key: cord-0695948-0qcrdijm authors: Frederick, Amy; Roumeliotis, Michael; Grendarova, Petra; Quirk, Sarah title: Performance of a knowledge‐based planning model for optimizing intensity‐modulated radiotherapy plans for partial breast irradiation date: 2021-12-22 journal: J Appl Clin Med Phys DOI: 10.1002/acm2.13506 sha: ded81b11646ac4866bb8b687b1bd9833834e88df doc_id: 695948 cord_uid: 0qcrdijm PURPOSE: To evaluate a knowledge‐based (KB) planning model for RapidPlan, generated using a five‐field intensity‐modulated radiotherapy (IMRT) class solution beam strategy and rigorous dosimetric constraints for accelerated partial breast irradiation (APBI). MATERIALS AND METHODS: The RapidPlan model was configured using 64 APBI treatment plans and validated for 120 APBI patients who were not included in the training dataset. KB plan dosimetry was compared to clinical plan dosimetry, the clinical planning constraints, and the constraints used in phase III APBI trials. Dosimetric differences between clinical and KB plans were evaluated using paired two‐tailed Wilcoxon signed‐rank tests. RESULTS: KB planning was able to produce IMRT‐based APBI plans in a single optimization without manual intervention that are comparable or better than the conventionally optimized, clinical plans. Comparing KB plans to clinical plans, differences in PTV, heart, contralateral breast, and ipsilateral lung dose–volume metrics were not clinically significant. The ipsilateral breast volume receiving at least 50% of the prescription dose was statistically and clinically significantly lower in the KB plans. CONCLUSION: KB planning for IMRT‐based APBI provides equivalent or better dosimetry compared to conventional inverse planning. This model may be reliably applied in clinical practice and could be used to transfer planning expertise to ensure consistency in APBI plan quality. Hypofractionated whole-breast irradiation (WBI) and accelerated partial breast irradiation (APBI) have been investigated for early-stage breast cancer in multiple phase III clinical trials over the last two decades. [1] [2] [3] [4] [5] [6] [7] Moderate hypofractionation for WBI, delivering 40-42.5 Gy in 15-16 fractions over 3 weeks, has become an international standard. 8 Recent studies 3, 4 have shown that five-fraction regimens for WBI are also safe and effec-techniques and dose constraints may alter the therapeutic ratio of APBI regimens and contribute to the conflicting complication profiles. [5] [6] [7] [12] [13] [14] [15] [16] The most common techniques used to deliver external beam APBI are three-dimensional conformal radiotherapy (3D-CRT) and intensity-modulated radiotherapy (IMRT). Several APBI planning studies 17, 18 have shown that IMRT improves dose conformity, dose homogeneity, and normal tissue sparing compared to 3D-CRT. These improvements may come at the cost of an increase in the number of monitor units and low-level radiation exposure of normal tissues, 17, 18 but secondary cancer risk remains lower than for WBI. 19, 20 Furthermore, there is strong clinical evidence supporting IMRT as a technique to deliver APBI. 7 However, inverse planning can be time-consuming, resource intensive, and subjective. 21 Dose-volume constraints are guided by clinical experience and recommended values from the literature. 22 Geometric variations in patient anatomy lead to large patient-to-patient variation in organ at risk (OAR) sparing and the need for additional dose control structures. As a result, the final plan quality is variable and dependent on time constraints, available resources, and the planner's (or institution's) experience. Knowledge-based (KB) planning is a data-driven approach to inverse optimization that has been shown to improve planning efficiency, consistency, and quality compared to conventional inverse planning. [23] [24] [25] [26] KB planning aims to generate the best plan based on historical, high-quality treatment plans. RapidPlan (Varian Medical Systems, Palo Alto, CA) is a commercial KB optimization engine that uses geometric and dosimetric features from a library of treatment plans to estimate the achievable range of OAR dose-volume histograms (DVHs) for new patients. 