key: cord-0705368-ergdmd8q authors: Stokes, William; Berenger, Byron M.; Singh, Takshveer; Adeghe, Ifueko; Schneider, Angela; Portnoy, Danielle; King, Teagan; Scott, Brittney; Pabbaraju, Kanti; Shokoples, Sandy; Wong, Anita A.; Gill, Kara; Turnbull, LeeAnn; Hu, Jia; Tipples, Graham title: Acceptable performance of the Abbott ID NOW among symptomatic individuals with confirmed COVID-19 date: 2021-07-26 journal: J Med Microbiol DOI: 10.1099/jmm.0.001372 sha: 75d9b852e5a66a84f9d5f37224718305f2889ef9 doc_id: 705368 cord_uid: ergdmd8q INTRODUCTION: The ID NOW is FDA approved for the detection of SARS-CoV-2 in symptomatic individuals within the first 7 days of symptom onset for COVID-19 if tested within 1 h of specimen collection. GAP STATEMENT: Clinical data on the performance of the ID NOW are limited, with many studies varying in their study design and/or having small sample size. AIM: In this study we aimed to determine the clinical performance of the ID NOW compared to conventional RT-PCR testing. METHODOLOGY: Adults with COVID-19 in the community or hospital were recruited into the study. Paired throat swabs were collected, with one throat swab transported immediately in an empty sterile tube to the laboratory for ID NOW testing, and the other transported in universal transport media and tested by an in-house SARS-CoV-2 RT-PCR assay targeting the E gene. RESULTS: In total, 133 individuals were included in the study; 129 samples were positive on either the ID NOW and/or RT-PCR. Assuming any positive result on either assay represents a true positive, positive per cent agreement (PPA) of the ID NOW compared to RT-PCR with 95 % confidence intervals was 89.1 % (82.0–94.1%) and 91.6 % (85.1–95.9%), respectively. When analysing individuals with symptom duration ≤7 days and who had the ID NOW performed within 1 h (n=62), ID NOW PPA increased to 98.2 %. CONCLUSION: Results from the ID NOW were reliable, especially when adhering to the manufacturer’s recommendations for testing. The ID NOW (Abbott) is approved by the United States Food and Drug Administration (FDA) for the detection of severe acute respiratory syndrome-coronavirus-2 (SARS-CoV-2) in individuals who are within the first 7 days of symptom onset. The ID NOW assay uses isothermal nucleic acid amplification of a region of the viral RNA-dependent RNA polymerase (RdRp) to detect the presence of SARS-CoV-2, with results available in under 15 min. Clinical specimens approved for testing include nasal, throat and nasopharyngeal swabs, which must be tested on the Abbott ID NOW either immediately or within 1 h of collection. Specimens placed in viral/universal transport media (UTM) are not valid for testing by the Abbott ID NOW [1] . Current limitations of the Abbott ID NOW include the paucity of strong data to determine its effectiveness in detecting SARS-CoV-2 in clinical settings. The studies used to obtain FDA approval were in vitro. These studies demonstrated that the limit of detection of the Abbott ID NOW is similar to other nucleic acid amplification tests at approximately 125 genome equivalents per millilitre. Of the clinical studies reported in the literature, the Abbott ID NOW has been shown to have excellent specificity (~100 %) but sensitivity varies widely between studies (48.0-94.1 %) [2] [3] [4] [5] [6] [7] [8] [9] [10] [11] [12] [13] [14] [15] [16] [17] . In addition, many of these studies vary in their study design such as comparing nasopharyngeal to nasal specimens or having major delays in testing specimens on the ID NOW. Some studies were also conducted prior to Abbott's updated guidance on ID NOW specimen transportation that recommended against using UTM [3] . Only two studies adhered to the FDA's recommendations for ID NOW specimen collection and transportation, and both studies had a small sample size (<20 samples positive for SARS-CoV-2) [16, 17] . We sought to assess the positive per cent agreement (PPA) of the ID NOW by comparing its performance to an in-house validated real-time reverse transcriptase PCR (RT-PCR) among individuals with recently confirmed COVID-19 while adhering as closely as possible to the manufacturer's recommendations. We also tested the accuracy of the ID NOW with samples taken from asymptomatic individuals at low risk for COVID-19 (i.e. no exposures), and on retrospective clinical samples previously positive for common respiratory pathogens. Community and hospitalized individuals within the Calgary and Edmonton Health Zones of Alberta, Canada, who recently tested positive for SARS-CoV-2 at Alberta Precision Laboratories (APL) and confirmed as cases by Alberta Health Services (AHS) Public Health were recruited. Diagnostic testing was performed by a Health Canada-approved SARS-CoV-2 assay or a laboratory-developed real-time RT-PCR assay (see below for details). Participants were identified by an