key: cord-0301580-11qqceyf authors: Williams, D. M.; Hornsby, H.; Shehata, O. M.; Brown, R.; Zafred, D.; Shun-Shion, A. S. M.; Hodder, A. J.; Bliss, D.; Metcalfe, A.; Edgar, J.; Gordon, D. E.; Sayers, J. R.; Nicklin, M. J.; Collini, P. J.; Brown, S.; de Silva, T. I.; Peden, A. A. title: A high content microscopy-based platform for detecting antibodies to the nucleocapsid, spike and membrane proteins of SARS-CoV-2 date: 2021-10-18 journal: nan DOI: 10.1101/2021.10.14.21264873 sha: 43c803449911190a81aa77e825f5a9ad0099fdf8 doc_id: 301580 cord_uid: 11qqceyf The strong humoral immune response produced against the SARS-CoV-2 nucleocapsid (N) and spike (S) proteins has underpinned serological testing but the prevalence of antibody responses to other SARS-CoV-2 proteins, which may be of use as further serological markers, is still unclear. Cell-based serological screening platforms can fulfil a crucial niche in the identification of antibodies which recognise more complex folded epitopes or those incorporating post-translation modifications which may be undetectable by other methods used to investigate the antigenicity of the SARS-CoV-2 proteome. Here, we employed automated high content immunofluorescence microscopy (AHCIM) to assess the viability of such an approach as a method capable of assaying humoral immune responses against full length SARS-CoV-2 proteins in their native cellular state. We first demonstrate that AHCIM provides high sensitivity and specificity in the detection of SARS-CoV-2 N and S IgG. Assessing the prevalence of antibody responses to the SARS-CoV-2 structural membrane protein (M), we further find that 85% of COVID-19 patients within our sample set developed detectable M IgG responses (M sensitivity 85%, N sensitivity 93%, combined N + M sensitivity 95%). The identification of M as a serological marker of high prevalence may be of value in detecting additional COVID-19 cases during the era of mass SARS-CoV-2 vaccinations, where serological screening for SARS CoV-2 infections in vaccinated individuals is dependent on detection of antibodies against N. These findings highlight the advantages of using cell-based systems as serological screening platforms and raise the possibility of using M as a widespread serological marker alongside N and S. Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), the causative agent of Coronavirus Disease 2019 , has impacted the lives of nearly everyone in the developed world (Wu, Zhao et al. 2020) . Controlling SARS-CoV-2 infections within communities and across populations requires accurate knowledge of both current and previous cases of COVID-19. Serological assays fulfil critical roles towards the latter, identifying individuals previously exposed to the virus who may potentially be immune whilst also contributing towards the construction of accurate epidemiological estimates of infection rates. Analysis of COVID-19 patient serum obtained through serological sampling has also provided invaluable information on the kinetics and profile of antibody responses produced against SARS-CoV-2 in relation to disease outcomes (Long, Liu et al. 2020 , Long, Tang et al. 2020 , Ripperger, Uhrlaub et al. 2020 . Antibody tests of high sensitivity and specificity capable of providing a comprehensive overview of the humoral immune landscape induced after exposure to SARS-CoV-2 are therefore desirable to track the spread of the virus within communities and gain deeper insight into the interplay between virus and host immune system during acute and convalescent phases of infection. To date, SARS-CoV-2 population-level serological testing relies upon detection of antibodies against either the SARS-CoV-2 nucleocapsid (N) or spike (S) proteins as both induce detectable humoral immune responses in the vast majority of those infected (National 2020) . The serological assays currently in use for SARS-CoV-2 display varying levels of sensitivity and specificity and come with a variety of strengths and weaknesses . Screening for reactivity of patient sera against purified versions of the SARS-CoV-2 N and S proteins by ELISA is considered the gold standard for identification of patients who have previously been infected with SARS-CoV-2 (Amanat, Stadlbauer et al. 2020) . However, this technique is limited in the scope of viral antigens which can be assayed due to the difficulty of producing viral proteins with physiochemical properties compatible with their production in a purified form (Johari, Jaffe et al. 2021) . Although not employed for mass serological testing, additional techniques such as viral peptide and protein arrays have been helpful in identifying other immunogenic SARS-CoV-2 proteins. In contrast to ELISA based methods, viral peptide and protein arrays can provide almost complete coverage of the viral proteome allowing a more global overview of the humoral immune response and fine mapping of antibody epitopes, the latter being of particular of importance for looking at antibody crossreactivity with other human coronaviruses and understanding the mechanism of action of neutralizing antibodies (Jiang, Li et al. 2020 , Shrock, Fujimura et al. 2020 , Ladner, Henson et al. 2021 , Li, Xu et al. 2021 , Stoddard, Galloway et al. 2021 . The use of peptide fragments as bait however, limits detection of antibodies to those dependent on binding to contiguous stretches of amino acids. An additional drawback is that the proteins and peptide fragments used in both ELISA and protein or peptide arrays often do not include post translational modifications which may be important features of viral epitopes recognised by host antibodies. Cell based methods such as immunofluorescence microscopy or flow cytometry can overcome some of the limitations associated with ELISA and viral peptide arrays and as such have been applied to different degrees to screen for the presence of antibodies induced after pathogenic infection (Meyer, Drosten et al. 2014 , Grzelak, Temmam et al. 2020 , Hambach, Stahler et al. 2021 . In contrast to ELISA and protein or peptide arrays, viral proteins expressed in mammalian cells are produced in their three dimensional fully folded state and are likely subject to the same post translational modifications viral proteins