 0.001 for both). Increased inspiratory pause or recruitment manoeuvres significantly reduced the gradient to −1.2 ± 3.3 mm Hg (−2.3 to −0.05) (p = 0.043) in the COVID group, and −1.9 ± 3.3 mm Hg (−3.1 to −0.07) (p < 0.003) in the anaesthetised patients, with no significant difference between the two groups. Table 2 . Both PetCO 2 and PACO 2 showed statistically significant linear correlation with PaCO 2 for the two groups. In both the separate groups and combined groups, increased inspiratory pause or recruitment manoeuvres improved this correlation. COVID-PACO 2 showed the best correlation (R 2 = 0.92). The Bland-Altman diagram for ∆(a-et)CO 2 confirmed the poor estimation of PaCO 2 by PetCO 2 . Considering all patient data, all but two values were within a large 95% confidence interval (−2 to +19), and the gradient increased with PaCO 2 , with great imprecision beyond 50 mm Hg (Figure 1 ). Increased inspiratory pause or recruitment manoeuvre induced a decrease in ∆(a-A)CO 2 , with a lower dispersion from −8 to +8 mm Hg. All but one value (from the anaesthesia group) were in the 99% confidence interval around the mean of the differences. These findings were verified when considering the separate groups ( Figure 2 ). Five measurements in the COVID group had a gradient higher than +2 mm Hg associated with suspected respiratory movements identified on the respiratory flow curves. Keeping only the measurements on apnoeic patients, the average and 99% con- Five measurements in the COVID group had a gradient higher than +2 mm Hg associated with suspected respiratory movements identified on the respiratory flow curves. Keeping only the measurements on apnoeic patients, the average and 99% confidence interval of ∆(a-A)CO 2 decreased slightly (mean −2.1 ± 2.5 mm Hg (−3.1 to −1.0)). All patients with a PACO 2 greater than 50 mm Hg were hypercapnic, with a PaCO 2 higher than 45 mm Hg. For both COVID and anaesthetised patients, the PACO 2 threshold of 48 mm Hg yielded the best performance in identifying hypercapnic patients, with Sensitivity ( Measuring PCO 2 at the end of expiration following a recruitment manoeuvre or an inspiratory pause significantly decreased the gradient ∆(a-et)CO 2 in both COVID and anaesthetised patients. The correlation between PACO 2 and PaCO 2 was better than the correlation between PetCO 2 and PaCO 2 , especially in the COVID group. The measurement of PACO 2 after inspiratory pause is completely non-invasive and does not require the patient to disconnect. It nevertheless justifies hygiene precautions such as those commonly observed during the management of suspected, probable, and confirmed cases of COVID-19 [12] . In the presence of correct measurement conditions, the absence of leakage and a respiratory rate < 30 per minute, PetCO 2 allows the estimation of PaCO 2 in healthy subjects in spontaneous ventilation [13] . Under anaesthesia, especially in mechanical ventilation, the ∆(a-et)CO 2 cannot be overlooked. This gradient varies from patient to patient and increases with age, smoking, ASA class, lung disease (especially in cases of pulmonary embolism), and bradycardia. ∆(a-et)CO 2 is not stable during anaesthesia, ranging from 4.5 to 13 mm Hg [2, 14] . In the ARDS, the ∆(a-et)CO 2 tends to increase with lung heterogeneity, as shown by Yousuf and colleagues in 2017: a greater gradient in moderate vs. mild ARDS was reported but no significant difference was found in severe vs. moderate ARDS [15] . A possible explanation could be that severe ARDS was associated with pulmonary hypertension and lower cardiac output, increasing dead-space areas that attenuate the gradient. Many different methods have been used to estimate the PaCO 2 . PetCO 2 presents a weak correlation in healthy lungs [14, [16] [17] [18] [19] [20] , worsening in sick lungs, with no difference in the accuracy using a main-or side-stream sensor [21, 22] . In diseased lungs, results vary greatly between studies, with moderate correlations and large dispersion around the reference PaCO 2 value [8, 15] . While the measurement of transcutaneous PCO 2 yields better results, its use is essentially limited to the paediatric intensive care unit [23] . Inspiratory time influences the PetCO 2 value [24] . Some manoeuvres, such as a simple prolonged exhalation, have been proposed to reduce P(a-et)CO 2 [9] . Out of a series of 16 patients undergoing thoraco-abdominal oesophagectomy, Tavernier and colleagues [9] showed a decrease in the P(a-et)CO 2 gradient after prolonged expiration or prolonged expiration preceded by lung