key: cord-0331683-gha4rgh9 authors: Grasdal, M.; Walter, M.; Krassioukov, A. V. title: The microbiological and physical properties of catheters for intermittent catheterization: A systematic review on the impact of reuse and cleaning methods date: 2021-05-02 journal: nan DOI: 10.1101/2021.04.28.21256230 sha: 2e942699871ce490f478ebc94743a0c2eeed518f doc_id: 331683 cord_uid: gha4rgh9 This systematic review provides an up to date and comprehensive summary of the clinical evidence of the effectiveness of various cleaning methods of intermittent catheterization that have been proposed to prepare catheters for reuse. This systematic review is registered at PROSPERO (registration number: CRD42020176065). A key word search of Medline (OVID), Excerpta Medica dataBASE (EMBASE, OVID), Cumulative Index to Nursing and Allied Health Literature (CINAHL), Web of Science and the Cochrane Central Register of Controlled Trials (CENTRAL), in addition to manual searches of retrieved articles, was undertaken to identify all English, Russian and German language literature evaluating the effectiveness of various cleaning methods of intermittent catheterization. Studies selected for review included analytical experimental, prospective cohort, cross-sectional and case series study designs. Prospective cleaning methods analyzed included heat-based sterilization, chemical cleaning solutions, mechanical abrasion, photocatalytic sterilization, and combined methods. Studies that failed to assess the bacterial colonization or physical properties of catheters following cleaning were excluded. In total, 12 studies (i.e. 9 analytical experimental, 1 cohort study, 1 cross-sectional and 1 case series) were included. Two cleaning methods were identified as likely being most promising: five-minute submersion in 70% alcohol and the 'Milton method'. Each eliminated bacterial colonization without affecting the physical properties of the catheters. All other cleaning methods were either non-bactericidal or caused gross visual or microscopic damage to the catheters, rendering their reuse unsafe. Additional higher-powered studies confirming the safety and efficacy of these cleaning methods must be obtained before we would feel comfortable challenging current clinical recommendations. Intermittent catheterization (IC) is a method of urinary drainage commonly used by individuals with neurogenic lower urinary tract dysfunction (NLUTD) 1 , often as a result of a spinal cord injury (SCI). It is performed by inserting a catheter into the bladder via the urethra, allowing for the drainage of urine. Once the bladder is empty, the catheter is removed and discarded. IC was traditionally performed using an aseptic technique 2 , however in 1972, Lapides et al. first demonstrated that using a clean technique, which they coined clean IC (CIC), was equally as effective at preventing urinary tract infections (UTIs). 3 Thus, CIC is considered the "gold standard" for bladder emptying in individuals with NLUTD and sufficient manual dexterity. [4] [5] [6] Presently, apart from only one manufacturer 7 that we are aware of, the majority of manufactured catheters are clearly labeled as single-use devices. As such, these catheters are distributed without instructions for cleaning, although it is well known that due to their cost and limited availability in many countries, they are commonly reused. 7 Catheter reuse is especially common in developing countries where it is not unheard of for some individuals to use the same catheter for greater than a year. 7 Some consumers have stated that they would value having the option to reuse in concert with single-use if there was a safe and practical method of cleaning them following IC. 8 In 2014 Prieto et al. suggested that the reuse of catheters was as safe as the single use of catheters 9 , although this work has since been retracted after an additional independent review of the data found crucial discrepancies of data extraction and analyses within the review. 10 At the time of writing, the Center for Disease Control and Prevention (CDC) clinical recommendations 11 state that further work must be done in this area before the reuse of catheters can be safely recommended. In Canada, the Canadian Urological Association (CUA) 12 recommends that catheters should only be reused in specific clinical scenarios, and that users should be made aware that there is limited evidence to support the cleaning and safe reuse of single-use catheters. There are many advantages to reusing catheters, namely convenience, reduced economic burden for consumers, and less environmental impact due to reduced waste. 13,14 These advantages are offset by the burden of cleaning the catheters, cost of cleaning supplies, the concern that reusing catheters may cause the rate of UTIs to increase, and the added cost associated with a potential rise in lower urinary tract (LUT) complications. 8, 13, 15, 16 There is recent evidence to suggest that structural damage of the catheter itself from cleaning could put users at risk of contracting UTIs by their reuse due to an increased risk of urethral and bladder trauma. 