key: cord-0268920-25fb3iom authors: Lelie, N.; Koppelman, M.; Van Drimmelen, H.; Bruisten, S. title: Analytical sensitivity and effectiveness of different SARS-CoV-2 testing options date: 2021-11-28 journal: nan DOI: 10.1101/2021.11.26.21265946 sha: 3daa96357838bbcb0414cfc24fb279fffa9356b0 doc_id: 268920 cord_uid: 25fb3iom We prepared severe acute respiratory syndrome-coronavirus-2 (SARS-CoV-2) working standards and reference panels from a pool of swab fluid samples before and after inactivation by beta-propiolactone and quantified viral load in nucleic acid amplification technology (NAT) detectable RNA copies/mL using limiting dilution analysis. The following 50% lower limits of detection (LOD) were estimated by probit analysis as compared to detection limits of rapid antigen tests on 1.5 fold dilutions of the native material: Roche cobas PCR 1.8 (1.0-3.3), Hologic Aptima TMA 6.6 (4.4-9.9), DRW SAMBA 15 (7-30), Molgen LAMP 23 (13-42), Fluorecare antigen 50,000, Abbott Panbio antigen 75,000 and Roche antigen 100,000 copies/mL. One 50% Tissue Culture Infectious Dose (TCID50)/mL of culture fluid was estimated to be equivalent to approximately 1000 RNA copies/mL (2700- 4300 International Units) in our working standard. When assuming this level as start of contagiousness in a log-linear ramp up viremia model with 10-fold rise of viral load per day for the B.1 (Wuhan) type we estimated relative time points of first detectability of early infection by the different SARS-CoV-2 assays from the LODs mentioned above. The four NAT assays would be able to detect early viremia 40-66 hours earlier than the 1000 copies/mL infectivity threshold, whereas the three antigen tests would become positive 41-48 hours later. Our modeling of analytical sensitivity data was found to be compatible with clinical sensitivity data of rapid antigen tests and confirms that NAT assays are more reliable than antigen assays for identifying early infected asymptomatic individuals who are potentially infectious. Several studies compared the clinical sensitivity of molecular assays and rapid antigen tests on nasopharyngeal or nasal swab samples for detection of severe acute respiratory syndrome-coronavirus 2 (SARS-CoV-2) infection [1] [2] [3] [4] [5] [6] . The limitation of these comparison studies is that separate swabs must be taken for detection of SARS-CoV-2 RNA and antigen respectively, whereas these individuals were usually tested because of (mild) symptoms and already had higher viral loads. Only limited data is available from studies that directly compared the analytical sensitivity of different nucleic acid amplification technology (NAT) systems and lateral flow devices for rapid detection of SARS-CoV-2 antigen using serial dilutions of swab fluid with a known viral load [6] [7] [8] . In this report we used reference preparations from pooled swab samples in Viral Transport Medium (VTM) before and after inactivation by betapropiolactone 9,10 for comparison of the analytical sensitivity of different SARS-CoV-2 assays. These working standards were first quantified in NAT detectable SARS-CoV-2 RNA copies/mL using limiting dilution analysis and later -at the time the World Health Organization (WHO) International Standard became available -also in International Units (IU/mL). Additionally, by comparison of NAT detection limits on USA-WA1-2020 culture fluid 11 the amount of NAT detectable RNA copies per 50% Tissue Culture Infectious Dose (TCID 50 ) was estimated. Since the rapid antigen tests are known to miss a considerable proportion of infected individuals 1,2 we also evaluated two NAT methods that were developed for fast detection of SARS-CoV-2 infection i.e. a loop-mediated isothermal amplification (LAMP) assay and a so called Simple amplification based assay (SAMBA) 12 and compared their sensitivity with two widely used NAT systems as reference methods i.e. a real time polymerase chain reaction (PCR) assay 11 and a transcription . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) preprint The copyright holder for this this version posted November 28, 2021. ; https://doi.org/10.1101/2021.11.26.21265946 doi: medRxiv preprint Version November 25 2021 4 mediated amplification (TMA) assay 13 . By testing dilution series of our native and inactivated working standards the lower limits of detection (LOD) of these NAT systems were compared with those of three rapid antigen assays. We then used the analytical sensitivity data, including the impact of testing minipools of six samples (MP6) using the PCR assay 11 , to evaluate the effectiveness of these different SARS-CoV-2 detection options in identifying infected individuals during the early asymptomatic phase of infection. For this evaluation a log-linear ramp up viremia model was assumed using best estimates for the viral doubling time and start of potential contagiousness 14, 15 . . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) preprint Ten-fold standard dilutions of both the native and inactivated SARS-CoV-2 standards were made in 2% plasma-PBS and tested in duplicate in the cobas PCR assay (Roche Molecular Systems) to compare the SARS CoV-2 RNA Ct values for both the ORFa/b and E gene targets. The yield after inactivation was determined by parallel line analysis on Ct values for the two PCR targets separately. A panel composed of 3 and 10-fold dilutions of the inactivated standard was tested in multiple replicate tests in different NAT methods to allow for estimation of the limit of detection (LOD) by probit analysis. The concentration in NAT detectable RNA copies/mL that was assigned to the inactivated standard was based on the 63% LOD in the most sensitive NAT method. We assumed that this assay reached 100% NAT efficiency or a 63% LOD of 1 detectable RNA copy per amplification reaction as follows from Poisson distribution in limiting dilution analysis. The concentration in RNA copies/mL in the inactivated working standard was then calculated by correcting for the input volume in the NAT method. In addition, 3 and 10-fold dilutions of the inactivated standard were tested in duplicate against similar concentrations of the WHO International Standard 20/146 in the cobas PCR assay. The conversion factor between NAT detectable RNA copies and International Units (IUs) was calculated by parallel line analysis on Ct values for both ORFa/b and E gene targets in the cobas assay. The concentration in the native standard was derived from the value assigned to the inactivated standard and the measured yield after inactivation, whereby the average was taken for the two PCR targets in the cobas assay. The yield of SARS-CoV-2 antigen after inactivation was determined by comparison of geometric mean . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) preprint The copyright holder for this this version posted November 28, 2021. ; https://doi.org/10.1101/2021.11.26.21265946 doi: medRxiv preprint detection endpoint titers on 1.5-fold dilutions of both the native and inactivated standard in three rapid antigen tests. We compared the 95% LOD expressed in TCID 50 /mL on USA-WA1-2020 culture fluid reported in the package insert of the Roche cobas assay 11 with the 95% LOD expressed in NAT detectable RNA copies/mL assigned to our inactivated working standard to estimate the dose of SARS-CoV-2 RNA copies per TCID 50 . To determine the 95% and 50% LOD by probit analysis the inactivated SARS-CoV-2 standard dilution panel with estimated viral RNA concentrations varying between 33,784 and 1.1 copies/mL was tested in multiple replicates by the following NAT assays: Roche cobas PCR 11 , Hologic Aptima TMA 13 is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) preprint The copyright holder for this this version posted November 28, 2021. ; https://doi.org/10.1101/2021.11.26.21265946 doi: medRxiv preprint visible line were scored as + whereas the next panel member(s) showing a faint line was recorded as ± or indeterminate. The viral concentration in the 1:1 mixed sample buffer with the last + and ± reactivity were recorded as endpoint titers, whereas the input copy numbers in the volumes added to the reaction holes were recorded as well. If the duplicate reactions were discrepant (+ and ± or ± and -) the series with the lowest endpoint titers were taken as final positive (+) and indeterminate (±) result. The 95%, 63% and 50% LODs on the BioQ SARS-CoV-2 analytical sensitivity panel were calculated by probit analysis and relative sensitivities of assays by parallel line probit analysis using SPSS software. The potency of the inactivated SARS-CoV-2 standards against the native standard and the WHO International Standard 20/146 was calculated from the Ct values in the cobas assay using parallel line analysis. . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) preprint The copyright holder for this this version posted November 28, 2021. ; https://doi.org/10.1101/2021.11.26.21265946 doi: medRxiv preprint Version November 25 2021 9 The 50% LOD NAT conversion point of the most sensitive assay was arbitrarily set at time point zero and the 50% LOD NAT conversion points found with (the most sensitive targets of) the other assays, including a MP6 PCR option 11 , were calculated using the assumptions for the viral growth curves mentioned above. The rapid antigen detection limits with a faint reaction line (±) on the 1.5 fold native standard dilutions were used to calculate the time intervals between the 50% NAT conversion points and the rapid antigen test conversion points. . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) preprint The copyright holder for this this version posted November 28, 2021. ; https://doi.org/10.1101/2021.11.26.21265946 doi: medRxiv preprint 1 0 Three 10-fold dilutions of SARS-CoV-2 standards before and after inactivation were tested in duplicate in the cobas SARS-CoV-2 assay and Ct values for the two PCR gene targets are shown in Supplemental From the proportions of reactive tests in three different NAT assays on inactivated standard dilutions the 63% LOD (and 95% confidence interval (CI)) was calculated by parallel line probit analysis and the results are shown in Table 1 . We assumed that the NAT method with the highest sensitivity (the cobas assay for ORFa/b target) reached 100% NAT efficiency or a 63% LOD of 1 NAT detectable RNA copy per amplification reaction as follows from Poisson distribution statistics. Since for each replicate cobas PCR test a volume of 400 uL was used as input in the amplification reaction the concentration at the 63% LOD was thus set at 2.5 copies/mL for the cobas PCR ORFa/b assay. Therefore initially assigned concentrations to the standard dilution panel (data not shown) needed to be adjusted with a recalibration . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) preprint The copyright holder for this this version posted November 28, 2021. ; https://doi.org/10.1101/2021.11.26.21265946 doi: medRxiv preprint 1 1 factor and from the dilution factors we calculated that the viral load in the undiluted inactivated standard was 3.38 x 10 6 NAT detectable RNA copies/mL and 4.96-fold higher (see above) in the native standard (1.68 x 10 7 copies/mL). With these assigned RNA copies/mL to the SARS-CoV-2 standards the calculated NAT efficiency was 100 (53-187)% and 54 (24-94)% in the cobas ORFa/b and E PCR tests respectively, whereas it would be 67 (42-102)% in the Aptima TMA assay and 40 (41-73)% in the LAMP assay (Table 1) . When comparing the dilutional titers in three rapid antigen tests on the 1.5-fold dilution panel we estimated that the SARS-CoV-2 antigen concentration was 1.8 (range 1.3-2.5) fold lower in the beta-propiolactone-inactivated standard than the untreated standard, whereas the RNA concentration was 5.0 (3.9-6.3)-fold lower after inactivation. Hence, the antigenicity relative to the RNA concentration was found to be 2.7 (range 2.5-3.0) fold higher in the inactivated standard than in the native standard. For calibration of the inactivated SARS-CoV-2 standard against the WHO 20/146 standard we tested dilution series in the cobas SARS-CoV-2 assay in duplicate (Supplemental Table 2a ) and analyzed the Ct values using parallel line analysis with exception of the lowest concentration to improve parallelism. The potency difference or the amount of IUs per NAT detectable RNA copy was 4.29 (4.44-5.36) for the ORFa/b target and 2.68 (2.19-3.29) for the E gene target (Supplemental Table 2b ). . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) preprint The 95% LOD (with 95% confidence interval (CI)) in the package insert of the cobas assay was estimated in TCID 50 /mL using culture fluid of the SARS-CoV-2 Strain USA-WA1-2020. For the ORFa/b gene the Roche package insert reported a 95% (CI) of 0.007 (0.005-0.036) TCID 50 /mL, whereas on our working standard we estimated the 95% LOD at 8.3 (4.5-18.6) copies/mL for this PCR gene target. According to this comparison 1 TCID 50 would be equivalent to 1186 (847-6098) PCR detectable RNA copies in our working standard. Table 2 gives the proportion of reactive results and mean measurement values on the inactivated standard dilutions for the two most sensitive NAT assays, i.e. the cobas real time PCR and Aptima TMA assay. Similar data are also available for the LAMP and SAMBA assays (data not shown). Table 3 shows the reactivity of three rapid antigen tests on 1.5 fold serial dilutions of the native and inactivated working standards. In Table 4 we combined the analytical sensitivity data of the four NAT systems and three rapid antigen assays. The cobas assay for ORFa/b was the most sensitive with 50% and 95% LOD of 1. fold less sensitive than cobas PCR with 50% and 95% LODs of 6.6 (4.4-9.9) and 29.7 (18.4-60.1) copies/mL. The LAMP assay was 12.4 (6.7-25) fold less sensitive than the Roche assay with estimated 50% and 95% LODs of 23 (13-42) and 102 (54-251) copies/mL. The SAMBA II assay was evaluated later using the same reference . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this this version posted November 28, 2021. ; https://doi.org/10.1101/2021.11.26.21265946 doi: medRxiv preprint 1 3 panel and in a separate probit analysis the 50% LOD and 95% LOD were 15 and 133 (44-447) copies/mL respectively. The analytical sensitivity of the rapid antigen assays on 1.5-fold dilutions of the native standard dilutions were 28,000 to 56,000 fold lower than the cobas PCR assay when comparing the lowest concentration giving indeterminate antigen reactivity with the 50% LOD in PCR, whereas the LODS with rapid antigen test positive reactivity were 9,000 to 21,000-fold higher than the 95% LOD in the cobas assay (Table 4 ). Interestingly the Chinese Fluorecare assay targeting the spike protein was slightly more sensitive than the Abbott and Roche antigen assays targeting the nucleocapsid protein. In the present study we directly compared the analytical sensitivity of four different NAT assays and three lateral flow devices for rapid SARS CoV-2 antigen detection on dilution series of a pool of swab fluid samples before and after inactivation by beta-propiolactone. The viral load in the native and inactivated material was quantified in NAT detectable RNA copies/mL by limiting dilution analysis and in IU/mL against the WHO International Standard. Moreover our working standards were used to calculate viral load in culture fluid of known infectivity and we estimated that 1 TCID 50 was equivalent to approximately 1000 NAT detectable RNA copies in our working standard. We were surprised that the antigen concentration relative to the RNA concentration was found to be 2.7 (range 2.5-3.0) fold higher in the inactivated standard than in the native standard. We speculate that the beta-propiolactone treatment has destroyed or modified subgenomic RNA fragments from human cells that were present in the pool of swab GLY samples, but had less impact on full length RNA genomes packaged in virions and on the antigenicity of the nucleocapsid or spike protein. For discussion on further details of the standardization in RNA copies or IUs in swab fluid before and after inactivation and the relation to TCID 50 we refer to the Appendix. On 1.5 fold dilutions of the native working standard the rapid antigen tests were found to be 28,000 to 56,000 fold less sensitive than the 50% LOD of 1.8 (1.0-3.3) copies/mL achieved by the most sensitive NAT assay in our comparison study, which was the cobas PCR assay. The Aptima TMA assay was 3.6 fold less sensitive than this PCR assay, whereas the SAMBA and LAMP assays were 8-12 fold less sensitive. . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) preprint Assuming a ramp up viremia model of 1 Log10 per day 14, 15 we estimated that the Fluorecare, Abbott and Roche antigen assays were able to detect early viremia 4-5 days later than the Roche cobas and Hologic Aptima NAT assays, whereas the cobas PCR in MP6 format, SAMBA and LAMP would be approximately 0.5-1 day slower in detecting early viremia than the most sensitive NAT options. More importantly, we estimated that the NAT assays were able to detect early viremia 40-66 hours before the estimated tissue culture infectivity conversion point at approximately 1000 copies/mL in the ramp up phase of viremia, whereas the antigen assays became positive 41-48 hours later. Rapid antigen tests have long been promoted as an effective tool in preventing SARS-CoV-2 transmission since they detected the majority of PCR positive symptomatic individuals and it was assumed that the probability of contagiousness was highest in antigen positive individuals 1,2 . It is possible that the airborne transmissibility of SARS-CoV-2 between humans in close contact starts at a viral load that is lower than the estimated tissue culture infectivity limit of 1000 copies/mL in swab fluid. Taking considerable uncertainty ranges in the exact conversion time points of assays and the start of contagiousness into account we believe that our modeling of analytical sensitivity data helps to understand what the effectiveness is of using different SARS-CoV-2 test options in timely identifying potentially contagious subjects. Our modeling of the time points of assay conversion during early viremia matches with evaluation data of laminar flow antigen detection methods in large clinical studies in the Netherlands 1,2 . One of these studies 2 found a clinical sensitivity of 59% with two widely used rapid antigen tests when asymptomatic individuals were tested 5 days after potential exposure, whereas a higher proportion of 73-84% of the exposed subjects with symptoms were antigen reactive. These proportions are . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. We therefore recommend the use of NAT methods rather than antigen assays for more reliable identification of contagious individuals. In a fast NAT screening setting where one receives the test results not immediately at the testing site but a few hours later by e-mail, one could just as well opt for a laboratory that uses a highly sensitive NAT method and receive the test result within the same day. Finally -if the infection rate in the target population is <4% one could also make use of more cost-effective NAT testing of pooled swab samples as recently demonstrated for MP4 using the . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) preprint The copyright holder for this this version posted November 28, 2021. ; https://doi.org/10.1101/2021.11.26.21265946 doi: medRxiv preprint 1 8 SAMBA assay 18 and shown in the present report for MP6 using the cobas PCR assay. In conclusion, we estimated up to 56,000 fold differences in detection limits between several SARS-CoV-2 detection methods on a pool of swab fluid samples before inactivation. Translating these analytical sensitivity data to assay conversion time points using a ramp up viremia model helped to better understand what the effectiveness is of these testing options in detecting early (asymptomatic) infection in potentially contagious individuals. . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) preprint 75000 9036 according to comparison of 50% and 95% LODs respectively for NAT methods as calculated by probit analysis. For antigen assays detection endpoint titers with indeterminate (±) reactivity were compared with the 50% LOD in cobas PCR, whereas endpoint titers with positive (+) reactivity were compared with the 95% LOD in cobas PCR #MP6: minipools of six samples . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. . CC-BY-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) preprint The copyright holder for this this version posted November 28, 2021. ; https://doi.org/10.1101/2021.11.26.21265946 doi: medRxiv preprint Clinical evaluation of the Roche SD Biosensor rapid antigen test for SARS-CoV-2 in a municipal health service testing site. The Netherlands Diagnostic accuracy of rapid antigen tests in asymptomatic and presymptomatic close contacts of individuals with confirmed SARS-CoV-2 infection: cross sectional study Head to head comparison of SARS-CoV-2 antigen-detecting rapid test with self-collected nasal swab versus professional-collected nasopharyngeal swab Diagnostic accuracy of two commercial SARS-CoV-2 antigen-detecting rapid tests at the point of care in community-based testing centers SARS-CoV-2 rapid antigen test: High sensitivity to detect infectious virus Analytical sensitivity and clinical sensitivity of the three rapid antigen detection kits for detection of SARS-CoV-2 virus Analytical Sensitivity of the Abbott Binax NOW COVID-19 Ag CARD Comparative sensitivity evaluation for 122 CE-marked rapid diagnostic tests for SARS-CoV-2 antigen Inactivation of the Hutchinson strain of hepatitis non-A, non-B virus in intravenous immunoglobulin by β -propiolactone Inactivation of viruses by beta-propiolactone in human cryo poor plasma and IgG concentrates Package insert: cobas® SARS-CoV-2 Qualitative assay for use on the cobas® 6800/8800 Systems Performance evaluation of the SAMBA II SARS-CoV-2 test for point-of-care detection of SARS-CoV-2 Package insert: Aptima® SARS-CoV-2 Assay Viral infection and transmission in large, well-traced outbreak caused by SARS-CoV-2 Delta variant SARS-CoV-2 B.1.617.2 Delta variant replication and immune evasion Nature Virological and serological kinetics of SARS-CoV-2 Delta variant vaccine-breakthrough infections: a multi-center cohort study Outbreak of SARS-CoV-2 infections, including COVID-19 vaccine breakthrough infections, associated with large public gatherings Evaluation of sample pooling using the SAMBA II SARS-CoV-2 Test