key: cord-0647599-lvwng1ms authors: Krishnan, S.; Collins, S.; Smoors, D.; Webster, C.; Baroncelli, T.; Brooks, G.; Mould, J.; Dix, W. J.; McNamara, P.; Scutti, F.; Lane, G.; Urquijo, P.; Duffy, A. R. title: A scalable and reconfigurable industrial-grade Slow Control System for SABRE-South Dark matter experiment date: 2021-10-01 journal: nan DOI: nan sha: 013b8e6c5678e0ca7dbd0045c489d8d4834e4efc doc_id: 647599 cord_uid: lvwng1ms The Sodium iodide Active Background Rejection Experiment-South (SABRE-South) is a direct dark matter detector soon to be deployed in the Stawell gold mine, in Victoria, Australia. Monitoring of external environmental and experimental conditions, (temperature, barometric pressure, relative humidity, high voltage, and seismic vibration) is vital to ensure the data quality of the SABRE search for dark matter via direct detection. These parameters are generally non-time-critical in the range of Hz to a few kHz and constitute a slow control system. We present the design of a novel compact, industrial-grade, and self-contained slow control system for SABRE-South. This system, featuring innovative hardware and software architecture based on National instruments compact RIO (NI-cRIO) and LabVIEW can be scaled up at low-cost and is capable of implementing the functionalities available in high-end SCADA systems while maintaining the flexibility to integrate custom software code (i.e. C++, PYTHON) for bespoke interfacing needs. needs. Sodium iodide Active Background Rejection Experiment-South (SABRE-South), the first ultra-low background direct dark matter experiment in the southern hemisphere will be located in an active gold mine in Stawell, Victoria, Australia [3] . There is a need to monitor the experimental environment to ensure the data quality of the dark matter particle detection. The environment of the active gold mine is subject to seismic vibrations adding further complexity for maintenance, calibration and health of the detector. Large particle physics research centres such as Laboratori Nazionali del Gran Sasso (LNGS) have a well-supported technical infrastructure and use traditional supervisory control and data acquisition (SCADA) system for slow control. An example is the Xenon1T dark matter experiment [4] at LNGS. We present a self-contained, industrial-grade slow control system using National Instrument's Compact RIO (NI-cRIO) [14] , for SABRE-South, which has a functional performance similar to the traditional SCADA system, can operate reliably in active mine environments and is easily scalable as the experiment evolves. The NI-cRIO system is comprised of a real-time controller, reconfigurable input/output (I/O) modules (RIO), field-programmable gate array (FPGA) module and an Ethernet expansion chassis. This modular design enables researchers to add a large number of slow control parameters quickly and cost-effectively. The software design allows the storage of a large volume of slow-control data entirely in its local memory for the duration of the life of the experiment, without the need for any additional external servers and ensures the integrity of slow-data in the event of unexpected disruptions. The data is exported periodically to a remote database (i.e. a PostgresSQL server) for monitoring. The system caters to low-speed data acquisition recorded on a scale of seconds (i.e. temperature, barometric pressure, relative humidity) as well as high-speed, resource-intensive data (e.g. vibration) that requires a 5 kHz data rate. Some of the parameters currently identified for slow control moni-toring for SABRE-South include seismic vibrations, high-voltage (HV) supply for the Photomultiplier Tubes (PMTs), temperature, pressure and relative humidity. The design has been set up to monitor local parameters close to the SABRE vessel as well as remote parameters (up to a distance of 100m) through EtherCAT/(TCP/IP) in real-time. The data can be accessed by a remote database. In addition, the flexible design allows for the integration of bespoke systems such as muon flux monitors [11] for surface condition monitoring of seasonal modulations of cosmic muons, and purpose-built industry-standard power monitoring system to perform reliable long term measurements of the mains power supply. This open-ended and robust system has been operating (since March 2020), in the remote-access mode in the laboratory at the University of Melbourne, monitoring a muon coincidence detector composed of three EJ200 plastic scintillators (each of dimension 60cm x 30cm x 5cm), and read out by Hamamatsu R7724 PMTs, powered by a six-channel VME HV power supply module V6533. This set-up is planned to be deployed to the Stawell gold mine, at depths of 310m and 1025 m below the surface, to measure the muon flux as a function of depth. The slow control set up at the University of Melbourne is depicted in Figure 1 . One of the primary considerations in this design was to ensure reliable operation in the active mine environment for the lifetime of the experiment, and this mandated industrial-grade performance. For underground experiments such as SABRE that operate without on-site support, the system needs to be able to automatically reboot, in the event of power failure, and handle network disruptions gracefully. The system should be able to handle slow control sensors with different communication interfaces and data collection rates and also have the provision for easy expandability to accommodate additional parameters, as the experiment evolves. The Stawell Underground Physics Laboratory, (SUPL) is newly established, and should have the ability to adapt to multiple disparate software environments and application interfaces (APIs) and be able to leverage emerging technology trends to meet the current and future