key: cord-0689787-5d4oqzle authors: Kotani, Toru; Shono, Atsuko title: Roles of Electrical Impedance Tomography in Determining a Lung Protective Strategy for Acute Respiratory Distress Syndrome in the Era of Coronavirus Disease 2019 date: 2021-04-02 journal: JMA J DOI: 10.31662/jmaj.2021-0014 sha: 0e1038a8ce45321ffa155ba65e15ffa3a4e74d32 doc_id: 689787 cord_uid: 5d4oqzle Electrical impedance tomography (EIT) is noninvasive and can be used at the bedside for real-time evaluation to identify ventilation distribution of infected lungs. This review briefly describes the basic principle of EIT and summarizes the latest findings on its potential contribution to lung protective strategies in coronavirus disease 2019 patients. Additionally, experimental approaches for detecting the distribution of pulmonary blood flow in coronavirus disease 2019 patients are presented. The findings underscore the role of EIT in determining lung protective strategies for coronavirus disease 2019-associated acute respiratory distress syndrome. Acute respiratory distress syndrome (ARDS) is characterized by life-threatening hypoxemia caused by hyperpermeability of pulmonary capillaries resulting from various background diseases or conditions. Although mechanical ventilation is required for ARDS diagnosis and treatment, it has become obvious that ventilator-associated lung injury plays an important role in the development of multiple organ failure, which is the cause of death for ARDS. Large-scale clinical trials have demonstrated that low tidal volume ventilation (LTV), which limits both tidal volume and plateau pressure, improves ARDS prognosis. Currently, LTV is one of lung protection strategies and is essential for treating ARDS. When gas exchange cannot be maintained by LTV, early implementation of prone positioning acts as rescue therapy and improves the prognosis. It is well known that prone positioning recruits collapsed regions and improves oxygenation. Recent studies have suggested that prone positioning promotes homogeneous ventilation distribution and might have a lung protective effect. If the lungs are evenly ventilated by reducing regional overdistention and recruiting the collapsed region, local stress and strain can be ameliorated, resulting in the prevention of ventilator-induced lung injury. The importance of homogeneity of regional ventilation has been increasingly recognized. It has been reported that lung inhomogeneity dur-ing tidal ventilation increases dynamic lung strain (1) . In an animal model, increased dynamic strain was more associated with the development of pulmonary edema, derangement of lung mechanics, and higher mortality than static strain (2) . However, monitoring measures to assess regional ventilation homogeneity in critically ill patients at bedside are limited. Computed tomography is a standard test for examining ventilation and aeration of the lungs, but transportation of critically ill patients, radiation exposure, and limitations to static evaluation are the major obstacles for its frequent use in the clinical practice. Developing a new testing method that can be performed repeatedly at bedside is warranted. Body tissue consists of specific components such as water, lipids, and electrolytes. These components have their own responses to an externally applied electric current. Gas in the lungs acts as an electrical resistor and increases regional impedance. Ventilation causes cyclic regional changes in impedance depending on the regional difference in aeration. Electrical impedance tomography (EIT) measures and visualizes the impedance changes generated by ventilation in real time. The principle of measurement is described elsewhere (3) . Briefly, the impedance value is measured by applying a small alternating current through electrodes implemented in the EIT belt, which is normally placed at the fifth to sixth intercostal level of the patient's chest (Figure 1) . Sequential cross-sectional images are reconstructed using raw impedance data, and the user watches the dynamic image of ventilated lungs. Figure 2 presents typical sequential images of EIT. With a rate of 20-30 frames per second, it is easy to recognize the spatial localization and tem-poral discordance or asynchrony of ventilation as a dynamic image. EIT is radiation free, and a dynamic regional ventilation profile can be monitored at the bedside every time the ventilator settings are changed. EIT provides reliable information to detect alveolar overdistention and collapse. To assess the homogeneity of ventilation distribution, the patient was paralyzed, and incremental and decremental PEEP trials were performed at a fixed driving pressure (10-12 cmH 2 O). The whole procedure was monitored and recorded using EIT, and the saved impedance data were analyzed using software (EIT diag, Draeger, Luebeck, Germany). In severe ARDS patients, EIT is used to optimize mechanical ventilation settings (4) , providing useful information when titrating personalized