key: cord-0698930-8sse0jsv authors: Wang, C.-G.; Li, Z.; Liu, S.; Ng, C.T.; Marzuki, M.; Jeslyn Wong, P.S.; Tan, B.; Lee, A.; Hui Lim, C.F.; Bifani, P.; Fang, Z.; Ching Wong, J.C.; Setoh, Y.X.; Yang, Y.Y.; Mun, C.H.; Fiona Phua, S.Z.; Lim, W.Q.; Lin, L.; Cook, A.R.; Tanoto, H.; Ng, L.-C.; Singhal, A.; Leong, Y.W.; Loh, X.J. title: N95 respirator decontamination: a study in reusability date: 2021-06-23 journal: Mater Today Adv DOI: 10.1016/j.mtadv.2021.100148 sha: bc64938d2095dfb0ea6a820b78ef10f7a88cd253 doc_id: 698930 cord_uid: 8sse0jsv The coronavirus disease 2019 (COVID-19) pandemic had caused a severe depletion of the worldwide supply of N95 respirators. The development of methods to effectively decontaminate N95 respirators while maintaining their integrity is crucial for respirator regeneration and reuse. In this study, we systematically evaluated five respirator decontamination methods using vaporized hydrogen peroxide (VHP) or ultraviolet (254 nm wavelength, UVC) radiation. Through testing the bioburden, filtration, fluid resistance, and fit (shape) of the decontaminated respirators, we found that the decontamination methods using BioQuell VHP, custom VHP container, Steris VHP, and Sterrad VHP effectively inactivated Cardiovirus (3-log(10) reduction) and bacteria (6-log(10) reduction) without compromising the respirator integrity after 2โ€“15 cycles. Hope UVC system was capable of inactivating Cardiovirus (3-log(10) reduction) but exhibited relatively poorer bactericidal activity. These methods are capable of decontaminating 10โ€“1000 respirators per batch with varied decontamination times (10โ€“200 min). Our findings show that N95 respirators treated by the previously mentioned decontamination methods are safe and effective for reuse by industry, laboratories, and hospitals. The coronavirus disease 2019 (COVID-19) pandemic has spread worldwide. Various solutions have been proposed to combat the spread of the virus, including sanitization [1] . Other novel and interesting solutions to handle the pandemic have also come up, such as the use of ionizer plants to enhance ventilation or a smart mask to monitor the vital signs of COVID-19 patients [2, 3] . However, there has been no other more urgent need than the increasing need for personal protective equipment (PPE). PPE is the primary line of defense for health care workers in their daily routines. The novel severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) remains viable for hours in aerosols and droplets generated by coughing or sneezing and could be transmitted person-to-person via direct (inhalation or contact with mucosal surfaces) or indirect routes (contact with virus-contaminated surfaces followed by transmitting to a person) [4e6] . Human exhalation droplets are generally less than 5 mm in diameter (dry state) or larger (wet state) [7, 8] . Filtering facepiece respirators (FFR), such as N95 respirators, are required PPE for health care workers because they are able to filter 0.3-mm-sized particles with 95% efficiency and provide high protection to filter out human exhalation droplets [9e11] . As with other PPEs, N95 FFRs were in serious shortage because of the COVID-19 pandemic. To address this issue, the development of regeneration approaches to decontaminate and reuse N95 FFRs was an essential and pragmatic interim solution to prolong the supply of critical PPE [12e14] . N95 FFRs are negative-pressure air-purifying particulate respirators and need to be approved by National Institute for Occupational Safety and Health (NIOSH). Although these respirators are designed and recommended for single use, the evaluation of their safety and reusability post-decontamination is of strategic interest internationally. Several dry and wet decontamination methods have been reported to disinfect masks using heat, ozone, ultraviolet (UV) radiation, steam, alcohol, ethylene oxide vapors, and vaporized hydrogen peroxide (VHP) [15e19] . Based on the guidance document from the Food and Drug Administration (FDA) 'Enforcement Policy for Face Masks and Respirators During the COVID-19 Public Health Emergency (PHE)' (FDA Guidance, May 2020), an effective decontamination method must (1) inactivate the target organism, such as the virus that causes COVID-19; (2) not damage the respirator's filtration efficacy; (3) not affect the respirator's fit; (4) not affect the liquid barrier performance if it is a fluid resistance respirator; and (v) be safe for the person wearing the respirator [20e22]. The main N95 FFR manufacturer 3M reported a few recommended methods, including the use of VHP, steam, heat, and UV radiation, for decontamination without deterioration of the protection properties. Research has been conducted using industrial systems or home appliances for N95 FFR decontamination. Nguyen et