needs of the researcher. A mains power supply monitoring system has also been integrated into the slow control design, which could be used to measure the unpredictable mains power in the active mine. FPGA is used for data communication between the I/Os and cRIO controller. In some cases, FPGA performs pre-processing as well. in Figure 2 , the reconfigurable FPGA backplane of the slow control data logging system collects the data from the I/O modules and communicates it onto the host cRIO real-time processor (RTP). In some cases, such as vibration monitoring, the FPGA pre-processes the data before logging it to the host processor. The real-time processor and FPGA are programmed using LabVIEW (NI LabVIEW) [14] software. The program enables the RTP to store the data on the local onboard memory (SD card) for offline processing and also provide regular, periodic uploads to the remote database. The slow control outputs are displayed on the local human-machine interface (HMI). Figure 3 Vibration Sensors: Three independent vibration sources in the X, Y and Z directions are measured using three micro-electromechanical systems (MEMS) based accelerometers (IFM-VSA001) [8] , which generate a current output in the range of 0-10 mA. The FPGA sets the vibration event detection condition and performs low-level processing of the high-speed vibration data and streams the data to RTOS for data logging in the onboard memory and local display. A terminating resistor converts the current outputs of the vibration sensors to voltage, which is then fed to a 4-Channel, 51.2 kS/s/channel, +/−5 V, NI Vibration dynamic signal acquisition input module (NI-9234 [26] ), featuring inbuilt antialiasing filters, which allows for automatic adjustment of the sampling rate. The internal delta-sigma ADC of the NI-9234 is configured for a span of 0-5V and a sampling rate of 10.24 KS/s. Application of the Nyquist criterion [27] sets the maximum vibration frequency to ∼5 kHz, which meets the performance requirements of SABRE-South experiments and SUPL, as the maximum expected SUPL vibration frequency is ∼1 kHz. As shown in Figure 4 , the FPGA pre-processes the input from the sensors, calculates the G force parameters and determines if a valid event has occurred. G force is calculated by G force = I/S where I = V adc /R t , I is the output current of the vibration sensor, S is the sensitivity of the vibration sensor (S = 142µA/g) for IFM-VSA001, V adc is the measured ADC voltage and R t is the terminating resistor. R t is chosen to be 500Ω for an output current range (0 to 10 mA) and positive full span of ADC (0 to 5V). The resultant vibration amplitude G sum is generated by taking the vector sum of the vibration data in the X,Y and Z direction, . The moving window average G avg generates a filtered output of the vector sum which is compared to a set threshold and is used to determine valid vibration events. Each sample is bundled up into a data frame of G x , G y , G z , the vector sum amplitude G sum , mean amplitude (moving window average)(G avg ), and a marker ( 55 ) for simple frame validation. This data frame is stored in a shared First In First Out (FIFO) buffer in the FPGA, which can be read and processed by the host processor. This parallel and modular approach of using FPGA to pre-process the data reduces the CPU load of the RTOS. LabVIEW has been used to program For SUPL, as shown in eq. (1), the maximum expected barometric pressure, at a depth of 795m below sea-level, is 1150hPa, with an expected measurement accuracy of +/− 1hPa stipulated by the SABRE collaboration. To meet these high pressure requirements, high accuracy and reliable piezoresis- Two types of universal multichannel HV supplies, CAEN A1535SP and CAEN DT 5534 have been integrated into the slow control system. Each of the HV supplies communicates through different software and connection topologies. CAEN A1535SP HV supplies will be used to power the SABRE vessel PMTs which form a part of the active veto system [5] . The HV model The cRIO host processor (RTP) continuously logs data to three locally stored CSV files. The data include the slow environment data updated every 500 ms (2Hz), periodic logging of valid vibration data, and a log of system events as needed. A simplified flow-diagram of vibration data logging is shown in the solution for a simple interfacing task was ultimately deemed unnecessary. We explored options to implement a simple, compact, and energy-efficient solution using a more portable programming language. We have tested a shown in Figure 9 . The data transfer rate is slower with the RaspberryPi than with the PC, as shown in Representative results obtained with the slow control system during testing are discussed in this section. As shown in Figure 10 , the raw G-force vector sum represents the vibration amplitude and is logged in the cRIO onboard memory, while the filtered moving window average G-force is used as the source to generate an event detection threshold. As shown in Figure 11 , the slow control system displays the independently controlled HV of the four-channel DT 5534 HV module. In the Figure 11 , the pre-configured settings in channel 3 override user settings and protect the PMTs from accidental over-voltage. A representative time-stamped data logging CSV output is shown in Table 2 . The monitored value of temperature, barometric pressure and relative humidity is shown in Figure 12 and Figure 13 . The slow control system handles local (i.e.directly connected) and remote (i.e.connected through EtherCAT) sensors interchangeably, without additional configuration. A representation of the power meter display is shown in Figure 14 and We are currently testing the slow control system with a threechannel muon coincidence system at the University