key: cord-0287999-sn6u81ln authors: Ashe, W. B.; Innis, S.; Shanno, J.; Hochheimer, C.; Williams, R. D.; Ratcliffe, S. J.; Moorman, R.; Gadrey, S. M. title: Analysis of Respiratory Kinematics: a method to characterize breaths from motion signals date: 2021-09-12 journal: nan DOI: 10.1101/2021.09.06.21263179 sha: d278e0514be39ce5e3542efcfa09bb243f3c979e doc_id: 287999 cord_uid: sn6u81ln Rationale: Breathing motion (respiratory kinematics) can be characterized by the interval and depth of each breath, and by magnitude-synchrony relationships between locations. Such characteristics and their breath-by-breath variability might be useful indicators of respiratory health. Objectives: To enable breath-by-breath characterization of respiratory kinematics, we developed a method to detect breaths using motion sensor signals. Methods: In 34 volunteers who underwent maximal exercise testing, we used 8 motion sensors to record upper rib, lower rib and abdominal kinematics at 3 exercise stages (rest, lactate threshold and exhaustion). We recorded volumetric air flow signals using clinical exercise laboratory equipment and synchronized them with kinematic signals. Using instantaneous phase landmarks from the analytic representation of kinematic and flow signals, we identified individual breaths and derived respiratory rate signals at 1Hz. To evaluate the fidelity of kinematics-derived respiratory rate signals, we calculated their cross-correlation with the flow-derived respiratory rate signals. To identify coupling between kinematics and flow, we calculated the Shannon entropy of the relative frequency with which kinematic phase landmarks were distributed over the phase of the flow cycle. Measurements and Main Results: We found good agreement in the kinematics-derived and flow-derived respiratory rate signals, with cross-correlation coefficients as high as 0.94. In some individuals, the kinematics and flow were significantly coupled (Shannon entropy < 2) but the relationship varied within (by exercise stage) and between individuals. The final result was that the phase landmarks from the kinematic signal were uniformly distributed over the phase of the air flow signals (Shannon entropy close to the theoretical maximum of 3.32). Conclusions: The Analysis of Respiratory Kinematics method can yield highly resolved respiratory rate signals by separating individual breaths. This method will facilitate characterization of clinically significant breathing motion patterns on a breath-by-breath basis. The relationship between respiratory kinematics and flow is much more complex than expected, varying between and within individuals. A clinician's bedside physical examination is never complete without careful characterization of the breathing motion (respiratory kinematics) (1) (2) (3) . Features of interest include breath intervals, overall depth of breathing and the magnitude-synchrony relationships between key anatomical locations (4) . They contain information about the subject's health which can be useful in wide range of respiratory and non-respiratory illnesses. Tachypnea (short breath interval) and hyperventilation (increased depth of breaths) in the setting of infection can herald the onset of sepsis (5) . Abdominal paradox (inward abdominal motion, asynchronous with rib cage expansion), in the setting of pneumonia, suggests inspiratory muscle overload and is a risk marker for imminent acute respiratory failure (4, 6) . In many clinical scenarios, breath-to-breath variability of motion patterns is also an important feature. Respiratory alternans is a sign of inspiratory muscle overload where ribcagepredominant breaths alternate irregularly with abdomen-predominant ones (4, 6) . Tidal volume variability (shallow breaths alternating with deep ones) precedes hypoventilation (consistently shallow breaths) in opiate-induced respiratory depression and can provide early warnings of an overdose (7) (8) (9) . Despite well-established clinical significance, quantitative characterization of respiratory kinematics has proven elusive. Some methods use chest impedance or electrocardiograms to characterize respiratory rates and rhythms (10) (11) (12) . These can be deployed in intensive care and other telemetry enabled units. Other methods like pneumography, inductance plethysmography, optoelectronic plethysmography, and magnetometry involve obtrusive hardware and/or cumbersome calibration protocols (4, 13) . They remain limited largely to sleep, exercise and physiology laboratories. In most clinical settings (home, clinics, ambulances and low acuity wards), an experienced clinician's visual examination remains the sole method to characterize breathing motion patterns. We have developed the Analysis of Respiratory Kinematics (ARK) system for quantitative characterization of breathing motion in an unobtrusive manner in any clinical setting. Signals are recorded using a network of microelectromechanical system inertial measurement units (MEMS-IMUs) placed on key anatomical landmarks of the chest and abdomen. To pave the way for characterization on a breath-by-breath basis, a method was needed to accurately detect individual breaths using ARK system signals. The aim of this study was to develop a reliable and reproducible method for breath separation