key: cord-0331731-z3x0dfzk authors: Mei, K.; Geagen, M.; Roshkovan, L.; Litt, H. I.; Gang, G. J.; Shapria, N.; Stayman, J. W.; Noel, P. B. title: Three-dimensional printing of patient-specific lung phantoms for CT imaging: emulating lung tissue with accurate attenuation profiles and textures date: 2021-08-01 journal: nan DOI: 10.1101/2021.07.30.21261292 sha: 087c50ad339568bd06333a85a47cf59871b39d11 doc_id: 331731 cord_uid: z3x0dfzk Purpose: Phantoms are a basic tool for assessing and verifying performance in CT research and clinical practice. Patient-based realistic lung phantoms accurately representing textures and densities are essential in developing and evaluating novel CT hardware and software. This study introduces PixelPrint, a 3D printing solution to create patient-based lung phantoms with accurate attenuation profiles and textures. Methods: PixelPrint, a software tool, was developed to convert Patient DICOM images directly into printer instructions (G-code). The density was modeled as the ratio of filament to voxel volume to emulate attenuation profiles for each voxel. A calibration phantom was designed to determine the mapping between filament line width and Hounsfield Units (HU) within the range of human lungs. For evaluation of PixelPrint, a phantom based on a human lung slice was manufactured and scanned with the same CT scanner and protocol used for the patient scan. Density and geometrical accuracy between phantom and patient CT data was evaluated for various anatomical features in the lung. Results: For the calibration phantom, measured mean Hounsfield units show a very high level of linear correlation with respect to the utilized filament line widths, (r > 0.999). Qualitatively, the CT image of the patient-based phantom closely resembles the original CT image both in texture and contrast levels, with clearly visible vascular and parenchymal structures. Regions-of-interest (ROIs) comparing attenuation illustrated differences below 15 HU. Manual size measurements performed by an experienced thoracic radiologist reveal a high degree of geometrical correlation of details between identical patient and phantom features, with differences smaller than the intrinsic spatial resolution of the scans. Conclusion: The present study demonstrates the feasibility of 3D printed patient-based lung phantoms with accurate organ geometry, image texture, and attenuation profiles. PixelPrint will enable applications in the research and development of CT technology, including further development in radiomics. densities are essential in developing and evaluating novel CT hardware and software. This study 31 introduces PixelPrint, a 3D printing solution to create patient-based lung phantoms with accurate 32 attenuation profiles and textures. Anthropomorphic phantoms, geometric image quality phantoms, and mathematical phantoms are 55 fundamental tools for developing, optimizing, and evaluating novel methods in computed 56 tomography (CT) research and clinical practice. Common CT phantoms are typically 57 manufactured by casting, forming, and molding homogenous materials such as resin or plastic. 58 While many different phantoms are available commercially and in research laboratories, there is 59 a lack of patient-based phantoms that fully represent attenuation profiles and textures seen in 60 clinical CT acquisitions, for example, for healthy and diseased lungs. Additionally, the academic 61 and clinical CT community would benefit from a rapid and inexpensive manufacturing process 62 compared to current commercial solutions. 63 Over the last decade, fused deposition modeling (FDM)-based 3-dimensional (3D) printing of 64 various tissue-mimicking phantoms has been widely explored for validation and evaluation of CT 65 imaging technology 1-6 . Studies have focused on several areas, including manufacturing 66 geometrically correct models of organs 7-11 , generating realistic texture samples 12-14 , and creating 67 accurate attenuation profiles 15-18 . The general procedure to 3D print an anthropomorphic phantom 68 from CT image data includes: (i) segmentation of regions/organs of interest in CT images, (ii) 69 conversion of selected regions from volumetric data to triangulated surface geometry models (e.g., 70 STL or SLA files), and (iii) use of printer-specific slicing software to apply proper parameters (e.g., 71 extrusion rate, print speed, infill ratios, etc.) and generate instructions (G-code) for printers to 72 create 3D products. While this approach produces phantoms that better resemble true anatomical 73 structures, it still has shortcomings. First, spatial resolution is largely lost due to segmentation of 74 regions and conversion to surface models. Second, for each region/surface model, the slicer 75 software assigns unique infill and exterior walls (or perimeter), creating abrupt, unrealistic 76 transitions between regions of different densities in the final product. Third, due to its reliance on 77 segmentation, this method is susceptible to boundary placement errors. 78 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 1, 2021. ; https://doi.org/10.1101/2021.07.30.21261292 doi: medRxiv preprint different densities in 3D printed CT phantoms, these methods utilize a pixel-by-pixel change in 80 the filament extrusion rate, while maintaining a constant printing speed. Although this approach 81 enables generation of sophisticated phantoms with realistic attenuation profiles, it falls short when 82 printing high-resolution features. This reduction in spatial resolution is an important concern when 83 generating natural image textures. Therefore, there is an unmet need for a quick and cost-efficient 84 process for generating patient-based phantoms with accurate organ geometry, image texture, 85 and attenuation profiles. 