17 Therefore, it is important to understand how proposed methods of cleaning catheters for their reuse impacts the catheter's structural integrity. All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted May 2, 2021. ; https://doi.org/10.1101/2021.04.28.21256230 doi: medRxiv preprint A variety of catheters have been marketed for IC over the years. Historically, the most commonly used catheters were constructed of materials including rubber, latex, silicone, polyvinyl chloride (PVC) and polyethylene. 18 Today, PVC catheters are the most commonly used 19 , with some having a hydrophilic outer coating for ease of use. Still, because of the continued variability in materials used, cleaning methods that are effective for some catheters may not be suitable for others. To date, there is no reported consensus on the preferred method of cleaning, nor the period of time that a single catheter may be reused. 20 Obtaining consensus is further complicated by the fact there is high variability and poor compliance when performing cleaning techniques, leading to an increased risk of bacterial colonization 20, 21 and LUT symptoms. Consequently, it is of great interest to determine if there are effective catheter cleaning methods to help ensure patient safety for those who choose to reuse them, and to potentially reduce the environmental impact and financial burden of single-use catheters onto the healthcare system. To that end, this systematic review provides a comprehensive analysis of the effectiveness of various proposed methods of cleaning of catheters for IC to prepare them for their reuse. We were specifically interested in comparing the effectiveness of cleaning and reuse methods by evaluating the outcomes of bacterial colonization and change in physical properties. First, we examined the effectiveness of cleaning methods in terms of their ability to eliminate or diminish bacterial colonization; and second, we examine the impact of cleaning techniques on the physical properties of catheters. All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines were used to report our systematic review. 22 A PROSPERO review protocol exists and can be accessed (registration number CRD42020176065). We included all published studies thus far which investigate the properties of intermittent catheters and how they are affected by various cleaning procedures during re-use. We were specifically interested in studies focusing on either the relationship between cleaning during reuse and bacterial colonization of an intermittent catheter or the effects of cleaning and reuse on the physical properties of an intermittent catheter. population key words were: "intermittent catheter", "bladder catheter" and "urinary catheter". These were paired with the terms: "clean", "cleaning methods", "cleaning solution", "disinfect", "decontaminate", "sterilize" OR "physical", "microscopic", "mechanical" OR "reuse". See Supplementary Appendix SA1 for the full search strategies used for each database. Studies were included for qualitative analysis if they met the following criteria: published in English, German or Russian language that evaluated the impact that various methods of cleaning catheters have on catheter bacterial colonization and changes in catheter physical properties. We included all study designs except reviews, book chapters, opinion papers, non-peer-reviewed work, conference abstracts or papers and studies where the full text was unavailable. Studies were collated and uploaded into Covidence. 23 Independent reviewers (author 1 and 3) screened titles, abstracts, and full-texts; only eligible studies were included in the qualitative analysis. A third reviewer (author 2) resolved discrepancies. Figure 1 illustrates the PRISMA flow diagram. A consensus was achieved (between authors 1-3) on data to be The literature search yielded 9613 articles. After eliminating duplicates and reviewing the remaining titles and abstracts (Figure 1 ), a total of 12 studies 7, 16, 19, 21, [28] [29] [30] [31] [32] [33] [34] [35] that included outcomes of catheter bacterial colonization or changes in physical properties following cleaning of catheters for reuse were eligible and included ( Table 1) . Description of the studies: Of the 12 studies included in this systematic review, nine were analytical experimental studies 19,28-35 , one of which was also partially a cohort study. 30 The remaining were either a cohort 16 or cross-sectional 21 study and a case series. 