that is based on fundamentally sound analytic principles. We recruited 34 healthy volunteers for maximal oxygen consumption (VO2-Max) testing on a bicycle ergometer in the University of Virginia's Exercise Physiology Core Laboratory. We placed 8 inertial sensors on key anatomical locations of the chest and abdomen (Figure 1a ). Each motion sensor captured 9 streams of signals (3 orthogonal axes each from accelerometers, rate gyroscopes, and magnetometers). To identify breath intervals in this study, we used the linear acceleration signal that was normal to the body surface. We will refer to this signal as the accelerometer signal. To study the relationship of kinematic signals with the air flow cycle, we recorded volumetric flow rate ("flow") using the laboratory's standard clinical equipment (Vyaire V-Mask 29-CE; Hans Rudolph metabolic masks). We used the Rated Perceived Exertion (RPE) scale to monitor the level of exertion (14) , with lactate threshold defined as RPE of 13 (15) and exhaustion as RPE of 20 and/or failure to maintain a pedaling cadence of 60 revolutions per minute. We obtained 2-minute recordings of all signals at rest (pre-exercise), lactate threshold, and exhaustion ( Figure 1b) . During recording periods, volunteers assumed a supine position with 30 o head elevation. Each of the eight sensors was connected via USB to a host application running on a computer. Data collection on the sensors was stagger-started over eight consecutive ticks of a common 100Hz timer on the host machine. Once started, each sensor recorded signals at a sampling frequency of 100Hz based on a local timer on the sensor. Data were realigned to a common time frame using the initial host clock start time and the stagger order of the devices. Due to minor inconsistencies between local timers on the sensors, the data were interpolated to identical sampling times across all sensors. For interpolation, data were upsampled to a common 10μS clock by duplicating samples and filtering, then decimated to fall on hundredthsecond boundaries. We filtered the flow and accelerometer signals to preserve the respiratory content and to reduce the amplitude of the non-respiratory components. We used a Butterworth bandpass filter (fourth order low pass and sixth order high pass; zero-phase non-causal) with corner frequencies of 0.05Hz and 1Hz. We based this choice on the fact that the frequency of human respirations can reasonably be expected to range between the corresponding rates of 3 and 60 breaths per minute in most circumstances, including after maximal exercise. Since both flow and accelerometer measured the effects of the same underlying cyclical phenomenon -the subject's respiration -we used the same filter specifications for both signals. of the wave monotonically increases to match the progression of the complex signal around the origin. For quasi-cyclic processes with repetitive but not strictly periodic behavior, this instantaneous phase angle can be used to reliably detect consistent landmarks, where a given phase closely tracks the same point on the original signal (e.g. a specific peak or zero-crossing) across different cycles. We used this property to identify breath intervals on flow and accelerometer signals as the intervals between successive occurrences of a particular phase angle ( Figure 2 ). We derived respiratory rate signals from accelerometer and flow signals by adapting the method that was originally proposed by Berger and coworkers to derive heart rate signals from electrocardiograms (17) . We used intervals between instantaneous phase landmarks on the accelerometer and flow signals in place of R-R intervals on electrocardiograms ( Figure 4a ). In this study, we chose the phase angle of π radians as our landmark for both flow and accelerometer signals, and we resampled the respiratory rate at 1Hz. To evaluate the fidelity of accelerometer-derived breath intervals, we calculated cross-correlation between accelerometer-derived and flow-derived respiratory rate time series. We calculated the relative frequency with which kinematic phase landmarks were distributed over the phase of the air flow cycle. We split the air flow cycle into 10 bins (bin width of 5 radians). Similar methods have been used in the past to create cardiorespiratory synchrograms (18) . To quantify synchronization, we calculated the Shannon Entropy of the histograms: Here ( ) is the probability (relative frequency) of a kinematic phase landmark occurring in the j th bin of air flow phase. The utility of this measure in quantifying cardiopulmonary coupling has been described (19) . In our application, the Shannon entropy could range between a maximum of 3.32 bits (log210) signifying uniform distribution across 10 bins, and a minimum of 0 bits signifying localization to a single bin (Figure 4b ). We used MATLAB version 2020b for signal processing (20) . We used REDCap electronic data capture tools hosted at the University of Virginia (21, 22) . Of the 34 volunteers, five had no usable data and seven had only partial data. The final sample consisted of 22 healthy volunteers. The reasons for missing data were related to unanticipated hardware and software malfunction in our locally produced prototype. Our subjects, a third of whom were women, had