key: cord-0324953-yidnba9f authors: Boonya-ananta, Tananant; Rodriguez, Andres J.; Ajmal, Ajmal; Du Le, Vinh Nguyen; Hansen, Anders K.; Hutcheson, Joshua D.; Ramella-Roman, Jessica C. title: Synthetic Photoplethysmography (PPG) of the radial artery through parallelized Monte Carlo and its correlation to Body Mass Index (BMI) date: 2020-07-23 journal: bioRxiv DOI: 10.1101/2020.07.22.214015 sha: 1fd540da48d0580669b8de63359d9300592af5f6 doc_id: 324953 cord_uid: yidnba9f Cardiovascular disease is one of the leading causes of death in the United States and obesity significantly increases the risk of cardiovascular disease. The measurement of blood pressure (BP) is critical in monitoring and managing cardiovascular disease hence new wearable devices are being developed to make the BP metric mode accessible to physicians and patients. Several wearables utilize photoplethysmography from the wrist vasculature to derive BP assessment although many of these devices are still at the experimental stage. With the ultimate goal of supporting instrument development, we have developed a model the photoplethysmographic waveform derived from the radial artery at the volar surface of the wrist. To do so we have utilized the relation between vessel biomechanics through Finite Element Method and Monte Carlo light transport model. The model shows similar features to that seen in PPG waveform captured using an off the shelf device. We observe the influence of body mass index (BMI) on the PPG signal. A degradation the PPG signal of up to 40% in AC to DC signal ratio was thus observed. Elevated blood pressure (BP) is considered one of the highest risk factors for cardiovascular disease. In fact, it has been estimated that 47% of all coronary heart disease worldwide is attributable to high BP 1 . BP is less than 120/80 mmHg in normotensive individuals, between 120-129/80 mmHg in prehypertensive individuals, between 130-139/80-89 mmHg in hypertensive stage 1, above 140/90 mmHg in hypertensive stage 2, and above 180/120 mmHg in hypertensive 2 . In adults, the risk of cardiovascular disease is significantly increased with obesity [3] [4] [5] , defined as having a BMI of over 30 kg/m 2 6-9 . BMI is found to correlate strongly with increased 24 hour blood pressure as well as non-dipping nocturnal blood pressure 10, 11 . Physiological changes that occur in individuals with obesity may increase the measurement uncertainties, and optical means for monitoring blood pressure will have to account for obesity-related changes to the skin optical properties 12 and increased subcutaneous tissue thickness. There are several ways to evaluate BP. Sphygmomanometers and oscillometric devices are standard techniques 13 , and the most common clinical method of measuring blood pressure is through auscultations using a sphygmomanometer with the arm maintained at heart level 14 . Invasive arterial line blood pressure measurement is often used in the Intensive Care Unit (ICU), where an arterial catheter is placed inside the radial artery of a patient to directly measure vessel internal pressure 15, 16 . Figure 1 . Typical PPG waveform with constitutive components. PPG wave components shows the systolic upstroke to systolic peak, then systolic decline into the dicrotic notch and the diastolic peak. Research into cuff-less and continuous BP devices based on photoplethysmography (PPG) is rapidly expanding. These systems provide unique diagnostic opportunities, such as monitoring of nocturnal hypertension 10 a condition strongly associated with cardiovascular events and organ damage. The PPG waveform can provide information such as pulse wave velocity, the relative change in blood volume and pressure wave reflection 17 . The systolic peak and diastolic peak are the primary and secondary features respectively as seen in Figure 1 . The primary feature is directly representative of systolic performance during the cardiac cycle 17, 18 . The vessel dilates due to the pressure wave generated by ventricular contraction. The pressure profile inside the arteries changes as the pressure wave propagates downstream from the heart to peripheral vasculature [19] [20] [21] . This is due to vessel compliance, vessel branch and bifurcations, and wave reflections encountered by the pulse during its journey from the heart. The dicrotic notch moves further away from the systolic peak as arterial line measurement location moves further down the arterial tree. Measurement of blood pressure at the radial artery located at the wrist is desirable since it is easily accessible, and instrumentation can be developed for a wearable device. The blood pressure in the radial artery at the volar surface of the wrist is more representative of blood pressure in the main arterial network as opposed to pressure in superficial vasculature of the skin 22 . In order to derive a better understanding of devices based on PPG, several groups have conducted light propagation models using Monte Carlo (MC) simulations [23] [24] [25] . Yet to our knowledge, these studies have not attempted to capture the PPG waveform in its entirety and have focused on an idealized skin model. The design of wearable devices must account for the population diversity, including