key: cord-0691635-qbe1zgrm authors: Khalid, Muhammad Fazli; Selvam, Kasturi; Jeffry, Alfeq Jazree Nashru; Salmi, Mohamad Fazrul; Najib, Mohamad Ahmad; Norhayati, Mohd Noor; Aziah, Ismail title: Performance of Rapid Antigen Tests for COVID-19 Diagnosis: A Systematic Review and Meta-Analysis date: 2022-01-04 journal: Diagnostics (Basel) DOI: 10.3390/diagnostics12010110 sha: 9b882a0360b449461060a97b4aae5b9cf33a1534 doc_id: 691635 cord_uid: qbe1zgrm The identification of viral RNA using reverse transcription quantitative polymerase chain reaction (RT-qPCR) is the gold standard for identifying an infection caused by SARS-CoV-2. The limitations of RT-qPCR such as requirement of expensive instruments, trained staff and laboratory facilities led to development of rapid antigen tests (RATs). The performance of RATs has been widely evaluated and found to be varied in different settings. The present systematic review aims to evaluate the pooled sensitivity and specificity of the commercially available RATs. This review was registered on PROSPERO (registration number: CRD42021278105). Literature search was performed through PubMed, Embase and Cochrane COVID-19 Study Register to search studies published up to 26 August 2021. The overall pooled sensitivity and specificity of RATs and subgroup analyses were calculated. Quality Assessment of Diagnostic Accuracy Studies 2 (QUADAS-2) was used to assess the risk of bias in each study. The overall pooled sensitivity and specificity of RATs were 70% (95% CI: 69–71) and 98% (95% CI: 98–98), respectively. In subgroup analyses, nasal swabs showed the highest sensitivity of 83% (95% CI: 80–86) followed by nasopharyngeal swabs 71% (95% CI: 70–72), throat swabs 69% (95% CI: 63–75) and saliva 68% (95% CI: 59–77). Samples from symptomatic patients showed a higher sensitivity of 82% (95% CI: 82–82) as compared to asymptomatic patients at 68% (95% CI: 65–71), while a cycle threshold (Ct) value ≤25 showed a higher sensitivity of 96% (95% CI: 95–97) as compared to higher Ct value. Although the sensitivity of RATs needs to be enhanced, it may still be a viable option in places where laboratory facilities are lacking for diagnostic purposes in the early phase of disease. Coronavirus disease 2019 (COVID- 19) , an infectious disease caused by severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), has evolved into a global pandemic and is still a major health concern around the world. The initial outbreak of SARS-CoV-2 started in Wuhan, China, in December 2019 [1, 2] . As of 17 August 2021, SARS-CoV-2 infected over 211 million people and killed over 4.4 million people worldwide. Those infected with the SARS-CoV-2 experience a variety of symptoms, including fever, cough, exhaustion, shortness of breath, headache, sore throat, and loss of smell and taste [3] [4] [5] . The symptoms develop normally after 2 days to 2 weeks and last up to 3 weeks or longer for patients with mild to severe COVID-19 infection [1] . The SARS-CoV-2 infection can be divided into five stages. These stages are asymptomatic, mild, moderate, severe, and critical. The peer reviewed articles were searched on 26 August 2021 without restricting the publication date. The articles were searched through PubMed, Embase and Cochrane COVID-19 Study Register by using several keywords with the following search string: Three authors (M.F.K., A.J.N.J. and M.F.S.) individually extracted the number of true positives (TP), true negatives (TN), false positives (FP) and false negatives (FN) from each study and entered them into an Excel datasheet. Disagreeable findings were discussed, and when in doubt, the authors sought verification. The sensitivity and specificity for each of the studied antigen were calculated using RT-qPCR as the reference standard. The number of true positive outcomes was divided by the sum of true positive and false negative outcomes to calculate the sensitivity (1) . The number of true negative outcomes was divided by the sum of true negative and false positive outcomes to calculate the specificity (2) . Forest plots were used to demonstrate the comparative performance of the rapid antigen tests: Speci f icity = TN TN + FP (2) The quality of each study was assessed using Quality Assessment of Diagnostic Accuracy Studies 2 (QUADAS-2) tool [19] . Patient selection, index test, reference standard and flow and timing are the four core domains of the QUADAS-2 tool Table 1 ). The risk of bias was classified as low, high or unclear for each domain. Five authors (M.F.K., A.J.N.J., M.F.S., M.A.N. and K.S.) independently performed the assessment to judge the quality of each study. Disagreements among the authors were resolved by discussion. Table 1 . QUADAS-2 risk of bias assessment criteria. Patient selection Patient enrolment strategy is specified and free of bias. A case-control design and inappropriate exclusions are avoided. The index test results are interpreted without knowledge of the results of the reference standard. The conduct or interpretation of the index test does not introduce bias. The reference standard correctly classifies the target condition. The reference standard results are interpreted without knowledge of the results of the index test. The reference standard, its conduct or its interpretation do not introduce bias. There is an appropriate interval between the index test(s) and reference standard. All patients receive the same reference standard. All patients included in the analysis and patient flow do not introduce bias. A total of 1732 articles were identified through databases and register searching ( Figure 1 ). Of these, 271 duplicates were identified