AHS Public Health confirmed case list. Oral consent by phone was obtained for collection of samples in the participant's home or in a hospital (if hospitalized). Individuals under the age of 18 years and individuals in supportive or congregate living facilities were excluded. Individuals who lived farther than a 30 min drive from the laboratory were also excluded. Eligible patients who consented to the study were recruited to have two throat swabs collected by trained healthcare professionals within their homes or inpatient unit. Individuals were asked to confirm their symptoms and date of symptom onset at the time of study swab collection. Healthcare workers performing the collection were given instructions on how to perform throat swabs using the Clas-siqSwabs (COPAN Diagnostics) and the throat swab provided in the ID NOW testing kits (Abbott) [18] . Throat swabs were collected from both sides of the oropharynx and the posterior pharyngeal wall under the uvula. Throat swabs were collected approximately 1 min apart, and the order in which throat swabs were collected was recorded. For each paired throat swab, one was placed into a dry 15 ml conical centrifuge tube (Fisher Scientific) for ID NOW testing and the other into a tube containing UTM (COPAN) for RT-PCR testing. After testing one household, samples were transported to the APL Public Health Laboratory as quickly as possible, at room temperature, and tested upon receipt. Testing on the ID NOW instrument was done immediately upon receipt as per the manufacturer's instructions. Throat swabs in UTM collected for RT-PCR testing were stored at 4 °C and tested within 72 h. Two hundred microlitres of UTM was extracted on the MagMAX Express-96 or Kingfisher Flex (ABI) using the MagMAX-96 Viral RNA Isolation Kit (Ther-moFisher) or the PurePrep Pathogen Kit (MolGen) according to the manufacturers' instructions, and eluted into a volume of 110 µl. RT-PCR testing included an assay targeting the envelope (E) gene of SARS-CoV-2, developed and validated at APL, and the Cobas SARS-CoV-2 (Roche Diagnostics) test on the Cobas 6800 instrument (dilution studies compared to other RT-PCR platforms provided in Table S1 within the supplementary material) [19] . For our lab-developed test, the samples were considered positive for SARS-CoV-2 when the cycle threshold (Ct) value was <35. If the Ct was ≥35, amplification from the same eluate was repeated in duplicate and was considered positive if at least 2/3 results had a Ct <41. Testing for SARS-CoV-2 on the Cobas 6800 instrument was performed according to the manufacturer's instructions. For the Cobas SARS-CoV-2 test, a positive result was defined as 2/2 targets positive or one or more targets positive in duplicate. If 1/2 targets were positive and duplicate testing was negative, the result was considered indeterminate. For discrepant results, the specimens were retested in triplicate with our lab-developed RT-PCR test and in triplicate with the N2 assay from the CDC 2019-Novel Coronavirus (2019-nCoV) Real-Time RT-PCR Diagnostic Panel using the UltraPlex 1-Step Toughmix (Quantabio) [20] . If still negative, the test was run on the Cobas 6800. PPA was calculated with Clopper-Pearson 95 % confidence intervals. Statistical analysis was performed using a Pearson Chi-squared test for categorical variables and t-test for continuous variables using STATA (version 14.1). Two throat swabs were collected from asymptomatic individuals at low risk of having COVID-19 (no recent travel, no exposures). One throat swab was tested immediately (<2 min) on the ID NOW instrument. The other throat swab was tested by RT-PCR as explained above. To assess for cross-reactivity, retrospective samples containing high concentrations of various respiratory viruses, stored in UTM, were tested by aliquoting 400 µl of sample into the ID NOW blue specimen receiver. These samples were previously tested by either the NxTAG Respiratory Pathogen Panel (Luminex) or the CDC influenza SARS-CoV-2 multiplex assays. The ability of the ID NOW to process this volume of UTM was confirmed by testing four retrospective positive SARS-CoV-2 samples, in UTM, and showing that they could be detected (data not shown, Ct values ranging from 21 to 28). In total, 152 patients were recruited for this study. Fourteen individuals were asymptomatic at the time of COVID-19 diagnosis and at time of study collection and were therefore excluded from analysis (sub-analysis provided in the supplementary material). The date of symptom onset for two individuals was not captured at time of consent and they could not be reached later for clarification, and therefore were excluded. Three samples were excluded as samples were lost for RT-PCR testing (all were ID NOW-positive). Symptom details were not recorded for two individuals but they were still included in the analysis