key: cord-0694192-3qcrbbyk authors: Datta, Pratik; Ukey, Rahul; Bruiners, Natalie; Honnen, William; Carayannopoulos, Mary O.; Reichman, Charles; Choudhary, Alok; Onyuka, Alberta; Handler, Deborah; Guerrini, Valentina; Mishra, Pankaj K.; Dewald, Hannah K.; Lardizabal, Alfred; Lederer, Leeba; Leiser, Aliza L.; Hussain, Sabiha; Jagpal, Sugeet K.; Radbel, Jared; Bhowmick, Tanaya; Horton, Daniel B.; Barrett, Emily S.; Xie, Yingda L.; Fitzgerald-Bocarsly, Patricia; Weiss, Stanley H.; Woortman, Melissa; Parmar, Heta; Roy, Jason; Dominguez-Bello, Maria Gloria; Blaser, Martin J.; Carson, Jeffrey L.; Panettieri, Reynold A.; Libutti, Steven K.; Raymond, Henry F.; Pinter, Abraham; Gennaro, Maria Laura title: Highly versatile antibody binding assay for the detection of SARS-CoV-2 infection and vaccination date: 2021-10-09 journal: J Immunol Methods DOI: 10.1016/j.jim.2021.113165 sha: ef3b4ce77e005e682a9fd55fcffc1c624a33ce2f doc_id: 694192 cord_uid: 3qcrbbyk Monitoring the burden and spread of infection with the new coronavirus SARS-CoV-2, whether within small communities or in large geographical settings, is of paramount importance for public health purposes. Serology, which detects the host antibody response to the infection, is the most appropriate tool for this task, since virus-derived markers are most reliably detected during the acute phase of infection. Here we show that our ELISA protocol, which is based on antibody binding to the Receptor Binding Domain (RBD) of the S1 subunit of the viral Spike protein expressed as a novel fusion protein, detects antibody responses to SARS-CoV-2 infection and vaccination. We also show that our ELISA is accurate and versatile. It compares favorably with commercial assays widely used in clinical practice to determine exposure to SARS-CoV-2. Moreover, our protocol accommodates use of various blood- and non-blood-derived biospecimens, such as breast milk, as well as dried blood obtained with microsampling cartridges that are appropriate for remote collection. As a result, our RBD-based ELISA protocols are well suited for seroepidemiology and other large-scale studies requiring parsimonious sample collection outside of healthcare settings. Diagnosis of infection with the novel coronavirus SARS-CoV-2, the causative agent of the ongoing COVID-19 pandemic, has relied on two classes of assays. One comprises the methods for detecting the presence of the virus in upper respiratory specimens, either by viral nucleic acid amplification tests (NAAT) or immunodetection of viral antigen. NAATs based on Real-time PCR represent the gold standard for diagnosis of acute SARS-CoV-2 infection while the antigen tests, which are comparatively less sensitive, are critically important for public health purposes, since they have a very rapid turn-around and detect infectious cases (Mallett et al., 2020; Mina et al., 2020; Ravi et al., 2020; van Kasteren et al., 2020; Mina and Andersen, 2021; Yüce et al., 2021) . The second class of assays comprises methods for detecting virus-specific antibodies in peripheral blood. These antibodies are reliable indicators of viral exposure, since they become detectable approximately two weeks after initiation of productive infection and typically persist for 6-12 months or longer, well beyond the time in which virus detection assays return to negativity (Fig. 1) . Thus, antibody-based assays are most valuable as metrics of infection burden in the population for epidemiological purposes and largescale studies. Antibody-based assays for SARS-CoV-2 infection are based on two SARS-CoV-2 antigens. One is Spike (S), a two-subunit protein that decorates the surface of the virion and establishes contact with the host cell receptor, angiotensin-converting enzyme 2 (ACE2), through the receptor-binding domain (RBD) in the S1 subunit, thus determining J o u r n a l P r e -p r o o f Journal Pre-proof host range and tissue tropism (Li, 2016) . The second viral antigen is the Nucleocapsid (N), which interacts with the viral genomic RNA inside the viral envelope. Both antigens have been used for SARS-CoV-2 antibody detection, with an initial preference for the N antigen in most commercial antibody detection assays utilized in clinical settings (for example, (Kohmer et al., 2020; Padoan et al., 2020) ). The S protein has been adopted as antibody capture antigen in research settings since the beginning of the pandemic ]e.g., (Amanat et al., 2020; Robbiani et al., 2020; Gaebler et al., 2021) ], primarily because the S1 RBD region is particularly immunogenic