and removed using data tool in Excel. After screening the titles, 961 articles that were not primary research articles and unrelated to antigen-based detection for COVID-19 were excluded. Of the 500 abstracts screened, 369 articles that did not meet the basis of selection criteria were excluded. After screening the full text articles, 37 articles were excluded. The remaining 94 studies that fulfilled our selection criteria were included in this systematic review. The index test results are interpreted without knowledge of the results of the reference standard. The conduct or interpretation of the index test does not introduce bias. The reference standard correctly classifies the target condition. The reference standard results are interpreted without knowledge of the results of the index test. The reference standard, its conduct or its interpretation do not introduce bias. There is an appropriate interval between the index test(s) and reference standard. All patients receive the same reference standard. All patients included in the analysis and patient flow do not introduce bias. A total of 1732 articles were identified through databases and register searching (Figure 1 ). Of these, 271 duplicates were identified and removed using data tool in Excel. After screening the titles, 961 articles that were not primary research articles and unrelated to antigen-based detection for COVID-19 were excluded. Of the 500 abstracts screened, 369 articles that did not meet the basis of selection criteria were excluded. After screening the full text articles, 37 articles were excluded. The remaining 94 studies that fulfilled our selection criteria were included in this systematic review. The characteristics of the 94 studies reporting the performance of rapid antigen tests (RATs) for SARS-CoV-2 were summarized in Table 2 . Most of the studies (n = 84) were published in 2021, and the remaining studies were published in 2020. The majority of the studies employed clinical samples from European countries which include Germany (n = 14), Spain (n = 11), Italy (n = 10), the Netherlands (n = 7), France (n = 6), Belgium (n = 5) and Switzerland (n = 3). In Asia, most studies were conducted in Japan (n = 7), followed by China (n = 3) and India (n = 2). The United States (n = 8) and Chile (n = 3) were among the American countries reported the performance of RAT for SARS-CoV-2. A total of 25 studies used Panbio™ COVID-19 Ag RDT (Abbott, Jena, Germany), 11 used STANDARD Q COVID-19 Ag Home Test (SD Biosensor, Seoul, South Korea), 10 used SARS-CoV-2 Rapid Ag Test (Roche, Basel, Switzerland) and 9 used Lumipulse G SARS-CoV-2 Ag (Fujirebio, Tokyo, Japan). Most of the studies (n = 71) used nasopharyngeal swabs, 12 used pools of nasopharyngeal and throat swabs, 4 used nasal swabs, 3 used saliva, 2 used throat swabs, 1 study used a pool of nasopharyngeal and nasal swabs and 1 study compared the performance of nasopharyngeal swabs, saliva and sputum. The QUADAS checklist was completed for all included studies (Supplementary Material Table S1 ). The QUADAS-2 criteria for 94 studies included in this systematic review were presented in Figure 2 . Majority of the included studies (83%) have low risk of patient selection bias. Twelve (13%) studies have high risk of patient selection bias due to the case-control study design. These studies specifically recruited clinical samples known to be uninfected or infected with coronavirus. The remaining studies (4%) have unclear risk of patient selection bias. These studies were not case-control but provided insufficient details about the inclusion and exclusion criteria. About 68% (64 out of 94 studies) have unclear risk of index bias due to unclear information on whether the index test results were interpreted with knowledge of reference test results. The remaining studies (32%) have low risk of index bias as the reference tests ware blinded from each other and results were recorded independently by two readers. With regards to the reference standard risk of bias, almost all the studies (87%) have low risk of bias, as these studies used similar RT-PCR as the reference standard and the reference tests results were interpreted without knowledge of index test results. Eleven studies (12%) have unclear risk of reference standard bias due to these studies did not provide enough information about whether reference standard results were interpreted without knowledge of the results of the index test. Only 1% of the studies has high risk of reference standard bias because the reference standard results were interpreted with knowledge of the results of the index test. Most of studies (87%) have low risk of flow and timing bias. Seven studies (8%) have unclear risk of flow and timing bias due to no information on whether the samples for a reference test and the index test were taken at the same time. Only five studies (5%) have high risk of flow and timing bias. One of the five studies has high risk of flow and timing bias because samples were collected from the same patients at multiple time points. Another three studies have high risk of flow and timing bias due to the use of different samples for index and reference test and one more due to use of different standard references. selection bias. Twelve (13%) studies have high risk of patient selection bias due to the case-control study design. These studies specifically recruited clinical samples known to be uninfected or infected with coronavirus. The remaining studies (4%) have unclear risk of patient selection bias. These studies were not case-control but provided insufficient details about the inclusion and exclusion