27, 28 From this prediction, dose-volume constraints and priorities are generated to drive optimization of the new plan. The goal is to produce a plan of similar quality to those used to train the RapidPlan model, with minimal human interaction. RapidPlan has demonstrated the ability to create improved or equivalent plans for WBI 24, 26 but there are no published experiences for APBI. Our institution's experience in developing a five-field IMRT class solution beam strategy 29 and more rigorous dosimetric constraints 30 compared to major phase III trials 5-7 was used to train and validate a RapidPlan model for left-and right-sided APBI. This study reports on the dosimetry achievable with large-scale retrospective application of this RapidPlan model, and compares the results to conventional clinical plans and dosimetric constraints from phase III APBI trials. The five-field IMRT class solution beam strategy 29 and this RapidPlan model are useful tools for institutions looking to offer APBI as a routine treatment option for appropriately selected early-stage breast cancer patients. A total of 184 early-stage breast cancer patients who received APBI (27 Gy in 5 fractions) as part of phase II prospective clinical trial 31 (the ACCEL trial) were selected for model training and validation. This study was determined to be of minimal risk and consistent with a quality improvement project using A pRoject Ethics Community Consensus Initiative (ARECCI) screening tool 32 provided by the Health Research Ethics Board of Alberta. A formal research ethics board review and approval were not required according to institutional mandate. Planning and treatment protocol details are published elsewhere. 29, 31 All patients underwent a free-breathing CT simulation (Philips Big Bore, Philips, Andover, MA) in the supine position on a standard wing board with both arms raised. CTs were acquired with 120 kVp, 200 mAs, and a slice thickness of 3 mm. The seroma was delineated by the treating radiation oncologist, based on the seroma/surgical cavity and surgical clips visible on the CT. The clinical target volume (CTV) is the seroma plus a margin of 10 mm, cropped to the chest wall and 5 mm inside the body contour. The CTV is expanded by 7 mm in all directions to create the planning target volume (PTV). A dose evaluation volume (DEV), used to evaluate target coverage,is defined as the PTV trimmed to the chest wall and 5 mm inside the body contour. Contoured OARs include ipsilateral and contralateral breast 29,31 , ipsilateral lung, and the heart. 33 A clinical IMRT plan was created for each patient by a certified dosimetrist using the five-field planning strategy outlined by Quirk et al. 29 , which achieves highly conformal dose distributions with improved OAR sparing. Key components of this planning strategy include: gantry and couch angle class solutions for left-and right-sided breast seroma locations, clearance charts of permissible gantry and couch angle combinations based on patient body habitus and seroma location, and rigorous dosimetry guidelines 29, 30 (Table 1) . All treatment plans used five 6 MV non-coplanar sliding window IMRT fields: four off -axis tangents (two medial and two lateral) and a field from an anterior oblique direction. Gantry and couch angles were selected according to Quirk et al.'s 29 class solution approach, which were adapted for patient body habitus and to maximize the angle between couch positions. Clearance charts for a TrueBeam linear accelerator (Varian Medical Systems, Palo Alto, CA) with Exact IGRT couch, detailing all possible gantry/couch combinations and account for seroma location and body habitus, were consulted to ensure a deliverable configuration. All patients were planned such that at least 98% of TA B L E 1 Planning constraints (minor variations) used in this study compared to those from phase III trials Heart (left-sided, lower inner quadrant) Heart (left-sided, other quadrants) Abbreviations: BID, twice daily; CI, conformity index, volume of tissue receiving at least 95% of the prescription dose divided by the volume of the PTV; DEV, dose evaluation volume; NSABP, National Surgical Adjuvant Breast and Bowel Project, RTOG, Radiation Therapy Oncology Group; PTV, planning target volume; RAPID, Randomized Trial of Accelerated Partial Breast Irradiation. the DEV received at least 95% of the prescription dose. All contoured volumes and clinical plans were peerreviewed per trial protocol. 31 The RapidPlan model configuration and evaluation processes are described in detail in the literature. 