undergo in infected cells. Screening of serum samples by immunofluorescence microscopy has demonstrated high sensitivity and specificity when applied previously during the SARS-CoV and MERS outbreaks (Chan, Ng et al. 2004 , Meyer, Drosten et al. 2014 , however this methodology has to date largely been used as an adjunct to those techniques described above during the current SARS-CoV-2 pandemic (Grzelak, Temmam et al. 2020 , Wolfel, Corman et al. 2020 . In this paper, we describe the development of a cell-based expression platform combined with automated high-content immunofluorescence microscopy (AHCIM) to identify the presence of antibodies to specific SARS-CoV-2 proteins and test its performance in comparison to ELISA based COVID-19 serological testing. Calibration of our automated system using StrepTagged SARS-CoV-2 N and S proteins demonstrated that automated immunofluorescence-based antibody screening could detect N and S antibodies in sera collected from COVID-19 patients with high sensitivity and specificity. Most interestingly, further application of this system identified antibodies in a significant number of COVID-19 serum samples against the SARS-CoV-2 membrane (M) protein, opening up the possibility of using M as a third high prevalence serological marker alongside N and S. Together, our findings demonstrate the versatility and utility of a cell-based immunofluorescence assay for use as both a tool in seroprevalence studies and as an investigative tool to gain greater insight into humoral immune responses to pathogenic infection. is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. ; https://doi.org/10.1101 https://doi.org/10. /2021 For immunofluorescence microscopy experiments StrepTagged proteins were detected using StrepTactin-DY549 (1:1000, IBA Lifesciences 2-1565-050) and human IgG with a rabbit anti-human-IgG AlexaFluor-488 (1:500, Life technologies A1101). Primary antibodies used for western blotting were mouse anti-SARS-CoV-2 nucleocapsid (Genetex, GTX632269), mouse anti-strep-mAb (1:1000, IBA Lifesciences) and rabbit anti-GAPDH (Proteintech, 60004-1-Ig). Secondary antibodies for western blotting were HRP conjugated donkey anti-human IgG (1:2000, Biolegend 410902), HRP conjugated anti-mouse (1:2000, Jackson Immuno Research 115-035-008) and HRP-conjugated goat anti-rabbit (1:2000, Jackson Immuno Research 111-035-144). StrepTagged SARS-CoV-2 S, N and M expression constructs were a kind gift from Professor Nevan Krogan, University of California. All SARS-CoV-2 proteins were human codon optimised and expressed from pLVX-EF1alpha-IRES-puro (N and M) or pTwist-EF1alpha-IRES-puro (S) plasmids as previously described (Gordon, Jang et al. 2020 ). Serum samples used were from healthcare workers (HCWs) recruited at Sheffield Teaching Hospitals NHS Foundation Trust (STH) as part of the COVID-19 Humoral Immune Responses in front-line HCWs (HERO) study, sampled in May and June 2020 (Hodgson, Colton et al. 2021) . Regulatory approval was provided by HRA and Health and Care Research Wales (20/HRA/2180, IRAS ID: 283461). Anonymised serum samples from hospitalised COVID-19 patients and serum collected before 2017 during routine clinical care were used for assay validation purposes with approval from the STH R&D office as per standard practice. HEK-293T cells (CRL-3216) originally obtained from ATCC, were grown in Dulbeccos modified Eagles medium (Merck, D6429) supplemented with 10% fetal bovine serum (Gibco, 16140071), 100IU / ml penicillin, 100 µg/ ml streptomycin and 2mM glutamine (Merck, G1146-100ML). Cells were cultured at 37°C in a 5% CO2 humidified incubator. . CC-BY-NC-ND 4.0 International license It is made available under a perpetuity. is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. ; https://doi.org/10.1101 https://doi.org/10. /2021 For initial pilot experiments to assess the ability of immunofluorescence microscopy to detect SARS-CoV-2 antibodies in patient sera, cells were seeded onto poly-l-lysine (Sigma, P8920) coated glass coverslips and left to adhere overnight at 37°C before transfection. The next day, plasmid DNA was mixed with transfection reagent (Fugene HD, E2311) at a ratio of 1 µg DNA : 3ul Fugene in Optimem (GIbco, 31985-070) and incubated at room temperature for 10 minutes. The DNA-Fugene mixture was then diluted 1:10 in complete DMEM (10% fetal bovine serum (FBS), 1% penstrep) and added to cells. Cells were left at 37°C for a minimum of 16 hours before being processed for immunofluorescence microscopy. Following transfection, the cell culture media was aspirated, and the cells and coverslips washed twice with phosphate buffered saline (PBS) to remove residual media. Cells were then fixed for 15 minutes at room temperature with 4% paraformaldehyde (PFA; Park Scientific Limited, 04018-4) in PBS, PFA quenched with 100mM glycine (Fisher Scientific, 10070150) for 5 minutes and cells permeabilised and blocked with a 0.1% Saponin (Sigma, S4521) and 1% bovine serum albumin (BSA) (Fisher Scientific, BP1600-100) solution (diluted in PBS) (IF buffer) for 10 minutes. For detection of intracellular proteins (N and M), coverslips were incubated with patient serum diluted at 1:125 in IF buffer for 1 hour at room temperature. For detection of cell surface S, coverslips were incubated with patient serum (1:125 in DMEM) for 1 hour at 37 °C prior to fixation. The patient serum was aspirated, and the coverslips were washed three times over 15 minutes using IF buffer. Bound antibodies and StrepTagged viral proteins were detected by incubating the coverslips with an antihuman IgG secondary antibody conjugated to AlexaFluor-488 (1:500, Life technologies A11014) and StrepTactin-DY549 (1:1000, IBA lifesciences 2-1565-050, 1 in 500 in IF buffer) for 1 hour at room temperature. Cell nuclei were then counterstained with DAPI (Sigma, D9542), coverslips washed a further 3 times for 5 minutes each with IF buffer and mounted on microscope slides with Prolong Gold Antifade Mountant (ThermoFisher scientific, P36930). Patient samples were imaged using a 20x objective (Olympus BX61 motorised wide-field epifluorescence microscope) and images collected with a Hamamatsu Orca monochrome camera. Non-transfected cells served as internal specificity controls for each patient serum incubation and were used to calculate background non-specific antibody staining. For manual quantification of signal strength from patient antibodies recognising SARS-CoV-2 N and S proteins, average 488-. CC-BY-NC-ND 4.0 International license It is made available under a perpetuity. is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. ; https://doi.org/10.1101 https://doi.org/10. /2021 fluorescence intensity measurements were obtained in FIJI-ImageJ (Schindelin, Arganda-Carreras et al. 2012 ) from a minimum of six non-transfected and six transfected cells by drawing around the perimeter of each cell of interest and measuring the average 488-fluorescence intensity. Background fluorescence was then accounted for as follows. 