key: cord-0673430-2dwceqmv authors: Mazhar, Rashid; Chowdhury, Muhammad E. H.; Ramzee, Ahmed Faidh; Raza, Tasleem; Surgery, Abdulaziz Ahmed Al-Hashemi Pulmonology-Thoracic; Hospital, Hamad General; Doha,; Qatar,; Engineering, Department of Electrical; University, Qatar title: Dynamic Ventilatory circuit divider: Safer way to multiply the number of available ventilators date: 2020-05-13 journal: nan DOI: nan sha: db1e746b351c93934d66bf36531eed5323644bda doc_id: 673430 cord_uid: 2dwceqmv Background: Recent pandemic has brought a sudden surge in the requirement of mechanical ventilators all over the world. In this backdrop, there is wide interest in looking for ways to support multiple patients from a single ventilator. Various solutions, based upon simple mechanical division of the ventilator tubings are described. However, recently warnings have been issued by multiple international professional societies against these plumbing solutions as it can seriously harm the patients. Methods: We have bifurcated the inspiratory and expiratory conduits from a single ventilator with addition of one way valves, pressure and flow sensors along with volume and PEEP control. A purpose-built software and a low-cost microcontroller based control system integrates and displays the data in the familiar ventilatory graphic and numerical format onto a generic screen. The system is calibrated with simulated lungs with varying compliance. In addition to the standard microbial, heat and moisture exchanger filters we design to add UV-C lights at 254-260 nanometre wavelength in the expiratory channel for its virucidal effect. Results: The dynamic ventilatory divider system is capable of providing and controlling individual flow, tidal volume (TV) and Positive End-expiratory pressure (PEEP) for individual patients. Furthermore, it would also display the ventilatory parameters of both the patients, on a single split screen, in a familiar format. FiO2 and rate are still controlled by the mother ventilator. Conclusions: The prototype system has a potential to provide safe ventilation to at least two individuals from a single ventilator, while maintaining the unique requirements of each patient. With the ongoing pandemic of COVID-19, mechanical ventilators have become a rate-limiting therapeutic tool 1 . Manufacturing large numbers of ventilator machines is time consuming, besides the economic consideration in low-income countries. To fulfil this shortage, the popular media and some professionals have propagated an immediate solution by simple division of the ventilator tubings with suitable plumbing of the ventilator tubing with suitable connectors. Historically, this type of passive mechanical, divider solution is described in literature. These endeavours/reports fall in three categories: In-vitro experiments on lung simulators, In-vivo animal experiments and human study. Sommer et al. 2 and Neyman et al. 3 describe simple connectors dividing the ventilator circuit on plumbing fashion, without any individual provisions of pressure, volume sensing, control and display. These tests were carried with the presumption of equal ventilation to all four lung-simulators, presuming equal lung physiology. Branson 4 wrote a letter to the editor in criticism to the study reported by Neyman et al. 3 . Moreover, Branson et al. 5 have reported similar experiments to monitor individual parameters where the divided circuits were attached to dedicated pneumotachograph connected to an individual respiratory monitor. The basic fact to be noted here that gas flow would follow the path of least resistance. The tidal volume for individual patients would be distributed according to the variability of compliance of lung rather than theoretical equal distribution. This type of apparent solution will cause severe lung injuries rather than supporting the patients as better lung (with higher compliance) will get more tidal volume and the stiffer lung (with lower compliance) will get less volume even though the requirement might be completely opposite. Other critical respiratory problem and a cause of concern was the lack of control of positive endexpiratory pressure (PEEP). Hence, the apparently attractive concept of supporting multiple patients from a single ventilator by simple division/branching of the pipes is flawed and dangerous. Paladino et al. 6 carried out animal study using same simple division concept on healthy sheep for 12 hours. Again, there were no individual provisions of pressure, volume sensing, control and display. Furthermore, they were criticised 7 as being good only for supporting paralyzed, size-matched, and sedated individuals with normal lungs who are hemodynamically stable. The solution was deemed biologically irrelevant for mass respiratory failure, except possibly botulism (neurotoxicity with healthy lungs). One published report 8 does describe the use of split ventilation in two volunteers (using a facemask interface rather than intubation). However, there was no provision for monitoring, controlling and display of individual ventilatory parameters. Thus all the above, apparently one ventilator-multiple patients solutions remained out of practical use. With the dire current need of more and more ventilators, there is a resurgence in reporting of similar passive divider solutions. However, this apparent simple divider solution is not recommended and recently warnings have been issued by individuals as well as multiple international professional societies against these as it can seriously harm the patients. 