a mean age of 47 (range: 21-71), a mean BMI of 25 (range: 21-33) and a mean VO2-max of 42 ml/kg/min (range: 15-60). We found a strong relationship between respiratory rate time series derived from accelerometer and flow signals. In light of the variability we observed in the kinematic-flow cycle synchronicity, we quantified the degree of agreement using the maximal cross-correlation value without regard for lag. We observed cross-correlation coefficients as high as 0.94. Of note, the strength of this relationship was preserved despite trends, cyclical fluctuations or transient disturbances in true respiratory rate ( Figure 5 ). When analyzed across individuals and exercise states, landmarks from the phase of kinematic signals were uniformly distributed across the phase of the air flow cycle ( Figure 6 ). The average Shannon entropy across the lower rib and abdominal sensors was 3.31, which is comparable to the entropy limit of 3.32 for a completely uniform distribution. . The clinical significance of breathing motion patterns is well established, but there are no unobtrusive methods to quantitatively characterize such patterns at the bedside. Clinicians must rely on visual inspection to assess breathing motion, and qualitatively describe their findings in notes. This is problematic for several reasons. First, there is considerable inaccuracy in manual recordings of even the simplest kinematic feature, the respiratory rate (23, 24) . For more complex features, even highly trained physicians achieved only moderate inter-rater reliability (25) . Less experienced clinicians may simply miss red-flag signs or fail to recognize the risk they portend (26) . Second, attempts to standardize terminology in narrative descriptions have not achieved wide adoption (27) . Such an approach would require large scale continuous training of staff at every level of care. Finally, only a small proportion of variance in respiratory kinematic features is explained by conventionally monitored vital signs (28) . Timely detection of respiratory kinematic aberrations, therefore, depends on frequent bedside assessments by experienced clinicians. This is not possible in most settings due to resource and/or staffing constraints. Patient isolation requirements in illnesses like COVID-19 further reduce the frequency of bedside assessment (29) (30) (31) . Visualizations like the ones in Figure 3 or 6 can facilitate in-depth inspection of kinematic features by clinicians. They can be reviewed remotely and stored as a reference to ascertain the trajectory of respiratory kinematics over time. Quantitative characterization of the signals will yield novel features which may prove useful in predictive modelling of respiratory failure. We foresee considerable clinical utility in such remote, quantitative assessments of respiratory effort. We found that the relationship between kinematic and air flow cycles was rich with complexity -varying between and within individuals. For practical breath-detection applications, this finding has an important implication -no landmark point is superior to any other in terms of optimizing alignment between the kinematic and air flow cycles. We selected the phase angle of π radians to identify each nadir in the accelerometer signals. Any other landmark (like the phase angle of 0 radians to identify each peak of the accelerometer signal) might be selected. We demonstrated that the ARK signals can yield highly resolved respiratory rate time series from kinematic signals. This finding paves way for two types of analyses in the future. First, a large number of well-established mathematical operations can be used to characterize respiratory rate time series in any clinical setting (32) . The regular sampling on a real-time axis also allows for meaningful analysis in the frequency domain. Second, well-defined breath intervals will allow breath-by-breath characterization of signal features, i.e., comparing magnitude-synchrony relationships between different sensor locations within a single kinematic cycle. This is important to quantify patterns like respiratory alternans or opiate-induced ataxic breathing, where the breathing pattern varies from one breath to the next. Our future work will build on this breath detection method to describe the clinical significance of a large and diverse set of metrics. The overarching vision is to create a set of novel respiratory vital signs that can be conveniently measured in any setting and improve medical decision making in common clinical scenarios. A strength of this work is that it is grounded in fundamentally sound analytic methods like the Hilbert transform, the analytic representation of a signal, Shannon entropy, and Berger's interpolation algorithm. We do not rely on any arbitrary assumptions, decisions, thresholds or models. This enhances the reliability, reproducibility of our method and its generalizability to other signals that record respiratory motion (impedance, inductance plethysmography etc.). Another strength is the unique data set. To our knowledge, this is the 13 first instance where multi-focal kinematic signals are compared to synchronously-recorded volumetric air flow signals in varying states of respiratory physiology. This is, therefore, among the first descriptions