key: cord-0687215-ptdgqjb4 authors: Angulo, Martín; Beltramelli, Rodrigo; Amarelle, Luciano; Alzugaray, Pedro; Briva, Arturo; Santos, Cristina title: Mechanical Risks of Ventilator Sharing in the COVID-19 Era: a Simulation- Based Study date: 2020-06-26 journal: Arch Bronconeumol DOI: 10.1016/j.arbres.2020.06.006 sha: 4b9282d38eda7921e288f3a1264c0f925213be4c doc_id: 687215 cord_uid: ptdgqjb4 nan Reviewer #1 I think is a very interesting article, above all in the current situation. My only comment is that taking into consideration the special circumstances that we have recently lived, it would be adequate, to add at the end an small sentence referring to the necessity of further analysis ,instead of rejecting completely the possible use of sharing ventilators in these exceptional circumstances. We are grateful about the Reviewer's interest in our manuscript. We also agree that the sharing of a mechanical ventilator should not be so firmly rejected. We therefore included the following phrase at the end of the manuscript: "Further research and safety measurements are required before it could be recommended in exceptional circumstances". Reviewer #2 Un trabajo muy sencillo pero sin embargo muy interesante y con aplicabilidad clínica. A pesar de lo complejo del tema que tratan, esta explicado de manera entendible y sobre todo con el apoyo de la gráfica. En estos tiempos donde tan importante se ha demostrado la ventilación mecánica (esto no quiere decir que antes no lo fuese) creo que los profesionales sanitarios que atendemos a pacientes respiratorias debemos entender la fisiopatología del paciente que estamos tratando. Felicidades a los autores. Muchas gracias por el interés y la interpretación realizada acerca de nuestro trabajo. Esperamos que la versión actual del manuscrito se encuentre apta para su publicación en Archivos de Bronconeumología. La pandemia por SARS-CoV-2 ha determinado una situación de emergencia sanitaria, desafiando las capacidades de los sistemas de salud pública. La escasez de ventiladores mecánicos representa una preocupación importante, al punto que ha llevado a ciertos centros a considerar el uso de un único ventilador mecánico para ventilar dos pacientes en forma simultánea. Esto ha desencadenado numerosas preocupaciones a nivel de distintas sociedades científicas. La distribución desigual del volumen corriente entre los pacientes constituye uno de los principales riesgos inherentes a esta práctica. En este estudio basado en simulación reprodujimos diferentes escenarios clínicos que podrían presentarse al ventilar dos pacientes con un mismo equipo de ventilación mecánica. En particular, nos centramos en la distribución del volumen corriente a medida que se modificaba la mecánica respiratoria, como podría esperarse en pacientes con COVID-19. Si bien gran parte de la comunidad médica percibe los riesgos que podrían esperarse al utilizar esta técnica, este estudio de prueba de concepto documenta de forma contundente estos hechos. En momentos en que se plantean diversas terapéuticas por fuera de los usos habituales, creemos que la publicación de este manuscrito en Archivos de Bronconeumología sería de importancia para crear conciencia sobre los riesgos que la ventilación múltiple puede representar para los pacientes. Dejamos constancia que ninguno de los resultados han sido previamente publicados. Todos los autores del manuscrito están de acuerdo con su contenido y ceden los derechos I think is a very interesting article, above all in the current situation. My only comment is that taking into consideration the special circumstances that we have recently lived, it would be adequate, to add at the end an small sentence referring to the necessity of further analysis ,instead of rejecting completely the possible use of sharing ventilators in these exceptional circumstances. We are grateful about the Reviewer's interest in our manuscript. We also agree that the sharing of a mechanical ventilator should not be so firmly rejected. We therefore included the following phrase at the end of the manuscript: "Further research and safety measurements are required before it could be recommended in exceptional circumstances". Un trabajo muy sencillo pero sin embargo muy interesante y con aplicabilidad clínica. A pesar de lo complejo del tema que tratan, esta explicado de manera entendible y sobre todo con el apoyo de la gráfica. En estos tiempos donde tan importante se ha demostrado la ventilación mecánica (esto no quiere decir que antes no lo fuese) creo que los profesionales sanitarios que atendemos a pacientes respiratorias debemos entender la fisiopatología del paciente que estamos tratando. Felicidades a los autores. Muchas gracias por el interés y la interpretación realizada acerca de nuestro trabajo. Esperamos que la versión actual del manuscrito se encuentre apta para su publicación en Archivos de Bronconeumología. To the Editor: The coronavirus disease 2019 (COVID-19) pandemic has created a public health emergency challenging the health care system capabilities. The shortage of medical resources, in particular of mechanical ventilators represents a major concern, leading to some centers considering the use of a single mechanical ventilator for two patients (co-venting). Protocols designed to co- To describe the impact that different C and R would have on VT during co-ventilation, a mechanical ventilator (Puritan Bennett 840, Medtronic, Minneapolis, MN) was connected to a dual-chamber lung simulator (Training and Test Lung, Michigan Instruments, Grand Rapids, MI) using two tubing sets connected through T-tubes, as previously described. 