488-fluorescence intensity measurements from the six non transfected cells per patient were averaged, and this value was then subtracted from each of the six 488-intensity measurements obtained from transfected cells measurements per patient. A ratio of cell associated fluorescence from non-transfected cells compared to transfected cells was obtained by dividing 488-intensity measurements from transfected cells identified by StrepTactin 549 staining by 488-intensity measurements from nontransfected cells. For manual quantification of antibodies recognising M, 488-signal associated with Golgi localised M identified by the StrepTag staining was quantified from a minimum of six transfected cells. For non-transfected cells 488-signal from a small perinuclear region of equivalent size was measured and a ratio calculated as described above for N and S. Of note, our initial experiments assaying IgG responses against N suggested patient sera N antibody levels were limiting as we observed that having too many transfected cells significantly reduced the number of virus specific antibodies bound to transfected cells, as indicated by a weaker anti-human IgG-1 Alexa-488 signal at higher transfected cell densities (Supplementary Figure 1) . To avoid using larger volumes of patient sera to boost the transfected cell specific Alexa-488 signal, in all other experiments using tagged SARS-CoV-2 proteins we instead controlled the ratio of transfected to non-transfected cells so that there was a significant excess of non-transfected cells (ratio of 10:1 non transfected: transfected cells). HEK cells seeded in 6 well plates were transfected with SARS-CoV-2 plasmid DNA encoding either S, N or M proteins or mock transfected as described above, Cells were detached by rinsing with conditioned medium 24 hours after transfection. Mock transfected cells were mixed with SARS-CoV-2 protein transfected cells at a ratio of 10:1 before being seeded into a poly-l-lysine coated 96 well plate (Greiner-bio-one, 655098) at a density of 30,000 -40,000 cells per well. Cells were then left to adhere overnight at 37°C The next day, cells were processed for immunofluorescence microscopy as described above with a maximum of 96 serum samples analysed per plate. Following labelling of bound human IgG and StrepTagged SARS-CoV-2 proteins, images were acquired with a 20x objective on a Molecular is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. ; https://doi.org/10.1101/2021.10.14.21264873 doi: medRxiv preprint Devices Image Xpress Micro high content microscope, collecting signal from DAPI, FITC and Texas red channels. For image acquisition, wells to be imaged were selected within the microscope software and a minimum of 4 sites per well imaged using a 20x objective. Captured images were analysed using the Molecular Devices Image Xpress Micro software package. Transfected and nontransfected cells were identified as described in Figure 2 . Total fluorescence signal for each cell identified by the software was divided by cell area and an average fluorescence intensity for transfected and non-transfected cells calculated from multiple cells per serum sample assayed. HEK-293T cells expressing the nucleocapsid protein grown in 6 or 12 well dishes were detached by rinsing with conditioned media and cell suspensions collected in 15ml falcon tubes. Cells were then pelleted by centrifugation at 1000 x g for 10 minutes and the conditioned media aspirated from the pellet. Following one wash with PBS, the cell pellet was lysed in SDS sample buffer containing 5% βmercaptoethanol and boiled for 10 minutes at 95°C. Cell lysates were then loaded onto a Tris-Glycine polyacrylamide gel and proteins resolved by SDS-PAGE gel electrophoresis. Gels were transferred overnight via wet transfer in 1xTobin buffer (25 mM Tris, 192 mM glycine, 20% Methanol) onto PVDF membranes (Cytiva, 10600023) before blocking in PBS containing 0.1% Tween20 (Sigma, P7949) (PBST) and 5% skim powder milk (Sigma, 70166) for 1 hour at room temperature. Membranes were then probed with either patient sera diluted 1:1000 in blocking solution or an anti-StrepTag monoclonal antibody (1:1000, IBA lifesciences 2-1507-001) overnight or for 1 hour at room temperature. Membranes were washed three times for 5 minutes each with PBST and incubated with either donkey anti-human (1:2000, Biolegend 410902) or anti-mouse (1:2000, Jackson Immuno Research 115-035-008) HRP conjugated secondary antibodies for a further 1 hour. After three 5 minute washes with PBST, membranes were incubated with Clarity Western ECL substrate (Bio-rad, 170-5061) and signal detected using a LICOR c-DiGiT imaging system. Quantification of western blots was performed using the Image Studio Lite software package. ELISA for S and N proteins were performed as previously described (Hodgson, Colton et al. 2021 ). High binding microtiter plates (either Immulon 4HBX; Thermo Scientific, 6405, or Nunc MaxiSorp; Thermo Scientific, 442404) were coated overnight at 4oC with 50 µl/well SARS-CoV-2 protein diluted in PBS (pH 7.4). Either full-length spike produced in mammalian cells (Johari, Jaffe et al. 2021) , or . CC-BY-NC-ND 4.0 International license It is made available under a perpetuity. is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. ; https://doi.org/10.1101/2021.10.14.21264873 doi: medRxiv preprint nucleocapsid produced in E.coli were used as antigens. Recombinant nucleocapsid protein was produced as previously described (Hodgson, Colton