key: cord-0706335-k6mzg9v1 authors: Stocking, Christopher; de Miguel, Laura; Suteu, Gabriela; Dressel, Alexander; Soricelli, Andrea; Roskos, Martin; Valor, Santiago; Mutschmann, Caren; März, Winfried title: Evaluation of five widely used serological assays for antibodies to SARS-CoV-2 date: 2021-10-29 journal: Diagn Microbiol Infect Dis DOI: 10.1016/j.diagmicrobio.2021.115587 sha: e770cbd7ec1ef229831df81e5a0a0b23b42ee75c doc_id: 706335 cord_uid: k6mzg9v1 BACKGROUND AND AIMS: Reliable diagnostic technologies are pivotal to the fight against COVID-19. While real-time reverse transcription-polymerase chain reaction (rRT-PCR) remains the gold standard, commercial assays for antibodies against (SARS-CoV-2) have emerged. We sought to examine five widely used commercial methods. METHODS: We measured antibodies against SARS-CoV-2 with assays, Abbott-IgG, Roche-IgT (total antibodies, isotype-unspecific), EUROIMMUN-IgG, EUROIMMUN-IgA, DiaSorin-IgG, in 191 serum samples from patients with rRT-PCR proven COVID-19 between days 0 and 47 after the onset of clinical symptoms and in biobank samples collected in 2018. The assays were calibrated using the manufacturers´ instructions; results are in multiples of the assay specific cut-offs (Abbott, Roche, EUROIMMUN) or in arbitrary units (AU/mL, DiaSorin). RESULTS: The assays for IgG and IgT have approximately the same sensitivity and specificity for detecting seroconversion which starts at approximately day 3 after symptom onset, sensitivity reached 93 percent on day 16 and was 100 percent for each assay on day 20. The assay for IgA antibodies was superior in sensitivity and had a lower specificity than the others. Bivariate non-parametric correlation coefficients ranged between 0.738 and 0.991. CONCLUSIONS: Commercial assays for IgG or total antibodies against SARS-CoV-2 are largely equivalent for establishing seroconversion but differ at high antibody titres. Increased sensitivity to detect seroconversion is afforded by including IgA antibodies. Further international efforts to harmonise assays for antibodies against SARS-CoV-2 are urgently needed. Coronavirus disease 2019 has quickly spread worldwide from Wuhan (China) since December 2019 (Zhu et al., 2020) . Reliable diagnostic technologies are pivotal to the fight against COVID-19 (Younes et al., 2020) . While real-time reverse transcription-polymerase chain reaction (rRT-PCR) testing of respiratory specimens remains the gold standard, a series of commercial assays for antibodies against SARS-CoV-2 (Severe Acute Respiratory Syndrome Coronavirus 2) has been launched (Younes et al., 2020) . Their main indications are: to determine the prevalence rate of SARS-CoV-2 infections in epidemiological surveys and regional outbreaks, to re-classify ambiguous results of rRT-PCR, to establish the diagnosis of COVID-19 in acute infections once rRT-PCR results have become negative (Guo et al., 2020 , Kucirka et al., 2020 , Wikramaratna et al., 2020 , to retrospectively infer infection late after onset following the relief of symptoms and to potentially confirm immunity or the efficacy of vaccines. Available tests rely on the SARS-CoV-2 spike (S) protein and/or the nucleocapsid (N) protein (Ou et al., 2020) . Tests detecting IgG, IgA, or IgM isotypes or combinations thereof are available. The world-wide, explosive propagation of COVID-19 had prompted the emergency development and accelerated approval of tests by manufacturers and authorities, however, only a few robust evaluations of available products have been completed (Administration, 2020 , Hong KH, 2020 . The Cochrane Collaboration concludes that the numbers of individuals contributing data within pertinent studies are small and are usually not based on tracking the same groups of patients over time (Deeks et al., 2020) . We therefore sought to examine five widely used, commercial methods for antibodies to SARS CoV-2 in parallel. Most patients selected as part of this study provided at least two samples during follow-up. We selectively collected 191 serum samples from patients who had tested positive by real-time reverse transcription-polymerase chain reaction (rRT-PCR) for SARS-CoV-2 and in whom serologic testing had been ordered. The samples came from the Hospital Universitario La Moraleja and the Hospital Virgen del Mar, both located in Madrid, Spain. We liaised with the responsible physicians to collect information on the onset of clinical signs of COVID-19, symptoms, severity, and course