and the dominant target of neutralizing (protective) antibodies (Ju et al., 2020; Robbiani et al., 2020; Shi et al., 2020) . Moreover, mutations in RBD, which are important factors in the evolution of all major SARS-CoV-2 variants, increase affinity for the ACE-2 receptor and lead to resistance to monoclonal and polyclonal antibodies developed in response to infection or vaccination (Starr et al.; Greaney et al.; Greaney et al.) . More recently, use of S1 RBD for antibody testing has been extended to commercial assays and clinical applications (see list of emergency use authorized serology tests at fda.gov/medicaldevices) since the introduction of SARS-CoV-2 vaccines, which contain S but not N (Jackson et al., 2020; Polack et al., 2020) . Thus, it becomes increasingly important to identify all potential uses of S-based serological assays for SARS-CoV-2 infection. Here we describe key characteristics of our serological assay utilizing a novel S1 RBD antigen and its suitability for antibody detection from minimal (l scale) amounts of remotely collected peripheral blood, which is critical for seroepidemiological and largescale studies conducted outside of health care settings. We also show that the assay is equally suited for detecting antibodies in different liquid compartments of peripheral blood and other bodily fluids, making it adaptable to diverse study designs. The analyses presented in the present work draw upon data and biospecimens gathered during seven studies of COVID-19 in NJ, USA. All participants were enrolled after written informed consent was obtained from each participant after the nature and possible consequences of the study had been fully explained. All study subjects were >20 years of age. To assess antibody responses to SARS-CoV-2 infection we used plasma/serum samples obtained from 83 SARS-CoV-2 PCR-confirmed convalescent subjects (Mishra et al., 2021) ; 146 patients hospitalized for at Robert Wood Johnson University Hospital in New Brunswick, NJ; blood collected after >2 weeks from completion of full vaccination from 283 subjects vaccinated against SARS-CoV-2 between mid-December 2020 and mid-February 2021 among healthcare workers in Rutgers-affiliated hospitals (Barrett et al., 2020) and Rutgers employees (Mishra et al., 2021) ; and 148 residents living in the township of Lakewood, NJ in April 2020. Studies were approved by the Research Subjects Institutional Review Board at The reference protein was produced from the vector pCAGGS containing the SARS-Related Coronavirus 2, Wuhan-Hu-1 Spike Glycoprotein Receptor Binding Domain (RBD) (Catalog No. NR-52309, BEI Resources, Manassas, VA), utilizing the manufacturer's protocol. Spike Protein S1 from non-SARS-CoV-2 coronaviruses (HCoV-229E, HCoV-NL63, HCoV-HKU-1) and Spike Protein S1 and S2 extracellular domain (HCoV-OC43) were obtained from Sino Biologicals (Wayne, PA, USA) and pooled in equimolar amounts to a final concentration of 1 mg/ml. The resulting pool was used at 2 µg/ml (50 l per well) to coat the ELISA plates. 96-well ELISA plates (Nunc MaxiSorp, ThermoFisher, Rochester, NY) were coated with 2 µg/ml recombinant SARS-CoV-2 RBD (50 l per well) overnight at 4°C. Plates were washed four times with 100 l/well washing buffer (1x PBS containing 0.05% Tween 20) (Sigma-Aldrich, St. Louis, MO) and blocked with 100 l/well blocking buffer [2% Blotto (Nestle Carnation, US) in PBS] for 30 min at 37°C. Diluted plasma/serum (1:1 in 1x PBS) was heat-inactivated at 56°C for 1 hour prior to use. After blocking, plates were washed four times with 100 l /well washing buffer, and 50 l plasma/serum diluted in blocking buffer was added to each well and incubated for 1 hour at 37°C. For matrix equivalence studies, serum was diluted in the test matrices (breast milk or plasma obtained from blood collected in various anticoagulant tubes), as described in Results. Bound IgG was detected by adding alkaline phosphatase-conjugated goat anti-human IgG (Jackson ImmunoResearch, West Grove, PA) diluted 1:2,000 in blocking buffer (50 l/well). Enzyme activity was assayed by adding 50 l/well phosphate substrate (Sigma-Aldrich, St. Louis, MO) solubilized in developing buffer (2:1 diethanolamine:MgCl2.6H 2 O (Sigma-Aldrich), pH 9.8). The reaction was stopped after 30 min with 1M NaOH (50 l/well) and results were read as absorbance (OD 405 ) values. Each ELISA plate contained positive and negative serum/plasma controls and background control wells without primary antibody, and each sample was tested in duplicate. The protocol was automated, using