34, 35 The RapidPlan model was configured in the Eclipse treatment planning system (version 15.6.06) using 64 IMRT-based APBI plans. These plans were selected and verified to be high quality based on peer-review of contours and beam geometry, and meeting the planning constraints in Table 1 . These patients were also selected to represent a range of breast and target sizes, target positions within the breast, and body habitus (reported in Section 3; Table 4 ). RapidPlan model configuration consists of data extraction and model training. Geometric and dosimetric data were extracted from the training dataset, where OAR contours are divided into four subregions: targetoverlap, in-field, leaf -transmission, and out-of -field. 34, 35 For each OAR, a model is trained using a combination of principal component analysis and step-wise multiple regression for the in-field region, and a mean and standard deviation model of the DVH for the other OAR subregions. 34, 35 When applying the RapidPlan model to a new patient, their geometric data (target and OAR contours, and field geometry) are used to predict a DVH curve for each OAR subregion, which are added together and weighted by the corresponding relative volume of each subregion. In this study, the OARs trained were the ipsilateral and contralateral breast, ipsilateral lung, and heart. Within the RapidPlan workspace, a statistical summary describing the quality of the model is produced as Potentially influential cases were determined using Cook's distance, which measures the impact a single case has on the regression coefficients. Influential cases may not necessarily appear as an outlier in the regression and residual plots. A Cook's distance greater than 4 indicates an influential case that may be a geometric or dosimetric outlier. All cases identified as potential outliers were carefully reviewed and found to have anatomical differences with respect to the rest of the population in the training dataset. As a result, no patients were excluded from the training dataset to better capture the variation in patient anatomy. The iterative process described by Hussein et al. 25 was performed to refine the optimization objective parameters. Tables 2 and 3 Photon Optimizer were set to X = 50 and Y = 40. Target coverage, uniformity, and conformity were improved by increasing the priority of the upper and lower target objectives and changing the normal tissue objective to our local optimized settings. For OARs, line-type objectives with generated priorities are placed along the inferior border of the predicted DVH range. In some patients where the target was located in an inner breast quadrant, it was found that the heart and lung DVHs were overestimated by RapidPlan. Additional point objectives were added for these organs so that in the event of any cases where RapidPlan estimated a higher value for the line objective than the fixed point objectives, then the point objectives would take priority and achieve better sparing. In the following sections, the results refer to the final refined RapidPlan model. RapidPlan model performance was evaluated using 120 consecutively treated APBI patients that were not included in the training dataset. Herein, RapidPlanoptimized plans will be denoted as KB plans. The clinical and KB plans were generated with the same prescription dose, beam energy and geometry, and optimization (Photon Optimizer, version 15.6.06) and dose calculation algorithms (Anisotropic Analytical Algorithm, version 15.6.06). All dose calculations were performed on a 2.5 mm grid with heterogeneity corrections turned on. To facilitate dosimetric comparison, all plans were normalized such that 98% of the DEV received 95% of the prescription dose. The dosimetry of KB plans was compared to clinical plans, and both the clinical 30 and phase III trial 5-7 planning constraints. The conformity index (CI) was used to score the conformity of the high dose region to the target. The CI is defined as the volume of tissue receiving at least 95% of the prescription dose divided by the PTV volume. Dose homogeneity within the target was assessed using the near-maximum dose (D1cc) to the PTV. Differences between clinical and KB plans were The characteristics of patients included in the training and validation datasets are provided in Table 4 . Figure 1a ,b compares the PTV D1cc and CI, respectively, between clinical and KB plans in the validation dataset. A statistically significant improvement in PTV dose homogeneity was observed (median difference: 2.0%, range: −1.6%-7.6%, p < 0.05) for KB plans compared to clinical plans. All PTV D1cc values in the KB plans meet