9-11 The reasons against this "blind, uncontrolled division" of the ventilatory circuit include 9 : • Inability to control individual volumes. With the branching circuits, the gas would follow the path of least resistance and the patient with stiffer lungs would take less and less volume while the patient with more compliant lungs would preferentially accumulate the incoming volume, harming both types of patients. • Inability to manage individual PEEP. • Inability to monitor individual patient's vital pulmonary mechanics. • In case of sudden deterioration of a single patient (e.g., pneumothorax, kinked endotracheal tube), balance of ventilation would be distributed to the other patients. • Risk of cross infection to all the patients attached to a single branching circuit. It is suggested that to circumvent these shortcomings, with a volume-cycled mode, the deeply sedated patients having similar severity of lung injury, and preferably of the same body size are grouped together. The ventilator should be set up to pressure-cycled ventilation with a high PEEP and a low driving pressure (to achieve lung protection) and the ventilatory efficacy of each patient to be tracked using an end-tidal CO2 monitor placed in-line with their own endotracheal tube, allowing permissive hypercapnia. Furthermore, it is suggested that viral filters should be attached to prevent crosscontamination of pathogens between different patients 10 . Keeping these practical concerns in mind, we have devised the functional prototype of a "dynamic ventilatory circuit divider", whereby a single ventilator could be used to ventilate multiple patients, with individual display and control of the essential, required parameters of ventilation. Furthermore, it provides an anti-microbial provision, which is inherently anti-viral as well. In the proposed system, mechanical, electronic and anti-microbial modifications were done to the ventilatory circuit. In principle, we have based our solution on the following functional aspects: C) PEEP control: Mechanical, gated valves were placed along the expiratory conduits to graduate the PEEP in the target lungs. This is possible with the help of one-way valve placed in the inspiratory and expiratory limbs for individual circuit, which provides a closed path outside mother ventilator. The mother ventilator along with its inspiratory and expiratory interfaces and display was interfaced to an add-on system, where an electronic control system and additional graphical display were used ( Figure 1 ). This additional display can be computer, laptop or any other single-board computer with screen and universal serial bus (USB) interfaces. The electronic control system is a low-cost microcontroller based system, which is the brain of the dynamic bifurcated system and the graphical display was used as a user-interface only. In future, both of the system can be combined to have a control system along with graphical interface. A) Electronic sensors: Two different sensors were used for each patient: Bi-directional flow sensor (SFM 3300-AW, Sensirion AG, Switzerland) and a Pressure sensor (MS4525-A5, TE connectivity). A bidirectional flow sensor placed distal (towards the patient) to the inspiratory control gate valve ( Figure 1) . A pressure sensor placed at the end of the expiratory tube along with the one-way valve (as shown in Figure 1) . The GUI shows the real-time flow, volume and pressure graphs and numerical values like a typical ventilator for individual patient. It immediately reflects any change happened to any of the divided system independently. A familiar digital display on a generic digital screen, which is divided to show both/all the patients ventilatory parameters on one screen, at the same timeakin to central monitoring display (Figure 2 ). The innovation has the capability to a wireless display mode where multiple patients' management could be supervised by tale-medicine. Furthermore, each individual circuit, coming out of the ventilator and the patients, is provided with a suitable standard bacterial filters as well as UV-C lights at 254-260 nanometre wavelength for its Virucidal effect. These tubing segments, carrying the UV light source are lined by internally reflective coating to prevent external scatter and plastic decay. The control system of the prototype system gathers flow and pressure information from the sensors independently. The control unit is connected to the computer using USB interface to send data to the computer to display the individual patient parameters and graphs using the in-house built graphical user-interface using processing application. Both the flow and pressure sensors were calibrated using the mother ventilator while connected to two lung simulators of varying compliance (Figure-2) . The calibration process is beyond the scope of this article, can be reported in a separate manuscript. All the conversion and calculation were done in the microcontroller and graphical user-interface (GUI) was used to provide an interactive but familiar interface to the respiratory therapists. Since the flow sensor is a bi-directional sensor, it can provide real-time inspiratory and expiratory flow directly in L/Min unit. This was directly used to show the real-time flow measurement in the GUI, however, the tidal, inspiratory and expiratory volume were calculated from the flow data in the microcontroller using the time of inspiration and expiration. On the other hand, the pressure sensor measures the pressure in the expiratory tube with respect to the air pressure and provides analogue voltage