et al. 2021) . Once coated, plates were washed 3x with 0.05% PBS-Tween, and blocked for 1 hour with 200 µl/well casein blocking buffer. Plates were emptied (no wash step) and loaded with 100 µl/well of samples and controls, diluted to 1:200 for a 2 hour incubation. Plates were washed 3x and loaded with 100 µl/well goat anti-human IgG-HRP conjugate (Invitrogen, 11594230) at 1:500 dilution for a 1 hour incubation. After a final 3x wash, plates were developed with 100 µl/well TMB substrate (KPL, 5120-0074) for 10 minutes in the dark. The reaction was stopped by addition of 100 µl/well 1% HCl stop solution (KPL, 5150-0021). Absorbances were read immediately at 450 nm. All graphs included in figures were made using the GraphPad Prism software package. Statistical comparisons between groups using an unpaired t-test and pearsons correlation coefficient analysis of N, S and M antibody relationships were all performed using GraphPad Prism. A recent complication in SARS-CoV-2 serological testing has arisen from the use of S as the immunogen applied in SARS-CoV-2 vaccination programs. This prevents the use of S as a marker of past infection and makes detection of prior SARS-CoV-2 infection in those vaccinated dependent on assaying for antibodies against N. The identification of additional SARS-CoV-2 serological markers of high seroprevalence would be beneficial under these circumstances to ensure robust high sensitivity serological testing. Due to limitations involved in the use of other methods applied to characterise SARS-CoV-2 humoral immune responses, the seroprevalence and strength of antibody responses against other SARS-CoV-2 proteins is unclear with no other serological markers of high prevalence having been identified to date (Jiang, Li et al. 2020 , Shrock, Fujimura et al. 2020 ). In addition to being of use as an additional method to screen for the presence of antibodies to N and S, cell-based expression platforms are an ideal system in which to test the antigenicity of other SARS-CoV-2 proteins; SARS-CoV-2 proteins transiently expressed in mammalian cells (Gordon, Hiatt et al. 2020 , Gordon, Jang et al. 2020 can be assayed for reactivity with patient sera in a state likely to closely resemble that encountered by the humoral immune system. Such a system also bypasses the requirement for purified antigens. Given its relative simplicity, we investigated the utility of an immunofluorescence microscopy-based method to detect antibodies against specific SARS-CoV-2 proteins in human serum samples ( Figure 1A and B). To test the effectiveness of this approach, we transfected HEK-293 cells with codon optimised plasmids encoding StrepTagged SARS-CoV-2 N and S ( Figure 1C and F). Based on reports from the previous SARS-CoV pandemic of antibody responses directed towards the SARS-CoV M protein (Wang, Wen et al. 2003 , He, Zhou et al. 2005 , we also investigated whether COVID-19 patients develop IgG responses against SARS-CoV-2 M ( Figure 1I ). Cells transfected with either StrepTagged N or M were fixed, permeabilised and incubated with sera from either pre pandemic individuals or from samples confirmed to have been infected with SARS-CoV-2 by RT-PCR. Cells transfected with a StrepTagged S protein were incubated with sera from the same groups prior to fixation. Transfected cells were identified using StrepTactin labelled with DY-549 and the antigen specific antibodies using anti-human IgG1 secondary antibody conjugated to Alexa488. As the transfection efficiency for the viral proteins was less than 100% we were able to use nontransfected cells as a control for non-specific signal ( Figure 1B) . As an indirect readout of relative . CC-BY-NC-ND 4.0 International license It is made available under a perpetuity. is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. ; https://doi.org/10.1101/2021.10.14.21264873 doi: medRxiv preprint antibody concentrations in the serum samples we assayed, we measured the fluorescence intensity of anti-human IgG-1 Alexa-488 signal associated with both transfected and non-transfected cells and calculated a ratio of transfected to non-transfected 488 fluorescence intensity. Pre-pandemic serum samples had ratios close to one, reflecting equal amounts of signal associated with transfected and non-transfected cells (non-specific background signal). As indicated by ratios of greater than one, antibodies specific to the N and S proteins were only observed in sera from individuals previously infected with SARS-CoV-2 ( Figure 1D and G) . Interestingly, we could also see that many serum samples from COVID-19 patients contained detectable IgG against the SARS-CoV-2 M protein ( Figure 1J ). To see how the results from our microscopy-based assay for N and S compared with those obtained using ELISA for the same set of samples, we normalised the data generated from both methods by expressing values as a percentage of the maximum signal within each dataset. Similar trends could be seen between ELISA absorbance readings and immunofluorescence ratios ( Figure 1E These preliminary results confirm that the combined use of transiently expressed tagged SARS-CoV-2 proteins with immunofluorescence microscopy is a viable approach to detect antibodies against SARS-CoV-2 antigens without the need for purified viral proteins. Whilst screening serum samples by manual imaging and quantification provided proof of principle that immunofluorescence microscopy could be used to detect antibodies to viral antigens, the amount of work required to image and analyse samples manually would make it difficult to efficiently screen large sample numbers. To overcome this, we used a high content microscope in conjunction with automated image analysis software (automated high content immunofluorescence microscopy; AHCIM) ( Figure 2A ). Cells were identified in each captured image using a mask based on DAPI staining of the nuclei ( Figure 2B ). From the nuclear mask a series of additional masks were generated that were used to: 1) determine whether the cells were transfected based on a threshold value of StrepTactin549 signal and 2) measure the amount bound antibody. Applying these masks, the software was able to analyse hundreds of transfected and