of the disease (Table 1) . A timepoint relative to the onset of symptoms was assigned to each sample based upon the clinical information provided by physicians and patients, with day 0 marking symptom onset. Samples were collected from 66 patients, 59 of the patients had samples collected at more than one timepoint throughout the course of the infection and recovery up to day 47. All results from patient samples were included in this analysis. Analysis of outliers has not been performed nor have any samples been excluded. The imprecisions of the assays were evaluated using at least two levels of quality control material and one positive human serum pool, over six independent batches with six measurements per batch, for a total of 36 measurements for each material. To calculate mean values and standard deviations we used assay readouts after normalizing them against the assay specific cut-off. Specificities were examined in 262 to 337 frozen samples collected before 2018 for which serology of respiratory infections had been requested. The ethics board of the Medical Association of Berlin was informed of the study and confirmed in writing that there was no need for further ethical advice, since no additional analyses were performed other than the medically warranted ones. rRT-PCR of pharyngeal swaps was performed at SYNLAB Barcelona, Spain, using the VIASURE SARS-Cov-2 Real Time PCR Detection Kit. We measured antibodies against SARS-CoV-2 with five laboratory-based immunoassays under the conditions shown in Table 2 . As these methods were not available at a single laboratory site, we distributed aliquots as indicated. Shipment was between 4 to 8°C; the cold chain was not interrupted at any time. All assays were calibrated using the manufacturers´ specified procedures and single measurements were performed. While four of the five methods measure either IgG or IgA, the Elecsys Anti-SARS-CoV-2 (Roche-IgT) is not isotype-specific for IgG; rather, the package insert states that it is a test for the "qualitative detection of antibodies (including IgG) against" SARS-CoV-2. The EUROIMMUN-IgG and the EUROIMMUN-IgA test may be used simultaneously. We therefore also provisionally combined the two tests by assigning an overall "positive" result if one of the two tests was "positive" and treating the remaining results as "negative" analyses (EUROIMMUN-IgG/A). The results of the DiaSorin LIASON SARS-CoV-2 S1/S2 IgG (DiaSorin-IgG) test are in Arbitrary Units per millilitre (AU/mL). For the purpose of this comparison, they have been normalised against the assigned cut-off of 13.5 AU/mL. All other assays provide relative units normalised against the assay specific cut-off. In our analyses, we used those relative units, also for "negative" samples, which then scored below 1. Imprecision. Imprecision (within-batch, between-batch) was calculated using a two-factor analysis of variance and expressed as the coefficients of variation (CV). For this purpose, total variance was decomposed into intra-batch and between-batch variance components from which the intra-batch and between-batch standard deviations were calculated (Supplementary Table1). Bivariate comparisons. Spearman correlation coefficients and the non-parametric regression method of Passing and Bablok (Passing and Bablok, 1983) were performed using MedCalc (Supplementary Table 2 ). Pearson´s correlation coefficients and least squares regressions were not calculated because the CUSUM test (included in MedCalc) detected highly significant deviations from linearity for each of the pairwise comparisons. We also produced scatter diagrams along with restricted cubic splines (with knots at the 5th, 27.5th, 50th, 72.5th and 95th percentiles, respectively, of the values of the abscissa variable), whereby we provisionally plotted the SARS-CoV-2 IgG assay from Abbott Diagnostics on the abscissa ( Figure 1A -D). This does not prejudice the interpretation of the data nor indicate that we consider the Abbott Diagnostics assay as a reference method. In order to examine whether differences between the methods are heterogenous across the concentration range, we generated difference plots according to Bland and Altmann (Bland and Altman, 1986) for descriptive purposes only (Supplementary Time course of antibody concentrations. We plotted medians (and interquartile ranges) against the day of blood sampling after the onset (day 0) of clinical symptoms. If less than 10 samples were available for a