hyperinflation, from 9.8 ± 3.0 to 6.0 ± 3.8 and 4.5 ± 3.8 mm Hg, respectively. The authors concluded, however, that due to extreme inter-individual variability, these manoeuvres did not improve PetCO 2 's assessment of PaCO 2 . However, PetCO 2 can identify the most severe hypercapnia and hypocapnia. A PetCO 2 between 30 and 35 mm Hg most often corresponds to normocapnia (35 to 45 mm Hg). Several pathophysiological mechanisms could be involved to explain the effectiveness of our manoeuvre, as follows. First, maintaining high pressure in the airways for a period of time allows some collapsed alveoli to open, homogenizing the distribution of the ventilation/perfusion ratio. The alveoli recruitment mechanism was evidenced by the greater decrease in the carbon dioxide gradient in the COVID group, demonstrated by a mean difference ∆(a-A)CO 2 − ∆(a-et)CO 2 of −12.2 vs. −7.5 mm Hg in the anaesthesia group. Several studies have shown that prolongation of the inspiratory time decreases alveolar dead-space [9, 25] . Diffusion of CO 2 is time-dependent, and this prolongation increases the time available for alveolar gas exchange. Second, during apnoea, increased inspiratory time allows homogenisation of PCO 2 in the alveoli and between the alveoli and blood. During prolonged apnoea, alveolar PCO 2 approaches pulmonary venous blood PCO 2 due to the absence of alveolar gas movement. This effect has certainly contributed to the reduction of CO 2 gradients. Under ARDS conditions, tidal volume was set to 6 mL/kg with a high respiratory rate to maintain alveolar ventilation and reduced expiratory time. This increased the consequences of inequalities in regional respiratory time constants. Hence, the PetCO 2 value depended on short-and long-time constant alveoli gas mix. PACO 2 tends toward the central venous PCO 2 due to reduced expired volume of poorly ventilated alveoli. Setting the respiratory rate to 10 cycles per minute prior to each PACO 2 measurement improved the expiration of long-time constant alveoli. The recruitment of high CO 2 -concentrated alveoli cumulated with the prolongation of expiration, inducing an increased expiratory CO 2 peak, with PACO 2 over-estimating the PaCO 2 [26, 27] . Late emptying of well-perfused alveoli with higher CO 2 tensions and better overall ventilation/perfusion matching helped to reduce the gradient between PaCO 2 and PetCO 2 , which became negative in some cases. Negative (a-et) PCO 2 gradients have been reported in infants and children, in pregnant patients, and during exercise [11, 26] . This late mechanism may play an important role. Fletcher and Jonson have studied the Vd/Vt ratio and P(a-et)CO 2 gradient at two levels of Vt (450 and 750 mL) in anaesthetised patients [28] . Increasing Vt and decreasing the respiratory frequency did not change the airway dead-space but decreased the alveolar dead-space ratio. This improvement with increasing Vt was attributed to beneficial effects on gas distribution and diffusion time. At large Vt, the P(a-et)CO 2 gradient decreased from 4.5 to 2.5 mm Hg, with a negative gradient in some patients. It is difficult to distinguish between what amounts to an increase in inspiratory time and expiratory time. Increased inspiratory time improves the distribution of the ventilation/perfusion ratio, and increased expiratory time decreases alveolar dead-space. Third, PetCO 2 and PACO 2 are measured by computer analysis of the highest point reached by the capnogram curve at the end of the expiration. The overall performance (response time) of the capnograph including the sample line may be insufficient when high respiratory frequencies are used [24, 29] . It is possible that some PetCO 2 values may have been underestimated due to insufficient capnograph response time relative to the respiratory rate used. Therefore, the 4 s extended expiration time reduced the expiratory slope and thus reduced the bias in the computer analysis of PetCO 2 . Our study also had some limitations. First, the lack of hypocapnic patients does not allow us to draw conclusions about the capacity of the measurement to detect a PaCO 2 less than 35 mm Hg under such conditions. Second, some measurements were suspected to have been collected under not completely apnoeic conditions. However, the accuracy of the ∆(A-a)CO 2 was relatively unaffected by the inclusion or not of these measures in the Bland-Altman diagram. Initially, the data analysis protocol did not anticipate how to deal with the PACO 2 values obtained using both the inspiratory pause and recruitment methods. We chose to