7 16 or combination of two methods 19, 35 ). The majority of these studies (n = 10) 16, 19, [28] [29] [30] [31] [32] [33] [34] [35] used culture analysis as a microbiological technique to assess for the presence of bacteria. Several studies (n = 7) 7, 16, 19, 21, 29, 31, 32 examined the impact of cleaning methods on physical properties of catheters, however only a minority (n = 4) 7,16,19,31 examined the microscopic physical properties of the catheter. All 12 studies were grouped based on the type of cleaning methods analyzed, which included the following categories: heat-based, mechanical-based, chemical, combined and others. Next, the effectiveness of cleaning based on levels of bacterial colonization after using specific cleaning methods was determined (Tables 2-6 ). In the majority of studies, the bacterial colonization was either measured in terms of colony forming units (cfu) or log reduction depending on the quantification method reported. Changes in physical properties of catheters following various cleaning methods was then determined (Tables 2-6 ). These changes were evaluated macroscopically (i.e., gross visible inspection of changes and change in stiffness) and/or microscopically (by some form of Electron Microscopy (EM) or Differential Scanning Calorimetry (DSC)). Overall, quality was fair for all included studies. See Supplementary Appendix SA2 for full details of quality assessment. The risk of bias was assessed in accordance with ROBINS-I for all eligible studies, i.e. the two cohort studies (see Figure 2 ). Risk-of-bias assessment showed either serious or critical risk of bias. All other studies did not qualify for risk of bias analysis since only catheters and not patients were reported. All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. Heat-based methods: Among heat-based cleaning methods the microwave treatment (n = 6) 19,21,29,31-33 was the most commonly used followed by boiling (n = 1) 19 and steam sterilization (n = 1). 19 Overall, it was generally agreed that upon increasing the length of microwaving time resulted in a reduction in colony counts. Rubber catheters were reportedly able to be sterilized by microwave treatment 29,32,33 , but none of these studies fully analyzed the impact on the structural integrity of the catheters. When analyzing non-rubber catheters, none of the microwave treatments studied were as effective at reducing bacterial colony counts of Escherichia (E.) coli, Pseudomonas (P.) aeruginosa and Staphylococcus (S.) aureus, while also maintaining catheter structural integrity as were the chemical and combined methods (e.g. alcohol, "Milton method") 19, 31 . In two studies 19,32 , investigators were not able to obtain bacterial colonization data as the physical properties of the catheters changed to dramatically for the study to continue. Boiling in tap water for two minutes resulted in undetectable levels of E. coli but also led to the PVC catheter being damaged. Other uropathogens were not tested in this study due to evidence of catheter damage. 19 Steam cleaning also resulted in undetectable E. coli or Klebsiella (K.) pneumoniae. However, there was evidence of some viable but non culturable bacterial population (VNCB). 19 Mechanical-based methods: The use of detergent and water was the most common mechanical cleaning method (n = 4) 19,28,32,35 , followed by ultrasonic jewelry cleaner (n = 1). 19 Although these methods have demonstrated some reduction in bacterial counts, neither of these methods were effective for sterilisation. For example, Chan et al. showed that 44% of PVC catheters still had a positive culture following soaking in a 1.5% solution of antibacterial soap for 5 minutes followed by a 1-minute water wash. 35 Wilks et al. reported that the use of a domestic jewelry cleaner was also not an effective cleaning method as the action of the cleaner was not bactericidal and damaged the catheter surface. 19 Chemical methods. There were a variety of cleaning solutions examined (n = 9). 7, 19, 28, 31, 34 Submerging PVC catheters in a 70% alcohol solution for 5 minutes (n = 1) 31 was the most effective as there was both bacterial sterilization (6-log colony count reduction) and showed no catheter damage. Another study determined that soaking plastic catheters for 5 minutes in 3% peroxide solution (n = 1) 28 was not fully bactericidal but reduced the bacterial colony count by nearly 3 log. Similarly, it was found that soaking non-latex catheters in 0.6% hydrogen peroxide (n = 1), 1:4 solution of bleach in tap water (n = 1), or a 1:2 solution of betadine in tap water (n = 1) for 30 minutes were all effective at preventing E.coli growth 48 hours post All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted May 2, 2021. ; https://doi.org/10.1101/2021.04.28.21256230 doi: medRxiv preprint cleaning, but there was no analysis for structural damage. 34 Soaking in vinegar (n = 2) 19, 28 was not effective at eliminating most uropathogens, although it did not affect the physical properties of the catheters. Additionally, it was found that the use of Milton solution (diluted Milton concentrate with tap water as per manufacturer's instructions, 0.6% sodium hypochlorite final concentration) (n = 1) 19 resulted in no detectable culturable bacteria after repeated contamination and decontamination events over 24 hours when tested with a full range of uropathogens; however, there was evidence of some VNCB population remaining. A cross-sectional study by Kovindha et al. followed participants who kept their reusable silicone catheters in a Savlon solution between uses over the course of 1-2 years. These reused catheters showed encrustation of the lumen and an increase in stiffness of 20% relative to a new catheter. 