key: cord-0227314-9pa14cmt authors: Ghane, Behrooz; Karimian, Alireza; Mostafapour, Samaneh; Gholamiankhak, Faezeh; Shojaerazavi, Seyedjafar; Arabi, Hossein title: Quantitative analysis of image quality in low-dose CT imaging for Covid-19 patients date: 2021-02-16 journal: nan DOI: nan sha: abb66f075ca5955cfbd1ee748b0e46dd203e4c7d doc_id: 227314 cord_uid: 9pa14cmt We set out to simulate four reduced dose-levels (60%-dose, 40%-dose, 20%-dose, and 10%-dose) of standard CT imaging using Beer-Lambert's law across 49 patients infected with COVID-19. Then, three denoising filters, namely Gaussian, Bilateral, and Median, were applied to the different low-dose CT images, the quality of which was assessed prior to and after the application of the various filters via calculation of peak signal-to-noise ratio (PSNR), root mean square error (RMSE), structural similarity index measure (SSIM), and relative CT-value bias, separately for the lung tissue and whole-body. The quantitative evaluation indicated that 10%-dose CT images have inferior quality (with RMSE=322.1-+104.0 HU and bias=11.44-+4.49% in the lung) even after the application of the denoising filters. The bilateral filter exhibited superior performance to suppress the noise and recover the underlying signals in low-dose CT images compared to the other denoising techniques. The bilateral filter led to RMSE and bias of 100.21-+16.47 HU, -0.21-+1.20%, respectively in the lung regions for 20%-dose CT images compared to the Gaussian filter with RMSE=103.46-+15.70 HU and bias=1.02-+1.68%, median filter with RMSE=129.60-+18.09 HU and bias=-6.15-+2.24%, and the nonfiltered 20%-dose CT with RMSE=217.37-+64.66 HU and bias=4.30-+1.85%. In conclusion, the 20%-dose CT imaging followed by the bilateral filtering introduced a reasonable compromise between image quality and patient dose reduction. The World Health Organization (WHO) identified Coronavirus 2 (SARS-CoV-2), which causes a severe acute respiratory syndrome, as a global public health concern. 1 Up to this date (January 30, 2021), more than 100 million cases have been diagnosed with Coronavirus disease. 2 Coronavirus or Covid-19 affects the respiratory system and causes mild to severe respiratory syndromes, including pneumonia. The management, outbreak prevention, and treatment of this disease are currently the most critical global challenges in medicine. 3, 4 Therefore, an accurate and quick diagnosis of this disease is of essential importance. Real-time reverse transcription-polymerase chain reaction (RT-PCR) test is regarded as a standard method for the diagnosis of Covid-19. However, this method has severe limitations such as the preparation time, considerable false positive and false negative rates 5 , and relatively low sensitivity 6 . Preliminary studies have confirmed that computed tomography (CT) imaging is an effective approach for diagnosing Covid-19 disease. 7, 8 CT imaging is an accurate and noninvasive method for detecting abnormalities, including tumors, bone fractures, and cardiovascular diseases. [9] [10] [11] [12] Recently, an extended range of clinical studies has been performed on the potentials of CT imaging for the diagnosis and management of patients with disease. These studies, in fact, continue to fully identify the underlying Covid-19 features in CT images. 8 Although CT imaging is a rapid method for the diagnosis of Covid-19 disease, the absorbed radiation dose by patients, due to the x-ray exposure, is one of its major concerns. The radiation risks of CT scans are seriously high for those required to undergo multiple scans, pregnant women, and pediatrics. 13 A recent study showed that DNA double-strand breaks and chromosome aberrations would increase in patients who undergo standard-dose CT scans, while there was no DNA damage in people who have been examined with low-dose CT scans. 14 Despite the significant advances made in software and hardware technology of CT scanners 15, 16 , it is not yet considered as a low-dose safe imaging modality 17 . Therefore, regarding CT scan's side-effects 18 , the application of low-dose CT protocols would be of crucial importance in routine CT imaging. However, increased image noise is a major limitation in this imaging modality. 19 The diagnosis of Covid-19 (infection) patterns within the lung region would be highly challenging in CT images with increased noise levels, which can skew the correct diagnosis. Reducing noise levels in CT images and enhancing image quality has always been one of the important research topics. 