key: cord-0317067-mprw4t0y authors: Johnson, C.; Ferguson, K.; Smith, T.; Wallace, D.; McConnell, K.; Conserve, D.; martin, i. title: EVALUATION OF THE PANBIO SARS-COV-2 RAPID ANTIGEN DETECTION TEST IN THE BAHAMAS date: 2021-07-15 journal: nan DOI: 10.1101/2021.07.13.21260402 sha: b956721a6b7a5cd057d1b3a6ea55cb084e9de02d doc_id: 317067 cord_uid: mprw4t0y Identification and isolation of persons infected with SARS-COV-2 are key mitigation strategies in the current pandemic, and rapid antigen detection tests (RADTs) offer the promise of decreased turnaround time to diagnosis when compared with gold standard RT-PCR testing. We evaluated the analytical performance of the Abbott Panbio RADT on nasopharyngeal samples stored at the Ministry of Health National Reference Lab in Nassau, Bahamas. The Panbio demonstrated a test sensitivity of 94% and a specificity of 100% on 50 PCR negative samples and 50 samples presumed to be infectious based on having PCR cycle threshold values below 30. Additionally, in our examination of operator results there was low interpersonal variation (1%) among three blind operators and significant correlation between sample Ct value and perceived signal strength on the RADT device. However, three PCR positive samples below Ct 30 were misdiagnosed by RADT, including one sample with Ct value less than 20. These results support the use of the Panbio in symptomatic persons to detect active SARS-COV-2 infections, with the caveat that RADT-negative samples should be confirmed by RT-PCR. that has been added to the World Health Organisation (WHO) Emergency Use Listing (EUL) (2) and became commercially available in the Bahamas in November 2020. The primary goals of SARS-COV-2 diagnosis and contact tracing are to accurately identify and rapidly isolate "contagious" persons in order to limit community transmission (3) . In this regard, it has emerged that samples with low viral loads/ high cycle threshold (Ct) values do not replicate in viral culture, which has been used as a surrogate for contagiousness or infectivity. Significantly, several studies have demonstrated negligible cultivability of samples with N or E gene Ct values above 27-35 (with the wide range probably due to variability in the RT-PCR and viral culture assays used) (4-9). However, while it has been established that RADT test sensitivity decreases at around Ct 25-30 (10-18), recent reports find that RADT results correlate more to viral cultivability than do RT-PCR results, supporting the postulate that the RADT detects active SARS-COV-2 infections, and that there is minimal clinical significance to their reduced sensitivity at higher Ct values compared to gold standard RT-PCR testing (19-21). We sought in this study to assess the analytical performance of the Panbio™ in a selection of 100 frozen stored samples at the Ministry of Health National Reference Lab (NRL) in the Bahamas, including 50 RT-PCR negative samples and 50 RT-PCR positive samples with presumed infectious quantities of virus as garnered by their Ct value (<30). The objective of this analysis was to specifically assess the ability of trained operators using the Panbio™ RADT to detect SARS-COV-2 active infections in symptomatic, presumed contagious individuals with high viral loads. Samples obtained from symptomatic persons meeting the clinical case definition for SARS-COV-2 infection (3) from community primary public health clinics on islands throughout the Bahamas archipelago are routinely sent to NRL. 100 de-identified retrospective samples stored frozen in viral transport medium (VTM) at the NRL were selected using the following criteria: a) 50 PCR positive samples from symptomatic persons (E gene Ct value <30) and b) 50 PCR negative samples from persons with ILI . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) The copyright holder for this preprint this version posted July 15, 2021. ; https://doi.org/10.1101/2021.07.13.21260402 doi: medRxiv preprint symptomatology. These samples were selected from the month of November 2020, during the 'second wave' of the covid-19 epidemic in the Bahamas. 150ul of each thawed VTM sample was combined with 150ul of the Abbott® extraction buffer (dilution ratio 1:1) and 130ul of the mixed solution was dispensed onto the Abbott® Panbio™ rapid antigen test device as per manufacturer's product insert (22). Results were recorded in 15-20 minutes by three independent testers. Samples were previously identified as SARS-COV-2 positive by detection of the E gene using the Charite (Berlin) RT-PCR protocol (23) on the ABI QuantStudio 6 Flex at the NRL. RADT performance and results interpretation was performed by three independent operators concurrently performing blind testing; operators were asked to score their results as negative (score of 0), weakly detected-1, or strongly detected-2. In the case of discordance among the three operators, the consensus result was taken. Sensitivity Spearman's rank coefficient was calculated for operator-ranked results and graphs constructed in Graphpad Prism Version 9.0.2. (Table 1) . Samples were collected during the month of November 2020, during the peak of the 'second wave' of the SARS-COV-2 epidemic in the Bahamas. . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted July 15, 2021. ; https://doi.org/10.1101/2021.07.13.21260402 doi: medRxiv preprint LOD for signal strength=1 (weak) was between E gene approximate Ct values 27 and 32, and for signal strength=2 (strong) was between approximate Ct 23 and 27 ( Figure 2A ). Operator-ranked signal strength (i.e. 0, 1, 2) of the 50 PCR positive samples was moderately inversely correlated with sample Ct value (r s = -0.59, -0.57 and -0.59, p=0.001 for each of the three operators)-see Figure 2B . Of 3 independent operators reading qualitative results (i.e., RADT positive or negative) on blinded samples, overall there was 1% discordance (described as a dissenting result among the three submitted), due to one of the three operators misdiagnosing a PCR positive sample (Ct 28.35) as negative on the Abbott® RADT. Overall, the Panbio™ RADT performed well in our study samples derived from symptomatic individuals throughout the Bahamas archipelago, with a sensitivity and specificity of 94% and 100% respectively in samples with Ct value <30. Notwithstanding this, and in common with previous studies, we found that high Ct value is not the sole . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted July 15, 2021. ; https://doi.org/10.1101/2021.07.13.21260402 doi: medRxiv preprint determinant of false negative test results on the Panbio™ RADT, since one sample with Ct value <20 was given a false negative result by all three test operators. This phenomenon has been seen in other settings where the Panbio™ RADT produced false negatives in a few of these high viral load samples (10,13). Interestingly, this issue of missing some low Ct samples seems to be a general feature of RADTs, as revealed in a recent head-to-head comparison between 4 RADTs and an automated antigen test, in which the latter was the only among those evaluated that did not exhibit this phenomenon (13). Though an apparently relatively rare and a minor contributor to overall test performance in this and previous studies, false negative results in low Ct/high viral load samples may be significant in real-world contexts in which missing active infections can lead to preventable transmission events (although the clinical relevance of this remains to be seen, since RADT results correlate better with viral cultivability than PCR results (19-21)) In our opinion, this observation reveals an important limitation to use of RADTs for mass screening activities and also lends strong support to following an algorithm in which symptomatic RADT negative persons are re-tested (by RADT or preferably RT-PCR), as is recommended by the WHO (1). Indeed, previous evaluations of RADT accuracy in asymptomatic groups, such as children (24) and household contacts of known cases (25) have reported low sensitivity of the Panbio™ in these populations, probably due to low viral load (high Ct value). In previous analyses the Panbio™ test LOD has been described as correspondent to a Ct value of between 24 and 29 (10-18). Our LOD results using sample serial dilution were consistent with these published LODs, since we estimate the LOD as between approximate Ct 28 and 32, as there was a weak line detected in a sample with Ct 28 but no line at Ct 32 (Figure 2A ). In addition, blinded operator-ranked results of RADT signal intensity were significantly inversely correlated in the nonparametric analysis to sample Ct value ( Figure 2B) . Notably, however, samples as low as Ct 23 were concordantly ranked as weak (score=1) by all operators ( Figure 2B) . Therefore, despite a concordance of 99% among 3 independent operators, and the fact that the single discordant result recorded by one of the operators was reported on a sample with a . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted July 15, 2021. ; https://doi.org/10.1101/2021.07.13.21260402 doi: medRxiv preprint relatively high Ct value (28.35) and faint RADT band, there remains the possibility that samples with Ct values <25 could produce "weak" signal strength on the RADT and be subject to interpersonal variation leading to misdiagnosis of potentially infectious samples. The impact of interpersonal variation in Panbio™ test result interpretation can be minimised by measures such as implementation of two-reader policy, or device automation. In addition, expansion of quality assurance services for SARS-COV-2 antigen testing, such as proficiency testing, can help to identify sources of error in test administration and results interpretation particularly in borderline/faint samples (26). In totality, the present analysis supports the utility of the Panbio™ RADT within the prescribed parameters of early symptomatic infection, demonstrating a high sensitivity in presumed infectious samples. However, use of this test must be tempered with the understanding that it may (on relatively rare occasions) misdiagnose low viral load/high Ct value samples and that these samples may produce faint test lines on the rapid test device leading to false negative results due to variability in results interpretation. In addition to this, there are instances where low Ct value/high viral load samples produce weak RADT signal strength and/or false negative results for yet unknown reasons. Thus, while RADT is a useful tool to enhance access to SARS-COV-2 testing, this data also supports cautious rollout, limited to detection of cases in the early symptomatic phase, and supported by quality oversight and RT-PCR confirmation of RADT-negative symptomatic cases. The authors declare no conflict of interest Ethics approval for this study was granted by the Bahamas National Covid-19 Ethics Committee, April 2021. . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted July 15, 2021. is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) The copyright holder for this preprint this version posted July 15, 2021. Valdivia A, Solano de la Asunción C, Ferrer J, Colomina J, Navarro D. (2020) Clin Microbiol Infect. Nov 13:S1198-743X(20)30697-2. Antigen-detection in the diagnosis of SARS-CoV-2 infection using rapid immunoassays: interim guidance License: CC BY-NC-SA 3.0 IGO 2. WHO Emergency Use Listing for In vitro diagnostics (IVDs) Detecting SARS-CoV-2 Public health surveillance for COVID-19: interim guidance: Interim guidance Viral RNA load as determined by cell culture as a management tool for discharge of SARS-CoV-2 patients from infectious disease wards Evaluation of a rapid antigen test (Panbio COVID-19 Ag rapid test device) for SARS-CoV-2 detection in asymptomatic close contacts of COVID-19 patients