specific day, samples from several days were pooled together and the day number displayed then is the mean for the group (Figure 2 ). Sensitivity, specificity. As the concentration of antibodies present are known to rise immediately following the onset of symptoms, the sensitivity of each assay was evaluated relative to the day of onset of symptoms using the same strata as in Figure 2 (Table 3A ) and the cut-offs provided by the manufacturers. For this purpose, we treated "indeterminate" results as "negative". We provisionally defined overall sensitivity as the cumulative sensitivity on day 16, which likely marks the timepoint at which the window for positive rRT-PCR results closes (Younes et al., 2020) . To scrutinize potential differences between the assays, we performed two multivariate logistic regressions using time strata (less than 3 days, 3 to 7 days, 8 to 13 days, 14 days or more) as independent and the test result as the dependent variables. In the first model, we used the five assays separately, in the second one, we replaced the two EUROIMMUN tests by their combination EUROIMMUN-IgG/A (Table 3B) . We also visualized the time-dependency of the assays´ sensitivities by creating monotonic cubic splines with four knots at the centres of time quartiles ( Specificity was expressed as the proportion of subjects in the biobank samples with negative test results in the total of samples tested (Table 3C) . 191 samples were collected from 66 patients. The detailed clinical characteristics of these patients are shown in Table 1 . Their mean age was 68 years with a range of 30 to 97 years. Among the comorbidities, hypertension, dyslipoproteinaemia, diabetes mellitus, chronic obstructive lung disease and ischemic heart disease were prevailing. The mean time interval between the onset of symptoms and hospital admission was five days. As expected, the most frequent clinical features of COVID-19 were pneumonia, cough, shortness of breath, and fever (Table 1) . 57 patients recovered, 9 died. Characteristics of the deceased and the surviving patients are also provided with Table 1 . Supplement Table 1 shows coefficients of variation (CVs) obtained in control sera at low (negative) and high (positive) antibody titres and in an in-house pool of positive sera. At concentrations below the diagnostic cut-off, total CVs were around 10%, with the exception of the EUROIMMUN-IgA assay showing a CV of 48.7%. At high concentrations, the CVs ranged from 1.7 to 9.3% for Abbott-IgG and EUROIMMUN-IgG and between 2.5 to 6.8% for Roche-IgT and EUROIMMUN-IgA. The scatterplots shown in Figure 1 indicate substantial scattering of the results in all pairwise comparisons. The CUSUM test for linearity indicated significant (p < 0.01) deviations from linearity for all comparisons. In the light of the scattering of the data, we dispensed with the calculation of Pearson correlation coefficients and least squares regression lines and visualised the relationships using splines (Figure 1 ). Spearman non-parametric bivariate correlation coefficients were calculated for descriptive purposes and ranged between 0.738 to 0.861 (Supplementary Table 2 and at day 2, and then increased to approach 100 percent at a concentration of 5 and day 6, respectively. The median concentrations obtained with Abbott-IgG, Roche-IgT, EUROIMMUN-IgG and EUROIMMUN-IgA started to rise between days 3 and 5 post onset of symptoms, the median concentrations of the DiaSorin-IgG started to increase between days 5 and 6 ( Figure 2 ). All tests plateaued at day 20 after the onset of symptoms. After this timepoint, Abbott-IgG did not show any change, Roche-IgT showed some fluctuation, EUROIMMUN-IgG a slight decline, EUROIMMUN-IgA a strong decline and DiaSorin-IgG a further increase between day 35 and 45, but this may be due to a smaller number of measurements available for day 45 (n=5, Figure 2 ). We analysed the sensitivities of each of the tests in relation to the day post onset of symptoms. As displayed in Table 3A , the cumulative sensitivity increased along the time axis. Until about day 12 the EUROIMMUN-IgA appeared as the most sensitive, whereas DiaSorin-IgG appeared the least sensitive assay. Sensitivity was then 93 percent across all tests on day 16 post onset of clinical symptoms. One patient became negative after testing positive for one assay. This patient with moderate clinical symptoms had a negative titre on the Abbott-IgG assay at day 37 and day 45 after four positive results