the planning constraint of D1cc < 107%. The PTV CI is comparable between clinical and KB plans (median difference: 0.0, range: −0.1-0.2, p = 0.05), and are within the planning constraints or minor variations. The ipsilateral breast volume receiving 50% and 95% of the prescription dose for clinical and KB plans are plotted in Figure 1c . All ipsilateral breast V50% and V95% values are within the planning constraints or minor variations. The KB plans provided a statistically significant improvement in the ipsilateral breast V50% (median difference: 3.2%; range: −1.2%-17.4%; p < 0.05) and V95% (median difference: 0.1%; range: −1.3%-3.3%; p < 0.05) compared to the clinical plans. The ipsilateral lung volume receiving 10% and 30% of the prescription dose for clinical and KB plans are plotted in Figure 1d . KB planning improved the lung V10% (median difference: 1.2%; range: −5.7%-5.9%; p < 0.05) and V30% (median difference: 0.4%; range: −2.3%-4.0%; p < 0.05) compared to conventional inverse planning for most patients. The median (range) contralateral breast V3% in the clinical and KB plans were 0.0% (0.0%-1.6%) and 0.0% (0.0%-1.9%), respectively (median difference: 0.0%, range: −0.3%-0.4%, p = 1.00). For all right-sided cases, the heart V5% was 0% in clinical and KB plans. For left-sided lower inner quadrant cases, the median (range) heart V15% in the clinical and KB plans were 0.0% (0.0%-3.5%) and 0.0% (0.0%-3.8%), respectively (median difference: 0.0%, range: −0.3%-1.3%, p = 0.44). For all other quadrants in left-sided cases, the median (range) heart V10% were 0.0% (0.0%-1.9%) and 0.0% (0.0%-1.1%) in the clinical and KB plans respectively (median difference: 0.0%, range: 0.0%-0.9%, p = 0.06). All clinical and KB plans met the planning constraints for the contralateral breast and heart, and differences were not statistically significant. Variation in the technical details, efficacy, and cosmesis results of phase III APBI trials have left the community with questions regarding the optimal prescription dose and fractionation, treatment modality, patient selection, and volume to irradiate. 16 Due to the confounding nature of these variables, the consequences of the selected dose, fractionation, and patient population will likely be clearer if dosimetry is well-controlled. In this study, our institutional experience in APBI was used to train and validate a RapidPlan model. The RapidPlan model is able to produce IMRT plans in a single optimization without manual intervention that are comparable or better than the conventionally optimized clinical plans. Table 1 ). Wilcoxon signed-rank tests show that differences between CP and KBP are statistically significant for the PTV D1cc; ipsilateral breast V50% and V95%; and ipsilateral lung V10% and V30% KB planning provided a statistically significant reduction in the PTV D1cc compared to conventional inverse planning. Previous studies have shown that the delivery of excessive radiation dose to regions within the breast is associated with acute and chronic toxicities and less optimal cosmetic results. 37, 38 The maximum PTV D1cc values in KB and clinical plans were 104.6% and 110.3%, respectively. The improvement in dose homogeneity with KB planning is unlikely to be clinically significant but KB planning consistently met the planning requirement of D1cc < 107%. 30 KB and conventional inverse planning achieved similar PTV CIs, with all plans fulfilling the planning constraints or minor variations. 30 The conformity of the KB and clinical plans was comparable and has previously been reported 30 to be superior to that accepted in recently published phase III APBI trials. 5, 6 Early results from the ACCEL trial report favorable toxicity and cosmesis for this planning strategy. 39 The difference in the ipsilateral breast V95% between KB and clinical plans was statistically significant, but small, with a median difference < 1%. Small differences in ipsilateral breast V95% between KB and clinical plans are expected because dose conformity for APBI has a stronger dependence on patient anatomy and the IMRT beam arrangement. 29 In contrast, the ipsilateral breast V50% was significantly reduced in the KB plans, with a maximum difference of 17.4%. Several studies have investigated the relationship between ipsilateral breast dose and normal tissue toxicity or cosmesis after APBI. 13, 14 Statistically significant associations have been demonstrated between the ipsilateral breast V50% and the risk of grade 2-4 subcutaneous fibrosis, fat necrosis, and fair/poor cosmetic outcome. 