7 Finally, it should be noted that in general cleaning was not as effective without rinsing and drying the catheter prior to submersion in the cleaning solution. 28 Combined, mixed and other methods: There were two combined methods reported. The first combined an antibacterial soap wash for 5 minutes followed by microwaving in an 800W microwave for 5 minutes (n = 1). 35 This was found to be more effective than washing with antibacterial soap alone, but still found that 24% of catheters had a positive E. coli culture. The other, coined the "Milton method" by Wilks et al, was found to be more effective and was performed by soaking a catheter in hot soapy water for 5 minutes, rinsed with water, left to soak in Milton solution for 15 minutes and then rinsed off with tap water again (n = 1). 19 This resulted in undetectable E. coli levels, no evidence of a VNCB population and showed no structure damage to the uncoated PVC catheters. Sekiguchi et al. have developed a titanium dioxide (TiO 2 ) coated silicone catheter which has a photocatalytic antimicrobial effect using only UVA illumination. 30 After 60 minutes of black light illumination using a 15W 352nm black lamp with intensity of 1000uW/cm 2 , it was demonstrated that the photocatalytic effect of the TiO 2 catheters reduced the survival of all bacteria tested (Serratia marcescens, methicillin-sensitive S. aureus, methicillin-resistant S. aureus, P. aeruginosa) to a negligible level. These catheters reportedly maintained their antibacterial effect after 200 uses. There was no data collected about how the sterilization affected the structure of the catheters. Another recent study by Newman et al. did not control for cleaning methods (participants choice and a mix of methods were reportedly used) but found that after a mean reuse of 21 days, 74% of catheters had microbial contamination and 20 different bacterial species were found in total. A biofilm was found in 20% of the reused catheters and viable microorganisms could be cultured in 67% of the samples. There was debris contamination found on all catheters All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted May 2, 2021. ; which EM showed to be a mixture of residuals from urine, lubricant, soap, detergents and cells. 16 Controls Unused catheters were found to have no microbial contamination (n=2) 16, 35 , except in one study where growth of contaminants was found due to the packaging materials of the catheters. 35 Likewise unused catheters had no surface damage as determined by EM (n=2). 7, 19 Catheters that were inoculated and not cleaned, or inoculated and rinsed with tap water only, had consistently higher bacterial colonization than catheters that had been cleaned by heat, mechanical, chemical, combined, or other methods (n=7). 19 This systematic review provides an up to date and comprehensive summary of the clinical evidence of the effectiveness of various cleaning methods of IC that have been proposed to prepare catheters for reuse. This analysis will help guide future clinical recommendations and is applicable to both health care providers and individuals requiring the use of catheters for IC. The available evidence suggests that the effectiveness of microwave sterilization greatly depends on the material of the catheter. Rubber catheters seem to be able to withstand repeated microwave sterilizations with a common household microwave without incurring macroscopic changes. 29 ,32,33 It should be noted that to date, no assessment of microscopic structural damage has been published after microwaving rubber catheters. Reported time to sterilize seems to be a minimum of 5 minutes and depends on the wattage of the microwave used, however there was an inverse relationship between the duration of microwave sterilization and the amount of viable bacterial colonies present on the catheter. 29 Polyethylene catheters were shown to also be completely sterilized by a microwave sterilization method; however, this material of catheter was only evaluated in one study and no data regarding the physical properties of the catheter was present. 29 Additionally, it has been determined that plastic catheters melted in the microwave. 28 Microwave sterilization is also not a recommend cleaning method for other catheter material such as latex or PVC as there is evidence of melting and incomplete sterilization. 19, 28 Although gross changes in physical properties of catheters were documented in a few studies, less is known regarding the microscopic changes and changes in stiffness of catheters. It is crucial to acknowledge that based on expert consensus opinion, the physical properties of catheters are a crucial risk factor for ureteral micro trauma and development of UTIs. 17 Other heat-based methods (steaming and boiling) were not effective and damaged PVC catheters. 