PEEP (5) . EIT might help evaluate the possibility of lung recruitment maneuvers and estimate recruitable alveolar collapse (6) . Prone positioning is a rescue therapy for patients who do not respond to lung protective ventilation. We performed prone positioning for a patient with ARDS who could not maintain gas exchange with LTV and observed the process of the change in regional ventilation distribution using EIT (7) . Initially, ventilation was shifted to the mid-ventral region, and the dorsal area completely collapsed. Twenty minutes after the initiation of prone positioning, ventilation distribution became more homogeneous, and oxygenation improved dramatically. The homogeneity of ventilation was stabilized after the second session, and the patient survived. Prone positioning basically provides homogeneous ventilation distribution in ARDS lungs, but temporary overdistention in the dorsal region has also been observed immediately after initiation (8) . The response to prone positioning varies among cases and the close monitoring using EIT provides information to improve its efficacy. The outbreak of severe acute respiratory syndrome coronavirus-2 (SARS-CoV-2) is an emergent threat to the current global health. In Japan approximately 1.6% of patients with coronavirus disease 2019 (COVID-19) are severe, defined by the requirement of tracheal intubation and mechanical ventilation. It has been reported that COVID-19 pneumonia can cause ARDS. Therefore, it is necessary for COVID-19-associated ARDS to implement the same lung protection strategy described earlier. However, a recent study paid special attention to lung protection strategies for COVID-19. EIT plays a role in determining lung protective strategies for COVID-19associated ARDS. There are two phenotypes that have a decisive effect on mechanical ventilation strategies in COVID-19 pneumonia (9) . Conceptually, type L is characterized by low elastance, low ventilation-perfusion ratio, and low lung weight and recruitability, whereas type H is defined as high elastance, high right to left shunt, and high lung weight and recruitability. Recruitability is defined as the possibility to open up collapsed areas of the lung by positive pressure. A lung protective strategy of limiting tidal volume and plateau pressure is not required for type L but should be actively implemented in type H. However, it is not easy to distinguish between the two types because the lung weight cannot be measured clinically, and the method for evaluating recruitability is unclear. Measuring ventilation-perfusion matching at bedside has not been established. PEEP titration using EIT can not only evaluate recruitability but also determine the optimum PEEP at the time of measurement and provide regional ventilation homogeneity. As a result, it is considered possible to attenuate the regional dynamic strain and inhomogeneity of transpulmonary pressure. To assess the phenotype, regional ventilation distribution was analyzed using EIT in mechanically ventilated patients with acute hypoxemic respiratory failure (AHRF) and ARDS due to COVID-19. Morais and colleagues (10) presented three cases of COVID-19 AHRF that had similar levels of oxygenation but variable respiratory system compliance. EIT indicated different characteristics of the regional ventilation profile and was helpful in understanding the etiology of hypoxemia at bedside. Tomasino and colleagues (11) used EIT to identify the characteristics of COVID-19 pneumonia and to decide whether to use high PEEP or prone positioning. In a cohort study of mechanically ventilated non-COVID-19 patients, EIT was used to assess the optimal PEEP at which both overdistention and collapse are minimized because both phenomena were equally harmful for ARDS lungs (12) . Perier and colleagues (13) compared the response to PEEP titration between COVID-19 associated ARDS and ARDS from other causes. They defined optimal PEEP using EIT as the smallest sum of overdistention and collapse. COVID-19associated ARDS lungs required 12 cmH 2 O of PEEP and had more collapse at low PEEP levels. Van der Zee and colleagues (14) reported that EIT monitoring during the decremental PEEP trial to seek the level of lowest relative alveolar overdistention and collapse provided personalized PEEP that did not result in high driving pressure or transpulmonary pressure. Shono and colleagues personalized PEEP levels and applied prone positioning sessions to obtain homogeneous ventilation distribution based on EIT analysis (15) . PEEPset corresponded better with the higher PEEP-FIO 2 table (used in ALVEOLI study) than the lower PEEP-FIO 2 table (used in ARMA study), especially in patients with a higher body mass index. However, Sella and colleagues (16) compared EIT-guided PEEP and lower PEEP-FIO 2 table-guided PEEP and found a correlation. The researchers mentioned that the opposite results might be due to differences in the interpretation of the results DOI: 10.31662/jmaj.2021-0014 JMA