of Melbourne. The set up has been functioning reliably in an unstaffed environment (due to COVID-19 conditions) since 23 rd March 2020. We are monitoring the temperature, Figure 10 : Representative graph showing the G force vector sum and its moving window average for the tri-axial vibration sensor. A tri-axial sensor was mounted on a structural member of the building at Swinburne University and data was generated by light tapping of hammer on the building . Fig (a) shows the data logging of 4 consecutive data frames of the the G force vector sum and moving window average of the G force vector sum in case of a single valid vibration event. Fig (b) shows the magnitude vs time of occurrence of the G force vector sum and moving window average of the G force vector sum. and current values for CAEN DT 5534 four channel HV. An example warning status is shown in CH3. In this instance, a warning is issued when the user-set value (1800V) exceeds the preset HV values (1700V) which is set during pre-commissioning using GECO2020. barometric pressure, relative humidity and the environmental vibrations in the laboratory through a remote database. There were network connectivity issues which caused a disruption in data transmission to the remote database from 26 th March to 29 th April 2020. However, the onboard memory of the cRIO system is configured to keep a backup of all the slow control data. As a result, when network communication was restored, the system was able to automatically re-synchronise the data collected over this period with the external database. A representative sample of the acquired slow control data is shown in Figure 16 . The planned deployment of the slow control system to the Stawell mine, and long-term testing to cross-validate the slow control system data against the standard real-time mine environmental data, will be undertaken when lockdown restrictions are relaxed. The SABRE experiment needs to be sensitive enough to reliably measure dark matter search data in the presence of background. As shown in Figure 17 , the SABRE-South vessel will house a NaI(Tl) detector array having seven crystals, and fourteen PMTs, one on each end of the crystal, and a further 18 PMTs mounted in the active liquid scintillator veto system, Linear Alkyl Benzene (LAB). Since the PMT dark rates are highly susceptible to temperature variations [7] , High relative humidity could negatively impact the function, stability, and the life of the electronics equipment and must be monitored for safe operation [10] . For testing purposes, an industrial-grade relative humidity sensor with a tolerance of ±2% tolerance has been used and can be easily replaced with meteorological grade RH sensor such as Vaisala-Hmt310 [19] , having ±1% tolerance if required, as both sensors have standard 4-20mA current outputs. The air-conditioning system will maintain the relative humidity of the SABRE laboratory in the 30%-50% range. When subjected to external vibrations and shocks, PMTs are known to exhibit changes in signal levels and microphonic noise, and excessive vibrations could result in their breakage [7] . As the active mine environment in Stawell will experience vibrations due to heavy machinery operation and explosions for mining excavations, the environmental vibrations must be closely monitored, to minimise the damage to sensitive equipment such as PMTs. As the veto PMTs will be submerged in the LAB, the corresponding response of the PMTs to the changed environmental conditions need to be considered. One challenge that the Thallium doped Sodium Iodide crystal (NaI(Tl)) presents as a dark-matter target is its highly hygroscopic nature. A crystal must be protected from moisture at all times to prevent degradation of its optical properties. The modular detector array has crystals sealed in enclosures, that can be thoroughly outgassed using a continuous flow of high purity dry nitrogen (DN 2 ) to remove trace amounts of water clinging to the surfaces [18] . Monitoring the moisture content of (DN 2 ) at the inlet and outlet of the detector array is critical. The slow control system allows for this operation through integration with industry grade sensors such as Pura Pure Gas Trace Moisture Transmitter [12] or bespoke design built using I 2 C Sensirion SHT21 [17] . The DN 2 gas flow-rate, volatile organic compound (VOC) vapours on N 2 exhaust and in the catchment at the base of SABRE vessel to detect LAB leaks, DN 2 gas-pressure in the LAB and SABRE vessel shielding, level of LAB fluid, are just some of the additional identified parameters which the slow control system could monitor. The prototype slow control system has been designed with a modular architecture to enable scaling by simply daisy-chaining it to other Field- This open-ended real-time slow control system design is capable of industrial-grade performance/reliability and is expected to meet the key current and future emerging needs of the experiments in a subterranean environment. APAR 2018, Humidity transducer EPJ Web of Conferences Proceedings of ICALEPCS Photomultiplier tubes-Basics and Applications Temperature Input Module Extreme-temperature and harsh-environment electronics Pure Gas Trace Moisture Transmitter Barometric Formula The LabVIEW RIO Architecture: A Foundation for London Underground Improves Reliability for 200 Million Annual Passengers with Remote Condition Monitoring Digital Humidity Sensor SHT2x 2009 International Conference on Measuring Technology and Mechatronics Automation C Series Sound and Vibration Input Module Sound and Vibration data acquisition The authors acknowledge the support received from Swinburne University of technology and the University of Melbourne for making available the equipment and resources to carry out the development and testing.