al. reported the use of dry heat (100 C, 5% relative humidity, 50 min) generated by an electric cooker to decontaminate N95 FFRs. The filtration performance and shape of the FFR did not degrade after 20 cycles of the dry heat treatment [17] . Kumar et al. investigated different decontamination methods for N95 FFRs. The results demonstrated that peracetic acid dry fogging, VHP, autoclaving, and moist heat treatment are effective against SARS-CoV-2 and maintain the FFR integrity up to five decontamination cycles [18] . Song et al. reported that the use of dry heat at 60 C or 70 C for 1 h could effectively kill or inactivate respiratory bacteria, one fungi species, and the H1N1 indicator virus to decontaminate surgical face masks and N95 FFRs [19] . The decontaminated N95 FFRs exhibited filtering efficiency of aerosol as 98%, which is similar to the value (99%) of new FFRs. Warriner et al. recently reported a continuous process using hydroxyl-radicals for FFR decontamination. The process was effective in inactivating bacterial and human coronavirus without negative effects on the FFR functionality [23] . Although these studies demonstrated effective methods for N95 FFR decontamination and reuse, a majority of the studies used home electronics or small-scale laboratory equipment, which are not suitable for largescale application. In addition, the decontaminated FFRs were generally evaluated by a single property, such as microbial inactivation or filtration efficacy. A comprehensive evaluation of different decontamination methods is desired but less studied. In this report, we investigated the validity and efficacy of five decontamination methods using VHP and UV radiation for N95 FFR reuse. The VHP decontamination methods were either authorized by FDA for emergency use or were adapted from authorized methods and able to decontaminate 10 to 1,000 FFRs per operation cycle. The UV radiation method requires only 10 min per decontamination cycle without using chemicals (e.g. hydrogen peroxide). Therefore, these methods were considered promising for urgent deployment and with good accessibility by laboratory, hospital, and industry. Through testing the bioburden, filtration, fluid resistance, and fit of the decontaminated FFRs, the usefulness and limitations of the five decontamination methods are evaluated. Surgical fluid-resistant N95 FFRs (Model 1860 [resistance: 120 mm Hg] and 1860S [resistance: 80 mmHg]; 3M Co. Ltd., USA) were used for all tests describe in the following sections. These non-cellulose N95 FFRs are commonly used in hospitals in Singapore. Respirator samples were prepared and analyzed in biosafety level 2 (BSL-2) laboratories after decontamination tests. Decontamination methods were evaluated and performed by using a chemical VHP method and a photoradiation short-wavelength UV (UVC; wavelength 254 nm) method. Three commercially available VHP systems (BioQuell, Steris, and Sterrad), one customized VHP container (VHPC), and a Hope UVC system are established for the tests. The customized VHPC uses a Steris VHP generator, which is an adaptation of the Battelle solution (USA). Table 1 summarizes the decontamination methods used in this study. Bioburden tests were performed to determine the viricidal as well as sporicidal or mycobacterial activities before and after decontamination of the FFRs. Mengovirus (vMC 0 , belonging to genus cardiovirus) as well as spores of Gram-positive bacteria Geobacillus stearothermophilus and bacteria of Mycobacterium genus viz. Mycobacterium smegmatis were used for these tests. The agents were selected for the decontamination tests because of their biological safety, experimental relevance, and accessibility [24] . Mengovirus, which is a non-enveloped BSL-2 Cardiovirus, was selected as non-enveloped viruses typically have greater resistance to inactivation than enveloped viruses. G. stearothermophilus was chosen because of its ability to form spores, whereas Mycobacterium spp. was selected because it is known to be more chemically resistant than other bacterial species. The bioburden tests were evaluated with two Emergency Use Authorization recommended criteria as (1) 3-log 10 (99.9%) reduction in virucidal activity and (2) 6-log 10 (99.9999%) reduction in mycobacterial or sporicidal activity to define the completion of decontamination. Filtration efficiency of the FFRs was determined via bacterial filtration efficiency (BFE) and particle filtration efficiency (PFE) tests in line with the guidelines of the FDA and Ministry of Health of Singapore [25e27]. BFE test determines the filtration efficiency by comparing the bacterial control counts to test article effluent counts. The BFE tests were evaluated with ASTM F2101-01 standard as over 98% filtration efficiency for particles with or below 3 mm diameter [28] . PFE