criteria. About 68% (64 out of 94 studies) have unclear risk of index bias due to unclear information on whether the index test results were interpreted with knowledge of reference test results. The remaining studies (32%) have low risk of index bias as the reference tests ware blinded from each other and results were recorded independently by two readers. With regards to the reference standard risk of bias, almost all the studies (87%) have low risk of bias, as these studies used similar RT-PCR as the reference standard and the reference tests results were interpreted without knowledge of index test results. Eleven studies (12%) have unclear risk of reference standard bias due to these studies did not provide enough information about whether reference standard results were interpreted without knowledge of the results of the index test. Only 1% of the studies has high risk of reference standard bias because the reference standard results were interpreted with knowledge of the results of the index test. Most of studies (87%) have low risk of flow and timing bias. Seven studies (8%) have unclear risk of flow and timing bias due to no information on whether the samples for a reference test and the index test were taken at the same time. Only five studies (5%) have high risk of flow and timing bias. One of the five studies has high risk of flow and timing bias because samples were collected from the 0% 20% 40% 60% 80% 100% Flow and timing risk of bias A total of 74,445 samples tested using 30 different rapid antigen tests (RATs) and confirmed with RT-PCR as reference test were included in the meta-analysis. The sensitivity and specificity of the RATs ranged from 37% to 90% and 65% to 100%, respectively ( Figure 3 ). Of the 30 RATs analyzed, 52% (n = 16) reported at least 70% sensitivity. CoviNAg ELISA Kit (XEMA, Moscow, Russia) and Sienna-Clarity COVID-19 Ag RTC (Salofa Oy, Salo, Finland) showed highest sensitivity of 90% while MEDsan SARS-Cov-2 Ag (MEDsan GmbH, Hamburg, Germany) showed lowest sensitivity of 37%. Nonetheless, when those kits tested using small number of samples were excluded from analysis, SARS-CoV-2 Ag Test (LumiraDx GmbH, Cologne, Germany) showed the highest sensitivity of 86% followed by Lumipulse G SARS-CoV-2 Ag (Fujirebio, Tokyo, Japan) and BinaxNOW™ COVID-19 Ag Self-Test (Abbott, Jena, Germany) with sensitivity of 83% and 79%, respectively. Meanwhile, for specificity, almost all RATs reported over 90% specificity with only four (13%) showed specificity below 90%. These include CoviNAg ELISA Kit (XEMA, Moscow, Russia), E25Bio Rapid Diagnostic Test (E25Bio, Cambridge, USA), GSD NovaGen SARS-CoV-2 (Eurofins Technologies, Budapest, Hungary) and Aegle Coronavirus Ag RTC (Bio-Rad, California, United States). Panbio™ COVID-19 Ag RDT (Abbott, Jena, Germany) was observed to be the most studied kit for SARS-CoV-2 detection with 16,207 samples were included from 25 different studies. The pooled sensitivity and specificity of Panbio™ COVID-19 Ag RDT was 75.9% and 99.6%. The overall pooled sensitivity and specificity of RATs were 70% (95% CI: 69-71) and 98% (95% CI: 98-98), respectively (Table 3) . For subgroup analysis of the RATs based on specimens, a total of 65 studies were included after excluding studies utilizing combination of nasopharyngeal and throat swabs. Nasal swabs showed the highest sensitivity of 83% (95% CI: 80-86), whereas nasopharyngeal swabs showed a sensitivity of 71% (95% CI: 70-72), throat swabs 69% (95% CI: 63-75) and saliva 68% (95% CI: 59-77). Comparably high specificities were observed among all four specimens, ranging from 97% to 99%. (MEDsan GmbH, Hamburg, Germany) showed lowest sensitivity of 37%. Nonetheless, when those kits tested using small number of samples were excluded from analysis, SARS-CoV-2 Ag Test (LumiraDx GmbH, Cologne, Germany) showed the highest sensitivity of 86% followed by Lumipulse G SARS-CoV-2 Ag (Fujirebio, Tokyo, Japan) and BinaxNOW™ COVID-19 Ag Self-Test (Abbott, Jena, Germany) with sensitivity of 83% and 79%, respectively. Subgroup analysis based on the presence of symptoms showed RATs gives higher sensitivity (82%) among symptomatic as compared with asymptomatic (68%). The specificity of RATs was similar in both symptomatic and asymptomatic patients (98%). Cycle threshold (Ct) values are inversely correlated with the viral load in a specimen. Among those patients with higher viral load (Ct value ≤25), the RATs showed sensitivity and specificity of 96% and 99%, respectively. Meanwhile, the sensitivity of RATs dropped to 69% when used on patients with low viral load (Ct value >25). A subgroup analysis of the RATs based on countries showed sensitivity and specificity ranging from 58% to 87% and 96% to 100%, respectively. The sensitivity was observed to be highest in China (87%) followed by Italy (81%), Chile (77%), the United States (77%), Belgium (73%), Japan and the Netherlands (72%) and Spain (71%). On the other hand, the sensitivity was observed to be lower than average (70%) in Germany, France and India with all showed sensitivity of 58%. In regard to specificity, comparably high specificities, ranging from 96% to 100% were observed in all countries, except in Belgium (84%). After excluding the case-control studies, the overall pooled sensitivity and specificity of RATs were 72% (95% CI: 71-73) and 98% (95% CI: 98-98), respectively (Table 4) . Meanwhile, the diagnostic performance of different specimens showed similar result to the subgroups analysis without excluding case-control studies. For analysis based on symptoms, the RATs performance after excluding case-control studies showed lower sensitivity of 