19 The quality of evidence for the efficacy of chemical cleaning is limited because the effectiveness of most have not been duplicated by more than one study, and many do not have corresponding data regarding their effects on catheter physical properties. To date, it seems that the most promising chemical cleaning solutions are 70% alcohol and Milton solution. It was demonstrated by one study that soaking a PVC catheter in the 70% alcohol solution for five minutes reduced bacterial colony counts by 6 log and did not affect the physical properties of PVC catheters. Soaking for longer period in alcohol solution is not recommended, as damage was seen in the catheters soaked for 45 minutes. 31 Milton solution also did not affect the properties of PVC catheters, but did show some evidence of VNCB present. 19 Other cleaning solutions such as 0.6% peroxide, a 1:4 solution of bleach with tap-water or a 1:2 solution of betadine in tap water were shown to be bactericidal All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted May 2, 2021. ; https://doi.org/10.1101/2021.04.28.21256230 doi: medRxiv preprint against for E. coli but no structural analysis was performed. 34 Another study supported the suggestion that peroxide could be a potential cleaning solution as it reported that a 3% peroxide solution reduced the bacterial colony count by 3 log; however, the same study also noted that it was less effective than rinsing with water and drying alone, which reduced the bacteria colony count by 5 log. 28 Vinegar, while it appears to not affect the physical properties of the catheters, does not seem to be a suitable bactericidal cleaning solution. 19, 28 Kovindha et al. examined the effects of long term reuse with a Savlon cleaning solution as a case series by imaging catheters with EM. 7 They found that there was encrustation on the lumen, and that the catheters had a 20% increase in stiffness. It should be noted that the two non-control catheters used in this case series were used for 1.5 and 2 years. Given the extent of time of reuse, one cannot make a conclusion on the short-term burden of reuse form this study. Future studies, such as randomized-controlled that incorporate larger sample sizes and long-term follow-up, should be conducted before these chemical cleaning solutions can be recommended for the reuse of catheters. Mechanical cleaning methods such as washing with detergent or using a domestic jewelry cleaner left much to be desired when used on their own. When washing with detergent, one study reported that 44% of catheters still resulted in a positive culture 35 , while another showed that there was still 100,000 bacteria/mL present. 32 A combined approach of washing with detergent and microwaving was better, but still resulted in 22% of catheters having a positive culture swab, and it was limited by the fact that microwaving greatly damages some catheter materials. A third study reported that just rinsing with tap water and drying alone resulted in a greater reduction in bacteria on catheters than washing with detergent alone. 28 Just soaking in detergent also lowered counts of E. coli and K. pneumonia and did not damage PVC catheters, although it was reported to leave behind some residue in the lumen of the catheters. 16 Soaking in detergent followed by soaking in Milton solution ("Milton method") was reported as totally bactericidal and left no evidence of damage on PVC catheters. 19 The other mechanical cleaning method analyzed, domestic jewelry cleaner, was not an effective as it was not bactericidal and caused surface damage to the catheters. 19 Photocatalytic sterilization is an interesting proposal as it was reportedly bactericidal for all bacteria tested but no structural analysis was performed. 30 A major limitation to this method is that a special TiO 2 coated silicone catheter and ideally a black lamp is required, although it was reported that sunlight could be used as an alternative to the black lamp. In addition, we had to grade the study done by Sekiguchi et al. in terms of risk of bias as critical because there was missing data. 15 TiO 2 catheters were analysed but only 10 control catheters analysed. There should have been the same number of control catheters as TiO 2 catheters analyzed as there should be one of each per participant that completed the study hence there was missing data and therefore a risk of bias. All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. 