test evaluates the non-viable particle retention or filtration efficiency of the respirators at submicron levels. The PFE tests were evaluated with NIOSH method as over 95% filtration efficiency for airborne particles with 0.3 mm diameter. The experimental details of BFE and PFE tests are described in Supporting Information. The fluid resistance test for the N95 FFRs was conducted using synthetic blood to simulate the splatter of bodily fluids onto the samples. The sample was visually inspected to determine the penetration of the fluid into the inner surface. The experimental details are described in Supporting Information. A clinical trial involving 10 healthy volunteers were recruited for the quantitative mask fit test. The clinical trial testing protocol and ethics review and safeguard were approved under the DSO National Laboratories and Singapore Armed Forces (DSO-SAF) Institutional Review Board (IRB No: 0022/2020, Singapore). The respirator fit examination was conducted using the procedure from Occupational Safety and Health Administration (OSHA) protocol no. 29 CFR 1910.134. Briefly, each volunteer donned a decontaminated N95 FFR that was connected to the PortaCount (TSI PortaCount Pro รพ Respirator FitTester, Model 8048-1) and performed a series of physical activities to obtain an overall fit factor score. The physical activities include breathing normally and deeply, moving head up-and-down and sideto-side, bending over and talking. Fit score for each physical activity was then averaged. An overall fit score greater than 100 is required for N95 respirators as stated in the previously mentioned OSHA protocol. The difference between the test and control masks was calculated from the difference in mean log 10 TCID 50 per carrier value for Mengovirus or mean log 10 pfu/carrier for G. stearothermophilus and M. smegmatis. The value for the limit of detection of the respective titration assays was used for masks that had no detectable growth. For each experiment, a one-sided hypothesis test that the difference in mean log 10 TCID 50 per carrier or log 10 pfu per carrier scores was at least 3-log 10 (Mengovirus) or 6-log 10 (G. stearothermophilus and M. smegmatis) was conducted. These differences correspond to at least a thousand-fold and million-fold difference on the linear scale. The analysis was performed using R (Bell Labs). According to the FDA guidance described previously, the viral and bacterial decontamination, filtration, fluid resistance, and fit tests were performed to evaluate the usefulness of the decontaminated N95 FFRs. Currently, most of the studies are applicable for laboratory scale and did not evaluate the five requirements comprehensively. The requirement and description of the tests are shown in Table 2 . For decontamination, filtration, and fluid resistance tests, the decontaminated FFRs and the decontamination processes were examined and evaluated based on the FDA guidance. For the fit examination, the evaluation was based on the feedback and observation of volunteers. Five decontamination methods, namely, BioQuell VHP, VHPC, Steris VHP, Sterrad VHP, and Hope UVC, were used for the decontamination of FFRs. Mengovirus and two types of bacteria, G. stearothermophilus and M. smegmatis, were spiked respectively onto experimental FFRs and six control FFRs with or without simulated body fluids (0.2% mucin and 0.2% sodium chloride; Fig. 1 , L1 and L2 positions, respectively). Decontamination of the experimental FFRs was conducted once via the different decontamination methods, and the FFRs were then subjected to bioburden tests to determine the load of inactivated viruses or bacteria that remained on the surface after decontamination, and thus, the efficacy of the decontamination method was assessed. Mengovirus, which is a non-enveloped BSL-2 Cardiovirus, was used to evaluate the virus decontamination efficiency of the five decontamination methods. In this test, 30 experimental FFRs and six control FFRs were spiked with Mengovirus (location L1, Fig. 1) and Mengovirus with 0.2% saline (sodium chloride solution) as synthetic sweat and 0.2% mucin as synthetic saliva (at location L2, Fig. 1 ). The experimental FFRs were decontaminated once and subjected to bioburden tests. From an original spike concentration of 5.56-to 8.89-log 10 TCID 50 /carrier, Mengovirus titers of 2.15-to 6.90-log 10 TCID 50 / carrier were recovered from the inoculation sites of the control FFRs. As shown in Fig. 2 and Tables S2 and S3, BioQuell VHP and Steris VHP successfully achieved >3-log 10 reduction of Mengovirus at location L1 (without saline and mucin), whereas only VHPC displayed >3-log 10 reduction of Mengovirus at location L2 (with saline and mucin). It was noted that the decontamination efficacy of several VHP systems could not be fairly evaluated