only consider the first of the two measures, hypothesising that the residual effect of the first might bias the interpretation of the second. According to the Berlin definition [10] , there was no severe ARDS in our study, and patients were not haemodynamically unstable. We cannot draw conclusions about the effectiveness of our methods in patients with more severe lung diseases. To conclude, measuring PCO 2 after a prolonged inspiration and expiration improves the estimation of PaCO 2 from PetCO 2 and the identification of hypercapnic patients me-chanically ventilated during anaesthesia and for COVID-related ARDS. Measuring endtidal PCO 2 after prolonged inspiratory time reduced the PaCO 2 -PetCO 2 gradient to the point of obtaining values close to PaCO 2 . This non-invasive measure seems particularly interesting when using small Vt in the protected ventilation. However, further studies are needed to explore its usefulness in other conditions, such as in brain-injured patients, severe ARDS, and/or hypocapnic patients. Informed Consent Statement: Patients consent was waived as no change in current practice was performed. Data Availability Statement: Data available on demand (arthur.salome1995@gmail.com). The authors declare that they have no conflict of interest. Autopsy findings in COVID-19-related deaths: A literature review Variation in PCO 2 between arterial blood and peak expired gas during anesthesia Correlation of arterial PCO 2 and PETCO 2 in prehospital controlled ventilation Development of atelectasis and arterial to end-tidal PCO 2 -difference in a porcine model of pneumoperitoneum Concordance between side-stream end-tidal carbon dioxide and arterial carbon dioxide partial pressure in respiratory service setting End-tidal arterial CO 2 partial pressure gradient in patients with severe hypercapnia undergoing noninvasive ventilation Estimating alveolar dead space from the arterial to end-tidal CO 2 gradient: A modeling analysis Capnography in the Emergency Department: A Review of Uses, Waveforms, and Limitations Can prolonged expiration manoeuvres improve the prediction of arterial PCO 2 from end-tidal PCO 2 ? ARDS Definition Task Force Acute respiratory distress syndrome: The Berlin Definition Arterial to end-tidal Pco 2 difference during exercise in normoxia and severe acute hypoxia: Importance of blood temperature correction Management of the corpse with suspect, probable or confirmed COVID-19 respiratory infection-Italian interim recommendations for personnel potentially exposed to material from corpses, including body fluids, in morgue structures and during autopsy practice Accuracy of capnography in nonintubated surgical patients Arterial to end-tidal carbon dioxide gradient and physiological dead space monitoring during general anaesthesia: Effects of patients' position Establishing a gradient between partial pressure of arterial carbon dioxide and end-tidal carbon dioxide in patients with acute respiratory distress syndrome Accuracy and precision of three different methods to determine Pco 2 (Paco 2 vs. Petco 2 vs. Ptcco 2 ) during interhospital ground transport of critically ill and ventilated adults Estimating Arterial Partial Pressure of Carbon Dioxide in Ventilated Patients: How Valid Are Surrogate Measures? Evaluation of the Microstat TM sublingual PCO 2 monitor in ambulatory patients A comparison of indirect methods for continuous estimation of arterial PCO 2 in men End-inspiratory rebreathing reduces the end-tidal to arterial PCO 2 gradient in mechanically ventilated pigs Mainstream vs. sidestream capnometry for prediction of arterial carbon dioxide tension during supine craniotomy Accuracy of postoperative end-tidal Pco 2 measurements with mainstream and sidestream capnography in non-obese patients and in obese patients with and without obstructive sleep apnea Comparison of arterial CO 2 estimation by end-tidal and transcutaneous CO 2 measurements in intubated children and variability with subject related factors Ventilatory frequency influences accuracy of end-tidal CO 2 measurements. Analysis of seven capnometers An appropriate inspiratory flow pattern can enhance CO 2 exchange, facilitating protective ventilation of healthy lungs Negative arterial to end-tidal CO 2 gradients in children End-tidal carbon dioxide measurement in infants and children during and after general anaesthesia Deadspace and the single breath test for carbon dioxide during anaesthesia and artificial ventilation. Effects of tidal volume and frequency of respiration Respiratory Gas Analysis-Technical Aspects