from day 10 to 17, however, the maximum titre was only 3.83. All other assays remained positive. Multivariate logistic regression accounting for the time after symptom onset suggests that the EUROIMMUN-IgA is approximately two-fold more sensitive than the other tests, with no material differences between them (Table 3B) . Also, the combined use of the EUROIMMUN-IgA and the EUROIMMUN-IgG was more sensitive than the three remaining tests. This was confirmed by visualization of the assays´ sensitivities as a function of the time after symptom onset using monotonous cubic spline interpolation ( Figure 3 ): The EUROIMMUN-IgA and the EUROIMMUN-IgG/A had a higher sensitivity than the other tests up to 10 days after the onset of symptoms while later the assays´ sensitivities converged. We used biobank samples collected before 2018 to examine specificity. The numbers of samples used for each test and results are displayed with Table 5B . The specificities were 1.00 for the Abbott-IgG and the Roche-IgT test; they ranged from 0.953 through 0.989 to 0.996 for EUROIMMUN-IgA, DiaSorin-IgG and EUROIMMUN-IgG, respectively. Figure 4 shows the PPVs and NPVs according to assumed prevalence rates in the populations between 0 and 0.05 calculated using the sensitivities at day 16 displayed in Table 3A . At an assumed prevalence rate of 0.003 for Germany, the PPV and NPV of the Abbott-IgG and the Roche-IgT were 1.0. Due to the low prevalence rate assumed, the PPVs of the EUROIMMUN-IgA, EUROIMMUN-IgG and the DiaSorin-IgG were 0.41, 0.06 and 0.20, respectively. The NPVs of all tests were 1.00 at the prevalence rate of 0.003. This is a systematic comparison of contemporary commercial assays to measure antibodies against SARS-CoV-2. We made the following key observations: At concentrations way above the decision range and/or 20 days after onset of symptoms our comparison reveals substantial random scattering of results. We need to recognize, however, that only results of the DiaSorin-IgG are claimed to be quantitative by the manufacturer, while EUROIMMUN provides semi-quantitative and Abbott and Roche qualitative results. Anyway, the vast scattering of results at concentrations way above the decision ranges suggests that the methods may not be interchangeable for the long-term follow-up of SARS-CoV-2 antibody titres, at least as long quantitative tests will become widely available. Only few studies have compared isotype-specific (IgM, IgG, and IgA) sensitivities of antibodies for SARS-CoV-2 (Guo et al., 2020, Xiao et al.) . Here, seroconversion started at day 3 after the onset of clinical symptoms (15 to 38 percent of samples positive), reached 50 percent on day 5 (42 to 54) and 93 percent on day 16. On day 20 seroconversion was complete for all assays (Table 5 ). This is in line with other observations (Ai et al., 2020 , Gudbjartsson et al., 2020 , Kabesch et al., 2020 , Okba et al., 2020 , Tan et al., 2020 . In line with previous work (Okba et al., 2020), the EUROIMMUN-IgA a higher sensitivity as the IgG-and IgT-based tests. There was no material difference in sensitivity between the Roche-IgT and those tests capturing IgG only. This is in agreement with the observation that there is only a tiny, if any, time shift between the emergence of specific IgM and IgG in the plasma upon SARS-CoV-2 exposure , as it has also been shown for SARS and MERS. The obvious difference in the detection of early seropositivity between EUROIMMUN-IgA and the other tests may have clinical relevance if one accounts for the documented limitations of rRT-PCR. While rRT-PCR is widely accepted as the gold standard in the diagnostics of acute COVID19 infections it has been argued that the overall sensitivity of PCR is limited to 80% and can be improved by combination with antibody testing (Zhao et al., 2020) . This may be of special importance if the clinical situation is suggestive of COVID-19 while PCR results are negative. While all tests were highly consistent for negative samples obtained early in the course of the infection, there was some inconsistency around the decision cut-offs and the days of seroconversion (Supplementary Figure 2) . The reason for this may lie in the comparatively low reproducibility at low concentrations within the so called "grey zone". It is hence obvious that single results close to the assay cut-offs need to be interpreted cautiously. Also, and consequentially, it might be prudent to reproduce such results by duplicates, in an independent