key: cord-0684433-an85m397 authors: González-López, Andrea; Fernández Abedul, M. Teresa title: Chapter 21 Genosensor on gold films with enzymatic electrochemical detection of a SARS virus sequence date: 2020-12-31 journal: Laboratory Methods in Dynamic Electroanalysis DOI: 10.1016/b978-0-12-815932-3.00021-8 sha: a18739beb3a69592f01e65188c99154897a5363f doc_id: 684433 cord_uid: an85m397 Abstract Biosensors are one of the best examples of miniaturization and simplification trends in analytical chemistry. Among them, DNA sensors, with oligonucleotides as recognition elements, are being widely utilized for many applications, mainly diagnosis of infectious diseases. In this experiment, a hybridization-based genosensor was designed on a 100-nm sputtered gold film. This material worked as an immobilization and transduction surface for the electrochemical diagnosis of a specific DNA sequence of SARS (severe acute respiratory syndrome) virus. The simplicity of the electrochemical instrumentation is combined with a selective methodology to detect the hybridization event enzymatically. This chapter explains all the necessary steps for the construction of the genosensor allowing a low detection limit for the chosen DNA sequence. Protocols for studying the influence of parameters affecting the methodology are given. This chapter also includes questions for students and indications for writing a lab report. . On the other hand, the use of indicator molecules that interact differentially with singleor double-stranded DNA has been widely exploited in indirect approaches. Bioconjugation is commonly not required because electrostatic or hydrophobic interactions take place easily. A different possibility for indirect detection is the use of enzymes, alternative that enhances assay sensitivity enormously because of their inherent amplification. Although enzymes can be directly conjugated to DNA strands, the use of a well-known affinity interaction such as avidinebiotin is an easy alternative. Horseradish peroxidase has been widely used as enzyme label as well as glucose oxidase and alkaline phosphatase (AP). AP (EC 3.1.3.1) is a hydrolase that converts orthophosphoric monoesters into alcohols and thus various phosphate esters can act as substrates. As the name suggests, they are more effective in alkaline media. 3-Indoxyl phosphate (3-IP) has been proposed as an adequate substrate for AP in immunoelectrochemical approaches [6] . Because of its satisfactory electrochemical behavior (see Chapters 5, 30 , and 31), it has been used for the development of AP-based enzyme immunoassays. In this experiment, AP is used to detect DNA hybridization events on gold electrodes. The surface of the genosensor is of paramount importance in the performance of the analytical device because it is where immobilization and transduction take place. Gold has always been an appropriate material for the electrochemical detection of substances, while also allowing different formats: disks, wires, films, etc. Thin-film (thickness in the nm-scale) gold electrodes can be produced by sputtering, a physical vapor deposition methodology in which gold is ejected from a target in a vacuum chamber onto a substrate. Gold also offers the relevant possibility of generating self-assembled DNA monolayers through thiol groups because of the strong binding between sulfur and gold, which may be considered almost covalent. In this chapter, a DNA hybridization assay with enzymatic electrochemical detection (see Fig. 21 .1) is carried out on a 100-nm sputtered gold film that allows working with small volumes. Reducing the cell volume has several advantages, the first being the decrease in the diffusion distances required for analytes to reach their surface-bound receptor partners. A simple, cheap, and easy-to-handle homemade device is presented allowing the performance of the hybridization procedure and sequential detection. Square wave voltammetry (SWV) (see Chapters 3 and 24) is employed for measuring in a fast and sensitive way. This experiment is adapted from Ref. [7] . The sequence chosen as target for the development of this genosensor corresponds to a gene that encodes the nucleocapsid protein (422 amino acids) of the SARS (severe acute respiratory syndrome) virus, specifically a short lysine-rich region that appears to be unique to SARS and suggestive of a nuclear localization signal [8] . A 30-mer oligonucleotide with bases comprised between numbers 29,218 and 29,247, both included, was chosen. A complementary strand to the chosen SARS sequence is labeled with a thiol group and immobilized on the gold surface. The target (30-mer oligonucleotide with a sequence included in the SARS coronavirus) is conjugated to biotin and