as symptom onset was known and they were still symptomatic at the time of collection. This resulted in the inclusion of 133 individuals in our analysis, and their characteristics are provided in Table 1 . Mean duration of symptoms at the time of collection was 6.9 days (median=7, range=1-17 days). Seventy-five per cent (n=100) of individuals were within the 7 day symptom onset window, 62.4 % (n=83) of individuals had their samples tested on the ID NOW within 1 h of collection, and 46.6 % (N=62) met the regulatory agency approved criteria for testing on the ID NOW (symptom onset ≤7 days and ID NOW test conducted within 1 h of collection). The mean E gene Ct value for positive results from the RT-PCR assay was 30.9 (median=31.0, range=16.4-37.9). Of the 133 samples, 96 were positive on both the ID NOW and the RT-PCR (Table 2) . Assuming any positive result represents a true positive, the PPA of the ID NOW compared to RT-PCR with 95 % confidence intervals was 89.1 % (82.0-94.1%) and 91.6 % (85.1-95.9%), respectively. There were 13 false negatives on the ID NOW instrument, with 11/13 (91.7 %) having Ct values >30 on RT-PCR (or indeterminate on the Cobas 6800) and 6/13 (41.7 %) with Ct values >37 on RT-PCR (or indeterminate on the Cobas 6800) ( Table 3) . Eight of the 13 false negative samples (61.5 %) were from individuals within 7 days of symptom onset, four (30.7 %) were samples that were tested on the ID NOW within 1 h of collection, and only one (7.7 %) was from an individual with symptom duration ≤7 days and had the ID NOW performed within 1 h. Samples tested on the ID NOW instrument were more likely to be positive if the sample was tested within 1 h of collection (P=0.031) and from individuals who had lower Ct values on RT-PCR testing (P<0.001) ( Table 4 ). RT-PCR samples were more likely to be positive if the sample was tested on individuals with symptom duration ≤7 days (Table S2 , available in the online version of this article). When tested in triplicate using RT-PCR followed by triplicate testing by the CDC method and testing on the Cobas 6800, 5/10 (50.0 %) RT-PCR-negative samples, with paired positive ID NOW swabs, resolved as positive (Table S3) . One sample was unable to undergo additional testing. PPA between the ID NOW and RT-PCR, stratified based on individual characteristics, is provided in Fig. 1 . The highest ID NOW PPA (98.2 %) was in individuals with symptom duration ≤7 days and who had the ID NOW performed within 1 h, followed by individuals with symptom duration ≤7 days (91.2 %), and in samples with ID NOW performed within 1 h (89.8 %). Further details are provided within the supplementary material (Table S4 -S15). Twenty asymptomatic individuals at low risk of COVID-19 were tested, all of whom were negative on the ID NOW and RT-PCR. All four retrospective samples positive for SARS-CoV-2 were positive on the ID NOW. All 11 retrospective samples containing other respiratory viruses tested were negative. These samples were previously positive for one of either human metapneumovirus, adenovirus, parainfluenza virus, other coronavirus (NL63, HKU1, NL63), enterovirus, respiratory syncytial virus, influenza A H3N2, influenza A H1N1 or influenza B. Only one false negative sample (coloured in grey) was from an individual who had duration of symptoms ≤7 days at time of collection and had their sample tested within 1 h of collection. Our study compared the PPA of the ID NOW to RT-PCR among individuals confirmed to have SARS-CoV-2 infection. Overall, PPA of the ID NOW instrument was high at 89.1 % and comparable to our laboratory's RT-PCR (91.6 %) for this population. In the cohort of patients with symptom onset ≤7 days prior to collection, and whose samples could be tested within 1 h of collection, the PPA increased to 98.2 %. This is a result of higher Ct values being observed among those with symptom onset >7 days (Table 3) , resulting in decreased PPA of the ID NOW compared to RT-PCR (Table 4 ). The decreased ID NOW positivity among samples tested after 1 h from collection probably corresponds to viral degradation over time when transported in the absence of viral transport media. Although our study detected a high PPA of the ID NOW compared to our lab-developed RT-PCR, it is important to note that this is only when testing a population that is actively symptomatic and within the first 7 days of symptom onset. It is not meant to replace standard RT-PCR testing, particularly when testing individuals who are more likely to have lower viral loads, such as individuals with symptom duration greater than 7 days or asymptomatic close contacts. In patient populations where there is a high risk of a severe consequence of missing a case, such as hospitalized or continuing care facilities, RT-PCR or confirming ID NOW negatives with RT-PCR may be warranted. The ID NOW is best suited at