non- is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. ; https://doi.org/10.1101/2021.10.14.21264873 doi: medRxiv preprint data to the manual quantification. Ratios calculated automatically from images captured by AHCIM were comparable to those generated through manual quantification of the same serum samples (Figure 2E, F and G) . These results demonstrate that the automated image capture and analysis workflow can accurately identify and quantify bound human IgG in COVID-19 patient sera. This methodology also significantly increases the throughput of immunofluorescence-based detection of SARS-CoV-2 antibodies in patient serum samples. Having determined that AHCIM image capture and analysis could accurately quantify bound IgG against SARS-CoV-2 proteins, we applied this system to a larger sample set consisting of 258 samples containing serum from 62 individuals acquired prior to the SARS-CoV-2 pandemic (pre-pandemic negative controls) and 196 patients confirmed to have been infected by the SARS-CoV-2 virus by RT-PCR (serum samples taken 10 to 120 days from positive RT-PCR result). Using our automated platform, we observed a significant difference in signal intensity between the pre-pandemic Based on the IgG ratios calculated for each sample, we constructed receiver operating characteristic (ROC) curves for N and S to assess the overall sensitivity and specificity of our automated immunofluorescence-based platform in detecting N and S IgG ( Figure 3D is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint ELISA inpatient AUC = 0.99 vs AHCIM inpatient AUC = 0.99. ELISA outpatient AUC 0.98 vs AHCIM outpatient AUC 0.99). As described for manual and automated quantification of the original training set 15 differences in antibody titres could also be reliably discerned within this larger collection of serum samples by AHCIM. There was high level of concordance between the strength of ELISA absorbances and AHCIM IgG ratios for both N and S in each serum sample tested (Supplementary Figure 5A & B) . Collectively, these results demonstrate that AHCIM provides high specificity and sensitivity for detection of both SARS-CoV-2 N and S IgG. As a high percentage of COVID-19 patients in our training set had antibodies against the SARS-CoV-2 M protein, we also tested our larger sample set to better establish the prevalence of antibody is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. ; https://doi.org/10.1101/2021.10.14.21264873 doi: medRxiv preprint responses against SARS-CoV-2 M. As predicted from our pilot experiments we observed a significant is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. ; https://doi.org/10.1101/2021.10.14.21264873 doi: medRxiv preprint individually (M sensitivity 84.7%, N sensitivity 92.9%) (Figure 5C-D, Supplementary Table 3) . The combined detection of N and M antibodies therefore provides a boost to the sensitivity of AHCIM based antibody screening with these results suggesting that samples misclassified as negative by AHCIM based on N antibody levels can often be picked up screening for M antibodies and conversely, samples with a lack of detectable M antibodies can be picked up by screening for N. In our earlier analysis, we found that 5.61% of COVID-19 samples were misclassified as negative by N ELISA, owing either to limitations in the sensitivity of the method, or the absence of antibodies against N in these samples. To see if detection of M could complement ELISA based testing for N, we first identified cases in our sample set which were positive by AHCIM for M IgG but negative for N IgG. 5 samples matched these criteria (Figure 5A) . Two of these samples had below threshold While establishing our serological platform we noticed that in a few instances we observed a crossreaction with N when using pre-pandemic sera or samples collected from individuals exposed to seasonal human coronaviruses (H-CoVs) ( Figure 5E) . Interestingly, each of these serum samples contained no detectable antibodies against either SARS-CoV-2 S or M (Figure 5F-I) . Consistent with the results seen by AHCIM, 3 of the pre-pandemic N positive samples also registered high ELISA readings, with additional high value ELISA absorbance values for N from the pre-pandemic serum set also showing high AHCIM ratios ( Figure 5J) . As the negative pre-pandemic sample set was collected prior to the emergence of SARS-CoV-2, these results suggest that the signal we are observing is caused by the presence of antibodies to other coronaviruses. We are not the only group to have observed cross-reactivity of pre-pandemic samples with SARS-CoV-2 N; early studies on . CC-BY-NC-ND 4.0 International license It is made available under a perpetuity. is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. ; https://doi.org/10.1101/2021.10.14.21264873 doi: medRxiv preprint population level infection rates in Hong Kong reported a 1.5% false positive rate in pre-pandemic sera assayed for antibodies against SARS-COV-2 N (To, Cheng et al. 2020) , whilst many commercial platforms see significant overlap in N signal between weak COVID-19 positive serum samples and pre-pandemic negative serum samples that show higher signal relative to the average (National 2020 , Whitcombe, McGregor et al. 2021 . Thus, it is likely that population based serological data based solely on N will over-estimate the number of individual infected with SARS-CoV-2. N antibody levels have been reported to decline more quickly over time in comparison to S antibody levels, raising doubts over the usefulness of N as a long-term indicator of SARS-CoV-2 infection (Long, Tang et al. 2020 , Ripperger, Uhrlaub et al. 2020 , Fenwick, Croxatto et al. 2021 , Whitcombe, McGregor et al. 2021 ). To investigate whether M antibodies may persist for longer than N post infection, we examined N and M antibody levels over time by grouping samples based on the timepoint at which they were taken post PCR test and calculating an average IgG ratio per group. We focussed our analysis to COVID-19 outpatients samples as the inpatient sample values were more variable for both N and M (Figure 6A-D) . As expected, we observed a time dependent increase in IgG levels for both N and M over the first 41 days , Long, Liu et al. 2020 ). However, the levels of N start