of the enormous variability encountered in the flow-kinematics relationship. The small size and inclusion of only healthy individuals are limitations and further development in larger and more clinically diverse samples will be needed. The results at this stage serve as demonstration of feasibility of our breath detection method and lays the groundwork for detailed breath characterization in the future. We describe a reliable and reproducible method to detect individual breath cycles from respiratory kinematic signals. Despite a complex relationship between respiratory kinematics and air flow, our method resulted in highly resolved respiratory rate time series. This will facilitate quantitative characterization of clinically significant breathing motion patterns in any care setting. which varies between and within individuals. Panel A shows findings from an individual whose kinematics were strongly coupled with flow within a stage. In each stage, however, the kinematic phase landmarks localized to different areas of the air flow phase. As 28 a result, the overall Shannon entropy for that individual was higher (2.83) than their stage-specific Shannon Entropy (1.87, 1.55, 1.24). Panel B shows findings from an individual whose kinematics weakly coupled with flow even within a stage. Across individuals and exercises stages, therefore, the overall Shannon entropy (3.31) was very close to the maximal possible value of 3.32 (panel C). This figure displays results obtained from a lower rib sensor. Similar results were noted at all sensor locations. [PIF: Peak Inspiratory Flow; OE: Onset of Exhalation; PEF: Peak Expiratory Flow; OI: Onset of Inhalation; φ = Phase angle] Breathing patterns. 1. Normal subjects Breathing patterns. 2. Diseased subjects Respiratory Rate and Pattern ATS/ERS Statement on respiratory muscle testing Bench-to-bedside review: ventilatory abnormalities in sepsis Clinical manifestations of inspiratory muscle fatigue Opioid-induced respiratory depression is associated with increased tidal volume variability Chronic Opioid Use is a Risk Factor for the Development of Central Sleep Apnea and Ataxic Breathing An Automated Algorithm Incorporating Poincaré Analysis Can Quantify the Severity of Opioid-Induced Ataxic Breathing Recent development of respiratory rate measurement technologies Quantification of periodic breathing in premature infants ECG Derived Cheyne-Stokes Respiration and Periodic Breathing are Associated with Cardiorespiratory Arrest in Intensive Care Unit Patients ERS statement on respiratory muscle testing at rest and during exercise Perceived exertion as an indicator of somatic stress Perceptual responses and blood lactate concentration: effect of training state Phase synchronization of chaotic oscillators An efficient algorithm for spectral analysis of heart rate variability Heartbeat synchronized with ventilation Breath-by-breath analysis of cardiorespiratory interaction for quantifying developmental maturity in premature infants Research electronic data capture (REDCap)--a metadata-driven methodology and workflow process for providing translational research informatics support The REDCap consortium: Building an international community of software platform partners The accuracy of respiratory rate assessment by doctors in a London teaching hospital: a cross-sectional study Accuracy and interobserveragreement of respiratory rate measurements by healthcare professionals, and its effect on the outcomes of clinical prediction/diagnostic rules The validity and reliability of the clinical assessment of increased work of breathing in acutely ill patients Inadequacies of Physical Examination as a Cause of Medical Errors and Adverse Events: A Collection of Vignettes DiapHRaGM: A mnemonic to describe the work of breathing in patients with respiratory failure Association Between Rating of Respiratory Distress and Vital Signs, Severity of Illness, Intubation, and Mortality in Acutely Ill Subjects. Respiratory Care Telehealth in the COVID-19 Era: A Balancing Act to Avoid Harm Will the Traditional Physical Examination Be Another Casualty of COVID-19? The Disappearance of the Primary Care Physical Examination-Losing Touch Highly comparative time-series analysis: the empirical structure of time series and their methods We thank David Chen, director of UVA's Coulter Translational Research Partnership, for his role in establishing and supporting this multidisciplinary collaboration. We thank Sharon Krueger, program director of the Ivy Foundation for her advice and feedback. We thank UVA's Division of General, Geriatric, Palliative and Hospital Medicine, UVA's Center for Engineering in Medicine and the Ivy Foundation for the seed funding which made this project possible. We thank Dr.Arthur L. Weltman, Director of UVA's Exercise Physiology Core Laboratory (EPCL), for generously lending his domain expertise to optimize our study design. We thank EPCL's manager, Lisa Farr, and her entire team (Olivia Hazelrigg, Nicole Huebner, Courtney Paxton and Jackie Dosik) for their enthusiastic support of our work. We thank Melissa Goldman, Fabrication Lab manager at the UVA School of Architecture, and Trevor Kemp, assistant manager, for their advice and assistance throughout prototype design and construction. None