hybridized with the probe. Addition of AP-labeled streptavidin (St) allows enzymatic detection via the electrochemical signal of the product. The parameters affecting all assay steps are studied, and analytical characteristics, including selectivity, can be discussed. The electrochemical cell (see in more detail in Section 21.5.1) consists of 5 cm  5 cm supports of a 0.125 mm thick polyimide substrate that are sputtered with gold (Emitech sputter coater model K550) and after delimiting areas, act as working electrodes. A miniaturized Ag/ AgCl/saturated KCl electrode is employed as reference and a platinum wire as auxiliary electrode. All the three electrodes are connected with alligator clips to the potentiostat. -Hybridization buffer (2x SSC): 30 mM sodium citrate buffer with 300 mM NaCl, pH 7 (Table 21 .1). -Milli-Q water is employed for preparing solutions. -Electrode fabrication: Polyimide substrate, conductive wire, epoxy resin (CW2400), selfadhesive washers with 5 mm internal diameter (19.6 mm 2 internal area), crocodile clips, silver wire, platinum wire, syringe rubber piston, micropipette tip, insulating tape. -Apparatus: Weighing scale, pH meter, and incubator. -Other materials: 100-mL and 1-mL micropipettes with corresponding tips, 1.5-mL microcentrifuge tubes. Students are required to wear lab coat, appropriate gloves, and safety glasses. Special care should be taken when handling concentrated acids. The three-electrode potentiostatic system used for this experiment is schematized in Fig. 21. 2. Working electrodes are made on 5 cm  5 cm supports of a 0.125 mm thick polyimide substrate called Kapton HN. They are covered with gold by a sputtering process explained on Section 21.5.2. A conductor wire is attached to the center of one of the sides by means of an epoxy resin (CW2400) and it is cured at room temperature. The working area is limited by self-adhesive washers of 5 mm internal diameter. The gold film is placed on a support to which an alligator connection is fixed. Reference and auxiliary electrodes are coupled in a micropipette tip. The reference electrode is an anodized silver wire introduced in the tip through a syringe rubber piston. In turn, the tip is filled with saturated KCl solution and contains a low-resistance liquid junction. The platinum wire acts as auxiliary electrode and it is fixed with insulating tape. For measurement recording, the tip is fixed on an electrochemical cell Metrohm support, allowing horizontal and vertical movement. The working electrodes made on polyimide supports are covered with gold by a sputtering process. The Kapton slide (5 cm  5 cm) must be cleaned with ethanol and left drying. Once dried, the slide is covered with gold. Gold atoms are deposited (from the cathode) over the Kapton (placed on the anode) in a vacuum chamber filled with argon. The thickness of the gold layer is controlled by means of the duration and intensity of the discharge (possible values are a 35 mA discharge applied for 220 s). wait 10 min for its enzymatic hydrolysis. 11. Add, then, 5 mL of concentrated H 2 SO 4 to stop the reaction. 12. Finally, add 5 mL of Milli-Q water and perform the electrochemical measurements. The tip with the reference and auxiliary electrodes is introduced into the 30-mL drop (20 mL from the 3-IP solution plus 5 mL of the concentrated H 2 SO 4 and 5 mL of Milli-Q water) deposited on the gold film. A potential of À0.35 V is applied for 30 s before scanning the potential between À0.15 and þ0.3 V following a square wave format with frequency of 50 Hz and amplitude of 50 mV. Measure the peak current, which is the analytical signal that can be correlated with the concentration. Evaporation seems to be a critical condition in the immobilization of SH-DNA. A more rigorous study (step 1e2 of Section 21.5.3) could be done. Using a drop of 5 mL and a strand concentration of 1 mM, different immobilization times can be employed at room temperature (22 C approximately), 37 and 47 C. Note down when evaporation occurred and compare the signals obtained. Select the best immobilization conditions. Blocking the surface is one of the most important steps to minimize and control nonspecific adsorptions. Two main types of agents can be considered: proteins and sulfur-containing compounds. The signal/background ratio, and therefore the blocking capacity, should be studied for each compound. Proteins such as albumin are common blocking agents in bioanalysis and adsorb well onto gold surfaces. Evaluate the nonspecific signal obtained after blocking (step 4 in Section 21.5.3) with aqueous solutions of BSA, casein, and a gelatine derivative called Perfect Block. The other group of blocking agents are compounds that contain sulfur atoms and therefore