16 41% of participants stated they just rinsed with water and 41% reported they use both soap and water to clean their catheters before reuse. These catheters were collected, and it was found that 74% of them had microbial contamination and 20% had biofilm formation. Debris from residuals of urine, soap, detergent and cells were seen on EM. It should be noted that the mean number of days these catheters were reused prior to analysis was 21 days, but reuse ranged from one day to 270 days. Because of this wide range, there was some bias of generalizing the results of the impact of bacterial load. A major limitation of this systematic review was that every experimental study used a different approach for analyzing the effectiveness of cleaning methods. For example, the catheter material, method of cleaning, indicator organisms, and outcome measures were not constant among studies, making quality assessment and meaningful comparisons a challenge. The non-experimental studies also did not standardize the type of catheter or cleaning method used. Many of the studies also only performed the cleaning method once before analysis, which generated outcomes that were not practical for real-life application 36 as those who reuse catheters often do so more than once. Many studies also did not completely analyze the impact of cleaning on the physical properties of the catheters, which also decreased the practicality of their outcomes. The amount of published data on the topic is also a limitation on our results. While there were 12 studies that met our inclusion criteria, many of them were examining different cleaning methods on different catheter types, so in many cases cleaning methods were only examined in one study. For example, the two most promising cleaning methods reported, 70% alcohol immersion and "Milton method" were only examined in one study which decreased our confidence. As a result, we were not able to perform any meaningful metaanalysis. All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. There were two cleaning methods reported that reduced all of the bacteria tested to negligible levels while confirming that the structural integrity of the catheter was maintained: soaking catheters in 70% ethanol solution for 5 minutes, and the combined approach of soaking catheters for 5 minutes in hot detergent water followed by soaking for 15 minutes in Milton solution. The structural integrity of the two cleaning methods were confirmed by calorimetric analysis and EM respectively. Major limitations to this result are that these methods of cleaning have not been confirmed effective by more than one study, as only PVC catheters were examined, and outcomes were measured after only a single reuse. While these cleaning methods are promising, additional data must be obtained before we would feel comfortable challenging current clinical recommendations. The authors believe that this topic requires further consideration and additional research that could provide better insight on cleaning methods, reuse of catheters, and potential benefits versus harm to individuals who use of catheters for IC. All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted May 2, 2021. ; https://doi.org/10.1101/2021.04.28.21256230 doi: medRxiv preprint Acknowledgments Matthias Walter and Andrei V. Krassioukov had full access to all the data in the study and takes responsibility for the integrity of the data and the accuracy of the data analysis. No competing financial interests exist. All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted May 2, 2021. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The risk of bias analysis only addresses the cohort part of this publication. All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted May 2, 2021. ; 2 0 Ultrasonic cleaning using a domestic jewellery cleaner followed by water rinsing; Detergent and water (mixed and left to soak for 5 minutes followed by water rinsing). White vinegar (50% solution, 30-minute soak followed by water rinsing); Milton solution (15-minute followed by water rinsing). Bacterial culture analysis (colony forming units). EM for evidence of catheter surface damage. Gross visual inspection. at 37 degrees Celsius for 5 minutes. Detergent and water (mix and 5-minute soak) followed by Milton solution (15 minute soak followed by water rinsing), also known as the Milton method. Rinsed with cold tap water for bacterial culture analysis. Not specified. Aliquots of a 24-hour broth culture of Escherichia coli were spread over the lumen and surface of the catheters. Antibacterial soap wash (5-minute soak in 1.5% antibacterial soap for 5 minutes followed by 1 minute of water rinsing). Combination of antibacterial soap (5-minute soak in 1.5% antibacterial soap for 5 minutes followed by 1 minute rinse of water rinsing) followed by microwave (5 minutes, 800W with a 8oz water heat sink). Handling controls were uninoculated and not cleaned. Wash controls were uninoculated and cleaned by washing. Wash + microwave controls were uninoculated and cleaning by washing + microwave. Positive controls were inoculated and not cleaned. Catheter type: Silicone. Human use. One catheter was reused for period of 1.5 years and another for a period of 2 years. 1:100 Savlon solution (chlorhexidine 1.5% and cetrimide 15%). One new unused catheter was used as comparator to reused catheters for EM. Vinyl (48%), rubber (24%), polyvinyl chloride (18%) and other (10%). These percentages refer to all 129 participants that returned the survey. Human use. Microwave (Variable length but 73% reported 6 minutes. 