physiological changes due to obesity 26-28 as well as skin tones. Our study has two main goals. First, we develop a more realistic representation of light interaction with the radial artery during a pulse as seen by a commercially available photoplethysmographer (Nellcor™ by Covidien). This location and instrumentation are the focus of our experimental efforts toward developing low-cost cuffless BP sensors. Second, we use our model to study the influence of increased BMI on the PPG waveform. Recent studies have shown that obesity creates variations in skin physiology 5, 29, 30 . These changes include but are not limited to skin barrier function, epidermal changes, dermal changes, and changes to the vascular and capillary recruitment function. In our model, both skin tone and BMI diversity will be considered. The behavior of arterial pulsatile flow can be modeled as a pulsatile flow through an elastic walled tubing 19 . The key distinction between rigid versus elastic walled flow is the fluid streamline velocity profile (Figure 2A, 2B) . The typical Poiseuille velocity profile for ideal, fully developed fluid flow in a rigid tube is constant along the length of the tube (Figure 2A ), whereas in elasticwalled tubing, the velocity profile is dependent upon the location along the streamline ( Figure 2B ). In elastic tubes, pressure waves cause a local change in fluid pressure that is then propagated downstream. Wall compliance results in oscillation of the wall and changes to tube diameter with induced pulsatile flow, as seen in the vascular system. The mechanics of large vessels such as the aorta, carotid artery, and large coronary arteries has been studied extensively [31] [32] [33] [34] [35] [36] but several authors have noted that smaller peripheral vessels, such as the one under study here, exhibit similar mechanical behaviors to larger ones 37, 38 . Considering the biomechanical changes in the arterial vascular bed, our model of PPG is divided in two parts. First, a finite element model is used to observe the dilation of the radial artery under applied pressure. Then, a Monte Carlo model is applied to this dynamic geometry to observe loss of optical signal in relation to the pulse. We have used Finite Element Analysis (FEA) to model the mechanical behavior of an arterial vessel. Our model allows us to visualize the motion of an arterial wall under an applied internal pressure. The artery is modeled in Solidworks computer-aided design (CAD) software. We have chosen the volar surface of the distal forearm as the location of our virtual PPG instrument. This choice is governed by our interest in the development of wrist-based wearable devices targeting that vessel. A drawing of the vessel with associated dimensions is shown in Figure 3A . The radial artery has been reported to have 2.5mm inner diameter and 0.2 wall thickness [38] [39] [40] . Vessel wall mass density and Poisson's ratio were set to 1160 kg/m 3 and 0.49 36, 38, 41 respectively. A radial artery wall Young's modulus of 0.70MPa was used 38 . Arterial mechanical stiffness ranges between 0.70MPa to 1.10MPa 38 . Radial artery pressure has been shown to be 5 to 15 mmHg 42 higher than the brachial artery, where blood pressure is often measured using the sphygmomanometer. An internal pressure of 130 mmHg 42 was imposed to the vessel representing the radial artery. The result of the FEM model at peak applied force is shown in Figure 3B . The figure shows a time snapshot of the vessel shape as a time-dependent pressure pulse propagates from the heart to the radial artery. This is different from a quasi-static pressure vessel representation often used to analyze vascular wall mechanics. The arterial wall is often treated as incompressible 36, 43 , pressure changes applied internally translates to vessel wall dilation, but the total wall volume is conserved. This incompressibility estimation is seen in the FEM model. Changes to the inner wall and outer wall radii are relatively small in comparison to total vessel dilation. Vessel walls have been reported to experience 10-20% diametral strain 44, 45 . Applying 0.70MPa stiffness, vessel wall shows total dilation of 0.4 mm in diameter. The mechanical behavior of the artery during pulse propagation is integrated into the light transport model (Monte Carlo) through changes in the vessel geometry. This was done by deriving the mathematical envelope of the FEA derived pulse geometry using a double ellipsoid shown in Equations 1-3. In Equations 1 and 2, x, y, and z represent each direction in the frame, locates the ellipsoid in the vertical, z, direction in the media. In Equation 1, IDx defines the dilation of the inner diameter of the vessel, 1 is the location of the primary ellipsoid in the y-axis, and 1 defines the width of the primary ellipsoid in the y direction. In Equation 2, IDxd represents the dilation of the inner diameter of the secondary ellipsoid, 2 is the location of the secondary ellipsoid in the y-axis, and 2 defines the width of the secondary ellipsoid in the y direction. Equation 3 defines the parameter as the separation between the first and second ellipsoids. The geometry developed with these equations is shown in Figure 3E . Aside from the radial