and may be underestimated in our Phase I study because of lower-than-expected recovery of virus from the control FFRs. Hope UVC carries a small chamber for decontaminating 10 respirators in 10 min without using chemicals (e.g. hydrogen peroxide), offering a facile and cost-effective decontamination process. Hope UVC was effective at inactivating Mengovirus and achieved 3-log 10 virucidal reduction on the FFRs in the presence (3.82-log 10 reduction) or absence (4.41-log 10 reduction) of saline and mucin (Fig. 3) . The TCID 50 data and microscopic images of Mengovirus recovery are shown in Tables S6eS11 and Figs. S1eS4. We also used G. stearothermophilus and M. smegmatis for the decontamination and bioburden tests. G. stearothermophilus or M. smegmatis were spiked onto the experimental and control FFRs. The experimental FFRs were similarly decontaminated using the five methods described previously. Although there was a nearly 6-log 10 reduction when compared with the spiked load, the decontamination did not meet the FDA requirement when compared with the bacterial recovery from the control FFRs. The results indicated that all four VHP methods were efficient to achieve certain reduction (ranging from 3.23-to 5.17-log 10 reduction) for G. stearothermophilus (48 h incubation; Fig. 4A ; Table S4 ) and M. smegmatis (72 h incubation; Fig. 4B ; Table S5 ). However, the decontamination effectiveness of the VHP systems could not be demonstrated to meet the recommendations from the FDA guidance. The low recovery (5.73-log 10 CFU/carrier) of bacteria from the control FFRs inevitably limited the achievement of decontamination tests. The reduction thus cannot be attributed to the decontamination process. Apart from VHP methods, Hope UVC was also applied for FFR decontamination of G. stearothermophilus and M. smegmatis. After treatment using Hope UVC, the viable load of G. stearothermophilus (Fig. 5A ) that remained on the experimental FFRs was similar to that of control FFRs (4.64-log 10 CFU/carrier [treated] and 5.31-log 10 CFU/ carrier [control] ). This result indicated that the Hope UVC system was not effective at decontaminating G. stearothermophilus on the FFRs. Hope UVC was also found to be less effective in inactivating M. smegmatis (Fig. 5B ) than the other VHP methods, achieving only 4.44-log 10 reduction after UVC treatment. The recovery colonies were dispersive from 0.69-to 3.31-log 10 CFU/carrier, indicating that the decontamination is less effective for all tested FFRs. Although Hope UVC was effective at inactivating Mengovirus, it is less effective against the bacteria strains that we tested. The poor bacterial decontamination efficacy of Hope UVC did not support further investigation in our Phase II examination. In our Phase I FFR decontamination test, we noted poor recovery of the challenge organisms from the control (untreated) FFRs, which limited the achievement of the >3-log 10 reduction requirement for Mengovirus and >6-log 10 reduction requirement for bacteria. We were also motivated by the observation that no hydrogen peroxide or other toxic residues remained on the VHP-treated FFRs, suggesting that the N95 FFRs are safe for reuse if efficacy was supported. Therefore, a Phase II study was conducted using a higher initial concentration of challenge organism to re-evaluate the efficacy of the VHP systems. BioQuell VHP, custom VHPC, Steris VHP, and Sterrad VHP were again tested against Mengovirus using a higher virus concentration of 7.15-to 8.15-log 10 TCID 50 /carrier and was conducted as two independent runs (Run I and Run II, six control and six test masks [12 masks in total] per run for BioQuell and VHPC; three control and three test masks [six masks in total] per run for Steris and Sterrad). In Phase II, the recovery of virus from control FFRs was much higher, with a mean recovery of 7.05-to 7.36-log 10 TCID 50 per carrier of Mengovirus across all four VHP methods ( Fig. 6 and Table S12 ). In the treated FFRs, between 1.4-to 1.9-log 10 TCID 50 per carrier Mengovirus were recovered, resulting in a 5.13-to 5.54-log 10 reduction in Mengovirus across all four VHP methods and fulfilling the >3-log 10 reduction recommendation from the FDA guidance. Similarly, the M. smegmatis challenge was repeated using a higher initial concentration of 7.72-log 10 CFU/carrier. The M. smegmatis experiment was similarly conducted as two independent runs (see Table S13 for details). In Phase II, the recovery of M. smegmatis from the control FFRs was in the range of 7.21-to 7.67-log 10 cfu/carrier. In the treated FFRs, only 0.20-to 0.22log 10 cfu/carrier could be recovered, resulting in a 6.99-to 7.45log 10 reduction of M. smegmatis across all four VHP methods ( Fig. 7 and Table S13), and similarly fulfilling the >6-log 10 reduction recommendation set by the FDA