80% (95% CI: 78-81) and higher specificity of 99% (95% CI: . Similarly, the RATs performance showed lower sensitivity and higher specificity when tested for samples from those patients with Ct value >25. The performance of RATs for asymptomatic and patients with Ct value ≤25 remain similar. Further analysis, which includes comparative performance of studies that blinded the index test and studies that did not blind the index test were performed (Table 5 ). Studies where the index text was blinded to the results of the reference standard had a sensitivity and specificity of 72% and 97%, respectively. Meanwhile, studies where the index test was not blinded to the results of the reference standard has a sensitivity and specificity of 72% and 97%, respectively. Table 5 . Comparative performance of studies that blinded the index test and studies that did not blind the index test. Recently, there have been two systematic reviews on the performance of RATs for SARS-CoV-2 detection [16, 17] . These reviews only included studies up to 13 January and 30 April 2021. Our review provided more comprehensive analysis on the performance of RATs until 26 August, following the inclusion of 94 studies. Nonetheless, most of the studies still from European countries such as Germany, Spain, Italy, the Netherlands, France, Belgium and Switzerland. Lack of studies from West and Southeast Asian countries, South America and Africa, highlighting the current gap pertaining to understanding the RATs performance in such geographical areas. A similar observation was reported by the previous study, in which, most of the included studies were from Germany, Spain and Italy [17] . The plausible reason may be attributed to the fact that most kits are manufactured in European countries; thus, such test kits were easier to obtain in those countries as compared to others where supply shortages are commonly reported. The second reason is that these countries were badly affected by COVID-19 in the beginning of the outbreak. Therefore, the COVID-19 RAT is becoming popular across European countries as governments' efforts to slow the spreading of the virus by tracking infected individuals. Panbio™ COVID-19 Ag RDT (Abbott, Jena, Germany) and STANDARD Q COVID-19 Ag Home Test (SD Biosensor, Seoul, South Korea) were used in the majority of studies (25 and 11 studies, respectively) followed by SARS-CoV-2 Rapid Ag Test (Roche, Basel, Switzerland) and Lumipulse G SARS-CoV-2 Ag (Fujirebio, Tokyo, Japan). This could be due to the fact that Panbio™ COVID-19 Ag RDT (Abbott, Jena, Germany) and STANDARD Q COVID-19 Ag Home Test (SD Biosensor, Seoul, South Korea) are two RAD kits that are currently included under the 'WHO Emergency Use Listing for In vitro diagnostics (IVDs) Detecting SARS-CoV-2 [113] . There is still a lack of evaluation for newly developed test kits such as CoviNAg ELISA Kit (XEMA, Moscow, Russia) and Sienna-Clarity COVID-19 Ag RTC (Salofa Oy, Salo, Finland). Therefore, future diagnostic evaluation studies should include the newly developed test kits so that more data can be obtained for future comparative analyses. Immunochromatography, which involves spotting antibodies onto nitrocellulose membranes that interact with specific antigens in patient samples is the basis of RATs. The antigen-antibody interaction can be visualised manually or by using an immunofluorescence machine reader. The genome of SARS-CoV-2 comprises genes the responsible for four structural proteins such the spike (S), envelope (E), membrane (M) and nucleocapsid (N) [114] . N protein is frequently employed as a target analyte in RATs for COVID-19 diagnosis, as shown in Table 2 . N-protein is mostly expressed during the early stages of SARS-CoV-2 infection and has the least amount of variation in its gene sequence, indicating that it is a stable protein [115] . Most studies used nasopharyngeal swabs as specimens examined for evaluating the performance of the test kits. Only a few studies used nasal swabs, saliva and throat swabs. This observation signaling the need for more studies using such specimens so that alternative sampling approaches for the rapid detection of SARS-CoV-2 could be identified. Subgroup analysis of the test kits based on specimens showed nasal swabs gave the highest sensitivity of 83% (95% CI: 80-86) followed by nasopharyngeal swabs 71% (95% CI: 70-72), throat swabs 69% (95% CI: [63] [64] [65] [66] [67] [68] [69] [70] [71] [72] [73] [74] [75] and saliva 68% (95% CI: [59] [60] [61] [62] [63] [64] [65] [66] [67] [68] [69] [70] [71] [72] [73] [74] [75] [76] [77] . Interestingly, nasopharyngeal swabs are currently considered as the gold standard specimen for SARS-CoV-2 laboratory diagnosis. This finding may suggest that nasal swab could replace nasopharyngeal swabs as the gold standard specimen. However, it is important to note that the average sensitivity of the nasal swabs was calculated only based on 2148 samples as compared to 38548 samples for nasopharyngeal swabs. Thus, evaluation using more samples are still needed to provide a comprehensive conclusion regarding the nasal swabs' performance. Lower sensitivity for throat swabs was in agreement with a recent study on the performances of the different sampling approaches for SARS-CoV-2, which reported lower sensitivity by throat swabs (68%) as compared to nasal swabs and saliva [116] . The meta-analysis shows that the sensitivity and specificity of the 30 RATs ranged from 37% to 90% and 65% to 100%, respectively, whereas the overall pooled sensitivity and specificity of RATs were 70% (95% CI: [69] [70] [71] and 98% (95% CI: , respectively. Based on sample size, the