key: cord-0698935-0v4tlvdc authors: Zhou, Elsa S.; Bhatia, Sujata K. title: Decrease in Reported Rates of Cardiovascular Device-Related Adverse Events During the Coronavirus Disease 2019 Pandemic date: 2021-10-31 journal: Am J Cardiol DOI: 10.1016/j.amjcard.2021.10.006 sha: f94c71754b5f26470fdefd5e985e02abe1483a8a doc_id: 698935 cord_uid: 0v4tlvdc nan Given the disruption of cardiology practices by the coronavirus disease 2019 (COVID-19) pandemic, we sought to investigate whether the number of reports of adverse events attributed to cardiovascular medical devices changed over the course of the pandemic, specifically examining implantable cardioverter defibrillators (ICDs) and coronary drug-eluting stents (DESs). Using data from the Food and Drug Administration (FDA) Manufacturer And User Facility Device Experience (MAUDE) database, we compared weekly reported rates of adverse events for each device during the year immediately preceding the pandemic (March 2019 to March 2020) with those during the first year of the pandemic (March 2020 to March 2021). We report a 46% decrease in reported ICD-related deaths during the pandemic compared with before the pandemic, and a 27% decrease in reported coronary DESrelated injuries during the pandemic compared with before the pandemic. Because adverse event reporting plays a critical role in postmarket surveillance and risk assessment for medical devices, this report assesses the impact of the COVID-19 pandemic on the number of weekly reports of adverse events attributed to 2 cardiovascular medical devices: ICDs and coronary DESs. We used the FDA MAUDE database, which lists reports from manufacturers, distributors, clinicians, and other voluntary reporters and is publicly accessible. 1 We filtered the MAUDE data by device and adverse event type, examining malfunction, injury, and death reports with the filter "Implantable Cardioverter Defibrillator (Non-Crt)" for ICDs and the filter "Coronary Drug-Eluting Stent" for coronary DESs. Since the World Health Organization officially declared COVID-19 a pandemic on March 11, 2020, 2 we chose to record the number of reports given each week over the course of 3 years (March 2018 to March 2019, March 2019 to March 20, and March 2020 to March 2021). For clarity, March 2020 to March 2021 will be herein called "pandemic data" or "2020 to 2021," March 2019 to March 2020 will be called "prepandemic data" or "2019 to 2020," and March 2018 to March 2019 will be called "2018 to 2019." We performed paired t tests for the differences between weekly reported adverse event types for each event type. Comparing the data from 2019 to 2020 with the data from 2020 to 2021, we found that there were, on average, approximately 2.6 fewer weekly reports of ICD-attributed deaths during the pandemic than there were in the prepandemic year, a decrease of approximately 45.8% (p <0.0001) (Figure 1 ). To determine whether this trend was isolated, we also compared the weekly reports of ICD-attributed deaths for 2018 to 2019 with the data for 2019 to 2020. In this case, we found a decrease in the average number of weekly reported ICD-attributed deaths by approximately 1.6 weekly reports between 2018 and 2019 and 2019 to 2020; however, the decrease was not as large as that from the prepandemic to the pandemic and the p value was approximately 0.0023. We also compared the differences in reported ICD-attributed deaths (prepandemic subtracted from pandemic) with the number of reported COVID-19-attributed deaths and found no significant correlation. We additionally examined the number of weekly reports of ICD malfunctions and ICD-attributed injuries in the pandemic year versus the prepandemic year and found no significant differences (ICD malfunctions: mean difference [pandemic À prepandemic] approximately equal to 10.8 reports, p value approximately equal to 0.19; ICD injuries: mean difference [pandemic À prepandemic] approximately equal to 9.9, p value approximately equal to 0.52). We examined adverse event data for coronary DESs and found that there were, on average, approximately 14. cally significant result (p value approximately equal to 0.63). We again compared the differences in weekly coronary DES-attributed injuries (prepandemic subtracted from pandemic) with the weekly number of COVID-19 deaths but found no significant correlation. Our statistical analyses for coronary DES-attributed malfunctions and deaths did not yield significant differences (coronary DES malfunctions: mean difference [pandemic À prepandemic] approximately equal to À3.0, p value approximately equal to 0.44; coronary DES deaths: mean difference [pandemic À prepandemic] approximately equal to 1.0, p value approximately equal to 0.38). In summary, our examination of the FDA MAUDE database revealed that the reported rates of ICD-attributed deaths and coronary DES-attributed injuries significantly decreased during the pandemic. There are several potential explanations for the observed decreases. One explanation is that fewer cardiovascular devices were implanted during the pandemic because of undertreatment, leading to fewer adverse events. An Italian study of 84 arrhythmia centers revealed that 92.9% of centers reported a significant reduction in ICD implantation for primary prevention in the first 2 months of the pandemic. 3 Similar trends have been recorded in Catalonia, 4 Greece, 5 Iran, 6 and Romania. 7 The early stages of the pandemic were also associated with a 23.4% reduction in hospital admissions for acute coronary syndromes in two Italian high-volume centers, 8 with similar trends in Austria 9 and China. 10 A study in the United States of 9 high-volume centers revealed a 38% reduction in ST-segment cardiac catheterization laboratory activations, 11 with a similar trend in Spain. 12 Another explanation is that there was a lower burden of arrhythmia and major coronary events during the pandemic. A US study of 2,458 patients with ICDs demonstrated a 32% reduction in ventricular arrhythmias needing device therapies during the pandemic. 13 However, a different US study of 14,665 patients with ICDs demonstrated an increase in defibrillator shocks in 3 major cities during the pandemic. 14 A third explanation is that the observed decrease in ICD-attributed deaths and coronary DES-attributed injuries is a result of underreporting, thus raising concerns about adverse event surveillance during the pandemic. Further investigation is needed to determine the extent of cardiovascular undertreatment, underreporting, or both during the COVID-19 pandemic. The authors have no conflicts of interest to declare. Food and Drug Administration, MAUDEmanufacturer and user facility device experience WHO declares COVID-19 a pandemic Impact of COVID-19 pandemic on the clinical activities related to arrhythmias and electrophysiology in Italy: results of a survey promoted by AIAC (Italian Association of Arrhythmology and Cardiac Pacing) Reduction in new cardiac electronic device implantations in Catalonia during COVID-19 Impact of COVID-19 pandemic on cardiac electronic device implantations in Northwestern Greece The survey of electrophysiology lab activity during the COVID-19 pandemic Impact of COVID-19 pandemic on the implantation of intra-cardiac devices in diabetic and non-diabetic patients in the western of Romania Impact of the COVID-19 pandemic on coronary invasive procedures at two Italian high-volume referral centers Decline of acute coronary syndrome admissions in Austria since the outbreak of COVID-19: the pandemic response causes cardiac collateral damage Impact of coronavirus disease 2019 (COVID-19) outbreak on STsegment-elevation myocardial infarction care in Hong Kong Reduction in ST-segment elevation cardiac catheterization laboratory activations in the United States during COVID-19 pandemic Impacto de la pandemia de COVID-19 sobre la actividad asistencial en cardiolog ıa intervencionista en España Ventricular arrhythmia burden during the coronavirus disease 2019 (COVID-19) pandemic Implantable cardioverter-defibrillator shocks during COVID-19 outbreak ARTICLE IN PRESS 2 The American Journal of Cardiology