artery, the Monte Carlo simulation also considered three skin layers: epidermis, dermis and subcutaneous tissue with the radial artery inserted into the adipose tissue layer. The Monte Carlo model used in this work is an adaptation of MCMatlab developed by Marti, D et al 23 . MCMatlab converts S. Jacques' "mcxyz.c" 24 from C-based code to a compact tool usable through Matlab interface. Monte Carlo framework simulates a 3-layer model to include epidermis, dermis and subcutaneous tissue with the radial artery inserted into the adipose tissue layer. The epidermis consists of multiple sublayers with various components, including melanin and melanocytes 46 . The dermis is comprised of connective tissue along with blood vessels (capillaries and arterioles) with blood volume ratios ranging from 0.2% to 7% and water volume concentration around 65% 47, 48 . The subcutis consists of subcutaneous adipose tissue and other connective tissue 49 . In this work, we modeled a commercial PPG (Nellcor™ Covidien, this device also provide pulse oximetry). This is a clinical system that incorporates two wavelength sources at 660 nm peak 22 nm FWHM and 890nm peak 54nm FWHM. The sensor provides and effective beam waist radius of 0.06 cm. The detector is located 0.97 cm away from the sources providing reflectance PPG signal. Light collector area is measured to be 0.35cm x 0.35cm which is an effective collector area of 0.1225cm 2 with numerical aperture of 0.866. The sensor schematic is shown in Figure 3C . The geometry of our Monte Carlo simulation is shown in Figure 3D . The first layer mimics the epidermis. The thickness of human epidermis is approximately 0.10 mm 50, 51 . To avoid using a very small voxel size in the simulation, we chose to scale the optical properties of the epidermis to match its optical thickness 52 . The dermis layer was 1.0 mm in thickness. Below the dermis, a subcutaneous adipose tissue layer was added. Within this layer, the radial artery is modeled as a cylinder situated at 2.5mm depth from the top surface 39, 40 . The target radial artery is constructed with a vessel wall and the internal lumen is filled with blood. Optical properties used in Monte Carlo simulations are extracted from previous work 47, 53 Analyses of the performance of commercial systems and functionality of arterial PPG must account for factors of skin tone and physiological changes with the development of obesity 26-28 . The melanin content of the skin increases with darkening skin tone. Melanin volume fraction variation can be used to model the various pigmentation levels. Lightly pigmented skin adults have a melanin volume fraction of 1.3-6.3%, moderately pigmented skin adults have a 11-16% melanin volume fraction, and darkly pigmented skin adults have an 18-43% melanin volume fraction in the epidermis 47 . Dermal thickness variations with BMI is shown to range from 1.0mm to over 2.5mm thickness 50 . Along with changes to the skin, the artery itself is situated deeper in the subcutaneous adipose tissue as adipose tissue layer thickness with increasing level of obesity. The signal from the radial artery is discovered to vary in depth with different levels of obesity from 2.5mm for nonobese individuals at BMI of 25 to 3.5mm for BMI of 45 39, 40 . Trans-epidermal water loss (TEWL) the process where water evaporates through the epidermis from the dermis is disruption in the obese 30 . As BMI increases so does TWEL 58 , and this effect directly contributed to dry skin and skin irritation often seen in the obese population. A summary of all the parameters utilized in the modeling is shown in Table 3 . Table 3 are applied to the MC geometry through two different methods. The changes to TWEL and dermal blood contented are applied to the media layer properties. TWEL changes the amount of water in the dermal layer. The water loss percentage is multiplied to standard dermal water content. Similarly, dermal blood content percentage changes are multiplied to blood content in the dermis. For geometric changes of dermal thickness and artery depth, these properties are changed directly in the geometry construction. A power absorption map of our geometry is shown in Figure 4A and Figure 4B . This is a single slice in the three-dimensional geometry at a single location of the pulse waves. Figure 4C shows an example of PPG waveform generated for the non-obese case. The returned signal is comprised of an AC and a DC baseline signal, where the AC signal represents the pulsatility of the vasculature. A comparison of the AC to DC signal ratio of each waveform shows the degradation of the signal with increasing BMI. Different PPG signals are generated through varying four different physiological changes observed with obesity. These changes can be seen in Table 2 . Preliminary experimental data using the PPG sensor ( Figure 5) shows features similar to that seen in simulation curve. The PPG of three different individuals was captured experimentally ( Figure 5A ). There is significant variation between individuals ( Figure 5A green, red, blue), however the features of the shape can be identified in all three cases. The PPG data is adjusted from pulse shape in time to PPG with respect to pulse position ( Figure 5B, 5C ). This is done by matching the start and end point of the two sets as well as the primary peak and overlay on top of each other. Figure 6 shows changes AC/DC signal from four different changes attributed to obesity across different BMI levels. Figure 6A shows a 14.0% total signal decline associated with changes to trans-epidermal water loss between normal to highest BMI level. Figure 6B shows an 18.0% change with dermal blood content. Figure 6C shows total percentage change between smallest and largest dermal thickness is 41.4%. In Figure 6D , percentage change between shallowest and deepest arterial depth is 32.1%. Similar trends of signal degradation are seen across the four different changes which occur with obesity: trans-epidermal water loss, blood perfusion, dermal thickness, and radial artery depth. Each scenario shows changes isolated to each feature of obesity. Figure 7 highlights changes seen at four different skin tones classified by epidermal melanin concentration. Total AC to DC signal ratio change is seen at 17.1% between 3% epidermal melanin concentration and 42% epidermal melanin concentration. Waveforms for different levels of obesity are generated through varying the combined dermal thickness and arterial depth to indicate corresponding physiological changes. Figure 8 shows generated waveforms at non-obese, obese 1, obese 2 and obese 3 at two different photon counts. Changes to BMI and obesity level is presented through the increase of dermal thickness from 1.0mm to 2.5mm and arterial depth from 2.5mm to 4.5mm. A comparison of the waveform shows the degradation of the baseline signal as well as changes to the wave shape itself. Across each level of obesity, at both 100 million and one billion photon counts, there is a significant diminishing of the amplitude of the signal. Figure 9A shows the four different signals superimposed on top of each other on the same curve. Baseline signal change between each obesity level and non-obese is shown in Table 2 . The baseline change ranges from 12.7% to 54.5%. Similar changes are seen in both the 100 million photons set as well as the one billion photons set as to be expected. The advantage of using lower number of photons is the computational time for each simulation in exchange for a higher standard deviation at each point. The non-obese BMI of 25kg/m 2 , the AC/DC signal ratio is 8.1% and at the maximum level calculated the ratio is 4.66%. This accumulates to the total percentage change of 43% between the Non-Obese waveform and Obese 3 waveform. The degradation of the return signal gives an indication to the decrease in quality of the signal that will be received by the PPG device. A loss in signal quality is shown to increase as BMI and obesity level increase with changes in dermal thickness and arterial depth. PPG devices designed to periodically or continuously probe arterial blood pressure as a wearable device at the wrist must be designed to be mindful of the significant effects of obesity on optical performance. The advantage of the development of synthetic waveform is understanding these changes and the significance of their impact. We have demonstrated an approach to modeling PPG signals utilizing a combination of FEM and Monte Carlo modeling. Ultimately, our intent is to develop a wearable health device capable of assessing continuous cuff-less blood pressure with comparable level of performance across populations of varying body size. Cuff-less, continuous blood pressure monitoring can provide critical information concerning nocturnal cardiovascular function, which is a significant indicator of overall cardiac health. This work focuses on a commercial PPG sensor measurement of pulsatile flow in the radial artery. Our simulations demonstrate that with no changes to the underlying vessel mechanical behavior, an increase in BMI significantly impacts the PPG waveform. The AC/DC ratio changes from 14% to 41% as BMI increases from 20 to 40 for isolated changes attributed with increasing level of obesity and BMI. Increase to dermal thickness increases the total volume of absorbers and scatterers through which photons must traverse to reach the artery and return to the detector. Increased radial artery depth also increases the total travel distance for photons to reach the vessel wall as and return to detector. Vessel expansion and contraction, under pressure changes of the cardiac cycle, is directly associated with the pulsatility of the PPG waveform 59 , the presence of obesity degrades the quality of the signal even when the vessel mechanics remain unchanged. Synthetic waveform generation provides an insight and prediction of optical device functionality. Future work will focus on the understanding of waveform morphology through connecting various cardiovascular conditions with the corresponding PPG. The addition of synthetic waveform generation can allow for study of isolated effects from a variety of factors on the output PPG. Post-processing of the PPG such as first and second derivative and frequency domain information can be used as tools for feature extraction for desired information. This work is limited to the detection of the radial artery and does not consider the pulsatile