20 Noise reduction in low dose CT imaging can be achieved in three ways 21 : denoising in the sinogram or projection space prior to the image reconstruction 22, 23 , using the dedicated reconstruction algorithms which model the noise within image formation 24, 25 , and denoising in the image domain using post-reconstruction algorithms 19, 26 . The induced-noise in low-dose CT images normally follows a Poisson noise distribution, Gaussian noise distribution, or a combination of the two. 27 The commonly used denoising techniques in medical imaging are Gaussian, median, and bilateral filters which could be employed either in the projection or image space. 28, 29 Implementation of these techniques in the projection space requires access to the raw acquisition data which is not commonly provided by CT scanners. 21 Iterative CT image reconstruction algorithms tend to model the underlying causes of noise within the image reconstruction process to suppress the noise in the resulting images, commonly through the maximum likelihood formulation of the Poisson noise. These techniques, in addition to highcomputation time, also require raw acquisition data. Moreover, they should be dedicatedly developed/optimized for each scanner and acquisition protocol to properly perform noise suppression. 21, 30, 31 The denoising techniques in the image domain, contrary to the other approaches, do not require raw sinogram data and can directly be applied to CT images. In addition to low computation time, these techniques are almost independent of CT scanners, acquisition protocols, and image reconstruction algorithms. Moreover, they could easily be adapted to the new acquisition protocols and, or noise levels through the adjustment of a few hyperparameters. In addition to the above-mentioned approaches, deep learning-based denoising techniques have exhibited promising performance in low-dose CT imaging. 32, 33 These methods require an extensive training dataset to create a robust model to inter-subject image quality and, or anatomical variations. Moreover, these techniques are very sensitive to abnormal and, or odd cases wherein they may lead to gross errors and misdiagnosis. 19, 34, 35 The aim of this work is to investigate low-dose CT imaging for Covid-19 patients, and in particular, how reduced patient dose would affect the quality of the CT images. To this end, the low-dose CT imaging of Covid-19 patients will be simulated for different dose levels up to 10% of the standard dose to measure the increased noise levels and quantitative bias. Thereafter, different post-reconstruction denoising approaches, namely Gaussian, median, and bilateral filters, will be applied to the low-dose CT images to suppress the noise and recover the image quality. The goal is to determine which reduced dose level and post-reconstruction denoising approach would lead to an optimal compromise between image quality and patient radiation dose. the entire group of patients, the lung region was semi-automatically segmented from the standard CT images using Pulmonary Toolkit (PTK) software. 36 The lung segmentation was visually verified and the miss-segmentation errors were manually corrected. In this study, low-dose CT imaging was simulated for reduced doses of 60%, 40%, 20%, and 10% of the standard CT images (reference). To this end, the method proposed by 37 was employed to simulate low-dose CT imaging as described in the following steps. The parallel beam geometry estimates the spatial coordinates of the x-ray beam passing through the subject in the field of view. Figure 1 depicts the concept of parallel beam geometry and its key parameters. A single point in space is defined by a combination of φ and s parameters which are the angle between the x-ray beam and the distance of the x-ray beam from the center, respectively. Since the Ls, φ line is perpendicular to the x-ray beam, it can be described by a point (x, y) on Ls, φ using a real number t ℝ and φ ϵ (0,2π) as follows: Ls, φ (t)=(x(t),y(t)) = s (cosφ, sinφ)+ t (-sinφ, cosφ). (1) Then, function f (Ls, φ) can be defined on ℝ 2 as: In the above equation, R is the Radon transform which is applied to the f function. Based on Beer-Lambert's law, the result of applying the Radon transforms in Eq. (2) would be I0 is the intensity at the x-ray source and I1 is the intensity (signal) recorded in the detector. HUs= (PixelValue × Slope) + Intercept 2. Considering the tube voltage in standard CT imaging, we assume µair=0 and the linear attenuation coefficients were computed using Eq. (5) 3. The projection data (ρs) is obtained from the application of the Radon transform to the reference CT images (converted to the linear attenuation coefficients). 