SARS-CoV-2 Ag Test (LumiraDx GmbH, Cologne, Germany) showed the highest sensitivity of 86% followed by the Lumipulse G SARS-CoV-2 Ag (Fujirebio, Tokyo, Japan) and BinaxNOW™ COVID-19 Ag Self-Test (Abbott, Jena, Germany) with sensitivities of 83% and 79%, respectively. Furthermore, most of the studies showed a high level of specificity (greater than 90%). The WHO recommended that the minimum performance requirements of RATs must be ≥ 80% sensitivity and ≥97% specificity. The studies included in this review have showed a wide range of sensitivity and specificity. The different results could be related to various study methods, RAT kits manufacturers, patient selection, types of specimens and the stage of disease at the time of sample collection. Based on the metaanalysis, the RAT kits that meet WHO criteria are Sienna-Clarity COVID-19 Ag RTC (Salofa Oy, Salo, Finland), Biosynex COVID-19 Ag BSS Test (Biosynex, Strasbourg, France), Innova Rapid SARS-CoV-2 Ag Test (Xiamen Biotime Biotechnology, Fujian, China), Dräger Ag Test SARS-CoV-2 (Dräger, Lübeck, Germany), SARS-CoV-2 Ag Test (LumiraDx GmbH, Cologne, Germany), Lumipulse G SARS-CoV-2 Ag (Fujirebio, Tokyo, Japan), S-PLEX SARS-CoV-2 N Kit (MesoScale Diagnostics, Maryland, USA) and BinaxNOW™ COVID-19 Ag Self-Test (Abbott, Jena, Germany). Subgroup analysis of the RATs based on countries showed sensitivity and specificity ranging from 58% to 87% and 96% to 100%, respectively. The sensitivity was observed to be highest in China (87%) followed by Italy (81%), Chile (77%), the United States (77%), Belgium (73%), Japan and the Netherlands (72%) and Spain (71%). On the other hand, the sensitivity was observed to be lower than average (70%) in Germany, France and India with all showing a sensitivity of 58%. The previous systematic review reported that the sensitivity of RATs in the population of Europe and America was higher as compared to that of Asia and Africa [17] . Their finding was different with our finding. One plausible reason is that the RATs evaluated previously were manufactured from Europe and America which may affect the test performance in Asia and Africa after repeated freeze-thaw procedures during transportation [117] . In regard to specificity, comparably high specificities, ranging from 96% to 100%, were observed in all countries, except in Belgium (84%). According to the WHO, RATs should be prioritized for use in symptomatic individuals who meet the COVID-19 case definition, as well as to test asymptomatic individuals at high risk of infection, particularly in settings where NAAT testing capacity is limited. Thus, this review also analysed the sensitivity and specificity of RATs based on symptoms. RATs showed similar specificity (98%) for symptomatic and asymptomatic patients. However, the sensitivity of RATs for symptomatic patients was greater than asymptomatic patients (82% vs. 68%). A similar finding was reported by the previous systematic reviews, in which, the sensitivity of RATs is higher when used for symptomatic patients [16, 17] . In addition, there is a clear association between Ct values of RT-qPCR and RATs' sensitivity and specificity. The lower the Ct value (≤25), the greater the sensitivity and specificity of RATs, whereas the higher the Ct value (<25), the lower the sensitivity and specificity of RATs. Ct values, on the other hand, cannot be directly compared between tests and must be interpreted with caution because they are impacted by sample type, sample collection timing, and assay design [118] . The severity of the disease, the timing of sample collection, the types of samples, and sample handling techniques all influence antigen levels in samples [20] . It is hard to determine if the difference in observed sensitivity is due to the test's performance or the qualities of the samples utilized in the test without this information. Unfortunately, the majority of the studies included in this review did not provide information on antigen levels in the samples. Information about disease severity and sample collection timing are often missing. Future research should include this information to enable for a more accurate assessment of diagnosis test performance as well as the identification of their actual limitations. Throughout this systematic review, our study identified the relevant peer-review articles published to reach the objective findings about the performances of the antigen detection for diagnostic of COVID-19. This study followed the PRISMA guidelines to reduce the risk of bias and meet the objective of the review. The use of several predetermined keywords and guidelines in this study during the screening process can assure the reproducibility of this systematic review. The screening process that started from the title screening followed by abstract and full text screening was documented properly to avoid any risk of bias in the systematic review. Based on the quality assessment of the QUADAS-2, most of the included studies showed a low of bias. However, there were several studies that indicate high risk of bias for the QUADAS-2 assessment. The high risk in the patient selection was due to the casecontrol study design, while the unclear risk of the patient selection was due to insufficient details about the patient inclusion and exclusion criteria. For the risk of the index test, most of the included studies were unclear as the authors did not mention whether the index test results were interpreted with knowledge of the reference test results. The high risk of the reference standard bias was due to the results that were not interpreted with knowledge of the result of the index test. For