a Hamilton Microlab STAR liquid handler (Hamilton Company, Reno, NV) for sample handling and dilution, and a BioTek EL406 combination washer dispenser and a Synergy Neo2 microplate reader (BioTek, Winooski, VT) for ELISA. Work involving blood products from SARS-CoV-2-infected subjects was performed in a biosafety level 2+ (BSL-2+) laboratory utilizing protocols approved by the Rutgers Institutional Biosafety Committee. The Roche Elecsys ® Anti-SARS-CoV-2 assay utilizing the Roche Cobas e601 instrument and the Abbott Architect SARS-CoV-2 IgG assay utilizing the Abbott Architect c4000, which both use SARS-CoV-2 N protein as capture antigen, were performed by specialized personnel following the manufacturer's instructions. For phlebotomy, standard venipuncture was performed, and 10 mL of blood was collected in a serum separator tube with inert clot activator (BD367861, Franklin Lakes, NJ) or in a tube containing the anticoagulant sodium heparin (BD366480). For the matrix equivalence study, tubes containing other anticoagulants [potassium/EDTA (BD367861), lithium heparin (BD367960), or sodium citrate (BD363083)] were also used. Serum tubes were maintained in an upright position at 4°C for 1-2 hr to allow for coagulation prior to centrifugation. For plasma separation, blood samples were processed within 2-6 hours after collection. Plasma and serum samples were centrifuged in a swinging bucket rotor at 1,260 xg for 100 minutes at room temperature with low acceleration and no brake. Plasma phase was aspirated carefully from the top. Sera and plasma were sub aliquoted into cryo-vials and stored at −80°C. Volumetric absorptive microsampling (VAMS) (Mitra Collection Kit; Neoteryx, CA) was performed following the manufacturer's instructions. Prior to sample collection, the lateral portion of the participant's finger was cleaned with an alcohol swab and punctured with a lancet device provided in the kit. A hydrophilic 30-μL VAMS microsampler was held against the blood drop until filled. Two microsamplers were utilized per subject. Blood-filled microsamplers were returned to the protective cartridges, which were placed in sealed containers with silica desiccant packets and stored at room temperature for up to 2 weeks from the collection date. One microsampler tip (30 μl) was added to 300 μl VAMS buffer (1x PBS) (Corning, Manassas, VA), supplemented with 1% bovine serum albumin (Roche Diagnostics, J o u r n a l P r e -p r o o f Journal Pre-proof Mannheim, Germany) and 0.5% Tween 20 (Sigma, MO) in a 1 ml-deep 96-well plate (Greiner Bio-One, Monroe, NC). The plate was covered with an adhesive seal and maintained shaking at 250 rpm for 16 hours at 4°C. The resulting eluates were heatinactivated at 56°C for 60 minutes and clarified by centrifugation at 3,500 rpm for 5 minutes. Supernatants were used for ELISA or aliquoted and stored at -80°C. The SARS-CoV-2 S1 RBD antigen used in this study. The assays described in this report were performed using a novel gp70-fusion protein form of the S1 RBD antigen. The gp70 domain possesses chaperone-like qualities, and this fusion protein system has been shown to facilitate the correct folding and glycosylation of conformational subdomains of the HIV-1 gp120 glycoproteins and to efficiently express epitopes recognized by HIV-1 patient sera that are dependent on native structures (Kayman et al.; Wu et al.; Krachmarov et al.) . The structure and properties of the gp70-RBD antigen are described in Fig. 2 . The gp70 carrier domain has a His8 affinity tag inserted near its N-terminus to facilitate purification and an HRV-3C protease cleavage signal (LEVLFQGP with a GS linker) inserted before the RBD sequence (aa 316-544 of the Wuhan sequence) to allow cleavage and removal of the carrier domain, if desired ( Fig. 2A) . The purity of the intact fusion protein and the isolated RBD domain is shown in Fig. 2B (the reference RBD antigen, which was utilized in the first published SARS-CoV-2 ELISA protocol (Amanat et al., 2020) , contains a His6 affinity tag and thus J o u r n a l P r e -p r o o f Journal Pre-proof appears slightly larger than the cleaved product; lanes 3 and 4). When we compared binding curves obtained utilizing equal protein concentrations of gp70-RBD fusion protein and reference RBD antigen against two convalescent sera (Fig. 2C) , we observed that, despite its larger molecular weight, the fusion protein yielded a more sensitive signal than the reference RBD protein with both sera, under all antigen concentrations and serum dilutions tested. This