4. The normal-dose transmission data (Tnd) is calculated using Eq. (6) . 5. The low-dose transmission data (Tld) is generated by adding Poisson noise to the original data where I0 is the incident flux for simulating low-dose scans. 6. The low-dose projection data (ρld ) is calculated in the sinogram domain using Eq. . 7. Finally, the inverse of Eq. (5) is applied to convert the attenuation coefficients to the CT values and the final low-dose image is reconstructed using the inverse Radon transform and filter back-projection algorithm. Based on Beer-Lambert's law, the intensity index I0 indicates the number of photons per detector bin in the air scan, which is chosen 6000 for 60% low-dose simulation. According to Ref. 37 , I0 was assumed 10 4 in the full-dose imaging, thus I0 =6000 was considered as 60% low-dose, I0 =4000 as 40%, I0 =2000 as 20%, and I0 =1000 as 10% low-dose CT imaging. Gaussian filter is a simple and intuitive approach of noise suppression that replaces the value of each pixel with the weighted average of surrounding ones. This approach may introduce noticeable signal loss and blur in the CT images. 29 The Gaussian filter is defined as where, x and y are the vertical and horizontal distances from the target pixel (in twodimensional implementation) and σ is the standard deviation of the Gaussian distribution. The kernel obtained from this equation is convolved with the noisy CT images to diminish the noise levels. The standard deviation of the Gaussian kernel is the key parameter that defines the levels of smoothness in the resulting images. The procedure to find optimal standard deviation values for the different noise levels (low-dose CT images) is described in the following section. The median filter, as a non-linear denoising approach, is commonly used for noise suppression in natural as well as medical images. The median filter is functionally similar to the moving average filter, but it calculates the median value of the pixels/voxels rather than the mean value. Median filters might exhibit better performance regarding the edge preservation in the filtered images than the Gaussian filter. The median filter is formulated as where x × y defines the size of the filter window for calculation of the median value. The filter window size should be optimized independently for the different low-dose CT images. Bilateral filter, regarded as an edge-preserving smoothing approach, tends to suppress noise while maintaining the prominent edges/patterns of the image through penalizing the smoothing function over the edges. 38 The Bilateral filter, employed to reduce the noise levels in the different low-dose CT images, is mathematically formulated as Here, (i,j) and (k,l) indicate the location of the target voxel to be filtered as well as its neighboring voxels. w(i,j,k,l) returns the weighting factors for each of the neighboring voxels (k,l) based on spatial closeness (first term in Eq. 11) and intensity difference (second term in Eq. 11) with respect to the target voxel (i,j). I(i,j) and I(k,l) denote the intensity of (i,j) and (k,l) voxels, respectively, and σd and σr are the smoothing parameters. The calculated weights from Eq. 11 are used in Eq.12 to estimate the noise-free voxel values (ID) in the filtered images. The performance of the Gaussian, Median, and Bilateral denoising techniques was evaluated on the CT images with different dose levels using the following four image-quality metrics. Peak Signal-to-Noise Ratio (PSNR) presents the ratio between the maximum intensity of the noisy or reference image, and the mean square error (MSE), which is usually expressed in the unit of logarithmic decibel as follows: Here, x is the full-dose CT image (reference), y is the low-dose (or filtered) CT image and MSE (y, x) stands for the mean squared error between the reference and low-dose (filtered) CT images as follows: Here, m×n (row×column) is the number of voxels in the CT images and i and j stand for the voxel indices. Structural Similarity Index Measure (SSIM), which reflects the perceived quality of the images, was used to compare the overall structural similarity between low-dose (or filtered) and the reference CT images. Here, x is the full-dose CT image (reference), y indicates the low-dose (or filtered) CT image, and μx and μy are the mean intensity values of the x and y images, respectively. Similarly, and are the variance of the x and y images, respectively, and stands for the covariance of x