guidance. PFE and BFE tests were conducted on the FFRs after the respirators were subjected to the various decontamination methods. The maximum cycles of the decontamination systems were used to FFRs, and the filtration efficacy of the decontaminated respirators were evaluated. The testing results are summarized in Table 4 . All respirators treated with BioQuell VHP (20 cycles), VHPC (20 cycles), Steris VHP (10 cycles), or Hope UVC (10 cycles) passed the PFE tests (95%) with average PFE values over 99%. For FFRs treated with Sterrad VHP (two cycles), 28 FFRs out of 30 passed the PFE tests. The appearance of crumpling FFRs at the inner surfaces of the FFRs that failed the PFE test was observed (Fig. S5) . The crumpling may occur either during transportation or as a result of the vacuum treatment during the decontamination process. The FFRs with crumpling exhibited inferior PFE values (<95%). Therefore, Sterrad decontamination method may affect the respirator PFE although no visible tear or puncture to the outer and inner parts of the FFRs. It is also worthy to note that no crumpled FFR was generated when the Sterrad VHP decontamination cycle was repeated with the ALLClear function turned off. (The ALLClear is an optional function using a small amount of plasma to increase the decontamination productivity in Sterrad VHP system. Plasma may deteriorate the filter materials and function.) All the decontaminated FFRs (30 over 30) using Sterrad VHP but without ALLClear function passed the PFE test (95%). Further bioburden tests are necessary to confirm the decontamination efficacy of Sterrad VHP without ALLClear function. Similarly, for BFE evaluation, FFRs were decontaminated under the maximum cycles within each decontamination method. The BFE test setup is shown in Fig. 8 , and the results are summarized in Table 3 . All FFRs passed the BFE test (>98%), with an average BFE value over 99%, except for two FFRs after decontamination with BioQuell VHP for 20 cycles. However, the failure of the BFE tests could be attributed to experimental error. As shown in Fig. 8 , droplets spiked with bacteria from the glass column may have entered and contaminated the stage 1 impactor plate during the removal of the FFR sample, leading to false positives. Typically, the aerosol particles measuring 2e3 mm would have been captured in stages 3e5 of the impactor, resulting in significant growth of bacteria colonies as was evident in the measurement of the control. Nevertheless, bacteria colonies were not detected at stages 3e5 in the failed samples, indicating that these were false positives caused by experimental error. Therefore, BioQuell VHP is also an effective decontamination method without deteriorating the BFE of FFRs. In summary, BioQuell VHP, VHPC, Steris VHP, Sterrad VHP (without ALLClear function), and Hope UVC methods exhibited no negative effect on PFE and BFE of FFRs. Sterrad VHP with ALLClear function would occasionally render crumpling on the FFRs, causing the decrease of PFE. ASTM F1862/F1862 M standard was used to evaluate the fluid resistance of FFRs, as shown in (Fig. 9 ). At least 29 of 32 FFRs should exhibit no penetration of synthetic blood to pass the fluid Table 4 . It is found that for each decontamination method, more than 30 decontaminated FFRs maintain their fluid resistance over 120 mm Hg for preventing synthetic blood penetration. The result emphasized no deterioration in the fluid resistance of the FFRs regardless of the decontamination method. A quantitative fit examination was conducted with 10 participants, of which five were males and five were females. Before examination, N95 FFRs were decontaminated with BioQuell VHP (5, 10, 15, and 20 cycles), VHPC (5, 10, 15, and 20 cycles), Steris VHP (10 cycles), Sterrad VHP (two cycles), or Hope UVC (five cycles). Face seal and odor checking was evaluated to determine the reusability of decontaminated FFRs in real case application. Decontaminated FFRs are deemed to provide good face seal when they passed the fit test. The fit examination results are summarized in Table 5 . BioQuell VHP and VHPC decontaminated FFRs maintained their original forms when the decontamination cycles were below 15. A few FFRs were observed with a deformed shape when the decontamination cycles reached 20 (Fig. S6a) . FFRs could maintain the shapes after applying Steris VHP and Sterrad VHP decontamination. However, the inner side of a few FFRs was observed to be crumpled after treating with Steris VHP (Fig. S6b ). N95 FFRs decontaminated by Hope UVC were deformed because of the physical stretching when mounting onto the frame of Hope UVC (Fig. S6c) . N95 FFRs are highly important and effective PPE to prevent the spread of SARS-CoV-2 and other airborne infectious diseases. In this research, we studied five decontamination