the risk of the flow and timing bias, high risk of the bias due to the several reference standards or collected samples for the index and references tests at separate time. Nevertheless, our systematic review has three main limitations. First, this systematic review revealed that there was considerable publication bias in the included studies as the study on the performances of the antigen detection of the SARS-CoV-2 were still in the early stages of development and almost all reported diagnostics performance assessments were done and published by the same research group. We anticipated that such bias will be reduced after these antigen performances were completely evaluated by various and independent research teams. Second, the presence of the SARS-CoV-2 antigen in the patients does not always indicate the presence of the viable virus, and we need to examine whether SARS-CoV-2 antigen-positive patients are contagious to other people. Lastly, the possibility of protein mutations in SARS-CoV-2 variants, including the newly revealed Omicron, affecting the sensitivity and specificity of RATs is not reported in this review. In conclusion, our systematic review and meta-analysis revealed the current performance of RATs for SARS-CoV-2 detection. Overall diagnostic sensitivity and specificity of these RATs were 70% and 98%, respectively. Quality assessment showed majority of the studies have low risk of bias. However, several studies showed high risk of patient selection bias due to case-control study design and high risk of flow and timing bias due to the use of different samples for index and reference test and use of different standard references. Regarding index test risk of bias, the majority did not mention whether or not the authors performed double-blinded index test. Future study should attempt to perform double-blinded index test as such improvement in study design would reduce index test risk of bias. CoviNAg ELISA Kit (XEMA, Moscow, Russia) and Sienna-Clarity COVID-19 Ag RTC (Salofa Oy, Salo, Finland) had the highest diagnostic sensitivity among all RATs, while MEDsan SARS-Cov-2 Ag (MEDsan GmbH, Hamburg, Germany) had the lowest diagnostic sensitivity. More studies from the Middle East and Southeast Asian, South American, and African countries are warranted so that a comprehensive subgroup analysis based on regions can be performed. Nasal swabs showed the highest sensitivity followed by nasopharyngeal swabs, throat swabs and saliva. The RATs showed higher sensitivity for those patients with symptoms and Ct value ≤25. Comparative performance of the RATs using less invasive sampling approaches is still lacking. Improvement in these key areas would help to boost acceptability and accessibility for large practical of RATs for rapid surveillance of SARS-CoV-2, allowing immediate isolation of the infected individuals and reducing the disease's transmission. A Pneumonia Outbreak Associated with a New Coronavirus of Probable Bat Origin A Novel Coronavirus from Patients with Pneumonia in China Clinical Features of Patients Infected with 2019 Novel Coronavirus in Prevalence of Olfactory Dysfunction in Coronavirus Disease 2019 (COVID-19): A Meta-analysis of 27,492 Patients Prevalence and Characteristics of Taste Disorders in Cases of COVID-19: A Meta-Analysis of 29,349 Patients. Otolaryngol.-Head Neck Surg Reverse Transcriptase Loop Mediated Isothermal AmplifiCation (RT-LAMP) for COVID-19 Diagnosis: A Systematic Review and Meta-Analysis Community Transmission of Severe Acute Respiratory Syndrome Coronavirus 2 Escalating Infection Control Response to the Rapidly Evolving Epidemiology of the Coronavirus Disease 2019 (COVID-19) Due to SARS-CoV-2 in Hong Kong The Coronavirus Pandemic and Aerosols: Does COVID-19 Transmit via Expiratory Particles? The Mental Health Impact of the COVID-19 Pandemic on People with and without Depressive, Anxiety, or Obsessive-Compulsive Disorders: A Longitudinal Study of Three Dutch Case-Control Cohorts Mental Health before and during the COVID-19 Pandemic: A Longitudinal Probability Sample Survey of the UK Population US CDC Real-Time Reverse Transcription PCR Panel for Detection of Severe Acute Respiratory Syndrome Coronavirus 2 Recent Advances and Perspectives of Nucleic Acid Detection for Coronavirus Diagnosis of SARS-CoV-2 Infection Accuracy of Novel Antigen Rapid Diagnostics for SARS-CoV-2: A Living Systematic Review and Meta-Analysis Diagnostic Accuracy of Rapid Antigen Test Kits for Detecting SARS-CoV-2: A Systematic Review and Meta-Analysis of 17,171 Suspected COVID-19 Patients The PRISMA Statement for Reporting Systematic Reviews and Meta-Analyses of Studies That Evaluate Healthcare Interventions: Explanation and Elaboration QUADAS-2: A Revised Tool for the Quality Assessment of Diagnostic Accuracy Studies Rapid SARS-CoV-2 Antigen Detection Assay in Comparison with Real-Time RT-PCR Assay for Laboratory Diagnosis of COVID-19 in Thailand Evaluation of Rapid Antigen Test for Detection of SARS-CoV-2 Virus Low Performance of Rapid Antigen Detection Test as Frontline Testing for COVID-19 Diagnosis Clinical Evaluation of Self-Collected Saliva by Quantitative Reverse Transcription-PCR (RT-QPCR), Direct RT-QPCR, Reverse Transcription-Loop-Mediated Isothermal Amplification, and a Rapid Antigen Test To Diagnose COVID-19 Evaluation of a Novel Antigen-Based Rapid Detection Test for the Diagnosis of SARS-CoV-2 in Respiratory Samples Comparison of the SARS-CoV-2 Rapid Antigen Test to the Real Star Sars-CoV-2 RT PCR Kit Field Evaluation of a Rapid Antigen Test (Panbio TM COVID-19 Ag Rapid Test Device) for COVID-19 Diagnosis in Primary Healthcare Centres