13, 14 These relationships suggest that stricter dose constraints for the ipsilateral breast may be appropriate and can consistently be achieved by implementing the RapidPlan model. Early-stage breast cancer patients have excellent long-term breast cancer-specific survival and it is critical to minimize normal tissue doses to reduce the risk of long-term toxicities. Heart and contralateral breast doses in KB plans were not significantly different compared to clinical plans, and the ipsilateral lung dose was lower in KB plans for most patients. In general, APBI provides very low normal tissue doses and dose-volume parameters for the ipsilateral lung, heart, and contralateral breast are substantially lower than in WBI 29 or QUANTEC recommendations. 40, 41 The large-scale retrospective application of our RapidPlan model for IMRT-based APBI has demonstrated the model's validity for clinical implementation. This RapidPlan model has several potential applications. RapidPlan models can be shared with other institutions, providing an opportunity to transfer planning expertise from our more experienced center to less experienced centers looking to adopt APBI. 26, 29, 30, 42 This RapidPlan model could be supplied to participating centers for future multi-institutional clinical trials to standardize the planning technique and dosimetric criteria 29, 30 so that they are better positioned to answer remaining questions regarding optimal prescription dose, fractionation, and patient selection. Alternatively, the RapidPlan model could be used as a tool for planning quality assurance in clinical trial treatment plan audit/credentialing and to benchmark different techniques. [42] [43] [44] Tol et al. 42 proposed a workflow utilizing RapidPlan to quickly evaluate whether a treatment plan submitted to a clinical trial provides sufficient OAR sparing by comparing the submitted plan OAR DVHs to those predicted by RapidPlan. Overall, clinical implementation of this RapidPlan model for treatment planning or as a quality assurance tool provides the opportunity to standardize plan quality and could benefit any future phase III APBI trials looking to address the shortcomings of previous trials. A RapidPlan model was successfully trained and validated for left-and right-sided APBI planning. Model validation results show an improvement in plan quality and consistency for KB plans compared to conventionally optimized clinical plans. These results suggest that this model can be reliably applied in clinical practice and may be used to transfer planning expertise to ensure consistency in APBI plan quality. A. Frederick received postgraduate scholarships from the Natural Sciences and Engineering Research Council of Canada (NSERC) and Alberta Innovates. The authors declare no conflict of interest. None. Amy Frederick and Sarah Quirk conceived the project. Amy Frederick developed and validated the RapidPlan model, and collected and analyzed the data. Michael Roumeliotis, Petra Grendarova, and Sarah Quirk provided clinical expertise and supervision of the project. Amy Frederick drafted the manuscript and all co-authors revised and approved the final manuscript. The RapidPlan model and relevant data are available from the authors upon reasonable request. S. Quirk can be contacted to request data. O R C I D Amy Frederick https://orcid.org/0000-0003-2171-032X The UK Standardisation of Breast Radiotherapy (START) Trial B of radiotherapy hypofractionation for treatment of early breast cancer: a randomised trial Long-term results of hypofractionated radiation therapy for breast cancer Ten-year results of fast: a randomized controlled trial of 5-fraction wholebreast radiotherapy for early breast cancer Hypofractionated breast radiotherapy for 1 week versus 3 weeks (FAST-Forward): 5-year efficacy and late normal tissue effects results from a multicentre, non-inferiority, randomised, phase 3 trial Long-term primary results of accelerated partial breast irradiation after breast-conserving surgery for early-stage breast cancer: a randomised, phase 3, equivalence trial External beam accelerated partial breast irradiation versus whole breast irradiation after breast conserving surgery in women with ductal carcinoma in situ and node-negative breast cancer (RAPID): a randomised controlled trial Accelerated partial-breast irradiation compared with whole-breast