methods using VHP or UV radiation for N95 FFR decontamination and reuse. Decontamination efficacy, filtration efficacy, fluid resistance, and shape (fit) tests of the decontaminated FFRs were examined to evaluate the decontamination methods. BioQuell VHP and VHPC are effective in inactivating M. smegmatis and Mengovirus without compromising the filtration efficacy, fluid resistance, and fit of the FFRs up to 15 decontamination cycles. Furthermore, BioQuell VHP and VHPC are capable of decontaminating more than 400 FFRs per batch, which are beneficial for industry use. Steris VHP and Sterrad VHP exhibited superior decontamination abilities for M. smegmatis and Mengovirus. Steris VHP also exhibited the highest performance to maintain FFR integrity (filtration, fluid resistance, and shape) among the five decontamination methods. Sterrad VHP could also highly maintain the FFR performance but slightly deteriorate PFE. Hope UVC was effective in inactivating Mengovirus (>3-log 10 reduction); however, it was not amenable for inactivating G. stearothermophilus and M. smegmatis. FFRs with Hope UVC decontamination showed high filtration efficacy and fluid resistance, but a few FFRs could not maintain their shapes and thus failed the FFR fit test. Steris VHP and Sterrad VHP are compact equipment with small throughput (10 respirators per batch) for rapid (28 min) and effective FFR decontamination and reduction of virucidal activity, which is practical for laboratory or hospital use. Sanitizing agents for virus inactivation and disinfection The efficacy of plant-based ionizers in removing aerosol for COVID-19 mitigation Lab-on-mask for remote respiratory monitoring Respiratory virus shedding in exhaled breath and efficacy of face masks Aerosol and surface stability of SARS-CoV-2 as compared with SARS-CoV-1 It is time to address airborne transmission of coronavirus disease 2019 (COVID-19) Turbulent gas clouds and respiratory pathogen emissions: potential implications for reducing transmission of COVID-19 On coughing and airborne droplet transmission to humans Critical supply shortages -the need for ventilators and personal protective equipment during the covid-19 pandemic To mask or not to mask: modeling the potential for face mask use by the general public to curtail the COVID-19 pandemic Protecting healthcare workers from SARS-CoV-2 infection: practical indications Face masks in the new COVID-19 normal: materials, testing, and perspectives Filtering facepiece respirator (N95 respirator) reprocessing: a systematic review Cort es, Decontamination and reuse of N95 filtering facemask respirators: a systematic review of the literature Can N95 respirators be reused after disinfection? How many times? Evaluation of regeneration processes for filtering facepiece respirators in terms of the bacteria inactivation efficiency and influences on filtration performance Dry heat as a decontamination method for N95 respirator reuse Decontamination of N95 masks for re-use employing 7 widely available sterilization methods Decontamination of surgical face masks and N95 respirators by dry heat pasteurization for one hour at 70 C, Am Recommendations for Sponsors Requesting EUAs for Decontamination and Bioburden Reduction Systems for Face Masks and Respirators during the Coronavirus Disease 2019 (COVID-19) Public Health Emergency, Food and Drug Administration Implementing Filtering Facepiece Respirator (FFR) Reuse, Including Reuse after Decontamination, when There Are Known Shortages of N95 Respirators Recommended Guidance for Extended Use and Limited Reuse of N95 Filtering Facepiece Respirators in Healthcare Settings, US National Institute for Occupational Safety and Health Decontamination of N95 and surgical masks using a treatment based on a continuous gas phase-advanced oxidation process Antiseptics and disinfectants: activity, action, and resistance Department of Health and Human Services, Food and Drug Administration The National Infection Prevention and Control Guidelines for Acute Healthcare Facilities, Ministry of Health of Singapore Filtration performance of FDAcleared surgical masks Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus Aereus This work was financially sponsored by Temasek Foundation and partially supported by the Agency of Science, Technology, and Research (A*STAR) and the National Environment Agency, Singapore. The authors would like to thank Dixun Lim and Keng Wai Mak for supporting the assessment of decontamination efficacy in the Phase II trials. The authors also acknowledge the contributions of Xi Ping Ni, Poh Chong Lim, Jian Li and Kangyi Zhang in performing mask decontamination and testing. The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper. Supplementary data to this article can be found online at https://doi.org/10.1016/j.mtadv.2021.100148.