86 We propose a 3D printing solution that is capable of achieving accurate organ geometry, image 87 texture, and attenuation profiles while eliminating the complexities and limitations of previous 88 methods. Our solution is a one-step method for translating CT images into printer instructions (G-89 code) that can be used by any FDM 3D printer. It combines varying printer speeds with a constant 90 filament extrusion rate to control the density of each printed voxel. In the following sections we 91 present a complete description of the proposed method, as well as results from successful proof-92 of-principle experiments with geometrical and patient-based lung phantoms. 93 Conventional 3D printing utilizes slicing software to convert 3D models (e.g., STL files) to printer 96 instructions written in G-code, a widely used machine language defining 3D printing parameters 97 (e.g., layer height, retraction, print speed, etc.). We present a solution that accepts volumetric CT 98 DICOM data as input and converts these data directly into G-code without segmentation or 99 intermediate 3D models. 100 Applied to common FDM 3D printers, PixelPrint produces multiple 2D layers, one layer at a time, 101 to create 3D objects or phantoms. Each printed layer is mapped from the corresponding DICOM 102 slice, with the physical scale controlled to ensure that the resulting phantom has the same dimensions as the scanned patient. For each layer in the 3D printed phantom, PixelPrint 104 generates an array of spaced, parallel filament lines. The lines are at fixed spacing but are of 105 varying widths, creating a partial volume effect to form varying densities in the CT scan, i.e., wider 106 line widths for high-density areas and narrower line widths for low-density areas. PixelPrint 107 computes the density of the input image at closely spaced intervals along each line and maps it 108 into appropriate extrusion and printhead speeds over each interval. It then records one G-code 109 command that defines a starting point, an end point, the filament extrusion, and speed for that 110 interval. This process is repeated for every interval over every line, in every layer, until the whole 111 phantom is fully encoded in a G-code file. Since layers are deposited in alternating directions, the 112 varying line widths create a matrix of high-and low-density regions that correlate with the original 113 3D input image volume. The matrices are shifted in angle and location at each layer, and the 114 printed layer height is much smaller than the typical CT slice thickness. Each CT slice of the 115 printed phantom will therefore contain multiple shifted layers, ensuring that reslicing of the CT 116 image data will not result in sampling or moiré patterns. 117 In our experiments, we found that altering the line width by varying the extrusion rate alone does 118 not provide sufficient spatial resolution due to the inherently slow response time of the extrusion 119 process. Instead, we maintain a constant filament flow rate while changing the speed of the 120 printhead to control the extrusion width. 121 Calibration Phantom. The partial volume effect created by varying the filament line width with fixed 123 line spacing determines the (local) mean material density and thus the x-ray attenuation in a CT 124 scan. A calibration phantom was designed to determine the mapping between filament line widths 125 and Hounsfield Units (HU) within the range of human lungs. A multi-sector phantom (cylinder 126 divided into radial slices) with a diameter of 100 mm and a height of 10 mm was printed using 127 PixelPrint. The phantom comprises ten sections with different material densities, from 10% to 100% 128 at 10% intervals ( Figure 1 ). CT measurements of this phantom were used to determine the 3D 129 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 1, 2021. Table 1 presents acquisition and 142 reconstruction parameters utilized for imaging. Since this proof-of-principle study focuses on lung 143 imaging, the right lung of the patient was selected as input to reduce printing time and complexity. 144 A 20 mm diameter ring surrounding the lung was added for better positioning of the phantom 145 within the bore of a 300 x 400 mm 2 oval phantom representing a medium-sized patient (see details 146 below). HU values were converted into filament line widths using the mapping calculated from the 147 calibration phantom described above. A lower cut-off value of 10% and an upper cap of 100% 148 material density were applied. The phantom is shown in Figure 3 . 149 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. For imaging, printed phantoms were placed inside the 20 cm bore of a technical phantom 160 (Gammex multi-energy CT phantom, Sun Nuclear Corporation, Melbourne, FL, USA) to mimic 161 attenuation profiles of an average-sized patient (300 x 400 mm 2 ). Imaging was performed with 162 the same CT scanner using the same protocol as the clinical acquisition (Table 1) . 