irradiation for early breast cancer: long-term results of the randomized phase III APBI-IMRT-florence trial Radiation therapy for the whole breast: executive summary of an American Society for Radiation Oncology (ASTRO) evidence-based guideline Adapting radiation therapy treatments for patients with breast cancer during the COVID-19 pandemic: hypo-fractionation and accelerated partial breast irradiation to address world health organization recommendations Five-fraction radiotherapy for breast cancer: FAST-forward to implementation Accelerated partial breast radiation: information on dose, volume, fractionation, and efficacy from randomized trials Interim cosmetic and toxicity results from RAPID: a randomized trial of accelerated partial breast irradiation using three-dimensional conformal external beam radiation therapy The effect of dose-volume parameters and interfraction interval on cosmetic outcome and toxicity after 3-dimensional conformal accelerated partial breast irradiation Unacceptable cosmesis in a protocol investigating intensity-modulated radiotherapy with active breathing control for accelerated partial-breast irradiation Accelerated partial breast irradiation: executive summary for the update of an ASTRO evidence-based consensus statement Accelerated partial breast irradiation: more questions than answers? Dosimetric comparison of four different external beam partial breast irradiation techniques: three-dimensional conformal radiotherapy, intensity-modulated radiotherapy, helical tomotherapy, and proton beam therapy Dosimetric comparison of 3D conformal, IMRT, and V-MAT techniques for accelerated partial-breast irradiation (APBI) Secondary malignancies after partial versus whole breast irradiation: a systematic review and meta-analysis Long-term risks of secondary cancer for various whole and partial breast irradiation techniques The physical basis of IMRT and inverse planning Accelerated partial breast irradiation using 3D conformal radiation therapy (3D-CRT) Knowledge-based planning for intensity-modulated radiation therapy: a review of data-driven approaches Performance of a knowledge-based model for optimization of volumetric modulated arc therapy plans for single and bilateral breast irradiation Clinical validation and benchmarking of knowledge-based IMRT and VMAT treatment planning in pelvic anatomy Is it possible for knowledge-based planning to improve intensity modulated radiation therapy plan quality for planners with different planning experiences in leftsided breast cancer patients? Quantitative analysis of the factors which affect the interpatient organ-at-risk dose sparing variation in IMRT plans Predicting dose-volume histograms for organs-at-risk in IMRT planning Five-field IMRT class solutions and dosimetric planning guidelines for implementing accelerated partial breast irradiation Results of the ACCEL trial: dosimetry in accelerated partial breast irradiation A phase II trial to determine the cosmetic outcomes and toxicity of 27 Gy in five-fraction accelerated partial breast irradiation:the ACCEL trial Hayward S. Innovations in the ethical review of health-related quality improvement and research: the Alberta Research Ethics Community Consensus Initiative (ARECCI) Development and validation of a heart atlas to study cardiac exposure to radiation following treatment for breast cancer On the pre-clinical validation of a commercial model-based optimisation engine: application to volumetric modulated arc therapy for patients with lung or prostate cancer Creation of knowledge-based planning models intended for large scale distribution: minimizing the effect of outlier plans SciPy 1.0: fundamental algorithms for scientific computing in Python Optimizing breast cancer treatment efficacy with intensity-modulated radiotherapy A multicenter randomized trial of breast intensity-modulated radiation therapy to reduce acute radiation dermatitis One-year cosmesis and fibrosis from ACCEL: accelerated partial breast irradiation (APBI) using 27 Gy in 5 daily fractions Radiation dosevolume effects in the heart Radiation dose-volume effects in the lung Can knowledge-based DVH predictions be used for automated, individualized quality assurance of radiotherapy treatment plans? Does quality of radiation therapy predict outcomes of multicenter cooperative group trials? A literature review Applying a RapidPlan model trained on a technique and orientation to another: a feasibility and dosimetric evaluation