result is likely due to more efficient binding of the antigen to the ELISA plate wells and better exposure of RBD epitopes resulting from the presence of the gp70 tag. Characteristics of the S1 RBD-based antibody binding assay. The limited sequence conservation between the SARS-CoV-2 S1 RBD with that produced by non-pathogenic human coronaviruses (Li, 2016) is expected to minimize the potential detection of crossreactive antibodies. Indeed, pre-COVID-19 sera did not react with SARS-CoV-2 S1 RBD but reacted with a mixture of non-SARS-CoV-2 coronavirus N antigens, presumably due to exposure to non-pathogenic human coronaviruses (Fig. 3A) . We next compared detection of virus-specific antibody responses to SARS-CoV-2 infection by our RBD-based ELISA vis-à-vis two assays (Roche Elecsys Anti-SARS-CoV-2, Basel, Switzerland, and Abbott Architect SARS-CoV-2 IgG assay, Chicago, IL, USA) that use SARS-CoV-2 N antigen as the capture reagent, and the Roche Elecsys Anti-SARS-CoV-2 S, Basel, Switzerland, which is Spike-based. All three commercial assays For the commercial assays, cut-off values for positive or negative test determination were as established by the manufacturers. For our ELISA, the cut-off was established as the mean OD 405 + 3SD value obtained with independent samples from SARS-CoV-2 PCR-negative subjects (n=103) that remained SARS-CoV-2 PCR-negative for at least 16 weeks (Barrett et al., 2020 ) (see next section). The results of the four parallel assays are shown in Fig. 3B . While the sample size was somewhat limited by the requirement of the commercial assays for relatively large volumes (typically >100 l per assay) to accommodate the dead volume of the system, the results suggest that our RBD-based ELISA has excellent sensitivity and specificity (Fig. 3B ) and compares favorably with commercial assays that are widely utilized in clinical settings (R 2 = 0.97 by Spearman correlation of the comparison between our in-house ELISA and the Roche Spike-based assay). Applications of the S1 RBD-based antibody binding assay. The high accuracy of our RBD-based ELISA makes it an ideal tool to study the antibody response to SARS-CoV-2 infection and vaccination and for seroepidemiological purposes. As a proof of principle, our assay was used for the detection of RBD-specific IgG antibodies in sera collected from convalescent subjects within nine months since the first positive SARS-CoV-2 PCR test (n=83), hospitalized COVID-19 patients during the first two weeks of hospitalization (n=146), and subjects fully vaccinated with SARS-CoV-2 RNA vaccines within five months post-vaccination (n=283) (Fig. 4) . The overall higher reactivity of the convalescent group relative to the hospitalized patients is presumably due to a larger J o u r n a l P r e -p r o o f Journal Pre-proof proportion of recently infected subjects in the latter group who may not have seroconverted. As expected (Goel et al., 2021; Mishra et al., 2021) , the antibody response to mRNA vaccination was generally stronger than that to natural infection. Moreover, no antibodies were detected in the negative control subjects (n=104) and SARS-CoV-2 PCR-negative subjects (n=103) that remained uninfected for at least 16 weeks after the blood draw tested in the assay (Barrett et al., 2020) ] (Fig. 4) . However, studies conducted outside of healthcare settings, such as seroprevalence studies, require blood draws by non-specialized personnel or even by the study subjects themselves, possibly at remote sites. One such procedure involves microsampling by appropriate tool for this task, since virus-derived markers are not reliably detected outside of the acute phase of the infection. Here we show that our S1 RBD-based ELISA is well suited to detect the antibody response to SARS-CoV-2 infection and vaccination. We also provide a proof-of-principle demonstration of the value of SARS-CoV-2 seroepidemiology, since our assay can identify individuals who were infected with SARS-CoV-2 among those exposed or perceived to have been exposed because they live in a high-burden area. Thus, notwithstanding immune impairment limiting the sensitivity of the immunoassay, seroepidemiology has the potential to yield more accurate estimates of prevalence of infection than epidemiological tools based on clinical symptomatology or reported exposure. We also demonstrate that our ELISA is accurate, versatile, and highly suited for research and clinical applications. Our protocol is performed utilizing robotic sample handling and dilution and automated