sample obtained a few days later to prove titre changes, or to verify them by another assay and/or a confirmatory test like Western blotting once this approach becomes widely available. Titres produced by Abbott-IgG stayed constant after 20 days, while there was a fluctuation of titres obtained with Roche-IgT (decrease first, increase thereafter), which might relate to an isotype shift. Similarly, EUROIMMUN-IgG showed a moderate and the EUROIMMUN-IgA and a sharp decline of signals beyond day 20, in agreement with Tan et al. (Tan et al., 2020) . These data of course need to be interpreted with a sense of proportion, since they rely on the small number of samples available beyond day 20. However, they illustrate the current uncertainty as to whether and how long antibodies against SARS-CoV-2 would circulate in the serum at detectable concentrations late after an infection. In any case, the findings suggest that the most reliable timepoint to test for SARS-CoV-2 exposure would be no earlier than day 20 after onset of symptoms (World Health Organization. Laboratory testing for 2019 novel coronavirus (2019-nCoV) in suspected human cases), while the possibility exists that samples become negative thereafter. It is completely consistent with this estimate that we attained 100 % sensitivity at day 20. A recent validation study including lateral flow tests and ELISAs reported specificities between 0.84 and 1.0 (Whitman et al., 2020) . As assessed using frozen biobank samples collected from 2018 and earlier, we found two tests producing specificities of 1.00 (Abbott-IgG and Roche IgT). This corresponds to the specificity reported previously for the Roche-IgT (0.998 obtained by analysing 5272 serum samples) (https://www.fda.gov/media/137605/download). The specificities of two other tests came close to unity (EUROIMMUN-IgG and DiaSorin-IgG), whereas the specificity of the EUROIMMUN-IgA was materially lower (0.95). While more true negative samples would be required to draw definite conclusions in regard to the specificities, even a very low rate of false positives would translate into a considerable restriction of the predictive value of a positive test in a low prevalence setting as it is encountered with COVID-19 ( Figure 4 ). If applied to epidemiological studies, even very low false positive rates would hence produce an artefactual overestimation of the disease prevalence rate paired with an underestimation of the case fatality. Tentatively, and until more data for test specificity become available, the Abbott-IgG and the Roche-IgT might in this respect have some advantages over other tests in epidemiological surveys. This may, however, not apply to any clinical situation in which the pre-test probability is higher than in the general population. We studied symptomatic patients only and cannot infer on the strength of the immune response in patients without or with few symptoms. In severe COVID-19 cases seroconversion may occur early (Okba et al., 2020) while mild or asymptomatic infections may even lack seroconversion (Gudbjartsson et al., 2020 , Yongchen et al., 2020 , Zhou et al., 2013 . Approximately, seroconversion was seen in 50 percent of the samples at day five after symptom onset. It is a limitation of the current work, that the timepoint of symptom onset was obtained retrospectively, so that evidently the estimate of the time to seroconversion depends on the precision of the information conveyed by patients to treating physicians. Also, as a consequence of the limited clinical information and the small number of severe COVID-19 cases, we were not able to reliably correlate the kinetics of antibody concentrations with disease severity (Huang et al., 2020 , Okba et al., 2020 , Tan et al., 2020 We also have no knowledge on the long-term persistence of antibodies beyond the time frame of our study which extends to 48 days only after the onset of clinical symptoms. It thus remains open whether any of the tests would be suitable to monitor the immune status in persons presenting later after an infection and/or with mild or no symptoms (Gudbjartsson et al., 2020) . It is also a limitation that lateral flow tests, legions of which have been launched for on-site testing, have not been included into the current comparison. Finally, neutralization tests (Haveri et al., 2020 , van der Heide, 2020 , Wolfel et al., 2020 , Wu et al., 2020 are not currently available to our laboratories. It hence remains outside the scope of this