key: cord-0198981-w625pegv authors: Das, Abhimanyu; Robinson, Matthew L.; Warsinger, D. M. title: Converting Hemodialysis (HD) membranes to Extracorporeal Membrane Oxygenation (ECMO) machines date: 2020-05-06 journal: nan DOI: nan sha: f6d0a94dc699c1e65526d6b62b5e7a4f237865a6 doc_id: 198981 cord_uid: w625pegv Crises like COVID-19 can create a massive unmet demand for rare blood oxygenation membrane machines for impaired lungs: Extracorporeal Membrane Oxygenation (ECMO) machines. Meanwhile, Hemodialysis (HD) machines, which use membranes to supplement failing kidneys, are extremely common and widespread, with orders of magnitude more available than for ECMO machines. This short study examines whether the membranes for HD can be modified for use in blood oxygenation (ECMO). To do so, it considers mass transfer at the micro-scale level, and calculations for the macro-scale level such as blood pumping rates and O2 supply pressure. Overall, while this conversion may technically be possible, poor gas transport likely requires multiple HD membranes for one patient. Background on HD and ECMO HD and ECMO machines both operate by removing large volumes of blood from the body and passing them through membranes that rely on diffusion to transfer desirable or undesirable solutes. HD machines focus on adding and removing compounds to assist with kidney failure, including removing salts and small organic compounds like urea. Meanwhile, ECMO uses membranes to replace O2 and help remove CO2. Custom HD machines made for blood oxygenation [2] have been successfully demonstrated in the past; the question is whether existing HD machines can be converted readily. The blood flow paths of the two machines are similar, although flow rates and pipe diameters are smaller in HD: this may require multiple machines. Catheter insertion points for ECMO are chosen to provide O2 where it is needed; the same placement would be necessary for a HD to ECMO conversion. The dialysate side of the HD membrane presents numerous problems that limit the use of a liquid dialysate. While it is possible to use an oxygenated dialysate for ECMO, it would need a separate oxygenation chamber along with high pressure O2 supply. On the contrary, directly connecting a O2 supply to the dialyzer ports allows for a far simpler arrangement that requires a lower O2 supply pressure. Therefore, it was assumed that a pressurized O2 gas would be far more viable, although it presents its own concerns like bypassing the machine warning and shut-down systems triggered by no flow through the dialysate pump. Such pressurized O2 tanks are common in hospitals. Simply described, such a conversion would resemble the invasive setup of venovenous ECMO (VV-ECMO), with blood connected to HD machine(s) that house the membranes, pump and controls, and the dialysate side connected to a high pressure O2 tank. The number of HD machines required to be connected in parallel for the conversion depends on the maximum flow rate through the blood pump. With a flow rate of 1.63 l min -1 through the blood pump, an array of at least three HD machines is needed for the conversion. The number of membrane modules required depends on the blood flow rate through them. While the blood circuit is the same as in a conventional dialysis machine, the dialyzer circuit is modified to operate with O 2 gas from a compressed cylinder or a medical gas supply line. Gas exchange occurs inside the dialyzer as deoxygenated blood gets oxygenated while rejecting CO 2 to the gas stream due to partial pressure differences of O 2 and CO 2 across a hollow fiber. Modified human body graphic from http://www.learnecmo.com/vv-ecmo There are numerous concerns and risks to address for attempting to modify HD machines as ECMO machines.  Reliability: putting a system together with multiple machines and membranes, calculated here as 3 HD machines and 12 membranes modules  Each patient takes up at least 3 dialysis machines  High O2 supply pressure may lead to bubble induction in blood  Pumps in parallel may lead to flow instabilities  Peristaltic blood pumps in HD machines may have a maximum flow rate less than 1.6 l min -1  Modifications to dialysis machines may be difficult to make  Continuous renal replacement therapy (CRRT) systems might be needed as these systems will be used for long periods of time as compared to typical dialysis machines that are used for a few hours  Calculations have been carried out using values from multiple sources and topics; actual membrane properties of commercially available membranes have not been used  Mass transfer resistance on blood side has been neglected; actual membrane gas permeability may be lower than calculated NOMENCLATURE Maximum flow rate through HD blood pump ml min -1 , Rated blood flow rate through a membrane module ml min -1 Number of HD machines required in parallel [-] , Actual flow rate through HD blood pump ml min -1 Number of membrane modules required in parallel [-] , Actual blood flow rate through a membrane module ml min -1 Rate of blood oxygenation in a membrane module ml min -1 Partial pressure of O 2 on gas side kPa Volume fraction of O 2 in gas supply [-] Total pressure of gas supply kPa Viscosity of whole blood Pa s ∆ Pressure drop along a hollow fiber kPa The approach analyzes blood flow rates for O2 supply and performs simple transport calculations to examine whether sufficient gas exchange can be provided. The concentrations of O2 present are described in partial pressures. Representative values of HD membrane properties were obtained. A large number of commercially available synthetic HD membranes today are made of polysulfone [3] which is inherently hydrophobic. Hydrophilicity is imparted in the membrane material by adding polyvinylpyrrolidone (PVP) for reasons stated in [4] . Each membrane module consists of tens of thousands of hollow fibers through which blood flows while the dialysate flows over the fibers [4] . The two flows are usually counter-current, with radial concentration gradients driving mass transfer of various necessary and waste solutes. This arrangement differs from ECMO in which blood flows over the hollow fibers which carry oxygen [5] . We adhere to the former arrangement with O2 flowing on the dialysate side to keep the number of necessary modifications at a minimum. The membrane dimensions used in the calculations are listed in Table 1 . For properties that have been reported with a range, average values were used. The value for membrane area was chosen from the largest available membrane from Fresenius Medical Care to assess a system with least number of membrane modules. The external radius of the hollow fiber calculated as while the number of fibers in a membrane module were calculated using (2) The transport of gases between the blood and the oxygen streams, across the membrane, is essentially a diffusion problem driven by concentration gradients. Since O2 and CO2 are carried by the blood as species bound to hemoglobin, we need to the convert the concentration gradients to partial pressure gradients in order to be able to use the oxygen dissociation curve (ODC). The ODC is a non-linear relationship between the volume fraction of O2 in blood and its partial pressure. [8] (Volume fraction of O2 in systemic arterial blood) 0.208 [8] In order to solve for the rate of diffusion of O2, the diffusion coefficient or diffusivity of O2 in the membrane material needs to be calculated. Given the partly hydrophilic nature of the membrane material, we assume that the pores in the hollow fibers are mostly filled with plasma. The diffusivity of O2 in pore confined plasma, however, would be lower than its bulk value, , . Given the porosity and the tortuousness of the transport path in a polymer material, an effective diffusivity can be calculated as , [9] (3) The effective diffusivity can then be multiplied with the solubility of O2 in plasma to obtain Krogh's diffusivity given as [10] (4) This along with the partial pressures of O2 allows us to use the same functional forms of the diffusion equations to calculate O2 transport across the membrane. Reported and calculated values of the diffusion parameters are listed in Table 3 . , at 37℃ 2.18e-5 cm 2 s -1 [11] 0.3 [12] 3.33 [13] 1.98e-6 cm 2 s -1 0.003 ml O 2 (100 ml plasma) -1 m Hg -1 [10] 5.89e-11 cm 2 s -1 ml O 2 ml plasma -1 mm Hg -1 The transport calculations must consider the micro-scale, describing diffusion through membranes into the blood, and the macro-scale, describing changing O2 levels with length. Relative diffusion in the different channels must be compared to neglect those that are an order of magnitude smaller. On the micro-scale, O2 permeation can be calculated by assuming a negligible gas side resistance and ignoring the blood side resistance owing to a high blood flow rate [13] . With mass transfer being dominated by the membrane resistance, the overall mass transfer coefficient for a single hollow fiber can be written as On the macro-scale, the variation of gas concentration with length along the membranes resembles heat transfer in a shell and tube heat exchanger with phase change on the shell side, so a logarithmic mean pressure difference is calculated as (6) The partial pressure of O2 on the gas side and is assumed to be constant along the length of a hollow fiber. This appears to be a reasonable assumption provided the rate of O2 transport across the membrane is small as compared to the bulk O2 flow rate on the gas side. The net rate of blood oxygenation in one membrane module with fibers can then be written as (7) The blood flow rate from the body necessary to satisfy its oxygen demand can be calculated as , The flow rates listed in Table 4 can be used to calculate the required O2 supply pressure and the number of HD machines and membranes modules. (Rated blood flow rate through a membrane module) 400 ml min -1 [3] The number of HD blood pumps needed is obtained using , , , The flow rate through each HD blood pump is then Similarly, the total number of membranes required in parallel , , The above expression for ensures that the blood flow rate through each membrane module is higher than the rated value and that each dialysis machine handles the same number of membranes allowing for identical arrangement. The actual blood flow rate through each membrane module is then The required rate of oxygenation in each bundle is then calculated as , The required partial pressure of O2 on gas side is obtained by equating with ; the calculated values have been listed in Table 5 along with other parameters. The value obtained is less than the usual maximum supply pressure of O2 in hospitals [14] . The calculated value of can be achieved by either using pure O2 at the same pressure or a mixture with lower volume fraction and higher total pressure according to (14) Assuming blood to be a Newtonian fluid [4] , [15] , the pressure drop along the hollow fiber is calculated as In equation above, the viscosity of whole blood used instead of that of plasma to estimate maximum pressure drop. The limiting factor in the above arrangement is the maximum flow rate through the HD blood pump, which in turn dictates the number of dialysis machines required to be put in parallel (3 here). If the maximum blood pump flow rate is lower than 1.63 l min -1 , we will need another dialysis machine and so on. Another major concern would be to reprogram a hard-coded HD console that might be not function as desired in the absence of the dialysate loop. Overall, there a significant challenges in gas exchange in this conversion, but these likely can be overcome with engineering to use multiple membranes per patient, modified flow and catheters, and the inclusion of high pressure O2. EXTRACORPOREAL MEMBRANE OXYGENATION ( ECMO ) ACCESS IN THE 30 LARGEST U . S . METROS Dialysate with High Partial Pressure of Oxygen Enhances Oxygenation in Blood during Simulated Circulation of Hemodialysis High flux dialyzers Haemodialysis membranes Evolution of Gas Permeable Membranes for Extracorporeal Membrane Oxygenation Solute transport in hemodialysis: advances and limitations of current membrane technology Medical phisiology: Principles for clinical medicine Relating oxygen partial pressure, saturation and content: the haemoglobin-oxygen dissociation curve Restricted diffusion in liquids within fine pores Regulation of tissue oxygenation Diffusion of Oxygen in Plasma and Blood BT -Oxygen Transport to Tissue -II Solvent extraction with immobilized interfaces in a microporous hydrophobic membrane Use of polysulfone hollow fibers for bubbleless membrane oxygenation/deoxygenation of water The anaesthesia gas supply system We would like to thank Maher Amadi, William Clark, Samir C Gautam, and Steve Sozio for their feedback on this work.