3 Each of the simulator chambers represented a different patient (simulated case #1 and #2, respectively). Stable, relatively normal C (50 mL/cmH2O) and R (5 cmH2O/sec) were maintained for case #1 Manuscrito con cambios resaltados/Manuscript with hgihlighted changes 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 throughout the experiment, while different abnormal conditions were simulated for case #2. Pressure, flow and VT were registered for each chamber individually (SAMAY MV16, Uruguay). During PCV the ventilator was set at peak pressure of 18 cmH2O, positive end-expiratory pressure (PEEP) of 10 cmH2O, respiratory rate of 15 breaths/min, inspiratory-expiratory ratio of 1:2. Mechanical ventilation was initiated with identical C (50 mL/cmH2O) and R (5 cmH2O/sec) for both simulated patients and baseline measurements were obtained. Afterwards, different pathological scenarios were simulated to occur to case #2. Progressive reduction of lung C (maintaining R = 5 cmH2O/sec) resulted in a substantial contraction of VT for case #2, leading to a decrease of up to 18% from baseline when C was 10 mL/cmH2O. Case #1 presented a gradual but modest reduction of VT as C of case #2 declined (Figure 1a) . Later, airway R of case #2 was increased while maintaining C at 50 mL/cmH2O (Figure 1b) . Tidal volume was relatively preserved for case #1 and case #2 at R = 20 cmH2O/sec (98% and 89% from baseline, respectively). However, a severe increase in R (50 cmH2O/sec) resulted in a drastic reduction of VT for case #2, while a minor decrease was observed for case #1 (52% and 91% from baseline, respectively). The same experimental protocol was repeated in volume-controlled ventilation (VCV) with VT set at 800 mL while maintaining the other settings unchanged. As observed in PCV, the decrease in lung C or increase in airway R determined a progressive reduction of VT for case #2. More importantly, this reduction was paralleled by an increase in VT for case #1 (Figure 1c-d) . Therefore, case #1 and case #2 could potentially receive highly unequal VT such as 177% and 32% from baseline, respectively (C = 10 mL/cmH2O). Ventilating two patients with a single mechanical ventilator has been proposed as a last resort in a crisis standard of care, as could occur during COVID-19 pandemic. This strategy obviously 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 presents significant limitations that could expose both patients to an excessive risk of adverse events. Changes in respiratory mechanics may occur unexpectedly as a result of diverse situations (bronchospasm, secretions, hyperinflation, lung edema, pneumothorax, etc.). Branson et al. have already shown the disparity of VT distribution among four simulated patients connected to a single ventilator, as C and R were modified. 4 Here, we aimed to reproduce a scenario that we believe is more likely to occur during the COVID-19 outbreak, co-ventilating two simulated patients that might present relatively preserved or extremely abnormal respiratory mechanics. 2, 5, 6 As we demonstrated in this simulation-based analysis, variation of a single characteristic (C or R) on one patient can drastically affect the way gas is distributed. Of note, we observed similar results when simulation was performed at different PEEP levels (5, 10 and 15 cmH2O). Our study design represents an oversimplification of what might occur on a clinical setting, in which both patients could present changes in C and R, in similar or opposite directions. In this scenario, both patients (in different ways) could be exposed to a significant risk of hypo or hyperventilation, with hypercapnia and volutrauma among the most feared consequences. Despite a thoughtful setting of alarm parameters, without individual respiratory monitoring the entailed risk of late detection of these phenomena is too high. As expected, VCV was associated to a more uneven distribution of VT, significantly increasing the risks. In summary, ventilator sharing could result in deleterious effects related to inadequate VT distribution. This study addressed a single aspect of the issues related to patient co-venting, using a simulation experimental setting, while many other concerns remain to be studied. Notwithstanding, ventilating two patients with a single mechanical ventilator appears to be an unsafe practice. Further research and safety measurements are required before it could be recommended in exceptional circumstances. 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 1 To the Editor: The coronavirus disease 2019 (COVID-19) pandemic has created a public health emergency challenging the health care system capabilities. The shortage of medical resources, in particular of mechanical ventilators represents a major concern, leading to some centers considering the use of a single mechanical ventilator for two patients (co-venting). Protocols designed to co- To describe the impact that different C and R would have on VT during co-ventilation, a mechanical ventilator (Puritan Bennett 840, Medtronic, Minneapolis, MN) was connected to a dual-chamber lung simulator (Training and Test Lung, Michigan Instruments, Grand Rapids, MI) using two tubing sets connected through T-tubes, as previously described. 