article whether or to what extent the assays examined would reflect resilience against subsequent SARS-CoV-2 infections. Due to the threatening spread of COVID-19 serological tests for COVID-19 have been approved and launched rapidly. Hence, extensive evaluations were skipped for comprehensible reasons. In this respect, the current report represents a snapshot of the status quo awaiting confirmation and extension. In the light of the still low prevalence rate of COVID-19 in many countries, our preliminary data on the specificity of the assays evaluated will need further scrutiny in diverse environments. Further, the long-term course of antibody concentrations will have to be clarified and research into simple assays specifically indicating protection from SARS-CoV-2 infections or immunity after vaccination is warranted (Whitman et al., 2020) . In this regard, surrogate neutralisation assays avoiding the work with viable viruses might be of particular interest (Grzelak L and 10.1101/2020.04.21.20068858) . The scattering of results in our bivariate method comparisons suggests that there is need for reference materials covering the entire concentration range of antibody concentrations and multiple point calibration may be needed to eliminate such differences. Sufficiently large and challengeable validation studies will be needed which will be facilitated by the formation of international research networking involving the major stakeholders like manufacturers, laboratory service providers and clinical centres. All authors except AD are employed with SYNLAB International GmbH or its local subsidiaries. There are no other known conflicts of interest associated with this publication. There was no financial support to SYNLAB or any of its employees from the manufacturers of the assays used in this evaluation and there has been no other financial support for this work that could have influenced its outcome. CM and SV designed the study; UH, LDM, and GS collected data; CS performed statistical analysis and surveyed laboratory analyses; GS surveyed laboratory analyses; WM wrote the manuscript; all authors validated, reviewed and edited the manuscript. The 2 If less than 10 samples were available for a specific day, samples from several days were combined. The displayed day number is then the mean for the group. 3 The Anti-SARS-CoV-2 ELISA (IgG) EUROIMMUN and Anti-SARS-CoV-2 ELISA (IgA) EUROIMMUN were not considered separately. Rather, the combined test was considered positive if any of the tests was positive. 4 There are 6 less samples for LIASON SARS-CoV-2 S1/S2 IgG, the number of samples for this test are in brackets in the second column. Administration, Food and Drug. Policy for Coronavirus Disease-2019 Tests During the Public Health Emergency (Revised) Correlation of Chest CT and RT-PCR Testing in Coronavirus Disease Statistical methods for assessing agreement between two methods of clinical measurement Antibody tests for identification of current and past infection with SARS-CoV-2 SARS-CoV-2 serological analysis of COVID-19 hospitalized patients, pauci576 Spread of SARS-CoV-2 in the Icelandic Population Profiling Early Humoral Response to Diagnose Novel Coronavirus Disease (COVID-19) Serological and molecular findings during SARS-CoV-2 infection: the first case study in Finland Antibody Responses in Patients with COVID-19 Temporal profiles of viral load in posterior oropharyngeal saliva samples and serum antibody responses during infection by SARS-CoV-2: an observational cohort study Neutralizing antibody response in mild COVID-19 Test performance evaluation of SARS-CoV-2 serological assays Estimating false-negative detection rate of SARS-CoV-2 by RT-PCR Virological assessment of hospitalized patients with COVID-2019 World Health Organization. Laboratory testing for 2019 novel coronavirus (2019-nCoV) in suspected human cases Ig Preprint at medRxiv Profile of specific antibodies to SARS-CoV-2: The first report Early Viral Clearance and Antibody Kinetics of COVID-19 Among Asymptomatic Carriers Different longitudinal patterns of nucleic acid and serology testing results based on disease severity of COVID-19 patients Challenges in Laboratory Diagnosis of the Novel Coronavirus SARS-CoV-2 Antibody Responses to SARS-CoV-2 in Patients With Novel Coronavirus Disease First infection by all four non-severe acute respiratory syndrome human coronaviruses takes place during childhood A Novel Coronavirus from Patients with Pneumonia in China