the point of care for testing community members who develop early symptoms suggestive of COVID-19, particularly in remote locations where access to standard laboratory testing is limited. While there were instances in our study where the ID NOW was positive and the RT-PCR was negative, we believe these are true positives for several reasons. Participants recruited in our study were all recently diagnosed with COVID-19; none of the samples from the asymptomatic individuals at low risk of COVID-19 gave false positive results throughout the study; when retested in triplicate or on an alternative platform, 40 % of the discrepant RT-PCR samples had detectable SARS-CoV-2 RNA present; and no issues with false positive results have been identified by the ID NOW manufacturer or among previous publications within the literature [2] [3] [4] [5] [6] [7] [8] [9] [10] [11] [12] [13] [14] [15] [16] [17] . We feel that the discrepancies between the ID NOW and RT-PCR are probably related to other factors, such as the variability that comes with collecting multiple specimens. PPA of the ID NOW varies widely in the literature from 48.0 to 94.1 % [2] [3] [4] [5] [6] [7] [8] [9] [10] [11] [12] [13] [14] [15] [16] [17] . However, most of these studies varied in their study design or were conducted when the manufacturer still considered placing swabs in UTM as appropriate for sample collection (prior to April 2020). Several studies did not adhere to the recommended time limit of 1 h for time of collection to result which, as confirmed in this study, is a statistically significant factor in ID NOW's performance [2, 11, 15] . One of the major studies [3] that detected poorer performance of the ID NOW compared dry nasal swabs (tested on the ID NOW) to nasopharyngeal swabs (tested on Cepheid Xpert Xpress), which is an inappropriate comparison given the superiority of positivity rate among nasopharyngeal specimens to nasal specimens [21] . Furthermore, Basu et al. tested patients with symptom onset up to 1 month from time of sample collection, and it is unclear how many patients with confirmed COVID-19 had symptom duration ≤7 days [3] . Among the studies that adhered to current ID NOW manufacturer recommendations for specimen collection and transportation, ID NOW PPA was 66.7 and 94.1 % [16, 17] . However, both studies had small sample sizes and did not provide details as to whether all patients tested were symptomatic and within the first 7 days. The study with ID NOW PPA of 66.7%, for instance, was done primarily for 'surgical screening' , which suggests that many, if not all, patients were actually asymptomatic. The limitations of our study include the low number of hospitalized patients recruited such that we cannot make strong conclusions about the ID NOW performance among this population. The majority of throat swabs tested on the ID NOW were collected before the comparator swab (77.4%). However, we did not observe any difference in ID NOW or RT-PCR positivity rate when comparing patients who had ID NOW throat swab collected first vs second (Table 4 and supplementary material). There were discrepancies between ID NOW-positive, RT-PCRnegative specimens that could have resulted from multiple factors, such as intra-collector variability in the throat swab collections and degradation of virus during transportation/storage, as opposed to false positive ID NOW results. Our cross-reactivity study was also limited by using retrospective samples in UTM, as this is not an appropriate transport medium for the ID NOW. We determined that SARS-CoV-2 could be detected by the ID NOW in UTM-containing specimens, but we recognize that some analytical sensitivity may be lost by doing so. The strengths of our study include the large number of COVID-19-positive individuals recruited, particularly those residing within the community. We adhered to the manufacturer's recommendations, as much as possible, to recruit COVID-19-positive individuals with symptom duration ≤7 days at time of collection and to test samples on the ID NOW within 1 h of collection. Adhering to these requirements was difficult, as it often took several days from symptom onset for an individual in the community to get swabbed and test results reported. Consequently, symptom onset was often near 7 days. Meeting the 1 h criterion was challenging as we had to drive back and forth from participants' households to our laboratory, and many participants were located in distant parts of the city (e.g. >30 min drive to the laboratory) and other obstacles (e.g. traffic) increased transit time. Another strength of our study was the concurrent testing of asymptomatic individuals at low risk of COVID-19, and retrospective samples positive for other respiratory viruses, throughout the study to ensure there were no issues with false positive results (e.g. caused by contamination