effect of the superficial arterioles, this will be the focus of future addition to the present model. This work was supported by the National Science Foundation Engineering Research Center for Precise Advanced Technologies and Health Systems for Underserved Populations (PATHS-UP) (#1648451). High blood pressure and all-cause and cardiovascular disease mortalities in community-dwelling older adults Understanding blood pressure readings Deaths: leading causes for 2017 Epidemiology of cardiovascular disease in the 21st century: updated numbers and updated facts Obesity and subtypes of incident cardiovascular disease Quetelet's index (W/H2) as a measure of fatness A comparison of the Slaughter skinfold-thickness equations and BMI in predicting body fatness and cardiovascular disease risk factor levels in children Estimates of excess deaths associated with body mass index and other anthropometric variables Comparison of bioelectrical impedance and BMI in predicting obesity-related medical conditions Nocturnal hypertension: new technology and evidence Impact of obesity on 24-hour ambulatory blood pressure and hypertension Obesity and the skin: skin physiology and skin manifestations of obesity Blood pressure response to the daily stressors of urban environments: Methodology, basic concepts, and significance The Japanese Society of Hypertension Guidelines for the Management of Hypertension. Chapter 2. Measurement and clinical evaluation of blood pressure Methods of blood pressure measurement in the ICU Oscillometric brachial mean artery pressures are higher than intra-radial mean artery pressures in intensive care unit patients receiving norepinephrine Hypertension assessment using photoplethysmography: a risk stratification approach Kaplan's Essentials of Cardiac Anesthesia E-Book Aortic input impedance in normal man: relationship to pressure wave forms McDonalds blood flow in arteries: Theoretical, experimental and clinical principles MCmatlab: an open-source, userfriendly, MATLAB-integrated three-dimensional Monte Carlo light transport solver with heat diffusion and tissue damage Monte Carlo simulations of light transport in 3D heterogenous tissues Monte Carlo analysis of optical interactions in reflectance and transmittance finger photoplethysmography Fitbits and other wearables may not accurately track heart rates in people of color Quick analysis of optical spectra to quantify epidermal melanin and papillary dermal blood content of skin Characterization of skin function associated with obesity and specific correlation to local/systemic parameters in American women Skin changes in the obese patient Non-contact Quantification of Jugular Venous Pulse Waveforms from Skin Displacements Understanding Arterial Pressure Waveforms Pulse wave analysis of the aortic pressure waveform in severe left ventricular systolic dysfunction Arterial viscoelasticity: role in the dependency of pulse wave velocity on heart rate in conduit arteries Wall stress analysis in small asymptomatic, symptomatic and ruptured abdominal aortic aneurysms Mechanical properties of arteries in vivo Significance of source and size in the mechanical response of human cerebral blood vessels Arterial stiffness identification of the human carotid artery using the stress-strain relationship in vivo Differences in the properties of the radial artery between Cun, Guan, Chi, and nearby segments using ultrasonographic imaging: a pilot study on arterial depth, diameter, and blood flow Association of obesity with anatomical and physical indices related to the radial artery in Korean adults Cardiovascular fluid mechanics-lecture notes Brachial and Radial Systolic Blood Pressure Are Not the Same: Evidence to Support the Popeye Phenomenon Constitutive modelling of arteries Finite element modeling of blood flow in arteries Finite element modeling and simulation of arteries in the human arm to study the aortic pulse wave propagation Pheomelanin as well as eumelanin is present in human epidermis Origins of tissue optical properties in the UVA, visible, and NIR regions Hemoglobin concentration, total hemoglobin mass and plasma volume in patients: implications for anemia The optics of human skin Effects of age, gender, BMI, and anatomical site on skin thickness in children and adults with diabetes Measurement of epidermis, dermis, and total skin thicknesses from six different body regions with a new ethical histometric technique. Türk Plastik, Rekonstrüktif ve Estetik Cerrahi Dergisi Method for depth-resolved quantitation of optical properties in layered media using spatially modulated quantitative spectroscopy Optical properties of biological tissues: a review Dynamics and Fluctuations in Biomedical Photonics VIII. 78980Z (International Society for Optics and Photonics) Influence of oxygen saturation on the optical scattering properties of human red blood cells in the spectral range 250 to 2000 nm Optical properties of skin, subcutaneous, and muscle tissues: a review Dermal component-based optical modeling of skin translucency: impact on skin color The influence of body mass index on skin susceptibility to sodium lauryl sulphate McDonald's blood flow in arteries: theoretical, experimental and clinical principles