Journal: Volume 4, Issue 2 https://www.jmaj.jp/ presented by the same device and differences in the degree of obesity. In addition, marked hypoxemia is often observed even in cases without dorsal collapsed areas in COVID-19-infected lungs. Ventilation-perfusion mismatch is considered to be the major cause of this phenomenon. It is believed that COVID-19 hypercoagulability and vascular endothelial cell damage are responsible for this phenomenon, but the details are unclear. Because ventilation-perfusion mismatch cannot be improved by mechanical ventilation, responding to hypoxemia by increasing ventilation pressure, including PEEP, results in considerable lung injury risk due to overdistention. In summary, investigation of ventilation-perfusion matching on the bedside by EIT is useful for lung protection. Recently, EIT has been used to assess regional lung perfusion distribution in a clinical setting. Rapid injection of a contrast agent such as 10 ml of hypertonic sodium chloride enables the generation of a lung perfusion profile during an endexpiratory hold maneuver (4) . Mauri and colleagues reported that the mismatching rate was elevated to 34% (32%-45%) of lung units in COVID-19 lungs (17) . Furthermore, in six of seven patients, ventilated nonperfused units (dead space) represented a much larger proportion than perfused nonventilated perfused units (shunt), although the potential for lung recruitment had large variability. The researchers concluded that an elevated dead space fraction might be a specific pathophysiological characteristic of COVID-19 ARDS. Strong inspiratory effort is often observed as a characteristic of COVID-19 pneumonia. Damaged lungs with reduced compliance generate large negative pressure in the thorax, increase transpulmonary pressure, and can cause or aggravate lung injury, patient self-inflicted lung injury (P-SILI). The pathological changes caused by P-SILI are irreversible and worsen the prognosis. Neuromuscular blockade to attenuate inspiratory drive is effective in protecting lungs against P-SILI. However, the use of neuromuscular blockade leads to alveolar collapse, especially in the dorsal regions, with inappropriate PEEP levels, resulting in inhomogeneous ventilation distribution. The efficacy of EIT to prevent P-SILI has not been tested. Further studies are warranted. This review briefly described the basic principle of EIT and summarized the latest findings on its potential contribution to lung protective strategies in COVID-19 patients. EIT is noninvasive and can be used for real-time evaluation to identify the ventilation distribution of infected lungs. The latest approach indicated the feasibility of detecting the distribution of pulmonary blood flow and evaluating ventilation-perfusion matching at bedside. Concept and design: TK. Drafting of the manuscript: TK, AS. The authors read and approved the final manuscript. Stress and strain within the lung Lung stress and strain during mechanical ventilation: any difference between statics and dynamics? Clinical implication of monitoring regional ventilation using electrical impedance tomography Electrical impedance tomography in acute respiratory distress syndrome Variation of poorly ventilated lung units (silent spaces) measured by electrical impedance tomography to dynamically assess recruitment Bedside estimation of recruitable alveolar collapse and hyperdistension by electrical impedance tomography Electrical impedance tomography-guided prone positioning in a patient with acute cor pulmonale associated with severe acute respiratory distress syndrome Regional overdistension during prone positioning in a patient with acute respiratory failure who was ventilated with a low tidal volume: a case report COVID-19 pneumonia: different respiratory treatments for different phenotypes? Bedside electrical impedance tomography unveils respiratory "chimera" in COVID-19 Electrical impedance tomography and prone position during ventilation in COVID-19 pneumonia: Case reports and a brief literature review Individual positive end-expiratory pressure settings optimize intraoperative mechanical ventilation and reduce postoperative atelectasis Electrical impedance tomography to titrate positive end-expiratory pressure in COVID-19 acute respiratory distress syndrome Electrical impedance tomography for positive end-expiratory pressure titration in COVID-19-related acute respiratory distress syndrome Personalisation of therapies in COVID-19 associated acute respiratory distress syndrome, using electrical impedance tomography Positive endexpiratory pressure titration in COVID-19 acute respiratory failure: electrical impedance tomography vs. PEEP/FiO2 tables Potential for lung recruitment and ventilation-perfusion mismatch in patients with the acute respiratory distress syndrome from coronavirus disease 2019 JMA Journal is an Open Access journal distributed under the Creative Commons Attribution 4.0 International License. To view the details of this license