Evaluation of a Rapid Antigen Test (Panbio TM COVID-19 Ag Rapid Test Device) for SARS-CoV-2 Detection in Asymptomatic Close Contacts of COVID-19 Patients Performance of an Antigen-Based Test for Asymptomatic and Symptomatic SARS-CoV-2 Testing at Two University Campuses-Wisconsin Accuracy of a Nucleocapsid Protein Antigen Rapid Test in the Diagnosis of SARS-CoV-2 Infection Evaluation of a Rapid Diagnostic Assay for Detection of SARS-CoV-2 Antigen in Nasopharyngeal Swabs Panbio Antigen Rapid Test Is Reliable to Diagnose SARS-CoV-2 Infection in the First 7 Days after the Onset of Symptoms Evaluation of Two Rapid Antigen Tests to Detect SARS-CoV-2 in a Hospital Setting Clinical Validation of Quantitative SARS-CoV-2 Antigen Assays to Estimate SARS-CoV-2 Viral Loads in Nasopharyngeal Swabs Field Evaluation of the Performance of a SARS-CoV-2 Antigen Rapid Diagnostic Test in Uganda Using Nasopharyngeal Samples Correlation of SARS-CoV-2 Nucleocapsid Antigen and RNA Concentrations in Nasopharyngeal Samples from Children and Adults Using an Ultrasensitive and Quantitative Antigen Assay Diagnostic Accuracy of the Panbio Severe Acute Respiratory Syndrome Coronavirus 2 Antigen Rapid Test Compared with Reverse-Transcriptase Polymerase Chain Reaction Testing of Nasopharyngeal Samples in the Pediatric Population Evaluating a Novel, Highly Sensitive, and Quantitative Reagent for Detecting SARS-CoV-2 Antigen Efficacy Comparison of Three Rapid Antigen Tests for SARS-CoV-2 and How Viral Load Impact Their Performance Comparison of SARS-CoV-2 Detection by Rapid Antigen and by Three Commercial RT-QPCR Tests: A Study from Martin University Hospital in Slovakia Evaluation of a Rapid Antigen Test To Detect SARS-CoV-2 Infection and Identify Potentially Infectious Individuals Diagnostic Performance of CerTest and Panbio Antigen Rapid Diagnostic Tests to Diagnose SARS-CoV-2 Infection Comparative Analysis of Antigen and Molecular Tests for the Detection of Sars-CoV-2 and Related Variants: A Study on 4266 Samples Parallel Detection of SARS-CoV-2 RNA and Nucleocapsid Antigen in Nasopharyngeal Specimens from a COVID-19 Patient Screening Cohort Evaluation of Lumipulse®G SARS-CoV-2 Antigen Assay Automated Test for Detecting SARS-CoV-2 Nucleocapsid Protein (NP) in Nasopharyngeal Swabs for Community and Population Screening Comparative Performance of SARS-CoV-2 Lateral Flow Antigen Tests and Association with Detection of Infectious Virus in Clinical Specimens: A Single-Centre Laboratory Evaluation Study Comparison of Roche and Lumipulse Quantitative SARS-CoV-2 Antigen Test Performance Using Automated Systems for the Diagnosis of COVID-19 Efficacy and Validity of Automated Quantitative Chemiluminescent Enzyme Immunoassay for SARS-CoV-2 Antigen Test from Saliva Specimen in the Diagnosis of COVID-19 Diagnostic Accuracy of Panbio Rapid Antigen Tests on Oropharyngeal Swabs for Detection of SARS-CoV-2 Evaluation of a Laboratory-Based High-Throughput SARS-CoV-2 Antigen Assay for Non-COVID-19 Patient Screening at Hospital Admission Comparison of Five SARS-CoV-2 Rapid Antigen Detection Tests in a Hospital Setting and Performance of One Antigen Assay in Routine Practice: A Useful Tool to Guide Isolation Precautions? Diagnostic Performance and Characteristics of Anterior Nasal Collection for the SARS-CoV-2 Antigen Test: A Prospective Study Clinical Application of a New SARS-CoV-2 Antigen Detection Kit (Colloidal Gold) in the Detection of COVID-19 Performance of the LIAISON®SARS-CoV-2 Antigen Assay vs. SARS-CoV-2-RT-PCR. Pathogens Evaluation of the AMP SARS-CoV-2 Rapid Antigen Test in a Hospital Setting The Abbott PanBio WHO Emergency Use Listed, Rapid, Antigen-Detecting Point-of-Care Diagnostic Test for SARS-CoV-2-Evaluation of the Accuracy and Ease-of-Use Analytical Performances of the Point-of-Care SIENNA TM Antigen Rapid Test for the Detection of SARS-CoV-2 Nucleocapsid Protein in Nasopharyngeal Swabs: A Prospective Evaluation during the COVID-19 Second Wave in France Performance of a Rapid Antigen Test in the Diagnosis of SARS-CoV-2 Infection Validation and Implementation of the Panbio COVID-19 Ag Rapid Test for the Diagnosis of SARS-CoV-2 Infection in Symptomatic Hospital Healthcare Workers Clinical Assessment of the Roche SARS-CoV-2 Rapid Antigen Test Evaluation of an Antigen-Based Test for Hospital Point-of-Care Diagnosis of SARS-CoV-2 Infection Performance of SARS-CoV-2 Rapid Antigen Test Compared with Real-Time RT-PCR in Asymptomatic Individuals Performance Evaluation of a Lateral Flow Assay for Nasopharyngeal Antigen Detection for SARS-CoV-2 Diagnosis Validation of the STANDARD Q COVID-19 antigen test in Vojvodina, Serbia Comparative Evaluation of Panbio and SD Biosensor Antigen Rapid Diagnostic Tests for COVID-19 Diagnosis Rapid Antigen Detection Kit for the Diagnosis of SARS-CoV-2-Are We Missing Asymptomatic Patients? High Performance of a Novel Antigen Detection Test on Nasopharyngeal Specimens for Diagnosing SARS-CoV-2 Infection Rapid Antigen Test for SARS-CoV-2 Has Acceptable Accuracy in Symptomatic Patients in Primary Health Care Evaluation of the Test Accuracy of a SARS-CoV-2 Rapid Antigen Test in Symptomatic Community Dwelling Individuals in the Netherlands Saliva Is a Valid Alternative to Nasopharyngeal Swab in Chemiluminescence-Based Assay for Detection of SARS Diagnostic Accuracy of Two Commercial SARS-CoV-2 Antigen-Detecting Rapid Tests at the Point of Care in Community-Based Testing Centers Accuracy of a Novel SARS-CoV-2 Antigen-Detecting Rapid Diagnostic Test from Standardized Self-Collected Anterior Nasal Swabs Clinical Performance Evaluation of SARS-CoV-2 Rapid Antigen Testing in Point of