3 Each of the simulator chambers represented a different patient (simulated case #1 and #2, respectively). Stable, relatively normal C (50 mL/cmH2O) and R (5 cmH2O/sec) were maintained for case #1 Manuscrito (anónimo)/Manuscript (anonimous ) 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 throughout the experiment, while different abnormal conditions were simulated for case #2. Pressure, flow and VT were registered for each chamber individually (SAMAY MV16, Uruguay). During PCV the ventilator was set at peak pressure of 18 cmH2O, positive end-expiratory pressure (PEEP) of 10 cmH2O, respiratory rate of 15 breaths/min, inspiratory-expiratory ratio of 1:2. Mechanical ventilation was initiated with identical C (50 mL/cmH2O) and R (5 cmH2O/sec) for both simulated patients and baseline measurements were obtained. Afterwards, different pathological scenarios were simulated to occur to case #2. Progressive reduction of lung C (maintaining R = 5 cmH2O/sec) resulted in a substantial contraction of VT for case #2, leading to a decrease of up to 18% from baseline when C was 10 mL/cmH2O. Case #1 presented a gradual but modest reduction of VT as C of case #2 declined (Figure 1a) . Later, airway R of case #2 was increased while maintaining C at 50 mL/cmH2O (Figure 1b) . Tidal volume was relatively preserved for case #1 and case #2 at R = 20 cmH2O/sec (98% and 89% from baseline, respectively). However, a severe increase in R (50 cmH2O/sec) resulted in a drastic reduction of VT for case #2, while a minor decrease was observed for case #1 (52% and 91% from baseline, respectively). The same experimental protocol was repeated in volume-controlled ventilation (VCV) with VT set at 800 mL while maintaining the other settings unchanged. As observed in PCV, the decrease in lung C or increase in airway R determined a progressive reduction of VT for case #2. More importantly, this reduction was paralleled by an increase in VT for case #1 (Figure 1c-d) . Therefore, case #1 and case #2 could potentially receive highly unequal VT such as 177% and 32% from baseline, respectively (C = 10 mL/cmH2O). Ventilating two patients with a single mechanical ventilator has been proposed as a last resort in a crisis standard of care, as could occur during COVID-19 pandemic. This strategy obviously 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 presents significant limitations that could expose both patients to an excessive risk of adverse events. Changes in respiratory mechanics may occur unexpectedly as a result of diverse situations (bronchospasm, secretions, hyperinflation, lung edema, pneumothorax, etc.). Branson et al. have already shown the disparity of VT distribution among four simulated patients connected to a single ventilator, as C and R were modified. 4 Here, we aimed to reproduce a scenario that we believe is more likely to occur during the COVID-19 outbreak, co-ventilating two simulated patients that might present relatively preserved or extremely abnormal respiratory mechanics. 2, 5, 6 As we demonstrated in this simulation-based analysis, variation of a single characteristic (C or R) on one patient can drastically affect the way gas is distributed. Of note, we observed similar results when simulation was performed at different PEEP levels (5, 10 and 15 cmH2O). Our study design represents an oversimplification of what might occur on a clinical setting, in which both patients could present changes in C and R, in similar or opposite directions. In this scenario, both patients (in different ways) could be exposed to a significant risk of hypo or hyperventilation, with hypercapnia and volutrauma among the most feared consequences. Despite a thoughtful setting of alarm parameters, without individual respiratory monitoring the entailed risk of late detection of these phenomena is too high. As expected, VCV was associated to a more uneven distribution of VT, significantly increasing the risks. In summary, ventilator sharing could result in deleterious effects related to inadequate VT distribution. This study addressed a single aspect of the issues related to patient co-venting, using a simulation experimental setting, while many other concerns remain to be studied. Notwithstanding, ventilating two patients with a single mechanical ventilator appears to be an unsafe practice that should not be encouraged. 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 (Figure) Anesthesia Patient Safety Foundation, American Association of Critical-Care Nurses COVID-19 pneumonia: Different respiratory treatments for different phenotypes? A single ventilator for multiple simulated patients to meet disaster surge Use of a single ventilator to support 4 patients: Laboratory evaluation of a limited concept Lung recruitability in SARS-CoV-2 associated acute respiratory distress syndrome: A single-center, observational study COVID-19 in critically ill patients in the Seattle region -Case series Anesthesia Patient Safety Foundation, American Association of Critical-Care Nurses, and American College of Chest Physicians COVID-19 pneumonia: Different respiratory treatments for different phenotypes? A single ventilator for multiple simulated patients to meet disaster surge Use of a single ventilator to support 4 patients: Laboratory evaluation of a limited concept Lung recruitability in SARS-CoV-2 associated acute respiratory distress syndrome: A single-center, observational study COVID-19 in critically ill patients in the Seattle region -Case series ¿En su trabajo intervienen pacientes o sujetos humanos?: No 3. ¿Su trabajo incluye un ensayo clínico?: No 4. ¿Todos los datos mostrados en las figuras y tablas incluidas en el manuscrito se recogen en el apartado de resultados y las conclusiones? Page 16 of 17