output according to the Our solution overcomes the "blind, uncontrolled division" approach suggested in the social media by a "dynamic ventilatory circuit divider" which provides each individual circuit branch, and hence each patient, with volume, pressures and flow rate sensing as well as controlling capability. Two demo lungs' data/graphs were displayed on a single screen of a computer making it possible for one caregiver to manage both patients more ergonomically. Different system components with the mechanical control, electronic sensing and control and userinterface is shown in Figure 3 (A), while Figure 3 (B) shows the dynamic bifurcated system is in operation. It can be clearly seen that individual patient's graphs and numerical data are shown side- A B by-side in the similar template of the conventional ventilator ( Figure-3 B) , which would make it easy to be used by the doctors, nurses and the respiratory therapists. The proposed system has been evaluated for different clinically relevant scenarios to evaluate its functionality. Initially, mother ventilator was set to 800 mL tidal volume with peak pressure of 24 cmH2O and PEEP of 5 cmH2O. Although both the lungs were of similar compliance one lung was getting tidal volume of ~400 mL, while other is getting ~360 mL. This discrepancy is due to the imperfect circuitry of the prototype where the two limbs had slightly different lengths and ensuing dead space. Our setup can be used for both pressure and volume controlled ventilation in controlled mode. Although the data in COVID-19 pulmonary involvement is still evolving, preliminary data in a significant proportion suggest a phenotypic pattern, which is different from classical acute respiratory distress syndrome (ARDS). These patients show severe hypoxemia and high dead space fraction, but lung compliance is relatively preserved. Therefore, patients generally do not require high PEEP, but high minute ventilation and oxygen supplementation. Non-invasive ventilation and high flow oxygen use in COVID-19 is discouraged, as they are high aerosol generating procedures and high infection risk for staff. Simple ventilatory modes may be adequate in these patients to provide oxygen support and therefore multiple-patient-single-ventilator may be a feasible strategy. The provided solution is a method of dividing the ventilatory circuit with provision of monitoring, controlling and displaying the main ventilatory parameters of individual patients. It is to be noted that this set-up is NOT being presented as a substitute for one ventilator-one patient solution, which remains the gold standard. The authors do see the only possible utility of this innovation in situations where there is a dire need to ventilate more patients than the available ventilators. Even under these circumstances, we DO NOT recommend attaching more than two patients to a single ventilator. Minute ventilation requirements of patient may be drastically different, but this can be controlled with adjustment of tidal volume despite same respiratory rate for all patients. Patients will also require adequate sedation to avoid spontaneous breaths and ventilator patient asynchrony. During cardiac arrest situation, affected patient will require discontinuation from the circuit and main ventilator delivered volume will need readjustment for the remaining patient, in volume-controlled mode. This is to re-emphasize that any and all such solutions are suboptimal as compared to a one patientone ventilator setting. However, in the situation of gross mismatch of ventilator supply/demand, we believe that the presented solution is an economical, easy-to-adopt and relatively safer choice. One suggestion of immediate utility could be application of this solution to stable, long-term ventilated patients in the society to free desperately needed ventilators for seriously ill patients. Asian Critical Care Clinical Trials Group. Intensive care management of coronavirus disease 2019 (COVID-19): challenges and recommendations Improvised automatic lung ventilation for unanticipated emergencies A single ventilator for multiple simulated patients to meet disaster surge A single ventilator for multiple simulated patients to meet disaster surge. (author reply Use of a single ventilator to support 4 patients: laboratory evaluation of a limited concept Increasing ventilator surge capacity in disasters: ventilation of four adult-human-sized sheep on a single ventilator with a modified circuit One ventilator multiple patients-What the data really supports (Letter to the Editor) Simultaneous ventilation of two healthy subjects with a single ventilator Joint Statement on Multiple Patients Per Ventilator PulmCrit -Splitting ventilators to provide titrated support to a large group of patients Splitting one ventilator for multiple patientstechnical assessment This manuscript, including related data, figures and tables has not been previously published and the manuscript is not under consideration elsewhere. The authors declare no conflict of interest. We describe a safer method of using one ventilator for at least two patients. A bifurcated ventilatory circuit is provided with an array of one way valves, pressure and flow sensors, along with flow controllers. A microcontroller and specific software processes the data to display the ventilatory parameters of each patient on a single generic screen. Thus, barring the FiO2 and the respiratory rate, the presented design allows volume, pressures and flow rate sensing as well as controlling capability for individual patients.