or cross-reactivity). The ID NOW was found to be a comparable method to our RT-PCR for the detection of SARS-CoV-2 among individuals with symptomatic COVID-19 infection. PPA was enhanced when tested on individuals with symptom onset ≤7 days and when time from collection to testing was within 1 h. These results reassure us that the ID NOW is a reliable test method in symptomatic individuals, especially when adhering to the FDA-approved indications and recommendations for testing. Given the speed and low complexity of ID NOW testing, these instruments can truly be used as a point-of-care device. As such, they will play an impactful role in combating the COVID-19 pandemic by improving testing in settings where low-volume, rapid (<1 h) turnaround times are much needed, such as among difficult to reach populations (e.g. homeless) and in rural areas where access to a laboratory is limited because transportation delays are significant. The authors received no specific grant from any funding agency. United States Food and Drug Administration. ID now COVID-19 -instructions for use How many are we missing with ID NOW COVID-19 assay using direct nasopharyngeal swabs? Findings from a midsized academic hospital clinical microbiology laboratory Performance of Abbott ID NOW COVID-19 rapid nucleic acid amplification test using nasopharyngeal swabs transported in viral transport media and dry nasal swabs in a New York city academic institution Clinical evaluation of three sample-to-answer platforms for the detection of SARS-CoV-2 Comparison of Abbott ID Now, DiaSorin Simplexa, and CDC FDA emergency use authorization methods for the detection of SARS-CoV-2 from nasopharyngeal and nasal swabs from individuals diagnosed with COVID-19 Comparison of Abbott ID Now and Abbott m2000 methods for the detection of SARS-CoV-2 from nasopharyngeal and nasal swabs from symptomatic patients Comparison of Cepheid xpert xpress and Abbott ID now to Roche Cobas for the rapid detection of SARS-COV-2 Direct comparison of SARS-CoV-2 analytical limits of detection across seven molecular assays Comparison of two commercial molecular tests and a laboratory-developed modification of the CDC 2019-nCoV Reverse transcriptase PCR assay for the detection of SARS-CoV-2 Evaluation of the COVID19 ID NOW EUA assay Comparative study of four SARS-CoV-2 Nucleic Acid Amplification Test (NAAT) platforms demonstrates that ID NOW performance is impaired substantially by patient and specimen type An extended laboratory validation study and comparative performance evaluation of the Abbott ID NOW COVID-19 assay in a coastal California tertiary care medical center Clinical evaluation and utilization of multiple molecular in vitro diagnostic assays for the detection of SARS-CoV-2 A comparison of five SARS-CoV-2 molecular assays with clinical correlations Comparison of Abbott ID NOW COVID-19 rapid molecular assay to cepheid Xpert Xpress SARS-CoV-2 assay in dry nasal swabs Commercial severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) molecular assays: Superior analytical sensitivity of cobas SARS-CoV-2 relative to NxTAG CoV extended panel and ID NOW COVID-19 test Performance characteristics of the ID NOW COVID-19 assay: a regional health care system experience Alberta health services. Collection of nasopharyngeal and throat swab for detection of respiratory infection Development and validation of reverse transcriptase-PCR assays for the testing of SARS-COV-2 CDC 2019-novel Coronavirus (2019-NCoV) real-time RT-PCR diagnostic panel: For emergency use only: Instructions for use Comparing nasopharyngeal and midturbinate nasal swab testing for the identification of severe acute respiratory syndrome Coronavirus 2 This work was funded using internal operating funds of Alberta Precision Laboratories and Alberta Health Services (AHS). Test kits and instruments were paid for by the Public Health Agency of Canada. We thank the AHS mobile integrated health team for collecting samples in the community and Alberta Precision Laboratory staff for assistance with testing of samples. The manufacturer had no role to play in the study. The authors have no conflict of interests to disclose pertaining to this study The University of Calgary Research Ethics board approved this study (REB20-444). Five reasons to publish your next article with a Microbiology Society journal 1 . The Microbiology Society is a not-for-profit organization. 2. We offer fast and rigorous peer review -average time to first decision is 4-6 weeks. 3. Our journals have a global readership with subscriptions held in research institutions around the world. 4. 80% of our authors rate our submission process as 'excellent' or 'very good'. 5. Your article will be published on an interactive journal platform with advanced metrics.Find out more and submit your article at microbiologyresearch.org.