and y. Constants c1 and c2 are utilized to avoid division by zero (or very small numbers). Root Mean Square Error (RMSE) measures the difference between CT values in the noisy and reference CT images. MSE was defined in Eq. 14. Relative Bias (BAIS) indicates the percentage of error in the low-dose or filtered CT images with respect to the reference CT images. Here, x and y denote the CT values in the full-dose and low-dose CT images, respectively. One of the major challenges in the application of the post-reconstruction denoising techniques is finding the optimal values for the hyper-parameters to achieve the best image quality. Optimal sets of the hyper-parameters would lead to maximal PSNR and SSIM indices as well as minimum RMSE and BIAS. To optimize the hyper-parameters of the Gaussian, Median, and the bilateral filters, a reasonably wide range of values were set for these parameters, and the ones leading to the maximum PSNR and minimum RMSE were selected for the final implementation of these filters. It should be noted that this procedure was repeated independently for each level of the low-dose CT images. Figure 2 shows a representative example of this optimization procedure for the Gaussian hyper-parameter at the low-dose level of 60%. The BIAS, PSNR, and RMSE metrics were calculated for the whole-body, lung, and whole-body without lung, separately. Fig. 2 The optimal value for the Gaussian hyper-parameter (sigma) at the low-dose level of 60%. Table 1 . Table 2 presents the quantitative metrics calculated after the application of the Gaussian filter to the different low-dose CT images. The SSIM and PSNR indices increased after the application of the Gaussian filters, which indicates the overall improved quality of the resulting images. Figure 4 presents the low-dose CT images after the application of the Gaussian filter (for the same views presented in Fig. 2 ). Visual inspection revealed effective noise reduction; however, some streak-like artifacts are seen in the 10% low-dose image. filtering led to a maximum bias of less than 5% compared to the Gaussian and median filters with a bias of more than 7% and -9%, respectively. Figure 6 illustrates the effective noise reduction in the Bilateral filtered images, particularly in the 20% low-dose CT image wherein the pattern of infection can be clearly seen in the lung region with minimal signal loss and, or artifact. To compare the performance of the different denoising techniques for the different low-dose levels, boxplots of the RMSE and BIAS parameters are presented in Fig. 7 . It is clearly seen that the Bilateral filter resulted in noticeably lower BIAS and RMSE in the four low-dose CT images within the lung region. Although the Bilateral filter enabled remarkable noise reduction, the BIAS in the 10% low-dose CT images within the lung region is yet significantly high, which might skew the correct diagnosis of the infection in the lung region. as well as the resolution of the CT images. Therefore, the optimization of these hyperparameters should be repeated when the input CT images are from different scanners or acquired through the use of different acquisition protocols. Overall, considering Fig.7 , the Bilateral filter led to a noticeable smaller quantitative bias and RMSE compared to the other denoising techniques, regardless of the noise levels in the input images. The quantitative bias observed in the Bilateral filtered images was less than 5% even in 10%-dose images, which is a clinically tolerable error. However, due to the presence of the streak-like artifacts in the lung, which might skew the correct classification of the Covid-19 infection, 10%-dose imaging is not recommended even after bilateral filtering. On the other hand, 20%-dose CT imaging would offer an acceptable compromise between dose reduction and image quality wherein the quantitative bias in the lung region did not exceed 5% and -1% before and after the application of the bilateral filter. Regarding the quantitative errors in the lung region and the image artifacts, which might be introduced in the lung region due to the low-dose imaging and, or post-reconstruction denoising, no more than 40% low-dose imaging is recommended if no denoising approach is supposed to be applied. Nevertheless, the application of the bilateral filter to the 20% low-dose CT images led to a similar image quality of the 60% low-dose images without applying any denoising filters. Dose reduction beyond 20% is not recommended for Covid-19 patients as the quantitative bias as well as image artifact would impair