163 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 1, 2021. Printing of the multi-sector calibration phantom required 16 hours of printing time. The CT image 180 of the phantom shows homogeneous appearances within the ten different material density regions. 181 Pearson's correlation coefficient r > 0.999 (see Figure 2 ). The maximum HU (corresponding to 184 100% material density) was measured as 115 and the minimum HU (corresponding to 10% 185 material density) was measured as -867. 186 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 1, 2021. structures. Figure 4 shows the identical regions in patient and phantom data selected for density 197 measurements. Although the patient image appears noisier than the phantom image, due to 198 higher attenuation from the patient body, five ROIs show very similar mean values, with 199 differences less than 15 HU (see Table 2 ). In general, lower density areas have a slightly higher 200 difference due to the 10% material density cut-off used with PixelPrint. 201 Figure 5 presents the three manually measured anatomical features. Manual size measurements 202 performed by the radiologist illustrate a high degree of geometrical correlation of details between 203 the patient image and the phantom images, with differences smaller than the intrinsic spatial 204 resolution of the scans (see Table 3 ). 205 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 1, 2021. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 1, 2021. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 1, 2021. The present study has some limitations. Only one patient-based phantom was evaluated for this 264 proof-of-concept study. Follow-up studies will provide additional data and measurement to 265 describe specific lung diseases such as COVID-19 pneumonia. For our initial study, we have 266 focused on generated realistic models of the human lung. Future studies will be essential to add 267 the capability to print soft tissue and bone to cover a broader range of anatomical regions and 268 tissue x-ray attenuation. 269 Finally, we would like to allow the larger medical, academic, and industrial CT community to have 270 access to PixelPrint. We can make copies of our phantoms available as well as customized 271 phantoms based on specific CT images, which could include various lung diseases. For additional 272 information, please see the project homepage: www.pennmedicine.org/CTResearch/PixelPrint. 273 In conclusion, the present study is the first to illustrate the possibility of creating 3D printed patient-275 based lung phantoms with accurate organ geometry, image texture, and attenuation profiles. This 276 may lead to a paradigm change for the development of novel CT hardware and software by 277 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 1, 2021. ; https://doi.org/10.1101/2021.07.30.21261292 doi: medRxiv preprint enabling accelerated evaluation and validation with realistic patient-based data. Ultimately this 278 will shape the clinical day-to-day routine and benefit patients with novel and standardized CT 279 Acknowledgement 281 We acknowledge support through the National Institutes of Health (R01CA249538). 282 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 1, 2021. ; https://doi.org/10.1101/2021.07.30.21261292 doi: medRxiv preprint RadioGraphics Update: Medical 3D doi:10.1148/rg.2020190217 288 2. Filippou V, Tsoumpas C. Recent advances on the development of phantoms using 3D 289 printing for imaging 3D printing based on imaging 292 data: Review of medical applications An overview on 3D printing technology A systematic review on 3D-Printed imaging 298 and dosimetry phantoms in radiation therapy Anatomic modeling using 3D printing: quality assurance 301 and optimization. 3D Print Med Experimental feasibility of spectral photon-303 counting computed tomography with two contrast agents for the detection of endoleaks 304 following endovascular aortic repair Evaluation of a preclinical photon-counting CT 307 prototype for pulmonary imaging Spectral Photon-counting cT: Initial Experience 310 with Dual-Contrast Agent K-Edge Colonography Development of a 3D printed anthropomorphic lung phantom for image quality 314 assessment in CT Development of an organ-specific insert protocols 3D-Printed Textured Phantoms for 320 Assessment of High Resolution CT Performance assessment of texture 322 reproduction in high-resolution CT Comparison of low-contrast detectability between 326 two CT reconstruction algorithms using voxel-based 3D printed textured phantoms 3D printing of 329 anatomically realistic phantoms with detection tasks to assess the diagnostic 330 performance of CT images Using inkjet 3D printing to create contrast-enhanced 333 textured physical phantoms for CT A novel 3D printing method for accurate anatomy replication in patient-336 specific phantoms The OpenCV Library. Dr Dobb's J Softw Tools Joemai RMS, Geleijns J. Assessment of structural similarity in CT using filtered 344 backprojection and iterative reconstruction: A phantom study with 3D printed lung 345 vessels Radiopaque three-dimensional printing: A method 347 to create realistic CT phantoms Paper-based 3D 350 printing of anthropomorphic CT phantoms: Feasibility of two construction techniques. Eur 24. Willemink MJ, Noël PB. The evolution of image reconstruction for CT-from filtered back 353 projection to artificial intelligence Comprehensive Investigation on Controlling for CT Measuring computed tomography scanner variability of 359 radiomics features