present a high affinity for gold. Ethanolic solutions of thioctic acid, 6-mercapto-1-hexanol, and 1-hexanethiol could be tested. It is very interesting to note the effect of these agents on both capacitive and faradaic currents. A calibration curve for the complementary strand (c-DNA) can be performed under optimized conditions, employing a 1-mM solution of SH-DNA for immobilization. As an example, concentrations in the 0.1e10 nM interval could be tested (step 6 in Section 21.5.3). SWV measurements for different concentrations are included in Fig. 21.3 . Plot the values of the peak current for different concentrations and fit the data to a regression curve. Take into account that it can be nonlinear and fit into logarithmic or a different model. Determine the dynamic range, sensitivity, and limit of detection of the assay. Precision studies of the platform could be made by obtaining measurements with the genosensor on different working areas, polyimide substrates, days, or groups. Evaluate the results through the value of the relative standard deviation of the peak current. Write a lab report following the typical scheme of a scientific article, including a brief introduction, experimental part (materials, equipment, and protocols), results and discussion, and conclusions. The following points should be beard in mind: 1. In the introduction, explain the purpose of the experiment and do a short review of the methods described to tackle this problem. 2. Protocols must be suitably detailed including schemes preferentially where appropriate and the necessary calculations. 1. Oligonucleotide solution aliquots must be prepared and maintained at À20 C and working solutions must be conserved at 4 C. 2. IC solutions can be employed to know the electrochemical behavior of the enzymatic product (and the analytical signal). They must be protected from light and kept refrigerated at 4 C. Working solutions must be prepared daily. 3. 3-IP solutions must be prepared daily and kept at 4 C, protected from light. 4. AP-labeled St aliquots must be prepared and maintained at À20 C; working solutions are conserved at 4 C. 5. A drop of 5 mL and a strand concentration of 1 mM are employed in the immobilization steps (step 1 in Section 21.5.1), but both variables could be varied to study their influence. 6. The blocking of the active surface remaining after immobilization is very important in bioassays. Different agents could be checked, as commented in Section 24.5.2. The influence of different concentrations could also be evaluated. 7. SWV is employed for measurement because it is a fast and sensitive electrochemical technique. However, cyclic voltammetry or differential pulse voltammetry could also be evaluated. In particular, cyclic voltammetry should be made initially to know the electrochemical behavior and processes of 3-IP. 8. Selectivity of the genosensor can be studied by evaluating the signal of a, e.g., 3-base mismatch strand: 5 0 -ACA-GCG-CCT-AAA-AAC-GAC-AAA-AAG-AG-AAG-3 0 -biotin. Mismatches are located in base numbers 5, 15, and 26. It is also biotinylated at the 3 0end to allow hybridization detection by interaction with AP-labeled St. Adding agents that increase stringency (e.g., 50% of formamide) should be considered. 9. The sensitivity could be improved using different conditions (drop volume, time of the different steps, buffer composition, etc. [7] ). Students are encouraged to discuss and evaluate the different variables. 10. In this case, a proof of concept of a biosensor able to detect SARS DNA is presented. The target is labeled with biotin. Then, in a real assay, DNA would have to be amplified using biotinylated primers. Alternatively, a sandwich format (thiolated capture probeetargetebiotinylated detection probe) should be employed. 1. Indicate clearly all the steps of the procedure. The use of a scheme is encouraged. 2. Why a DNA strand is functionalized with a thiol group? 3. What is the role of the blocking agent? Enumerate the different possibilities. 4. What is the aim of employing the biotineavidin interaction here? 5. Explain how the analytical signal is obtained, especially in what concerns to the electrochemical technique employed. Smartphone-based biosensors: a critical review and perspectives DNA biochip arraying, detection and amplification strategies Electrochemical biosensors Direct electrochemistry of DNA, guanine and adenine at a nanostructured filmmodified electrode 3-Indoxyl phosphate: an alkaline phosphatase substrate for enzyme immunoassays with voltammetric detection Genosensor on gold films with enzymatic electrochemical detection of a SARS virus sequence Genosensor on gold films with enzymatic electrochemical detection of a SARS virus sequence