Care Usage in Comparison to RT-QPCR Evaluation of Three Rapid Lateral Flow Antigen Detection Tests for the Diagnosis of SARS-CoV-2 Infection Clinical Evaluation of Rapid Point-of-Care Antigen Tests for Diagnosis of SARS-CoV-2 Infection Comparison of the Quantitative DiaSorin Liaison Antigen Test to Reverse Transcription-PCR for the Diagnosis of COVID-19 in Symptomatic and Asymptomatic Outpatients Clinical Evaluation of Roche SD Biosensor Rapid Antigen Test for SARS-CoV-2 in Municipal Health Service Testing Site, the Netherlands Real-life Evaluation of a COVID-19 Rapid Antigen Detection Test in Hospitalized Children Prospective Analytical Performance Evaluation of the QuickNavi TM -COVID19 Ag for Asymptomatic Individuals Rapid Antigenic Tests Compared to RT-PCR in a Student Population: The StudyCov Study Performance and Implementation Evaluation of the Abbott BinaxNOW Rapid Antigen Test in a High-Throughput Drive-Through Community Testing Site in Massachusetts Performance Characteristics of a Rapid Severe Acute Respiratory Syndrome Coronavirus 2 Antigen Detection Assay at a Public Plaza Testing Site in San Francisco Diagnostic Accuracy of a SARS-CoV-2 Rapid Antigen Test in Real-Life Clinical Settings Comparison of Seven Commercial SARS-CoV-2 Rapid Point-of-Care Antigen Tests: A Single-Centre Laboratory Evaluation Study Evaluation of COVID-19 Antigen Fluorescence Immunoassay Test for Rapid Detection of SARS-CoV-2 Multicenter Evaluation of a Fully Automated High-Throughput SARS-CoV-2 Antigen Immunoassay Performance Evaluation of an Automated SARS-CoV-2 Ag Test for the Diagnosis of COVID-19 Infection on Nasopharyngeal Swabs Evaluation of a Rapid Antigen Detection Test (Panbio TM COVID-19 Ag Rapid Test Device) as a Point-of-care Diagnostic Tool for COVID-19 in a Pediatric Emergency Department Diagnostic Accuracy of SARS-CoV-2 Antigen Detection Test in Children: A Real-Life Study Comparison of Four Commercial, Automated Antigen Tests to Detect SARS-CoV-2 Variants of Concern Validation of a Rapid Antigen Test as a Screening Tool for SARS-CoV-2 Infection in Asymptomatic Populations. Sensitivity, Specificity and Predictive Values Evaluation of Accuracy, Exclusivity, Limit-of-Detection and Ease-of-Use of LumiraDx TM : An Antigen-Detecting Point-of-Care Device for SARS-CoV-2 Evaluation of Three Immunoassays for the Rapid Detection of SARS-CoV-2 Antigens Clinical Evaluation of Sofia Rapid Antigen Assay for Detection of Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) among Emergency Department to Hospital Admissions Diagnostic Performance of the Elecsys SARS-CoV-2 Antigen Assay in the Clinical Routine of a Tertiary Care Hospital: Preliminary Results from a Single-center Evaluation Real-Life Validation of the Panbio TM COVID-19 Antigen Rapid Test (Abbott) in Community-Dwelling Subjects with Symptoms of Potential SARS-CoV-2 Infection The Effectiveness of Rapid Antigen Test-Based for SARS-CoV-2 Detection in Nursing Homes in Valencia, Spain Head-to-Head Comparison of Rapid and Automated Antigen Detection Tests for the Diagnosis of SARS-CoV-2 Infection Evaluation of a SARS-CoV-2 Antigen-detecting Rapid Diagnostic Test as a Self-test: Diagnostic Performance and Usability Evaluation of Two Fluorescence Immunoassays for the Rapid Detection of SARS-CoV-2 Antigen-New Tool to Detect Infective COVID-19 Patients Performance of a Point-of-Care Test for the Rapid Detection of SARS-CoV-2 Antigen. Microorganisms SARS-CoV-2 Antigen Detection to Expand Testing Capacity for COVID-19: Results from a Hospital Emergency Department Testing Site Evaluation of a High Volume Antigen Test for Detection of SARS-CoV-2 Diagnostic Accuracy and Feasibility of Patient Self-Testing with a SARS-CoV-2 Antigen-Detecting Rapid Test Performance of the Diasorin SARS-CoV-2 Antigen Detection Assay on the LIAISON XL Field Performance and Public Health Response Using the BinaxNOW TM Rapid Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) Antigen Detection Assay During Community-Based Testing Performance of a SARS-CoV-2 Antigen Rapid Immunoassay in Patients Admitted to the Emergency Department Implementing SARS-CoV-2 Rapid Antigen Testing in the Emergency Ward of a Swiss University Hospital: The INCREASE Study. Microorganisms Multi-Site Assessment of Rapid, Point-of-Care Antigen Testing for the Diagnosis of SARS-CoV-2 Infection in a Low-Prevalence Setting: A Validation and Implementation Study Diagnostic Performance of Rapid Antigen Test for COVID-19 and the Effect of Viral Load, Sampling Time, Subject's Clinical and Laboratory Parameters on Test Accuracy Diagnostic Accuracy of Rapid Antigen Tests in Asymptomatic and Presymptomatic Close Contacts of Individuals with Confirmed SARS-CoV-2 Infection: Cross Sectional Study Multicenter Evaluation of the PanbioTM COVID-19 Rapid Antigen-Detection Test for the Diagnosis of SARS-CoV-2 Infection Performance of Repeat BinaxNOW Severe Acute Respiratory Syndrome Coronavirus 2 Antigen Testing in a Community Setting Evaluation of Rapid Antigen Detection Kit from the WHO Emergency Use List for Detecting SARS-CoV-2 The Architecture of SARS-CoV-2 Transcriptome The SARS-CoV Nucleocapsid Protein: A Protein with Multifarious Activities Diagnostic Performance of Different Sampling Approaches for SARS-CoV-2 RT-PCR Testing: A Systematic Review and Meta-Analysis Effect of Freeze-Thaw Cycles on the Molecular Weight and Size Distribution of Gluten Association Between SARS-CoV-2 Cycle Threshold Values and Clinical Outcomes in Patients With COVID-19: A Systematic Review and Meta-Analysis. Open Forum Infect Data Availability Statement: Not applicable. The authors declare no conflict of interest.