to decline and reach a level which is approximately 1.7-fold lower than at its peak. In contrast, M antibody levels remained relatively stable throughout the time course and showed only a very minor decrease in signal. . CC-BY-NC-ND 4.0 International license It is made available under a perpetuity. is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. ; https://doi.org/10.1101/2021.10.14.21264873 doi: medRxiv preprint The initial impetus for this paper was to investigate the utility of an immunofluorescence based serological screening platform for detection of N and S antibodies in sera from individuals infected with SARS-CoV-2. The method we used is routinely applied by countless laboratories worldwide for many purposes, involving transfection of plasmid encoded proteins into cells for transient expression, followed by a standard immunofluorescence staining protocol. Thus, if of acceptable sensitivity and specificity, we reasoned other laboratories could apply this methodology for a variety of purposes related to SARS-CoV-2 serology. A high throughput microscopy-based method has been described to assay total SARS-CoV-2 antibody levels (Pape, Remme et al. 2021) , involving infection of cells with SARS-CoV-2 and detection of total bound human IgG following incubation with patient serum. However, this method did not allow antibody responses against individual SARS-CoV-2 proteins to be analysed, and to date no studies have extensively examined antibody responses in human serum samples against individual SARS-CoV-2 proteins using a high throughput immunofluorescence microscopy-based system. In our study, we found the use of automated high content microscopy to be an efficient method of rapidly screening patient sera for antibodies against SARS-CoV-2 N, S and M. The method we employed here is semi-quantitative, with IgG ratios generated from mean fluorescence intensity values per transfected and non-transfected cell. Strong antibody responses were indicated by high IgG ratios and these values were frequently associated with strong absorbance readings obtained for the same samples by ELISA. Using these ratios, we assessed the sensitivity and specificity of AHCIM for detection of N and S IgG. Sensitivity and specificity values for detecting S IgG were equivalent to those seen for absorbances measured by ELISA, whilst our immunofluorescence-based platform had a slightly lower sensitivity than that seen by ELISA for detection of N IgG, likely related to higher background staining of transfected cells in permeabilised samples (Supplementary Figure 4A ). Detection of SARS-CoV-2 N and S antibodies by automated high content immunofluorescence microscopy may thus be a cheaper, faster, alternative to ELISA based methods in the absence of access to the purified proteins required for ELISA. The fast, high throughput, semi-quantitative method we present here is an upgrade on previous immunofluorescence based serological screening methods which have largely involved low throughput qualitative assessment of antibody titres (Peiris, Lai et al. 2003 , Wu, Chiu et al. 2004 , Wolfel, Corman et al. 2020 . is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. ; https://doi.org/10.1101 https://doi.org/10. /2021 Validation of our automated system through detection of N and S gave us the confidence to investigate antibody responses in infected samples against additional SARS-CoV-2 proteins. We chose to focus on the transmembrane protein M, one of four SARS-CoV-2 structural proteins, as studies conducted during the SARS-CoV pandemic identified detectable antibody responses against the SARS-CoV M protein (Wang, Wen et al. 2003 , He, Zhou et al. 2005 . We could clearly detect antibodies against SARS-CoV-2 M in a significant number of COVID-19 patients; 84.7% of COVID samples tested had detectable IgG against M (166 out of 196 PCR validated COVID-19 patients). Consistent with the results presented here, a recent study investigating serum reactivity to the structural proteins of SARS-CoV-2 by flow cytometry also found a high prevalence of M antibodies in COVID-19 positive patient sera, although only approximately 50% of COVID-19 patients tested contained M antibodies (Martin 2021) . The high number of COVID-19 patients with antibodies to M detected in this report and not in other studies (Jiang, Li et al. 2020 , Shrock, Fujimura et al. 2020 is likely related to the state in which viral proteins are assayed for reactivity with patient sera. The M protein is a multi-spanning transmembrane protein proposed to be phosphorylated at its C terminus (Bouhaddou, Memon et al. 2020) . The high amino acid sequence similarity (91% identity) to the membrane glycoprotein M of SARS-CoV suggests SARS-CoV-2 M also has the potential to be glycosylated (Oostra, de Haan et al. 2006) . Both the folded architecture of M and the post translational modifications it is subjected to are likely to be key parts of epitopes recognised by circulating M antibodies. Underestimates of the seroprevalence of SARS-CoV-2 M antibodies from studies using viral peptide or proteins arrays are potentially related to the absence of post-translational modifications and folded epitopes in the versions of proteins and peptides used in these investigations (Hachim, Kavian et al. 2020 , Jiang, Li et al. 2020 , Shrock, Fujimura et al. 2020 , Li, Xu et al. 2021 ). Thus, plasmid-based expression of M in mammalian cells holds several advantages in this respect as binding of antibodies to M is assayed with the protein in its native state. In addition, the StrepTagged-M used in isolation here is Golgi localised and therefore amenable to the same post translation modifications M undergoes in infected cells following its synthesis and transport through the secretory pathway (Oostra, de Haan et al. 2006 , Bouhaddou, Memon et al. 2020 . is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. ; https://doi.org/10.1101 https://doi.org/10. /2021 confer sterilizing immunity have also been proposed and tested (Dangi, Class et al. 2021 ). This strategy would prevent the use of both N and S antibody tests and thus further highlights the potential value of M as an additional serological marker. Currently however, detection of M antibodies would