ELISA. Moreover, it compares favorably with accurate commercial tests that have been widely used in clinical practice to determine exposure to SARS-CoV-2. Furthermore, our protocol accommodates use of various blood-and non-blood-derived biospecimens as well as dried blood obtained with microsampling cartridges that are appropriate for remote sampling and transportation. Additionally, the suitability for samples obtained in small volumes constitutes a distinct advantage of our in-house, fully automated ELISA over commercial assays widely used in clinical settings that, as mentioned above, require relatively large volumes (typically >100 l per assay) to accommodate the dead volume of the system. Thus, our RBD- Cut-off values for the commercial assays were as per manufacturer's instructions. For the in-house ELISA, the cut-off value (OD 405 = 0.3) was calculated as the mean + 3 SD obtained with sera from 103 SARS-CoV-2 PCR-negative subjects who remained negative for at least 16 weeks after the blood draw utilized in the assay (data shown in Fig. 4) . It is noted that, for clarity purposes, a single scale (Index on the right y axis) was used for both commercial assays. However, the index calculation is different in the two assays; therefore, the relative numbers cannot be compared across assays. CI, confidence interval. A serological assay to detect SARS-CoV-2 seroconversion in humans Prevalence of SARS-CoV-2 infection in previously undiagnosed health care workers in New Jersey SARS-CoV-2 RNA and antibody detection in breast milk from a prospective multicentre study in Spain Antibodies in the breast milk of a maternal woman with COVID-19 Distinct antibody and memory B cell responses in SARS-CoV-2 naïve and recovered individuals following mRNA vaccination Coronavirus disease 2019 vaccine response in pregnant and lactating women: a cohort study Comprehensive mapping of mutations in the SARS-CoV-2 receptor-binding domain that affect recognition by polyclonal human plasma antibodies Antibodies elicited by mRNA-1273 vaccination bind more broadly to the receptor binding domain than do those from SARS-CoV-2 infection COVID-19 Vaccine immunodeficiency virus type 1 clade a and clade B v3 domains are common in patient sera from Cameroon, but their neutralization activity is usually restricted by epitope masking Structure, Function, and Evolution of Coronavirus Spike Proteins At what times during infection is SARS-CoV-2 detectable and no longer detectable using RT-PCR-based tests? A systematic review of individual participant data COVID-19 testing: One size does not fit all Rethinking Covid-19 Test Sensitivity -A Strategy for Containment Vaccination boosts protective responses and counters SARS-CoV-2 Analytical and clinical performances of five immunoassays for the detection of SARS-CoV-2 antibodies in comparison with neutralization activity Safety and Efficacy of the BNT162b2 mRNA Covid-19 Vaccine Diagnostics for SARS-CoV-2 detection: A comprehensive review of the FDA-EUA COVID-19 testing landscape Convergent antibody responses to SARS-CoV-2 in convalescent individuals A human neutralizing antibody targets the receptor-binding site of SARS-CoV-2 Deep Mutational Scanning of SARS-CoV-2 Receptor Binding Domain Reveals Constraints on Folding and ACE2 Binding Comparison of seven commercial RT-PCR diagnostic kits for COVID-19 Application of volumetric absorptive microsampling (VAMS) to measure multidimensional antiinfluenza IgG antibodies by the mPlex-Flu assay Characterization of neutralization epitopes in the V2 region of human immunodeficiency virus type 1 gp120: role of glycosylation in the correct folding of the V1/V2 domain We thank the PHRI biosafety officers and the RBHS Institutional Biosafety committee for fast-track review and approval of laboratory protocols and practices related to handling of SARS-CoV-2 and infected biospecimens; the Rutgers Institutional Review Board for timely review and approval of COVID-19-related protocols for human subject protection; Daniel Fine and Steven Libutti for supporting the start of our COVID-19 work;and Nancy Reilly and the entire Rutgers Corona Cohort team that established a cohort from which some of the serum samples utilized in this study were obtained. This work was funded by National Institutes of Health grants R01 HL149450, R01 HL149450-S1, U01 AI122285-S1, P30 ES005022, K23 AR070286, and UL1 TR003017.Role of the funding source. The funding sources have no involvement in study design, data analysis, and preparation and submission of the manuscript.