the correct diagnosis or grading of Covid-19 disease. This work set out to investigate low-dose CT imaging for Covid-19 patients, and in particular, to what extent acquisition dose could be reduced with no significant quantitative bias or image artifact. For this purpose, different low-dose CT imaging, namely 60%, 40%, 20%, and 10% of the standard-dose, were simulated and quantitatively assessed before and after applying the common denoising techniques. Dose reduction beyond 20% introduced considerable quantitative bias as well as image artifact, which may skew the correct diagnosis or grading of Covid-19 infection in the lung region; thus dose reduction beyond 20% is not recommended for Covid-19 patients. The quantitative evaluation of the low-dose CT images indicated the superior performance of the bilateral filter, particularly in the lung region compared to the Gaussian and median filters, which resulted in less than 5% quantitative bias in the lung for the Characteristics of and Important Lessons From the Coronavirus Disease 2019 (COVID-19) Outbreak in China Summary of a Report of 72 314 Cases From the Chinese Center for Disease Control and Prevention Coronavirus disease (COVID-19) pandemic Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) and coronavirus disease-2019 (COVID-19): The epidemic and the challenges Chest-CT findings of COVID-19 in patients with pre-existing malignancies; a pictorial review Laboratory Diagnosis of COVID-19: Current Issues and Challenges Sensitivity of Chest CT for COVID-19: Comparison to RT-PCR Coronavirus Disease 2019 (COVID-19) CT Findings: A Systematic Review and Metaanalysis Diagnostic Performance of CT and Reverse Transcriptase Polymerase Chain Reaction for Coronavirus Disease 2019: A Meta-Analysis Deep Learning for Low-Dose CT Denoising Using Perceptual Loss and Edge Detection Layer Monte Carlo-based assessment of the trade-off between spatial resolution, field-of-view and scattered radiation in the variable resolution X-ray CT scanner Novel detector design for reducing intercell x-ray cross-talk in the variable resolution xray CT scanner: a Monte Carlo study Design and construction of a variable resolution cone-beam small animal mini-CT prototype for in vivo studies Radiation in medicine: Origins, risks and aspirations Biological Effects of Low-Dose Chest CT on Chromosomal DNA The effect of focal spot size on the spatial resolution of variable resolution X-ray CT scanner Feasibility study of a new approach for reducing of partial volume averaging artifact in CT scanner CT is still not a low-dose imaging modality Recent Advances in CT Image Reconstruction Ultra-low-dose chest CT imaging of COVID-19 patients using a deep residual neural network Improvement of image quality in PET using post-reconstruction hybrid spatialfrequency domain filtering Methods for Clinical Evaluation of Noise Reduction Techniques in Abdominopelvic CT Projection Space Implementation of Deep Learning-Guided Low-Dose Brain PET Imaging Improves Performance over Implementation in Image Space An experimental study on the noise properties of xray CT sinogram data in radon space Regularization strategies in statistical image reconstruction of low-dose x-ray CT: A review Resolution-recovery-embedded image reconstruction for a high-resolution animal SPECT system Spatially guided nonlocal mean approach for denoising of PET images Non-local mean denoising using multiple PET reconstructions Projection space denoising with bilateral filtering and CT noise modeling for dose reduction in CT Image filtering for improved dose resolution in CT polymer gel dosimetry Abdominal CT with model-based iterative reconstruction (mbir): initial results of a prospective trial comparing ultralow-dose with standard-dose imaging Validation of dualsource single-tube reconstruction as a method to obtain half-dose images to evaluate radiation dose and noise reduction: phantom and human assessment using CT colonography and sinogramaffirmed iterative reconstruction (safire) A Review on CT and X-Ray Images Denoising Methods Applications of artificial intelligence and deep learning in molecular imaging and radiotherapy Comparative study of algorithms for synthetic CT generation from MRI: Consequences for MRI-guided radiation planning in the pelvic region Deep learning-guided joint attenuation and scatter correction in multitracer neuroimaging studies Evaluation of open-source software for the lung segmentation Dose Reduction Potential with Photon Counting Computed Tomography Performance of bilateral filtering on Gaussian noise