be most powerful when deployed alongside N to confirm prior SARS-CoV-2 infection in individuals vaccinated using S. In support of the use of M for this purpose, we identified cases from our sample set classified as COVID-19 negative based on the lack of detectable N antibodies which could be re-classified as positive through screening for M antibodies by AHCIM ( Figure 5A ). In addition, we also found that 11% of pre-pandemic COVID-19 negative samples contained antibodies that cross-reacted with SARS-CoV-2 N, but not S or M (Figure 5H-I) . Detection of N antibodies in serum collected from both vaccinated and un-vaccinated patients may therefore not indicate prior infection with SARS-CoV-2 in 100% of cases. With these caveats in mind, complementation of ELISA based detection of N antibodies with detection of M antibodies by AHCIM could provide an immediate strategy to enhance the sensitivity and specificity of SARS-CoV-2 serological testing during the era of mass COVID-19 vaccinations. Future work to identify the key immunodominant epitopes of M detected by AHCIM may simplify serological testing using M even further, allowing the production of a SARS-CoV-2 M protein compatible for use with an ELISA format, thereby helping to facilitate widespread serological testing using M. Another key issue in SARS-CoV-2 serology relates to the length of time that each serological marker persists for post infection (Meyer 2021 ). Long lived serological markers are desirable for optimal COVID-19 sero-surveillance as the loss of detectable antibody levels after infection may lead to an underestimation of infection rates and indicate a loss of protective immunity. Whilst detection of sero-conversion due to SARS-CoV-2 infection in those vaccinated is reliant on assaying for N antibodies, the utility of N as a long-lived serological marker is uncertain (Long, Tang et al. 2020 , Ripperger, Uhrlaub et al. 2020 , Fenwick, Croxatto et al. 2021 , Whitcombe, McGregor et al. 2021 , although recent data suggests that N antibodies levels remain high for longer than previously believed (Shrotri, Harris et al. 2021) . Elucidating the kinetics of M antibody levels over time may help to establish whether M may be a more durable serological marker of SARS-CoV-2 infection than N. A preliminary analysis of N and M IgG ratios plotted against days post PCR test when each serum sample was collected showed a trend in N, but not M, for antibody levels to decline over time, however the number of samples used in this analysis at crucial later timepoints (71-120 days) was low (Figure 6) . While this data is suggestive of greater temporal stability in M antibody levels, is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. ; https://doi.org/10.1101/2021.10.14.21264873 doi: medRxiv preprint longitudinal profiling of N and M antibody titres in individual COVID-19 serum samples would likely give a clearer answer as to the utility of M as a long-lived serological marker in comparison to N. Our observations regarding the seroprevalence of M IgG also raise several other important questions which warrant further investigation. Correlations between the presence of M antibodies and disease outcomes and whether antibodies against M may serve to provide and bolster the neutralising activity of humoral immune responses generated against SARS-CoV-2 are currently unknown and are clearly important questions to address. Of note, M IgG ratios were significantly higher in inpatient samples analysed in this study ( Figure 3J ) suggesting higher M antibody levels correlate with more severe COVID-19 cases, as previously reported for N and S (Roltgen, Powell et al. 2020 , Batra, Tian et al. 2021 , Yates, Ehrbar et al. 2021 ) and seen again here in this study ( Figure 3B & F). The synergistic use of proteome-wide tagging of viral proteins with automated high content immunofluorescence microscopy could provide a rapid standardised format to quickly identify the most antigenic proteins of newly emergent pathogens. This would be of value for both vaccine development and guiding which antigens to use as markers of prior infection in diagnostic methods such as ELISA or AHCIM. One could in theory quickly progress from obtaining the complete genome of a previously uncharacterised pathogen, to generating a codon optimised tagged library of proteins for expression in mammalian cells, to identifying the strongest antigenic targets across the whole proteome using an automated immunofluorescence microscopy-based platform like the one described here. The identification of M as a key target of the humoral immune response to SARS-CoV-2 provides a first-hand example of the advantages of such an approach. . CC-BY-NC-ND 4.0 International license It is made available under a perpetuity. is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. ; https://doi.org/10.1101 https://doi.org/10. /2021 amount overlap between positive and negative IgG ratios, consistent with AHCIM based detection of S giving the highest specificity and sensitivity values. ROC curves comparing the sensitivity and specificity of IF and ELISA based detection of (D) nucleocapsid and (H) spike IgG in COVID-19 positive and pre-pandemic sera. ****P<0.0001, **P<0.01 (P=0.0030) (Unpaired t-test). No cross reactivity with SARS-CoV-2 spike or M proteins were observed. However, one sample H3 (highlighted with an asterix) had IgG signal to N which was above the threshold. Inset in (E) shows representative images for this sample. To explore whether other pre-pandemic samples also showed a similar cross-reactivity we plotted the IgG ratios for N against S (H) and M (I). Samples highlighted in blue represent sera that is positive for N but negative for either SARS-CoV-2 S or M suggesting cross-reactivity to a seasonal coronavirus. (J) To examine whether high ELISA values were also seen in the same pre-pandemic samples with seasonal coronavirus cross reactivity identified by AHCIM, IgG ratios and ELISA absorbances from these samples were plotted against one another. Dashed lines between data points link IgG ratios and ELISA absorbances from individual samples. Red dashed lines indicate samples with above threshold N signal identified by AHCIM. Half-length black dashed lines in the