39% reported using 500 to 1,000 W microwave, 37% reported using greater than 1,000 W microwave and remaining 24% did not know wattage. All reported using a heat sink between 125mL to 1L water). This refers to the 40 participants from the follow-up survey. None. water heat sink). Detergent and water (Washed for 15 seconds and then rinsed off with water). None. analysis (bacterial colony count). Analytical experimental; n =312 catheters. Plastic. Catheter tips were submerged in All catheters were rinsed and dried before cleaning. 3% hydrogen peroxide (5-minute soak); Vinegar (5-minute soak). Bacteria culture analysis (bacterial colony count). n/a All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted May 2, 2021. contaminated urine (10 7 cells/mL Pseudomonas aeruginosa and 10 7 cells/mL Escherichia coli) for 5 minutes. Detergent (10 second wash with detergent followed by 10 second rinse with tap water). Washing with water (10 seconds) and rinsing (10 seconds). Kurtz et al. 1995 32 ; Analytical experimental; n = 48 catheters. Non-latex catheters. Escherichia coli isolates at concentrations ranging from 4.8x10 5 -1.0x10 8 cfu at 37 degrees Celsius for 2 hours. Catheters were rinsed for 1 minute with tap water and then soaked in one of three cleaning solutions for 30 minutes: 0.6% Hydrogen peroxide; Bleach in a 1:4 solution with tap-water; Betadine in a 1:2 solution with tap water. One catheter tap water rinsed. Catheter type: Polyethylene. Catheters were immersed in a solution of Proteus mirabilis, bacteria (10 6 -10 7 cfu/mL) for 30 seconds. Additionally, the bacterial solution was drawn through the lumen of the catheter with a sterile syringe. Microwave (Either 2, 4, 6, or 8 minutes (each six catheters), 650W microwave with a 300mL water heat sink). Six non-microwaved catheters (0 minutes). Bacterial culture analysis (colony forming unit). Gross visual inspection. Douglas et al. 1990 34 ; Analytical experimental; n = 6 catheters. Rubber. Catheters were incubated in the solution of microorganisms (Escherichia coli, Klebsiella pneumoniae, Proteus mirabilis, Enterobacter cloacae, Pseudomonas aeruginosa, Streptococcus pneumoniae, Staphylococcus aureus, and Candida albicans) of a concentration of 10 4 -10 6 cells /mL for 60 minutes at a temperature of 37 degrees Celsius. Microwave (12 minutes,10650W microwave with a 250mL heat sink). One non-microwaved catheter. Bacterial culture analysis (number of viable microorganism s). n/a Abbreviations: EM = electron microscopy; PVC = polyvinyl chloride; TiO 2 = titanium dioxide; VNBC = viable but non-culturable. All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. Abbreviations: EM = electron microscopy; PVC = polyvinyl chloride; VNBC = viable but non-culturable; TiO 2 = titanium dioxide All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. Clean intermittent catheterisation performing abilities, aversive experiences and distress The value of intermittent catheterisation in the early management of traumatic paraplegia and tetraplegia Clean, intermittent self-catheterization in the treatment of urinary tract disease Best practices for the treatment and prevention of urinary tract infection in the spinal cord injured population: The Alberta context Bladder management in the context of a spinal cord injury Summary of European Association of Urology (EAU) Guidelines on Neuro-Urology Clean intermittent catheterization and urinary tract infection: Review and guide for future research Reused silicone catheter for clean intermittent catheterization (CIC): Is it safe for spinal cordinjured (SCI) men? Spinal Cord Reuse of intermittent catheters: A qualitative study of IC users' perspectives WITHDRAWN: Intermittent Catheterisation for Long-Term Bladder Management. Cochrane Database Syst Rev Intermittent catheterization: The devil is in the details Guideline for Prevention of Catheter-Associated Urinary Tract Infections Prevalence and Cost of Catheters to Manage Neurogenic Bladder Single-use Versus Multi-use Catheters: Pro Single-use Catheters Intermittent catheterization with single-or multiple-reuse catheters: clinical study on safety and impact on quality of life Adult Neurogenic Lower Urinary Tract Dysfunction and Intermittent Catheterisation in a Community Setting: Risk Factors Model for Urinary Tract Infections Recent updates in urinary catheter products for the neurogenic bladder patients with spinal cord injury An effective evidence-based cleaning method for the safe reuse of intermittent urinary catheters: In vitro testing Reuse versus single-use catheters for intermittent catheterization: What is safe and preferred? Review of current status Variability in catheter microwave sterilization techniques in a single clinic population The PRISMA statement for reporting systematic reviews and meta-analyses of studies that evaluate health care interventions: Explanation and elaboration Covidence systematic review software. Veritas Health Innovation The Joanna Briggs Institute critical appraisal tools for use in JBI systematic reviews: checklist for quasi-experimental studies (nonrandomized experimental studies) Study quality assessment tools: quality assessment tool for observational cohort and cross-sectional studies