ELISA or AHCIM columns indicate the respective thresholds for classification of samples as COVID-19 positive or negative by ELISA or AHCIM. is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. ; https://doi.org/10.1101/2021.10.14.21264873 doi: medRxiv preprint Representative images of StrepTagged SARS-CoV-2 nucleocapsid transfected cells imaged at different ratios of transfected to non-transfected cells. Serum samples with strong (A09), intermediate (C01) and weak (C08) SARS-CoV-2 nucleocapsid IgG responses were selected. Transfected cells plated at 100% confluency were split at the indicated dilutions, fixed, incubated with patient sera and processed for immunofluorescence as described. (B) Automated quantification of 488 intensity ratios for COVID19 serum samples A09, C01 and C08 and the COVID19 negative serum sample D06. Decreasing cell density increases the IgG ratio for strong, intermediate and weak COVID-19 positive samples. is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. ; https://doi.org/10.1101/2021.10.14.21264873 doi: medRxiv preprint is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. ; https://doi.org/10.1101/2021.10.14.21264873 doi: medRxiv preprint is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. ; https://doi.org/10.1101/2021.10.14.21264873 doi: medRxiv preprint is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. ; https://doi.org/10.1101/2021.10.14.21264873 doi: medRxiv preprint is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. ; https://doi.org/10.1101/2021.10.14.21264873 doi: medRxiv preprint is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. ; https://doi.org/10.1101/2021.10.14.21264873 doi: medRxiv preprint is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. ; https://doi.org/10.1101/2021.10.14.21264873 doi: medRxiv preprint is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. ; https://doi.org/10.1101/2021.10.14.21264873 doi: medRxiv preprint is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. ; https://doi.org/10.1101/2021.10.14.21264873 doi: medRxiv preprint is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. ; https://doi.org/10.1101/2021.10.14.21264873 doi: medRxiv preprint is the author/funder, who has granted medRxiv a license to display the preprint in (which was not certified by peer review) preprint The copyright holder for this this version posted October 18, 2021. ; https://doi.org/10.1101/2021.10.14.21264873 doi: medRxiv preprint A serological assay to detect SARS-CoV-2 seroconversion in humans Role of IgG against N-protein of SARS-CoV2 in COVID19 clinical outcomes The Global Phosphorylation Landscape of SARS-CoV-2 Infection Immunofluorescence assay for serologic diagnosis of SARS Combining spike-and nucleocapsid-based vaccines improves distal control of SARS-CoV-2 Changes in SARS-CoV-2 Spike versus Nucleoprotein Antibody Responses Impact the Estimates of Infections in Population-Based Seroprevalence Studies Comparative host-coronavirus protein interaction networks reveal pan-viral disease mechanisms A SARS-CoV-2 protein interaction map reveals targets for drug repurposing A comparison of four serological assays for detecting anti-SARS-CoV-2 antibodies in human serum samples from different populations ORF8 and ORF3b antibodies are accurate serological markers of early and late SARS-CoV-2 infection A simple, sensitive, and low-cost FACS assay for detecting antibodies against the native SARS-CoV-2 spike protein Identification of immunodominant epitopes on the membrane protein of the severe acute respiratory syndrome-associated coronavirus Risk factors for SARS-CoV-2 seroprevalence following the first pandemic wave in UK healthcare workers in a large NHS Foundation Trust SARS-CoV-2 proteome microarray for global profiling of COVID-19 specific IgG and IgM responses Production of trimeric SARS-CoV-2 spike protein by CHO cells for serological COVID-19 testing Serology assays to manage COVID-19 Epitope-resolved profiling of the SARS-CoV-2 antibody response identifies cross-reactivity with endemic human coronaviruses Dynamic changes in anti-SARS-CoV-2 antibodies during SARS-CoV-2 infection and recovery from COVID-19 Antibody landscape against SARS-CoV-2 reveals significant differences between nonstructural/accessory and structural proteins Clinical and immunological assessment of asymptomatic SARS-CoV-2 infections SARS-CoV-2 envelop proteins reshape the serological responses of COVID-19 patients Waning antibodies to SARS-CoV-2 -Don't panic Serological assays for emerging coronaviruses: challenges and pitfalls Performance characteristics of five immunoassays for SARS-CoV-2: a headto-head benchmark comparison Glycosylation of the severe acute respiratory syndrome coronavirus triple-spanning membrane proteins 3a and M Microscopy-based assay for semi-quantitative detection of SARS-CoV-2 specific antibodies in human sera: A semi-quantitative, high throughput, microscopy-based assay expands existing approaches to measure SARS-CoV-2 specific antibody levels in human sera Coronavirus as a possible cause of severe acute respiratory syndrome Orthogonal SARS-CoV-2 Serological Assays Enable Surveillance of Low-Prevalence Communities and Reveal Durable Humoral Immunity Defining the features and duration of antibody responses to SARS-CoV-2 infection associated with disease severity and outcome Fiji: an open-source platform for biological-image analysis Persistence of SARS-CoV-2 N-Antibody Response in Healthcare Workers Epitope profiling reveals binding signatures of SARS-CoV-2 immune response in natural infection and cross-reactivity with endemic human CoVs Seroprevalence of SARS-CoV-2 in Hong Kong and in residents evacuated from Hubei province, China: a multicohort study Assessment of immunoreactive synthetic peptides from the structural proteins of severe acute respiratory syndrome coronavirus Comprehensive analysis of SARS-CoV-2 antibody dynamics in New Zealand Virological assessment of hospitalized patients with COVID-2019 A new coronavirus associated with human respiratory disease in China Serologic and molecular biologic methods for SARS-associated coronavirus infection Serological analysis reveals an imbalanced IgG subclass composition associated with COVID-19 disease severity